Professional Documents
Culture Documents
ANALYZER / CONTROLLER
START-UP BY ____________________________________
ORDERING INFORMATION
In order for us to fill your order immediately and correctly, please order material by description and part number, as shown
in this book. Also, please specify the serial number of the equipment on which the parts will be installed.
WARRANTY
Seller warrants for a period of one year after shipment that the equipment or material of its manufacture is free from defects
in workmanship and materials. Corrosion or other decomposition by chemical action is specifically excluded as a defect
covered hereunder, except this exclusion shall not apply to chlorination equipment. Seller does not warrant (a) damage
caused by use of the items for purposes other than those for which they were designed, (b) damage caused by unauthorized
attachments or modifications, (c) products subject to any abuse, misuse, negligence or accident, (d) products where parts
not made, supplied, or approved by Seller are used and in the sole judgement of the Seller such use affects the products’
performance, stability or reliability, and (e) products that have been altered or repaired in a manner in which, in the sole
judgement of Seller, affects the products’ performance, stability or reliability. SELLER MAKES NO OTHER WARRANTY OF
ANY KIND, AND THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING ANY WARRANTY OF MERCHANTABILITY OR OF FITNESS OF THE MATERIAL OR EQUIPMENT FOR
ANY PARTICULAR PURPOSE EVEN IF THAT PURPOSE IS KNOWN TO SELLER. If Buyer discovers a defect in mate-
rial or workmanship, it must promptly notify Seller in writing; Seller reserves the right to require the return of such defective
parts to Seller, transportation charges prepaid, to verify such defect before this warranty is applicable. In no event shall
such notification be received by Seller later than 13 months after the date of shipment. No action for breach of warranty
shall be brought more than 15 months after the date of shipment of the equipment or material.
LIMITATION OF BUYER’S REMEDIES. The EXCLUSIVE REMEDY for any breach of warranty is the replacement f.o.b.
shipping point of the defective part or parts of the material or equipment. Any equipment or material repaired or replaced
under warranty shall carry the balance of the original warranty period, or a minimum of three months. Seller shall not be
liable for any liquidated, special, incidental or consequential damages, including without limitation, loss of profits, loss of
savings or revenue, loss of use of the material or equipment or any associated material or equipment, the cost of substitute
material or equipment, claims of third parties, damage to property, or goodwill, whether based upon breach of warranty,
breach of contract, negligence, strict tort, or any other legal theory; provided, however, that such limitation shall not apply
to claims for personal injury.
Statements and instructions set forth herein are based upon the best information and practices known to Siemens Water
Technologies Corp., but it should not be assumed that every acceptable safety procedure is contained herein. Of necessity
this company cannot guarantee that actions in accordance with such statements and instructions will result in the complete
elimination of hazards and it assumes no liability for accidents that may occur.
Water Technolgies
1901 West Garden Road, Vineland, NJ 08360
WT.050.580.0AA.UA.IM.1206
MFC ANALYZER / CONTROLLER
INTRODUCTION
WT.050.580.0AA.UA.IM.1206 Introd.
MFC ANALYZER / CONTROLLER
Intended Use
The MFC may only be used in buildings and under the operating condi-
tions described inthe technical area. The device is not designed for any
application other than that described in this manual. Compliance with the
intended use of this device also includes reading this operating manual
and observance of all instructions which it contains, particularly the safety
instructions. Furthermore, all inspection and maintenance work must be
performed at the prescribed intervals. If the device is not employed in
accordance with its intended use, safe and reliable operation cannot be
guarunteed.
Table Of Contents
This page provides very important safety information related to safety in installation, operation, and
maintenance of this equipment.
WARNING
ALL USERS OF THIS EQUIPMENT SHOULD BE MADE AWARE OF THE PROBLEMS ASSO-
CIATED WITH HANDLING HAZARDOUS MATERIALS IN EITHER LIQUID OR GASEOUS
FORM AND OF THE EFFECTS OF EXPOSURE TO THEIR FUMES. REFERENCE SHOULD BE
MADE TO THE LITERATURE AVAILABLE FROM THE SUPPLIERS OF THESE CHEMICALS,
PARTICULAR ATTENTION BEING PAID TO THE INFORMATION AND ADVICE ON PROTEC-
TIVE CLOTHING.
WT.050.580.0AA.UA.IM.1206 SP-1
MFC ANALYZER / CONTROLLER
NOTE
Minor part number changes may be incorporated into Wallace & Tiernan® (W&T) products from time
to time that are not immediately reflected in the instruction book. If such a change apparently has been
made in your equipment and does not appear to be reflected in your instruction book, contact your local
W&T sales office for information.
Please include the equipment serial number in all correspondence. It is essential for effective commu-
nication and proper equipment identification.
WT.050.580.0AA.UA.IM.1206 SP-2
MFC ANALYZER / CONTROLLER
REGIONAL OFFICES
Direct any questions concerning this equipment that are not answered in the instruction book to the
Reseller from whom the equipment was purchased. If the equipment was purchased directly from
Siemens Water Technologies Corp., Vineland, NJ, contact the office indicated below.
UNITED STATES
CANADA
If the equipment was purchased directly from Siemens Water Technologies Canada, Inc., contact the
nearest office indicated below.
ONTARIO QUEBEC
MEXICO
If the equipment was purchased directly from Mexico, contact the office indicated below.
WT.050.580.0AA.UA.IM.1206 1.010-1
MFC ANALYZER / CONTROLLER
List of Contents
PARA. NO.
WT.050.580.0AA.UA.IM.1206
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206
MFC ANALYZER / CONTROLLER
Housing
Dimensions (W x H x D): 8.5” x 14.8” x 6.1”
(215mm x 375mm x 155mm)
Weight: approx. 3.3 Lbs (1.5 kg)
Connections
Sample water: PVC hose 1/4” OD x 3/16” ID
or 1/4” OD x 1/16” ID
Thread connection: 1/2” NPT
Flow control valve
Flow rate: approx. 33 l/h (controlled)
Control range: 2.9 - 58 PSI
Back-pressure: Open drain connection
Sample Water
Temperature: max. + 122 °F
Conductivity: > 200 µS/cm
Multi sensor
Switching point: 21 l/h ± 3 l/h
Swicthing hysteresis: 2 l/h
Temperature sensor: PT 1000
WT.050.580.0AA.UA.IM.1206
MFC ANALYZER / CONTROLLER
Sample water
Back-pressure
temperature
pH/mV Open drain connection max. + 122 °F
Fluoride Open drain connection max. + 122 °F
Conductivity Open drain connection max. + 122 °F
pH electrode
Max. measuring range: pH 0 - 12
Operating temperature range: 32 °F - 122 °F
Storage temperature range: 23 °F - 86 °F
Sample water conductivity: 200 µS/cm - 200 mS/cm
Redox electrode
Max. measuring range: -1000 - +1000 mV
Operating temperature range: 32 °F - 122 °F
Storage temperature range: 23 °F - 86 °F
Sample water conductivity: 200 µS/cm - 200 mS/cm
Fluoride electrode
Max. measuring range: 0.2 - +1000 mV
Operating temperature range: 32 °F - 122 °F
Storage temperature range: 23 °F - 86 °F
Sample water conductivity: 200 µS/cm - 200 mS/cm
Conductivity electrode
Measuring system: 4 electrode system LF325
Accuracy: <= 1.5 % FS
(incl. reproducibility)
Principle of operation: 4 conductor measurement
Operating temperature range: 23 °F - 194 °F
Storage temperature range: 23 °F - 194 °F
Response time: < 20 seconds
Cell constant: 0.48 cm-1 ± 1.5 %
WT.050.580.0AA.UA.IM.1206
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206
MFC ANALYZER / CONTROLLER
1.5 Design
The MFC unit is a modular design and can be equipped with up to four
sensor measuring modules. The modules can be different or the same.
This component arrangement determines the version and number of flow
block assemblies. It is also possible to install several different versions
of flow block assemblies.
A B C
WT.050.580.0AA.UA.IM.1206
MFC ANALYZER / CONTROLLER
• Motherboard with power unit and five slots for the sensor measuring
module cards
• Relay card
• Plug-in card with analog outputs
• Terminal strip
• Glands for the cables of the sensor measuring modules
• RS232 interface
• Cable unions
A Plastic housing B
H
B Slots
C Relay card
D Terminal strip
E Cable union
C
F RS232 interface
D
G Sensor measuring module
housing duct and glands
H Analog outputs
E
G F
WT.050.580.0AA.UA.IM.1206
MFC ANALYZER / CONTROLLER
• Graphic display
• Operator controls
• A green LED to indicate the power supply
• Eight red LEDs to indicate the operating conditions
A Cover B
E
B Green LED (Power) C
C Red LED (operating and
control indicator)
D
D Operator controls (keys)
E Graphic display
Two clips are installed in the housing cover. These are used to facilitate
pH and fluoride calibration. Insert these clips into the rear panel of the
housing. Insert one clip into the upper catch to attach the sensor. Insert
the second clip into the center catch inorder to measure the buffer or
calibration solution using the plastic cups provided.
WT.050.580.0AA.UA.IM.1206 10
MFC ANALYZER / CONTROLLER
E
G
Two clips are installed in the housing cover. These are used to facilitate
pH and fluoride calibration. Insert these clips into the rear panel of the
housing. Insert one clip into the upper catch to attach the sensor. Insert
the second clip into the center catch inorder to measure the buffer or
calibration solution using the plastic cups provided.
WT.050.580.0AA.UA.IM.1206 11
MFC ANALYZER / CONTROLLER
E
G
WT.050.580.0AA.UA.IM.1206 12
MFC ANALYZER / CONTROLLER
D
A Sensor cable
B Housing cable gland
C Sensor B
D Plug-in card
WT.050.580.0AA.UA.IM.1206 13
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 14
MFC ANALYZER / CONTROLLER
Section 2 - Installation
List of Contents
PARA./DWG. NO.
WT.050.580.0AA.UA.IM.1206 15
MFC ANALYZER / CONTROLLER
The device must be protected against rain, frost and direct sunlight and
may not be installed outdoors. It must be mounted horizontally on a flat
wall with an ambient temperature of 32 to 122°F. The air in the room
should be non-condensing.
2. Loosen the five screws on the cover of the MFC electronic module.
3. Carefully remove the cover of the MFC electronic module and leave
to hang on the strain relief cables.
1. Fasten the mounting rail to the wall, using screws and inserts compat-
able with wall material and mounting rail.
2. Hook the electronic module onto the mounting rail so that it is flush
to the right and fasten to the wall with a screw.
3. Hook the DEPOLOX® 5 flow block assembly onto the mounting rail
to the left of the MFC electronic module and fasten to the wall with
two screws.
NOTE: If the flow block assembly is not mounted directly next to the
MFC, it can also be mounted without the mounting rail.
If the electronic module and the DEPOLOX® 5 flow block assembly are
to be mounted in different places, the modules can be hooked onto suit-
able tallow-drop screws by the top holding fixtures instead of onto the
mounting rail. Proceed with the installation as described above.
WT.050.580.0AA.UA.IM.1206 16
MFC ANALYZER / CONTROLLER
NOTE: If the electronic module and the flow block assembly are
mounted at separate locations, the Wallace & Tiernan® sensor cable
extensions with a maximum length of 50m must be used. An impedance
converter for the Redox, fluoride and pH sensors is also required .
WT.050.580.0AA.UA.IM.1206 17
MFC ANALYZER / CONTROLLER
2.2 Start-Up
Once the device is mounted, connect to the power source. Insert the sensor
measuring modules (if not factory installed) and select the application.
The cable descriptions with the numbers 1–4 must be attached to the
sensor cables according to the selected slot. The cable descriptions are
included with the sensor measuring modules and ensure that cables will
not be confused while calibrating multiple similar measurements (e.g. 2
pH measurements).
WT.050.580.0AA.UA.IM.1206 18
MFC ANALYZER / CONTROLLER
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
Soft Keys
� � �
WT.050.580.0AA.UA.IM.1206 19
MFC ANALYZER / CONTROLLER
If module 1 is available:
1 Set dosing output, and adjust posi- Display
tioner running time, Tp, max. pulses if 1.1.1
necessary
2 Calibrate “Ym” on positioner with Display
feedback 1.1.2
3 Select control mode Display
1.1.2
4 Check setpoint and dosing factor, ad- Display
just if necessary 1.1.2
5 Check setpoint and dosing source, Display
adjust if necessary 1.1.2
6 Check flow rate source, adjust if nec- Display
essary 1.1.3
7 Check flow rate direction, adjust if Display
necessary 1.1.3
8 Check control variable 2, adjust if Display
necessary (ratio control only) 1.1.3
9 Check X direction, adjust if necessary Display
(ratio control only) 1.1.3
10 Check X factor, adjust if necessary Display
(ratio control only) 1.1.3
11 Adjust values for Xp and Tn on con- Display
trol loop (single feedback closed-loop 1.1.4
control only)
WT.050.580.0AA.UA.IM.1206 20
MFC ANALYZER / CONTROLLER
If module 2 is available:
1 Set dosing output, and adjust position- Display
er running time, Tp, and max. pulses if 1.2.1
necessary
2 Select control mode Display
1.2.2
3 Check setpoint and dosing factor, ad- Display
just if necessary 1.2.2
4 Adjust values for Xp and Tn on con- Display
trol loop 1.2.3
WT.050.580.0AA.UA.IM.1206 21
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 22
MFC ANALYZER / CONTROLLER
2. Fill half a cap from the plastic bottle with grit and pour it into the cell
body (approx. 1/2 cm³ grit).
NOTE: Make sure that the opening is clean; rinse off with distilled
water, if necessary.
WT.050.580.0AA.UA.IM.1206 23
MFC ANALYZER / CONTROLLER
D C B C B D A
Depolox 5 VariaSens
Keep the dust protection caps and watering caps of the sensors for sub-
sequent use.
Cable extension:
The sensor cable for free chlorine, conductivity and total chlorine may
be extended to a max. of 50 m.
If the pH, Redox or fluoride sensor cables must be extended (max. 50 m),
an impedance converter must be attached to the sensor. The impedance
converter converts the very high-resistance sensor signal into a low-resis-
tance signal. The impedance converter is supplied by an installed battery.
The life of the battery is approx. 5 years; the impedance converter should
be sent to the factory for battery replacement.
WT.050.580.0AA.UA.IM.1206 24
MFC ANALYZER / CONTROLLER
AB CD E F G H
WT.050.580.0AA.UA.IM.1206 25
MFC ANALYZER / CONTROLLER
4. Fit the supplied glands to ducts which are not in use in order to seal
the housing.
WT.050.580.0AA.UA.IM.1206 26
MFC ANALYZER / CONTROLLER
• Motherboard with power unit and five slots for the sensor measuring
module cards.
• Relay card
• Plug-in card with analog outputs
• Terminal Strip
• Glands for the sensor measuring module cables
• RS232 interface
• Cable unions
A
B
I
G D
H F
A Plastic housing
B Slots
C Relay card
D Terminal strip
E Cable union
F RS232 cable
G Gland
H Seansor measuring module housing
duct with glands
I Analog outputs
WT.050.580.0AA.UA.IM.1206 27
MFC ANALYZER / CONTROLLER
A B C
A Sample water inlet with ball valve
B Drain on the drain screw
C Sample water outlet
3. To prevent long loop lag times, ensure that the pipes in the sample
water inlet are as short as possible.
4. An external strainer with a mesh width of 0.02 inch is provided for
the sample water inlet.
WT.050.580.0AA.UA.IM.1206 28
MFC ANALYZER / CONTROLLER
D
C
A Union nut
B Hose
C Locking ring
D Hose bushing
3. Push the locking ring out until the union nut engages the connecting
threads.
A B
�
�
WT.050.580.0AA.UA.IM.1206 30
MFC ANALYZER / CONTROLLER
• 24 V DC
• 200–240 V AC, 50–60 Hz (switchable)
• 100–120 V AC, 50–60 Hz
WT.050.580.0AA.UA.IM.1206 31
MFC ANALYZER / CONTROLLER
Relay
Card
WT.050.580.0AA.UA.IM.1206 32
MFC ANALYZER / CONTROLLER
• Version
• MFC
• Analog scan
• Wallace & Tiernan GmbH
� � � ��� � � � ��� � � � � �� � �� � � � � � � � � �
� � �� � ���������� � � � � � � � �� � �� � �� � � � �� � ��
� � �� � �� � ���� � �� � �� � � � �� � ��� � � � � � � ��� �
� � � �� � �� � � �
� �� ��� �� � � � � �� � �� � � � � �� �� �
� �� ��� �� � � � ��� � � �� � � �� �� �
� �� ��� �� � � � ��� � � �� � � �� �� �
� �� ��� �� �
� �� ��� �� �
� �� ��� �� � �
� � ��� � � �� �� �� � � � � �� � � �
� � ��� � � �� � �� � �� �
� �� � � � � �� � � � � � ��
� � ���� � � � � � � �� � � �
WT.050.580.0AA.UA.IM.1206 33
MFC ANALYZER / CONTROLLER
1. Starting from the basic display, restart the system by selecting “Reset”
under the “Select” and “System” menus, and then selecting “System
Restart” ... “Yes”.
