Professional Documents
Culture Documents
Registration of 3D Trans-Esophageal Echocardiography To X-Ray Fluoroscopy Using
Registration of 3D Trans-Esophageal Echocardiography To X-Ray Fluoroscopy Using
a r t i c l e i n f o a b s t r a c t
Article history: Two-dimensional (2D) X-ray imaging is the dominant imaging modality for cardiac interventions. How-
Received 22 June 2010 ever, the use of X-ray fluoroscopy alone is inadequate for the guidance of procedures that require soft-
Received in revised form 20 April 2011 tissue information, for example, the treatment of structural heart disease. The recent availability of
Accepted 4 May 2011
three-dimensional (3D) trans-esophageal echocardiography (TEE) provides cardiologists with real-time
Available online 12 May 2011
3D imaging of cardiac anatomy. Increasingly X-ray imaging is now supported by using intra-procedure
3D TEE imaging. We hypothesize that the real-time co-registration and visualization of 3D TEE and X-
Keywords:
ray fluoroscopy data will provide a powerful guidance tool for cardiologists. In this paper, we propose
Image registration
Cardiac intervention
a novel, robust and efficient method for performing this registration. The major advantage of our method
Trans-esophageal echocardiography is that it does not rely on any additional tracking hardware and therefore can be deployed straightfor-
X-ray fluoroscopy wardly into any interventional laboratory. Our method consists of an image-based TEE probe localization
algorithm and a calibration procedure. While the calibration needs to be done only once, the GPU-accel-
erated registration takes approximately from 2 to 15 s to complete depending on the number of X-ray
images used in the registration and the image resolution. The accuracy of our method was assessed using
a realistic heart phantom. The target registration error (TRE) for the heart phantom was less than 2 mm.
In addition, we assess the accuracy and the clinical feasibility of our method using five patient datasets,
two of which were acquired from cardiac electrophysiology procedures and three from trans-catheter
aortic valve implantation procedures. The registration results showed our technique had mean registra-
tion errors of 1.5–4.2 mm and 95% capture range of 8.7–11.4 mm in terms of TRE.
Ó 2011 Elsevier B.V. All rights reserved.
1361-8415/$ - see front matter Ó 2011 Elsevier B.V. All rights reserved.
doi:10.1016/j.media.2011.05.003
G. Gao et al. / Medical Image Analysis 16 (2012) 38–49 39
can be achieved either by using prior calibration (Rhode et al., diac EP procedures and trans-catheter aortic valve implantations
2003, 2005; Yu et al., 2005), manual registration (De Buck et al., (TAVI), were used to examine the clinical feasibility of the proposed
2005), or automatic approaches (Sra et al., 2007). Recently, the algorithm. Although all the clinical data were currently processed
use of three-dimensional (3D) rotational X-ray angiography (3D off-line, a clinical workflow was proposed to demonstrate how
RXA) (Knecht et al., 2010; Manzke et al., 2010) has been shown our technique will be used live in the clinical environment.
to be useful for cardiac interventional guidance. This imaging
modality is perfectly suited to efficient clinical work flow since
2. Methods
the 3D image data are acquired just prior to the intervention with
the patient lying on the same X-ray table. Therefore, the registra-
2.1. Localization of the TEE probe
tion of the 3D data to the 2D fluoroscopy data is implicit. However,
with the limitation of the current technology the image quality and
Our method is based on the localization of the TEE probe. Since
FOV of 3D RXA are considerably lower than those of MRI and CT.
the addition of tracking devices, such as an EM tracker, is not ideal
The use of pre-procedural MR/CT and 3D RXA imaging will pro-
for the clinical cardiac catheterization environment, the key chal-
duce roadmap images that are static and do not update with the
lenge in our method is to design an efficient and robust image-
intra-procedural situation. Additional steps are required to com-
based TEE probe tracking algorithm. Fig. 1a and b shows the 3D
pensate for intra-procedural deformations, for example, those
TEE probe used in this study and an X-ray image acquired during
caused by respiratory motion of the heart (King et al., 2009). In
an aortic valve implant procedure, clearly showing the TEE probe.
contrast to MR and CT imaging, 3D echocardiography (echo) is a
Our hypothesis is that by using one or more X-ray images, the po-
real-time imaging modality that can be readily used in the catheter
sition and the orientation of the TEE probe can be automatically
laboratory environment. Because of its relatively low image quality
determined from the X-ray data.
and small FOV, 3D echo has been used as an intermediate imaging
modality to register X-ray fluoroscopy with high resolution imag-
ing modalities such as MRI and CT (King et al., 2010). Although pre- 2.1.1. 3D reconstruction of the TEE probe
vious works have successfully registered 3D echo with other Although the TEE probe is designed to be flexible, its transducer
imaging modalities such as MRI (King et al., 2010; Roche et al., is encapsulated in a rigid head. For simplicity, the rigid head of the
2001; Penney et al., 2004; Wein et al., 2009), the registration be- TEE probe is referred to as ‘‘the TEE probe’’ in the following text.
tween echo and X-ray fluoroscopy still relied on prior calibration The first step of our localization algorithm is to reconstruct a pre-
and tracking. With recently emerging technologies such as 3D cise 3D model of the TEE probe. Our approach is to use a nano-CT
trans-esophageal echo (TEE), the image quality and resolution of system (nano-Pet™/CT, Mediso Ltd., Budapest, Hungary) which is
echo has improved considerably. In comparison to trans-thoracic capable of reconstructing ultra-high-resolution 3D volumes.
