1.1 Characteristics
1.1.1 High Voltage Generating Device
a. Adopt high frequency contra-variant technology that produces stable high voltage output, bringing about
good diagnostic results.
b. Radiographic parameter adjustment of X-ray apparatus employs the three parameters of kV, mA and mAs,
which are chosen by pressing corresponding keys. According to the capacity limit of X-ray tube, the
equipment software determines interlocking protection conditions.
c. The adjustment of kV and mA parameters of X-ray apparatus is carried out by keys.
d. A television system is equipped with X-ray apparatus, which can conduct ABC radioscopy.
e. Work status (mode) shift of X-ray apparatus can be realized by pressing the keys on the control panel:
radiography, manual radioscopy and ABC radioscopy.
f. The exposure of X-ray apparatus employs two gears mechanism: i.e. gear I of manual brake for
preparation and gear II for exposure.
g. The radioscopy exposure of X-ray apparatus employs one gear mechanism: i.e. manual gear II, foot brake
and radioscopy button for exposure.
h. X-ray apparatus employs a computer to monitor various breakdowns or failures. Once a breakdown
occurs, the computer will cut off all output and display the code of breakdown.
1.1.2 Mechanical Devices
X-ray generating device and image intensifier of the X-ray apparatus are fixed at two ends on a C-type
arm separately. SID (the distance between the focus point and the receiving surface of image intensifier) is
fixed (not adjustable).
C-type arm can do the motion such as rotation, swinging, slide, ascend and descend, which ensure the
free shooting of X-ray in three dimensions.
1.1.3 Television System
a. Corollary television system of X-ray apparatus can do normal radioscopy (X-ray examination), which is
easy in observation, diagnosis, operation and teaching.
b. Corollary television system of X-ray apparatus can conduct ABC radioscopy.
1.2 Operation Range
X-ray apparatus can be employed for X-ray diagnosis (both radioscopy and radiography) in ward, bone
surgery and operation room.
This X-ray machine must not be used in environment with mixed inflammable
and explosive gases.
Danger
1.3 Operation Conditions
1.3.1 Environment Conditions
a. Environment temperature: +10℃~+ 40℃
b. Relative humidity: 30%~75%
c. Atmospheric pressure: 700hPa~1060hPa
1.3.2 Power Conditions
a. The power is featured by single phase 220V and 50Hz sine wave electric voltage. The voltage falls into
the range of 220V±22V with the frequency range of 50±1Hz.
b. The internal power resistance shall be no more than 0.5Ωwith power capacity no less than 4kVA.
c. The internal resistance of leakage plunger shall be no more than 0.5Ωwith power capacity no less than
4kVA.
1
Chapter 2 Operation Mechanism and Structural Characteristics
1
10
2
11 19
3 12
20
13 21
4
14
5
15
22
6
16 23
7 17
8 18
4. C-type arm slide lock handlebar 12. Vertical lock handlebar 20. Propelling handlebar
6. Base support 14. Swing lock handlebar 22. Monitor carriage door
2
In addition, hand brake, foot brake and image system remote control box are standard parts not
shown in the Figure.
a. System Components
·X-ray tube (including beam limiting device);
·Mobile C-type arm carriage (including control system);
·Television system of X-ray image intensifier;
·Monitor carriage (optional).
b. Electric components of high-voltage generating device
· Power control
· High-voltage (kV) control circuit
· Filament (mA) control circuit
· High-voltage (combination) oil tank
· Display and key control
· Motor control
2.1.2 Operation Mechanism
The operation mechanism of X-ray apparatus is illustrated in Figure 2.2.
a. High-voltage circuit
It adopts high-frequency contra-variant technology. The kV-SET and kV-SAMP high-voltage sampling
signal set by CPU form as pulse width modulation (PWM) control signal after error magnification. Then it
becomes the driving signal of the circuit.
Major contra-variant device adopts intellectual power module (IPM), and forms a complete circuit via
high-voltage transformer.
