Professional Documents
Culture Documents
Cardiac Pacemakers:
➢ Pacemaker is an electrical pulse generator for starting and/or maintaining the normal heart beat.
The output of Pacemaker is applied either externally to the chest or internally to the heart muscle.
➢ In case of cardiac standstill the use of pacemaker is temporary just long enough to start a normal
heart rate. In long term pacing pacemaker is surgically implanted in the body and its electrodes are
in direct contact with heart.
➢ In cardiac diseases where the ventricular rate is too low it can be increased to normal rate by using
pacemaker.
Energy requirements to excite heart muscle:
➢ The heart muscle can be stimulated with an electric shock. The minimum energy required to excite
heart muscle is 10µJ.
➢ For better stimulation and safety purposes 100µJ pulse energy is applied on heart muscle. During
ventricular fibrillation heart muscle contracts so rapidly and irregularly. The pulse to space ratio
1:10000
➢ The concentration of sodium ions inside the cell becomes much lower than outside. Since the sodium
ions are positive, the outside of the cell is more positive than inside.
➢ To balance the electric charge, potassium ions which are positive enters the cell causing a higher
concentration of potassium on inside than on outside.
(ii) External stimulation: It is employed to restart the normal rate of heart in case of cardiac stand still. The
paddle shaped electrodes are applied on the surface of chest current in the range of 20-150mA
Based on the placement of pacemaker there are two types
(i)External pacemaker
(ii)Implanted (Internal pacemaker)
It does not require the open chest surgery It requires an open chest minor surgery
During placement swelling and pain do not arise During placement swelling and pain arise
Mostly there are used for temporary heart damages. Mostly there are used for permanent heart damages
➢ A special circuit called voltage monitor senses the cell depletion and signals in rate slow down
circuit energy compensation circuit.
➢ The rate slowdown circuit shuts off some of the current to timing network to slowdown
8±3beats/min during cell depletion.
DEFIBRILLATOR:
➢ A Defibrillator is an electronic device that creates a sustained myocardial depolarization of a patient’s
heart in order to stop ventricular fibrillation or atrial fibrillation.
➢ Ventricular fibrillation is a serious cardiac emergency resulting from asynchronous contraction of
heart muscles. This results from electric shock or abnormalities of body chemistry.
➢ Hence it causes a steep fall of cardiac output and can lead to death if adequate steps are not taken
promptly.
➢ Ventricular fibrillation can be converted to a more efficient rhythm by applying a high voltage shock
to the heart. This voltage causes all muscle fibers to contract simultaneously. The instrument for
administering the electric shock is called defibrillator.
➢ The sudden cardiac arrest can be treated using a defibrillator and 80% of patient’s will be cured if
the treatment is given within one minute of attack.
➢ An atrial fibrillation causes reduced cardiac output but is usually not fatal. It happens for the young
people who are always smoking and can even be cured by drug therapy.
➢ Types of defibrillators: There are two types of defibrillators based on electrodes placement (i)Internal
defibrillator (ii) external defibrillator
CIT 5 DEPT OF ECE
OEC 754 MEDICAL ELECTRONICS
(i)Internal Defibrillator:
➢ It is used when the chest is opened. It uses large spoon shaped electrodes with insulated handle.
Since the electrodes are direct contact with heart the contact impedance is about 50Ω. The current
passes through the heart is of 1 to 20A.
(ii)External defibrillator:
➢ External defibrillator is used on the chest using paddle shaped electrodes. The bottom of the
electrode consists of a copper disc and is attached with highly insulated handle. The required
voltages are from 1000 to 6000V.
➢ When the electrodes are placed on the chest after the application of electrode gel the contact
impedance on the chest is about 100Ω.
➢ The D.C defibrillator is designed to deliver 50 to 400J of energy through thorax. The duration of
shock is about 1 to 5ms. The current flowing through the chest is about 10 to 60A.
➢ Depending upon the nature of voltage applied the defibrillators can be divided in to six groups.
1. A.C defibrillator
2. D.C defibrillator
3. Synchronized D.C defibrillator
4. Square pulse defibrillator
5. Double square pulse defibrillator
6. Biphasic D.C defibrillator
1. A.C defibrillator:
➢ It is the earliest and simplest type of defibrillator. It has appropriate voltages for both internal and
external defibrillation. It consists of a step-up transformer with various tappings on secondary side.
➢ An electronic timer circuit is connected to the primary of the transformer. The timer connects the
output of the electrodes for a preset time.