“RESET CPU” and then “MFC” appears.
In the following display “Analog Scan”, press both the “left” and the
“right” arrow at the same time for at least two seconds.
“Wallace & Tiernan GmbH” then appears.
� � ��� � � �� �� �� � � � � �� � � �
� � ��� � � �� � �� � �� �
� �� � � � � �� � � � � � ��
� � ���� � � � � � � �� � � �
������� �� ���� ��� � � � �� �� �� � � �� � � �
Another application can now be set with the “Up” or “Down” keys (See
3.4 “Applications”).
2.12 Ym Calibration
Automatic calibration:
1. Starting with the basic display in the main menu, open the “Dosing”
window from the “Module Type (1)” menu.
2. Select “Ym Calibration” and confirm the selection.
3. Select the “Auto” function and confirm the selection.
WT.050.580.0AA.UA.IM.1206 34
MFC ANALYZER / CONTROLLER
Manual calibration:
WT.050.580.0AA.UA.IM.1206 35
MFC ANALYZER / CONTROLLER
2.13 Decommissioning
If the installation site of the flow block assembly is not frost-free, the
system must be shut down prior to any possible frost formation.
A “KCI tank to store the sensors” may be used instead of the protection
cap for pH and Redox electrodes (see section 7, “Complete Devices,
Retrofit Kits and Spare Parts”).
� A Sensor
� B Sealing cap
� C O-ring
� D Tank
E Stand
�
The fluoride electrodes can be stored wet in the short term in a 100 mg/l
fluoride solution at pH 7. Fill the electrodes in the tank with 100 mg/l
fluoride solution and mount on the electrodes. However, the electrodes
should be stored in a dry condition if they are to be warehoused for several
months. In this case, drain the fill solution from the electrode chamber
and wash the membrane and chamber with distilled water.
WT.050.580.0AA.UA.IM.1206 37
”
(51.5mm)
5 56QE” 8 6 ” 6 556QE” ”
(150mm) (212mm) (170mm) (202mm)
WT.050.580.0AA.UA.IM.1206
(70mm) (70mm) (93.5mm) (108.5mm)
13 56QE”
(350mm)
(6QE”
(6QE” (10mm) 6QE”
(10mm) (7mm)
2 (6QE”
2 6QE”
(62mm)
ca. 27 6QE”
(55mm)
ca. (700mm)
6 56O”
10 6 ”
8”
(163mm)
(263mm)
11 (6 ”
(301mm)
(203mm)
ca. 13 (6 ”
6QE”
ca. (350mm)
(8mm)
38
1 56O”
2 6QE”
(40mm)
(65mm)
2”
(50mm)
2 6 ”
2 6 ”
6 556QE”
(30mm)
1 56M”
(30mm)
2 (6 ” (32mm) (170mm)
1 6 ” (72mm)
(35mm)
2 6 ” 2 6 ” 7 6 ” 6 6QE”
(60mm) (60mm) (193.5mm) (160mm)
5 56O” 5 56O”
(140mm) (140mm)
ca. 3156QE”
ca. (810mm)
MFC ANALYZER / CONTROLLER
1 (6QE”
6 ” (10mm)
(14.5mm)
56 ” 56 ”
(2 mm) (2.5 mm)
8 6QE” 12 (6QE”
ISSUE 0
1 56QE” 8 6QE”
(45mm) (210mm) (210mm) (316mm)
50.580.100.010
12-06
TOP HAT RAIL ASSEMBLY - DIMENSIONS
1 56QE” 4 56 ” 8 6 ” 6 556QE” ” ”
(45mm) (105mm) (212mm) (170mm) (202mm) (51.5mm)
Min. 13 (6 ”
Min. (350mm)
WT.050.580.0AA.UA.IM.1206
9 6M”
(247mm)
10”
8 6QE”
10 6 ”
(252mm)
(263mm)
(207mm)
(6QE” (6QE” 10 6 ”
(270mm)
(10mm) (10mm)
1 56QE”
(45mm)
2”
(50mm)
39
3,5
2 6QE”
(65mm)
569”
6QE”
(8mm)
(5mm)
2 6 ” 2 6 ”
(30mm) (30mm)
2 6 ”
2 6 ” (60mm) 1 56M”
(60mm) (32mm)
5 56O” 2 (6 ” 5 56O” 7 6 ” 6 6QE”
(140mm) (72mm) (140mm) (193.5mm) (160mm)
31 (6 ”
(810mm)
MFC ANALYZER / CONTROLLER
1 56M”
(6QE” (13.5mm)
(11mm)
6 556QE”
(170mm)
56 ” 56 ”
(2 mm) (2.5 mm)
8 6QE” 8 6QE” 12 (6QE”
ISSUE 0
(210mm) (210mm) (316mm)
(193.5mm)
1 56QE”
(45mm)
50.580.100.020
12-06
WALL MOUNT ASSEMBLY - DIMENSIONS
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 40
MFC ANALYZER / CONTROLLER
Section 3 - functions
List of Contents
PARA./DWG. NO.
General Information....................................................3.1
Measurement Inputs....................................................3.2
Output Modules..........................................................3.3
Applications................................................................3.4
Controller Function.....................................................3.5
Controller Outputs......................................................3.6
Control Parameters......................................................3.7
Alarms.........................................................................3.8
Adaption......................................................................3.9
Interfaces.....................................................................3.10
Special Features..........................................................3.11
WT.050.580.0AA.UA.IM.1206 41
MFC ANALYZER / CONTROLLER
The MFC is a special measuring and control device for use on potable
quality water.
Typical applications:
• Measured value
• Mode
• Bar graph with limit values
• Setpoint and measuring range
• Description of customised measuring points
The menus are easy to use, displayed in plain text and are selected using
softkeys.
A 7-day trend display enables you to view past measured values for up
to four selectable process variables.
Four mA outputs and an RS 485 bus interface including Wallace & Tier-
nan® protocol are available to connect visualization systems. Five different
process applications, which reflect the variety of on-site conditions, are
integrated into the MFC to simplify commissioning.
WT.050.580.0AA.UA.IM.1206 42
MFC ANALYZER / CONTROLLER
The value of the combined chlorine is calculated from the difference be-
tween the total chlorine and the free chlorine (optional). This requires a
free chlorine and total chlorine measurement in the same sample water.
The graphic display shows the measured data, limit values and setpoints
as numeric values, diagrams or a trend line.
WT.050.580.0AA.UA.IM.1206 43
MFC ANALYZER / CONTROLLER
NOTE: Not all the same controller functions are available in all ap-
plications (see 3.4 “Applications”).
WT.050.580.0AA.UA.IM.1206 44
MFC ANALYZER / CONTROLLER
• Safety cut-off if dosing tank signals empty and also if the sample
water supply fails
• Dosing time delay
• Alarms
• External stop for all controllers with digital input
• “Positioner closed” function in the event of a power failure (only if
positioner has external power supply)
• Password protection on two levels
3.1.5 Applications 1 to 5
The MFC provides the option to customise the system to the desired on-
site controller functions and systems using up to 5 integrated applications
(see section 3.4 “Applications”).
3.1.6 Links
• CMS 3.0:
Visualisation software for archiving and display of measured values
on PCs with Windows operating systems
• SECO-S7:
PLC driver for data links to Siemens PLC, Type S7-300
• OPC-Server Data Access V2.0:
Server software for Windows operating systems for data links to
visualisation system with OPC client capability
• ChemWeb-Server:
Measured value archiving and display, remote diagnosis, remote ac-
cess with standard browser with Internet and e-mail capability
• Process control systems of different manufacturers
(refer to the manual “RS-485 Bus Interface” IM 050.570-2DE for
description, specification and protocol)
WT.050.580.0AA.UA.IM.1206 45
MFC ANALYZER / CONTROLLER
• Robust sensors
• Constant flow rate with the aid of the flow control valve
• Hydrodynamic grit cleaning of the DEPOLOX ® 5 flow block
electrodes
• Optimum flow around all sensors
The multi-sensor integrated into the flow block assembly monitors the
constant sample water flow rate, registers the sample water temperature
and ensures wide-spread equipotential grounding (sample water ground-
ing).
When the device is switched on, the menus are initialised according to
the installed sensor modules. Even if the sensor modules are changed at
a later date, the user menus are automatically initialised when the device
is switched on.
WT.050.580.0AA.UA.IM.1206 47
MFC ANALYZER / CONTROLLER
By connecting the 3 electrode cell to the MFC (DES sensor module for
3 electrode cell), a variable Upot cell voltage can be output between
the working electrode (red) and reference electrode (white) via the po-
tentiostatic control circuit. A measuring cell current (µA signal), which
is evaluated using the MFC, sets itself proportional to the disinfectant
concentration in the sample water.
A special cleaning grit is filled into the flow block assembly, which
is circulated by the sample water current and continuously cleans the
platinum electrodes.
WT.050.580.0AA.UA.IM.1206 48
MFC ANALYZER / CONTROLLER
• 0–70 µA
• 0–100 µA (factory setting)
• 0–200 µA
• 0–1000 µA
The “µA Meas. Range” parameter can be modified in the “Meas. Range”
menu of the respective module.
Ambient conditions
The following must be taken into account when installing the 3 electrode
measurement:
WT.050.580.0AA.UA.IM.1206 49
MFC ANALYZER / CONTROLLER
3.2.2 pH Measurement
WT.050.580.0AA.UA.IM.1206 50
MFC ANALYZER / CONTROLLER
Installation Notes
Ambient conditions
WT.050.580.0AA.UA.IM.1206 51
MFC ANALYZER / CONTROLLER
Installation Notes
Ambient conditions
The following must be taken into account when installing the Redox
measurement:
Installation Notes
Ambient conditions
The following must be taken into account when installing the fluoride
measurement:
WT.050.580.0AA.UA.IM.1206 53
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 54
MFC ANALYZER / CONTROLLER
Installation Notes
Ambient conditions
The following must be taken into account when installing the con-
ductivity measurement:
WT.050.580.0AA.UA.IM.1206 55
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 56
MFC ANALYZER / CONTROLLER
Installation Notes
Ambient conditions
The following must be taken into account when installing the mem-
brane sensors:
I A
H B
G C
F
D
A Counter electrode
B Electrode finger
C Membrane cap
D Valve opening
E Membrane
F Elastomer seal
G O-ring
H Gold working electrode
I Reference electrode
J Electrode shaft with amplifier
K Electrode screw connector
WT.050.580.0AA.UA.IM.1206 58
MFC ANALYZER / CONTROLLER
The FC1 membrane sensor is a special sensor for measuring the concentra-
tion of free chlorine in water. The sensor incorporates a pH buffer in the
electrolyte and is suitable for water with various pH values. The sensor
may only be used in clear water of potable water quality.
WT.050.580.0AA.UA.IM.1206 59
MFC ANALYZER / CONTROLLER
6. Then place the electrode shaft on the filled membrane cap and
slowly screw the membrane cap in place by hand.
Inserting the Membrane Sensor for Free Chlorine FC1 Into the
Flow-Through Adapter
1. Push the sensor through the cover into the flow-through adapter
until it touches the inflow mating connector. It may be necessary to
position the flow-through adapter cover correctly by turning it.
WT.050.580.0AA.UA.IM.1206 60
MFC ANALYZER / CONTROLLER
NOTE: The membrane may not touch the gold working elec-
trode.
1. Remove air bubbles on the membrane (e.g. tap the inflow mat-
ing connector); air bubbles prevent the chlorine from diffusing
through the membrane and distort the measurement.
2. Replenish electrolyte.
3. Open membrane sensor. To do this, push the elastomer seal to one
side to permit the inflow of air and unscrew the membrane cap.
The CD7 membrane sensor enables the chlorine dioxide content in all
types of water to be determined selectively, without cross-sensitivity to
chlorine, bromine and hydrogen peroxide, however, with cross-sensitiv-
ity to ozone and peracetic acid. The sensor does not demonstrate a pH
value dependency and is therefore very suitable for water with various
pH values.
1. Screw off the membrane cap from the electrode shaft and fill with
the included gel electrolyte.
2. Tap on the membrane cap with the electrode shaft until air bubbles
are removed.
3. Rub gold working electrode with the included lapping paper
(special emery). To do this, lay the lapping paper on a paper
towel, take hold of the corner and while holding the measuring
cell vertically, slide the electrode end one to two times over the
rough side of the lapping paper.
WT.050.580.0AA.UA.IM.1206 62
MFC ANALYZER / CONTROLLER
Inserting the Membrane Sensor for Chlorine Dioxide CD7 Into the
Flow-Through Adapter
1. Push the sensor through the cover into the flow-through adapter
until it touches the inflow mating connector. It may be nec-
essary to position the flow-through adapter cover correctly by
turning it.
CAUTION: The membrane may not touch the gold working elec-
! trode.
WT.050.580.0AA.UA.IM.1206 63
MFC ANALYZER / CONTROLLER
1. Remove air bubbles on the membrane (e.g. tap the inflow mating
connector); air bubbles prevent the chlorine dioxide from diffusing
through the membrane and distort the measurement.
2. Replenish electrolyte.
3. Open membrane sensor. To do this, push the elastomer seal to one
side to permit the inflow of air and unscrew the membrane cap.
4. Pour out the electrolyte.
5. Wash the electrode finger and the membrane cap with clean (dis-
tilled) water and allow it to air dry.
The OZ7 membrane sensor enables the ozone content in all types of
water to be determined selectively, without cross-sensitivity to chlorine,
bromine and hydrogen peroxide, however, with cross-sensitivity to chlo-
rine dioxide and peracetic acid. The sensor does not demonstrate a pH
value dependency and is therefore very suitable for water with various
pH values.
WT.050.580.0AA.UA.IM.1206 64
MFC ANALYZER / CONTROLLER
NOTE: Ensure that only one elastomer seal is in the slot and that
the valve opening closes.
5. Then place the electrode shaft on the filled membrane cap and
slowly screw the membrane cap in place by hand.
6. Then push the second elastomer seal in the slot over the first
one.
WT.050.580.0AA.UA.IM.1206 65
MFC ANALYZER / CONTROLLER
Inserting the Membrane Sensor for Ozone OZ7 into the Flow-Through
Adapter
1. Push the sensor through the cover into the flow-through adapter
until it touches the inflow mating connector. It may be necessary to
position the flow-through adapter cover correctly by turning it.
1. Remove air bubbles on the membrane (e.g. tap the inflow mating
connector); air bubbles prevent the ozone from diffusing through
the membrane and distort the measurement.
WT.050.580.0AA.UA.IM.1206 66
MFC ANALYZER / CONTROLLER
2. Replenish electrolyte.
3. Open membrane sensor. To do this, push the elastomer seal to
one side to permit the inflow of air and unscrew the membrane
cap. Pour out the electrolyte.
4. Wash the electrode finger and the membrane cap with clean (dis-
tilled) water and allow to air dry.
The membrane sensor measures the total amount of free and combined
chlorine (chloramine). The sensor incorporates a pH buffer in the elec-
trolyte and is suitable for water with various pH values. The sensor may
only be used in clear water of potable water quality.
1. Screw off the membrane cap from the electrode shaft and fill
to the top of the threads on inside of cap with the included gel
electrolyte.
2. Rub gold working electrode with the included lapping paper
(special emery). To do this, lay the lapping paper on a paper
towel, take hold of the corner and while holding the measuring
cell vertically, slide the electrode end one to two times over the
rough side of the lapping paper.
WT.050.580.0AA.UA.IM.1206 67
MFC ANALYZER / CONTROLLER
Inserting the Membrane Sensor for Total Chlorine TC1 Into the
Flow-Through Adapter
1. Push the sensor through the cover into the flow-through adapter
until it touches the inflow mating connector.It may be necessary to
position the flow-through adapter cover correctly by turning it.
WT.050.580.0AA.UA.IM.1206 68
MFC ANALYZER / CONTROLLER
CAUTION: The membrane may not touch the gold working elec-
! trode.
1. Remove air bubbles on the membrane (e.g. tap the inflow mat-
ing connector); air bubbles prevent the chlorine from diffusing
through the membrane and distort the measurement.
2. Replenish electrolyte.
3. Open membrane sensor. To do this, push the elastomer seal to one
side to permit the inflow of air and unscrew the membrane cap.
4. Pour out the electrolyte.
5. Wash the electrode finger and the membrane cap with clean (dis-
tilled) water and allow to dry in a dust-free place.
WT.050.580.0AA.UA.IM.1206 70
MFC ANALYZER / CONTROLLER
The relay module has over eight relays, each with a two-way switch.
These switches are assigned various switching tasks depending on the
selected application (see section 3.4 “Applications”). The corresponding
diagrams for the five applications are in the appendix under section 5,
“Wiring Diagrams”. The switches are wired with suppressor diodes to
protect from spikes. In order to switch larger inductive load, we recom-
mend installing an additional contact such as a contactor or load relay.
WT.050.580.0AA.UA.IM.1206 71
MFC ANALYZER / CONTROLLER
3.4 Applications
The MFC provides the option to customise the system to the various on-
site systems using five integrated applications.
