Professional Documents
Culture Documents
Q lit Control,
C t l Quality
Q lit Assurance,
A
Proficiency Testing I & II
HKU SPACE
Higher Certificate in Medical Laboratory Science
Laboratory Management
Vanessa Lo
Adjunct Lecturer
Clinical Scientist, FFSc(RCPA)
2&9S
September
t b 2022 (6 (6:30
30 – 9:30
9 30 pm))
Quality
Q lit Management
M t
System (QMS)
Internal
External
E ternal
Quality
Assurance
4
Two Types of Laboratory Error
1. Systematic Error
• Poor
P accuracy / Bias
Bi
• Definite cause
• Reproducible
Reprod cible
• Avoidable through adoption of the best
practice or state of the art technology
2. Random Error
• Poor precision
• Non specific cause
• NOT Reproducible
R d ibl
• Can happen anytime
• NOT avoidable
id bl
5
Commercially Prepared
• Clinical and Laboratory Standard Institute
(CLSI) subcommittee on media quality control
collected data over several years regarding
the incidence off QC failure
f off commonly
used microbiology media
• Based on the findings the subcommittee
published a list of media that did not require
retesting in the user's laboratory if purchased
from a manufactures who follow CLSI
guidelines
8
However:
1. Quality control tests should be carried out
by the end-user laboratory to ensure that
• the performance characteristics of the
medium are within specification and
• the methodology of medium preparation
is satisfactory
2. Each lot/batch of prepared medium should
b subjected
be bj t d tto a minimal
i i l testing
t ti program
which will
• ensure that
th t it is
i acceptable
t bl andd
• demonstrate a typical bacterial
performance
f
9
pH Value
• Check that pH of the prepared medium, when
tested in final form at ambient temperature
(25°C) lies within the range given on the
product label
• The medium should be discarded if the pH
value lies outside the specified range
Stability
• Periodically perform ALL checking
procedures on stored prepared media in order
to determine whether the storage conditions
will give optimal results
10
Sterility Check
• Check that pH of the prepared medium, when
tested in final form at ambient temperature
(25°C) lies a representative sample of each
lot/batch of medium should be incubated for 2
to 5 days at 35 to 37°C
• Testingg sample
p number depends
p on lot size
• <=100 units 3 to 5% samples
• > 100 units 10 random samples
• There should be no evidence of microbial
gro th after incubation
growth inc bation
• Discard all samples of the batch which did not
pass the sterility check
11
Growth Performance
• Test the growth support properties of the product
byy inoculating
g the medium with appropriate
pp p
stock cultures and/or fresh isolates
• Use a standard inoculation procedure
p
• Examine the quantitative and qualitative
results
• Test with organisms expecred to:
• grow
g
• give positive / negative reaction
• If testing new lots/batches of media –
inoculate old and new lots in one test and
compare the performance of the two lots side by
side 12
Blood agar
g
Test for growth and haemolysis with
Streptococcus
p Pyogenes
y g
Stock Culture
• Laboratory should maintain stock culture
• Source:
1 Commercial
1. C i l
2 Clinical specimens
2.
3. Proficiency testing samples
4. Reference laboratories
5. Official Culture Collection (American Type
Culture Collection (ATCC))
14
Stock Strain
• Appropriate ATCC control strains
• Suggested by CLSI
Strain ATCC
Escherichia coli 25922
P
Pseudomonas
d aeruginosa
i 27853
Staphylococcus aureus 25923
Streptococcus pneumoniae 49619
Enterococcus faecalis 29212
Haemophilus influenzae 49247
P
Pneumoniae
i kl b i ll
klebsiella 70063
15
Media-Antimicrobial Susceptibility
• Verified with approved reference organism
CLSI (NCCLS)-M2-A7
• Variables that can affect results accuracy:
• Antibiotic
t b ot c pote
potency
cy
• Agar depth (Kirby-Bauer Test = Disc
Diffusion Test)
• pH
• Inoculum
• Incubation time
• Incubation
I b ti ttemperature
t
• Moisture
• CO2 concentration 16
Frequency
• Everyday of testing
• Weekly y – Unless laboratory
y adopted
p in house
verified procedure that
1 the appropriate QC strains were tested for a
1.