➢ The timing device may be simple capacitor and resistor network which is triggered by a push button
switch. The duration of shock may vary from 0.1-1sec depending upon the voltage to be applied.
➢ For safety the secondary coil of transformer should be isolated from earth so that there is any shock
risk to anyone.
➢ For external defibrillation the voltages are in the range from 250 to 750V. For internal defibrillation
the voltage is from 60 to 250 V.
➢ External defibrillation requires large currents for the simultaneous contraction of heart muscle fiber.
CIT 6 DEPT OF ECE
OEC 754 MEDICAL ELECTRONICS
➢ This current also results in occasional burning of skin under the electrodes. Further it produces
atrium fibrillation while arresting ventricular fibrillation.
2. D.C defibrillator:
➢ D.C defibrillator would not produce undesirable side effects and at the same time it produces normal
heart beat effectively.
➢ Ventricular fibrillation is terminated by passing a high energy shock through discharging a capacitor
to exposed heart or chest of patient.
➢ A variable auto transformer T1 forms the primary of a high voltage transformer T2. The output
voltage of transformer T2 is rectified by a diode rectifier and is connected to a vacuum type-high
voltage change over switch.
➢ In position ‗A‗ the switch is connected to one end of an oil filled capacitor. In this position the
capacitor charges to a voltage set by the positioning of auto transformer.
➢ During the delivery of shock to patient a push button switch mounted on handle of electrodes
operated. The high voltage switch changes to position ‗B‗ and the capacitor is discharged across the
heart through electrodes.
➢ An inductor ‗L‗ is placed in one of the electrodes leads so that the discharge from the capacitor is
slowed down by the induced counter voltage.
➢ The shape of waveform that appears across the electrodes will depend upon the value of capacitor
and inductor and it’s the amplitude depends upon the discharge resistance.
➢ The success of defibrillation depends upon the energy stored in the capacitor and not with the voltage
used. For internal defibrillation 100J of energy is required whereas for external defibrillation 400J
are required. The discharging duration is from 5ms to 10ms.
Fig(b): Dual peak d.c defibrillator Fig(c): output of Dual peak d.c defibrillator
Truncated defibrillator:
➢ In this type the capacitor discharge is adjusted so that the effective defibrillation is obtained at the
desirable low voltage level. The voltage level of the wave is almost constant but its duration is
extended to obtain the required energy.
➢ Fig. shows the modern d.c defibrillator circuit consisting of defibrillator electrocardioscope and
pacemaker.
➢ The pacemaker is used in case of emergency as a temporary pacing. It includes
a) Diagnostic circuitry used to assess the fibrillation before delivering the defibrillation pulse.
b) Synchronizer circuitry used to deliver the defibrillation pulse at the correct time so as to
eliminate the ventricular or atrial fibrillation without inducing them.
Working:
1. The electrocardiogram is obtained by means of an ECG unit connected to the patient who is going
to receive defibrillation pulse.
2. The switch is placed in the defibrillator mode if ventricular fibrillation is suspected.
Fig: Equivalent circuit of square pulse defibrillator and its output waveform
Advantages:
• Using double square pulse defibrillator efficient and quick recovery of heart to beat in normal
manner without side effect like burning of myocardium or inducement of atrial or ventricular
fibrillation.
• The double square pulse with required pulse space ratio can be produced with the use of digital
circuits.
Dialysis:
➢ Both acute and chronic renal failures can be treated successfully by dialysis which is a process by
which the waste products in blood are removed and restoration of normal pH value of the blood is
obtained by an artificial kidney machine.
➢ In dialysis, three physical processes called diffusion, osmosis and ultrafiltration are used to remove
the waste products. There are two types of procedures for doing dialysis, they are extracorporeal
and intracorporeal dialysis.
Extracorporeal Dialysis (Hemodialysis) Intracorporeal Dialysis (Peritoneal
dialysis)
Blood is purified by an artificial kidney machine
The peritoneal cavity in our body is used as a
called hemodialyzer, in which the blood is taken out
semipermeable membrane and by passing
from the body and waste products diffuse through a
the dialysate into it, waste products are
semipermeable membrane which is continuously
removed from the blood by diffusion
rinsed by a dialyzing solution or dialysate
More effective to separate the waste products Less effective
Technically complex and risk one because the blood
Simple and risk free
is taken out from the body
Dialysing time is about 3 to 6 hours Dialysing time is about 9 to 12 hours
➢ The below figure shows the principle of extracorporeal dialysis using hemodialyzer.