Tube voltage sampling is conducted at two sides, which is obtained after high-voltage resistance
division.
VM4 VM5
High-voltage transformer
VM1 CM3
CM1
VM3 VM6
KV-SET
VFC
KV-SAMP Trigger
b. Filament Circuit
It adopts variable frequency technology. The mA value (mA-SET) and tube current sampling signal
(mA-SAMP) set by CPU form as pulse width modulation (PWM) control signal after error magnification.
Then it becomes the driving signal of the power circuit.
Filament contra-variant device adopts MOS tube, and forms a complete circuit via filament transformer.
Tube current sampling is two-sided.
3
V18 V21
Filament transformer
V1
C5
V19 V20
MA-SET
VFC
MA-SAMP
Trigger
1 2 3 4
5
7 10
8 9
14 15
13
11 6 12
Front Back
In addition:
The hoist motor and motor control board are installed on the base support, the beam limiting device at
the back of electric supporting board and the image control box of the intensifier installed on the column.
Displaying board and keys are on the panel.
4
2.2.2 Each Circuit Board and the Function of Each Component
Corresponding
Figure code Circuit board Function
socket
HMC3601 System connection sketch
HMC3602 Switch board Power control PS
HMC3603 8031 System
HMC3604 CPU board KV Control PC
HMC3605 mA、 Filament control
HMC3606 Sampling board Sampling following and changing-over P
HMC3607 Driving board IPM driving PI
HMC3608 Main contra-variant circuit
HMC3609 Filament board Filament driving PF
HMC3610 Combined high-voltage generator
Displaying
HMC3611 8279 Display and key control PD
board
HMC3612 Panel Keys and display PK
Motor control Control of hoist motor and beam limiting
HMC3613 PM
board device motor
Main contra-variant power testing and
HMC3614 Discharge board PP
discharge
5
f. Computer Testing
X-ray apparatus employs 8031 monoboard computer as the control device that monitors seven
parameters when the equipment is under loaded operation.
When the tested results exceed the permitted range, it will cut off the operation of the main machine and
display corresponding breakdown code. (See Table 5.1).
g. Breakdown Debugging
Breakdowns are divided into serious ones (hardware) and general faults (software). For general faults,
Press any key you can resume to general status. For serious breakdowns, operator shall turn off the power to
check. Do not re-start the equipment until the breakdown or failure has been debugged, then the equipment
enters normal status.
6
Chapter 3 Technical Characteristics
40~59 1~250
60~79 1~200
80~89 1~160
90~110 1~125
7
Table 3.2 Adjusting Range of mAs
Gears of mAs Permitted kV
1.0,1.25,1.6,2.0,2.5,3.2,4,5,6.3,8,10,12.5,16, 40~110
20,25,32,40,50,63,80,100,125
160 40~89
200 40~79
250 40~59
The classification of mA and mAs is in line with R’10 number groups of GB9706.3-2000 Standard.
8
c. Slide along the rail 115o (+90 o ~-25 o ), the propelling force shall be no less than 100N.
d. Horizontal extension ≥ 200mm, the propelling force shall be no more than 100N.
e. Hoist range ≥ 400mm, hoist motion shall be stable and reliable.
f. Swing around the main axis ≥ ±12.5 o , with good self-alignment.
g. Swing around horizontal axis ≥ ±180 o , with good self-alignment.
h. Motion in all directions of the equipment is manually controlled except the hoist being driven by
electricity.
i. Carriage moves freely, the initiating propelling force on the handlebar shall be no more than 100N.
j. The carriage is controlled with mechanic lock with locking force ≥ 150N.
9
Heating and Cooling Curve: (See the Figure)
1.Heat capacity
3.Cooling curve
2.Heating curve
4.Time (min.)
10
3.2.14 Parameter of the Control Switch
11
Chapter 4 Installation and Debugging
Please carefully read through this chapter before installation and debugging.
