Professional Documents
Culture Documents
• Respiratory Instrumentation
– Mechanism of respiration
– Spirometry
– Pneumotachograph
– Ventilators
PACEMAKERS
– When placed over the skin layer covering the pacer, the magnet activates a
magnetically operated switch that prevents the pacer from sensing R-wave
activity.
– sense each intrinsic R wave and pacer emits an impulse with the occurrence of
each sensed R wave.
– The pacing impulses are transmitted to the myocardium during its absolute
refractory period, however, so they will have no effect on normal heart activity.
– When the intrinsic (natural) heart rate fall below the preset rate of the pacer, the
pacer will automatically operate synchronously at its preset rate to pace the heart.
• Timing circuit output signal feeds into a pulse width circuit (second RC
network ), which determines the stimulating pulse duration.
• When cell depletion has occurred, Rate slowdown circuit shuts off some of the
current to the basic timing network to cause the rate to slow down 8 ± 3 beats
per minute
• The sensing circuit detects a spontaneous R wave and resets the oscillator
timing capacitor.
• In the absence of an R wave, this circuit allows the oscillator to pace at its
preset rate ± 1 beat per minute.
DEFIBRILLATORS
• The rapid spread of action
potentials over the surface of the
atria causes two chambers of the
heart to contract together and
pump blood through the two
atrioventricular valves into the
ventricles.
• Drawback AC defibrillation:
– It cannot be successfully used to correct atrial defibrillation.
– attempts to correct atrial fibrillation by this method often result in the more serious
ventricular fibrillation.
– Thus, ac defibrillation is no longer used.
• DC defibrillation:
– In this method, a capacitor is charged to a high dc
voltage and then rapidly discharged through
electrodes across the chest of the patient.
• Energy delivered is represented by the typical waveform shown in below Figure. (time plot of the current
forced to flow through the thoracic cavity).
• It can be seen that the peak value of current is nearly 20 A and monophasic.
• An inductor in the defibrillator is used to shape the wave in order to eliminate a sharp, undesirable current
spike.
• The amplitude of this waveform is relatively constant, but its duration may be varied to obtain
the amount of energy required.
• To properly deliver a large current discharge applied through the skin large electrodes are used.
• These electrodes, called paddles, have metal disks that usually measure from 8 to 10 cm (3 to 4
in.) in diameter for external (transthoracic) use.
• Conductive jelly or a saline-soaked gauze pad (the latter is preferred) is applied between each
paddle surface and the skin to prevent burning.
• The two defibrillator electrodes applied to the thoracic walls are called either
anterior-anterior or anterior-posterior paddles.
– With anterior-anterior paddles, both paddles are applied to the chest.
– Anterior-posterior paddles are applied to both the patient's chest wall and back so that
the energy is delivered through the heart.
– The optimum time for discharge is during or immediately after the downward slope of
the R wave
(when the heart is in its absolute refractory period).
– This synchronization will ensure that the counter shock is not delivered during the
middle of the T wave, which is called the heart's vulnerable period.
– During this time, since it is partially refractory, the heart is susceptible to ventricular
fibrillation by the introduction of artificial stimuli.
Shortwave diathermy
• Diathermy means ‘through heating’ or producing deep heating
directly in the tissues of the body.
• A second winding can provide heating current for the cathode of the triode valve.
• The tank (resonance) circuit is formed by the coil AB in parallel with the condenser
C1.
• There is another coil EF and a variable condenser C2 which form the patient’s
resonator circuit due to its coupling with the oscillator coil AB.
• The anode supply of such a circuit is around 4000 V.
• The conduction in the triode takes place during the positive half-cycle and the high frequency is generated only during this period.
• The supply voltage is rectified before supplying to the anode of the oscillator valve. In such a case, the oscillations produced are
continuous and more power thus becomes available.
• In order to ensure that the oscillator circuit and the patient’s resonator circuit are tuned with each other, an ammeter is placed in series
with the circuit.
• The variable condenser C2 is adjusted to achieve a maximum reading on the meter, the needle swinging back on either side of the tuned
position.
• A thermal delay is normally incorporated in the anode supply which prevents the passage of current through this circuit until the
filament of the valve attains adequate temperature.
• The patient circuit is then switched on followed by a steady increase of current through the patient.
• A mains filter is incorporated in the primary circuit to suppress interference produced by the diathermy unit itself.
• There are several ways of regulating the intensity of current supplied
to the patient from a short-wave diathermy machine. This can be done
by either
– (i) controlling the anode voltage, or
– (ii) controlling the filament heating current, or
– (iii) adjusting the grid bias by change of grid leak resistanceR1, or
– (iv) adjusting the position of the resonator coil with respect to the oscillator
coil.