minimum of 30 consecutive days
2 have demonstrated acceptable
2.
performance
• Use
U specific ifi strains
t i
1. Haemophilus influenzae – gram negative
coccobacilli
2. Nesisseria gonorrhoeae – gram negative
diplococcus 17
Weekly – Unless
laboratory adopted in
house verified
procedure that
1. the appropriate QC
strains were tested
for a minimum of 30
consecutive days
2. have demonstrated
acceptable
t bl
18
19
20
21
22
23
24
Reagent – Daily
In-use reagent vial is refrigerated at night but
usually left at room temperature during the day
• has the opportunity to degrade while in use
• should
h ld bbe tested
t t d each h day
d off use with
ith b
both
th
positive and negative controls
Reagent Controls
C t l
Catalase C t l
Catalase (+) vs. C
Catalase
t l ((-)) Staph.
St h
Coaglulase Coagulase (+) vs. Coagulase (-) Staph.
Cytochrome C Oxidase (+)
Oxidase Neisseria, Moraxella, Campylobacter
and Pasteurella species vs. (-)
25
Reagent – Weekly
Reagents that are documented to have
consistent and dependable
p results may
y be
tested less frequently
Reagent Controls
Gram stain Gram positive and negative
Acid fast stain Ziehl–Neelsen stain for TB
Bacitracin Cyclic peptides by Bacillus subtilis
Optochin Streptococcus pneumoniae
ONPG (ortho- Lactose fermenter – Escherichia
Nitrophenyl-β-
p y β coli,, Klebsiella spp,
pp, Enterobacter
galactoside) spp 26
Reagent – Weekly
Reagents that are documented to have
consistent and dependable
p results may
y be
tested less frequently
With each
h new vial
i l off reagentt
27
28
29
Reagent – Stain
30
Reagent – Monthly and As Required
• Typing sera should be tested with
• Each new lot number and
• Each month of use
• Examples
• Salmonella spp
spp.
• Escherichia Coli
• Vibrio Cholerae
31
Equipment
• Biological safety cabinets
• Autoclave
• Oven
• Incubator
• Refrigerator
• Freezer
• Water bath
• Mi
Microscope
• Anaerobic jar
• Centrifuge
• Thermometer
• MALDI-TOF 32
Autoclave
33
35
39
C lib ti failures
Calibration f il
1. Improper application of the calibrator –
laboratories can assess potential user error by
reapplying and re-analyzing the calibrator
2. Improper preparation of the calibration
3 P
3. Problems
bl crop up with
i h the
h
• Matrix
• Reagents
• Target
• Instrument
40
Total Testing Process and Associated
Laboratoryy Errors (%)
41
Post analytical
Post-analytical
• There is subjective element(s) in the report
content
• The interpretative reports are opinion of the
reporting pathologists
• Some diagnoses require the combined input
off a cytologist
t l i t andd histopathologist
hi t th l i t
• There are a variety of reasons why clinical
appearances, cytology, biopsy and excision
results may appear discrepant
• Multi-disciplinary team (MDT) meetings can
often resolve p
perceived discrepancies
p
44
Highlights for Histopathology
1. Sample ‘Chain of Custody’
• A robust ‘chain of custody’ across the
specimen pathway
• These involve cross-checking of a minimum
of three patient identifiers at each stage,
to typically include name, hospital number
and accession number
• Slide labels will include patient surname in
addition to the accession number
45
46
4. Slide Staining
• Stains
• Haematoxylin and Eosin
• Special stains and immunohistochemical
stains
t i
• Prepare and use according to
manufacturer’s instructions with appropriate
regard to both positive and negative
control slides
• Internal technical q qualityy checks should
be carried out routinely including quality of
stainingg and q
quality
y of p
preparation
p
47
5 Mi
5. Microscopici diagnosis
di i depends
d d
crucially on quality control
• Methods used for quality assessment have
to:
1. Incorporate a process of continuous
dialogue within the laboratory
2. Improve individual histopathology
reporting accuracy
• Internal quality control of reporting can be
monitored by a variety of methods include:
48
1. Performance evaluations
2 Periodic audit of histopathology outcomes
2.