echo, TEE has unique access to structures such as the aorta, the Fig. 1c shows the nano-CT reconstructed TEE probe (matrix:
valves of the heart and both atria. This makes TEE an ideal solution 342 342, resolution = 0.2 0.2 0.2 mm3). The 3D reconstruc-
for the guidance of cardiac interventions. Recently, the clinical fea- tion of the TEE probe does not need to be repeated unless a differ-
sibility of 3D TEE was evaluated for guiding a variety of cardiac ent model of TEE probe is used.
interventional procedures including atrial septal defect closure,
patent foramen ovale closure, mitral valve/aortic valve repair and 2.1.2. 2D–3D image registration
EP procedures (Barker et al., 2008; Mackensen et al., 2008). TEE ap- In order to determine the position of the TEE probe in C-arm
pears to be a promising imaging modality for cardiac interventions space (i.e. the local 3D coordinate system of the X-ray system),
and its routine clinical use is rapidly increasing. It is already well- the nano-CT volume of the TEE probe was registered to the X-ray
established in the clinical work flow of several interventions, such images using a 2D–3D image registration algorithm. Intensity-
as minimally-invasive valve implantation procedures. based 2D–3D image registration has been extensively studied
Combining the soft-tissue information from TEE with the excel- (Penney et al., 1998; Hipwell et al., 2003; van de Kraats et al.,
lent device visualization of X-ray fluoroscopy could prove to be a 2005; Turgeon et al., 2005). It has been used to register pre-opera-
powerful combination in the catheter laboratory. The purpose of tive CT or MR volume of bones, blood vessels and medical devices
this study was to develop a practical technique to combine 3D to intra-operative X-ray images (Hipwell et al., 2003; Turgeon et
TEE and X-ray fluoroscopy images for the guidance of cardiac cath- al., 2005). Fig. 2 shows a typical workflow of an intensity-based
eterization procedures. A fast and robust image registration of 3D 2D–3D registration algorithm. During the registration process,
TEE and X-ray fluoroscopy was developed for this purpose. Previous the algorithm repeatedly repositions the 3D volume in space and
approaches relied on the use of an electromagnetic (EM) tracking compares its projection, called the digitally reconstructed radiograph
device (Gao et al., 2009; Jain et al., 2009). However, the use of the (DRR), with one or multiple X-ray image(s). At each iteration, an
EM tracker has several disadvantages including the requirement image comparison metric is used to calculate the similarity be-
of additional hardware, the requirement of modifications to the tween the X-ray image(s) and the DRR. The three translation and
ultrasound probe, and the sensitivity of these systems to metallic three rotation parameters are changed according to the similarity
interference. Neither of the studies reported in (Jain et al., 2009; measurement. The registration process continues until the similar-
Gao et al., 2009) involved clinical evaluations but were limited to ity between the X-ray image(s) and the DRR is maximized.
phantoms, therefore, the performance of EM tracking in the routine In this study, we implemented an intensity-based 2D–3D regis-
clinical work flow is unclear. Our method was based upon an im- tration algorithm similar to the one described in (Penney et al.,
age-based 2D–3D registration algorithm. It does not rely on track- 1998). This work shows gradient difference (GD) and pattern
ing hardware or modification to the TEE probe and therefore can intensity (PI) are more robust and accurate than other intensity-
be deployed straightforwardly to any cardiac catheterization labo- based similarity measurements, such as mutual information or
ratory. Preliminary results were reported in (Gao et al., 2010). In cross correlation. Both GD and PI are largely unaffected by the
this paper, we will describe a more thorough study to examine presence of soft tissue and thin line structures, such as guide wires,
the performance of the 2D–3D registration algorithm and the over- stents and other interventional devices that appeared in the X-ray
all TEE and X-ray image registration using both phantom and clin- fluoroscopy. In this study, GD was used as the similarity measure-
ical data. Clinical data (n = 5), which were acquired from two ment function in our 2D–3D registration algorithm. A classic Pow-
different types of cardiac intervention procedures, including car- ell function was used as the optimizer. The speed of the original
40 G. Gao et al. / Medical Image Analysis 16 (2012) 38–49
Fig. 1. (a) The 3D TEE probe used (X7-2t, Philips Healthcare, Andover, Boston, USA); (b) A typical X-ray fluoroscopy image of the TEE probe; (c) A high-resolution 3D nano-CT
scan reveals the contours and the internal structure of the TEE transducer. The 3D nano-CT volume was used to localize the 3D position of the TEE probe in X-ray space.
2.2. Image registration of 3D TEE and X-ray fluoroscopy arm space. This matrix is generated by the 2D–3D registration algo-
rithm that positions the nano-CT volume in C-arm space. TUS_img?
The transformation matrix, TUS_img?X-ray_img, which transforms
tracked relates the position of the 3D TEE images with the position
from 3D TEE image space to 2D X-ray image space consists of a ri- of the TEE probe in nano-CT space. This is the TEE probe calibration
gid body transformation matrix, Trigid, and a projection matrix, Tproj, matrix and can be calculated pre-procedurally using a specifically
T US img!Xray img ¼ T proj T rigid : ð1Þ designed calibration phantom.
