You are on page 1of 11

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/41969858

Electro-spinning of PLGA/PCL blends for tissue engineering and their


biocompatibility

Article  in  Journal of Materials Science: Materials in Medicine · March 2010


DOI: 10.1007/s10856-010-4048-y · Source: PubMed

CITATIONS READS

143 1,214

2 authors, including:

Thi Hiep Nguyen


Vietnam National University, Ho Chi Minh City
107 PUBLICATIONS   1,696 CITATIONS   

SEE PROFILE

All content following this page was uploaded by Thi Hiep Nguyen on 23 March 2014.

The user has requested enhancement of the downloaded file.


J Mater Sci: Mater Med
DOI 10.1007/s10856-010-4048-y

Electro-spinning of PLGA/PCL blends for tissue engineering


and their biocompatibility
Nguyen Thi Hiep • Byong-Taek Lee

Received: 30 November 2009 / Accepted: 1 March 2010


Ó Springer Science+Business Media, LLC 2010

Abstract In this study, an electro-spun co-polymer biomaterials have been designed to direct the organization,
PLGA/PCL blend was fabricated using various percentages growth, and differentiation of cells in the process of
of PLGA in the blend PLGA/PCL solutions. The PLGA/ forming functional tissue by providing both physical and
PCL ratios used to fabricate the electrospun fibrous mats chemical cues [3]; The involvement of cells in the tissue
were reflected in the FT-IR (Fourier Transform Infrared engineering field include developing novel methods for the
Spectroscopy) data. Experimental results from the MTT proliferation and differentiation of cells. The biomechani-
assay showed that the biocompatibility of the electro-spun cal aspects of tissue engineering design includes properties
co-polymer increased at increasing percentages of PLGA. of native tissues, identification of minimum properties
In vitro cells adhesion and proliferation of fibroblast cells required for engineered tissues, mechanical signals regu-
on electro-spun mats were characterized by SEM mor- lating engineered tissues, and efficacy and safety of engi-
phology. In addition, we found that increasing PLGA neered tissues [4].
concentrations affected the mechanical properties of elec- The electro-spinning method is one of the simplest
tro-spun membranes and increased the biocompatibility of among all methods for the preparation of fibrous mats as
PLGA/PCL electro-spun fibrous mats. extra-cellular matrices. Electro-spinning offers a new
direction for biopolymers since this approach allows one to
decrease the diameter of the pore size from micrometers to
1 Introduction nanometers; thereby, increasing the surface area to volume
or mass ratio and the mechanical properties of the polymer
Tissue engineering is an emerging multidisciplinary field [5]. The advantages of the electro-spinning technique are
involving biology, medicine, and engineering and this field the production of very thin fibers, on the order of few
is likely to revolutionize the ways we improve health by nanometers or micrometers, with large specific surface
restoring, maintaining, or enhancing tissue and organ areas, which allow for easy functionalization for various
function [1]. The foundation of tissue engineering for applications, superior mechanical properties and ease of
diagnostic applications is based on the ability to exploit processing, as suggested by many experts in the tissue
living cells in a variety of ways. Tissue engineering engineering field [6]. Electro-spun scaffolds allow cells to
research includes the following areas: biomaterials, cells, grow while providing sufficient mechanical support. Tan
biomolecules, engineering design aspects, biomechanical et al. reported that the mechanical properties of fibers
aspects of design, informatics to support tissue engineering, increased when the inner diameter of the fiber was
stem cell research, etc. [2]. In terms of biomaterials, novel decreased [7]. The scaffolds currently used for arteries,
however, are still far from optimal in terms of mechanical
endurance and biocompatibility. The structure and prop-
N. T. Hiep  B.-T. Lee (&) erties of electro-spun biopolymer scaffolds are of critical
Department of Biomedical Engineering and Materials,
importance for their capable application in tissue engi-
College of Medicine, Soonchunhyang University,
Cheonan 330-090, Korea neering. It has been generally accepted that the electrospun
e-mail: lbt@sch.ac.kr mats need to be biocompatible, biodegradable with