3. Non-conformities monitoring
4. Slides review by MDT (Multi-Disciplinary
Team)
5. Histopathology
p gy detection and reporting
p g
rates monitoring
6 Correlation of cytology with
6.
clinical/histological outcome
49
1. Positive
• Reveal specific binding and true positive
results
• Specimens contain biomarker of interest
with specific characteristics in known
location
• Histomorphology and cytomorphology can
be visualized by fluorescent or
chromogenic stain
2. Negative
• Reveal non-specific
p bindingg and false
positive results
• Specimens do not express the biomarker
or protein of interest 52
3 N
3. No P
Primary
i Antibody
A tib d
• Reveal non-specific binding and false
positive results
• Samples are incubated with NO primary y
antibody but ONLY the antibody diluent
• To determine if secondary antibody binds
non-specifically to cellular components
that
a do noto co
contain
a theebbiomarker
o a e o or p
protein
oe
of interest
53
54
5. Isotype
yp – Particularlyy crucial for monoclonal
antibodies
• Antibody
• of the same isotype (IgG2 / IgM / IgY),
clonality conjugate,
clonality, conjugate and host species
as the primary antibody
• can target a protein or biomaker that is
not present in the sample
• To
T validate
lid t the
th observed
b d staining
t i i is i
1. specific to the protein of interest
2. not produced by non-specific
interactions of the antibody y with the
other components of the tissue samples 55
6. Adsorption
• To demonstrate that an antibody y is binding
g
specifically to the protein of interest
• Pre
Pre-incubate
incubate antibodies with the
immunogen / purified antigen
• inactivate the antibody
• the tissue show little or no staining
56
Regulation in Blood Bank
• In the years before the HIV epidemic, blood
banks were perceived as organizations that
provided a community service
• Increased occurrence of HIV and increased
public scrutiny resulted in stricter FDA
regulations
• FDA regulatory oversight has resulted in an
increased effort to provide a safe, high-
quality product at low cost
57
N
New Kit S
Sensitivity
iti it & S
Specificity
ifi it
61
Utilization Review
Agreed by Joint Commission – Assess
1 Blood ordering and transfusion
1. f practices
of medical staff
2. Cross match / Transfusion ratio
• Number of units cross matched divided by y
the actual number transfused
• Used as an indicator that too much blood
is being requested to be on hold
• Could result in high outdate or wastage
3. Number of autologous transfusions
62
4. Number of emergency releases
5. Justification of transfusion
6. Define and audit transfusion criteria
• Hematocrit < 24%
• Hemoglobin < 8 g/dL
• Symptoms due to anemia
• Recent estimated blood loss of > 10% of
total blood volume
7. Record of actions taken for physicians who
violated the transfusion criteria
63
Haematology
gy
Daily Internal Quality Control
Routine
1. Start up status and background count check
• Monitor for acceptable
p background
g
• If required take corrective action
2. QC monitoring g using
g multilevel controls
• Commercial – Low, normal, high level –
Monitor L-J chart and control rules
• Retained patient samples – Monitor %CV
• Dailyy routine patient
p samples
p – Monitor
moving averages
• If required
q take corrective action
64
Establish Commercial Multilevel Controls
Range (Mean and SD) – Commercial
• Far before adoption of new lot control material
material,
it should be analyzed in parallel with current lot
• This may be accomplished by running the new
controls twice a day for five days
• Determine
D t i mean and d SD off th
the tten runs as
tentative mean and SD
• Verify validity of the tentative mean and SD by
collecting more data
• Establish a running mean and SD for the lot
65
67
68
Multi-QC Rules – Rejection
j Rule
Bias
69
Imprecision
70
Other Internal Quality
y Control Rules
71
Head Tail
• Ensure g
good drying
y g of the smear
• Do not apply overused stain
• Maintain required buffer pH
• Compile with staining time
72
Highlights
g g ts for
o Flow
o Cyto
Cytometry
et y
1. Clinical flow cytometry has evolved from two-
parameter quantitative assessment of
peripheral blood lymphocytes to six-
parameter qualitative evaluation of bone
marrow for hematopathology
2. Leukemia
eu e a a and
d lymphoma
y p o a immuno-u o
phenotyping represent an extremely
important
p complement
p to morphology
p gy in the
diagnosis and monitoring of hematopoietic
malignancies
g
3. The complexity of five- and six-parameter
analyses
y and corresponding
p g data
interpretation rely on: 73
Laboratory
y Developed
p Genetic Test
1. Conduct a review of available scientific studies
and pertinent references
2. Define appropriate patient populations for which
the test should be performed
3 Select
3. S l t the
th appropriate
i t test
t t methodology
th d l f the
for th
disease or condition being evaluated
4 Establish analytic performance specifications
4.
