user manual
low temperature circulating water tank

[REFRIGERATING BATH CIRCULATOR]


RW-0525G/1025G/2025G/3025G
Model:
RW-0540G/1040G/2040G/3040G

Manual No : 00HAA0001182 ( Version : 5.0 )

Page One
Page2
Thank you for purchasing a product from J.O Tech. We, J.O Tech Co., Ltd.

We always do our best for customer satisfaction.

This device is a new product from Jeotech Co., Ltd., which is focusing on the localization of physical and chemical devices.

It is a product made with technology and materials.

This user manual describes the functions and handling precautions of the purchased device.

are describing. All users should read this user manual before using the device.

Please read carefully to the end. To use this unit effectively and correctly

We hope you will take full advantage of this device.

Before using this device, be sure to read the user's manual carefully and completely

Read on. Pay special attention to safety precautions when using.

Safety Precautions
WarningIf you operate this unit incorrectly,

Serious damage may occur to the user.

means the case.

CautionIf you operate this unit incorrectly,

If the user has suffered minor damage or

This may cause physical damage to the device.

means the case

Copyright 2002 J.O Tech. ALL RIGHT RESERVED.

Page3
[ Customer Support Contact ]

Seoul Branch

[Tel] 02-2627-3800
[Fax] 02-325-2335

Suwon Branch Daejeon Branch

[Tel] 031-254-0336~7 [Tel] 042-933-9462

[Fax] 031-254-0152 [Fax] 042-933-9460

Yeongnam Branch

[Tel] 053-381-9591
[Fax] 053-381-9594

www.jeiotech.com

e-mail : sales@jeiotech.com

If you have any questions while using the device, please contact the contact information above or the place of purchase.

please give it to me. Fast and accurate answers to faithfully utilize your devices

We're here to help.

Page4
index

1. Purpose and characteristics of the device ............................................... ............... 6

2. Equipment installation and precautions ............................................... …………………… 7

3. Name and operation of each part ............................... ............. 8

4. Using the device ............................................... ……………………………………………………………………………………………………………… 12

5. MAINTENANCE AND MAINTENANCE OF THE EQUIPMENT ............................... ............... 18

6. Measures to be taken in case of an abnormality ............................... ............... 19

7. Product Warranty Criteria ............................................... …………………………………………………………………………………………………… 20

8. Specifications ............................................................... …………………………………………………………………………………………………… 21

9. Monitoring Program Installation ............................................... .... 25

10. How to use the Monitoring Program ............................................... 26

Page5
1. Purpose and characteristics of the device

It is a multi-purpose low-temperature circulating water tank used in various fields such as pharmacy,
medicine, chemistry, and biology.CLS(Custom Logical Safe)-Control Systemhas been implemented

(One)CLS-Control Systemmeans “a control device equipped with a logical safety mechanism dedicated to a specific device”

Thermal safety is perfect like in a laboratory where various flammable chemicals exist.

Research by JEO Tech for the safe use of heating-type devices operated in a required environment

Logic and composition developed by JinThe highest safety level control deviceis.

(Patent registration number: 0328729, 0397583)

(2)CLS-Control SystemThe temperature control and heating amount of the silver device are capable of sophisticated PID calculation.

It is performed by the main CPU, and all operations for safety are separately built-in Logic of the preceding function.

The device is designed to perform in the IC so that it can withstand any electrical or electronic disturbances applied to the device.

It is configured to perform certain actions for your safety.

(3)CLS-Control SystemWhen the safety device applied to the silver device operates, power is supplied to various parts2

All phases are immediately completely blocked, and the activated unsafe element is presented to the user as a visual and auditory device.

In a safe state, de-energized and until restarted after the unsafe element has been removed

will keep.

(4)CLS-Control Systemwill continue to operate the device when power is restored after a power outage

You want to stop the operation of the device even if power is restored due to other factors nearby.

You can choose when to use it.

(5)CLS-Control SystemThere is a weak current within only 5V and 10mA at the contact point of the safety device used.

