You are on page 1of 8

CLINICAL SCIENCE

Evaluation of Possible Error Sources in Corneal Endothelial


Morphometry With a Semiautomated Noncontact
Specular Microscope
Michael J. Doughty, PhD

(ECD)2–4 as the number of cells per square millimeter. This


Purpose: To assess the corneal endothelium, particularly the measurement is usually undertaken by using image analysis
polymegethism feature, using the Topcon SP-3000P specular software incorporated into the instruments.
microscope with newer center-dot software. Other morphometric indices have been used to assess the
Methods: Forty-eight healthy, normal weight, noncontact lens endothelial mosaic,5 especially assessment of polymegethism.
wearers of Asian ethnicity were assessed. Single endothelial images This is based on a calculation of the variability in the area of
from each subject were processed with center-dot software, reeval- the individual cells, usually presented as the coefficient of var-
uated after correction of obvious errors, and then by manual border iation (COV) of the cell area.5 Even in the early studies of the
marking and planimetry. Endothelial cell density based on average cell corneal endothelium by specular microscopy, it was generally
area and the coefficient of variation (COV) of cell area (polymegeth- considered that the most normal and healthy endothelium would
ism) were calculated. be composed of uniformly sized cells with a highly symmetrical
shape (the “hexagonal” cell concept).6 Any increase in the var-
Results: Error sources are associated with erroneous location of cell iability in cell size in relation to age, for example, could be
borders (usually creating larger or smaller “cells”) or failure to assign considered to indicate some alterations in the functional state
cell borders to a marked cell. On the initial application of the center- of the cell layer. A low value for the COV is likely accepted
dot software, the endothelial cell density values ranged from 1822 to as the “norm” unless there are confounding factors, such as
3244 cells per square millimeter (mean, 2644 cells/mm2); this range corneal disease (degenerations), or various types of inflamma-
was reduced (eg, 1955–3054 cells/mm2; mean 2690 cells/mm2 on tion associated with surgery or contact lens wear, for example.1,2
editing or in manual planimetry). The COV values ranged from It has been shown that the actual average COV value
17% to 39% (mean, 27.5% 6 5.5%), with one third of the endothelia obtained from an image can be highly dependent on the number
yielding COV values of greater than or equal to 30%. On editing or in of cells used for the calculations; although this, in itself, should
manual planimetry, the COV values were reduced to between 17% not lead to a systematic error that results in predictably higher or
and 29% (mean, 24.5% 6 3.2%; P , 0.001). lower values.7 The repeatability of the COV estimates, however,
Conclusions: In the use of a center-dot endothelial analysis can be expected to get worse if lower numbers of cells are
program with cell border identification, it is likely that at least 1 analyzed.4,7,8 Even with evaluation of a reasonable number of
set of editing steps is required to produce reasonable results. cells per endothelial image and the application of a flexible image
analysis software (which will allow for substantial editing,
Key Words: endothelial cell density, specular microscopy, polyme- including cell–cell border relocation), COV assessments may still
gethism, image analysis software only be within around 65%,8,9 or perhaps even worse than this.10
(Cornea 2013;32:1196–1203) This editing is important because at least some commer-
cially available image software systems are prone to error in
that much larger COV values (from what might be considered
normal) can be readily obtained. This can sometimes be the
T he normal cornea is lined on its posterior aspect by a contin-
uous layer of endothelial cells,1,2 easily viewed at high mag-
nification by either a contact or noncontact specular microscope
result of just a single error in a set of cells.11,12 These analyses
were undertaken using a within-instrument software that does
and confocal microscope. The most widely used measure of the not allow for cell–cell border editing, a system that is likely
corneal endothelium is to assess the endothelial cell density a common option nowadays. With this type of software avail-
able on the Topcon SP-2000P noncontact specular microscope,
it was not possible to easily identify why the COV errors had
Received for publication February 7, 2013; revision received April 4, 2013;
accepted April 5, 2013. occurred11 because the older software only provided the user
From the Department of Vision Sciences, Glasgow Caledonian University, with an output that identified which cells had been marked
Glasgow, United Kingdom. (dotted) for cell area measurements. Reliance was placed on
The author has no funding or conflicts of interest to disclose. the user examining the reported COV and cell minimum (MIN)
Reprints: Michael J. Doughty, Department of Vision Sciences, Glasgow
Caledonian University, Cowcaddens Road, Glasgow G4 0BA, United
and maximum (MAX) area values.
Kingdom (e-mail: m.doughty@gcal.ac.uk). From this perspective, an opportunity was taken
Copyright © 2013 by Lippincott Williams & Wilkins to reevaluate the center-dot software on the updated Topcon