The projection matrix, Tproj, transforms from 3D C-arm space, i.e. 2.3. TEE probe calibration
the local coordinate system of the X-ray system, to 2D X-ray image
space. This can be calculated by using the intrinsic parameters of The TEE probe calibration procedure aims to determine
the X-ray system that are determined from a system calibration TUS_img?tracked (Mercier et al., 2005). The calibration phantom
and are available from the DICOM image header (Hawkes et al., consists of a 9-L water tank and two thin metal wires. Nine metal
1987). Trigid can be decomposed into two matrices landmarks which were visible in both X-ray and ultrasound were
T US ¼ T proj T tracked!C-arm T US ð2Þ placed on the wires. The TEE probe was rigidly fixed beneath the
img!Xray img img!tracked
wires during data acquisition. X-ray images were acquired from left
where in our case, Ttracked?C-arm is the transformation matrix from anterior oblique (LAO) 45°, right anterior oblique (RAO) 45° and
the 3D tracked coordinate system, i.e. the nano-CT space, to 3D C- posterio-anterior (PA) projections using a Philips Allura Xper FD10
G. Gao et al. / Medical Image Analysis 16 (2012) 38–49 41
Fig. 3. The proposed workflow of the X-ray and TEE image registration in clinical
2.4. Clinical workflow procedures.
The objective of this study was to develop a practical solution to anthropomorphic external and internal anatomy including left/
integrate X-ray and 3D echo for the guidance of cardiac interven- right ventricles, left/right atria and the valves (Fig. 4).
tional procedures. We propose a clinical workflow to demonstrate For TEE acquisition, we used an iE33 3D real-time echo system
how the proposed TEE and X-ray registration algorithm can be with an X7-2t 3D TEE probe (Philips Healthcare, Andover, Boston,
used in clinical environments. The workflow consists of seven com- USA). The TEE probe was placed on the acoustic surface of the heart
ponents (Fig. 3), some of which are currently manual but could be phantom. Six TEE volumes were acquired to cover different
automated in the future, for example, the initial 2D–3D registra- sections of the phantom by varying the TEE probe position. Nine
tion and the failure detection for the probe localization. The probe X-ray and MRI-visible markers (Multi-modality Radiographic
tracking component, which is designed to compensate the jittering Marker, IZI Medical Products, Baltimore, USA) were placed on the
motion of the TEE probe caused by cardiac and respiratory motion, surface of the phantom. The MR volume of the heart phantom
is not available at present. It is an open research topic. One solution was acquired a day before the TEE and X-ray scans in the catheter
would be to speed up the intensity-based 2D–3D registration by laboratory using a Philips Achieva 1.5T MR scanner. The MR proto-
incorporating more sophisticated implementations such as multi- col consisted of a 3D BTFE scan with the following parameters:
ple-resolution techniques and advanced GPU computing. More de- 512 512 matrix, 180 slices, resolution = 0.96 0.96 1.0 mm3,
tails about the jittering motion of the TEE probe will be discussed TR = 8.42 ms, TE = 4.2 ms, flip angle = 90°. Between the MRI scan
in Section 3.2.3. Technically, the 2D–3D registration algorithm can and the TEE scan, the heart phantom was carefully stored in order
work either with mono-plane or bi-plane X-ray. The workflow does to avoid displacement of the markers. For X-ray image acquisition,
not restrict the number of X-ray images used in the 2D–3D regis-
tration. The decision on whether mono-plane or bi-plane X-ray
images will be used should be made based on the clinical require-
ments and how the registration results will be utilized.
3. Experiments
we used a Philips Allura Xper FD10 C-arm X-ray system, the same The starting positions were created as follows:
X-ray system used for the TEE probe calibration.
(a) For each of the six degrees of freedom, a range was deter-
3.1.2. Cardiac EP procedures mined. Within the ranges, transformation matrices were
We collected data from two cardiac EP procedures. Both of the generated using all possible combinations of the six DOFs.
patients had left atrial flutter and were under general anesthesia The center of the rotation is the center of the nano-CT vol-
(GA) during the procedures. For the first patient, TEE and X-ray ume. New positions of the nano-CT volume were generated
data were acquired after two deca-polar catheters were inserted by transforming the original nano-CT volume using the
into the right atrium (RA), one forming a loop along the endocar- matrices;
dial surface of RA and the other inserted into the coronary sinus (b) For each of the newly generated nano-CT positions, its cor-
(CS). Both catheters were visible in the TEE volume. For the second ner point distance to the position of the original nano-CT
patient, a trans-septal puncture was performed to gain access to volume was calculated as the mean target registration error
the left atrium (LA). TEE volumes were acquired after a lasso cath- (Eq. (4)). The new nano-CT positions were sorted using their
eter and an ablation catheter were inserted into LA. The movement mTRE to the original nano-CT position;
of the C-arm was limited by other equipment such as the life sup- (c) Intervals were chosen for the starting position distance, for
port system and the ultrasound scanner. X-ray images covering 4– example, 0–1 mm, 1–2 mm. For each interval, 100 new
5 cardiac cycles were acquired from PA and either RAO 30° or LAO nano-CT positions were selected randomly (uniformly dis-
30° projections. Cardiac and respiratory phases were estimated tributed) as the starting positions.