WT.050.580.0AA.UA.IM.1206 72
MFC ANALYZER / CONTROLLER
3.4.1 Application 1
� � � � � � �
� � � ��� � ��� �
�
� � � � �
� � � ��� � ��� � ��
� � � � � � �� � � � �� �� � � ��� � � �� � �� � �� � ��� � �� � � �� � � � � ���
� � �� � � �� � � � ��� � � ��
� � �� � �� � �� � ���
� � � � � �� �� �� �
� � � � � � �� � � �� � � �� � �� � �� � ��� � �� � � �� � � � � ��� �
� � ��� � � � � � �
� � �� � �� ���� � � � � � � �� � �� � �� � � ��� � �� � � � �� � �� � �� � ��� � �� � � �� � � � � ���
� � �� � �� � ���
� � �� � �� � ��� ���� ��� � ��� � � �� � � �� �
� � �� � � �� � � � �� � �� � ��
� �� ���� � � �
� � ����� � � �� � � �� � �
� � �
� � �� � �� � ���
� �� ���� � � � � �� ���� � � �
�� � � � � � � ��� � ��� � � �� �� � �� �� � � ���
� � �� � �� � ���
� �� ���� � � �
�� � � � � � � ��� � ��� � � �� �� � �� �� � � ���
� � �� � �� � ���
� �� ���� � � �
��� � � ��
� � � �� � � � � � �� � ��� � � � � ��� �
� � � � � � � �� �� �
� � �� � � �� ��� � � �
� � � � � � �� � � � �� � � � �� � �� � ����� � � � � �� � ����
� � � � � � � �� � � � ���� � �
� �� � �� � �� � � � � �
�� � � �� � � � �� �� � ��� � � ��� � �� ��� � �
�� � �
�� � � �
WT.050.580.0AA.UA.IM.1206 73
MFC ANALYZER / CONTROLLER
3.4.2 Application 2
� � � � � � �
� � � ��� � ��� �
� ����
� � � � �
� �
�
� � � ��� � ��� � ��
� � � �� � � � �� � � �� � � �� � �� ��� �� �
� � � � � � �� � � � �� �� � � ��� � � �� � �� � �� � ��� � � � � � � � � �� � � � � � �
� � � � �� ��
� � � ��� �
� � �� � � �� � � � ��� � � ��
� � � � � � �� � � � �� � � �� � ��
�� � � ��� � � �� � �� � �� � ���
� � �� � �� � �� � ���
� � � � � �� �� �� �
� � � � � � �� � � �� � � �� � �� � �� � ��� � �� � � �� � �� � ��� �
� � ��� � � � � � �
� � �� � �� ���� � � � � � � �� � �� � �� � � ��� � �� � � � �� � �� � �� � ��� � �� � � �� � �� � ���
� � �� � �� � ���
� � �� � �� � ��� ���� ��� � ��� � � �� � � �� �
� � �� � � �� � � � �� � �� � ��
� �� ���� � � �
� � ����� � � �� � � �� � �
� � �
� � �� � �� � ��� � � �� � �
� �� ���� � � �
�� � � �� � � ��� � ��� � � �� ��� ���� � � ��� � �� ���� � � �
� � �� � �� � ���
� �� ���� � � �
�� � � �� � � ��� � ��� � � �� ��� ���� � � ���
� � �� � �� � ���
� �� ���� � � �
� � �� � � �� ��� � � � ��� � � ��
� � � �� � � � � � �� � ��� � � � � ��� �
� � � � � � � �� �� �
� � � �� � �� � � � �� � � � �� � �� � ����� � � � � �� � ����
� � � � � � � �� � � � ���� � �
� �� � �� � �� � � � � �
�� � � �� � � � �� �� � ��� � � ��� � �� ��� � �
�� � �
�� � � �
WT.050.580.0AA.UA.IM.1206 74
MFC ANALYZER / CONTROLLER
3.4.3 Application 3
� � � � � � �
� � � ��� � ��� �
�
� � � � �
� �
� �
�
• Factory setting
• Up to four sensor modules
• Sensor measuring module 1 has controller outputs for ratio control,
single feedback closed-loop control or combi-control
• Sensor measuring module 2 has controller outputs as single feedback
closed loop control or ratio control (same flowrate signal as MOD
1)
• mA output can be assigned
• 4 alarm contacts
• Switching functions configurable on digital inputs and limit values
� � � ��� � ��� � ��
� � � �� � � � �� � � �� � � �� � �� ��� �� �
� � � � � � �� � � � �� �� � � ��� � � �� � �� � �� � ��� � � � � � � � � �� � � � � � �
� � � � �� ��
� � � ��� �
� � �� � � �� � � � ��� � � ��
� � � � � � �� � � � �� � � �� � ��
�� � � ��� � � �� � �� � �� � ���
� � � �� � � � �� � � �� � � �� � �� � �� ��� � � �
� � � � � � � � �� � � � � � �
� � � � � � �� � � � �� �� � � ��� � � �� � �� � �� � ��� � � � � �� ��
� � � ��� �
� � � � � � �� � � � �� �� � � ��� � � �� � �� � �� � ��� � � �� � ��
� � �� � �� � �� � �� � ��
� � �� � �� � �� � ���
� � � � � �� �� �� �
� � � � � � �� � � �� � � �� � �� � �� � ��� � �� � � �� � �� � ��� �
� � ��� � � � � � �
� � �� � �� ���� � � � � � � �� � �� � �� � � ��� � �� � � � �� � �� � �� � ��� � �� � � �� � �� � ���
� � �� � �� � ���
� � �� � �� � ��� ���� ��� � ��� � � �� � � �� �
� � �� � � �� � � � �� � �� � ��
� �� ���� � � �
� � ����� � � �� � � �� � �
� � �
� � �� � �� � ��� � � �� � �
� �� ���� � � �
�� � � �� � � ��� � ��� � � �� ��� ���� � � ��� � �� ���� � � �
� � �� � �� � ���
� �� ���� � � �
�� � � �� � � ��� � ��� � � �� ��� ���� � � ���
� � �� � �� � ���
� �� ���� � � �
� � �� � � �� ��� � � � ��� � � ��
� � � �� � � � � � �� � ��� � � � � ��� �
� � � � � � � �� �� �
� � � �� � �� � � � �� � � � �� � �� � ����� � � � � �� � ����
� � � � � � � �� � � � ���� � �
� �� � �� � �� � � � � �
�� � � �� � � � �� �� � ��� � � ��� � �� ��� � �
�� � �
�� � � �
WT.050.580.0AA.UA.IM.1206 75
MFC ANALYZER / CONTROLLER
3.4.4 Application 4
� � � � � � �
� � � ��� � ��� �
�
� � � � �
� �
� � � ��� � ��� � ��
� � � �� � � � �� � � �� � � �� � �� ��� �� �
� � � � � � �� � � � �� �� � � ��� � � �� � �� � �� � ��� � � � � � � � � �� � � � � � �
� � � � �� ��
� � � ��� �
� � �� � � �� � � � ��� � � ��
� � � � � � �� � � � �� � � �� � ��
�� � � ��� � � �� � �� � �� � ���
� � �� � �� � �� � ���
� � � �� � � � � �� � � ���
� � � � � �� �� �� �
� � � � � � �� � � �� � � �� � �� � �� � ��� � �� � � �� � �� � ��� �
� � ��� � � � � � �
� � �� � �� ���� � � � � � � �� � �� � �� � � ��� � �� � � � �� � �� � �� � ��� � �� � � �� � �� � ���
� � �� � �� � ���
� � �� � �� � ��� ���� ��� � ��� � � �� � � �� �
� � �� � � �� � � � �� � �� � ��
� �� ���� � � �
� � ����� � � �� � � �� � �
� � �� � �� � ��� � � �� � �
� �� ���� � � �
�� � � �� � � ��� � ��� � � �� ��� ���� � � ��� � �� ���� � � �
� � �� � �� � ���
� �� ���� � � �
�� � � �� � � ��� � ��� � � �� ��� ���� � � ���
� � �� � �� � ���
� �� ���� � � �
� � �� � � �� ��� � � � ��� � � ��
� � � �� � � � � � �� � ��� � � � � ��� �
� � � � � � � �� �� �
� � � �� � �� � � � �� � � � �� � �� � ����� � � � � �� � ����
� � � � � � � �� � � � ���� � �
� �� � �� � �� � � � � �
�� � � �� � � � �� �� � ��� � � ��� � �� ��� � �
�� � �
�� � � �
WT.050.580.0AA.UA.IM.1206 76
MFC ANALYZER / CONTROLLER
3.4.5 Application 5
� � � � � � �
� � � ��� � ��� �
� ����
� � � � �
� �
�
� � � ��� � ��� � ��
� � � �� � � � �� � � �� � � �� � �� ��� �� �
� � � � � � �� � � � �� �� � � ��� � � �� � �� � �� � ��� � � � � � � � � �� � � � � � �
� � � � �� ��
� � � ��� �
� � �� � � �� � � � ��� � � ��
� � � � � � �� � � � �� � � �� � ��
�� � � ��� � � �� � �� � �� � ���
� � �� � �� � �� � ���
� � � �� � � � � �� � � ���
� � � � � �� �� �� �
� � � � � � �� � � �� � � �� � �� � �� � ��� � �� � � �� � �� � ��� �
� � ��� � � � � � �
� � �� � �� ���� � � � � � � �� � ��� � �� � � ��� � ���� � � � �� � �� � �� � ��� � �� � � �� � �� � ���
� � �� � �� � ���
� � �� � �� � ��� ���� ��� � ��� � � �� � � �� �
� � �� � � �� � � � �� � �� � ��
� �� ���� � � �
� � ����� � � �� � � �� � �
� � �� � �� � ��� � � �� � �
� �� ���� � � �
�� � � �� � � ��� � ��� � � �� ��� ���� � � ��� � �� ���� � � �
� � �� � �� � ���
� �� ���� � � �
�� � � �� � � ��� � ��� � � �� ��� ���� � � ���
� � �� � �� � ���
� �� ���� � � �
� � �� � � �� ��� � � � ��� � � ��
� � � �� � � � � � �� � ��� � � � � ��� �
� � � � � � � �� �� �
� � � �� � �� � � � �� � � � �� � �� � ����� � � � � �� � ����
� � � � � � � �� � � � ���� � �
� �� � �� � �� � � � � �
�� � � �� � � � �� �� � ��� � � ��� � �� ��� � �
�� � �
�� � � �
WT.050.580.0AA.UA.IM.1206 77
MFC ANALYZER / CONTROLLER
• Flow proportional
• Ratio control
• Residual only
• Compound loop control
• Setpoint trim
Online measurement can be transmitted directly from the MFC and ex-
ternal measuring system via mA input signal. External control signals
such as flow rate and external setpoint are recorded using the mA/V input
module. The MFC system can record up to four main measurements and
two external control signals. In addition, measuring inputs for temperature,
actuator feedback and three digital inputs are available.
Sensors Kreis 1 DI DI DI
I II III
12:5
+Temp. 0,65
1.00
0.
6,85 .00
.00
05 00
pH mV
mg/l
Flow MENU CAL
0,00
MODE mA ?
.00
100% 28°C
00
sensor
ESC
control output
Flowsensor Dosing device
positioner, pumps
WT.050.580.0AA.UA.IM.1206 78
MFC ANALYZER / CONTROLLER
• Dosing pumps
• Solenoid pump
• Positioner with feedback 1kOhm/5kOhm
• mA analog output
The flow rate is recorded and the dosing rate adjusted proportionally to
the flow rate using the flow rate sensor with linear mA/V output signal.
For the flow signal settings, see menu “Input/Output” – “Flow Wq“.
You can switch between internal and external dosing factor (DF) via the
digital input (“Dos.Fact.Source” = “external with DI3” or “internal with
DI3”).
WT.050.580.0AA.UA.IM.1206 79
MFC ANALYZER / CONTROLLER
Sensors Kreis 1 DI DI DI
I II III
12:5
+Temp. 0,65
1.00
0.
6,85 .00
.00
05 00
pH mV
mg/l
Flow MENU CAL
0,00
MODE mA ?
.00
100% 28°C
00
sensor
ESC
control output
Dosing device
positioner, pumps
Input signals:
WT.050.580.0AA.UA.IM.1206 80
MFC ANALYZER / CONTROLLER
• Dosing pumps
• Solenoid pump
• Positioner with/without feedback (1kOhm/5kOhm)
• Continuous
The setpoint can be set within the measuring range of module 1 (at “Set-
point Source” = internal).
An external setpoint from 0–100 % can be provided via the mA/V input
signal (“Setpoint Source” = external) You can switch between internal
and external setpoint via the digital input (“Setpoint Source” = “external
with DI3” or “internal with DI3”).
The control direction can be selected with the parameter “Control Direc-
tion” = direct or inverse (e.g. direct = chlorination, inverse = dechlorina-
tion).
WT.050.580.0AA.UA.IM.1206 81
MFC ANALYZER / CONTROLLER
Sensors Kreis 1 DI DI DI
I II III
12:5
+Temp. 0,65
1.00
0.
6,85 .00
.00
05 00
pH mV
mg/l
Flow MENU CAL
0,00
MODE mA ?
.00
100%
00
28°C
sensor
ESC
control output
Flowsensor Dosing device
positioner, pumps
Input signals:
Output parameter:
• Dosing pumps
• Solenoid pump
• Positioner with/without feedback (1kOhm/5kOhm)
• Continuous
• 4-20mA out
WT.050.580.0AA.UA.IM.1206 82
MFC ANALYZER / CONTROLLER
To detect control deviations, the sensor module 1 records the control vari-
able and a setpoint is specified, which are compared with the integrated
single feedback closed-loop control.
The internal setpoint can be set within the measuring range of sensor
module 1. “Setpoint Source” must be set to “internal”. An external set-
point from 0–100 % can be provided via the mA/V input signal. “Setpoint
Source” must be set to “external”. You can switch between internal and
external setpoint via the digital input. The “Setpoint Source” must be set
to “external with DI3” or “internal with DI3”.
The MFC operates internally with a dosing factor table for 0 – 105 %
flow. In 5 % intervals, the device determines the required dosing factors
automatically during operation based on the corresponding flow rate..
The single feedback closed-loop control corrections are transferred into
the dosing factor table during this process. Non-linearities in the control
loop are learned this way. This quickly activates the setpoint if flow rate
changes occur.
The control operating mode can be switched between ratio control and
single feedback closed-loop control via digital input.
The control direction can be selected with the parameter “Control Direc-
tion” = direct or inverse (e.g. direct = chlorination, inverse = dechlorina-
tion)
WT.050.580.0AA.UA.IM.1206 83
MFC ANALYZER / CONTROLLER
The single feedback closed-loop control remains switched off (Ypi stop
function) during the disturbance variables (flow rate change, positioner
running time, dead time from line lengths). This maintains a stable con-
trol, which means the control operates with the dosing factor from the
dosing factor table applicable for the new flow rate. The time the single
feedback closed-loop control is switched off is determined by the fuzzy
module and is therefore variable (“PI” display in seconds).
Control deviations that occur are quickly offset by the PI single feedback
closed-loop control during continuous flow.
A positive jump in the flow rate causes a brief drop below the setpoint
due to the running time of the positioner and the dosing delay. Therefore,
the PI controller freezes for a brief period (“PI” display in seconds).
A negative jump in the flow rate causes the setpoint to be briefly exceeded
due to the running time of the positioner and the dosing delay. Therefore,
the PI controller freezes for a brief period (“PI” display in seconds).
WT.050.580.0AA.UA.IM.1206 84
MFC ANALYZER / CONTROLLER
To adjust the control for compound loop control, the Tconst and Tvar
process times must be entered in the parameter menu path. These times
refer to control loop dead times,which on the one hand are independent
of the control variables, and on the other hand depend proportionally on
the control variables.
The variable dead time< Tvar > depends on the current control variable
and is entered in the menu at a control variable of 100%.
The following calculation examples apply for the use of the MFC for
chlorine dosing.
WT.050.580.0AA.UA.IM.1206 85
MFC ANALYZER / CONTROLLER
The control variable independent dead time Tconst consists of the mea-
suring dead time and the dosing dead time.
Calculation 1:
The sample water is extracted right after the mixture loop and fed to
the measuring cell.
The sample water dead time depends on the nominal diameter and
length of the sample water line and the flow rate to the measuring
cell. A flow rate of 36 l/h is assumed for the DEPOLOX® 5 measur-
ing cell.
Example
tmW = (0.6 x 0.6 x 10) : 7.65 min = 0.47 min, (i.e. approx. 28 sec.)
Calculation 2:
The sample comes from the sample water pump (bypass line).
Sample water dead time depends on the flow rate of the sample water
pump, nominal diameter of the bypass line and its length up to the
sample water branch pipe to the measuring cell.
WT.050.580.0AA.UA.IM.1206 86
MFC ANALYZER / CONTROLLER
Check whether the length of the sample water line to the measuring cell
can be neglected. If so, establish the sum from calculation 1 and 2.