123
J Mater Sci: Mater Med

degradation or resorbable rates that match the tissue hydrophilicity and lack of surface cell recognition sites
replacement and possess good mechanical properties to [16]. Even though the electrospun PCL mats mimic the
match those of the tissues at the site of implantation. The identity of ECM in living tissues, its poor hydrophilicity
characteristics of electrospun mats include an appropriate caused a reduction in cell adhesion, migration, prolifera-
extracellular matrix structure to facilitate cell adhesion, tion, and differentiation [17, 18]. The advantage of using
proliferation and differentiation, architecture similar to a PCL in this composite is PCL has high mechanical
real tissue made from biodegradable or biocompatible strength, which is a high valuable property for tissue
material, an internal architecture with fully interconnected engineering applications. PLGA was added to the com-
pores and mechanical strength that satisfies physiological posite to support cells attachment, adhesion and cells
requirements. In recent years, many polymers have proliferation on the PLGA/PCL electro-spun mats, which
undergone electro-spinning to fabricate nano-fibers for was previously demonstrated for human mesenchymal
different applications such as artificial skin [8], artificial stem cells and osteoblast cells [19, 20]. The morphology of
blood vessels [9], bone tissue engineering [10], bladder, the fine PLGA/PCL electro-spun fibrous mats was imaged
etc. The polymer combinations used in these processes by SEM. The composition and mechanical properties of
have included the use of a single natural polymer [11], the blend electrospun fibrous mats were characterized by
synthetic polymer [12], copolymer consisting of a natural FT-IR and tensile strength tests, respectively. In vitro cell
and synthetic polymer [13], etc. However, to overcome the adhesion and proliferation were tested by seeding cells for
limitation associated with a single polymer, a new method 60 min, 1 day and 3 days. The MTT assay showed that the
was developed that involves the simultaneous electro- PLGA/PCL electrospun mats were highly biocompatibility.
spinning of a blend of two polymer solutions delivered by
mixing the two polymer then electro-spinning. The blend
electro-spun method not only overcomes the limitations 2 Materials and methods
associated with a single polymer, but also creates a
new component without the need for synthesizing a new 2.1 Materials
co-polymer, which is difficult and complex to achieve
using co electro-spinning of two single biopolymers. In the PLGA (85:15), polycaprolactone (Mn 80 000), tetrahydro-
blend electro-spun method, one can blend the two polymers furan (THF, minimum 99%), Dimethylformamide (DMF,
using an electro-spun 2-eject system, which contains a mix 99%) were purchased from Sigma-Aldrich, USA. Methyl
of the polymers and then electro-spin the mix. For exam- Cloride (MC, SK Chemicals Co., Korea). DMSO
ple, Duan et al. [8] investigated electro-spun PLGA–PVA/ (Dimethylsulfoxide 99, 0%, Samchun Pure Chemical Co.,
chitosan blends for skin applications, where they mixed LTD, Korea), Ethanol (Merck, Germany). Fetal bovine
PVA and chitosan during the first electro-spinning injection serum (FBS), P.S. (penicillin/streptomycin [antibiotics]),
and PLGA was added during the second electro-spinning Dulbecco’s phosphate buffered saline (D-PBS) without
injection. Using this approach, they obtained PLGA-PVA/ calcium or magnesiumand, MTT solution and trypsin–
chitosan mats that had the strong mechanical properties of EDTA were purchased from GIBCO (Carlsbad, CA). The
PLGA, absorptive properties of PVA and antibacterial L-929 cells line was obtained from the ATCC Cell Line
activity of chitosan. (CCL-1TM, NCTC clone 929 [L cell, L-929, derivative of
In this research, electro-spun PLGA/PCL mats were Strain L], Korea). HDMS (Hexamethyldisilazane, Sigma,
fabricated from PLGA (85:15), which is an expensive USA).
synthesis polymer. Therefore, PCL was used as the main
component because it is relatively inexpensive and stable 2.2 Preparation of polymer solutions
in material handling and storage [14]. PCL is a flexible
biopolymer that has been used to overcome the brittle and 10 wt% PLGA was dissolved in THF: DMF (50:50) [8] and
low elongate properties of PLGA [15]. However, PLGA is 12 wt% PCL was dissolved in MC: DMF (20:80) [21].
better than PCL in respect to cell adhesion and proliferation Both solutions were prepared overnight until the solution
due to its more hydrophilic property [15]. Polycaprolactone became homogeneous. Then, several PLGA/PCL azeotrope
(PCL) is hydrophobic and, thus, does not have any physi- solutions were prepared by adding a homogeneous PLGA
ological active sites, which makes PCL unfavorable for cell solution to a homogeneous PCL solution.
growth when it comes into contact with the living body.
Although, when the behavior of epithelial cells was 2.3 Electro-spinning setting
evaluated on a PCL membrane, the PCL membrane was
shown to enhance epithelium growth. However, cell PLGA/PCL blend solutions were placed in a plastic syringe
affinity towards PCL is generally poor due to their low (lure-lock type, 12 ml) and connected through a metal

123
J Mater Sci: Mater Med

syringe needle, which was already connected to a nozzle 2.5 In vitro study
gauge (25 gauge, inner diameter 0.25 mm) on the pump.
The parameters of electrospinning were as follows: the 2.5.1 Sterilization approach and preparation sample
solution flow rate was 0.5 ml/h, a high voltage of 25 kV for in vitro
was supplied directly from a high DC voltage power supply
(NNC-30 kilovolts-2 mA portable type, NanoNC Korea) All electro-spun mats were placed in a vacuum oven for
and the samples were collected through a steel plate 20 cm 48 h at room temperature and sterilized using 70% EtOH
away from the tip of the syringe needle. for 30 min before conducting in vitro testing.