1196 | www.corneajrnl.com Cornea  Volume 32, Number 9, September 2013


Cornea  Volume 32, Number 9, September 2013 Semiautomated Corneal Endothelial Morphometry

SP-3000P, which now provides the user with an overlay of the the central region of the image or just slightly higher or lower
cell–cell borders identified after the cells have been selected with than this (see Results). The instrument software is a “center-dot”
the center-dot procedure. algorithm in which a “dot” is placed in the vicinity of the middle
of a cell by a mouse “click” operation. This procedure was
repeated for a contiguous set of cells within the region of interest
METHODS (ROI) until 100 cells had been marked. The “CALC” menu step
then returns the image of the cells to the screen with an overlay
Subjects to show where the cell borders have been considered to be
After approval by the local ethics committee, subjects present (Fig. 1B). In some cases, a few cells that had been dotted
were recruited from the university-wide student population by did not result in a cell domain (border) being drawn, but the
word of mouth. Any Asian student aged between 18 and 30 location of the “dot” was still marked on the cell (see Results).
years was considered suitable providing that they were in This outcome meant that the “N” for the cell count was less than
general good health and had a negative history for contact 100. In some other images, even though all cells in a group had
lens wear or any surgical or parasurgical procedures on the been “dotted” (as reported by the “N” value before the CALC
eye. All individuals were ambulatory and presented with no command was applied), some borders were not drawn in. This
major health problems (eg, cardiovascular disease, diabetes, results in small locations within the contiguous set of fully tes-
thyroid disorders, obesity). The ethnic grouping of the sellated cells where a single “cell” is now made of 2 or more
subjects was as communicated by them, along with a request actual cells and again the “N” was lower than 100. A variant on
for information on their age and whether they were born in this outcome was that a border of a cell on the periphery of the
the United Kingdom or in their country of ethnic identity. cell set was drawn over onto an adjacent cell that had not been
dotted. In this outcome, the “N” could remain the same, but
a cell border identification was obviously in error (again result-
Specular Microscopy ing in larger and even very large “cells”).
The protocol was to obtain a single image of the central As the primary outcome measure, the results from the first
region of the corneal endothelium using a Topcon SP-3000P presentation of the center-dot algorithm were recorded. This
noncontact specular microscope, which also provides a mea- generated data for ECD (designated as CD on the printout) and
sure of central corneal thickness as a general indicator of the COV (designated as CV on the printout). These 2 primary
corneal health.13 The images were sent to a thermal printer outcomes were therefore termed ECD1 and COV1.
(Sony Video Graphic Printer, UP-897). For most of the images (45 of 48), at least a single error in
cell border identification was obviously found (see Results). As
far as possible, the ANALYZE command was reapplied to the
Image Analysis and Processing marked images (stored temporarily in the instrument software),
As the first step, a rectangular area of the image deemed to efforts were made to correct these errors, and the CALC
include well-resolved cells was selected with the “ANALYZE” command was again used to generate a second result where
command of the menu. In normal corneas and on the use of the the “N” was as close to 100 as could be obtained. This generated
default option on the instrument, this rectangular box size can be the second set of data designated as ECD2 and COV2. This was
expected to contain a little more than 100 cells. This was from managed for all of the 48 images analyzed for this report. The

FIGURE 1. An example of a uniform


endothelium of an Indian man (A),
as found after the application of the
center-dot morphometry algorithm
(B), and with the same region of cells
identified by manual planimetry (C).