It is constructed to flow, so there is no damage to the contact point even after a long time of use, so its durability is very long.

and an electronic device that controls the current flow of a heater that uses a large current when the safety device is activated

After stopping the control of (thyristor) first, through the contact separation of the power switch (Magnetic Switch)

By cutting off the power, damage to the contacts of the power switch and occurrence of noise were prevented.

Prevent damage to the contact point of the power switch and the occurrence of noise by performing in the reverse order when starting.

I did.

Page6
2. Equipment installation and precautions

(1) After opening the package, check the contents of the device.

Device body (1set), user manual (1 copy), glass tube fuse (2EA), silicone hose, communication

CD(1EA), CABLE for communication(1EA)

(2) This device is designed to operate normally under the specified power supply. Before installing the device

Check if the power on the ID Plate (name plate attached to the back of the product) matches the power used.

Power must be level 2 or higherUse a grounded power supplyand the Power Cord must be

You must use an outlet with a ground terminal.

(3) A suitable thermal liquid should be as non-flammable as possible, and in the case of flammability, the flash point (Fp) is 40°C

with more than 20 liquids°Liquids that can be circulated and pumped even below C should be used.

5 above the flash point of the flammable fruit°It is safe to use at a temperature lower than C (Fp-5).

- Be sure to check the power supply of the device.

- Be sure to use it under the power supply condition that can be grounded.

- Fruit oil has a flash point of 5°C low (Fp -5°C) Use it as a condition.

(4) When transporting the product, use an appropriate tool or carry it with two or more people.

(5) The outlet should be installed near the device and should be easy to use.

(6) Install the device on a flat and hard floor that is not subject to vibration or impact.

(7) Avoid heat sources or direct sunlight, and ambient temperature 5°C to 40°C, relative humidity

Install at 80% or less.

(8) Where there is an organic solvent such as Acetone, Methylene Chloride, etc. or strong high-frequency noise

Avoid use in the vicinity of devices where they occur.

(9) Prevent moisture, organic solvents, dust, and corrosive gas from entering the control box of the device.

do it.

Page7
3. Name and operation of each part

1) Display

Equipped with Main CPU equipped with proven software capable of Digital PID-Auto tune, Pt-

100ΩEquipped with the highest safety level control device such as temperature compensation function and heating amount control function for the sensor.

I am saying

Page8
(1) RUN LED

This LED indicates whether the device is running or stopped. When the device is in normal operation

The LED lights up, and the LED turns off when the instrument is stationary.

(2) HEATER LED

This LED indicates that the heater is being output.

(3) Auto Tuning LED

This LED blinks during Auto Tuning.

(4) Wait On Timer LED

This is the LED that displays the timer with the function of starting the device after the set time has elapsed. LED

When is blinking, the timer is running; when the LED is fully lit, the timer is running.

Indicates that it is waiting.

(5) Wait Off Timer LED

A timer with a function to stop the device after the set time elapses after PV and SV match

LED to indicate. When the LED is blinking, the timer is running and the LED is fully lit.

indicates that the timer is waiting for operation.

(6) Low water level warning LED

It is a device that detects the low water level of the thermal liquid (distilled water, alcohol, oil, etc.) in the bath.

Because sus-ball is used, it can be used at high temperatures as well as durability, and when low water level is detected,

The machine will stop and the buzzer will continue to sound. In this case, after replenishing the solution, the Start/Stop

Press the button to restart.

(7) Overheat warning LED

The temperature of the heating liquid in the bath becomes higher than the set value of the mechanical overheat prevention device installed in the front part.

In this case, the machine stops, a buzzer sounds and the overheat warning LED blinks. device is insecure

Since it is in a stopped state due to an element (overheating), take an appropriate action and press the Start/Stop button 1

Press twice to remove the buzzer and LED flickering before using the device.

(8) Temp button

Used to set the temperature.

(9) Timer button

It is used to set the timer, auto run function, and adjust the pump circulation amount.