2.4 Characterization
2.5.2 Cell line and maintenance
2.4.1 Morphology
Cell culture studies were carried out with the L929 cell
line obtained from ATCC Cell Line (CCL-1TM, NCTC
The detailed structure of each polymer fibrous mat was
clone 929 [L cell, L-929, derivative of Strain L], Korea).
imaged by scanning electron microscopy (SEM) (SM-65F,
The L-929 cells were subcultured in flasks using RPMI
JEOL, and Japan) at an acceleration voltage of 25 kV. Before
media, supplemented with 10% (v/v) FBS (fetal bovine
SEM observation, all of the samples cut from the electrospun
serum) and 1% PS (penicillin/streptomycin (antibiotics)),
fiber mats (1 cm 9 1 cm) were sputter coated with gold
50 lg/ml ascorbic acid (Sigma) and 10 mM b-glycero-
under a JEOL JFC-1200 fine coater for 60 s. The average
phosphate (Sigma) at 37°C, 95% humidity and 5% CO2
fiber diameter of the electrospun fibers was measured from
(Incubator, ASTEC, Japan). Cells were dissociated with
the SEM images using Adobe Photoshop 5.0 software.
trypsin–EDTA (GIBCO) centrifuged and resuspended in
medium. The culture medium was replaced every 2 days.
2.4.2 FT-IR analyze
Prior to cell culture experiments, 24-well or 96-well tis-
sue culture test plates gamma sterilized (SPL Life Sci-
The electro-spun PCL, PLGA and blend PLGA/PCL nano-
ence, Becton-Dickinson Labware) were prepared. The
fibrous mats were characterized by attenuated reflectance
culture plates were then sprayed with 70% ethanol before
Fourier transform spectroscopy (Spectrum GX, PerkinEl-
being placing under UV light for 30 min to sterilize
mer, USA). The infrared spectra of the samples were
them.
measured over a wavelength range of 4,000–500 cm-1. All
spectra were taken in the spectral range by the accumula-
tion of 64 scans with a resolution of 4 cm-1. 2.5.3 MTT assay/cytotoxicity assay

2.4.3 Mechanical properties The cell samples were washed in PBS 1 time and medium 2
times before the cells were seeded. The cellular viability %
The dimensions of the samples were measure by SEM of fibroblast cells on each polymer fibrous mat was deter-
before measuring the tensile strength (thick ness 100 lm, mined using the Cytotoxicity or MTT (3-[4, 5-dimethyl-
wide 1 mm and length 27 mm). Mechanical characteriza- thiazol-2-yl]-2, 5-diphenyltetrazolium bromide) assays.
tion was achieved by applying tensile test loads to speci- The L-929 mouse fibroblasts cells were seeded in 96-well
mens prepared from the electro-spun ultra fine non-woven tissue culture plates at 1,000 cells/well in 100 ll RPMI.
fiber mats. Since a single polymer nano-fiber is very weak, All media contained 10% FBS and all cell lines were
the resulting non-woven mat was so delicate that any direct incubated overnight. Diluted extract solutions (100 ll) of
contact with the mat surface during manipulation could every fibrous mat at various concentrations (100, 75, 50, 25
damage the fibers. Thus, sufficient care was taken in pre- and 0%) were then added. The L 929 fibroblast cells were
paring and handling the tensile specimens in order to avoid treated for 1, 2, and 3 days and then 20 ll of filtered MTT
severe damage. The specimens were prepared using the solution (5 mg/mL in PBS) was added. After incubation at
following method. First, a white paper frame was cut as 37°C for 3.5 h, the medium was removed from the well and
shown in Fig. 4, and sample (1) was glued (2) onto the top 150 ll of DMSO was added to dissolve any insoluble
and bottom areas of one side. The double-sided tape (3) formazan crystals. The absorbance was measured at
was 2 mm and the hold (4) was used to remove the frame 560 nm using an ELISA reader (Turner Biosystems CE,
from the load-cell. The frame was then glued onto the Promega Corporation, USA) and the cell viability was
topside of the fiber mat, and was cut into rectangular pieces calculated as the percentage relative to the untreated con-
along the vertical lines. trol cells.