 2013 Lippincott Williams & Wilkins www.corneajrnl.com | 1197


Doughty Cornea  Volume 32, Number 9, September 2013

machine software allows for the entire outcome to be undone by For the most part, the images were considered to be the best
“undotting” some or even all the cells and starting again. Even (grade 1) in terms of clarity.
with this being tried, 2 other images (see Results) were rejected
as being unanalyzable.
An ID was attached to the image prints, and a JPEG Outcome Measures from Morphometry
image file was obtained by scanning at 400 dpi. An A3-sized Figure 1A shows a uniform endothelial mosaic of an 18-
print (39 · 26 cm) was made, on which the endothelial cell year-old man of Indian origin. The outcome of the center-dot
borders that were included within and deliberately slightly algorithm applied to the image is shown in Figure 1B and
extended beyond the ROI used in the center-dot analysis were yields an ECD value of 3042 cells per square millimeter,
marked with a pen ( Fig. 1C). The cell border marking on the a COV1 of 17.0%, and a percentage of 6-sided cells (HEX)
prints was deliberately done without looking at center-dot out- of 74%. Essentially, the same outcome was obtained by man-
come. The cells with marked borders were numbered, and then, ual planimetry of the same region of cells (Fig. 1C). It should
their areas were measured with a digitizer pad in stream mode be noted, however, that the cell–cell borders marked by the
(GridMaster DigiPro; Elstree Computing, London). An average center-dot algorithm are not identical to those manually marked
of 156 contiguous cells (range, 137–182) were measured, and for the planimetry, but the agreement appears reasonable.
the areas were converted into absolute values using the scale bar
on the image instrument mask. This generated the third set of
data, designated as ECD3 and COV3. As with the center-dot Small Errors in Border Location and Cells Not
processed images, the ECD was calculated from the average cell Identified With Borders
area value (ie, ECD3 = 1,000,000/AV), whereas the COV was Figure 2A shows high density of cells (3052 cells/mm2)
calculated from the SD of the cell area values (ie, COV = SD/ of a 27-year-old Chinese man, but the reported COV1 was
AV) and multiplied by 100 to give COV3. This approach higher at 26%. Evaluation of the cell borders reveals a very
ignores whether the area measurements from each endothelium unusual-shaped “cell” in the lower left and several unusual
were normally distributed or not, and no assessment of such slightly larger and elongated cells in the upper right. In the
normality was considered simply because a data set of the areas same vicinity are 2 “dots” showing cells that were marked,
is not produced by the center-dot software that was used. but no borders were drawn. The manual cell border marking
(data not shown) revealed the slightly elongated cells to be
errors, and COV3 was just 23%. Figure 2B shows the cells
Statistical Analyses from a Chinese woman aged 26 years. The ECD1 was just
The data were entered into spreadsheets in Systat v. 11 2401 cells per square millimeter, and quite a few cells have
(Systat, Evanston, IL) to allow generation of global statistics rather irregular outlines. The COV1 of the center-dot algo-
and graphical output. The data sets were checked for rithm was again slightly higher at 28%, and the cell overlay
normality using the Shapiro–Wilk default option in Systat, again indicates a number of slightly irregular cells and dotted
and then, comparisons were made where appropriate, either cells without border identification. The manual outcome (data
using paired t tests (assuming unequal variance) or a signed not shown) revealed more regular shaped cells and a COV3 of
rank-order Friedman analysis of variance. just 22.5%. Figure 2C shows the endothelium of similar den-
sity (reported as 2401 cells/mm2) of a 26-year-old Indian man
where the center-dot algorithm has clearly produced not only
RESULTS a slightly larger irregular shaped cell in the middle of the ROI
but also an adjacent smaller cell and a gap. The software
Subject Details and Overall Outcomes From reported a COV1 of 29%, but on manual planimetry (not
Specular Microscopy shown), the COV3 was just 24.5%.
The 48 Asian subjects were aged between 18 and 29 The algorithm error of not identifying cell–cell borders
years (mean 6 SD, 23 6 3 years). Most of the subjects were is sometimes easily correctable and sometimes not. Figure 3A
born in the country identified as their ethnic grouping and had shows a high value for ECD1 (of 3244 cells/mm2) of a 19-
been temporarily resident in the United Kingdom for 1 to 3 year-old Malay man. The main error is in a small group of
years. Overall, the subjects were from Mainland China (12), cells (upper right of ROI) with no borders marked and the
Malaysia (12), India (12), Srilanka (3), Bangladesh (2), Thai- correction was to “undot” these cells and replace them with
land (3), South Korea (2), and Philippines (2). All corneas cells from lower down in the image. With all cells obviously
appeared clear with a quality tear film and had central corneal being of similar size, the COV2 was the same as COV1 (at
thickness values basically within a normal range between 21%) with no notable change in the estimated ECD. Figure
0.478 and 0.585 mm (mean, 0.527 6 0.024 mm). Gross 3B shows a higher ECD of 2919 cells per square millimeter of
external eye evaluations indicated no substantial abnormali- a 21-year-old woman from Bangladesh, but there are 3 obvi-
ties of the conjunctiva or eyelid margins. Although several of ous errors of cell border identification with 3 larger cells
the Indian subjects did have low-grade hyperemia of the bul- (lower left and just to the right of this and also in the upper
bar conjunctiva, they indicated (on specific questioning) no left of the ROI) being produced by 2 “dots” resulting in just 1
major problems in terms of ocular discomfort and so on. All cell. The COV1 was 19% and reduced by 2.0% on the editing
of the endothelia showed an uninterrupted mosaic pattern (ie, COV2 was 17%) to be the same as that from manual
with no signs of inflammation (blebs, pseudoguttae, etc). planimetry. Figure 3C shows slightly more cells that have