Page9
(10) Up button

It is used to increase the set value.

(11) Enter button

After changing the setting value, press this button to save the setting value.

(12) Down button

It is used to decrease the set value.

(13) Lock button

A button that locks the buttons on the controller so they cannot be used.

(14) Auto Tuning button

When executing Auto Tuning, press this button for 3 seconds to activate the function.

(15) Start/Stop button

When operating or stopping the device, or when the operation of the device is interrupted due to an unsafe factor

In this case, it is used to remove the warning sound and LED blinking after removing the risk factor.

(16) SV display window

Displays the set temperature or the remaining time when the timer is running.

(17) PV display window

Displays the current temperature.

2) Over-Temp. protector
It is a safety device with an independent circuit to protect the device and test subject from overheating.

The function of the controller is unstable or the temperature is higher than the set temperature due to the user's inexperienced operation.

When it heats up, it automatically stops the machine and sounds an alarm so the user can recognize overheating.

The LED on the display blinks.

3) Communication Port

Connect to PC through COM 1 or COM 2 port and use RS-232C protocol.

You can monitor the device status and change settings through the program. Also, the device

The control status can be saved and output through the printer is also possible.

Page10
4) Main Power Switch

Used to turn the main power off or on.

5) Fuse

It is connected to both phases to increase safety.

6) Drain Valve (rear)

It is used to drain the solution in the bath with a specially designed valve structure. counterclockwise valve

The solution in the bath is drained by turning it 2 or 3 times in the opposite direction.

Please note that if you turn the valve more than 5 times, the valve may come off. Drainage after high temperature test

When doing this, please turn the valve after wearing gloves.

7) Frozen UNIT

It is a place where the optimal refrigeration system is configured, and the better the ventilation is, the more efficient it is.

8) CONDENSER COVER

In order to facilitate the cleaning of the condenser, it is a removable screen that is used to keep the refrigerator at all times.

It allows you to keep it in optimal condition.

Page11
4. Using the device

1) Basic use
(1) Before connecting the power, fill the bath with an amount of thermal fluid (distilled water, oil) suitable for the purpose of the experiment.

(2) In case of internal circulation, connect the silicone hose to IN/OUT LINE.

(3) Over Temp. 10% higher than the set value. Position the Protector Knob.

(4) Connect the power, turn on the Main Power Switch, and input the desired setting value.

(5) Press Start / Stop Key to operate.

2) How to set the temperature

(One) Press the button.

Two windows to display the current temperature and set temperature

The number on the window displaying the set temperature blinks during

I write.

(2) When the number on the window displaying the set temperature is flickering

0000key and Use the key to set the desired temperature.

After entering, Press key to finish setting.

(3) When there is no key input for about 10 seconds during temperature setting

In this case, the change is canceled and the status returns to the original state.

is.

(4) Press key to display the set temperature window

When the key is pressed again, the next

additional features are displayed.

Page12
3) additional function of the button

(1) Set temperature storage function

① Press the button twice in succession.

- You can save and recall frequently used temperatures in SV.1.

It's a feature. key with Press key to set the temperature.

go Press key to finish setting. The set temperature is stored in the memory.

It is stored and controlled at the set temperature.

② Press the button 3 times in a row.

- You can save and recall frequently used temperatures in SV.2.

is a function. key with Press the key to set the temperature and

0000Press key to finish setting. The set temperature is stored in the memory.

It is stored and controlled by the set temperature.

③ Press the button 4 times in a row.

- You can save and recall frequently used temperatures in SV.3.

is a function. key with Press the key to set the temperature and

0000Press key to finish setting. The set temperature is stored in the memory.

It is stored and controlled by the set temperature.

(2) Change of display unit

① Press the button 5 times in a row.

- This is a function to change the unit of the displayed temperature.

- The first display unit is ℃, button and use the button

You can change between ℃ and ℉. Click the button to complete the setting

I'm done.

(3) Bias function

① Press the button 6 times in a row.