using the motion of the interventional devices such as the cathe-
ters (Brost et al., 2010; Klemm et al., 2007). X-ray images acquired For the calculation of mTRE, the landmarks were the eight cor-
during the same cardiac and respiratory phases were used in the ners of the bounding box of the nano-CT volume. The testing data
2D–3D registration. used in this experiment were X-ray images acquired from the clin-
ical studies. For each of the seven testing datasets, 1400 starting
3.1.3. Trans-catheter aortic valve implant procedures positions (1 mm 6 mTRE 6 15 mm) were examined to calculate
TAVI data were collected from three patients with degenerative the capture range. The capture ranges using bi-plane X-ray images
aortic stenosis. All the three procedures were performed in a GE and mono-plane X-ray images were studied separately
catheter laboratory equipped with a GE Innova 2100IQ C-arm X- n
1X i i
ray system. Similar to the Philips Allura Xper FD10, this GE X-ray mTRE ¼ T registration Plandmark T ref Plandmark ð4Þ
n i¼1
system can precisely track its C-arm position automatically. The
TEE data were acquired using the same echo system and TEE probe
where P denotes the landmarks defined on the eight corners of the
described above. In the first two TAVI cases, the replacement
nano-CT volume. To evaluate the registration accuracy using mono-
valves were delivered using trans-femoral approach. In the third
plane X-ray images, the mean projection distance, mPD was calcu-
TAVI case, the replacement valve was delivered using trans-apical
lated (van de Kraats et al., 2005)
approach. X-ray images and TEE volumes were acquired in the
n
same way as the data acquisition in the EP procedures after the 1X i i
replacement valve reached the deployment site. mPD ¼ Mregistration Plandmark M ref Plandmark Sdet F ð5Þ
n i¼1
3.2. Evaluation of the nano-CT and X-ray image registration where Mregistration and Mref are the final perspective projection
matrices for the registration and the reference. Sdet is the detector
By using our method, a successful TEE and X-ray image registra- size of the X-ray system. F is the magnification factor, which can
tion depends on a fast and robust 2D–3D nano-CT and X-ray image be calculated using Eq. (6)
registration algorithm. In this experiment, we evaluate the robust-
Dsource!object
ness and the speed of the 2D–3D registration algorithm using stan- F¼ ð6Þ
Dsource!image
dard methodologies presented in previously published literature
(van de Kraats et al. 2005). Dsource?object and Dsource?image are source to object distance and
source to image distance. Sdet, Dsource?object and Dsource?image can be
3.2.1. Experiment 1 – Capture range and speed of the 2D–3D retrieved from either the live data stream from the X-ray
registration system or the header of the DICOM files. It should be noted that
The capture range was defined as the largest starting misalign- Dsource?object is an approximated value and therefore the mean
ment at and below which registration was successful 95% of the time projection distance calculated using Eq. (5) is also an approximated
(Huang et al., 2009; Shekhar and Zagrodsky, 2002). To evaluate the value.mTRE has been previously proposed and accepted as a metric
capture range, a reference transformation matrix was introduced to evaluate the capture range of 2D–3D registration in different
for each of the testing datasets. The reference transformation matrix studies (Huang et al., 2009; van de Kraats et al., 2005). However,
was prepared by the following procedures: it cannot clearly differentiate the errors in translation from the
errors in rotation. An additional step was used to examine the reg-
(a) the automatic 2D–3D registration, as described in Section istration error in translations (mm) and rotations (°) separately by
2.1, was employed to generate a starting position to align using the magnitude of the rotations and the translations
the nano-CT volume with bi-plane X-ray images; qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
(b) the registration was then refined using manual expert inter- t¼ t 2x þ t 2y þ t2z and h ¼ h2x þ h2y þ h2z ð7Þ
action allowing all six degrees of freedom during careful
visual examination of the result to ensure the closest possi- where t and h are the magnitude of the translations and the rota-
ble match; tions, tx, ty, tz and hx, hy, hz are the translations and rotations in x,
(c) steps (a) and (b) were repeated three times resulting in three y and z directions.
transformation matrices. The reference matrix was con- The computation time of every registration was recorded to
structed using the mean parameters of these three matrices. analyze the efficiency of our 2D–3D registration algorithm.
G. Gao et al. / Medical Image Analysis 16 (2012) 38–49 43
Fig. 8. (a) For each of the datasets, 1400 starting positions with mTRE ranging evenly from 1 mm to 15 mm were used to examine the capture range of the intensity-based
2D–3D image registration algorithm. (b) For all the successful registrations, landmarks were defined within the TEE FOV at different depths to examine the potential target
error caused by the error of the 2D–3D image registration. Presuming that the 2D–3D image registration was the sole error source, the red curve shows the success rate of the
TEE and X-ray image registration for the targets defined within the TEE FOV at different depths.
G. Gao et al. / Medical Image Analysis 16 (2012) 38–49 45
Table 2
The errors of all the successful registration in terms of rotation and translation.
Translation Rotation
tx (mm) ty (mm) tz (mm) rx (°) ry (°) rz (°)
Mean err. 0.06 0.16 0.04 1.16 0.17 1.05
Std. dev. 0.03 0.06 0.03 0.23 0.11 0.58
Max. err. 0.47 1.07 1.06 3.13 1.47 2.46
Fig. 10. (a) Comparison of 3D errors in term of mTRE for the 2D–3D registration
Fig. 9. The starting positions of all the failed 2D–3D registrations in terms of
using bi-plane X-ray and mono-plane X-ray. The testing data were acquired from
translation and rotation. Only 4.2% of the failed registrations started from within a
the first of the TAVI procedures reported in this paper. Potentially, the 2D–3D
region of 10 mm 10°.
registration using one X-ray image could produce considerable error in the X-ray
projection direction, reducing the overall success rate. (b) The majority of the error
is in the direction of the X-ray projection, the in-plane error mPD.is 1.8 ± 1.3 mm
the misalignment to within a 10 mm 10° region to be subse- compared to the overall 3D error of 6.8 ± 7.8 mm.
quently successful for automatic registration.