Calculation 3:
The exposure time in the delay tank must be added to the calculated
time.
Dosing dead times arise from long dosing lines and positioner running
times.
Calculation 1:
Calculation 2:
If rapid control variable changes are expected in the system, which the
dosing equipment cannot adjust to (e.g. positioner running times, dosing
pump cycle times), the dosing delay time should be assumed under all
circumstances to be half of the positioner running time ty or the cycle
time tp.
WT.050.580.0AA.UA.IM.1206 87
MFC ANALYZER / CONTROLLER
The sum of the measured dead time and the dosing delay is displayed in
the< Tconst > menu in minutes.
The control variable dependent Tvar dead time depends on the nominal
flow rate, the internal diameter of the line and the distance between where
the chlorine is added and the sample water extracted.
WT.050.580.0AA.UA.IM.1206 88
MFC ANALYZER / CONTROLLER
� � � � � �� � � � �� �� �
�
� �
� �
� �
� � �
�
� � � � �
� � � � � � � �� � � �� �
� �� � � �� �
� �� �
� � � � � �
� �� �
� � �� � � � �
� �� � � ���� � � � �� ���� � � ���
� �� �
��� � � � � � � ��
� �� �
� � � �
� � �
������� � � �
� � � � � �
� � �
� � � � � ��
� � � � � �
� �� � �
� � � � � �
� � � �� � ��� � �� � � � � � �� � ��� � � �
� � � �� � �� � � �� �
� �� � � � � � � � � � � �� �� � � � ��� � � � �
Input signals:
WT.050.580.0AA.UA.IM.1206 89
MFC ANALYZER / CONTROLLER
Output parameter:
• Dosing pump
• Solenoid pump
• Positioner with/without feedback 1kOhm/5kOhm
• Continuous mA output
The control circuit 1 operates as the compound loop control (see section
3.5.3 “compound loop control”). All settings for module 1 must also be
made as described under compound loop control.
The setpoint adjustment operates with a delay time that determines the
setpoint trim function Tconst and Tvar as well as the flow rate using the
process times. The flow must be constant during this delay, otherwise
the time restarts.
WT.050.580.0AA.UA.IM.1206 90
MFC ANALYZER / CONTROLLER
For example:
Setpoint 1 = 0.50 mg/l Measuring range 1.00 mg/l
5 % of measuring range = 0,05 mg/l =>”good range” =
0.45–0.55 mg/l
A delay time begins when the setpoint 1 is reached, after which the set-
point trim function is activated. This delay time depends on the Tconst.
and Tvar delay times and the current flow rate. The smaller the flow rate,
the longer the delay time.
Example:
Flow rate = 50 %
Tvar (at 100 % flow rate) = 2h:30min
Tconst = 2 min
=> Delay = 100 % / 50 % x Tvar + Tconst = 5h:02min
The setpoint trim compares setpoint2 with actual value2 once the time
has expired and changes the combi-control setpoint1 according to the
following equation, if necessary:
The delay time is also restarted if the actual value1 is more than 5 % of
the setpoint 1. The time starts again only after the 5 % range has been
reached.
The delay time is always reset after an automatic setpoint change and
only restarted once the setpoint range has been reached.
WT.050.580.0AA.UA.IM.1206 91
MFC ANALYZER / CONTROLLER
The following parameters must be set for this operating mode in menu
(2):
• Setpoint range
• Trim factor
• Xsh
• Tconst
• Tvar
This parameter describes the range within which the combi-control set-
point 1 can be adjusted by the setpoint trim.
Example:
Measuring range = 2.0 mg/l free chlorine
Setpoint 2 = 1.00 mg/l
Setpoint range = 20 %
If the setpoint trim function records a value outside this range, the maxi-
mum or minimum value of the range is entered (in this example: 0.60
mg/l and 1.40 mg/l).
Example:
Measured value = 0.60 mg/l free chlorine
Setpoint = 1.00 mg/l
Trim factor = 50 %
Error = 0.40 mg/l
WT.050.580.0AA.UA.IM.1206 92
MFC ANALYZER / CONTROLLER
3.5.4.5 Xsh
This parameter describes a neutral range for the measured value of mod-
ule2 and is indicated in percent on the measuring range end value. If the
measured value in module 2 is within this range around the setpoint, the
setpoint trim function is inactive. If the measured value of module 2 is
outside this neutral range, the setpoint trim function is active and opti-
mises the setpoint of module1.
To adjust the control for the setpoint trim function, the Tconst and Tvar
process times must bet entered in the setpoint trim menu. These times
refer to control loop dead times, which are independent of the control
variables, and depend proportionally on the control variables.
Tconst Menu 2
The constant dead time “Tconst” for the setpoint trim is the sum of
the module 2 measurement dead time and possible dosing delays
during the combi-control.
Tvar Menu 2
The variable dead time “Tvar” depends on the current flow rate control
variable, the internal diameter of the line and the distance between
where the chlorine is added and the sample water extracted. Tvar is
entered in the menu for a control variable at 100 % (flow rate = 100
%).
WT.050.580.0AA.UA.IM.1206 93
MFC ANALYZER / CONTROLLER
Wait until the actual value2 adjusts to the dosing rate and remains con-
stant.
Calculate as follows:
WT.050.580.0AA.UA.IM.1206 94
MFC ANALYZER / CONTROLLER
Parameter
Controller For Type Action
Description
Positioner with feed- 3-point Positioner with Ym Dosing or
back
Positioner without 3-point Positioner without Ym Dosing or
feedback
Motor dosing pump 2-point Dosing pump 2p Dosing or
(pulse duration
controller)
2 Motor dosing 3-point Dosing pump 3p Dosing and
pumps (pulse dura-
tion controller)
Pulse pump (pulse 2-point Pulse pump 2p Dosing or
frequency controller)
2 Pulse pumps (pulse 3-point Pulse pump 3p Dosing and
frequency controller)
Dosing pump with 2-point Analog output 2p Dosing or
mA-input
2 Dosing pumps with 3-point Analog output 3p Dosing and
mA-input
Dosing contact 2-point Enable contact Dosing
WT.050.580.0AA.UA.IM.1206 95
MFC ANALYZER / CONTROLLER
MFC Controller
With the selection of the integrated controller 4.
Outputsfor ex-
for “positioner”,
ample, it is possible to use chlorine overfeed control in connection with
a positioner as dosing equipment in a chlorinator.
2-point pulse Pulse pumps are controlled with 0 to 100 or 0 to 120 pulses per
3.6.4 2-Point Pulse Frequency Controller
frequency controller For Pulse
minute, depending onPumps
the specification of the connected pump.
for pulse pumps
Pulse pumps areThe duty cycle
controlled during
with 0 toeach
100 dosing
or 0 tois120
0.3pulses
s. The per
break time is
minute,
calculated between 0.2 and 60 s depending on the dosing rate.
depending on the specification of the connected pump.
Example for a solenoid pump with 120 pulses/min.:
The duty cycle during each dosing is 0.3 s. The break time is calculated
between100...
Yout in %
0.2 and84...
60 s depending
72... 56
on 50...
the dosing rate.
33... 25... 10... 5... 10
Pulses/min 120...
Example 96... 85...
for a solenoid pump75 60...pulses/min.:
with 120 40... 30... 12... 6... 10
Yout
100... 84... 72... 56 50... 33... 25... 10... 5... 10
in %
Pulses/
120... 96... 85... 75 60... 40... 30... 12... 6... 10
min
WT.050.580.0AA.UA.IM.1206 96
MFC ANALYZER / CONTROLLER
The control range is between -100 % (pump 1) and +100 % (pump 2);
this range can also be set in manual mode.
� � �� � ��� �� � � �
� � � �
� � � � �� � � � � � � � ��
� � � � � � �
� � � �
� � � �� � �� � ��
� � � � �
� � � � � � � � �
The MFC has up to four analog mA outputs. These can be assigned in-
dividually as registration or controller outputs.
WT.050.580.0AA.UA.IM.1206 97
MFC ANALYZER / CONTROLLER
� � � �� � �� � ��
� � � �
� � � � �� � � � � � � � ��
� � � �� � ��� � � �
� � �
� � � � � � � � �
� � � �� � �� � �� � �
� � � � � � � � �
WT.050.580.0AA.UA.IM.1206 98
MFC ANALYZER / CONTROLLER
Control parameters and setting values for determining the control functions
of a controller. Different parameters apply for each controller type.
Pulses Max./Min
Meaning: Maximum number of pulses.
Setpoint
Specified value at which the control variable can be maintained by
the controller. The setting range corresponds to the respective mea-
suring range.
Tn
Meaning: Integral action time (I-element).
WT.050.580.0AA.UA.IM.1206 99
MFC ANALYZER / CONTROLLER
T
Sampling time T is the time after which a change control variable or
setpoint is responded to. This value must be adjusted in the case of
delayed feedback signals.
Tp
Meaning: Cycle period.
Display: Seconds.
Dosing pump
up to 20 20-40 40-80 80-125 125-200
strokes/min
Tp value 120 100 60 30 15
Ts
Meaning: Loop rise time.
Display: Minutes.
WT.050.580.0AA.UA.IM.1206 100
MFC ANALYZER / CONTROLLER
Tu
Meaning: Loop dead time.
Display: Seconds.
Ty
Meaning: Running time of the positioner.
Display: Seconds.
Control Direction
Meaning: Direction of the control.
Example:
pH: Control direction “inverse”: Lowering pH-value
by adding acid.
pH: Control direction “direct”: Adding alkaline to
raise the pH value.
WT.050.580.0AA.UA.IM.1206 101
MFC ANALYZER / CONTROLLER
Xp
Meaning: Proportional factor.
Xsh
Meaning: Neutral zone.
WT.050.580.0AA.UA.IM.1206 102
MFC ANALYZER / CONTROLLER
Ymax
Meaning: Dosing rate limitation (single feedback
control-loop control only).
Ymin
Meaning: Dosing rate basic load (single feedback
control-loop control only).
NOTE: Ymin and Ymax is only available for the single feedback
closed-loop control. The control range is limited by the parameters
Ymax and Ymin. Do not select a Ymax value lower than Ymin.
At Ymin > 0 overdosing can occur.
WT.050.580.0AA.UA.IM.1206 103
MFC ANALYZER / CONTROLLER
Ym Calibration
This parameter switches off the flow input (off) and activates the flow
rate signal for the ratio control (factory setting = flow measurement)
as control variable.
Flow Direction
This parameter determines the direction of the flow rate signal directly
proportional to the actuator output:
WT.050.580.0AA.UA.IM.1206 104
MFC ANALYZER / CONTROLLER
Control Variable 2
This parameter activates and deactivates a second control variable
during the ratio control.
X Direction
Determines the direction of the second control variable during the
ratio control.
X Factor
This parameter is only available during ratio control, control variable
2 = measured value X.
Tconst
Defines the constant dead time in the combi-controller loop. Consists
of the sample water line dead time and the dosing delay time (for the
calculation, see section 3.5.3 “Compound loop control”).
Tvar
Defines the variable dead time in the combi-controller loop. The time
to be entered is based on 100 % flow rate (for the calculation, see
section 3.5.3 “Compound loop control”).
WT.050.580.0AA.UA.IM.1206 105
MFC ANALYZER / CONTROLLER
Max.lin.Corr
This parameter monitors changes to already learned dosing factors.
If new dosing factor changes are learned, which are larger than the
max. linearity correction, this dosing factor is used for all values in
the dosing curve = > initialisation of the curve.
Example:
Max.lin.Corr. = 50 % (based on dosing factor):
Previous dosing factor: 30 %
Newly learned dosing factor: 48 %
max. permissible correction range: 30 ± (50 % from 30 %)
= 30 % ± 15 %
Change in this case: 48 % - 30 % = +18 %
=> The new dosing factor is assumed for the entire curve
because the new dosing factor (+ 48 %) is greater than the
max.lin.Correction (+18 %).
Control Factor
Setting the ratio of control range and measuring range, in order to
adjust the control amplification Xp to the process.
Control factor =
(End of measuring range - start of measuring range): Control range
Example:
Start of measuring range: pH 4
End of measuring range: pH 9
Max. process control range: ± 1 pH (=> 2 pH increments)
=> Control factor =
(9 - 4) : 2 = 2.5
WT.050.580.0AA.UA.IM.1206 106
MFC ANALYZER / CONTROLLER
3.8 Alarms
The alarms are output via relay contacts and the red LED. The number
of up to eight alarms is stipulated in the application.
In all alarm types the response can be influenced by entering a delay (td)
(refer to the diagrams in this chapter).
The LED lights up in the event of an alarm and goes out automatically
when the alarm conditions are eliminated. The same applies for the
contact.
Relay without
delay
WT.050.580.0AA.UA.IM.1206 107
MFC ANALYZER / CONTROLLER
The LED flashes in the event of an alarm until the alarm is acknowledged.
The LED goes out, also if the set alarm conditions still apply when the
alarm is acknowledged.
WT.050.580.0AA.UA.IM.1206 108
MFC ANALYZER / CONTROLLER
The LED flashes in the event of an alarm until the alarm is acknowl-
edged.
WT.050.580.0AA.UA.IM.1206 109
MFC ANALYZER / CONTROLLER
3.9 Adaption
� � ����� �� �� � � � � � � � � � �
� � � �
� � �
� � �
� � � �� �
� � � �
� � � �
� � � � �� � � �� � �� � ��
� � � �� �
� � � � � � � � �
3.9.1 Use
3.9.2 Requirements
WT.050.580.0AA.UA.IM.1206 110
MFC ANALYZER / CONTROLLER
1. Starting from the basic display, select “Adaption” from the “Cl2free”
menu. The Tu and Ts loop parameters are displayed.
2. In the “Cl2 Adaption” menu, select the “ADAPT” softkey. This dis-
plays the diagram of the previous adaption.
3. Press “START” to start the adaption. Adaption starts.
��� � � � � �
� � � � � � � � � �� � ���� � � � � � � � � �
� �� �
� �� �
� � ��
� � �� �� �
� �� �� � �
�
� � � � �
� � � � � �
���� � � � �� ��� � � � � � ������ ��� � � � � ��
3.9.4 Displays
The diagram shows the chlorine value curve during the adaption phases.
The current phase of adaption (total of 13) is shown in the bottom line.
WT.050.580.0AA.UA.IM.1206 111
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 112
MFC ANALYZER / CONTROLLER
When the loop times (dead time Tu and rise time Ts) have been completed
without errors, calculation of the control parameters Xp and Tn commences.
This is indicated by the message “Adaption OK”. The calculated param-
eters are entered into the menus. When adaption has been concluded, the
measuring amplifier adjusts with the newly calculated control parameters
and continues in the selected operating mode (e.g. automatic).
These are entered into the “Tu” and “Ts” menus to monitor the deter-
mined loop times.
If errors occur in the control loop during adaption, incorrect loop times
and therefore incorrect control parameters can be determined.
If errors occur in the control loop during adaption (e.g. filter backwashing,
changes in the circulation) or if the reaction times of the control loop are
too long, adaption is interrupted.
WT.050.580.0AA.UA.IM.1206 113
MFC ANALYZER / CONTROLLER
If the loop times Tu and Ts are already known or if these cannot be de-
termined automatically due to specific system conditions, the loop times
can be entered into the “Tu” and “Ts” menus. When Tu or Ts are saved,
the control parameters Xp and Tn are also calculated and entered into
the menus.
WT.050.580.0AA.UA.IM.1206 114
MFC ANALYZER / CONTROLLER
3.10 Interfaces
3.10.1 RS232
Daily Log
Each daily log consists of:
• Log header
• Line diagrams
• Release of daily min. and max. values
Log Header
The supplied measurements and the corresponding parameters are
printed in the log header. The date of the last calibration is also
documented.
Trend Graphs
This prints the trend graphs including the measuring ranges and the
current operating mode as a bar graph. Trend graphs are recorded in
one minute intervals.
The daily min. and max. values per measurement are printed at the
end of the log.
WT.050.580.0AA.UA.IM.1206 115
MFC ANALYZER / CONTROLLER
Log Printout
To print logs, Wallace & Tiernan® offers the RS232 log printer as an
accessory as well as the 3 m and 15 m printer connecting cable. The
printer includes setup instructions. Refer to section 7, “Complete
Devices, Retrofit Kits and Spare Parts”.
WT.050.580.0AA.UA.IM.1206 116
MFC ANALYZER / CONTROLLER
3.10.2 RS485
Ru
Rt
Rd
WT.050.580.0AA.UA.IM.1206 117
MFC ANALYZER / CONTROLLER
You can select between options for manual, permanently set temperature
value or temperature measurement with PT 1000 on the motherboard for
compensation in the calibration menu for pH. If the PT 1000 measure-
ment of the mother board is switched off, only a manual value may be
set for compensation.
Cl2++ Measurement
The pH dependency of the chlorine measurement is compensated
if the pH value fluctuates within the range of pH 6.5 to pH 9. This
function is only guaranteed to a max. 10 mg/l free chlorine.
WT.050.580.0AA.UA.IM.1206 118
MFC ANALYZER / CONTROLLER
SECTION 4 - Operation
List of Contents
PARA. NO.