- It compensates the temperature error caused by the error of the temperature sensor and other causes.

function.

- The BAIS correction value is displayed on the current temperature display window (PV) and the set temperature table

The sight glass (SV) indicates BAIS status.

- button or Press the button to go to ADJUST mode.

Page13
The temperature displayed on the set temperature display window (SV) is

to the same value as the temperature on the thermometer. button and button

Use to change. Click the button to complete the setting.

4) How to set the timer

(1) Wait on Timer

① Press the button once.


Timer type (On Timer / Off Timer) is displayed on the

window displaying the current temperature, and the time is

displayed on the window displaying the set temperature.

button and Use the button to


After entering the time button to set
Finish.
② After completing the setting, the W/ON LED turns on with

a beep sound to inform the setting.

(2) Wait off Timer


① Off Timer can be set by pressing the
button twice in a row.

button and Use the button to

After entering the time set at the push of a button

to finish.

② After completing the setting, the W/OFF LED turns on with

a beep sound to inform the setting.

Page14
③ The function of Timer is as follows.

Tem pe ra tu re

T im e

W a it on T im er W a it o ff T im er

☞ Wait on Timer

- The device operates after the time set in On Timer has elapsed.

- The maximum settable value is 99 hours 59 minutes, and the minimum value is 1 minute.

☞ Wait off Timer

- Off Timer operates from the moment PV (current temperature) and SV (set temperature) coincide.

After the set time has elapsed, the device will stop.

☞ Mix of Wait on Timer and Wait off Timer

- If both timers are set together, the device operates as shown in the figure above.

is.

④ Cancel Timer setting

To cancel Timer setting When the button is pressed, the LED turns off and the

Timer is canceled. (Wait on/Wait off Timer are both canceled)

Or, if you want to cancel only one of the two timers, set the time of the timer you want to cancel.

Setting the interval to 0 cancels the Timer set to 0.

Page15
5) additional function of the button

(1) Auto Run function

① Press the button 4 times in a row.

- It is a kind of blackout compensation function. During operation of the device,

If the device is stopped due to a power outage, or if the power is accidentally turned off,

This is a feature that automatically enables the device to work once it is restored.

If set to Yes, the device will run as soon as the power is restored.

(2) Pump circulation volume control

① Press the button 5 times in a row.

- It is a function to control the circulation amount by adjusting the speed of the motor speed.

Control range: 1 ~ 5 units can be changed. button and friend

Use the button to set in units of 1 to 5 button to set

Finish.

(The motor speed is set to 5 when shipped from the factory.)

When setting the motor speed to 4 or less, the viscosity characteristics of the solution according to the temperature change should be considered.

In the case of low temperature, the viscosity of the solution increases and the motor becomes overloaded, which causes malfunctions and does not

function properly when used with external circulation.

6) Auto Tuning method


(1) Auto for more accurate and faster temperature control

Execute Tuning. After tuning, the PID value is

It is saved as

① Set the desired temperature.

② If you press and hold the button (about 3 seconds), the next

In the window that displays the temperature, such as the rim, Auto

A/T LED blinks with Tune indication.

It hangs.

③ Auto Tune time varies depending on usage environment

is. When Auto Tune is finished, the LED turns off,

The current temperature and the set temperature become the same.

Page16
7) Lock function

(1) This button locks the buttons on the controller so that they cannot be used.

① When the Lock button is pressed for about 3 seconds, the Lock function is set with the Lock indicator and a beep sound.

It becomes impossible to input button.

② Press the Lock button again for about 3 seconds to release it.

③ Unnecessary elements caused by accidentally pressing a button during temperature control are removed in advance.

This is a prevent function.

Page17
5. MAINTENANCE AND MANAGEMENT OF THE DEVICE

1) Device body

(1) Turn off the Main Power Switch and unplug the power plug from the power outlet.

(2) Remove the solution in the bath.

(3) Clean the bath with a soft cloth moistened with neutral detergent.

(4) Clean the bath with a soft cloth moistened with distilled water.