123
J Mater Sci: Mater Med

2.5.4 Adhesion, growth behavior of cells including concentration, surface tension and the nature of
the solvents [22]. However, the solubility of the polymer
Sample membranes 20 mm in diameter were sterilized should be examined in more detail when the electro-spun
with 70% EtOH (30 min), then washed with PBS and material is a co-polymer because the uniform electro-spun
suspended in conditioning medium (15 min). The fibro- material just appears as a homogenous dissolution. In this
blast cells (L-929) were cultured on TCPs (Tissue Culture study, we mixed several ratios of PLGA and PCL (10, 20
Polystyrene), control-like and samples. The cells were and 30% PLGA) that were from separate homogenous
seeded at a cell density of 104 cells/cm2 in RPMI and solutions. The mixed polymers were incubated together for
cultured for 1 and 3 days at 37°C in a humidified air 6 h at room temperature to create a homogenous azeotrope.
atmosphere with 5% CO2. After 1 and 3 days of culture, The electro-spun PLGA/PCL mats were then fabricated
cellular constructs were harvested, rinsed twice with PBS using an electro-spinning machine (NNC-30 kV-2 mA
to remove non-adherent cells and subsequently fixed with portable type, Nano NC. Korea.). The parameters used for
2% glutaraldehyde for 15 min and washed with DPBS two the fabrication of each polymer fibrous mats were as fol-
times (15 min/time). The samples were then dehydrated lows: voltage 25 kV, distance 20 cm, flow rate 0.5 ml/h
through a series of graded ethanol (EtOH) solutions (50, and nozzle 25 G. SEM images and the diameter distribu-
75, 90, 95 and 100%) and 5 min was used for every tions of each different type of fiber are shown in Fig. 1a, b
washing time. The cells were then washed three times in and c. Based on the analysis of the SEM images, the
100% EtOH. Finally, the samples were washed two times electro-spun mats were homologous and uniform over a
with HDMS. All samples were air-dried overnight. Dry large area. However, morphology of each sample showed
cellular constructs were sputtered with gold and imaged by differences in fibers; for example, in cases of PLGA/PCL
SEM. (10/90) (Fig. 1a), fibers were not cylindrical and were
attached to each other. Contrary to Fig. 1a, the case of
PLGA/PCL (20/80) (Fig. 1b) showed that electro-spun
3 Results fibers were cylindrical and separated with continuous
fibers. On the other hand, when the concentration of PLGA
3.1 SEM morphology of co-polymer PLGA/PCL was increased to 30, PLGA/PCL (30/70) (Fig. 1c) showed
electro-spun that the fibers were attached and had melted into each
other. Variation in fibers for each sample also changed in
The morphology of electro-spun ultra-fine fibers is influ- cases of PLGA/PCL (20/80), the diameter for over 60% of
enced by various parameters such as the applied voltage, the fibers was around 500 ± 100 nm. For the PLGA/PCL
solution flow rate, distance between capillary and collector, (10/90) electro-spun fibers, the diameter for over 60% of
and especially the properties of the polymer solutions the fibers was around 1,000 ± 200 nm (Fig. 1a). Finally,

Fig. 1 SEM morphology of the PLGA/PCL blend (10/90) (a), (20/80) (b), (30/70) (c) electrospun nano-fibrous mats

123
J Mater Sci: Mater Med

the PLGA/PCL (30/70) electro-spun fibers were attached to contact with the mat surface during manipulation could
each other and the diameter for over 50% of the fibers were damage the fibers. Therefore, the samples were collected
around 2,000 ± 200 nm and the remaining fibers were using alumina foil at a similar polymer concentration, and
very large. then the thickness of each sample was measured by SEM.
These samples were measured, cut carefully and glued on
3.2 FT-IR the frame (Fig. 3), and then the aluminum foil was
removed. The paper frame was cut before initiating the
To determine if the blended PLGA/PCL electro-spun fibers tensile strength measurements. The tensile strengths of the
were successfully fabricated, we analyzed the FT-IR electro-spun PLGA/PCL fiber mat at different PLGA
spectrum of PCL, PLGA, and PLGA/PCL at different concentrations were measured using a tensile tester (Uni-
PLGA/PCL ratios (Fig. 2). The FT-IR spectrum of the versal Testing Machine, R&D Co., Ltd., Korea). Figure 4A
blend PLGA/PCL electro-spun fibrous mats showed that shows the curves of PCL and the PLGA/PCL composites at
the structure of blend electro-spun mats contained all the different PLGA concentrations. In these experiments, we
peaks corresponding to PCL and PLGA. More specifically, found that the stress in the electro-spun PLGA/PCL elec-
all of the PCL peaks in the PLGA/PCL composite over- tro-spun mats increased as the amount of PLGA increased.
lapped with all of the PLGA peaks without the peak at However, the strain did not always increase following an
865 cm-1, which appeared separate from the PCL peaks. increase in the amount of PLGA. In the case of PCL
In this case, the PLGA peaks had large widths but not (Fig. 4Aa) and PLGA/PCL (10/90) (Fig. 4Ab), the stress
depths while the PCL peaks had large depths but not increased from 1.8 ± 0.05 to 2.5 ± 0.05 MPa and the
widths. In addition, the amount of PLGA was smaller than strain increased from 130 ± 5 to 134 ± 5 (%). However,
the amount of PCL. Therefore, a larger PCL width in the this same trend was not observed when the PLGA con-
blend electro-spun PLGA/PCL fibers was expected. Fig- centration was increased from 10% [PLGA/PCL (10/90)]
ure 2 also demonstrates that no reaction between PCL and (Fig. 4Ab) to 20% [PLGA/PCL (20/80)] (Fig. 4Ac). In this
PLGA occurred; these fibers were just blended composites. scenario, the stress increased from 2.5 ± 0.5 MPa to
Almost peaks of the blend PLGA/PCL electro-spun fibrous 6.1 ± 0.5 MPa but the strain decreased from 134 ± 5 to
mats overlapped PCL peaks, it is easy to confuse but the 88 ± 5 (%). When the PLGA concentration was increased
peaks at 865 cm-1 which asserted PLGA present. How-
ever, the intensity of this peak increased with increasing
amounts of PLGA. The combined results demonstrate that
the blend PLGA/PCL electros-pun fibrous mats were suc-
cessfully fabricated.

