Representation of the expected values on the Quo-Lab A1C Control Data Card

LEVEL 1 LEVEL 2
Quo-Lab A1C Control Kit
Unit LOWER LIMIT TARGET UPPER LIMIT
Unit LOWER LIMIT TARGET UPPER LIMIT REF 3101-6021_VS
DCCT (%)

IFCC
DCCT (%)

IFCC
Instructions For Use
(mmol/mol) (mmol/mol)
Read and familiarize yourself with the contents of the
JDS (%) JDS (%)
Instructions for Use before using the controls.
Level 1 = normal Level 2 = abnormal
Consult the Quo-Lab A1C Control Data Card for the expected
values.
Calibration Storage and Stability
Quality Control values are assigned using
equipment calibrated using samples provided by
the European Reference Laboratory, via the NGSP
network. They are traceable to the IFCC reference
Store unopened control kits in a refrigerator at 2
to 8°C (36 to 46°F). The expiry date is printed on
the outside of the pack. The controls cannot be
used after this date.
▲ Please note the modified reconstitution process of the controls.

method. Once controls have been reconstituted they can MEANING

Troubleshooting be used for up to one month, when kept in their


package in a refrigerator at 2 to 8°C (36 to 46°F) . Read the instructions for use

If the result of either control fails then review the In-vitro-Diagnostic


Instructions For Use and repeat the control again. Protect the controls from sunlight. Always store IVD
Medical Device
If the repeat test also fails then contact your local the controls and the reconstituted controls
distributor or call EKF-diagnostic GmbH customer within their original box. LOT Batch or lot number

support on +49 39203 511 414 or Limitations of Use CAL Calibration Barcode
support@ekf-diagnostic.de.
For use with the Quo-Lab Analyzer and Quo-Lab
You must not use the analyzer or report any test REF Product Reference
A1C Test Cartridges only.
results until the problem has been resolved.
CONTROL Control

Storage Temperature

Contents sufficient for <n> Number of Tests


WARNING! POTENTIALLY BIOHAZARDOUS MATERIAL
Human sourced materials were used in the manufacture of this product. Each donor unit was tested for hepatitis
B surface antigen (HbsAg), antibodies to hepatitis C (HCV), and antibodies to Human Immunodeficiency Expiry date
Viruses (HIV-1 and HIV-2), HIV-1 Antigen and Syphilis and found to be negative. CAUTION! Because no
test method can offer complete assurance that HIV, Hepatitis B or C viruses, or other infectious agents are
absent, these products must be handled as potentially infectious. Handle the Hemoglobin A1c Control with Manufacturer
the same precautions used with patient specimens.

▲ Caution

CE mark
EKF-diagnostic GmbH
Ebendorfer Chaussee 3
Biological Hazard
39179 Barleben
Germany
Keep away from sunlight
T+49 39203 511 0
Diagnostics
support@ekf-diagnostic.de
for life www.ekfdiagnostics.com 3101-9202-1115 EN.01.00 06/2019
Intended Use Accessories Required Follow these instructions to run the controls on a The result will remain on screen until the door is
(not included) Quo-Lab Analyzer: opened and the used cartridge removed.
The Quo-Lab A1C Control Kit is intended to
check the general functionality and accuracy of Quo-Lab Analyzer • Place the Level 1 Control, Quo-Lab A1C Test • Now run the Level 2 Control in the same way,
the Quo-Lab A1C Test system. Quo-Lab A1C Test Kit Cartridges, the Sampling Sticks, a small plastic making sure that you scan the Level 2 Control
tray (provided in the control kit) and the Quo- Barcode.
Lab Analyzer on to a clean, dry, level surface.
Description of the Quo-Lab A1C Preparing the Controls • Refer to the Expected Values on the Quo-Lab
The Quo-Lab Analyzer and Test Cartridges
Control Kit Follow these instructions to reconstitute the must be at the same working temperature
A1C Control Data Card.