1198 | www.corneajrnl.com  2013 Lippincott Williams & Wilkins


Cornea  Volume 32, Number 9, September 2013 Semiautomated Corneal Endothelial Morphometry

FIGURE 2. Examples of the outcome


of application of the center-dot soft-
ware showing small possible errors
that are usually easily correctable. See
text for subject details.

not had their borders marked and with 1 cell (just below the The reported ECD1 was 2916 cells per square millimeter, and
middle of the ROI) with an unusual shape in a 23-year-old the COV1 was 33%. In this example, the correction of the large
Malay man. The ECD1 was 3055 cells per square millimeter, cells (see upper right and lower left of ROI) and replacing the
and the COV1 was 25%. With 2 sequential correction steps, cells without borders (see right-hand side of ROI) changed the
all cells were obtained with borders and the COV2 went down ECD2 to 2994 cells per square millimeter and substantially
slightly to 22% with a small increase in ECD2. reduced COV2 to 24%. Planimetry yielded a COV3 of
22.5%. In Figure 4C, from a 23-year-old Indian man, the
ECD1 was lower at 2566 cells per square millimeter but again
More Substantial Errors in Border Location shows some larger cell domains (now up to 930 mm2) where
and Cells Not Identified With Borders borders were not drawn. The COV1 was 32%, which was
Figure 4A shows that the center-dot algorithm clearly reduced to just 25% on editing. That the center-dot software
produces several unusual very irregularly shaped cells (see can produce some very peculiar cell domains with the border
lower left and upper right of ROI) of an Indian student aged marking can be seen in the upper left of the ROI.
18 years, and in 1 place (see upper region of ROI), what appears
to be a very small cell. This was recognized in the cell MIN
value, and the overall COV1 was 31% despite a reported ECD1 Multiple Substantial Errors Difficult to Correct
of 3021 cells per square millimeter. After what were considered In a few cases, the editing needed to be repeated several
to be acceptable corrections (eg, to remove the very irregular times before a satisfactory outcome was obtained, and in 2
cells), the COV2 was still 29%, whereas, planimetry yielded cases, the outcome of the center-dot routine was considered
a COV3 of just 24.5%. Figure 4B shows that the center-dot uncorrectable. Figure 5A shows the endothelium of a 21-year-
algorithm missed several cells of a 20-year-old Chinese man. old Malay man, with a reported ECD1 of 2611 cells per

FIGURE 3. Examples of the outcome


of application of the center-dot soft-
ware with differences in extent of
missed cells that take time to correct.
See text for subject details.