(5) Dry with a dry cloth.

(6) Do not use organic solvents.

(7) If the device is cleaned or decontaminated by other methods not specified in the manual, the user must

Appropriate methods not to damage the equipment should be checked with the manufacturer prior to implementation.

(8) Toxic chemicals that damage human skin or substances that generate dangerous gas overflow

When it spills, wear gloves that are safe for toxic chemicals and a mask that is safe for toxic gas.

It must be completely wiped off with a different cloth.

2) Accessory

(1) Soak Spring Wire Rack and Test Tube Rack in neutral detergent.

Spring Wire Rack Test Tube Rack


division
(WXDXH)mm (WXDXH)mm

RW-0525G/40G - -

RW-1025G/40G - 100 X 145 X 157

RW-2025G/40G 185 X 159 X 170 100 X 230 X 157

RW-3025G/40G 249 X 219 X 198 100 X 230 X 157

(2) Wash thoroughly with distilled water.

(3) Store after drying completely.

Page18
6. Measures to be taken in case of abnormality

1) If the device does not work


(1) Check the power used.

(2) Check the fuse.

(3) Check that the Run LED on the display is off. Start/Stop button if off

Press to activate it.

2) When temperature control is not available

(1) When “Err.S” is displayed on the window displaying the current temperature Input sensor disconnection or sensor

It is not connected, so check it.

(2) In case of temperature control or other causes, request service.

- If the device does not operate normally, turn off the power immediately.

- If you need A/S, you must contact JEO Tech or the place of purchase to receive repairs.

Do not disassemble the device arbitrarily.

- When handling electrical parts inside the device, only qualified personnel must handle it.

- If replacement of parts is necessary, only genuine parts must be used.

- Failures outside the normal limits cannot be technically maintained.

Page19
7. Product Warranty Standards

1) Free warranty repair service period

It is one year from the date of purchase, and the free warranty repair service is not available after one year.

is. After the free warranty repair service period, you must pay the purchase price for the necessary parts.

you can get

During the free warranty repair service period, repair, replacement and

Refund service is available. In the case of export products, Jeio Tech is responsible for all services.

In principle, you pay the cost.

2) If you cannot receive the free warranty

The product may be damaged due to accidents, contamination, or neglect of precautions due to natural disasters such as fire or water disaster.

Even if it is, you cannot receive the service. In addition, due to overvoltage ignoring the rated voltage,

Operation and misuse under abnormal conditions, including damage to the product, or resulting

Depreciation and damage to parts may also prevent you from receiving this service.

3) When receiving free warranty repair service

Postage by entering the model name, date of purchase, and serial number of the damaged product along with a detailed description

and insurance premiums payable to the payee. Of course, during the service period, the return fee is also

In principle, Jeio Tech bears the cost. Request product repair after the service period has elapsed

If you do, Jeio Tech will give you an estimate in advance.

Please time.

This warranty repair service covers only the free repair, replacement and refund of the Product, and for any other specific purpose.

does not apply to And in case of data loss due to other causes,

and Jeio Tech is not responsible for any indirect accidental damage or loss.

As some of the above provisions do not make exceptions, the consumer who purchased this product

Repair service may not be available.

Page20
8. Specification 1
Model RW-0525G RW-1025G RW-2025G RW-3025G

Bath Volume 5.0L 10.0L 20.0L 30.0L

Working Temperature Range2) - 25℃ ~ 150℃

Temperature StabilityOne)
0.05℃ 0.05℃ 0.05℃ 0.08℃
At -10℃ (bath fluid: ethanol)

Controller PID Controlled microprocessor touch pad, Digital Display


Temperature : 5℃ to 40℃
Relative humidity : 50% to 80%
Permissible environment
Altitude : Up to 2,000m
condition
Indoor use pollution : Degree 2
Installation category (Over voltage category) : Ⅱ