Technically, it is possible to register the nano-CT volume with
mono-plane X-ray images. The capture range of the image registra-
jittering motion which clearly reveal the pattern of cardiac and
tion using one X-ray image is similar to those using bi-plane X-ray
respiratory motion. For both cases, the jittering motion introduces
images. However, it could introduce a considerable error in term of
a maximum error of around 4 mm in the probe localization
mTRE to the target defined within the FOV of TEE (Fig. 10a). The
(4.1 mm and 3.6 mm, respectively). The maximum errors contrib-
majority of the error occurs along the direction of the X-ray beam
uted by the rotational components of the jittering motion are
and does not impact considerably if the echo data is used as an
1.6 mm and 1.0 mm, respectively. Although our experiment shows
overlay image on top of the X-ray fluoroscopy as demonstrated la-
the majority of the jittering motion is translational, the error can
ter in Figs. 13 and 14. Fig. 10b shows that the average in-plane reg-
still make a considerable impact on the accuracy of the overall
istration error in terms of mPD is 1.8 ± 1.3 mm. We do not restrict
TEE and X-ray registration especially if the target object locates
the number of X-ray images used in our clinical workflow. The
at the far end of the echo FOV. In the proposed clinical workflow,
decision will be made based on the requirement of the interven-
a real-time TEE probe tracking algorithm will minimize the error
tional procedures. More details will be discussed in Section 5.
caused by the jittering motion. Before such an algorithm becomes
available, physicians will need to accommodate the registration er-
4.1.2. Registration speed ror using the motion pattern of the interventional devices observed
The computational time for the 2D–3D image registration de- in the X-ray fluoroscopy and in the echo volume.
pends on several factors including the starting position of the
nano-CT volume, the size of the X-ray images and the number of
the X-ray images used in the registration. From over 30,000 regis-
4.2. TEE probe calibration
trations performed using different starting positions, the computa-
tional time of our 2D–3D image registration was recorded and is
The residual calibration error from Eq. (3) was 2.0 mm. By using
summarized in Table 3.
the calibration matrix, the last two volumes that were not involved
in the calibration were transformed to the C-arm space as where
4.1.3. Jittering motion error the landmarks from the X-ray images. The mTREs computed in
X-ray images acquired from a TAVI procedure (Fig. 11a) and an C-arm space were 4.6 ± 1.1 mm and 5.0 ± 0.8 mm. The determina-
EP procedure (Fig. 11b) were used to assess the error caused by the tion of the landmark positions from the TEE volume was difficult
jittering motion of the TEE probe. Fig. 11 shows the curves of the due to the noise and the shadowing effects of ultrasound. The
46 G. Gao et al. / Medical Image Analysis 16 (2012) 38–49
Fig. 11. The TEE probe localization error caused by the jittering motion. X-ray images were acquired during (a) a TAVI procedure and (b) a cardiac EP procedure.
mTREs show not only the registration error but also the subjective stage, the data acquisition of TEE and X-ray data was not synchro-
error in the determination of the landmark positions. nized. The end systolic phase was manually selected from the
X-ray and TEE sequences to calculate the mPD shown in Table 5.
However, the catheter motion (caused by the cardiac motion
4.3. Phantom study
including cardiac contraction and respiratory motion) shown in
the X-ray images was more considerable than that shown in the
The 3D positions of the landmarks were reconstructed in C-arm
TEE volume. This is because the catheter motion shown in TEE
space using two X-ray images acquired from LAP 45° and RAO 45°.
was partly canceled by the motion of the TEE probe itself. At pres-
The mean reconstruction error in term of mPD was much less than
ent, our 2D–3D registration algorithm is not fast enough to com-
1 mm (0.44 mm using the image acquired from LAO 45° and
pensate for the TEE probe motion in real-time. This error will be
0.14 mm using image from RAO 45°). The mean landmark registra-
additive in the registration results.
tion error for the MR to C-arm registration was 1.03 mm. The
Fig. 14 shows the X-ray images and a TEE volume acquired in one
point-to-surface distance errors between TEE and MRI data are
of the TAVI procedures. The diameter of the TAVI catheter (18-22F or
summarized in Table 4. The mean TEE to C-arm registration error
6–7 mm) is considerably larger than the EP catheters (8F, 3 mm).
was in a range 1.5–2.4 mm. Fig. 12 shows the MRI-derived LV
Although both the TAVI catheters and the valves can be easily iden-
and the RV surfaces registered with two TEE volumes in C-arm
tified from the TEE volume, it is difficult to determine the exact loca-
space.
tion of the replacement valve in the TEE volume. Fig. 14a shows the
position of the replacement valve was clearly visible in the X-ray
4.4. Clinical studies images. The combined image of X-ray and TEE reveals that the
replacement valve had passed through the aortic valve into the left
Fig. 13 shows the registration results when different catheters ventricle. Without seeing the overlay image, the physician decided
were used for two patients undergoing EP procedures. At this to deploy the replacement valve in this sub-optimal position. Post-
deployment Doppler revealed severe paravalvular leakage and poor
hemodynamic performance. The motion of the mitral valve was also
obstructed with the mis-placed aortic valve. A second replacement
valve was inserted to stop the leakage. This case highlights the po-
tential clinical impact of our technique.
The registration accuracy for all the clinical cases (two EP and
three TAVI) is summarized in Table 5.
5. Discussion
Table 3
The average computation time of the 2D–3D registration.
Fig. 14. Co-registered echo and X-ray images for one of the TAVI patients. (a) The X-ray image shows the interventional devices including the catheter, the guide wire and the
replacement valve; (b) the 3D TEE reveals cardiac anatomy including the aorta and the valves as well as the catheter. However, it was difficult to identify the exact location of
the replacement valve; (c) the highlighted area in the combined image of TEE and X-ray shows the replacement valve had passed through the aortic valve into the left
ventricle.