3.3 Mechanical property

Since a single polymer nano-fiber is very weak, the Fig. 3 A schematic of the paper frame used to prepare the tensile
resulting non-woven mat was so delicate that any direct specimens from the electro-spun non-women mats

Fig. 4 A Stress–strain curves of electrospun PCL (a), PLGA/PCL


Fig. 2 FT-IR curves of electrospun PLGA (a), PCL (b), PLGA/PCL (10/90) (b), PLGA/PCL (20/80) (c), PLGA/PCL (30/70) (d). B
(10/90) (c), (20/80) (d), (30/70) (e) Variation of tensile strength of PCL and PLGA/PCL blend

123
J Mater Sci: Mater Med

from 20% [PLGA/PCL (20/80)] (Fig. 4Ac) to 30% [PLGA/


PCL (30/70)] (Fig. 4Ad) the strain and stress both suddenly
decreased from 6.1 ± 0.5 to 3.8 ± 0.5 MPa and 88 ± 5
(%) (strain) to 59 ± 5 (%), respectively. The variations in
tensile strength, which contains more detail on the rela-
tionship between the stress and strain of each mat, is shown
in Fig. 4B. As shown in Fig. 4Ad, the stress and strain
were not high even though the amount of PLGA was high.
These results most likely occurred because the electro-spun
fibers containing 30% PLGA were not fine as the PLGA/
PCL (20/80) and PLGA/PCL (10/90) fibers. In addition,
PLGA is a very brittle material [15], which can easily
break if the stress increases. Duan et al. [8]. reported a Fig. 5 MTT cytotoxicity of the blend electro-spun PLGA/PCL mats
stress of 9 MPa and strain of 30% for only PLGA elec- at different PLGA/PCL ratio
trospun mats. For this reason, the mechanical properties of
the electrospun PLGA/PCL mats increased at higher
amounts of PLGA. Based on the combined results pre- 3.5 Cellular proliferation
sented above, we concluded that the stress of the PLGA/
PCL fibers was dependent on both the amount of PLGA The proliferation of fibroblast cells on the blend electro-
and the uniformity of the electro-spun fibers on the mats. spun PLGA/PCL fibrous mats was evaluated by the
Because the fibers of the electro-spun PLGA/PCL (30/90) reduction of MTT to formazan by mitochondrial succinate
aggregated with each other and the composite was not dehydrogenase in complex II (succinate–ubiquinone oxi-
uniform, the stress was lower than the PLGA/PCL (20/80) doreductase complex) [23]. The level of reduction of
mats. The co-polymer containing 10 wt% PLGA had a MTT into formazan reflects the level of cell metabolism.
strain value that was higher than the composites not con- Figure 6 compares the time course of formazan absor-
taining PLGA. The stress–strain data showed that PCL was bance of blend PLGA/PCL nanofiber mats fabricated with
a flexible polymer and if it was added to brittle PLGA different concentrations of PLGA. Biomimetic tissue
materials (non-flexible polymer) in small amounts, the formed on the blend PLGA/PCL nanofibrous mats, which
mechanical strength under fine fibrous mats conditions was affected the viability and proliferation of cells. Cell
increased. attachment and proliferation on the electro-spun PLGA/
PCL (10/90) co-polymer doubled between day 1 and day
3.4 Cytotoxicity results 3. Moreover, the absorbance values showed that cell
attachment and proliferation increased with an increase
The cytotoxicity of PLGA/PCL nanofiber mats was tested of PLGA concentration in the PLGA/PCL composite;
by quantitative analysis using the MTT test. After allowing however, between PLGA/PCL (20/80) and PLGA/PCL
the cells to spread for 3 days at the various extract dilutions (30/70), a slight increase was observed. According to
(0, 25, 50, 75 and 100%), the proliferation of L-929 cells the results, we concluded that electro-spun PLGA/PCL
was measured by the MTT assay. In the case of the control, (20/80) had the greatest effect on ECM production when
the extract solution containing cultured cells was not compared to PLGA/PCL (30/70) and PLGA/PCL (10/90)
added. Absorbance values were used to measure cellular (Fig. 6).
proliferation. The extracts containing the blend electrospun
PLGA/PCL nanofiber mats did not inhibit cell metabolism
relative to the control group. We found that cell metabo-
lism of the blend electro-spun PLGA/PCL nano-fibrous
increased with an increase in the PLGA concentration.
Based on the quantitative scores, it was concluded that the
extracts of the nanofibers mats displayed no cytotoxic
reactivity in this test, as shown in Fig. 5. In addition, nei-
ther cell death nor morphological disorder was observed
throughout the incubation period. Thus, it was concluded
that PLGA/PCL nano-fibrous mats produced from this
process displayed no cytotoxic effect due to the presence of Fig. 6 Cellular proliferation of the blend electro-spun PLGA/PCL of
remnant solvents. fibroblast cells L-929 after seeding the cells for 1 day and 3 days

