6/4/2021 Monovalent and Divalent Cation Standards (6) (Comparison of YS-50 and YK-421) | Shodex/ HPLC Columns, Detectors,

Detectors, Standards

Monovalent and Divalent Cation Standards (6) (Comparison of YS-50


and YK-421)

The six main cations can be separated using a newly developed column, IC YS-50. Comparing with the conventional column IC
YK-421, the theoretical plate number of divalent cations becomes more than double and the separation of Na+ and NH4+
becomes about 1.2 times.

Sample :
1. Li +
2ppm
2. Na+
10ppm
3. NH4+
10ppm
4. K+
20ppm
5. Mg 2+
10ppm
6. Ca2+
20ppm

Columns : Shodex IC YS-50 (4.6mmI.D. x 125mm) Shodex IC YK-421 (4.6mmI.D. x 125mm)


Eluent : 4mM Methanesulfonic acid aq. 5mM Tartaric acid
+ 1mM Dipicolinic acid
+ 1.5g/L Boric acid aq.
Flow rate : 1.0mL/min 1.0mL/min
Detector : Non-suppressed conductivity Non-suppressed conductivity
Column temp. : 40°C 40°C

https://www.shodex.com/en/dc/07/03/32.html 1/1