CO2 fractional beauty instrument manual

CO2 fractional beauty instrument manual

Contents
1. Boot and main interface...........................................................................................................4

2. Normal mode................................................................................................................................ 5

2.1 Power Button............................................................................................................................................. 5

2.2 Mode Button.............................................................................................................................................. 5

2.3 Duration Button....................................................................................................... 错误!未定义书签。

2.4 On time Button......................................................................................................................................... 6

2.5 Off time Button..........................................................................................................................................6

2.6 standby Button.......................................................................................................................................... 6

2.7 Ready Button............................................................................................................................................. 6

2.8 Return Button............................................................................................................................................ 6

3. Fractional mode.......................................................................................................................... 7

3.1 Shape........................................................................................................................................................ 7

3.2 Graphic size adjustment......................................................................................................................... 7

3.2 Parameter saving function...................................................................................................................... 8

3.3 Power........................................................................................................................................................... 8

3.4 Distance...................................................................................................................................................... 8

3.5 Duration....................................................................................................................................................... 8

3.6 Scanning mode......................................................................................................................................... 8

1
CO2 fractional beauty instrument manual

3.7 Overlap........................................................................................................................................................ 9

3.8 Repeat......................................................................................................................................................... 9

3.9 Preview button...........................................................................................................................................9

3.10 Standby button........................................................................................................................................ 9

3.11 Ready button........................................................................................................................................... 9

3.12 Return Button.......................................................................................................................................... 9

4. Gynae mode................................................................................................................................. 11

4.1 Shape selection...................................................................................................................................... 11

4.2 Duration..................................................................................................................................................... 11

4.3 Scan rows................................................................................................................................................ 12

4.4 Point number........................................................................................................................................... 12

4.5 Power........................................................................................................................................................ 12

4.6 Overlap...................................................................................................................................................... 12

4.7 Preview button........................................................................................................................................ 12

4.8 Displacement............................................................................................................................................12

4.9 Standby Button........................................................................................................................................13

4.10 Ready Button........................................................................................................................................ 13

4.11 Return Button........................................................................................................................................ 13

5.Vulvae mode................................................................................................................................... 14

5.1 Shape selection...................................................................................................................................... 14

5.2 Duration..................................................................................................................................................... 14

5.3 Distance.................................................................................................................................................... 15

2
CO2 fractional beauty instrument manual

5.4 Power........................................................................................................................................................ 15

5.5 Overlap...................................................................................................................................................... 15

5.6 Repeat....................................................................................................................................................... 15

5.7 Displacement............................................................................................................................................15

5.8 Preview Button........................................................................................................................................ 15

5.9 Standby Button........................................................................................................................................16

5.10 Ready Button........................................................................................................................................ 16

5.11 Return Button........................................................................................................................................ 16

6.Password and alarm surface......................................................................................... 17

6.1 Password unit.......................................................................................................................................... 17

6.2 Voice switch............................................................................................................................................. 17

6.3 Language.................................................................................................................................................. 17

6.4 Red light adjustment..............................................................................................................................18

6.5 Alarm beacon.......................................................................................................................................... 18

7.Setting interface..........................................................................................................................18

3
CO2 fractional beauty instrument manual

1. Boot and main interface


Turn on the key switch and emergency stop, the screen enters the main
interface, please wait for 1 seconds to select the function.
The main interface includes normal mode, fractional mode, gynae mode,

vulvae mode. 4 functional selection and System settings button . Please


select different buttons to enter the corresponding function.

(Main interface)

4
CO2 fractional beauty instrument manual

2. Normal mode
The ultra-pulse/pulse function includes power, working mode, on time, off time,
standby, ready, and “+” “-”.Press “+”or “-” to adjust the value of the function. As
shown below

2.1 Power Button

Adjust the laser power, adjust from 1w to 60w, stepping 1w.

2.2 Mode Button

Modes include pulse mode, super pulse mode, single and continuous.
Pulse In the pulse mode, the adjustment range of the on time and the off time is
1-99ms, and the step is 1ms.
Super Pulse In the super pulse mode, the adjustment range of the on time and
the off time is 0.1-9.9 ms, and the step is 0.1 ms.
Single Only one light is emitted in a single time, the range is 1-99ms, and the
step is 1ms.
Continuous The laser is always emitted continuously, and the intensity of the
light is controlled by the power.

5
CO2 fractional beauty instrument manual

2.3 On time Button

Each laser exit time in the pulse and super pulse modes, and the off time
constitute a periodic cycle.

2.4 Off time Button

The pulse time and the super pulse mode do not emit laser each time, and the
working time constitutes a cyclic cycle.

