Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00

General Notes on Optical Calibration:

• Requires downloading of the Optics Utility from TECTA-PDS onto USB drive. Access to this
download file is restricted to TECTA service level 2 technicians.

• Calibration is a tool of last resort on chambers that fail TECTA B16 software Validation QC
criteria (red icon upon validation or red crossed chambers) to regain loss of optical performance.
The first attempt to improve optical performance should be to clean optical path components
using approved methods or replace optical cover(s) and re-validate using TECTA B16 software.

• Minimize risk by not frequently reading and writing to the EEPROM microchip of TECTA B16
instrument.

• Only perform calibration or validation on an instrument where optical components are at room
temperature (18 to 25°C). Significant deviation from room temperature such as 8°C (upon
delivery) or 44.5°C (immediately following a fecal coliform incubation) can have a +/-15% impact
on optical validation results.

• Only Calibrate individual chambers that fail the TECTA software Validation QA criteria after
proper cleaning or inspection.

• Calibrate a quadrant of 4 chambers if an optical cover is exchanged where one or more probe
fails TECTA software Validation QA criteria.

• Calibrate all 16 chambers if the paired Validation Cartridge is exchanged.

• The Optical utility can be used by service level 2 technicians to generate an optical performance
report for all 16 chambers of TECTA B16 instrument in one report file.

Optical Calibration Procedure:

1) Download the Optical Utility


from TECTA-PDS.

Use a dedicated, reliable


USB drive with minimum of
2GB of storage space.

Consider a backup USB


drive or have the download
file accessible when
servicing a remote site.

Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 2 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00

2) Insert Optics Utility USB


drive into either USB port of
the TECTA B16.

Instrument must be powered


on first before inserting USB
key to be recognized.

3) Press ‘Home’ icon to view


the ‘System’ tab.

4) Press ‘Upgrade Application’.

When prompted, press ‘Yes’


to proceed with upgrade.

Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 3 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00

5) Type ‘272’ as Pass Key and


press ‘Authorize’ button.

6) Do not press the LCD


touchscreen and do not
remove USB drive while the
Optics Utility is loading.

7) Press ‘Clear All’ button once


the Optics Utility user
interface is displayed.

Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 4 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00

8) Select ‘Chamber: 1’ by
pressing check box or
‘Select’.

9) Press ‘Unlock’ button to


release lid of a TECTA B16
Rev1 and open the lid of
instrument.

Pressing the ‘Unlock’ button


is unnecessary if servicing a
TECTA B16 Rev2
instrument.

10) Load the Validation Cartridge


into ‘Chamber: 1’ and close
lid.

If servicing a TECTA B16


Rev1, press ‘Lock’ button to
keep lid retained during test.

Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 5 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00

11) Press ‘Calibrate’ button and


wait a few seconds while
chamber icon is yellow.

12) If the chamber calibration


result is green and the bottom
line reads “SUCCESSFUL
CALIBRATION!”, press the
‘Calibrate to EEPROM’ button.

If the chamber calibration


result is orange and reads
“Out of Bounds!” DO NOT
perform any other actions on
this chamber. Uncheck
chamber and move on to
next chamber requiring
calibration.

13) After the bottom line reads


“SUCCESSFUL WRITE!”
press ‘Validate’ button and
wait a few seconds while
chamber icon is yellow.

Chamber will display % delta


and offset delta and chamber
icon will display green if QA
criteria passed and purple if
QA criteria failed.

Failed chambers should be


re-calibrated until they pass.

Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 6 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00

14) To move to next chamber


requiring calibration, uncheck
‘Chamber: 1’ by pressing
checkbox or ‘Select’ and
select next chamber by
pressing its checkbox or
‘Select’.

15) Open the lid of the


instrument (press ‘Unlock’ if
Rev1). Load validation
Cartridge into next chamber
to calibrate and close lid
(press ‘Lock’ if Rev1).

16) Repeat Steps 11 through 16


until all chambers requiring
calibration have completed
Calibration and Validation
with the utility.

‘Validate’ only can be


performed on all chambers to
produce a complete
instrument optical report.

Note: Validation QC criteria


for utility is 5% which is
different than TECTA
software at 20%.

Confidential Information - No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the written
consent of Pathogen Detection Systems, Inc.
Page: 7 of 7
Date: 18-Dec-16
Title: SI-16-01-Optical Calibration Instructions Doc Number SI-16-00

17) When complete, Press


‘Report’ button to save a
report of all old and new
calibration values and
validation results to a folder
labelled “reports” on the
Optics Utility USB drive

Reporting is complete once


bottom line reads
“SUCCESSFULLY WROTE
optical_XEND######”.

18) Press ‘Quit’ button to exit


optical utility and TECTA B16
instrument will reboot into
TECTA software

Once TECTA B16 splash


screen displayed, it is safe to
remove USB drive from
instrument.

19) Open the lid and remove the


Validation Cartridge.

Return Validation Cartridge


to its original packaging to
protect it from light exposure.