Fig: Hemodialyser
Working:
1. For short term use, a double lumen catheter is inserted into the femoral vein and for long term use,
an arterio-venous shunt which is a permanent connection between an artery and a vein and
inserted below the skin in the hand by a minor operation, are used to take the blood from the artery
to the dialyzing unit. There should be a perfect protection against bacterial infection. By this way
the arterio-venous shunt can be used upto two years.
2. The arterio-venous shunt is opened and connected to the dialyser. Using a blood pump, the blood
is pumped into a number of planar sheets of cellophane which are pressed together in a frame.
Blood flows in alternate spaces and the dialysate flows through the spaces is very small, then the
arterial pressure is enough to maintain the flow in the dialyzing unit where the blood pump is not
necessary.
3. The dialysate is an electrolyte. Through the cellophane sheets, urea, creatinine, uric acid and
phosphates are diffused from blood to dialysate.
4. There is a blood leak detector to detect rupture of a membrane. Further there are pressure
monitoring meters at the input and output. A thermostatic control is provided to maintain the blood
at 370C.
5. There are different forms of semipermeable membrane. It may be in the form of pressed planar
sheets or spiral tube in the form of a coil or a bundle of fibers.
Peritoneal Dialysis:
➢ The membrane in the peritoneal cavity in the abdomen is used as a semipermeable membrane. A
catheter is inserted in the abdomen through a puncture just below the navel. A sterile dialysate about
1.5 to 2 litres is allowed to flow into the peritoneal cavity
➢ The diffusion takes place in 10 to 30 minutes and the dialysate is then removed from the cavity. This
procedure is repeated 20 to 30 times to remove all the waste products from the blood.
➢ The above procedure is done in an automatic manner using electronic control circuitry. First the
dialysate is pumped into the abdominal cavity through the volume recording pump. The dialysate is
kept at 370C by thermostatic control, here thermistor is used as a sensing device.
➢ When a dialysate volume is about 2 litres, a timing circuit will first deliver a signal to stop the
dialysate flow into the abdomen. Next the timing circuit allows the diffusion time upto 30 minutes.
VENTILATORS:
➢ As a part of intensive care, the patients often require assistance with breathing. When artificial
ventilation is required for a long time, a ventilator is used to provide oxygen enriched, medicated air
to a patient at a controlled temperature.
➢ Ventilators can operate in different modes: controlled breathing where breathing is initiated by a
timing mechanism and Assisted breathing or patient-initiated breathing.
➢ Controlled breathing is an automatically timed breathing which is usually provided for patients who
cannot breathe on their own. It provides inspirations and expirations at fixed rates except during the
rest period for the patient.
➢ Assisted breathing the patients own spontaneous attempt to breathe in, causes the ventilator to cycle
on during inspiration. Thus, it is used for the patient who has difficult breathing due to high air way
resistance.
➢ There are servo-controlled ventilators which can switch automatically to any mode depending upon
the condition of the patient. This type of mode is called assist control mode. By this mode, the patient
controls his own breathing as long as he can, but if he should fail to do so, the control mode is able to
take over for him.
➢ The ventilator treatment gives the following
a) Adequate ventilation by which enough oxygen is supplied and the right amount of carbon
dioxide is eliminated.
b) Elimination of respiratory work
c) Increased intrathoracic pressure which prevents at electasis that is collapse of portions of
the lung and counteracts edema of lung.
1. Every ventilator operates cyclically. During inspiration air or some other gaseous mixture is pumped
into the lungs. During expiration the pressure ceases. This cycle is regulated by a mechanical,
pneumatic or electronic circuit. The regulation is obtained by pressure limited, volume limited and
servo-controlled systems.
a) Pressure limited ventilators are based on the principle that the inspiration is terminated
when the gaseous mixture pumped into the patients lungs reaches a pre set pressure. It is
➢ The basic components of an NMR imaging system are shown in Fig. These are:
• A magnet, which provides a strong uniform, steady, magnet field B0;
• An RF transmitter, which delivers radio-frequency magnetic field to the sample;
CIT 15 DEPT OF ECE
OEC 754 MEDICAL ELECTRONICS
• A gradient system, which produces time-varying magnetic fields of controlled spatial nonuniformity;
• A detection system, which yields the output signal; and
• An imager system, including the computer, which reconstructs and displays the images.
➢ There is a magnet which provides a strong uniform steady magnetic field B 0. Now a days
superconducting magnet are used in the MRI systems. The superconducting magnetic coils are cooled
to liquid helium temperature and can produce very high magnetic fields.