WT.050.580.0AA.UA.IM.1206 119
MFC ANALYZER / CONTROLLER
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� � � � � � � � � � � ������������������ � � ������������� � �� � �
The MFC is operated with nine keys. The software function is controlled
with the top three keys (softkeys).
� � �
Indicators
CONTROL 1 System name (Entered under “System” - “Common”
- “System name”)
� � � � � �� Digital inputs 1, 2, and 3 active.
� �� ����� The symbols indicate that a function has been selected
for the digital signal and that a signal is applied.
WT.050.580.0AA.UA.IM.1206 120
MFC ANALYZER / CONTROLLER
Password active
The defined password must be entered to permit
modification of parameters and for calibrating the
device.
�� � ��� � �
�
� �
��
� �
��� �� � �� � � Time
� �� � � �� � “AUTO” operating mode active
� � �� � �� � The control unit is running in automatic mode.
� �� � � �� ��� � �
��� �� ��
� ��
� �� � ��� �� �
� � � �� � Dosing is performed automatically.
“MANUAL” operating mode active
� � �� � � �� �
� �� � Dosing can be set manually.
� � � � ���
� � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
“System stopped” operating mode
� � � � Dosing is switched off.
WT.050.580.0AA.UA.IM.1206 121
MFC Display and Operator Controls
MFC ANALYZER / CONTROLLER
Select
Select one
one or
or more options from
more options fromthe
thelist
listprovided.
SELECT
provided.
CHANGE Change
Changethetheoperating
operatingmode.
mode.
ENTER Confirm
Confirm your
your selection.
selection.
CAL Select
Selectthe
the“Calibration” menu.
"Calibration" menu.
LOCK Activate
Activate password
password protection.
protection.
MENU Select
Selectaamenu.
menu.
MODE Select
Selectthe
the“Mode”
"Mode" menu.
menu.
UNLOCK Start
Startdeactivation
deactivationof
of password
passwordprotection.
protection.
BACK Move
Move up
up one
one level.
level.
---> Open
Opennext
nextdiagnosis
diagnosisdisplay.
display.
13s
1s Feed
Feeddelay delay
The
Thetime time until dosing resumes
until dosing resumes after
after inter-
interruption.
ruption.
100% Positioner feedback (Ym)
100% Positioner feedback (Ym)
IfIf the
the display
display blinks,
blinks,thethepositioner
positioneris isin in
manual mode
be activated.
manual mode and cannot be activated.
28.4°C
28.°C Sample
Samplewater watertemperature
temperature
mA? 1/5 Error indication active (display bottom
mA? 1/5 Error indication active (display bottom right)
right)
The system has detected an error. The error can be sp
The
the system
table inhas detected
chapter 5.5 an error. The
“Errors" on pageerror173. The num
can be specified
combination with the
states the table
seriesinnumber
section 4.5 of the error mes
the total number of error messages
“Errors”. The number combination states (in this case the fir
total of five).
the series number of the error message and
the total number of
Display error messages
(display (in this
bottom right)
case the first error of a total of five).a menu and stipula
This number enables allocation to
sequence within the menu.
Display number (display bottom right)
PI 85 s This
YPI stopnumber
timeenables
display.allocation
The time ittotakes
a menu after a spike in t
and stipulates the sequence within
for the single feedback closed loop the in the comb
control
reactivate.
menu.
PI 85 s YPI stop time display. The time it takes
after a spike in the flow rate for the single
feedback closed loop control in the combi-
control to reactivate.
WT.050.580.0AA.UA.IM.1206 122
MFC ANALYZER / CONTROLLER
Operator Controls
Softkey
• Activate the function shown on the
graphic display with the keys.
Up
• Move up one level.
• Display the previous option.
• Increase the value.
Down
• Move down one level.
• Display the next option.
• Decrease the value.
Left/right
• Change the column in the menu.
• Change the position in the displayed
value (cursor menu).
• Move forwards or backwards by seven
hours in the trend graph.
Escape
ESC • Cancel the entry without saving the
new value.
• Move up one menu level.
Enter/Acknowledge
• Acknowledge alarm message.
• Set the running delays to zero.
• Delete adaption error.
• Acknowledge max. dosing time to
reactivate dosing.
• Activate the function chosen pn the
graphic display to adjust values.
WT.050.580.0AA.UA.IM.1206 124
MFC ANALYZER / CONTROLLER
4.2.1 Password
WT.050.580.0AA.UA.IM.1206 125
MFC ANALYZER / CONTROLLER
4.2.2 Operation
You have the following options starting from the basic display:
WT.050.580.0AA.UA.IM.1206 126
MFC ANALYZER / CONTROLLER
• Main menu
• Module type 1, e.g. Cl2 free 1
• Module type 2, e.g. Cl-sat. 2
• Module type 3, e.g. pH 3
• Module type 4, e.g. mA/V input 4
• nputs/Outputs
• Alarms
• System
• Diagnosis
• Calibration
• Mode
The “Calibration” and “Mode” menus are opened with the corresponding
keys directly from the basic display. All other menus can be accessed
with the “MENU” softkey.
The following pages show the 11 individual menus. The displays contain
the default settings. If options are available for individual parameter
values, these are listed to the right of the illustration of the display in-
formation.
NOTE: The actual displays on your device can vary from those il-
lustrated. The displays and menus depend on the number of sensor
measuring modules installed and the selected settings. The display
numbers are featured on the bottom right of the display. These num-
bers are frequently referred to in the following chapter.
WT.050.580.0AA.UA.IM.1206 127
MFC ANALYZER / CONTROLLER
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
� � � � � �
� � � � �
Basic display 2
� � � � � � ��� � � � � � �
� �� � � � � �
� � � � � ��������
� � � � � � � � �������
� � �
� �� �
� �� � � � � �
�� � � � �
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ��� � �� � �
Basic display 3
� � � � � �
� � � � � � �� � � � � � � � � � � �
� �� � � �� � � � � � � ��
�
� � � � � � � � � � � �
� ���� � � � � � � � �
� � �
���� � � � �� ��� � � �� ��� � � � � � � ������������������ � � � � �� � � �� � � � �
� � � � � �
Trend Graph 1
� � � � � � ��� � � � � � � � � � � � � �� � �� � �
� �� �
� �� �
� �� �
� � � � � � � � � � � � � � � � � � � � �
���� � � � �� ���� � � ��� ��� � � � �� � � � � � � � � � � � �
� � � � � �
� � � � �� � �� � �
Trend Graph 2
� � � � � � ��� � � � � � �
� �� �
� �� �
� �� �
� � � � � � � � � � � � � � � � � � � � � � � �
���� � � � �� ���� � � ��� � � � �� � � � � �
�� � � � �
� � � � � �
Trend Graph 3
� � � � � � ��� � � � � � � � � � � �� � �� � �
� � �
� � �
� � �
� � � � � � � � � � � � � � � � � � � � �
���� � � � �� ���� � � ��� ��� � � � �� � � � � � � � �
� � �
�� � � � �
Trend Graph 4
� � � � � �
� � � � � � ��� � � � � � � � � � � �� � �� � �
� �
� �
�
� � � � � � � � � � � � � � � � � � � � �
���� � � � �� ���� � � ��� ��� � � � �� � � � � � � � � � � � � �
�� � � � �
WT.050.580.0AA.UA.IM.1206 128
MFC ANALYZER / CONTROLLER
Basic Display 1
Top status line
• Mode
• Measured value from module 1, e.g. free chlorine (mg/l) as a digital
display and bar graph with module designation (optional)
• Measured value from module 2, e.g. pH value (pH) as a digital
display and bar graph with module designation (optional)
• Flow rate display Wq
• Softkey display
• Feed delay(s) (set in display 1.7.2) e.g. after sample water stop
or change of mode from manual to automatic.
• Error message. The display alternates if there are several error
messages.
• Sample water temperature (°C)
Basic Display 2
Top status line
• Mode
• Measured value from module 3, e.g. Redox potential (mV) as a digi-
tal display and bar graph with module designation (optional)
• Measured value from module 4, e.g. conductivity (µS/cm) as digital
display and bar graph with module designation (optional)
WT.050.580.0AA.UA.IM.1206 129
MFC ANALYZER / CONTROLLER
Basic Display 3
Top status line
• Mode
• Free chlorine (mg/l) - module 1 measured value display
• pH value (pH) - module 2 measured value display
• Redox potential (mV) - module 3 measured value display
• Conductivity (mS/cm) - module 4 measured value display
• System name
• Digital inputs activated
• Password protection activated
• Selected measured parameter unit
• Date of the displayed diagram (set in display 1.7.1)
• Error message
Bottom line
• Softkey display
• Measurement parameters (can be selected in display 1.7.3)
WT.050.580.0AA.UA.IM.1206 130
MFC ANALYZER / CONTROLLER
Basic Display
� � � � � �
� � � � � � ��� � � � � � � � � � � � � �
� �� � � �� �
� � �� � �� �
��� �� � � �� �� � � �
� � � � � � � �
� ��
� �� �
� �
� � � � � � � �
� �� �
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� � � � � � � � � � � ������������������ � � ������������� � �� � �
�� � � � �
� � � � � � � � � � � �
Menus Select � �
� �
� �
� �
� �
� �
��
��
��
���
���
���
� �
� �
� �
��
��
��
� �
� �
� ��
��
�
�
� � ��
�� � �
� � �� �
���� � �� � ��
� � �� � � � �
� � � � �� ��� � � �� � � � �� � � � � ��
�
� � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � �
� � � � � � �
� � � � � � � �
Dosing
� � � � � � ��� � �� ��� � � � �� � �� ��� ��� � �� � ��
� � � � � � � �
� � � � �� � ��� � � � �� � � �
� � �� � ��� � � ��� � � � � � ��
� � �� � ��� � �� � � � � �
� � �
� � �� � ��� � �� � � � � �
�� � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � �
Control
� � � �� � �� � � � � � � � � � � �� � �� � � � �
� � �� � �� ��� � � �� � � � ��
� � �� � �� ��� � � �� � � �� �� �� � �
� � � �� � � �� � �� �� � � � � �
� � � ���� � �� ��� � � �� � � �� �� �� � �
� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � �
Setup
� �� � �� � � �� � � � � � �� �� � � �� �
� �� � �� �� � � ��� � � � � �� � � �
� �� � � �� � � � � �� � � � ��
� �� �� � � � �� � � � � �� � � �
� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � �
Parameter
� � � � � � � � � � �
� � � � � ��� � � � �� � �� �
� � � �� � � � � �� � �� �
� � � ���� � � � � � � � �� � � �
� � � �� � ���� � �� � � � � ��
� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � �
Measuring Range � �
� �
� �
� � �
� � �
�� �
�� � � � ���� � � � � �
���� � � � �
� �
�
�
�� �
�� � �� � �
� ��
� � � �� � � � � � � �
� � �
� � �� � � � � � � � � � �
�� � � � �
� � � � � � � � � � � � � � � � � � � �
Limit Value 1
� � � � � � � � � � � � � �
� �� � � � � �� � � � ��
� � � � � � � �� � � � ��
� � � �� � � � �� � � �� � � �� � � � ��
� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � �
Limit Value 2
� � � � � � � � � � � � � �
� �� �� � �� � � �� � � � � �
� � � �� � �� � � �� � � � � �
� � � �� � � � �� ���� � � � � �
� �� �� � �� � � � � �
� � � �� � �� � � � � � � � �
�� � � � �
� � � �� � � � �� ���� � � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � �
Adaption
� � � � � � � � �
� � � � � � � � ��
WT.050.580.0AA.UA.IM.1206 131
MFC ANALYZER / CONTROLLER
Basic Display
Refer to main menu
Selection Menu
Display of all available menus
Dosing
Control output
Positioner with feedback (5kOhm)
Positioner with Ym (1kOhm)
Positioner without Ym
Dosing pump 2p.
Dosing pump 3p.
Pulse pump 2p.
Solenoid pump 3p.
Analog output 2p.
Analog output 3p.
Dosing output
Tp 10 s – 180 s (60 s)
Ty 10 s – 180 s (90 s)
Sampling time T 1–20 s
Ym calibration auto
Ym calibration 0%
Ym calibration 100 %
Pulses max./min 100/120
Hysteresis Depends on measuring range
0.01 to 0.50 / 00.1 to 5.0 / 1 to 50
min. ON 1min00s – 59 min59s
Control
Control mode Combi (fuzzy)/ratio/feedback control
Wi setpoint Measuring range
Setpoint source internal / external / internal with DI 3 /
external with DI 3
Dosing factor 0–100 %
Dos. factor source internal / external / internal with DI 3 /
external with DI 3
WT.050.580.0AA.UA.IM.1206 132
MFC ANALYZER / CONTROLLER
Setup
Flow rate source out / flow rate measured value
Flow rate direction direct / inverse
Control variable 2 out / measured value X
X direction direct / inverse
Control direction direct / inverse
X factor 0.1 to 4.0
Ymin 0–100 %
Ymax 0–100 %
Parameter
Xsh 0.0 to 5.0 %
Tconst 30 s – 10 min
Tvar 30 s – 10 min
Max. lin. corr. 0–100 %
Control factor 0.1 to 10
Xp 1–1000 %
Tn 0.0 to 100.0 min
Measuring Range
Range Start pH mV mA/V
WT.050.580.0AA.UA.IM.1206 133
MFC ANALYZER / CONTROLLER
Format mA/V
Sensor modules:
000.0 / 00.00 / 0000
Upot Cl2
Signal mA/V
Factor mA/V
WT.050.580.0AA.UA.IM.1206 134
MFC ANALYZER / CONTROLLER
pH Compensation Cl2
Cl2 free (1) – CL2 free (4) (only equipped sensor modules selectable
for 3 electrode cells)
Reference Temp. LF
Limit Value 1
Min within measuring range
Max within measuring range
Hysteresis Depends on measuring range
0.01 to 0.25 / 00.1 to 05.0 / 1 to 50
Limit Value 2
Min Ym/Yout 0–100.0 %
(not in single feedback closed-loopcontrol)
Max Ym/Yout 0–100.0 %
(not in single feedback closed-loop control)
Ym is only output if the actuator feedback is
present, otherwise the controller output is Yout
Hysteresis 0.1 to 5.0 %
Min Ypi 0–100.0 % (for combi-controller only)
Max Ypi 0–100.0 % (for combi-controller only)
Hysteresis 0.1 to 5.0 % (for combi-controller only)
WT.050.580.0AA.UA.IM.1206 135
MFC ANALYZER / CONTROLLER
Adaption
Adaption is only available for single feedback closed-loop control
with “DES” modules.
Tu 1–3600 s (60 s)
Ts 0.1 to 480.0 min (10 min)
WT.050.580.0AA.UA.IM.1206 136
MFC ANALYZER / CONTROLLER
Basic Display
� � � � � �
� � � � � � ��� � � � � � � � � � � � � �
� �� � � �� �
� � �� � �� �
��� �� � �
� � � � � � � �
�� �� � � � � ��
� �� �
� �
� � � � � � � �
� �� �
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� � � � � � � � � � � ������������������ � � ������������� � �� � �
�� � � � �
Menu Select
� � � � � � � � � � � �
� � � � �� �� � � � �� � � � �� � � �� ���� � �� � ��
�� � � � �
� � � � �� �� � � � �� � � � �� � � � �
� � � � �� �� � � � �� � �� � � � �� �
� � � � �� �� � � � �� � � � �� � � � � ��
�
� � � � � � � � � � � � � � � � � � � � � � � � � �
� � � � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � �
� � � � � � �
Dosing
� � � � � � � �
� � � �� � ��� � �� ��� � � � �� � �� ��� � ��� �
� � � � � � � �
� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � �
Control
� � � �� � ��� � � � � � � �� � �� ��� � � ��� � �
� � �� � �� ���� � � �� � � �
� � � �� � ��� �� � � ���� � �� � � �
� � � �� � ��� � �� �� � � � � �
� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � � �
Parameter
� � � � � � � �
� � � � � � � � ��
� � � � � � � � � �
� � �� � � �� � �
� � � � � � ��� � � � �
� � �
� � � � � � � � � � � � � � � � � � � � �� � � � �
Time setting
� � � � � � � � � � � � �
� �� � � �� � � �� � ��� � � � ��� � �� � ���� � �� � ��� �
� �� � � �� � � �� � ��� � � � ��� � �� � ���� � �� � ����
� � �
� � � � � � � � � � � � � � � � � � � � �� � � � �
Trim Setpoint
� � � � � � � � � � � � �
� � �� � �� ��� � �� � � � �
� � � � � � � � � � �
� � � � � � � � � �
� � � � � � � �
� � � � � ��� � � � �� � �� �
� � �
� � � �� � � � � �� � �� �
� � � � � � � � � � � � � � � � � � � � �� � � � �
� � � � � � � � � � � �
� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � �
Limit Values
� � � � � � � � � � � �
� �� � � � � �� � � �
� � � � � � � �� � � �
� � � �� � � � �� ���� � � �� � � �
� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � �
Adaption
� � � � � � � �
� � � � � � � � �
� � � � � � � � ��
� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � �
WT.050.580.0AA.UA.IM.1206 137
MFC ANALYZER / CONTROLLER
Basic Display
Refer to main menu
Menu Select
Display of all available menus
Dosing
Control output
Positioner without Ym
Dosing pump 2p.
