Professional Documents
Culture Documents
Date: 18/08/2021
Tube Inspection JP06/ MFL/Gen
Rev # 03 Page 1 of 7
Document
Rev Issue Date Prepared by: Reviewed by: Approved by:
No:
Name & Name &
Name & Signature:
Signature: Signature:
Lalith R. Vipin Shetty Ganeshbabu.M
Level-II Level-III Level-III
QHSE/ P11/
ANDT/ JP06/ 03 18/08/2021
MFL/Gen
Rev # 03 Page 2 of 7
2. References------------------------------------------------------------------------------ 3
7. Calibration Standard------------------------------------------------------------------ 4
9. Calibration ----------------------------------------------------------------------------- 4
Rev # 03 Page 3 of 7
1.0 Scope:
This Procedure describes the method of applying Magnetic Flux Leakage (MFL) testing to test
Ferromagnetic heat exchanger, Air Cooler and Boiler tubes.
The personnel who operate the MFL test equipment with the exception of the probe positioner
and puller technician shall be a minimum Level II in Magnetic Particle Testing Method.
Certification of the personnel shall be according to RISS Written Practice
The Result Evaluation of the MFL must be conducted by Personnel qualified to at least Level II
The basic principle of MFL system is belongs to the Permanent magnets. The two strong
permanent magnets coupled to a steel core, generate the magnetic field that saturates the tube
wall
An absolute coil (ABS) is wound around the core to measure magnetic field variation caused
by general wall loss
When small flaw is located between the two magnets the magnetic lines in the tube wall is
disturbed, resulting in a small amount of flux leakage into the inner tube. This flux leakage is
detected by the differential coil (LEAD), located between the magnets
A trailing coil (TRAIL), at the end of the probe, detects the residual magnetism
MFL probe selected based on the diameter and thickness of the tube as per the table.
Laptop with MFL software i.e. Multiview software
Wall loss adaptor
Procedure For Magnectic Flux Leakeage For QHSE/ P11/ ANDT/
Date: 18/08/2021
Tube Inspection JP06/ MFL/Gen
Rev # 03 Page 4 of 7
All tubes inner surface shall be adequately cleaned to remove any dust, loose foreign particles,
internal deposits or contaminations by a suitable process such as pressurized water jet cleaning.
It is also recommended to clean the outer surface of tube if possible.
Calibration tube shall be manufactured from a tubing of the same material specification, same
heat treatment and same nominal size and thickness. The calibration tube shall be manufactured
according to the sketch provided in Attachment-1
9.0 Calibration:
The calibration standard tube shall be used for equipment calibration as per following steps.
· The wall loss adapter is placed between the probe and MS5800 to integrate the voltage
across the absolute coil
· Check wall loss adapter drift periodically
· The RESET button sets the integrator output to zero. During the inspection RESET is
recommended every 10 tubes
· When insert the probe into the tube in order to reduce the electromagnetic noise. Hold the
RESET button and click the Balance
The test equipment response shall be checked and verified by use of the calibration tube
standard as follows
Rev # 03 Page 5 of 7
If during any check, it is determined that the test equipment is out of calibration (as defined by
the work instruction) the equipment shall be re calibrated and a new data file opened. The data
analyst shall determine which tubes, if any are to be reexamined.
It is the responsibility of the data capture technician to ensure that the data is of the quality
required by the relevant procedure or as defined by the data analyst.
· The inspection is conducted with the probe located at the fair end of the tube. The assembly
is pushed to the end of the tube by feeding the hose through the tube
· To start the inspection the assembly is gradually pulled along the tube towards the operator.
· The above procedure can be repeated for as many tubes are to be inspected during the
operating session. The equipment dismantled procedures are to be inspected during the
operating session. The equipment dismantle procedure are the reverse of those carried out
during the setup of the inspection
· The Pulling speed of the probe shall be kept constant during the inspection. The pulling
speed shall be kept same as the speed of the calibration.(maximum speed is 1m/s)
· In all techniques measurements are taken whilst withdrawing the probe. The pushing and
withdrawing shall be done manually.
· The technician shall ensure that an even speed is maintained
For the purpose of this procedure all test results will be recorded on suitable data acquisition
system
· Each data storage unit shall contain the following information
· Positive identification of the item under test
· Data storage unit number
· A reference to the contents
· Data
· Probe identification
Each trace shall be identified according to zone / row / column and tube number as applicable
Analysis of all data shall be conducted by NDT level II qualified in data analysis.
All indications shall be evaluated. Signal amplitude and phase shall be taken into account
whenever defect indications are evaluated using the absolute or differential modes.
All indication must be evaluated and defects classified in a category of wall loss as follows
Procedure For Magnectic Flux Leakeage For QHSE/ P11/ ANDT/
Date: 18/08/2021
Tube Inspection JP06/ MFL/Gen
Rev # 03 Page 6 of 7
The default setup already includes defects codes and depth curves
Reporting of indication shall be in accordance with attachment 1, and shall be recorded on the
report format
The acceptance criteria or rejection of tubes based on wall loss is at the discretion of the client.
· Reporting shall be prepared with the information gathered using the computer data files
· Report all tubes wall with a wall loss reading, percentage of wall loss and carto diagram
with wall loss and all reports issued to the clients shall Be signed by the person performing
the inspection
Procedure For Magnectic Flux Leakeage For QHSE/ P11/ ANDT/
Date: 18/08/2021
Tube Inspection JP06/ MFL/Gen
Rev # 03 Page 7 of 7
Rev # 03 Page 8 of 7