 2013 Lippincott Williams & Wilkins www.corneajrnl.com | 1199


Doughty Cornea  Volume 32, Number 9, September 2013

FIGURE 4. Examples of the outcome


of application of the center-dot
software with errors in cell border
assignment and missed cells that are
not easily correctable. See text for
subject details.

square millimeter. The COV1 was “normal,” yet inspection of reprocessing of the image, the resultant outcome was considered
the overlay shows 2 regions (upper left and lower left) where unusable despite the numerical outcomes being more reasonable.
there are large spaces including several cells. In this case, the
“dot” has not been recognized, and so these large domains
have not been computed either (with the cell MAX only being
670 mm2). After 3 steps of correction (undotting and redot- Systematic Comparison of Outcome
ting), the resultant COV2 was only slightly altered (at 22%), Measures From Uncorrected and Corrected
but the reported ECD2 increased to 2786 cells per square Center-Dot Algorithm and Comparison to
millimeter. Figure 5B shows the outcome from applying the Manual Planimetry
center-dot algorithm to the endothelial image of a 24-year-old All the images analyzed for the present study were of
woman from Srilanka. The reported ECD1 was 2486 cells per good quality and with no obvious abnormalities or overt signs
square millimeter, and the COV1 was higher at 34%. There of polymegethism. Although most cells appeared to have
are clearly several larger cell domains (with the cell MAX similar sizes, the overall outcome from the above-detailed
being 767 mm2), and again, several dotted cells without bor- examples is that it is not uncommon for differences in COV to
ders. Four steps of correction increased the ECD2 to 2587 arise from a first run of the center-dot algorithm on the SP-
cells per square millimeter and reduced the COV to 27%. 3000P. For the most part, these differences are readily
Planimetry yielded a slightly lower COV3 of 24.5%. correctable, providing of course time is taken to identify them.
Figure 5C shows the endothelium of a 20-year-old Indian These differences are far easier to identify by visual inspection
man, with reported ECD1 of 2786 cells per square millimeter of a printout. The overall results are provided in Table 1.
and the COV1 of 38%. Subjective evaluation of the original The overall outcome for ECD assessments was that the
image revealed no signs of polymegethism. However, an inspec- first run of the center-dot algorithm yielded slightly lower
tion of the overlay reveals multiple irregularly shaped cell values (mean, 2644 cells/mm2) compared with the corrected
domains and some very irregular cell–cell borders. Even after version obtained by 1 or more steps of manual editing (mean,
several attempts to correct the errors and then a complete 2690 cells/mm2). A very similar value was obtained by

FIGURE 5. Examples of the outcome


of application of the center-dot
software showing substantial errors
that are hard to correct. See text for
subject details.

1200 | www.corneajrnl.com  2013 Lippincott Williams & Wilkins


Cornea  Volume 32, Number 9, September 2013 Semiautomated Corneal Endothelial Morphometry

This difference was statistically significant (P , 0.001; paired


TABLE 1. Comparison of Morphometry
t test). The variability on the COV values (as 6SD) was also
Morphological First Operation Corrected substantially higher from the first assessment (ie, 65.5%
Parameter of Center-Dot Center-Dot Planimetry
vs. 63.4%; P , 0.001; Friedman). That the COV values of
ECD (mean 6 SD) 2644 6 266 2690 6 236 2662 6 229 the uncorrected center-dot routine are often higher is supported
ECD range 1822–3244 1955–3054 2022–3027 by the results from the planimetry where the mean COV was
COV (mean 6 SD) 27.5 6 5.5 24.5 6 3.2 24.5 6 3.0 just 24.5%. A Bland–Altman analysis for the differences
COV range 17–39 17–29 18–29 between COV2 and COV1 (Fig. 7) highlights not only this
COV (in percent) is designated as CV on Topcon SP-3000P. ECD (in cells per substantial variability but also a very notable systematic error
square millimeter) is designated as CD on Topcon SP-3000P. in that the higher the initial COV1, the more likely it was to be
in error (regression analysis, not shown in the figure, P ,
0.001; r = 0.710). The SD on the differences was large at
3.3%, yielding a 95% confidence interval for COV
manual morphometry (Table 1). The net difference of just 46 of 29.5% to +3.5%.
cells per square millimeter between ECD1 and ECD2 was
only 1.7% in relative terms but was statistically significant
(P , 0.001; Friedman). A step-wise comparison of ECD2
with ECD1 is shown in Figure 6. This Bland–Altman analysis DISCUSSION
shows what appears to be reasonable agreement for most of The general aim of the present study was to systematically
the images, but there are clearly 3 very substantial outliers. As assess the reliability of the newer center-dot software on the
noted above, although the net difference was only 46 cells per Topcon SP-3000P specular microscope. This revised on-machine
square millimeter, the SD on the difference was 78 cells per software provides the user with a very clear visual output to
square millimeter. Based on the differences, a 95% confi- indicate the fidelity of the cell border/cell domain identification.
dence interval for ECD of 2107 to +199 cells per square This offers the user a substantial advantage over earlier versions
millimeter was obtained. There was a modest trend (regres- of this type of software and is different from older output formats
sion analysis on the difference in relation to the mean, not from the IMAGEnet software that can be used along with this
shown in the figure; P = 0.005, r = 0.395) for the larger specular microscope (as a separate option).8,9,14
differences to be seen at both lower and higher cell densities. This systematic analysis of possible sources of error in
For COV, the difference was more substantial at 3.0% endothelial morphometry broadly confirms previous studies,
(ie, a relative difference of 12.2%), with a mean COV1 of carried out on individuals of white Northern European origin.11,12
27.5% (range from 17% to 39%) compared with just 24.5% The same overall result has been found for other computer-based
(range, 17%–29% for COV2) when corrections were made. programs linked to both specular microscopes8,9,15,16 and