123
J Mater Sci: Mater Med

3.6 Cell attachment and cell proliferation (Fig. 7a3, b3, c3), which shows more detail, the L-929
cells were shown to be well attached to the PLGA/PCL
3.6.1 One cell morphology-cell attachment fiber. Therefore, the PLGA/PCL electro-spun fibrous
has high biocompatibility that is suitable for biomaterial
The cell morphology was examined to continually check applications.
the biocompatibility of all PLGA/PCL blend membranes.
Figure 7 shows the cell morphology of one fibroblast 3.6.2 Cell proliferation
L-929 after seeding the cell for 1 h on a tissue culture plate
(TCPs) (Fig. 7a1–a3). Figure 7a2, a3 shows an image of To test cell proliferation on the PLGA/PCL electro-spun
cell attachment on the PLGA/PCL (10/90) electro-spun fibrous mats, we seeded the cells at the same density and
mat, while Fig. 7b2, 3 shows an image of cell attachment incubated the samples for different time periods (short-
on the PLGA/PCL (20/80) electro-spun mat. Finally, time). Figure 8 shows the cell proliferation after seeding
Fig. 7c2, c3 shows an image of cell attachment on the the cells for 1 day, while Fig. 9 shows the cell proliferation
PLGA/PCL (30/70) electro-spun mat. By comparing the after seeding the cells for 3 days. After 1 and 3 days of cell
filopodium and lamellipodium in the control sample with seeding the cells were spread on all scaffolds. In Fig. 8, the
each condition, we found that filopodium and lamellipo- images showed that the fibroblast cells were spread on the
dium growth was as good as the control. However, by TCP and PLGA/PCL fibrous mats after seeding the cells
comparing (Fig. 7a2, b2, c2), we found that filopodium and for 1 day. These results demonstrated cellular attachment
lamellipodium growth increased as the content of PLGA and proliferation of mouse fibroblast L-929 cells under
increased. On the other hand, in the enlarged image static seeding on all PLGA/PCL electrospun mats.

Fig. 7 Cell morphology of fibroblast cell growth on control (a1, b1, c1), blend PLGA/PCL (10/90) (a2, a3), PLGA/PCL (20/80) (b2, b3) and
PLGA/PCL (30/70) (c2, c3) electro-spun after seeding cell 1 h

123
J Mater Sci: Mater Med

Fig. 8 SEM morphology of fibroblast cells grown on the control (a), blend PLGA/PCL (10/90) (b), PLGA/PCL (20/80) (c) and PLGA/PCL
(30/70) (d) electro-spun after seeding cell 1 day

However, cell proliferation on the samples [Fig. 8b–d] was PCL co-polymer is a good composite for tissue engineering
clearly better than the TCP surface [Fig. 8a] as evidenced applications.
by the development of lamellopodia, which is an indicator
of cell attachment to the surface. In addition, after 1 day of
seeding, cells were well grown and had spread uniformly 4 Discussion
and extensively, covering the surfaces of the biomaterials,
which was similar to the control sample. The presence of In this study, we investigated a new component for
lamellapodia and an interlaced fibrous network demon- biomaterials and tissue engineering applications; the elec-
strated that cell adhesion was very good. Cell spread- tro-spun PLGA/PCL blend. We fabricated PLGA/PCL
ing increased on the PLGA/PCL electrospun mats when electro-spun mats by combining different PLGA and PCL
the PLGA concentration was increased (Figs. 8b and ratios into a single PLGA/PCL component. Calvert et al.
8 [c, d]). [24] and Lee et al. [25] reported that PCL should not be used
After 3 days (Fig. 9), cells completely covered the at high percentages in PLGA/PCL composites due to its
electro-spun mats. To assess cell proliferation and cells high cost and stability during handling and storage [14].
attachment on the surface, we took images of each sample at Therefore, we combined PCL with different ratios of PLGA
different magnifications. At the highest magnification to fabricate PLGA/PCL electro-spun fibrous mats. On the
(Fig. 9a3, b3, c3), we found that cell proliferation increased other hand, Calvert et al. [24] also reported that a high
at higher PLGA concentrations. In addition, cell attachment percentage of PLGA was not always as good as a low
was more clearly observed on the PLGA/PCL (30/70) percentage; therefore, we only used a maximum PLGA
electro-spun fibrous mat. Even though the incubation time concentration of 30% in the PLGA/PCL blend. The SEM
was just 3 days, the shape of the fibroblast cells changed morphology shown in Fig. 1 clearly demonstrates that
from round to elongated (Fig. 9a3, b3, and c3). Based on PLGA concentration was very important in obtaining fine
these results, we can conclude that the electro-spun PLGA/ PLGA/PCL fibers mats. The PLGA/PCL electro-spun fiber