The equipment has been debugged before leaving the factory. It is strongly
recommended that operators should install and debug the equipment
according to this Instruction Manual.
WARNING If there are any problems beyond the capabilities of the operator, please stop
installation and debugging immediately and contact the manufacturer in
time.
4.1 Installation Conditions and Technical Requirements
4.1.1 Installation Conditions
a. Power
The power at the installation site is 50Hz and 220V, with the capacity no less than 4 kVA. Power
internal resistance shall be no more than 0.5Ω.
b. Installation Site
Installation site shall have enough space that C-type arm will not be subject to collision or knock during
operation. In addition, the installation site shall be clean, dry and away from disturbing sources.
c. Required Tools
Socket head wrench: 1 set Screw driver: 1 set Multi-meter: 1 set
100Mhz memory oscilloscope: 1 set Electric iron: 1 set
4.1.2 Technical Requirements
There shall be two people for installation and debugging, at least one of them must be the recognized
technician who takes charge of the installation, debugging, operation and maintenance of the equipment.
4.2 Installation Procedures and Precautions
The removed small components during installation shall be put onto a tray
(There will be no such reminding message in the Manual). As soon as the
installation and debugging is finished, check if there is any surplus or
WARNING shortage of small components.
As this apparatus is a precision instrument, after carefully remove the wood packaging box please check
carefully if all packages are intact or damaged. If there are any abnormal phenomenon, please contact the
manufacturer or local sales representatives immediately.
Put the main machine carriage and monitor carriage well into their installation place. If the humidity is
relatively high or the carriages are quite damp, the plastic cover shall be removed, employ clean, dry and soft
gauze to cleanse them. Wait the carriages to be completely dry and adapt the surrounding environment.
Put the carton packaged components such as image intensifier, television camera, television monitor (1
set) and X-ray tube into installation places with caution and care.
During the following installations, installing workers shall carefully fill out
installation and check record, and send them together with installation report
NOTICE back to the manufacturer for the need of establishing files.
4.2.1 Installing of CCD
a . Push the main machine carriage to appropriate position, employ the brake to secure the two foot
wheel. Release the C-type arm sliding lock handlebar and move C-type arm to horizontal position.
b . Remove the carton package for image intensifier. Check if there are four M6 installation nails
(external hexagonal) in the intensifier package.
c.Take out the image intensifier, remove the plastic cover. Use a wrench to install the intensifier onto
the C-type arm and make sure the installation is secure.
d.Remove the rear cover of the image intensifier, connect DB9 and DB15 plugs. Then fix the rear cover
well.
4.2.2 Installing X-ray Tube
a.Push the main machine carriage to appropriate position, use the brake to secure the two foot wheels.
Release the C-type arm sliding lock handlebar and move C-type arm to vertical position.
b.Remove the carton package for X-ray tube with care. Check if there are four M6 installation nails
(external hexagonal) in the X-ray tube package. If there is a film clip, place it in a proper place for final
installation
12
c.Place a piece of soft setting cloth on the side of C-type arm facing the ground. Put a pair of gloves
and take out the X-ray tube with caution and care. Remove the plastic cover, place the tube on the cloth.
d .Use small cross screw driver to remove the four screws fixing the upper cover. Then remove the
upper cover and put it on the cloth with the tube nose and photochopper disclosed. Use long-handle cross
screw driver to remove the four screws fixing the bottom cover. Then remove the bottom cover and put it on
the cloth.
e.Put gloves on your hand and take out X-ray tube with caution and care. Observe if there is any oil
stain. If there is heavy oil stain, carefully identify the cause, use a piece of gauze to clean it. Fasten the
installation nail or use adhesive material. If there still serious oil stain, please contact the manufacturer or
local sales representative.
f.Use M5 socket spanner to install the tube (without cover or shell) onto C-type arm. Make sure the
installation is secure.
g. Relax C-type rotation lock handle and put C-type arm to horizontal position and lock it. Use cross
screw driver to fix the lead terminal onto the terminal plate on tube cover in accordance with the marked
number.