➢ Hence the SNR of the received signals and image quality are better than the conventional magnets used
in the MRI systems. Different gradients coil systems produce a time varying, controlled spatial
nonuniform magnetic fields in different directions.
➢ The patient is kept in this gradient field space. There are also transmitter and receiver RF coils
surrounding the site on which the image is to be constructed. There is a superposition of a linear
magnetic field gradient on to the uniform magnetic field applied to the patient.
➢ When this superposition takes place, the resonance frequencies of the processing nuclei will depend
primarily on the positions along the direction of the magnetic field gradient. This produces a one-
dimensional projection of the structure of the three-dimensional object.
➢ By taking a series of these projections at different gradient orientations using X, Y and Z gradient coils a
two- or three-dimensional image can be obtained. The slice of the image depends upon the gradient
magnetic field.
➢ The gradient magnetic field is controlled by computer and that field can be positioned in three-time
invariant planes (X, Y and Z). The transmitter provides the RF signal pulses. The received nuclear
magnetic resonance signal is picked up by the receiver coil and is fed into the receiver for signal
processing.
➢ By two-dimensional Fourier Transformation, the image is constructed by the computer and is displayed
on the television screen.
➢ Detection System: The function of the detection system (receiver) is to detect the nuclear
magnetization and generate an output signal for processing by the computer. A block diagram of a
typical receiver is shown in Fig. The receiver coil usually surrounds the sample and acts as an antenna
to pick up the fluctuating nuclear magnetization of the sample and converts it to a fluctuating output
voltage V(t).
Where M(t,x)is the total magnetization in a volume and B c(x)the sensitivity of the receiver coil at
different points in space.
CIT 16 DEPT OF ECE
OEC 754 MEDICAL ELECTRONICS
➢ Bc(x)describes the ratio of the magnetic field produced by the receiver coil to the current in the coil.
The receiver coil design and placement is such that B c(x) has the largest possible transverse
component.
➢ The RF signals constitute the variable measured in magnetic resonance tomography. These are
extremely weak signals having an amplitude in the nV (nano-Volt) range thus requiring specially
designed RF antennas.
➢ The sensitivity of an MR scanner depends on the quality of its RF receiving antenna. For a given
sample magnetization, static magnetic field strengths and sample volume, the signal-to-noise-ratio
(SNR) of the RF signal at the receiver depends in the following manner upon the RF-receiving
antenna.
SNR ~ K (Q/Vc)
Where K is a numerical constant, specific to the coil geometry Q is the coil magnetization factor, and
Vc is the coil volume.
➢ This implies that the SNR of an MR scan can be improved by maximizing the ratio of magnetization
to coil volume.
➢ The receiver coil is a matching network which couples it to the pre-amplifier in order to maximize
energy transfer into the amplifier. This network introduces a phase shift f to the phase of the
signal.The pre-amplifier is a low-noise amplifier which amplifies the signal and feeds it to a
quadrature phase detector.
➢ The detector accepts the RF NMR signal which consists of a distribution of frequencies centered
around or near the transmitted frequency w and shifts the signal down in frequency by w.
➢ The detector circuit accepts the inputs, the NMR signal V(t) and a reference signal, and multiplies
them, so that the output is the product of the two inputs. The frequency of the reference signal is the
same as that of the irradiating RF pulse.
➢ The output of the phase-sensitive detector consists of the sum of two components, one a narrow
range of frequencies centered at 2w0,and the other, a narrow range centered at zero.
➢ The low pass filter following the phase-sensitive detector removes all components except those
centered at zero from the signal. It is necessary to convert the complex (two-channel) signal to two
strings of digital numbers by analog-to-digital converters. The A-D converter output is passed, in
serial data form to the computer for processing.
The advantages of the NMR Imaging System are:
(i) The NMR image provides substantial contrast between soft tissues that are nearly identical in
existing techniques. NMR images that display T1 andT2properties of tissue provide tremendous
contrasts between various soft tissues, contrasts approaching 150% are possible in T1 and T2
images, while contrasts of only a few percent are possible between soft tissues with X-rays.
(ii) Cross-sectional images with any orientation are possible in NMR imaging systems.
(iii) The alternative contrast mechanisms of NMR provide promising possibilities of new diagnostics
for pathologies that are difficult or impossible with present techniques.
(iv) NMR imaging parameters are affected by chemical bonding and, therefore, offer potential for
physiological imaging.
(v) NMR uses no ionizing radiation and has minimal, if any, hazards for operators of the machines
and for patients.
(vi) Unlike CT, NMR imaging requires no moving parts, gantries or sophisticated crystal detectors.