2.5 standby Button

The parameters can be adjusted in the standby state. When the standby button
is clicked or the parameters are adjusted, the system will automatically exit the
standby state and enter the standby state.

2.6 Ready Button

When pressed, the system exits the parameter adjustable state and enters the
standby mode.Only when the standby mode is turned off, the footswitch laser
will emit light.Press the standby button again, the system exits the standby
mode and enters the parameter editable state.

2.7 Return Button

Press Return button to exit the current interface and return to the previous
interface.

6
CO2 fractional beauty instrument manual

3. Fractional mode
The CO2 fractional mode mainly includes shape selection, parameter saving
and reading, graphic size, power, distance, duration, scanning, overlap, interval
time, and common plus or minus adjustment, as shown in the following figure.

3.1 Shape

Including five kinds of graphics: square, circle, triangle, hexagon, ellipse, click
the shape button, then adjust the plus and minus numbers, select the desired
graphic, and the selected graphic is displayed in the main display frame.

3.2 Graphic size adjustment

The up, down, left and right buttons are used to adjust the size of the b
asic graphics. The height Y adjustment range is 1mm-20mm, the height X ad
justment range is 1mm-20mm, and the step is 0.5mm.

7
CO2 fractional beauty instrument manual

3.2 Parameter saving function

Includes 1-5 numeric keys and Save, the number 1-5 of the motor can di
rectly retrieve the saved parameters. When the parameter is saved, first pres
s a number key, then adjust other parameters. When press Save, the numbe
r key saves all current parameters.

3.3 Power

Adjust the laser power, adjustable from 1w to 60w, the step is 1w.

3.4 Distance

It is used to adjust the spacing of the points in the current graphic. The
adjustment range is 0.1mm-2mm and the step is 0.1mm.The smaller the dot
pitch, the greater the density of dots.

3.5 Duration

It is used to adjust the action time of the laser at each point. The longer the time,
the more energy the laser acts on each point.The adjustment range is
0.1-9.9ms and the step is 0.1ms.

3.6 Scanning mode

(Include 3 modes: Order, half and random)


①order mode: Scans from the vertex of a dot pattern to the end point when
striking.
②half mode: When striking, the dot pattern is divided into four quadrants from
the middle. The dot-up method is to first finish the first point of each quadrant,
and then hit the second point of the four quadrants until the end. The last point
in each quadrant.
③Random scan mode: When the dot is clicked, the dot pattern is scanned in
an out-of-order manner.

8
CO2 fractional beauty instrument manual

3.7 Overlap

It is used to adjust the number of dotes in the dot pattern. All points in the graph
are finished as a complete dot.The adjustment range is 1-10 times.

3.8 Repeat

It is used to adjust the time between two adjacent dot processes when the
number of scans is greater than one. The adjustment range is 0.1s-2s, and the
adjustment range is 0.1s.

3.9 Preview button

When the preview button is pressed, the system exits the parameter adjustable
state and enters the preview mode. At this time, the dot hand tool scans the
contour of the current graphic, and the contour is the contour of the laser action
range during the doting, and the user determines the scanning hand tool
placement by using the contour. The location on the patient's skin.

3.10 Standby button

The parameters can be adjusted in the standby state. When the standby button
is clicked or the parameters are adjusted, the system will automatically exit the
standby state and enter the standby state.

3.11 Ready button

Press the ready button to enter the preview and prepare modes.Only when the
standby mode is closed, the footswitch laser will emit light, and the system will
only operate.When the pedal is lifted during a complete dot-up, the system
continues to complete the remaining dot-up process after the pedal is closed
again.

3.12 Return Button

Press the back key to exit the current interface and return to the previous

9
CO2 fractional beauty instrument manual

interface.

10
CO2 fractional beauty instrument manual

4. Gynae mode
Use a private handpiece when at the Gynae mode with a focal length of
152mm.The gynae functions include shape selection, scan rows, duration, power,
number, overlap, scan, plus and minus adjustments, and displacement adjustment.
As shown below

4.1 Shape selection

Including squares and circles, squares need to be used with 45° elliptical
reflectors, and rounds need to be used with tapered reflectors.

4.2 Duration

To adjust the action time of the laser at each point. The longer the time, the more
energy the laser acts on each point.The adjustment range is 0.1-9.9ms and the step
is 0.1ms.

11
CO2 fractional beauty instrument manual

4.3 Scan rows

Adjustable only in circles.The number of strokes can be adjusted only when a


circle. It can also be considered as a circular number of rings. The adjustment
range is 1-9 lines.