Read through carefully the relevant circuit theory and circuits in the
instruction manual before debugging.
WARNING
During the following debugging process, the debugging worker shall carefully
fill out installation report and send it together with installation and check
NOTICE records to the manufacturer for file.
13
After procedures 1 and 2, the main machine shall have a normal load-free operation without any
breakdown or failure.
After installation, X-ray apparatus shall have the inherent filtration no less
than 2. mmAl.
2.6mmAl.
NOTICE
14
Chapter 5 Operation Method
This apparatus is composed of three components: the main machine, image system and monitor.
Main machine: generating X-ray
Image system: receiving X-ray and turn the invisible X-ray into visible light.
Monitor: transmitting the visible lights to medical staff through fluorescent screen for disease diagnosis.
5.1 Preparation and Check before Operation
5.1.1 Familiarizing Operation Panel
The explanations for the keys on operation panel of HMC—36 Mobile C-Arm X-ray Machines are as the
followings.
After press the switch, if the system self-check is
Power indicator
OK, the light becomes bright.
Power-on key Press the key and the main machine gets electricity
Power-off key Press the key and the main machine turns off
C -type arm
Press this key, C-type arm ascends as a whole
ascend key
C-type arm
Press this key, the whole C-type arm descends
descend key
15
The light-up of big focus point light indicates the
Big focus big filament of the tube under operation, generally
indicator under radiography condition.
Small focus The light-up of small focus point light indicates the
indicator small filament of the tube under
KV Digital
Displaying the pre-set kV value.
display
KV adjusting
It can reduce and increase the pre-set kV value.
key
mA Digital
Display the pre-set mA value.
display
Radiography
mAs adjusting It can reduce and increase the pre-set mAs value.
key
Beam limiting Press this key, the circularity direction of the beam
device key A limiting device opens.
Beam limiting Press this key, the circularity direction of the beam
device key B limiting device closes.
Beam limiting Press this key, the beam limiting device has levo
device key C rotation.
16
Beam limiting Press this key, the beam limiting device has dextro
device key D rotation.
Collimator control This button can control the open and close of
button the collimator. The groups are on the cross
17
this button can control 3 levels of variable
Image intensifier fields. The normal field is 9 inches, which is
field control button indicated by an indicator.
Image storage button this button can control the image storage.
Black and white this button can control the monitor show the
reverse button image in reverse
B. Check
Make sure the power supply connected to the TV system has protective
earth to avoid damage to personnel and equipment, and to avoid the
NOTICE degradation of the image quality.
Before flouroscopy, please make sure the TV system has been correctly
connected and powered on. At this time, the monitor screen should display
NOTICE blanking circle.
storee image
C. Method to stor imagess and:
1 Make sure that all the image control cables are correctly and firmly connected, and make sure that C-arm
principal machine is working in a normal condition.
2 Power on the panel power, only the indicators of 9’ are on, and the monitor displays blanking circle.
3 Press the Flouroscopy button (yellow), the monitor displays real-time image; when the Flouroscopy
button is released, the last image remains on the monitor and can be flipped horizontally and vertically.
4 With the storage choose button choose the position of image which will be stored, and then press the
18
storage button. At this time, the monitor screen will instantly displays numbers (01~25), indicating the storage
serial number of the image. Thus, the storage of the image is completed. If more images are to be stored, please
repeat this step.
D. Load Image
1 Make sure the image control system is in stand-by state, i.e. no real-time flouroscopy image is displayed
on the monitor.
2 With the storage choose button look over the image in the storage, and screen shows corresponding
number(01~25) at the same time.
3 When the Power-off button is pressed, the stored images in the memory will be lost. Please pay attention
to it.
F. Inspection
Before radioscopy, please identify that the television system has correctly
NOTICE connected and contacted the power.