and determine quality-control procedures using
the appropriate number
number, type,
type and variety of
samples
5. Ensure that test results can be interpreted for an
individual patient or family
6. Ensure that the limitations of the test are well
defined and reported 76
Specimen
p Selection for Validation
1. Evaluate specimen adequacy of the
specimens for the prevalence of the disease
and the mutations or variants
2 Specimen types may include blood
2. blood, buccal
swabs, dried blood spots, fresh or frozen tissue,
paraffin-embedded tissue,
tissue or prenatal
specimens
3 For a multiplex genetic test,
3. test all the mutations
or variants to be detected should be included in
performance establishment
4. For rare genotypes/mutations, alternative
samples
l acceptable
t bl
77
78
Analytical
y Testing
g Phase
CLIA requires laboratories having procedures to:
1 monitor and minimize contamination during
1.
the testing process
2 ensure a unidirectional
2. idi ti l workflow
kfl f
for
amplification procedures that are not contained
i closed
in l d systems
t (42 CFR §493.1101)
§493 1101)
80
1. Commercial
1
2. In-house – remainder patient specimens or
synthesized
3. Rare genetic diseases – positive controls are
h d tto obtain
hard bt i
4. CLSI guideline Verification and Validation of
Multiplex Nucleic Acid Assays (MM-17A; CLSI,
2008)
5. CDC’s Genetic Testing Reference Materials
Coordination Program (GeT-RM)
6. Positive and negative controls should be
tested each time p
patient samples
p are assayed
y
81
Quantitative Assays
• Quality control statistics are performed
monthly to
1 define analytic imprecision and
1.
2. monitor trends over time
• The laboratory must use statistical methods
1 Calculate SD and CV monthly t
1.
2. Evaluate variance in numeric QC data
• Multi-QC rules are applied if appropriate
82
Multi-QC Rules
83
Qualitative Assays
y
1. Cut-off value distinguishes positive from
negative result
• Establish the cut-off value initially
• Verify with every lot change or at least
every 6 months
2 Limit
2. i i off detection
d i (LoD)
( D) differentiate
diff i a
positive from negative result
• Establish the LoD when the test is initially
placed in service
• Verify with every change in lot (e.g. new
master mix),
), instrument maintenance,, or
at least every six months thereafter 84
• A low-positive
p control that is close to the
LoD is required to:
1 Identify weak-positive
1. weak positive patient sample
2. Verify reference material was prepared in
appropriate matrix
85
92
Total Testing Process and Associated
Laboratory Errors
93
Periodic / Preventive
Maintenance
Follow Manufacturer
Manufacturer’s s
Instruction of Speed and Time
(w.r.t. different blood collection
tube)
94
Upgrade
D b
Debug
Information
gy (IT)
technology ( )
department /
Outsourced IT
service provider
95
96
Pipette Performance Verification
Gravimetric Method
Three /
Four
Volumes
97
98
99
AssureTraceability
Measurement Traceability
• Result of a measurement or the value of a
standard whereby it can be related to stated
references
1 Traceability
1.
2. Standardization
• Usually national or international standards
• Th
Through h an unbroken
b k chain
h i off comparisons
i
all having stated uncertainties
100
Specimen
p Traceability
y
• Important concept in quality assurance
programmes
p g – Chain of custody y
• An effective qualitative assurance procedure
should allow an audit trail to be followed back
from the laboratory report / result
• This should allow access to a complete
documented history
• FROM receipt of the specimen TO the
issue of the report
• Major Objectives:
• Identify the root cause of error
• Apply remedial measures
101
102
103
104
105
107
109
110
EQA
Analytical
111
5 Di
5. Dispatch
t h performance
f reports
t (in
(i
cumulative manner) to the participating
l b t i after
laboratories ft every time
ti reportt
submission
6. Assess the laboratory periodic
performance on regular
p g basis e.g.
g
frequency of late submission, wrong sample
identification, bias, imprecision
p etc..