Internal Stainless steel, 1.0t, Cubic Type

External Steel, 1.2t, Double painted and baked

Material Heater 1000W/230V 1500W/230V 2000W/230V 2000W/230V

Sensor Pt 100Ω

Insulation Polyurethane foam 30mm

Bath

opening/depth 150X99,160 211X154,180 290X214,200 300X264,230


Dimensions
(WXL,D)
(mm)
Overall
302X438X690 360X490X780 380X560X825 440X620X895
(WXLXH)

Step One 2 3 4 5

Flow Rate 5.6 (l/min) 8 (l/min) 12.8 (l/min) 18.8 (l/min) 28 (l/min)

Height 0.35(m) 0.75(m) 1.7(m) 3.5(m) 4.2(m)

Suction - - - 5 (l/min) 10 (l/min)

Material Stainless steel mold pump


Pump
In/Out 12.7(mm)

Max. pressure,
5.9 PSI, (28.0 liters/minute)
max.flow rate

Max. suction
10.0 (liters/minute)
flow rate2)

Cooling at -20℃ 30W 50W 260W 390W

Page21
at 0℃ 230W 250W 410W 570W

at 20℃ 320W 380W 560W 680W

Safety Device CLS (Custom Logical Safe) – Control System

Electric Requirement
6.8A 8.9A 12.5A 13.5A
220VAC, 60Hz

Weight (Net) 37.0kg 44.0kg 60.0kg 67.0kg

※ 1) Technical data (according to DIN 12876)

Page22
8. Specifications 2
Model RW-0540G RW-1040G RW-2040G RW-3040G

Bath Volume 5.0L 10.0L 20.0L 30.0L

Working Temperature Range2) - 35℃ ~ 150℃

Temperature StabilityOne)
0.05℃ 0.05℃ 0.05℃ 0.08℃
At -10℃ (bath fluid: ethanol)

Controller PID Controlled microprocessor touch pad, Digital Display


Temperature : 5℃ to 40℃
Relative humidity : 50% to 80%
Permissible environment
Altitude : Up to 2,000m
condition
Indoor use pollution : Degree 2
Installation category (Over voltage category) : Ⅱ

Internal Stainless steel, 1.0t, Cubic Type

External Steel, 1.2t, Double painted and baked

Material Heater 1000W/230V 1500W/230V 2000W/230V 2000W/230V

Sensor Pt 100Ω

Insulation Polyurethane foam 30mm

Bath

opening/depth 150X99,160 211X154,180 290X214,200 300X264,230


Dimensions
(WXL,D)
(mm)
Overall
302X438X690 360X490X780 380X560X825 440X620X895
(WXLXH)

Step One 2 3 4 5

Flow Rate 5.6 (l/min) 8 (l/min) 12.8 (l/min) 18.8 (l/min) 28 (l/min)

Height 0.35(m) 0.75(m) 1.7(m) 3.5(m) 4.2(m)

Suction - - - 5 (l/min) 10 (l/min)

Material Stainless steel mold pump


Pump
In/Out 12.7(mm)

Max. pressure,
5.9 PSI, (28.0 liters/minute)
max.flow rate

Max. suction
10.0 (liters/minute)
flow rate2)

Cooling at -20℃ 80W 150W 400W 440W

Page23
at 0℃ 150W 330W 660W 700W

at 20℃ 280W 450W 800W 880W

Safety Device CLS (Custom Logical Safe) – Control System

Electric Requirement
8.3A 10.4A 13.5A 14.5A
220VAC, 60Hz

Weight (Net) 38.0kg 44.0kg 60.0kg 67.0kg

※ 1) Technical data (according to DIN 12876)

Page24
9. Monitoring Program installation

(1) Insert the installation CD into the CD-ROM drive and the installation program will start automatically.

(If it does not run automatically, run the “SETUP.exe” file)

(2) If you click the “Next” button, you can specify the path to install the program.

(It is recommended to install in the path set by default.)

(3) Click the “Start Installation” button to start installing the program.

(4) After the installation is completed, the Lab Tracer icon is created on the desktop.