• These pads or electrodes do not directly come into contact with the skin
– Usually layers of towels (sometimes air spaced also) are interposed between the
metal and the surface of the body.
– The pads are placed so that the portion of the body to be treated is sandwiched
between them.
• This arrangement is called the ‘Condenser Method’ wherein the metal pads
act as two plates while the body tissues between the pads as ‘dielectric’ of
the capacitor.
• When the radio frequency output is applied to the pads, the dielectric
losses of the capacitor manifest themselves as heat in the intervening
tissues.
• The dielectric losses may be due to vibration of ions and rotation of dipoles
in the tissue fluids (electrolytes) and molecular distortion in tissues which
are virtually insulators like fats.
• This cable is coiled around the arm or knee or any other portion of the patient’s body where plate electrodes are
inconvenient to use.
• When RF current is passed through such a cable, an electrostatic field is set up between its ends and a magnetic field
around its centre
• Deep heating in the tissue results from electrostatic action whereas the heating of the superficial tissues is obtained by
eddy currents set up by a magnetic effect.
• Intensity of treatment is dependent on the subjective sensation of warmth felt by the patient.
Haemodialysis machine
• General scheme for operation of Haemodialysis machine is shown below:
Figure:
• An Artificial kidney periodically connected to the circulatory system of uremic
patients to remove metabolic waste products from the blood.
(Metabolic wastes are substances left over from metabolic processes which cannot be used
by the organism. This includes nitrogen compounds, water, CO₂, phosphates, sulphates, etc.)
• The standard spirometer consists of a movable bell inverted over a chamber of water.
• Inside the bell, above the water line, is the gas that is to be breathed.
• The bell is counterbalanced by a weight to maintain the gas inside at atmospheric pressure
– so that its height above the water is proportional to the amount of gas in the bell.
• A breathing tube connects the mouth of the patient with the gas under the bell.
• Thus, as the patient breathes into the tube, the bell moves up and down with each
inspiration and expiration in proportion to the amount of air breathed in or out.
• As the kymograph rotates, the pen traces the breathing pattern of the patient.
• Several other types of Spirometers are available.
– For example, waterless spirometers (shown in Figure), which are also used clinically, operate on a principle similar to
that of the spirometer.
• Principle:
– In this instrument the air to be breathed is held in a chamber enclosed by two parallel metal pans hinged to each
other along one edge.
– The space between the two pans is enclosed by a flexible bellows (Like a fireplace bellows) to form the chamber.
– One of the pans, which contains an inlet tube, is fixed to a stand and the other swings freely with respect to it.
– The instrument provides electrical outputs proportional to both volume and airflow, from which the required
determinations can be obtained.
Spirogram
• In order to produce a spirogram,
– the patient is instructed to breathe through the mouth piece of the spirometer.
– His nose is blocked with a clip so that all breathing is through the mouth.
• The recorder is first set to a slow speed to measure vital capacity (typically, 32 mm/min).
– The patient breathed quietly for a short time at rest so as to provide a baseline.
– He was then instructed to exhale completely and then to inhale as much as he could.
– This process produced the vital capacity record at the extreme right of the figure.
• With his lungs at the maximal inspirational level,
– the patient held his breath a short time while the recorder was shifted to a higher
chart speed (e.g., 1920 mm/min).
– The patient was then instructed to blow out all the air he could as quickly as
possible to produce the FEV, curve on the record.
– To calculate the FEV1, a 1 -second interval was measured from the beginning of the
maximum slope.
• For the determination of maximal voluntary ventilation (MVV), the recorder is set at an intermediate speed.
– After a short rest, a few cycles of resting respiration were recorded.
– The patient was then instructed to breathe in and out as rapidly as possible for about 10 seconds, producing the MVV
record in the figure.
• Some instruments require that the height of the tracings be converted to Liters by use of a calibration factor
for the instrument, called the spirometer factor.
– This calibration factor can be obtained from a table or chart.
• Most ventilators in clinical settings use positive pressure during inhalation to inflate the
lungs with various gases or mixtures of gases (air, oxygen, carbon dioxide, helium, etc.).
• Most respirators can be operated in any one of the three different modes:
– Assist mode
– Controller mode
– assist-controller mode
• In the assist mode,
– The assist mode is used for patients who are able to control their
breathing but are unable to inhale a sufficient amount of air
without assistance.
• In the control mode,