Dosing pump 3p.
Solenoid pump 2p.
Solenoid pump 3p.
Analog output 2p.
Analog output 3p.
Dosing output
Ty 10–180 s (90 s)
Tp 10–180 s (90 s)
Pulses max./min 100/120
Hysteresis Measuring range
min. ON 1 min 00s – 59 min 59 s
Control
Control mode Ratio / Feedback control
Setpoint Measuring range
Control direction direct / inverse
Dosing factor 0–100 %
Parameter
Xp 1–1000 %
Tn 0.0 to 100.0 min
Xsh 0.0 to 5.0 %
Ymin 0.0 to 100.0 %
Ymax 0.0 to 100.0 %
WT.050.580.0AA.UA.IM.1206 138
MFC ANALYZER / CONTROLLER
NOTE: The Ymin and Ymax parameters are not displayed if “Po-
sitioner without Ym” is selected and during ratio control.
Time setting
Time contact 1 Selection of 9 on-duty and off-duty cycles max.
Time contact 2 Selection of 9 on-duty and off-duty cycles max.
1 : 00:00 - 00:00 --
Switching Duty Off-duty Day Mo-Su
cycle 1-9 cycle cycle (If “--” is displayed,
hh:mm hh:mm the switching cycle
is inactive)
Setpoint trim
Trim setpoint Switch on/off
Trim factor 0–100 %
Setpoint range 5–50 %
Xsh 1–10 %
Tconst 30 s – 10 h
Tvar 30 s – 10 h
Measuring range
See module type 1 under “Measuring range”
Limit values
Min within measuring range
Max within measuring range
Hysteresis Depends on measuring range
0.01 to 0.25 / 00.1 to 05.0 / 1 to 50
Adaption
Adaption is only available for single feedback closed-loop control
with “DES” modules.
Tu 1–3600 s
Ts 0.1 to 480.0 min
WT.050.580.0AA.UA.IM.1206 139
MFC ANALYZER / CONTROLLER
� � ��
�
� �
� �
� � �
�
� �� �
�� � � � � �
� �� �
��� �� � � �� ��� � � � ��
� �� � � �� �
� �
���� �� � � � �� � ���� �� � ��
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
�� � � � �
Menu Select
� � � � � �� � �� � �
� � � � �� ��� � � ��� � � �� � � �� ���� � �� � ��
� � � � �� ��� � � ��� � �� � � � �
� �� � � � �
� � � � �� ��� � � ��� �� � � � �� �
� � � � �� ��� � � ��� � � �� � � � � ��
�� � � � � �� � � � � �� � � � � � ���� ���
�� � � � � �
� � � � �� � � � �� � �� � � �� � �
Module Type (3) Menu � � � � �� � � � �
� �� ���� � �� � �
�� � � � �
�� � � � � �� � � � � � ��� �� �
�� � � � � �
� � � � � � � � � � � �� � � � � �
Measuring Range � � � � � �� �� ��� �
� � � � � �� � � � �
� � �
� � �
� � ��� � � � �
� � �
�� � � � �
�� � � � � �� � � � � � �� �� �� �
� �� �� � � � �� � �
Limit Values � �� � �
� � � � �
�
�
� � � � �
� � � � �
� � � �� � � � �� ���� � � � � � �
� � �
�� � � � � �� � � � � � � � � � � � � �� � � � �
Basic Display
Refer to main menu
Menu select
Display of all available menus
Measuring range
See module type 1 under “Measuring range”
Limit values
Min within measuring range
Max within measuring range
Hysteresis Depends on measuring range
0.01 to 0.25 / 00.1 to 05.0 / 1 to 50
WT.050.580.0AA.UA.IM.1206 140
MFC ANALYZER / CONTROLLER
Basic Display
� � � � � �
� � � �� � ��� � � � � � � �� � � � � �
� �� � � �� �
� � �� � �� �
��� �� � � �� ��� � �
���� �� � ��
� ��
� �� �
� �
���� �� � �
� �� �
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
�� � � � �
� � � � � � � � �� � �
Menu Select � �
� �
� �
� �
��
��
���
���
� �
� �
���
���
�
�
� ��
�
� � �� ���� � �� � ��
�� � � � � �� � � � �
� � � � �� ��� � � ��� �� � � � �� �
� � � � �� ��� � � ��� � � �� � � � � ��
�� � � � � �� � � � � �� � � � � � ���� ���
�� � � � � �
� � � � �� � � � �� �� � � � �� � �
Module type (4) Menu � � � � �� � � � �
� �� ���� � �� � �
�� � � � �
�� � � � � �� � � � � � ��� �� �
�� � � � � �
� � � � � � � � � � � �� � � � � �
Measuring Range � � � �
� � � �
� �� �� ��� �
� �� � � � �
�
�
� �� �
� �� �
� � ��� � � � �
� � �� � �� � � � �� �
� �� � � �� � � � ���� � � � � � �
� � � �� �� � � � �� �� � � � �
�� � � � � �� � � � � � �� �� �� �
Limit Values
� �� �� � � � �� � �
� �� � � � � � �� � � �
� � � � � � � � �� � � �
� � � �� � � � �� ���� � � � �� � � �
� � �
�� � � � �
�� � � � � �� � � � � � �� �� �� �
Basic Display
Refer to main menu
Menu Select
Display of all available menus
Measuring Range
See module type 1 under “Measuring range”
Limit Values
Min within measuring range
Max within measuring range
Hysteresis Depends on measuring range
0.01 to 0.25 / 00.1 to 05.0 / 1 to 50
WT.050.580.0AA.UA.IM.1206 141
MFC ANALYZER / CONTROLLER
Basic Display
� � � � � �
� � � � � � ��� � � � � � � � � � � � � �
� �� � � �� �
� � �� � �� �
��� �� � � �� �� � � �
��� � �� � �
� ��
� �� � ��� �� � �
� � � �� �
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
�� � � � �
� � � � � � � � � � � �
Menu Select � �
� �
� �
� � �� ���
� � �� ���
� � �� ���
� �
� �
� �
��
��
��
� �
� �
� ��
���
�
�
� � �� ���� � �� � ��
�� � � � �
� � �� �
�� � � � �
� � � � �� ��� � � �� � � � �� � � � � ��
�� � � � � �� � � � � �� � � � � � � � � � � � �
�� � � � � �
�� � � � � �� � � � � � � � � � � �
�� � � � � �
� � � �
Flow Rate
� �� � � �� � � � ���� � � �
� � � �� �� � � � ��
� � ��� � � ���
� � �� � � �� � � � � � � �
� � � � � �� �� ��� � � � � � ��� � � �
� � � � � �� � � � � � � � � ���
�� � � � �
�� � � � � �� � � � � � � � � � � � �
� � � � � � � � � � � � � � � � �
� � �
�� � � � �
�� � � � � �� � � � � � � � � � � � �
Ext. Setpoint / DF
� � � � � � � � � � � � � � � � �
� �� � � �� � � � ���� � � �
� � � �� �� � � � ��
� � �
�� � � � � �� � � � � � � � � � � � � �� � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � �
� � � �� � � � �� ���� �
�
�
� �
� �
� � �
�
�
� � �
�� � � � �
�� � � � � �� � � � � � � � � � � � �
� � � � � � � � � � � � � �
Digital Input
� � � � � � � � � � � � � �
� ���� � � � � ��� ��� �� �
�� ��� � � � � � ��� ��� � � � �� � �
� ���� � � � �� � � � �� � �
� �� � �� ��� � � ��� �� � ��
� �� � �� ��� � � � � � � � � �
�� � � � �
�� � � � � �� � � � � � � � � � � � �
Interface
� � � � � � � � �
� � � � � �� �� � � �� �� �
� � � � � �� � � � � � � � � �
� � � �� � � � � � ��� � ��� � � �
� � �
�� � � � �
�� � � � � �� � � � � � � � � � � � �
WT.050.580.0AA.UA.IM.1206 142
MFC ANALYZER / CONTROLLER
Basic Display
Refer to main menu
Menu Select
Display of all available menus
Menu Inputs/Outputs
Display of all available input/output settings
Flow Rate
Signal 0–20 mA, 4–20 mA, 0–10 V
Factor 0.1 to 4.0
Unit Max. 5 digits (any combination)
Format Measurement display
000.0 / 00.00 / 0000
Range-start Unlimited
Range-end Unlimited
Ext. Setpoint/DF
Signal 0–20 mA, 4–20 mA
Factor 0.1 to 4.0
WT.050.580.0AA.UA.IM.1206 143
MFC ANALYZER / CONTROLLER
Digital Input
DI (1), DI (2), A function can be assigned to each
DI (3), DI (4) digital input.
Stop control Yout = 0 %, dosing, mA analog output =
0%
Constant control Yout remains constant (i.e., the control
signals are kept constant).
Constant mA All mA outputs remain unchanged,
while DI is active.
Ratio If the DI is active, the control mode
switches from MOD 1 to ratio control.
Only if the control mode changeover is
set to combi-control via DI3.
Feedback control If the DI is active, the control mode
switches from MOD 1 to single feed-
back closed-loop control. Only if the
control mode changeover is set to
combi-control via DI3.
Interface
RS232 Out, printer, IAP download, parameter
RS485 address Bus addresses 00 to 31 (0)
CAN segment addr. CAN bus segment addresses
0 – 63 (1)
WT.050.580.0AA.UA.IM.1206 144
MFC ANALYZER / CONTROLLER
Basic Display
� � � � � �
� � � � �� � � �� � � � � � � � � � � � � �
� �� � � �� �
� � �� � �� �
��� �� � � �� �� � � �
�� � �� � � �
� ��
� �� � �� � �� � � �
� � � �� �
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
�� � � � �
Menu Select
� � � � � � � � � � � �
� � � � �� �� � � �� � � �� � � � �� ���� � �� � ��
� � � � �� �� � � �� � � � �� � � � � �� � � � �
� � � � �� �� � � �� � �� � � � �� �
� � � � �� �� � � �� � � � �� � � � � ��
�� � � � � �� � � � � �� � � � � � � � � � � � �
�� � � � � �
� � � � � � � � � � �
Menu Alarms
� �� �� �� � � �� �� � � � � �� � � � � � �� � � � � � � � � � � �
� �� �� �� � � �� �� � � � � �� �
� �� �� �� �� ��� � � ��� �
� �� �� �� � � �� � � �� �
� � �
�� � � � � �� � � � � � � � � � � � � � � �� � � � �
�� � � � � �
� � � � � � � � � � � � � � � � � � � �
� � � � � � � � � � � � � � � � � � � �
� � �
�� � � � � �� � � � � � � �� � � � � � � � � � � � � �� � � � �
� �� � � �� � ��� � � ���� �� �
� ��� � � � �
� � � � � � � �
� � � ��� � � �
� � �
�� � � � � �� � � � � � � � � � � � � �� � � � �
Basic Display
Refer to main menu
Menu Select
Display of all available menus
Menu Alarms
Display of all available settings
WT.050.580.0AA.UA.IM.1206 145
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 146
MFC ANALYZER / CONTROLLER
Basic Display
� � � � � �
� � � �� � ��� � � � � � � �� � � � � �
� �� � � �� �
� � �� � �� �
��� �� � � �� ��� � �
���� �� � �
� ��
� �� � ��� �� �
� � � �� �
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
�� � � � �
Menu Select
� � � � ��� � �� � �
� � � � �� ��� � � ��� � � � � ���� �� �� � � ��
� � � � �� ��� � � ��� � � �� � � � ���� � �� � � � �
� � � � �� ��� � � ��� �� � � � �� �
� � � � �� ��� � � ��� � � �� � � � � ��
�� � � � � �� � � � � �� � � � � � ���� ���
�� � � � � �
Menu System
� � � � �� � � � � � �
� � � � � � � � � �� ��
� � �� �� �� � � � �
� �� � �� �� � � � �
�� �� ��� � �
�� � � � � �� � � � � � �� �� �
�� � � � � �
General
� � � � � �
� �� � � � � � � �� �
� � �� � � � � � �� � �� �
� ���� � � � � � ��� � �
� � �� ��� � � ��� � � � � �
� � � � � � � � � � � � � ��� �
� � �
� � � �� � � � � � � � � � � �� � ���
�� � � � �
�� � � � � �� � � � � � �� �� �� �
� � � � � �
Safety �� � � �� � �� � � �
� � � � ��� � �� � �� � �� � �
� �� � ��
� � ��
� � � �� � �� � � � � � �� �� � � � �
� � � �� � � � � � �� �� � � � �
� � �
�� � � � � �� � � � � � � � � � � � � �� � � � �
Trend Graph
� �� � �� �� � � � �
� � � � � � ��� � � � �� �� � �
� � � � � � ��� � � � � ��� �
� � � � � � ��� � � � � ��� �
� � � � � � ��� � � � � ��� �
� � �
�� � � � � �� � � � � � � � � � � � � �� � � � �
Module Designation
� � � � �� �� � � �� � � � �� � �
� � � � �� �� � � � �
� � � � �� �� � � �
� � � � �� �� � � �
� � � � �� �� � � �
� � �
�� � � � � �� � � � � � � � � � � � � � �� � � � �
System Reset
� � � � � � �� � � � � �
� � � �� � ��� � �� ��� � � �
� � �� �� �� � � � � � � � �
� �� � � � � � �� � �� � � � � �
� � �� �� �� � � �� � � �� �� � �
� � �� �� �� � �� �� � � � � � � � � �
� � �
�� � � � � �� � � � � � ��� �� �� � �� � � � �
Basic Display
WT.050.580.0AA.UA.IM.1206 147
MFC ANALYZER / CONTROLLER
Menu Select
Display of all available menus
Menu System
Display of all available system settings
General
Time (hh:mm) Current time
Date (dd.mm.yy) Current date
Dampening1) off / low / high
Hold function Off / On (see chapter 4.4 “Calibration”)
Language German, English, French
System name 12 characters, each with character set
A-Z and digits 1-9 including special
characters
Safety
Feed delay2) 00:00 – 10:00 (03min : 00s)
Samp. water delay3) 00:00 – 10:00 (01min : 00s)
(sample water delay)
System password four-digit numeric code
(activate with Softkey “LOCK” in the
“Menu Select” window)
Calibration password* four-digit numeric code
(activate with Softkey “LOCK” in the
“Menu Select” window)
Trend Graph
Channel 1 to 4 Assignment of a measured value to the
line diagram. The selected measured
value is plotted in the trend graph
(can be traced back up to 7 days).
All the following measured values are
possible: module 1 - 4, flow rate Wq,
external setpoint/dosing factor,
temperature, control signals Yout from
module 1, Ypi from module 1 or Yout
from module 2. If “Off” is selected, that
channel is not recorded.
WT.050.580.0AA.UA.IM.1206 148
MFC ANALYZER / CONTROLLER
Module Designation
Module 1 to 4 Max. 6 digits, customised entry.
If blanks are entered, the module
description is switched off.
System Reset
System restart4) yes / no
Delete graph5) yes / no
Standard value6) yes / no
Delete dos. avg7) yes / no
Delete DL diagram8) 0–100 %
1) 8)
NOTE: The system settings marked with to are explained
below.