7. Issue periodic reports with statistical
analysis to quantify the laboratory
performance e.g. quarter / biannual / annual
+/- professional advice / comment
+/
114
PT Provider
PT Organization / Laboratory
Provider
115
Laboratory
116
To receive
T i FULL VALUE off the th proficiency
fi i testing
t ti
program / external quality assurance program,
l b t
laboratory shall
h ll consider
id allll iinformation
f ti obtained
bt i d
from the proficiency testing program directed
t
towardd IMPROVEMENT.
IMPROVEMENT
117
118
Limitations of Proficiency Testing
Program
1. Results can be affected by variables not
related to patient samples e.g. Matrix Effect
2. Cannot detect all problems incurred in the
laboratory
y e.g.
g LIS
3. Cannot detect ALL problems with pre and
post analytical procedures
• Pre-analytical – Bed side order etc..
• Post-analytical – Delta check, phone
clinically
c ca y ccritical
t ca result
esu t etc
etc..
119
4. Reports
p are available 2 – 4 weeks after result
submission – cannot reflect real time
deficiency
y
5. Might not cover all laboratory tests –
laboratory cannot apply those tests for
accreditation
120
Rechecking
g / Retesting
g
Participating
Initiating Laborator
Laboratory
Laboratory
Sample
S l
Exchange
Result
Exchange
Report
p
Exchange
121
Procedure
1. One testing laboratory initiates
2 Other testing laboratory(ies) participates
2.
3. Testing samples
• Most of the time are patient samples
• Can also use proficiency testing samples
4. Evaluate mutual performance with simple
statistical analysis
y
5. Effective feedback
6 Primarily used to assess analytical
6.
performance of analyte(s) not available by
proficiency testing providers
122
Limitations of Rechecking / Retesting
1. Not blinded
2. Rechecking / retesting cannot detect all
problems incurred in the laboratory
p y e.g.
g LIS
3. Rechecking / retesting cannot detect ALL
problems with pre and post analytical
procedures
• Pre-analytical
P l ti l – Bed
B d side
id order
d etc..
t
• Post-analytical
y – Delta check, p
phone clinically
y
critical result etc..
123
Proficiency
y Testing
g vs. Rechecking/Retesting
g g
Rechecking /
Characteristic Proficiency Testing
Retesting
Inter-laboratory
Yes Yes
Comparison
Simulated Sample Yes No
Patient-matrix Sample Yes / No Yes
Analysis Evaluated Many Few
Laboratoryy Manpower
p
L
Less M
More
Needed
Monetary Spent More Less
Report Turn Around Time Short Long
Comprehensive both
Performance Evaluation Minimal
accuracy & precision 124
Choices of External Quality Assurance
Program
1. Scientific Validity y
• Reliable testing materials
• Valid assigned
g values
• Valid statistical analysis
2. Reliabilityy
• Confidentiality
• Frequent
q distributions
• Short cycle
• Intelligible
g reports
p
• Strictly execution of the schedule
• Scoring g system
y
• Participant number 125
130
The projections for future are:
1. Use third-party control materials.
2. Implement IQC based on patient samples
when no stable control material is available.
3. Use analytical performance specifications
based on biological variation.
variation
4. Encourage external quality assessment
organizers
i t provide
to id commutable
t bl controls
t l
5. Push laboratories to have freezers to maintain
external quality assessment controls
at −80 °C.
131
Appropriate SD/ CV
Low false rejection
and acceptance rate
132
False Acceptance
• Type II error
• The system accepts a wrong result
• High false acceptance rate
• LJ chart
h t with
ith iinappropriately
i t l ‘LARGE’ SD /
CV
• Present with ‘GOOD’ analytical performance
BUT actually not
False Rejection
• Type I error
• The system rejects a valid result
• Hi h false
High f l rejection
j ti rate
t
133
???
Mean SD Appropriate
ABC
Co.