Operate the mechanic handle in line with the arrow direction in the
Figure meaning locking the apparatus.
Operate the mechanic handle in line with the opposite direction of the
marked arrow in the Figure meaning releasing the lock.
coefficient: The elastic limit coefficient of this apparatus > 2.2, forced damage coefficient > 4.0,
Safety coefficient
which are in line with the national standard of GB9706.14.
5.2.1 Propelling the Carriage
There are three foot wheels (7) on the carriage of X-ray apparatus. Propel the push operation handle
(15), enabling X-ray apparatus move smoothly on the ground.
wheelss are on brake, please do not propel the carriage
In case the two rear wheel
NOTICE body otherwise mechanical damage may occur.
19
5.2.2 Braking
As soon as the carriage body moves to the required position, step on the brake (8) to secure the two
wheels making the X-ray apparatus having a safe and sound stop.
5.2.3 Hoist
A motor driven hoist column (18) is equipped in the main machine (17) in order to
accommodate to different bed height. The hoist range is no less than 400 mm. In case
the motor hoisting is out of control, please press the "Emergency Stop" switch so as to
prevent any accidents.
WARNING ● When operating C-type arm (2), observe the location of the objects in the
room so as to prevent the collision against them. When moving or locking the
C-type arm in various directions, please bear in mind that you must hold the
C-type arm handlebar well, do not relax until it is locked so as to avoid the
damages to both the apparatus and people.
● When C-type arm (2) is put at horizontal position for horizontal shooting, the
(24)) with an appropriate range
arm (16) shall be placed on the C-type bracket (24
of 0 ~15
o 15 . Please do not put it at the +90 or -25 o extreme positions.
o o
20
NOTICE ● When propelling X-ray apparatus, please notice the following
following:
● Coil the cable and then put it on cable stand;
● Put the hoist column (18) at the lowest position;
● Vertical expansion bracket (11) is withdrawn to the shortest position;
● C-type arm (2) in vertical 0o position;
● Vertical expansion bracket (11) in forward position (not doing ±12.5
12.5o swing)
swing);;
(4),, (10), (12) and (14);
● Secure all locking handles (4)
● Relax brake (3) enabling brake column (4) hoisting and leaving the ground;
● Propel the push handlebar (7) enabling X-ray apparatus move smoothly.
For the code numbers in the above brackets, please refer to 2.1.1 Structural Sketch.
5.3 Operation Procedures
5.3.1 Switching On
Take the power line from the apparatus, plug the terminal into the appropriate socket in the ward. Press
the power-on key "I", and then it contacts with electricity. Meanwhile, the computer is powered and starts
self-test. It displays “HELLO” after the self-testing results being OK. At the same time, it conducts other
checks. As soon as it confirms everything is OK, the green colour exposure preparation light turns on
indicating everything is normal and the apparatus is ready for operation.
The apparatus has three working status or modes: i.e. manual radioscopy, ABC automatic radioscopy
and radiography. Manual radioscopy is the normal working condition or status.
5.3.2 Starting Television System
Start the television system according to the Television System Instruction Manual, adjust the brightness
comparison button to an appropriate position. After 3 minutes, image intensifier television system is under
stable working condition. Then the apparatus can be operated for practical use.
The shell of the apparatus and the television system shall have a good earth
connection so as to avoid possible health damage caused by leakage electricity.
WARNING
5.3.3 Working Mode (Status) Selection
When pressing keys, please do not hold on for relatively long period.
Otherwise , the working mode will be "ON" and "OFF" or shift to other ones in
NOTICE cycles..
cycles
a. Manual Radioscopy
When the apparatus is started under normal status, the displaying of kV, mA and mAs shall be “60”,
“1.0” and“000”,indicating normal manual radioscopy status.
b. ABC Function
If you want to select ABC function, please press ABC status selection key. When the indication light
above ABC key becomes bright, it shows that the apparatus is under ABC status. But its mA selection needs
manual control.
c. Radiography Function
If you want to select radiography function, press Radiographic status selection key. When the indication
light above Radiographic status selection key becomes bright, it shows that the apparatus is under
radiography status. At this time, kV, mAs and mAs values display“60”,“10” and“40”.
d. Beam limiting device Operation
The electric beam limiting device can move in four directions.