WT.050.580.0AA.UA.IM.1206 149
MFC ANALYZER / CONTROLLER
Basic Display
� � � � � �
� � � �� � ��� � � � � � � �� � � � � �
� �� � � �� �
� � �� � �� �
��� �� � � �� ��� � �
���� �� � �
� ��
� �� � ��� �� �
� � � �� �
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
�� � � � �
� � � � � �� � �� � �
Menu Select
� � � � �� ��� � � ��� � � � � ���� �� �� � � ��
� � � � �� ��� � � ��� � � �� � � � ���� �
� � � � �� ��� � � ��� �� �� � � � �
� � � �� �
�� � � � � � � � � �� ��� � � ��� � � �� � � � � ��
�� � � � � �� � � � � �� � � � � � ���� ���
�� � � � � �
Diagnosis
� �� � � � � � �� �
� � � �� �� �� �� � � � ��� � � � � � � � �� � �
� � � �� � �� � � �� � �
�� � � �� �� � �� � �� �� � � � �
� � � �� � �� � �� ��
� � �
�� � � � � �� � � � � � ���� �� ��� �� � � � �
�� � � � � �
� �� � � � � � �� � �� � � �� ��
Diagnosis (1) Cl2 �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
�� � �� � �� � ����� � � � � ������� �� � � � �������� �� � � �� � �������� � �� � �
� � �� � �� � ����� � � � � ������� �� � � � �������� �� � � �� � �����������
� � �
�� � � � � ������ � � ��� � � � � � � � �� � � � �
������ � � ���
Diagnosis (2) pH
� �� � � � � � �� � �� � � �� �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � �� � �� � ������� � � � � ����� �� � � ����� � �� � � �� � ���������� � � �� � �
� � �� � �� � ������� � � � � ����� �� � � ����� � �� � � �� � �������������
� � �
�� � � � � ������ � � ��� � � � � � � � �� � � � �
������ � � ���
Diagnosis (3) mV
� �� � � � � � �� � �� � � �� �
� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
� � �� � �� � ����� � � � � ����� � �� � � ������������������������������ � � � �� � �
� � �� � �� � ����� � � � � ����� � �� � � ����������������
� � �
�� � � � � ������ � � ��� � � � � � � � �� � � � �
������ � � ���
� � �
�� � � � � � � � � � � � �� � � � �
WT.050.580.0AA.UA.IM.1206 150
MFC ANALYZER / CONTROLLER
Basic Display
Refer to main menu
Menu Select
Display of all available menus
Diagnosis
Display of all available diagnosis displays
pH Diagnosis
Date and time of the last 5 calibrations
pH7 Signal offset at pH 7 in mV
Spread/pH mV signal of the pH sensor based on
pH1 step
Offs manual offset in pH
(Menu 2.1.2 - Offset pH)
Cur. (U) Current mV sensor signal
Redox Diagnosis
Date and time of the last 5 calibrations
Offset Signal offset of the mV sensor in mV
Cur. (U) Current mV sensor signal
WT.050.580.0AA.UA.IM.1206 151
MFC ANALYZER / CONTROLLER
F¯ Diagnosis
Date and time of the last 5 calibrations
Zero point Established sensor zero point signal
Decade mV signal of the sensor based on
1 decade (log)
Cur. (U) Current mV sensor signal
Conductivity Diagnosis
Date and time of the last 5 calibrations
Spread Conductivity measuring cell calibration
factor
Cur Displays current sensor current in mA
Displays current sensor voltage in mV
Displays temperature of the conductivity
sensor
WT.050.580.0AA.UA.IM.1206 152
MFC ANALYZER / CONTROLLER
������ � � ���
� �� � � � � � �� � ��� � � �� �� � � � � �� � � � � �
� �� � � � � �� � � �� � �
� � �� � � �� � � �� � �� �
��� �� � � � �� �
� �� ��� � � � �� � � � �
� � �� � �
�
��
� ��
� ��
� �� �
� �
� �� � � � � � � ��
� � � � ���
� � � �� � �� � �� � �� �
���� � � � � ������ � � ��� ������������ ���� �� �� �
� � � � � � � � � � � � � � � � � � � � � � � � � �
Dosing Average Diagnosis �� ��� �� � �
�� ��� � � � �
�
�
� � �
� � �
���� � � �
� � �
�� � � � � � � � � � � � � �� � � � �
������ � � ���
������ � � ���
WT.050.580.0AA.UA.IM.1206 153
MFC ANALYZER / CONTROLLER
Inputs/Output Diagnosis
Information on
• The assignment of the mA outputs
• The current mA output in mA and %
• The current dosing output Yout for module 1 as well as Ypi
• The current dosing output for module 2
• The current switching conditions of the relays:
Relay off Relay on
• The selected application
• The send/receive condition of both interfaces RS485, RS232
and CAN
• The current switching conditions of the digital inputs 1, 2 and 3
• The current flow measurement (Wq) input signal
• he current input signal of the external setpoint (We) or external
dosing factor (ext. DF)
WT.050.580.0AA.UA.IM.1206 154
MFC ANALYZER / CONTROLLER
Basic Display
� � � � � �
� � � �� � ��� � � � � � � �� � � � � �
� �� � � �� �
� � �� � �� �
��� �� � � �� ��� � �
���� �� �
� ��
� �� � ���� �� �
� � � �� �
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
��� � � ���
Calibration
� � �� � � � � �� �
� � � � �� ��� � � ��� � � � � � �� �� � � ��� �
� � � � �� ��� � � ��� �
� � � � �� ��� � � ��� � �� � � � �
� � � � �� ��� � � ��� �
�� � � � � �� � � � � �� � � � � � ���� �� ���
�� � � � � �
� � ���� � � � �� � � � � �� � � �
Module Type (1) Calibration � � � � ������
� � �
��� �� � � � ��
� �� �
(Example Cl2) � � ��
� �� � � � � �
�� � � � � �� � � � � � � � � � � � � � � �� � � � �
� � �� �� � � � �� �� � � � �� � � �
Module Type (2) Calibration
� � ���� � � ��������������� �� �
� � ���� � � �� �
(Example pH)
� � ����� �
� �
� �
���� ����
� � ���
� � ��� �
� � � � � � ��
� � ��
� � � � �
� �
�� � � �
� � �
�� � � � � �� � � � � � �� �� �� � �� � � � �
� � ���� � � � �� �� � � � �� � �
Module Type (3) Calibration � � ���� � � ���������������� � � � �
� � �
(Example mV) � �
� � � � � � � �
�� � � � � �� � � � �
�� � � � � � �� �� �� �
Temperature Calibration
� � ��� � � � � � � �� � � � � � � � � � � �
� � ���� � � ����� � � �� � �
� � ��� � � � � � ��
� � � � � � � � � � ���� ����� �
�� � �� � �
� � �
�� � � � � �� � � � � � �� �� �� �
�� � � � �
Basic Display
Refer to main menu
Calibration
Display of all available calibration options
WT.050.580.0AA.UA.IM.1206 155
MFC ANALYZER / CONTROLLER
F-calibration
Cal. lower value within measuring range
Cal. upper value within measuring range
Temperature Calibration
Cal. °C 0–50°C
Unit °C/°F
PT1000 Switch automatic temperature
compensation on or off
If the PT1000 is switched off a
manual temperature value can be
entered into the calibration menu during
a pH measurement.
WT.050.580.0AA.UA.IM.1206 156
MFC ANALYZER / CONTROLLER
Mode - Menu 3
Display using an example (module 1 = Cl2 free and module 2 =
pH).
� � � � � �
� � � �� � ��� � � � � � � �� � � � � �
� �� � � �� �
� � �� � �� �
��� �� � � �� ��� � �
���� �� � �
� ��
� �� �
� �
���� �� � �
� �� �
� � �� � � �� �
� �� �
� � � � ���
���� � � � �� ���� � � ��� ��� � � � � � � � ������������������ � � ������������� � �� � �
�� � � � �
� � � � � � � � � � � � � � � � � � � � � �
� � �
� �
�� � � � �
�� � � � � � � � � � � � � � �
� � �
�� � � � � �� � � � � � �
� � � � � � � � � � � � � � � � � � � � � � � � � � �
� �� ��� � � �� � � �� � � �� � � � �� � �� ��� � � �� � � � � � �
� � ��� �� �� � � �� � � � � � ��� �� � � � �
� � � � � �
�� � � � � �� � � � �
�� � � � � �� � � � � � �� �� � �� � � � � �� � � � � � �� �� �
Basic Display
Refer to main menu
Manual / Automatic
AUTO / MANUAL toggle using the “CHANGE” softkey
Setpoints
Cl2 free(1) within measuring range
pH (2) within measuring range
MANUAL
In MANUAL mode dosing is not automatically controlled. The
values must be continuously monitored.
WT.050.580.0AA.UA.IM.1206 157
MFC ANALYZER / CONTROLLER
NOTE: When MANUAL mode is set: The pumps are off, the
positioner remains in its current position, if necessary unlock
the positioner and close either by hand or with the Man.dos.
menu.
AUTOMATIC
Automatically controls the measured variables acc. to the selected
application
STOP
STOP mode is automatically activated:
After activation:
ADAPTION
ADAPTION mode is activated, if the adaption for the single
feedback closed-loop control module 1 or module 2 is started.
WT.050.580.0AA.UA.IM.1206 158
MFC ANALYZER / CONTROLLER
4.4 Calibration
Calibration Aids
Two clips are installed in the housing cover. These clips can be inserted
into the rear panel of the housing. The clip (A) for the sensor will be
inserted into the upper catch. When the electrodes are calibrated in
the beaker with the calibration solution, the second clip (B) will be
inserted into the catch (left figure). The lower clip position is provided
for calibrating with the calibration solution bag (right figure).
A A
B
B
A Upper clip
B Lower clip
WT.050.580.0AA.UA.IM.1206 159
MFC ANALYZER / CONTROLLER
E
A Top clip with electrode mount
B Sensor
C Beaker
D Lower clip
E Ball valve on sample water inlet
1. Starting from the basic display in the main menu select the “Calibra-
tion” menu.
2. Select the “Temperature” menu item
WT.050.580.0AA.UA.IM.1206 160
MFC ANALYZER / CONTROLLER
Calibration of the 3 electrode cell for Cl2, KMnO4, O3, ClO2 and Cl2++
(pH-compensated). During calibration of the 3 electrode cell, perform
a zero point calibration and a measurement value span calibration. The
calibration process is nearly the same for chlorine, chlorine dioxide,
ozone and potassium permanganate.
1. Starting from the basic display in the main menu select the “Cali-
bration” menu.
2. Select the measurement to be calibrated from the menu, e.g. Cl2
free (1).
3. Select the “Cal.zero” parameter.
4. Close the ball valve on the sample water inlet.
WT.050.580.0AA.UA.IM.1206 161
MFC ANALYZER / CONTROLLER
For the selective ozone, chlorine dioxide and free chlorine measurements,
1-point or 2-point can be selected in the “Calibration mode” calibration
menu. 2-point calibration provides the option to compensate for possible
measuring cell zero point offsets.
Total Chlorine
WT.050.580.0AA.UA.IM.1206 162
MFC ANALYZER / CONTROLLER
1. Determine the values for free chlorine and total chlorine as de-
scribed under “Total chlorine”.
2. Determine the value for combined chlorine: Total chlorine minus
free chlorine.
3. Enter the determined value as described under “Total chlorine”
with the arrow keys under the parameter “Cal. combined”.
4. Confirm the entry.
Zero Point Calibration for Ozone, Chloride Oxide, Cl2 free (M)
Zero point calibration using 2-point calibration
Span Calibration Using 1-Point and 2-Point Calibration for O3, ClO2
and Cl2
WT.050.580.0AA.UA.IM.1206 163
MFC ANALYZER / CONTROLLER
4.4.4 pH calibration
pH-7 Alignment
1. Starting from the basic display in the main menu select the “Cali-
bration” menu.
2. Select the menu item “pH (1) ... pH (4)“ for the measurement
you wish to perform. The “Calibr. pH” window appears on the
graphic display.
3. Select the “Cal. pH7” parameter.
4. Place one of the supplied beakers into the bottom clip and fill
with the buffer solution “pH 7.00” or a bag with buffer solution
“pH 7.00” into the bottom clip.
5. Pull or unscrew the pH sensor from the lid of the cell body.
6. Dip the pH sensor through the top clip at least 2 cm deep into
the buffer solution and move slightly until the indicated pH value
remains constant.
7. Enter the pH value associated with the buffer temperature using
the arrow keys or or leave pH 7.00.
8. Save the value by pressing “Enter”.
Slope Alignment
9. Remove the buffer solution “pH 7.00” from the bottom clip.
10. Wash the sensor in distilled water to prevent carry-over of the
buffer solution.
11. Select the parameter “Cal. pH” and confirm the selection.
12. Place a second beaker into the bottom clip and fill with the buffer
solution “pH 4.65” or clamp a bag with buffer solution “pH 4.65”
into the bottom clip.
13. Dip the pH sensor at least 2 cm deep into the buffer solution until
the indicated pH value remains constant.
14. Use the four arrow keys to enter the determined value.
WT.050.580.0AA.UA.IM.1206 164
MFC ANALYZER / CONTROLLER
pH Calibration
1. Starting from the basic display in the main menu select the “Cali-
bration” menu.
2. Select the menu item “pH” for the measurement you wish to
perform.
3. Select the “Offset pH” parameter.
4. With the four arrow keys enter the difference between the com-
parative value and the displayed value (comparative value minus
the displayed value).
5. Save the value by pressing “Enter”.
WT.050.580.0AA.UA.IM.1206 165
MFC ANALYZER / CONTROLLER
1. Starting from the basic display in the main menu select the “Calibra-
tion” menu.
2. Select the menu item “mV (1) ... mV (4)“ for the measurement you
wish to perform. The “Cal. mV” parameter is selected.
3. Place one of the supplied beakers into the bottom clip and fill with
the buffer solution or clamp a bag with buffer solution into the bot-
tom clip.
4. Pull or unscrew the Redox sensor from the lid of the cell body.
5. Dip the Redox sensor through the top clip at least 2 cm deep into the
buffer solution until the indicated pH value remains constant.
6. Use the arrow keys to enter the determined value.
7. Save the value by pressing “Enter”.
8. Remove the Redox sensor from the top clip.
9. Insert or screw the Redox sensor into the lid of the cell body.
WT.050.580.0AA.UA.IM.1206 166
MFC ANALYZER / CONTROLLER
1. Starting from the basic display in the main menu select the “Calibra-
tion” menu.
2. Select the “Conductivity (1) ... (4)” menu item.
3. Select the “Calibration” parameter.
4. Place one of the supplied beakers into the bottom clip and fill with
the buffer solution “60.0mS/cm”* or clamp a bag with buffer solution
“60.0 mS/cm”* into the bottom clip.
5. Pull or unscrew the conductivity sensor from the lid of the cell
body.
6. Dip the conductivity sensor through the top clip into the calibration
solution to the bottom of the beaker.
7. Pull out the conductivity sensor and rinse off with distilled water.
8. Repeat the dipping and rinsing procedure several times.
9. Replace the calibration solution in the bottom beaker with a new
solution.
10. Repeat measurement. Move the conductivity sensor slightly until the
displayed value remains constant.
11. Use the arrow keys to enter the determined value.
12. Save the value by pressing “Enter”.
13. Insert or screw the conductivity sensor into the lid of the cell body.
14. Conductivity calibration is then concluded.
WT.050.580.0AA.UA.IM.1206 167
MFC ANALYZER / CONTROLLER
NOTE: Before use, the electrode must be placed in a 100 mg/l fluoride
solution at pH 7 for approx. 24 hours. This is necessary to guarantee
that the electrode functions properly.
1. Starting from the basic display in the main menu, select the “Calibra-
tion” menu.
2. Select the menu item “Fluoride (1) ... (4)“ for the measurement you
wish to perform.
3. Fill 10–20 ml of the lower concentrated standard solution in one of
the supplied beakers and place in the bottom clip. Dip the electrode
at least 2 cm.
4. Select “Cal. lower value” in the menu and press “Enter”.
5. Enter the calibration solution value using the arrow keys and wait
until the measured value on the display remains constant.
6. Save the value by pressing “Enter”.
7. Select “Cal. upper value” in the menu and press “Enter”.
8. Wash electrode with distilled water and fill 10–20 ml of the higher
concentrated standard solution into one of the supplied beakers. Dip
electrode.
9. Enter the calibration solution value using the arrow keys and wait
until the measured value on the display remains constant.
10. Save the value by pressing “Enter”.
11. Place the sensor back in the flow assembly.
WT.050.580.0AA.UA.IM.1206 168
MFC ANALYZER / CONTROLLER
4.5 Errors
Error Messages
The following table shows and explains all possible error messages which
can be displayed. If several errors occur at the same time, the correspond-
ing messages appear alternately in succession. When the error has been
remedied, the error message is automatically deleted.
If you are unable to remedy the error yourself, please contact the follow-
ing service address:
WT.050.580.0AA.UA.IM.1206 169
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 170
MFC ANALYZER / CONTROLLER
Calib Mod1 In 3 electrode cells and Check whether there are air
Calib Mod2 membrane sensors bubbles on the membrane
Calib Mod3 Slope error - the sensor sensor and remove, if nec-
Calib Mod4 current based on 1 mg/l has essary
fallen below the required
minimum Service membrane sensors
In measuring range: - replace electrolyte/mem-
70 µA: in. 0.2 µA/mg/l brane cap
100 µA:min. 0.4 µA/mg/l
200 µA: min. 2 µA/mg/l Clean 3 electrode cells, re-
1000 µA: min. 4 µA/mg/l place cell sand
In pH Check electrodes,
In pH 7 calibration, the Check buffer solutions,
sensor signal is outside Replace, if necessary
-100 to +100 mV or the
sensor sends a signal out-
side 46–70 mV pre pH
increment
The calibration point dis-
tance is smaller than 1 pH
increment
In mV Check electrodes,
The mV electrode correc- Check calibration solution,
tion offset is outside -50 Replace, if necessary
to +50 mV
Conductivity Clean sensor,
Conductivity measurement Inspect,
spread is smaller than 0.8 Replace sensor, if neces-
or larger than 1.2 sary
Fluoride Check electrode, cable and
The rate of change of the standard solution, use fresh
sensor curve is too small or standard solution, replace
the calibration limits have electrodes
been exceeded
0.2 mg/l: 40 to 160 mV
2.0 mg/l: -10 to 100 mV
20 mg/l: -60 to 40mV
The lower cal. value sen-
sor voltage must be 20
mV higher than the upper
cal. value
WT.050.580.0AA.UA.IM.1206 171
MFC ANALYZER / CONTROLLER
mA-OUT 3
mA-OUT 2
mA-OUT 1
W&T00056
WT.050.580.0AA.UA.IM.1206 172
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 173
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 174
MFC ANALYZER / CONTROLLER
Error
The following table shows and explains possible errors which can occur.