QC1
136
ESTABLISH THIRD PARTY QC INTERVAL
Ensure [QUALITY] related activities are FIT for
[CLINICAL APPLICATION]
Analytical
A l ti l SD and
d CV are nott an
"isolated" figure but clinically related to
analyte
l t dependent
d d t within-subject
ithi bj t or
intra-individual biological variation
To manage
g imprecision,
p , the laboratory
y
can involve in two ways [PROACTIVELY] &
[CLINICALLY]
Imprecision
Analytical
y ((CVa)) Biological
Uncertainty of
Measurement Biological Variation
(Intra and Inter-individual)
Maximum Analytical
Imprecision
p Allowable
Analytical Goal /
Level of Clinical Fitness
(Analyte Specific) 140
Relationship between Biological
Variation (CVi) and Analytical Goal
Analytical Goal / Level Allowable
of Clinical Fitness Imprecision Criteria
Optimum ¼ of CVi
D i bl
Desirable ½ off CVi
Minimum ¾ of CVi
141
Bilirubin – Conjugated
Serum 36.8 43.2
(Direct Bilirubin)
CVi
CVg 142
Clinically Applicable LJ Control Glucose
Chart
Intra-individual
Intra individual biological variation (CVi 4.5%)
4 5%)
Level of Clinical
Allowable
Fit
Fitness / Analytical
A l ti l
Imprecision Criteria
Goal
Optimum < 1.13%
1 13%
Desirable < 2.25%
Mi i
Minimum < 3.38%
3 38%
Three levels,
levels
choose
which one?
143
144
Clinically Applicable LJ Control Glucose
Chart
Intra-individual
Intra individual biological variation (CVi 4.5%)
4 5%)
Level of Clinical
Allowable
Fit
Fitness / Analytical
A l ti l
Imprecision Criteria
Goal
Optimum < 1.13%
1 13%
Desirable < 2.25%
Mi i
Minimum < 3.38%
3 38%
Glucose D-Bilirubin
148
Clinically Applicable LJ Control
Chart
Clinically Applicable
SD / CV Internal
I t l QC
Rangeg
149
151
153
DBIL 2
154
P f
Performance R
Review
i
TEa
Analyte CVa(%) Grade
(ALP)
Na 2 2 1 Bad
DBIL 2 20 10 Good
Remark: Sigma
S >=6 Good,
G <3 Bad
155
P f
Performance R
Review
i
<3σ
156
Relate Assay Required Performance
Clinically
157
P f
Performance R
Review
i
Lab A 1 20 20 Good
Lab B 3 20 7 Good
Remark: Sigma
S >=6 Good,
G <3 Bad
158
P f
Performance R
Review
i
Total No.
CVa of Daily
DBIL Grade Daily QC Resource
(%) QC
Analysis
Thrice & 3
Lab A 1 20 Good
levels@ 9
Once and 2
Lab B 3 7 Good
l
levels
l 2
Quality
Q y Achievements
160
161
162
Set-up Clear Achievable and Glucose
Realistic Target
Level of Clinical Allowable
Fitness / Imprecision
y
Analytical Goal Criteria
TEa (ALP)
Optimum < 1.13%
Desirable < 2.25% 8%
Minimum < 3.38%
TEa / CV
Is it
realistic?
164
Analyte(s) Need Attention
165
Review Peer’s
Peer s Performance
CV (%)
Bench
Mark
166
Review Peer’s
Peer s Performance
Our Laboratory
Sigma Value: 3
Peers Sigma =
ALP / CV
C
= 2% / 1% = 2
167
Un-necessaryy
troubleshooting
Un-necessary re-run
QC
168
Regular
electrode
l t d
replacement
169
170
P f
Professional
i l Recognitions
R iti
P f
Professional
i l Recognitions
R iti
Phosphate
P t i
Protein
Lactate
Urea
Glucose
LOCAL RECORD
173
175
176
Sigma
g Performance of > 50
Glucometers
(Low level internal quality control
materials
t i l off glucose
l llevell ~ 3.5
35
mmol/L) (n > 80,000)
Glucometers
177
Sigma Value
Sigma
g Performance of > 50
Glucometers
(High level internal quality control
materials
t i l off glucose
l llevell ~ 16.0
16 0
mmol/L) (n: > 80,000)
Glucometers
178
Sigma Value
Cli i ll Applicable
Clinically A li bl LJ Control
C t l Chart
Ch t