21
Do not press two keys at the same time, otherwise the 12V DC power may
gets short cut.
WARNING
Medical staff shall wear protective clothes. Put the manual brake line as
further away from radiation source as possible, being over 2 meters.
WARNING
The operation table for convenient operation will affect image quality of
X-ray apparatus. Please select the operation table with less attenuation
WARNING equivalent (< 2.3 mmAl). Please do not use iron and steel table in order to avoid
the penetration ability from decreasing.
22
In order to minimize X-ray absorbing dose by patient, the operator shall
patient’’s skin as distant as possible.
make the distance between the focus and patient
WARNING
Under service program, the computer no long monitor the working mode of
the main machine. Therefore, if the main machine fails to operate, it will lead the
apparatus out of control, bringing about greater damages. Non-professionals are
not allowed to do the work. At the same time, when the maintenance worker
WARNING employ service program in maintaining work, please remember to test the status
of each operation point of the equipment. If there is any problem, please resolve
it in time.
23
Chapter 6 Proper Combination of Testing and Test of Loading Factors
6.2 Loading Factor Indication Values and Result Consistency Test Combination
6.2.1 X-ray Tube Voltage:
X-ray tube voltage X-ray tube current
Test loading factor Current time product(mAs)
(kV) (mA)
40 60 1
Intermittent 40 60 6.3
110 32 3.2
99 Arbitrary Not applicable
Continuous
66 Arbitrary Not applicable
110 10 1
Intermittent
40 50 250
24
Testing Instrument and Their Precision
Direct-reading digital kilo-voltage meter: Model 4000M + ±2%
Storage oscillograph: 100M
Digital mAs meter
Digital multi-meter
6.6 Current Adjustment Range of Tube Voltage and Testing of Warning Device
6.6.1 Testing of Perspective Tube Voltage
The testing of perspective tube voltage shall be conducted simultaneously with that in Section 6.3. The
tube voltage shall be able to be adjusted continuously with upper limit 110kV and lower limit of 40 kV.
6.6.2 Testing of Tube Current
The testing of tube current shall be conducted simultaneously with that in Section 6.4. The tube voltage
shall be able to be adjusted continuously with upper limit 5mA and lower limit of 0.3mA.
25
Chapter 7 Maintenance
When replacing the fuse, firstly pull out the power plug, open the fuse socket,
then replace the broken fuse with a new one.
DANGER
26
Chapter 8 Breakdown Analysis and Debugging
8.1 Code Number and the Circuit of Each Circuit Board of X-ray Apparatus
See Attachment D.
The disposal of the oil from discarded high voltage transformer shall be
subject to the approval of local environment protection agency.
WARNING
27
Attachment A Indication Signs and Abbreviations
Signs
This sign means the electrical safety of the X-ray machine is
classified as Cat. B.