(5) Double-click the created icon to run the Monitoring Program.

basic specifications Recommended Specifications

Microsoft Windows 98 or later Microsoft Windows 2000, XP

CPU: P-II 233 or higher CPU: P-III 300 or higher

RAM: 32M Byte or more RAM: 64M Byte or more

caution: If you use Windows 95, 98 series OS, there may be an error in measurement time due to OS structure.

Page25
10. How to use Monitoring Program

(One)Comm → ConnectClick to connect the PC and the device through RS-232 communication.

(If the connection is notComm → Port settingClick to set the port number)

If the communication connection is successful, it is displayed at the bottom of the program window (circled).On Linemark appears.

The program window is again divided into two windows.

Displays the current temperature, and the window below shows the output of the heater as a graph.

(2)View menu

View → ParameterIf you select , the current temperature, set temperature, and heater output are displayed along with the graph.

A numerical window will appear.

View → StatusIf you select , another window appears at the bottom of the current temperature and set temperature display window.

is created. In this window, items indicating the operating status of the device are displayed. Operating,

Auto Tune, Program, Over Temp., Door, etc. are displayed.

Page26
The next picture isViewof the menuStatusWowParameterAfter selecting , monitor the device status

One screen.

Through three windows (Graph window, Parameter window, Status window), the operating status of the device is accurately monitored.

You can monitor.

① The contents of the Graph window are as follows.

The upper window among the graph windows indicates the current temperature (red line) and the set temperature (blue line).

② The contents of each item in the Status window are as follows.

- Operating indicates that the device is running.

When the blue line is Hi, it means On (operation), and when it is Low, it means Off (not operation).

indicates.

- Auto Tune indicates whether or not Auto Tuning is performed.

- Program item indicates whether program operation is performed.

- Over Temp. The item indicates whether the device is overheating.

- Door Open indicates whether the device door is opened or closed.

When the door of the device is opened, the blue line indicates Hi, and when it is closed, it indicates Low.

③ The contents of each item in the Parameter window are as follows.

- PV indicates the current temperature, SV indicates the set temperature, and Heat indicates the output of the heater.

- Run Time indicates the time the device has been operated.

- This is the time from the moment the Run LED turns on by pressing the button to start the device.

- Wait On Timer, Wait Off Timer is the remaining time of the set Timer. in other words,

It is the time minus the elapsed time from the set time.

- Power Frequency indicates the frequency of the power currently being used.

Page27
(3) Menu Icon Description

ⓐⓑ ⓒⓓⓔ ⓕⓖ ⓗⓘ ⓙⓚⓛⓜ

ⓐ File Open (Ctrl + O)

It is used to call up a graph that has been saved previously.

ⓑ File Save (Ctrl + S)

It is used to save the current graph as a file.

ⓒ Connect (Ctrl + C)

It is used when connecting the experimental device and PC through RS-232 communication.

ⓓ Disconnect (Ctrl + D)

It is used to disconnect RS-232 communication.

ⓔ Exit (Ctrl + X)

End program button. When this icon is executed, the graph so far

You will be asked to save.

ⓕ Print (Ctrl + P)

Print the graph.

ⓖ Preview

This is a preview function before printing the graph.

ⓗ ⓘ Scroll button

This function scrolls the graph forward or backward.

ⓙ ⓚ Auto Trace On/Off

Auto Trace if you want the end point of the graph (the current temperature point of the device) to always be in the center of the window

On.

ⓛ ⓜ Auto Span On/Off

Set the range of Y-axis (temperature range) of graph automatically and manually. set manually

You can put any range of values you want.

Page28
ⓝⓞⓟ ⓠⓡ ⓢⓣⓤⓥ ⓦⓧ ⓨ

ⓝ This function turns the Status window On/Off. (Ctrl + T)

ⓞ This function turns the parameter window on/off. (Ctrl + R)

ⓟ Panel View

The same screen as the front of the control unit's control panel is displayed on the PC screen.

ⓠ This is the function to set the pattern of program run and run the program.