48 G. Gao et al. / Medical Image Analysis 16 (2012) 38–49
Table 4 tem, although neither of the two studies were performed in the clin-
The point-to-surface error represents the error in the registration of TEE and MRI in C- ical environment. At present, there is no commercial EM tracker-
arm space. For LV and RV, the points were defined on all the six TEE volumes, 10 for
each volume. For LA, RA and the valves, points were defined on the volumes where
embedded TEE probe available. To use an EM tracking system with
the borders were clearly visible. The valves were not segmented from the MR volume. the TEE probe, the TEE probe must be modified so that an EM sensor
The point-to-surface distance for the valves was the distance of a point defined on the can be fitted within it. Compared to the use of an EM tracker, the first
valves to the top surface of the LV or the RV. advantage of our method is that it does not use any additional track-
Num. of points Mean error (mm) SD (mm) Max. error (mm) ing device. The accuracy of the EM tracking system relies on a static
LV 66 2.0 1.5 4.7
metal environment. With the presence of the moving X-ray C-arm
RV 60 1.9 1.2 4.9 and other medical devices including the TEE probe itself, the reli-
Valves 16 1.9 1.3 2.7 ability of the EM tracking system is questionable (Hastenteufel et
LA 22 1.5 0.8 2.8 al., 2006). Additionally, the methodology presented for the co-regis-
RA 16 2.0 1.4 4.4
tration of 3D TEE and X-ray fluoroscopy data is better suited for the
LVPM 20 2.4 1.8 3.2
Overall 200 1.9 1.2 4.9 routine clinical work flow.
Table 5
Registration error results from the clinical cases. Devices were projected from the TEE
The 2D–3D registration algorithm is not sensitive to detecting
volume data to the X-ray image data using the computed 2D–3D registration. The misalignment along the direction of the X-ray beam so that the
mPD was computed for each device. 2D–3D registration algorithm using mono-plane images often
Patient ID Procedure mPD (mm) SD (mm)
produces considerable error in the direction of projection.
Introducing one more X-ray image acquired from a different angle
1 EP 3.1 2.6
effectively solves this problem, allowing an accurate registration in
2 EP 1.5 1.6
2 EP 3.2 2.5 all six DOFs although the radiation dose will be double (but very
3 TAVI 2.7 1.8 small when compared to entire procedure dose). This can be
4 TAVI 4.2 3.0 performed using bi-plane X-ray systems but these are not com-
5 TAVI 2.3 2.1
monly available. A more conventional method is to use sequential
bi-plane images from a mono-plane X-ray system. However, with
sequential bi-plane acquisition, correct phase-matching (for both
devices are relatively straight in the FOV of the TEE images and respiratory phase and cardiac phase) of the X-ray data must be
therefore the tangential error along the interventional devices can- carried out.
not be captured. However, the validity of mPD as an accuracy met- Clinically, the 2D–3D registration with bi-plane X-ray is not al-
ric was supported in the unsuccessful TAVI case, where the clinical ways necessary. The clinical workflow includes two different ways
outcome matches the finding of the TEE and X-ray registration. The to visualize and utilize the registration results, Echo ? X-ray pro-
accuracy of the proposed registration method is also confirmed in jection and X-ray ? Echo projection. Echo ? X-ray projection
one of the EP cases where a circular catheter was used as the refer- overlays the echo volume on top of the X-ray images producing a
ence for mPD calculation. ‘roadmap’ to navigate the clinical procedures such as atrial fibrilla-
At this stage, all the clinical data we used were processed off- tion (AF) ablation and TAVI. In these cases, the 2D–3D registration
line. The X-ray images and the TEE volumes were not acquired at using mono-plane X-ray is sufficient as the accuracy in the direc-
the same time. Although we attempted to minimize the registra- tion of projection is not a major concern. However, in cases which
tion error by using the images acquired in the same cardiac and the 3D position of the interventional devices must be precisely
respiratory phases, the error caused in the cardiac phase estima- localized, bi-plane X-ray images will be needed to ensure a precise
tion will be propagated to the end registration result. Such error TEE and X-ray registration in all six degrees of freedom. For exam-
is likely to be further reduced in the future when real-time X-ray ple, in some AF ablation cases, the locations of the ablation sites
and ultrasound data can be acquired simultaneously. need to be recorded for reviewing purposes. An example for the
The jittering motion of the TEE probe during a clinical procedure application of X-ray ? Echo projection could be: (1) determine
is another source of error which potentially could affect the accu- the location of the ablation catheter tip from the X-ray image;
racy of the TEE and X-ray registration. Experimental results show (2) back-project the 2D location of the catheter tip to C-arm space,
the error introduced by the jittering motion can be as much as a producing a line intersecting the registered TEE volume; and (3)
4 mm error in the probe localization. Fig. 8a suggests this error determine the 3D position of the ablation catheter tip either man-
can be easily removed using the 2D–3D registration. The success ually or automatically from the TEE volume.
rate of the 2D–3D image registration is 100% given the starting po-
sition of the source volume is within 5 mm capture range. However, 5.4. Limitations and future work
the speed of the current 2D–3D registration algorithm needs to be
improved considerably to satisfy the requirement of real-time One of the main objectives of this study was to assess the clinical
tracking. feasibility of the proposed method. Although all the clinical data
were acquired specifically for this study, the data processing was
5.2. Comparison with EM tracking systems done off-line. However, the transition to real-time or near-real-
time capability is relatively straightforward in this case, with all
At present, our 2D–3D registration algorithm takes around 5 s to the required components already present, including live data
determine the 3D position of the TEE probe using two 512 512 X- streaming from both X-ray and ultrasound systems. Our future
ray images. EM tracking systems determine the location of objects work will focus on the implementation of the live functionality
that are embedded with the EM sensors in real-time. More impor- and also improvements in the visualization of the co-registered
tantly, it is not required that the EM sensor be in the line-of-sight data. The 3D TEE images were acquired in full volume mode. Our
of the magnetic field generator, as for optical tracking systems. Pre- TEE probe was calibrated using full volume echo images. The clini-
vious studies (Gao et al., 2009; Jain et al., 2009) confirmed that it cal procedures are usually guided by live 3D echo images whose
was possible to localize the TEE probe by using an EM tracking sys- FOV is about a quarter of the full volume images. Therefore,
G. Gao et al. / Medical Image Analysis 16 (2012) 38–49 49
co-registration with a global roadmap from either CT or MR image Clementy, Haissaguerre, M., Orlov, M., Jais, P., 2010. Prospective randomized
comparison between the conventional electroanatomical system and three-
data could assist in clarifying the anatomical context of the small
dimensional rotational angiography during catheter ablation for atrial
FOV echo data. fibrillation. Heart Rhythm 7, 459–465.