Abbreviations
kV
kV- Negative kV sampling of high voltage generator
mA
mA- Negative mA sampling of high voltage generator
28
Attachment B Circuit Theory Sketch
29
Attachment C Table of Key Components
30
IGBT 150A 2 Electric frame
Electrolytic
CD03-450V-3300uf
condenser 4 Electric frame
(Hitachi)
C1 ~ C4
6A (Small, medium YUHAO
Noise filter 1 Electric frame
ZYH-EF6A)
6A (Big, medium Shidianzi
Noise filter 1 Electric frame
FLHD53-6A)
Three-phase rectifier
MDS-90A-1200V 1 Electric frame
bridge
Contra-variant
3.6uf/1000V 1 Electric frame
capacitor
Filter capacitor
Filtercapacitor 2uf/1600V 2 Electric frame
High voltage Within the combined tube
1W/10M 20
resistance component
Switch power SP-T50B (Yongguang) 1 Electric door plate
Power plug EP10 (Qisheng) 1 Power line
Within the combined tube
X-ray tube 714 C 1
component
High-voltage Within the combined tube
(Homemade) 2
package primary
package—primary component
High-voltage Within the combined tube
(Homemade) 4
package secondary
package—secondary component
Filament package
package— Within the combined tube
(Homemade) 2
primary component
Filament package
package— Within the combined tube
(Homemade) 2
secondary component
Hoist motor 280 W 1 Bottom of electric frame
Position limit switch Omron 4 Back of the column
31
Attachment D Code of Breakdown
Code of
Cause Clearing of method
Breakdown
Err 01* Gear I of Manual brake Check the manual brake, CPU board E1, U11 and U1.
Err 02* Gear II of Manual brake Check the manual brake, CPU board E1, U11 and U1.
Err 03* Main contra-variant power Check BG1, C1 ~ C4 and discharging board E1.
Err 04* Filament power Check filament BG1, Q1, E3 and CPU board E4.
Err 16* Overheat of X-ray tube Stop adding load, let the tube rest.
Err 18* IPM breakdown Check IPM and its driver boards E1, E3.
Err 22 Perspective time limit 5 min. Press any key and the apparatus resume to the normal.
Over-heat at Radiography Press any key and the apparatus resume to the normal.
Err 23
manual brake gear I (15s)
32
Attachment E Specifications of Fuses
F1 20A Φ10
10 × 38 Left of electric frame
F2 20A Φ10
10 × 38 Left of electric frame
33
Attachment F Power Circuit Electric Materials
Contra-variant capacitor
3.6 uf/1000V 1 Electric frame
CM1
34
Attachment G
X-ray Protection Explanation for HMC-3
HMC -366 mobile C-arm X-ray system
HMC-36 mobile C-arm X-ray system features with C-type structure and high frequency contra-variant
X-ray generation. It has been confirmed that this apparatus has been tested by national testing institutes and in
line with national standard with each technical specifications meeting the design standards. When being
undertaken protective test, GB8279-87 "Health Protection Standard for Medical Diagnostic X-radiation".
Therefore, the provision of the National Standard of GB8279-87 shall be abided by when protective testing is
carried out. However, no consideration shall be made for those clearly are not applicable to this apparatus.
Because of the specialty of the apparatus, only the leaking radiation from the tube nose shall be tested
according to the requirements of GB8279-87, and the tube window needs to be sealed by lead. For specific
method, please refer to relevant context of Section 2.2.1 of GB8279-87.
The provisions or sections of GB8279-87 that are applicable to this apparatus are listed in the following
table. They act as a reference for local health testing departments in the protective testing for HMC-36 mobile
high frequency medical diagnostic X-ray apparatus.
Testing item Applicable provisions Testing method Remark
Leaking radiation of Shall be in line with the requirements of Comply with Section 2.2.1
X-ray tube Section 2.1.1.1 of GB8279-87 of GB8279-87
Shall be in line with the requirements of Check and make sure it Manual and foot
Radioscopy switch
Section 2.1.1.6 of GB8279-87 works well switch
In view of the special characteristics of mobile C-type X-ray apparatus, operator and operational doctor
have to work in close proximity with X-radiation when patient is diagnosed by C-type arm. Therefore, the
protection conditions shall be enhanced. All people involving the operation must wear protective device such as
lead clothes, lead cap, lead scarf and lead glasses. If the protection conditions need enhancement, please refer to
the "Restructuring Sketch for Bed Radiation Protection".
35
Bed Restructuring Sketch for Radiation Protection
Standing area
Protection device
Ground
Protection device
Standing area
Unit: cm
36