The system scans by superimposing electrically controlled magnetic fields. Consequently, scans
in any pre-determined orientation are possible.
(vii) With the new techniques being developed, NMR permits imaging of entire three-dimensional
volumes simultaneously instead of slice by slice, employed in other imaging systems.
➢ The reflected echoes are now displayed on the screen as a useful image. The various modes of display
are A mode, B mode, M mode or T-M mode.
➢ A mode (Amplitude modulation): The reflected echoes are depicted as vertical spikes along the
horizontal baseline. This mode is primarily used in ophthalmology and echoencephalography.
➢ B mode (brightness modulation): the reflected echoes are depicted as dots on the screen. Every
reflection produces a single dot. The brightness of the dot depends on the intensity of the reflected
echo.
➢ T-M mode (Time motion mode): We have unidimensional graphic representation of two-dimensional
structures. It is primarily used to study moving objects like the values and the walls of the heart.
Usually the machine has a line called the M- Cursor. By positioning the cursor at the required point
we can get an M mode tracing of the heart at that point. M mode is very useful in studying cardiac
function, by making various accurate measurements of the chambers and valve movements.
➢ The figure shows the block diagram of modern computer controlled ultrasonic image forming
system. It consists of so many peripheral sub units which are controlled by a computer through the
control buses.
➢ Transducer position data are fed to the computer. The computer sends this information to signal
processing unit which also receives the signal from the receiver.
➢ Then, it controls the receiver sensitivity also. Proper depth gain compensation is calculated by the
computer and given to the signal processing unit.
➢ The ultrasonic velocity is calculated and given to display unit. Here there is no drift, noise and
nonlinearity which are normally associated with analog devices.
➢ By using the image storage unit, the patient information data can be displayed again leisurely for
detailed examination.
➢ The received signals are converted into digital signals and are stored in the memory. Meanwhile the
scan converter control receives signals of transducer position and TV synchronous pulses and
generates X and Y address information which is fed to the digital memory.
➢ The stored digital image signals are processed and colour coded and are given to digital to analog
converter.
➢ Finally, they are fed into video section of the television monitor. By this circuit operations can be
performed quickly with higher frequency.
PART A
1. Give two important factors that demand internal pace maker’s usage. (Nov/Dec-2013)
The two important factors that demand internal pace maker’s usage are
(i). Type and nature of the electrode used
(ii). Nature of the cardiac problems.
(iii). Mode of operation of the pacemaker system.
6. Calculate the energy stored in 16μF capacitor of a DC defibrillator that is charged to a potential of 5000
Vdc.
Given Data: C = 16μF V= 5000 E= (1/2) CV 2 =(1/2) (16* 10-6) (25* 106)=200 Joules
10. Why asynchronous pacemakers (Fixed rate pacemakers) no longer used? (Nov/Dec-2016)
1. Using fixed rate pacemaker the heart rate cannot be increased
2. Simulation with a fixed impulse frequency results in the ventricles and atria beating at different rates.
This varies the stroke volume of heart and causes some loss in cardiac output.
3. Possibility of ventricular fibrillation will be more.
4. There may be competition between the natural heart beats and pacemaker beats.
12. When does the need for pacemaker arise? What is its function? (Nov/Dec 2015)
In cardiac diseases, where the ventricular rate is too low, it can be increased to normal rate by using
pacemakers. The various arrhythmias (rhythm disturbance) that result in heart block and Adams stokes
attacks represent a serious pathological condition. During that time, the patient becomes invalid because
of the constant risk of sudden losing consciousness. By fixing the artificial electronic pacemakers, the
above defects in the heart can be eliminated.
CIT 21 DEPT OF ECE
OEC 754 MEDICAL ELECTRONICS
15. What are the batteries used for implantable pacemakers? (Nov/Dec 2012)
The batteries used for implantable pacemakers are 1. Mercury cells 2. Lithium cells 3. Rechargeable cells
4. Nuclear cells 5. Bio Fuel cells 6. Bio Mechanical power generation sources.
19. What is the systolic and diastolic pressure of different areas of heart?
Left Ventricle: 130/5
Right Ventricle: 25/0
Left Atrium: 9/5
Right Atrium: 3/0
22. What are the two types of procedures for doing dialysis?
The two types of procedures for doing dialysis
1. Hemodialysis 2. Peritoneal dialysis
24. Which types of electrodes are used in internal stimulation and what is the current range?
The electrodes in the form of fine wires of Teflon coated stainless steel, spoon like electrodes are
used.The current range in 2 -15 mA