4.4 Point number

Adjustable only when rounded.It is used to adjust the number of points per circle
(circle) when the circle is adjusted. The adjustment range is 20-120 points.The
more points, the greater the density of points.

4.5 Power

Adjust the laser power, adjustable from 1w to 40w, the step is 1w.

4.6 Overlap

It is used to adjust the number of dotes in the dot pattern. All points in the graph are
finished as a complete dot.The adjustment range is 1-10 times.

4.7 Preview button

When the preview button is pressed, the system exits the parameter adjustable
state and enters the preview mode. At this time, the dot hand tool scans the contour
of the current graphic, which is the contour of the laser action range during the
doting.

4.8 Displacement

It is used to adjust the position of red light and laser in the field of
view, so that the figure is at the center of the field of view, and is adjusted in four
directions: up, down, left, and right.

12
CO2 fractional beauty instrument manual

4.9 Standby Button

The parameters can be adjusted in the standby state. When the standby button
is clicked or the parameters are adjusted, the system will automatically exit the
standby state and enter the standby state.

4.10 Ready Button

Press the ready button to enter the preview and standby modes.Only when the
standby mode is closed, the footswitch laser will emit light, and the system will
only operate.When the pedal is lifted during a complete dot-up, the system
continues to complete the remaining dot-up process after the pedal is closed
again.

4.11 Return Button

Press the back key to exit the current interface and return to the previous
interface.

13
CO2 fractional beauty instrument manual

5.Vulvae mode
With a vulvar handpiece when at the vulvae mode, and the focal length is
152mm.The vulvae function mainly includes shape selection, distance, duration,
power, overlap, interval and common plus and minus adjustment, as well as
graphic size and displacement adjustment.As shown below

5.1 Shape selection

Includes squares and circles.

5.2 Duration

It is used to adjust the action time of the laser at each point. The longer the time,
the more energy the laser acts on each point.The adjustment range is 0-10ms
and the step is 0.1ms.

14
CO2 fractional beauty instrument manual

5.3 Distance

It is used to adjust the spacing of the points in the current graphic. The
adjustment range is 0.1mm-1mm and the step is 0.1mm.The smaller the dot
pitch, the greater the density of dots.

5.4 Power

Adjust the laser power, adjustable from 1w to 40w, the step is 1w.

5.5 Overlap

It is used to adjust the number of dotes in the dot pattern. All points in the graph
are finished as a complete dot.The adjustment range is 1-10 times. When 10
times is selected, the system will be infinitely counted, that is, if the pedal is not
lifted after the pedal is pressed, it will always be hit.

5.6 Repeat

To adjust the time between two adjacent dot processes when the number of scans
is greater than one. The adjustment range is 0.1s-2s, and the adjustment range is
0.1s.

5.7 Displacement

To adjust the position of red light and laser in the field of view, so that the figure is at
the center of the field of view, and is adjusted in four directions: up, down, left, and
right.

5.8 Preview Button

When the preview button is pressed, the system exits the parameter adjustable
state and enters the preview mode. At this time, the dot hand tool scans the contour
of the current graphic, and the contour is the contour of the laser action range
during the doting, and the user determines the scanning hand tool placement by
using the contour. The location on the patient's skin.

15
CO2 fractional beauty instrument manual

5.9 Standby Button

The parameters can be adjusted in the standby state. When the standby button is
clicked or the parameters are adjusted, the system will automatically exit the
standby state and enter the standby state.

5.10 Ready Button

Press the reserve button to enter the preview and standby modes.Only when the
standby mode is closed, the footswitch laser will emit light, and the system will only
operate.When the pedal is lifted during a complete dot-up, the system continues to
complete the remaining dot-up process after the pedal is closed again.

5.11 Return Button

Press return button to exit the current interface and return to the previous interface.

16
CO2 fractional beauty instrument manual

6.Password and alarm surface


Click the gear-shaped setting button in the lower right corner of the main
interface to enter the password page, enter 123456 to confirm and enter the setting
interface.

6.1 Password unit

Enter 123456 to enter the settings page.

6.2 Voice switch

Responsible for the buzzer sound on the switch control board.

6.3 Language

Click to switch between Chinese and English.

17
CO2 fractional beauty instrument manual

6.4 Alarm beacon

When the system fails, the corresponding alarm icon will light up. Please contact
the factory for processing.

7.Setting interface
Please note that the parameters in the setting interface should not be tampering,
and non-professionals should not touch it.
The setting interface is mainly divided into hand scale proportional position
calibration and power calibration. If necessary, please contact the manufacturer for
guidance. Do not operate without permission.

18