Kovoor, P., Ricciardello, M., Collins, L., Uther, J., Ross, D., 1998. Risk to patients from
radiation associated with radiofrequency ablation for supraventricular
6. Conclusions tachycardia. Circulation 98, 1534–1540.
Linte, C.A., Moore, J., Peters, T.M., 2010. How Accurate is Accurate Enough? A Brief
In this study, we have described a novel TEE to X-ray fluoros- Overview on Accuracy Considerations in Image-Guided Cardiac Interventions.
In: Proceedings of 32nd Annual International Conference of the IEEE EMBS, pp.
copy registration technique. The method was successfully evalu- 2313–2316.
ated using phantom data and five clinical datasets acquired from Mackensen, G.B., Hegland, D., Rivera, D., Adams, D.B., Bahnson, T.D., 2008. Real-time
two different types of cardiac interventional procedures, including 3-dimensional transesophageal echocardiography during left atrial radio
frequency catheter ablation for atrial fibrillation. Circulation: Cardiovascular
cardiac EP procedures and TAVI procedures. The experimental re-
Imaging 1, 85–86.
sults show that our method is fast, robust and accurate. It is likely Manzke, R., Meyer, C., Ecabert, O., Peters, J., Noordhoek, N., Thiagalingam, A., Reddy,
that such a co-registration and visualization technology is going to V., Chan, C.C., Weese, J., 2010. Automatic segmentation of rotational X-ray
images for anatomic intra-procedural surface generation in atrial fibrillation
have a significant impact in the field of image-guided cardiac
ablation procedures. IEEE Transactions on Medical Imaging 29, 260–272.
interventions. Mercier, L., Langø, T., Lindseth, F., Collins, D.L., 2005. A review of calibration
techniques for freehand 3-D ultrasound systems. Ultrasound in Medicine &
References Biology 31, 449–471.
Modan, B., Keinan, L., Blumsterin, T., Sadetzki, S., 2000. Cancer following cardiac
catheterization in childhood. International Journal of Epidemiology 29, 424–
Barker, G.H., Shirali, G., Ringewald, J.M., Hsia, T.Y., Bandlsode, V., 2008. Usefulness of 428.
live three-dimensional transesophageal echocardiography in a congenital heart Penney, G., Blackalll, J.M., Hamady, M.S., Sabharwal, T., Adam, A., Hawkes, D.J., 2004.
disease center. The American Journal of Cardiology 103, 1025–1028. Registration of freehand 3D ultrasound and magnetic resonance liver images.
Brost, A., Liao, R., Hornegger, J., and Strobel, N., 2010. 3-D Respiratory Motion Medical Image Analysis 8, 81–91.
Compensation during EP Procedures by Image-Based 3-D Lasso Catheter Model Penney, G., Weese, J., Little, J.A., Desmedt, P., Hill, D.L., Hawkes, D.J., 1998. A
Generation and Tracking. Medical Image Computing and Computer Assisted comparison of similarity measures for use in 2-D–3-D medical image
Intervention (MICCAI’09) Medical Image Computing and Computer Assisted registration. IEEE Transactions on Medical Imaging 17, 586–595.
Intervention (MICCAI’09), pp. 394–401. Razavi, R., Hill, D.L., Keevil, S.F., Miquel, M.E., Muthurangu, V., Hegde, S., Rhode, K.S.,
De Buck, S., Meas, F., Ector, J., Bogaert, J., Dymarkowski, S., Heidbuchel, H., Suetens, Barnett, M., van Vaals, J., Hawkes, D.J., Barker, E., 2003. Cardiac catheterization
P., 2005. An augmented reality system for patient-specific guidance of cardiac guided by MRI in children and adults with congenital heart disease. The Lancet
ablation procedures. IEEE Transactions on Medical Imaging 24, 1512–1524. 362, 1877–1882.
Gao, G., Penney, G., Gogin, N., Cathier, P., Arujuna, A., Wright, M., Caulfield, D., Rhode, K.S., Hill, D.L., Edwards, P.J., Hipwell, J.H., Rueckert, D., Sanchez-Ortiz, G.,
Rinaldi, A., Razavi R., Rhode, K.S., 2010. Rapid Image Registration of Three- Hegde, S., Rahunathan, V., Razavi, R., 2003. Registration and tracking to
dimensional Transesophageal Echocardiography and X-ray Fluoroscopy for the integrate X-ray and MR images in an XMR facility. IEEE Transaction on
Guidance of Cardiac Interventions. Information Processing in Computer- Medical Imaging 22, 1369–1378.
Assisted Intervention (IPCAI’10) Information Processing in Computer-Assisted Rhode, K.S., Sermesant, M., Brogan, D., Hegde, S., Hipwell, J.H., Lambiase, P.,
Intervention (IPCAI’10). Rosenthal, E., Bucknall, C., Qureshi, S.A., Gill, J.S., Razavi, R., Hill, D.L., 2005. A
Gao, G., Reddy, K., Ma, Y., Rhode, K.S., 2009. Real-time Compounding of Three- system for real-time XMR guided cardiovascular intervention. IEEE Transaction
Dimensional Transesophageal Echocardiographic Volumes: The Phantom Study. on Medical Imaging 24, 1428–1440.