FIGURE 6. Assessment of the net differences in ECD estimates


when comparing the edited output from the center-dot soft- FIGURE 7. Assessment of the net differences in coefficient of
ware (ECD2) with that of the first output (ECD1). The middle variation, COV, estimates when comparing the edited output
line shows the mean difference, whereas the top and bottom from the center-dot software (COV2) with that of the first
horizontal lines show the 95% limits of agreement (61.96 SD). output (COV1). Other details are as in Figure 6.

 2013 Lippincott Williams & Wilkins www.corneajrnl.com | 1201


Doughty Cornea  Volume 32, Number 9, September 2013

confocal microscopes.17,18 The present studies extend previous that could be taken to indicate the presence of some degree of
ones in that, due to the new additions to the instrument-based endothelial polymegethism, that is, a COV of $30% as found
software, some progress can be made toward identifying when in 16 of the 48 subjects in this study in the initial image
these errors occur and then to be able to easily demonstrate the processing with the center-dot software. On editing and with
result of correcting these errors. also a comparison to the results from manual morphometry,
The most substantial impact of the errors is on the none of the endothelia analyzed were then considered to have
COV. As evaluated in detail elsewhere,19 the overall COV a COV above 30%.
cannot identify whether the increased variability in cell area The present study was one of convenience to take
values is due to the presence of smaller or larger cells. How- advantage of foreign national students attending a metropol-
ever, even just 1 extremely small or large “cell” can result in itan university, any of whom could be suitable subjects for
a skewed distribution (either negative or positive) for the cell various cornea or contact lens wear–related studies. The sub-
area values, which then results in an increased COV being jects were considered eligible for the study provided that they
reported for the image analysis outcome. The present assess- met the criteria especially that of not being contact lens wear-
ment, of a software option on a specular microscope for cor- ers. Because most of the subjects evaluated in the present
neal endothelial analyses, further indicates that errors in COV study were born in the country of their ethnic origin, they
are not uncommon. A specific average value for “normal” would have been exposed to the environment and lifestyle of
COV has yet to be defined. There have, however, been their particular Asian region.
numerous publications on younger adults (ie, between 20 In overall conclusion, the newer software option avail-
and 30 years of age) designated as healthy “normal” individ- able for the Topcon SP-3000P noncontact specular micro-
uals and where specular or confocal microscopy has indicated scope provides the user with a very useful on-screen
values of between 20% and 30% for the COV.14,18–29 These indication of the fidelity of the automated cell–cell border
studies cover a wide range of ethnic groupings. Although identification after application of a center-dot endothelial
some of these reports on those of Asian origin indicate analysis program. As illustrated in the present analyses, such
a slightly higher ECD, they do not indicate an unusual an option should greatly assist the user in identifying obvious
COV as based on both subjective assessments of published errors. It is likely, however, that at least 1 set of editing steps
images and values reported for the average COV.14,22,25,28 is required to produce reasonable results.
From this perspective, the reporting of average COV
values rather higher than 30% for normal individuals of
a younger age therefore represents special exceptions or
REFERENCES