The kit contains two controls: Level 1 (in the controls prior to use: before use. • R
 eturn the control material to the
normal value range) and Level 2 (in the abnormal refrigerator immediately after use. Always
• Place one Level 1 control bottle and one • Reconstituted controls can be used straight
value range) controls. Both controls, Level 1 and store the controls within their original box.
ampoule on to a clean, dry surface. from the refrigerator and do not need to be at
Level 2 should be used when Quality Control is room temperature before use. • Quality control results are stored in the
being carried out. Use one ampoule containing reconstitution
• 
analyzer memory as control samples. Control
solution for each control bottle. • Again swirl the bottle with the reconstituted
Quality controls should be run: sample data files include the test lot number,
control before use.
• Unscrew the cap of the control bottle. control lot number, whether the control was run
• With each new lot of Test Cartridges.
Open the ampoule by carefully breaking • Place the barcode scanner over the Level 1 as a Level 1 or Level 2 control, the A1C value and
• With each new shipment of Test Cartridges. the plastic cap. Invert the ampoule over the Control Barcode, which is printed on the Quo- if the control passed or failed. Refer to the Quo-
• At any time there is concern that the test control bottle. Fill the complete contents of Lab A1C Control Data Card. Press the barcode Lab Analyzer User Manual to access results held
result may be wrong, the Test Cartridges may the ampoule to the control bottle by gently scanner button until you hear a beep. in the memory.
not have been stored correctly, the user is squeezing the ampoule.
• Prepare and run an A1C test cartridge as per
unfamiliar with the test procedure or there
• When the reconstitution solution has been the Quo-Lab A1C Test Kit Instructions For Use.
is concern that a user may be performing


added, put the cap back on the control bottle.
the test incorrectly. • Whilst the analyzer prepares the reagent,
and dispose the ampoule.
Use the controls according to the quality unscrew the dropper cap, expel any control
• Leave the control bottle to stand for 15 solution in the dropper into the bottle, refit
Warnings and Precautions
standards set by your organization or laboratory.
Refer to the Expected Values on the Quo-Lab minutes. At the end of this time, gently the dropper cap and gently swirl the contents
A1C Control Data Card to check the results swirl the contents to ensure that they are before use. Avoid creating bubbles.
completely reconstituted. Avoid vigorously For in vitro diagnostic use only.
obtained. • Unscrew the dropper cap, gently squeeze the
shaking the bottle as this will create bubbles. Do not use an unopened control kit after the
Only Quo-Lab A1C controls are suitable for use bulb on the control cap, place the tip of the
• Leave the control bottle to stand for a further dropper into the control solution in the bottle expiry date or if the control kit has not been
with the Quo-Lab A1C Test System. Preservatives
15 minutes before use. and gently release the bulb on the control cap. stored in accordance with the recommendations.
and additives used in third party controls may
interfere with the running of the A1C test and Some of the control solution will be drawn into Do not use controls beyond one month once
• Remove the control bottle cap and replace it
EKF-diagnostic GmbH cannot guarantee the the glass dropper. Avoid taking up bubbles. reconstituted or if the reconstituted controls
with the white dropper cap.
results obtained with third party controls. • Place a droplet of the control solution on to have not been stored in accordance with the
• Write the date prepared on the control label. recommendations.
In addition to running the Quo-Lab A1C controls, the small plastic tray by gently squeezing the
The control can be used for up to one month
it is recommended that laboratories regularly bulb. Do not use the control kit if the vials containing
when stored at 2 to 8°C (36 to 46°F).
participate in an external quality control scheme. • Screw the dropper cap back on the control the control material or dropper bottle containing
• Follow the above instructions to prepare the the reconstitutrion solution are damaged.
bottle.
Level 2 control using the black dropper cap.
Use gloves - Used Test Cartridges, sampling
Quo-Lab A1C Control Kit Contents • T
 he controls are now ready. Refrigerate
• When the analyzer prompts the you to
‘Insert Sample’, touch the pointed end of the equipment and controls are potentially infectious.
2 bottles of Level 1 Control with white lids. immediately until required for use. Always
Sampling Stick to the drop of control solution Used Test Cartridges should be disposed of
store the controls within their original box.
2 bottles of Level 2 Control with black lids. on the plastic tray. Avoid taking up bubbles. immediately as clinical waste in accordance with
5 ampoules containing reconstitution solution for Using the Controls Insert the Sampling Stick in the Test Cartridge local / national regulations.
reconstituting the controls. and run the control sample, as defined in the
Refer to the Quo-Lab User Manual for setting Quo-Lab A1C Test Kit Instructions For Use. Protect the controls from sunlight by storing
4 color coded dropper caps (2 white, 2 black). up the analyzer and the Quo-Lab A1C Test Kit within their original box, whether unopened or
16 plastic trays. Instructions For Use for preparing Test Cartridges • When the test is finished, the result will be in reconstituted form.
for use. displayed on the screen. If the result is within
1 Quo-Lab A1C Control Data Card.
the acceptable range, the screen will show
1 Instructions For Use. “Control OK”. If the result is outside the range
then the screen will show “Control Fail”. The
measured A1C value is also displayed.
3101-9202-1115 EN.01.00 06/2019