123
J Mater Sci: Mater Med

Fig. 9 SEM morphology of fibroblast cells grown on blend PLGA/PCL (10/90) (a1, a2, a3), PLGA/PCL (20/80) (b1, b2, b3) and PLGA/PCL
(30/70) (c1, c2, c3) electro-spun after seeding the cells for 3 days

for PLGA/PCL (10:90) was very large and its mechanical strength because this fibrous mat was uniform and small in
strength was much higher than the composite not containing diameter, both of which are important factors dictating the
PLGA. However, at 20% PLGA, the PLGA/PCL electro- mechanical properties of electrospinning [7, 26]. The FT-IR
spun blend contained very fine fibers with small, smooth, curves showed that PLGA/PCL electro-spun mats were
cylindrical, and humongous fibers, at this composition the successfully fabricated using the PLGA/PCL blend due to
stress of the electro-spun mat was very high but the strain the presence of the 865 cm-1 peak. The PLGA/PCL elec-
was lower relative to the PLGA/PCL (10/90) electro-spun tro-spun mat was shown to not be toxic at all compositions
mat. As for the PLGA/PCL (30/70) composite, the high (Fig. 5). In addition, cell proliferation of fibroblast doubled
PLGA concentration changed the heat of fusion, which between day 1 and day 3 day (Fig. 6) through MTT results.
affected the blend solution, resulting in a morphology that On the other hand, the cells became attached to the PLGA/
showed a blend of PLGA/PCL (30/70) electro-spun fibers PCL electro-spun mats within 60 min (Fig. 7). The SEM
were larger. The morphology of PLGA/PCL (30/70) elec- images showed that filopodium and lamellipodium growth
tro-spun mat is a key to make the mechanical strength the on the electro-spun mats was as good as the control. But, the
PLGA/PCL (30/70) electro-spun mats were lower than the cell proliferation and attachment was better as the PLGA
PLGA/PCL (20/80) electro-spun mat. Lee et al. [25] fab- concentration in the PLGA/PCL electro-spun mats was
ricated PCL electro-spun nonwoven mats from PCL in a increased after seeding the cells for 1 day (Fig. 8) and
DMF: MC (25/75) solution, which displayed a strain of over 3 days (Fig. 9). However, in the case of PLGA/PCL (30/
200% and stress of approximately 1.7 MPa. In this study, 70), the cells did plate on the electro-spun mat but the cells
the PCL electro-spun mat was fabricated from PCL in a shrunk. These combined results clearly indicate that the
DMF: MC (80/20) solution and had a similar stress but presence of the PLGA support in the PLGA/PCL electro-
lower strain (130%). The PLGA/PCL (20/80) electro-spun spun mats improved biocompatibility, cell attachment and
mat showed the best mechanical strength and modulus cell proliferation.