If you are unable to remedy the error yourself, please contact the Wallace
& Tiernan® service department.
WT.050.580.0AA.UA.IM.1206 175
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 176
MFC ANALYZER / CONTROLLER
SECTION 5 - service
List of Contents
PARA. NO.
Maintenance Schedules...............................................5.1
Maintaining DEPOLOX® 5 Flow Block Ass’y...........5.2
Maintaining Membrane Sensors.................................5.3
Maintaining Redox Electrode.....................................5.4
Maintaining pH Electrode...........................................5.5
Maintaining Fluoride Electrode..................................5.6
Maintaining Conductivity Electrode...........................5.7
Replacing A Fuse........................................................5.8
Replacing the Battery..................................................5.9
Illustrations
MFC - Schematic Wiring.........................................50.580.155.010A-G
WT.050.580.0AA.UA.IM.1206 177
MFC ANALYZER / CONTROLLER
Task Period/Interval
pH measurement
Comparative pH measurement, calibrate if weekly/
necessary acc. to guidelines
Fluoride measurement
Comparative fluoride measurement, calibrate weekly/
if necessary acc. to guidelines
mV measurement
Check Redox in buffer solution every 4 to 6 weeks
Conductivity
Check conductivity every 4 to 6 weeks
WT.050.580.0AA.UA.IM.1206 178
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 179
MFC ANALYZER / CONTROLLER
NOTE: Screw the upper knurled nut back onto the electrolyte
container.
WT.050.580.0AA.UA.IM.1206 180
MFC ANALYZER / CONTROLLER
NOTE: You must calibrate each time the cell sand is replaced.
The calibration must be checked after one day. We recommend
checking and, if necessary, replacing the electrodes and dia-
phragms when replacing the cell sand (see “Replacing electrolyte,
electrodes and porous membranes”).
Replace Electrolyte
8. Unscrew the upper knurled nut on the electrolyte container.
9. Remove the electrolyte tank out of the cell body from below us-
ing the electrode mount.
10. Remove the electrolyte container from the electrode mount. To
do this, unscrew the knurled nut in the electrode mount.
11. Remove the electrolyte container out the electrode mount from
the top.
12. Remove the drain plug from the electrolyte container.
13. Turn the electrolyte container upside down and drain the electro-
lytes by lightly shaking it.
WT.050.580.0AA.UA.IM.1206 181
MFC ANALYZER / CONTROLLER
NOTE: The cell body’s dowl pin must be locked into place in the
appropriate hole in the electrode mount.
22. Screw the upper knurled nut back onto the electrolyte con-
tainer.
23. Reconnect the signal cable acc. to colour.
WT.050.580.0AA.UA.IM.1206 182
MFC ANALYZER / CONTROLLER
A Knurled nut
B O-ring
�
C Fine filter
D Filter unit
�
�
To do this, screw the M6 screw slightly into the fine filter and pull
the fine filter out of the filter unit
OR
Press the fine filter (C) with a suitable tool (not pointed) out of the
filter unit.
Depending on the quality of the sample water, the Redox electrode must
be cleaned at certain intervals. Unclean water causes the measurement
to be very slow. At the same time, the Redox voltage is frequently indi-
cated as too low. The surface of the platinum disc is usually soiled. This
may be to some extent in the form of an invisible coating. In particular,
the ceramic electrode diaphragm may also be soiled or coated with lime
deposits.
To clean the soiled metal disc, the Redox electrode is removed from the
flow-through adapter and cleaned with a paper towel; use diluted hy-
drochloric acid (up to 10 %), if necessary. Scouring cleaner may not be
used because it may clog the diaphragm. Then rinse well with water. Do
not touch the platinum electrode after rinsing in order to keep oil from
fingers off them. To remove lime deposits, immerse the electrodes just
past the diaphragms into hydrochloric acid (10 %); allow several minutes
for this to react and rinse with clear water.
NOTE: Do not use any other chemicals than those described here.
These could damage the electrode.
5.5 Maintaining pH Electrode
WT.050.580.0AA.UA.IM.1206 184
MFC ANALYZER / CONTROLLER
The electrode must be rinsed subsequently with nothing other than wa-
ter.
Remove lime deposits on the glass membrane and the diaphragm by im-
mersing the electrode into hydrochloric acid (up to 10 %). Rinse thoroughly
with water or distilled water here as well; pH electrodes age. This is often
the cause for a slow display of the pH value or a drop in the slope. pH
electrodes typically last 1 to 2 years. However, routine maintenance of
the electrodes recommended.
5.6 Maintaining Fluoride Electrode
WT.050.580.0AA.UA.IM.1206 185
MFC ANALYZER / CONTROLLER
The electrode does not contain any maintenance parts. However, depending
on how dirty the sample water is, the electrode can be cleaned at certain
intervals. A routine schedule cannot be given here because the cleaning
schedule depends heavily on the general condition of the sample water.
In general, calibrate approx. every 4 weeks.
If wiping with a soft, damp paper towel is insufficient, use one of the
following chemical cleaning methods depending on how the electrode
is soiled:
Time needed at
Soiling Cleaning agent
room temperature
Water soluble sub- Distilled water Any
stances
Grease and oil Warm water and Any
household washing
liquid
Lime and hydroxide Hydrochloric acid Any
deposits (0.1 n)
WT.050.580.0AA.UA.IM.1206 186
MFC ANALYZER / CONTROLLER
A
B
C
D
E
A Metal cover
B Sensor measuring module plug-in card
C plug-in card with mA-out
D Relay card
E Plug-in fuse
WT.050.580.0AA.UA.IM.1206 187
MFC ANALYZER / CONTROLLER
The battery is necessary for the supply of the data storage (e.g. for the
trend graphs) and of the real time clock in the case of a power failure.
Replace the battery if these functions are not working properly.
A
B
C
D
E
A Metal cover
B Sensor measuring module plug-in card
C plug-in card with mA-out
D Relay card
E Battery
WT.050.580.0AA.UA.IM.1206 188
MFC ANALYZER / CONTROLLER
1 2 3 4
Interfaces Power
RS232 RS485 CAN display
A A
GND
GND
GND
VCC
RXD
L2/N L1
TXD
A
H
B
B
Y
Z
L
(M) +24V DC
1314 15 16 17 18 19 20 21 22 23 24 25 26 27 29 30 31 32 28
WH
RD
BU
3 2 1
B B
PE L2/N L1
W&T-RS485-Interface (M) +24V
for CMS,ChemWeb-Server Power
usw. 1/N/PE AC 230/115V
W&T- to the graphik display
RXD
GND
TXD
DC 24V
CAN-bus (internal) prefuse max:
RS232-Printer max. 3m Length 230/115V AC 6A
or 24V DC 4A
PC for update power consumption: 30VA / 30W
C C
Inputs General
(AAC 2764)
ϑ Ym Digital-In mA/V Input
(module 5)
100%
0%
DI 1 DI 2 DI 3
S
1 2 3 4 5 6 7 8 9 10 11 12 1 2 3 4 5 6 7 8
GN
WH
YE
BN
D D
24V+
24V+
GND
max. 50mA
voltage output
+ -
mA Input
0% 100% external setpoint/
control stop feedback signal Dosing factor
control constant + + -
(only by Positioner
mA constant
0-10V
PT1000 with Ym by
single feed forward/
mA
AAC 5461
single feed back sensor measurement module 1)
E Input E
multi-sensor flow rate
Sample water-
Control/ External voltage must
temperature not be connected
to the digital inputs !
1 03.08.05 rb Datum
Bearb.
03.08.05
rb
Design Center GER
Prod. / Sales
Wiring diagram
Datum Date MFC
MFC - SCHEMATIC WIRING
Gepr. release
F Zustand Änderung Datum Name Norm Urspr. Ers. f. F
Ers. d.
Project 50.580.155.010A
=A1
ISSUE 0 10-06
WALLACE & TIERNAN MFC
WT.050.580.0AA.UA.IM.1206 189
order number Drawing number
+S1
Blatt 1
ETSP- AP 00026 7 Bl.
1 2 3 4
MFC ANALYZER / CONTROLLER
1 2 3 4
mA-Outputs
(module 6)
A A
mA 4 mA 3 mA 2 mA 1
+- +- +- +-
1 2 3 4 5 6 7 8
B B
0/4...20mA
Load: 400/1000 Ohm selectable
D D
E E
1 03.08.05 rb Datum
Bearb.
03.08.05
rb
Design Center GER
Prod. / Sales
Wiring diagram
Datum Date MFC
MFC- SCHEMATIC WIRING
Gepr. release
F Zustand Änderung Datum Name Norm Urspr. Ers. f. Ers. d. F
Project 50.580.155.010B
=A1
ISSUE 0 10-06
WALLACE & TIERNAN MFC
WT.050.580.0AA.UA.IM.1206 190
order number Drawing number
+S1
Blatt 2
ETSP- AP 00026 7 Bl.
1 2 3 4
MFC ANALYZER / CONTROLLER
1 2 3 4
WRK
WRK
GND
VCC
CNT
CNT
REF
REF
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8
WH
WH
WH
WH
RD
RD
RD
RD
BU
BU
BU
BU
B B
+
ϑ
-
PT100 AD592
C C
ϑ
I2
I1
1 2 3 4 5 6 7 8 1 2 3 4 5 6 7 8
D D
WH
GY
RD
YE
BU
BN
24V+
24V+
GND
max. 50mA
voltage output
+ + - for Sensor (optional)
0-10V
mA
signal
Input
E E
cable colour
YE = yellow WH = white
BN = brown RD = red
GN = green GY = grey
BU = blue
1 03.08.05 rb Datum
Bearb.
03.08.05
rb
Design Center GER
Prod. / Sales
Wiring diagram
Datum Date MFC
MFC - SCHEMATIC WIRING
Gepr. release
F Zustand Änderung Datum Name Norm Urspr. Ers. f. Ers. d. F
Project 50.580.155.010C
=A1
ISSUE 0 10-06
WALLACE & TIERNAN MFC
WT.050.580.0AA.UA.IM.1206 191
order number Drawing number
+S1
Blatt 3
ETSP- AP 00026 7 Bl.
1 2 3 4
MFC ANALYZER / CONTROLLER
1 2 3 4
Relay-Outputs (application 1)
A A
8 7 6 5 4 3 2 1
24 23 22 21 20 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
B B
C C
D D
E E
1 03.08.05 rb Datum
Bearb.
03.08.05
rb
Design Center GER
Prod. / Sales
Wiring diagram
Datum Date MFC
MFC- SCHEMATIC WIRING
Gepr. release
F Zustand Änderung Datum Name Norm Urspr. Ers. f. Ers. d. F
Project 50.580.155.010D
=A1
ISSUE 0 10-06
WALLACE & TIERNAN MFC
WT.050.580.0AA.UA.IM.1206 192
order number Drawing number
+S1
Blatt 4
ETSP- AP 00026 7 Bl.
1 2 3 4
MFC ANALYZER / CONTROLLER
1 2 3 4
Relay-Outputs (application 2 + 5)
A A
8 7 6 5 4 3 2 1
24 23 22 21 20 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
B B
L1
L2/N
PE
MAX
still be connected
MIN
PE
N
even when the main
switch is set to OFF !
M
max ~1 min
C C
Positioner module 1
2 1 2 1 1
6 5 4 3 2 1 6 5 4 3 2 1 3 2 1
D D
L1 L1
L2/N L2/N
PE PE
L1 N PE L1 N PE
L N PE L N PE
M M
1 ~ 1 ~ Pulse triggered Pulse triggered
Dosing pump Dosing pump dosing pump dosing pump
E module 1 ↓ module 1 ↑ module 1 E
module 1 module 1
1 03.08.05 rb Datum
Bearb.
03.08.05
rb
Design Center GER
Prod. / Sales
Wiring diagram
Datum Date MFC
MFC - SCHEMATIC WIRING
Gepr. release
F Zustand Änderung Datum Name Norm Urspr. Ers. f. Ers. d. F
Project 50.580.155.010E
=A1
ISSUE 0 10-06
WALLACE & TIERNAN MFC
WT.050.580.0AA.UA.IM.1206 193
order number Drawing number
+S1
Blatt 5
ETSP- AP 00026 7 Bl.
1 2 3 4
MFC ANALYZER / CONTROLLER
1 2 3 4
Relay-Outputs (application 3)
8 7 6 5 4 3 2 1
A A
24 23 22 21 20 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
L1
L2/N
PE
MAX
MAX
MIN
MIN
PE
PE
N
N
B B
External voltage may
still be connected
even when the main max
M
~1 min max ~1
M
min
switch is set to OFF !
Positioner Positioner
module 2 module 1
4 3 2 1 3 1
C C
12 11 10 9 8 7 6 5 4 3 2 1 9 8 7 3 2 1
L1
L2/N
PE
L1 N PE L1 N PE L1 N PE L1 N PE
M M M M
1 ~ 1 ~ 1 ~ 1 ~
Dosing pump Dosing pump Dosing pump Dosing pump
module 2 ↓ module 2 ↑ module 1 ↓ module 1 ↑ module 2 module 1
D D
4 3 2 1
12 11 10 9 8 7 6 5 4 3 2 1
L1
L2/N
PE
E E
L N PE L N PE L N PE L N PE
Pulse triggered Pulse triggered Pulse triggered Pulse triggered
dosing pump dosing pump dosing pump dosing pump
1 2 3 4
Relay-Outputs (application 4)
A A
8 7 6 5 4 3 2 1
24 23 22 21 20 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
MAX
MIN
PE
still be connected
N
even when the main
switch is set to OFF ! M
max ~1 min
Positioner module 1
C C
2 1 2 1 1
6 5 4 3 2 1 6 5 4 3 2 1 3 2 1
D D
L1 L1
L2/N L2/N
PE PE
L1 N PE L1 N PE
L N PE L N PE
M M
1 ~ 1 ~ Pulse triggered Pulse triggered
E E
Dosing pump Dosing pump dosing pump dosing pump
module 1 module 1 module 1 module 1 module 1
1 03.08.05 rb Datum
Bearb.
03.08.05
rb
Design Center GER
Prod. / Sales
Wiring diagram
Datum Date MFC
MFC - SCHEMATIC WIRING
Gepr. release
F Zustand Änderung Datum Name Norm Urspr. Ers. f. Ers. d. F
Project 50.580.155.010G
=A1
ISSUE 0 10-06
WALLACE & TIERNAN MFC
WT.050.580.0AA.UA.IM.1206 195
order number Drawing number
+S1
Blatt 7
ETSP- AP 00026 7 Bl.
1 2 3 4
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 196
MFC ANALYZER / CONTROLLER
SECTION 6 - illustrations
List of Contents
PARA. NO.
Parts
DEPOLOX® 5......................................................50.580.000.010A-D
VariaSens™..........................................................50.580.000.020A-C
WT.050.580.0AA.UA.IM.1206 197
8. MFC
WHEN ORDERING MATERIAL ALWAYS SPECIFY MODEL AND SERIAL NUMBER OF APPARATUS
WHEN ORDERING MATERIAL ALWAYS SPECIFY MODEL AND SERIAL NUMBER OF APPARATUS
WHEN ORDERING MATERIAL ALWAYS SPECIFY MODEL AND SERIAL NUMBER OF APPARATUS
WHEN ORDERING MATERIAL ALWAYS SPECIFY MODEL AND SERIAL NUMBER OF APPARATUS
WHEN ORDERING MATERIAL ALWAYS SPECIFY MODEL AND SERIAL NUMBER OF APPARATUS
List of Contents
PARA. NO.
WT.050.580.0AA.UA.IM.1206 205
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 206
MFC ANALYZER / CONTROLLER
Relay
VariaSens™ mA/V 4 mA module Infrared
DEPOLOX® 5
incl. PT100 input card output (8 two-way interface
switches)
AAD4378 AAD4381 AAC5308 AAC5317 AAC5320 AAC9118
plug-in cards plug-in cards
with terminals with terminals
only only
WT.050.580.0AA.UA.IM.1206 207
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 208
MFC ANALYZER / CONTROLLER
Extension pH
Membrane
Cable/ Redox Conductivity Multi
Sensors
Connector DEPOLOX® 5 Fluoride (with 1 Sensor
(with 1
Cable (with 2 connector) with Cable
connector)
Combination connectors)
WT.050.580.0AA.UA.IM.1206 209
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 210
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 211
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 212
8. MFC
MFC ANALYZER / CONTROLLER
7.1.10 5 Retrofit
DEPOLOX®DEPOLOX ® Set AAC6208
5 retrofit set (non-pressurised version) AAC6208
WT.050.580.0AA.UA.IM.1206 213
WT.050.580.0AA.UA.IM.1206 214
8. MFC
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 215
WT.050.580.0AA.UA.IM.1206 216
MFC ANALYZER / CONTROLLER
WT.050.580.0AA.UA.IM.1206 217