Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00
• Requires downloading of the Optics Utility from TECTA-PDS onto USB drive. Access to this
download file is restricted to TECTA service level 2 technicians.
• Calibration is a tool of last resort on chambers that fail TECTA B16 software Validation QC
criteria (red icon upon validation or red crossed chambers) to regain loss of optical performance.
The first attempt to improve optical performance should be to clean optical path components
using approved methods or replace optical cover(s) and re-validate using TECTA B16 software.
• Minimize risk by not frequently reading and writing to the EEPROM microchip of TECTA B16
instrument.
• Only perform calibration or validation on an instrument where optical components are at room
temperature (18 to 25°C). Significant deviation from room temperature such as 8°C (upon
delivery) or 44.5°C (immediately following a fecal coliform incubation) can have a +/-15% impact
on optical validation results.
• Only Calibrate individual chambers that fail the TECTA software Validation QA criteria after
proper cleaning or inspection.
• Calibrate a quadrant of 4 chambers if an optical cover is exchanged where one or more probe
fails TECTA software Validation QA criteria.
• The Optical utility can be used by service level 2 technicians to generate an optical performance
report for all 16 chambers of TECTA B16 instrument in one report file.
Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 2 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00
Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 3 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00
Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 4 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00
8) Select ‘Chamber: 1’ by
pressing check box or
‘Select’.
Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 5 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00
Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 6 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00
Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 7 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00
Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.