123
J Mater Sci: Mater Med

5 Conclusion 11. Lee KY, Jeong L, Kang YO, Lee SJ, Park WH. Electro-spinning
of polysaccharides for regenerative medicine. Adv Drug Deliv
Rev. 2009;61:1020–32.
In this study, we successfully fabricated electro-spun 12. Bini TB, Gao S, Tan TC, Wang S, Lim A, Hai LB, et al. Elec-
PLGA/PCL mats containing different concentrations of trospun poly(L-lactide-co-glycolide) biodegradable polymer
PLGA. PCL was used as the major material because it is nanofibre tubes for peripheral nerve regeneration. Nanotechnol-
inexpensive and stable during handling and storage. PCL is ogy. 2004;15:1459–64.
13. Bajgai MP, Aryal S, Bhattarai SR, Bahadur KCR, Kim KW, Kim
a flexible material and PLGA is a brittle material; thus, the HY. Polycaprolactone grated dextran biodegradable electrospun
PLGA/PCL electro-spun mats had a stronger mechanical matrix: a novel for tissue engineering. J Appl Polym Sci. 2008;
strength than PCL. The addition of PLGA as a doping 108:1447–54.
agent in the PLGA/PCL electro-spun mats increased the 14. Tang ZG, Callaghan JT, Hunt JA. The physical properties and
response of osteoblasts to solution cast films of PLGA doped
value of this composite for biomaterial and tissue engi- polycaprolactone. Biomaterials. 2005;26:6618–24.
neering applications because it increased biocompatibility, 15. Kim JY, Cho DW. Blended PCL/PLGA scaffold fabrication
cell attachment and cell proliferation. We investigated a using multi-head deposition system. Microelectron Eng. 2009;86:
new electro-spun component and demonstrated that it holds 1447–50.
16. Ciardelli G, Chiono V, Vozzi G, Pracella M, Ahluwalia A,
promise for use as a biomaterial for tissue engineering and Barbani N, et al. Blends of poly(3-caprolactone) and polysac-
biomedical applications. charides in tissue engineering applications. Biomacromolecules.
2005;6:1961–76.
17. Kim CH, Khil MS, Kim HY, Lee HU, Jahng KY. An improved
hydrophilicity via electrospinning for enhanced cell attachment
and proliferation. J Biomed Mater Res B. 2006;78B:283–90.
References 18. Li WJ, Cooper JA, Mauck RL, Tuan RS. Fabrication and charac-
terization of six electrospun poly(a-hydroxyester)-based nanofi-
1. Vunjak-Novakovic G, Freshney R. Culture of cells for tissue brous scaffolds for tissue engineering applications. Acta Biomater.
engineering. New York: Wiley; 2006. 2006;2:377–85.
2. Rickne A. Swedish possibilities within tissue engineering and 19. Curran JM, Tang Z, Hunt JA. PLGA doping of PCL affects the
regenerative medicine. VINNOVA—Swedish Governmental plastic potential of human mesenchymal stem cells, both in the
Agency for innovation systems/Verket för Innovations system; presence and absence of biological stimuli. J Biomed Mater Res
2009. A. 2008. doi:10.1002/jbm.a.31966.
3. Segvich S, Kohn DH. Biological interactions on materials sur- 20. Tang ZG, Hunt JA. The effect of PLGA doping of polycapro-
faces understanding and controlling protein, cell, and tissue lactone films on the control of osteoblast adhesion and prolifer-
responses. New York: Springer; 2009. ation in vitro. Biomaterials. 2006;27:4409–18.
4. Dawson E, Mapili G, Erickson K, Taqvi S, Roy K. Biomaterials 21. Khil MS, Shanta RB, Kim HY, Kim SZ, Lee KH. Novel fabri-
for stem cell differentiation. Adv Drug Deliv Rev. 2008;60:215– cated matrix via electrospinning for tissue engineering. J Biomed
28. Mater Res. 2005;72B:117–24.
5. Supaphol P, Chuangchote S. On the electrospinning of poly(vinyl 22. Deitzel JM, Kleinmeyer J, Harris D, Tan NCB. The effect of
alcohol) nanofiber mats: a revisit. J Appl Polym Sci. 2008;108: processing variables on the morphology of electrospun nanofibers
969–78. and textiles. Polymer. 2001;42:261–72.
6. Agarwal S, Wendorff JH, Greiner A. Use of electro-spinning 23. Karakecili AG, Satriano C, Gümüşderelioğlu M, Marletta G.
technique for biomedical applications. Polymer. 2008;49:5603– Enhancement of fibroblastic proliferation on chitosan surfaces by
21. immobilized epidermal growth factor. Acta Biomater. 2008;4:
7. Tan EPS, Ng SY, Lim CT. Tensile testing of a single ultrafine 989–96.
polymeric fiber. Biomaterials. 2005;26:1453–6. 24. Jay W, Calvert KGM, Lisa C, Prashant NK, Paul AD, Lee EW.
8. Duan B, Yuan X, Zhu Y, Zhang Y, Li X, Zhang Y, et al. A Characterization of osteoblast-like behavior of cultured bone
nanofibrous composite membrane of PLGA-chitosan/PVA pre- marrow stromal cells on various polymer surfaces. J Biomed
pared by electrospinning. J Eur Polym. 2006;42:2013–22. Mater Res. 2000;52A:279–84.
9. Vaz CM, Tuijil SV, Bouten CVC, Baaijens FPT. Design of 25. Lee KH, Kim HY, Khil MS, Ra YM, Lee DR. Characterization of
scaffolds for blood vessel tissue engineering using a multi-lay- nano-structured poly(caprolactone) nonwoven mats via electros-
ering electrospinning technique. Acta Biomater. 2005;1:575–82. pinning. Polymer. 2003;44:1287–94.
10. Zhang Y, Venugopal JR, El-Turki A, Ramakrishna S, Su B, 26. Li Y, Huang Z, Lu Y. Electro-spinning of nylon-6,66,1010 ter-
Lim C. Electrospun biomimetic nanocomposite nanofibers of polymer. Eur Polym J. 2006;42:1696–704.
hydroxyapatite/chitosan for bone tissue engineering. Biomateri-
als. 2008;29:4314–22.

123

View publication stats

You might also like