contradiction to the accepted “norm.”30–36 The group mean 1. Doughty MJ. Comparative anatomy and physiology of the cornea and
COV from these studies was 42.5% (range, 33.6%–58.3%), conjunctiva. In: Martin Herranz R, Corrales RM, eds. Ocular Surface.
and all of these studies were reported from Asian locations Boca Raton, FL: CRC Press; 2012: Chapter 3:32–78.
(India, Malaysia, Philippines, Taiwan, Thailand, and Turkey). 2. Edelhauser HF. The resiliency of the corneal endothelium to refractive
It needs to be specially noted that none of these reports30–36 and intraocular surgery. Cornea. 2000;19:263–273.
3. Doughty MJ, Müller A, Zaman ML. Assessment of the reliability of
provide images to illustrate the substantial polymegethism human corneal endothelial cell-density estimates using a noncontact
reported from the morphometry. As an overall outcome, it specular microscope. Cornea. 2000;19:148–158.
is recommended that corneal endothelial morphometry stud- 4. McCarey BE, Edelhauser HF, Lynn MJ. Review of corneal endothelial
ies include at least a representative example of the image specular microscopy for FDA clinical trials of refractive procedures,
surgical devices, and new intraocular drugs and solutions. Cornea.
output along with the results of the application of morphom- 2008;27:1–16.
etry software for that particular image. 5. Doughty MJ. Toward a quantitative analysis of corneal endothelial cell
The present studies were undertaken not only to assess morphology: a review of techniques and their application. Optom Vis Sci.
the newer Topcon SP-3000P software but also to use this 1989;66:626–642.
assessment as a platform on which to further assess the 6. Doughty MJ. Concerning the symmetry of the ‘hexagonal’ cells of the
corneal endothelium. Exp Eye Res. 1992;55:145–154.
endothelia of a group of individuals that some have reported 7. Doughty MJ, Fonn D, Trang Nguyen K. Assessment of the reliability of
to show quite remarkable endothelial polymegethism, that is, calculations of the coefficient of variation for normal and polymegethous
a COV of over 40% and which should be overtly obvious on human corneal endothelium. Optom Vis Sci. 1993;70:759–770.
visual inspection of the cell layer images. Of particular 8. Ding X, Huang Q, Zheng Y, et al. Measurement area and repeatability of
semiautomated assessment of corneal endothelium in the Topcon specu-
interest were those from the Indian subcontinent or Eastern lar microscope SP-2000P and IMAGEnet system. Cornea. 2012;31:
Asia. Although the present study is of too small a scale to 1111–1118.
draw definitive conclusions, it does further demonstrate that 9. Doughty MJ, Aakre BM. Further analysis of assessments of the coeffi-
young adult individuals of Asian origin can have what appear cient of variation of corneal endothelial cell areas from specular micro-
to be normal appearing corneal endothelia, that is, a mosaic of scopic images. Clin Exp Optom. 2008;91:438–446.
10. Abib FC, Holzchuh R, Schaefer A, et al. The endothelial sample size
cells with moderate-to-high cell density and having a gener- analysis in corneal specular microscopy clinical examinations. Cornea.
ally uniform cell size with no overt evidence of polymegeth- 2012;31:546–550.
ism. Despite this overall appearance, a semiautomated 11. Doughty MJ. Could the coefficient of variation (COV) of the corneal
software analysis reported rather high COV values in one endothelium be overestimated when a centre-dot method is used? Clin
Exp Optom. 2008;91:103–110.
third of the endothelial images. This outcome is presented in 12. Doughty MJ, Oblak E. A comparison of two methods for estimating
detail to show how it is not difficult to obtain outcome polymegethism in cell areas of the human corneal endothelium. Ophthal-
measures from semiautomated specular microscopy software mic Physiol Opt. 2008;28:47–56.