A maximum of 100 patterns and a maximum time of 99 hours can be tested.

- Program function can only be controlled using PC and cannot be set in Main.

ⓡ Program Run function.

This function is available only when Program Run is set on the device.

ⓢ ~ ⓥ Zoom In/Out function.

ⓦ ~ ⓧ This is a function to change the unit of temperature display in the program.

The temperature unit of the device does not change.

To change the temperature unit of the device, it must be changed in the control unit of the device.

ⓨ Clear the graph.

(4) Description of Print function

Page29
① Print range

· Print All: Prints the entire page.

· Print Screen: Prints the current screen.

(If Graph, Status, Parameter Frame is on the main screen, it is printed, if not

It does not print.)

· Print Current Page: Prints the page of the current main screen.

· Partial Print: Prints the page the user wants.

② Number of copies

Used when printing multiple copies.

If you press the Up and Down buttons, you can set from 1 to 100 sheets.

③ PV print interval

If Print at equal intervals is checked, PV and SV are converted to text according to the set print interval.

It will be printed.

④ Memo

This function allows you to make simple notes on the printed page.

A maximum of 60 characters is allowed.

⑤ Select Print

When you select a printer, it prints to the selected printer.

(5) Preview and Print

Page30
If the Print at equal intervals checkbox is checked on the print screen, the above

As shown in the figure, PV and SV are printed at regular intervals during preview and printing.

Also, if there is a particular location that the user wants to check, move the mouse cursor to the desired location.

When clicked, PV and SV are displayed along with the green line as shown in the picture above.

① Last Point Delete

Deletes the last set point.

② All Point Delete

Deletes all set points.

③ Zoom in/out

Enlarges or reduces the screen.

(6) Display

All operations of the display window are configured the same as those of the device.

If the PC and the device are connected, all control operations can be performed equally with the PC.

(7) Pattern Program

Figure 1 Main Lab Tracer Screen

Page31
Click the PRG icon or select Pattern -> Pattern settings from the menu to

The Programming window opens as shown in Figure 2.

Figure 2 Pattern Program screen

As shown in Figure 3 below, if you click the mouse on the SV Pattern Programming window, the set temperature and time,

The step number is displayed on the left.

Fig. 3 SV Pattern setting screen

If you want to edit the set value, drag & drop the selected Step (blue) to change the value.

If you change the temperature and time using the shortcut key, the temperature is changed by 1℃ and the time is changed in units of 1 minute.

Convenient.

① Shortcut

↑ : raise the temperature by 1℃ Alt + ↑ : Move the edit point to the right (next step)

↓ : Decrease the temperature by 1℃ Alt + ↓ : Move the position of the edit point to the left (previous step)

← : Decrease time by 1 minute Alt + ← : Move the position of the edit point to the left (previous step)

→ : Round up the time by 1 minute Alt + → : Move the edit point to the right (next step)

Page32
② Last step delete

Delete the last set Step.

③ All step delete

Delete all set steps.

④ Pattern save

Save the set pattern. The extension of the saved file is PIT.

Select the folder you want to save, name the file, and click the Save button to save it.

⑤ Pattern open

Open the pattern saved in the storage device.

Select the file you want to open and press Open to display the saved pattern.

⑥ Start

After pattern setting is completed, press this button to operate the device. At this time, the low water level, Over

Temp. If there is something wrong with the device, such as, it will not work.

Figure 4 Step information and control options

If the number of repetitions of the pattern is specified, the entire set pattern is repeated as many times as the specified number of times.

“Deleting the previous data”When you start after checking the checkbox, the previously tested data is

It starts by erasing everything, so use it with caution.

Page33
☞ Caution

The maximum pattern setting time is 99 hours.

If 99 hours is exceeded by increasing the number of repetitions, it does not run in Program Mode.

When setting the pattern, time and temperature must be set according to the performance of the device.

If you forcefully set the pattern beyond the maximum performance of the device, you will not be able to perform the desired experiment.

Page34
Page35