Professional Documents
Culture Documents
In general, measurements in dry dispersion are quicker than wet, enabling you to
Particle size measurements of measure a larger mass of material in a short time. This is particularly useful for
materials in the dry dispersed achieving reproducible results for coarse or polydisperse materials.
state are increasingly being used However, not every sample is suitable for dry dispersion. Very fine or cohesive
as an alternative to wet or liquid materials may be more suited to wet measurement, as surfactants and additives
dispersion methods. There are may be required to achieve full and reproducible dispersion. Dry dispersion may not
several advantages to using dry be suitable for extremely fragile materials due to the more aggressive nature of the
powder dispersion for laser dispersion mechanisms. Finally, while it is possible to measure very small sample
diffraction measurements: masses using a dry system, improved result reproducibility is often achieved using
If a material is processed in the wet dispersion.
dry state then measuring the Achieving reproducible results from any kind of particle characterization system
particle size in the same state may depends upon three factors:
be more representative than wet
dispersion. Representative sampling,
The correct state of dispersion
For some materials dry Appropriate measurement conditions.
measurement may be the only
option, due to dissolution in
common dispersants or changes The relative importance of each of these factors depends on the size of the particles
in the size of the particles in wet being measured. Figure 1 shows the relative risk factors associated with
dispersion (due to hydration). measurements of fine or coarse particles in dry dispersion.
Figure 1: Risk factors for coarse and fine particles in dry dispersion
Therefore, if you are measuring fine particles then the air pressure will have the
greatest influence on the results as it affects the samples state of dispersion.
However, if you are measuring coarse particles, the greatest possible source of
error will be sub sampling of the material.
The dry result showing the closest agreement to the well dispersed wet result will indicate the appropriate pressure. If the
results of a dry measurement are finer than the wet measurement this indicates that the primary particles are being broken at
that pressure. Figure 3 shows a dispersed wet result for the milk powder sample, compared to dry results measured at 0.5 bar
and 3 bar. The 0.5 bar measurement shows a larger size compared to the wet measurement, indicating that the sample is not
fully dispersed at 0.5 bar. However, the 3 bar measurement shows excellent comparability to the wet measurement. Therefore,
for this material the appropriate pressure to disperse the sample to its primary particle size is 3 bar.
Figure 3: Comparison of dispersed wet result and dry results at 0.5 bar and 3 bar
For the majority of materials these two mechanisms are sufficient to disperse the sample. However, some highly agglomerated
and robust samples may require the additional higher energy dispersion mechanism of particle-to-wall collisions, Figure 4 (c).
As the standard venturi is applicable to the vast majority of samples then this will be your starting point for developing a dry
method. However if full dispersion is not achieved for a robust material then the Aero’s high energy venturi , which makes use
of all three dispersion mechanisms, may be required.
If we measure the milk powder sample using the high energy venturi then we see a different pressure titration. Full dispersion
is achieved at a lower pressure, due to the higher energy dispersion mechanism. Figure 5 shows that the dry result at 1 bar
using the high energy venturi shows good agreement to wet result, whereas the dry result at 4 bar shows a smaller particle
size which indicates particle attrition.
Figure 5: Comparison of dispersed wet result and high energy venturi dry results at 1 bar and 4 bar
Figure 6 shows the Dv50 vs. pressure measured using both the standard venturi and high energy venturi, with the Dv50 from
the wet result added as a dashed line for comparison. This shows that results comparable to the wet measurement are
achieved at between 3 bar and 4 bar using the standard venturi and at 1 bar using the high energy venturi. In this case the
standard venturi would be recommended, as the dry results are comparable to the wet dispersion over a wider range of
pressure which improves the robustness of the method.
120
100
Size / m
80
60 Standard Dv50
High energy Dv50
Wet Dv50
40
0 1 2 3 4
Air pressure / bar
Figure 6: Dv50 vs. pressure using standard and high energy venturi compared to wet Dv50
Feed rate
The feed rate controls the rate at which material is fed through the venturi to the measurement cell by vibrating the sample
tray. The amount of material passing through the measurement cell is recorded via the obscuration (the percentage loss of
laser light through the measurement cell). The feed rate must be set so that the obscuration is in range for the majority of the
measurement duration, as illustrated in Figure 7. The obscuration filtering feature in the Mastersizer software ensures that
detector scans where the obscuration is outside the target range are not included in the results. The percentage of the
measurement duration where the obscuration was within range can be checked using the ‘Snap collection percentage’
parameter. The higher this percentage the more scans of the detector were within the obscuration range, and the more
representative and reproducible the results. This is one of the parameters checked by the Mastersizer 3000s data quality
report for dry measurements, which will show a warning when poor obscuration control is detected.
Figure 7: Variation in obscuration vs. time, showing good control of the feed rate in green, and poor control in red.
Sampling
If you are measuring coarse particles then amount of sample you measure and how you take the sub sample become the
largest possible source of error.
To get a representative result you need to measure a minimum number of particles. For example, to get the Dv90 to within a
5% standard error requires at least 400 particles to be measured during the analysis. Therefore, for a material with a density of
3
1.5g/cm and a Dv90 of 500μm you would need to measure about 0.5g to get the Dv90 to within a 5% standard error. As the
size of the particles increases the mass containing sufficient particles increases, and the minimum mass required to achieve
reproducible results also increases.
90
80
70
Minimum mass / g
60
50
40
30
20
10
0
0 500 1000 1500 2000 2500 3000
Dv90 /μm
Figure 1 shows the minimum mass required to achieve a 5% standard error as a function of particle size. You can test whether
your sample mass, or technique, is sufficient by measuring several separate sub samples and assessing the variability of the
results.
For example, Figure 9 shows several 5 second measurements of a sample with a median size of approximately 600μm. The
relative standard deviation of the percentiles shows quite a high variability which can be improved by increasing the
measurement duration and thereby improving the sampling of the material.
For finer, or monodisperse, samples it is possible to measure a very small mass of sample and still get representative results.
Figure 10 shows the results of dry measurements of a baby powder sample from a sub samples with masses between 50mg
and 5mg. The table in Figure 10 contains the resulting percentiles and relative standard deviations which show reproducibility
well within the expected repeatability for a laser diffraction measurement, according to the ISO standard [1].
Figure 10: Size distributions and variability for baby powder samples from 50mg to 5mg.
References
[1] ISO13320 (2009). Particle Size Analysis - Laser Diffraction Methods, Part 1: General Principles
Measurement precision
The precision of laser diffraction measurements can be assessed using the coefficient of variation.
%
(also referred to as the relative standard deviation)
The ISO standard for laser diffraction [1] recommends that the %COV should be less than 3% for central parameters such
as the Dv50, and less than 5% for parameters at the extremes of the distribution such as the Dv10 and Dv90. The
Mastersizer 3000 has a variability check in its data quality tab to allow you to check the variability of your measurements
against the limits set in the ISO standard.