1202 | www.corneajrnl.com  2013 Lippincott Williams & Wilkins


Cornea  Volume 32, Number 9, September 2013 Semiautomated Corneal Endothelial Morphometry

13. Doughty MJ, Zaman ML. Human corneal thickness and its impact on 25. Matsuda M, Yee RW, Edelhauser HF. Comparison of the corneal endo-
intraocular pressure measures: a review and meta-analysis approach. Surv thelium in an American and a Japanese population. Arch Ophthalmol.
Ophthalmol. 2000;44:367–408. 1985;103:68–70.
14. Cheung SW, Cho P. Endothelial cell analysis using the Topcon SP-1000 26. Tseng SH, Chen FK. Morphological and fluorophotometric analysis of
non-contact specular microscope and IMAGEnet system. Clin Exp Optom. the corneal endothelium after radial keratotomy. Cornea. 1998;17:
1998;81:1–7. 471–475.
15. Siertsema JV, Landesz M, van den Brom H, et al. Automated video 27. Yee RW, Matsuda M, Schultz RO, et al. Changes in the normal corneal
image morphometry of the corneal endothelium. Doc Ophthalmol. endothelial cellular pattern as a function of age. Curr Eye Res. 1985;4:
1993;85:35–44. 671–678.
16. Szalai E, Németh G, Berta A, et al. Evaluation of the corneal endothe- 28. Yunliang S, Yuqiang H, Ying-Peng L, et al. Corneal endothelial cell density
lium using noncontact and contact specular microscopy. Cornea. 2011; and morphology in healthy Chinese eyes. Cornea. 2007;26:130–132.
30:567–570. 29. Nieto-Bona A, González-Mesa A, Nieto-Bona MP, et al. Short-term
17. Jonuscheit S, Doughty MJ, Ramaesh K. In vivo confocal microscopy of effects of overnight orthokeratology on corneal cell morphology and
the corneal endothelium: comparison of three morphometry methods corneal thickness. Cornea. 2011;30:646–654.
after corneal transplantation. Eye (Lond). 2011;25:1130–1137. 30. Chiang CC, Lin JM, Bair PJ, et al. Effects of laser in situ keratomi-
18. Price MO, Fairchild KM, Price FW Jr. Comparison of manual and auto- leusis on the corneal endothelium. Kaohsiung J Med Sci. 2005;21:
mated endothelial cell density analysis in normal eyes and DSEK eyes. 272–276.
Cornea. 2013;32:567–573. 31. Mohammad-Salih PA. Corneal endothelial cell density and morphology
19. Doughty MJ. The ambiguous coefficient of variation: polymegethism of in normal Malay eyes. Med J Malaysia. 2011;66:300–303.
the corneal endothelium and central corneal thickness. Int Contact Lens 32. Padilla MD, Sibayan SA, Gonzales CS. Corneal endothelial cell density
Clin. 1990;17:240–247. and morphology in normal Filipino eyes. Cornea. 2004;23:129–135.
20. Deleted in proof. 33. Rao SK, Ranjan Sen P, Fogla R, et al. Corneal endothelial cell density
21. Carlson KH, Bourne WM, McLaren JW, et al. Variations in human and morphology in normal Indian eyes. Cornea. 2000;19:820–823.
corneal endothelial cell morphology and permeability to fluorescein with 34. Simaroj P, Kosalprapai K, Chuckpaiwong V. Effect of laser in situ
age. Exp Eye Res. 1988;47:27–41. keratomileusis on the corneal endothelium. J Refract Surg. 2003;19
22. Hashemian MN, Moghimi S, Fard MA, et al. Corneal endothelial cell density (suppl 2):S237–S240.
and morphology in normal Iranian eyes. BMC Ophthalmol. 2006;6:9. 35. Sopapornamorn N, Lekskul M, Panichkul S. Corneal endothelial cell
23. Kim KS, Park SY, Oh JS. Morphometric analysis of the corneal endo- density and morphology in Phramongkutklao Hospital. Clin Ophthalmol.
thelial cells in normal Koreans [in Korean]. J Korean Ophthalmol Soc. 2008;2:147–151.
1992;33:24–28. 36. Yeniad B, Yilmaz S, Bilgin LK. Evaluation of the microstructure of
24. Lee JS, Park WS, Lee SH, et al. A comparative study of corneal endo- cornea by in vivo confocal microscopy in contact lens wearing and
thelial changes induced by different durations of soft contact lens wear. non-contact lens wearing keratoconus patients. Cont Lens Anterior
Graefes Arch Clin Exp Ophthalmol. 2001;239:1–4. Eye. 2010;33:167–170.

 2013 Lippincott Williams & Wilkins www.corneajrnl.com | 1203

You might also like