IEEE Engineering in Medicine and Biology Society (EMBC’09) IEEE Engineering Roche, A., Pennec, X., Malandain, G., Ayache, N., 2001. Rigid registration of freehand
in Medicine and Biology Society (EMBC’09), pp. 499–502. 3D ultrasound with MR images: a new approach combing intensity and
Hastenteufel, M., Vetter, M., Meinzer, H.P., Wolf, I., 2006. Effect of 3D ultrasound gradient information. IEEE Transactions on Medical Imaging 20, 1038–1049.
probes on the accuracy of electromagnetic tracking systems. Ultrasound in Shekhar, R., Zagrodsky, V., 2002. Mutual information-based rigid and non-rigid
Medicine and Biology 32, 1359–1368. registration of ultrasound volumes. IEEE Transactions on Medical Imaging 21,
Hawkes, D.J., Mol, C.B., Colchester, A.C.F., 1987. The accurate 3-D reconstruction of 9–22.
the geometric configuration of vascular trees from X-ray recordings. Physics Silverstry, F.E., Kerber, R.E., Brook, M.M., Caroll, J.D., Eberman, K.M., Goldstein, S.A.,
and Engineering of Medical Imaging 250, 258. Herrmann, H.C., Homma, S., Mehran, R., Packer, D.L., Parisi, A.F., Pulerwitz, T.,
Hipwell, J.H., Penney, G., Mclaughlin, R.A., Rhode, K.S., Summers, P., Cox, T.C., Byrne, Seward, J.B., Tsang, T.S., Wood, M.A., 2009. Echocardiography-guided
J.V., Noble, J.A., Hawkes, D.J., 2003. Intensity based 2D–3D registration of interventions. Journal of American Society of Echocardiography 22, 213–231.
cerebral angiograms. IEEE Transactions on Medical Imaging 22, 1417–1426. Sra, J., Narayan, G., Krum, D., Malloy, A., Cooley, R., Bhatia, A., Dhala, A., Blanck, Z.,
Huang, X.S., Ren, J., Guiraudon, G., Boughner, D., Peters, T., 2009. Rapid dynamic Nangia, V., Akhtar, M., 2007. Computed tomography–fluoroscopy image
image registration of the beating heart for diagnosis and surgical navigator. IEEE integration-guided catheter ablation of atrial fibrillation. Electrophysiology
Transactions on Medical Imaging 28, 1802–1814. 18, 409–414.
Jain, A., Gutierrez, L., Stanton, D., 2009. 3D TEE Registration with X-ray Fluoroscopy Turgeon, G.-A., Lehman, G., Guiraudon, G., Drangova, M., Holdsworth, D., Peters, T.,
for Interventional Cardiac Applications. Functional Imaging and Modeling of the 2005. 2D–3D registration of coronary angiograms for cardiac procedure
Heart (FIMH’09) Functional Imaging and Modeling of the Heart (FIMH’09), pp. planning and guidance. Medical Physics 32, 3737–3749.
321–329. van de Kraats, E., Penney, G., Tomazevic, D., van Walsum, T., Niessen, W., 2005.
King, A.P., Boubertakh, R., Rhode, K.S., Ma, Y., Chinchapatnam, P., Gao, G., Tangcharoen, Standardized evaluation methodology for 2-D–3-D registration. IEEE
T., Ginks, M., Cooklin, M., Gill, J.S., Hawkes, D.J., Razavi, R., Schaeffter, T., 2009. A Transactions on Medical Imaging 24, 1177–1189.
subject-specific technique for respiratory motion correction in image-guided Wein, W., Carnus, E., John, M., Diallo, M., Duong, C., Al-Ahmad, A., Fahrig, R.,
cardiac catheterization procedures. Medical Image Analysis 13, 419–431. Khamene, A., Xu, C., 2009. Towards guidance of electrophysiological procedures
King, A.P., Rhode, K.S., Ma, Y., Yao, C., Jansen, C., Razavi, R., Penney, G., 2010. with real-time 3D intracardiac echocardiography fusion to C-arm CT. Medical
Registering preprocedure volumetric images with intraprocedure 3-D Image Computing and Computer Assisted Intervention (MICCAI’09) Medical
ultrasound using an ultrasound imaging model. IEEE Transactions on Medical Image Computing and Computer Assisted Intervention (MICCAI’09), pp. 9–16.
Imaging 29, 924–937. Yu, H., Fahrig, R., Pelc, N.J., 2005. Co-registration of X-ray and MR fields of view in a
Klemm, H., Steven, D., Johnsen, C., Ventura, R., Rostock, T., Lutomsky, B., Risius, T., hybrid XMR system. Journal of Magnetic Resonance Imaging 22, 291–301.
Meinertz, T., Willems, S., 2007. Catheter motion during atrial ablation due to the Yushkevich, P., Piven, J., Hazlett, H., Smith, R., Ho, S., Gee, J., Gerig, G., 2006. User-
beating heart and respiration: impact on accuracy and spatial referencing in guided 3D active contour segmentation of anatomical structures: significantly
three-dimensional mapping. Heart Rhythm 4, 587–592. improved efficiency and reliability. Neuroimage 31, 1116–1128.
Knecht, S., Wright, M., Akrivakis, S., Nault, I., Matsuo, S., Chaudhry, G.M., Haffajee, C.,
Sacher, F., Lellouche, N., Miyazaki, S., Forclaz, A., Jadidi, A.S., Hocini, M., Ritter, P.,