TomoTherapy
®
Hi Art System
®
·
The TomoTherapy Hi·Art treatment system incorporates unique design features that
combine to deliver highly conformal radiation therapy guided by a specialized multileaf
collimator (MLC). Four key advantages include:
• Complete system integration • Helical pattern delivery
• CTrue™ imaging technology • Discrete-angle, sliding beam delivery (optional)
At the heart of the Hi·Art treatment system is a common database that facilitates the entire
therapy process from planning to treatments to archive. The central database facilitates:
• Fast CT-guided dose targeting • Effective adaptive planning
• Precise treatment delivery • Easy treatment planning for common and complex cases
The Hi·Art treatment system creates verification CTrue images using low-intensity
megavoltage X-rays from the same linear accelerator used to treat the patient. This
allows clinicians to take an MVCT scan immediately prior to any treatment fraction
to verify the patient’s anatomy and position as compared to the original treatment plan,
and adjust as necessary.
Simplifying
radiation therapy through
total system integration
Physical Characteristics
Mechanical Features
S P E C I F I C AT I O N PERFORMANCE
Gantry
Direction of rotation Clockwise viewed from the foot of the couch
Rotational angle accuracy Within 0.5 degrees
Target to axis distance 85 cm
Mechanical isocenter stability <0.4 mm
Isocenter height 113 cm typical (dependent upon finished flooring)
Cooling Integrated onboard cooling system
Photon Beam
S P E C I F I C AT I O N PERFORMANCE
Accelerator type Standing Wave (0.3 meters)
Microwave power 2.5 MW (magnetron)
Redefined Nominal dose rate**
Nominal energy**
850 cGy/min
6 MV, single energy
accuracy Field size range at isocenter Selectable
1.0 cm x 0.625 cm to 1.0 cm x 40 cm
2.5 cm x 0.625 cm to 2.5 cm x 40 cm
5.0 cm x 0.625 cm to 5.0 cm x 40 cm
Treatment Maximum radiation field length 150 cm with Couch at height of isocenter plane
Treatment volume - TomoHelical 80 cm (transverse diameter) x 135 cm (longitudinal)
flexibility
Treatment volume - TomoDirect
for typical patient set-up*
40 cm (transverse diameter) x 135 cm (longitudinal)
for typical patient set-up* (TomoDirect is an optional feature)
Average MLC leakage 0.25% (typical)
Efficient Collimation
throughput S P E C I F I C AT I O N
Jaw collimation
PERFORMANCE
CTrue™ Imaging
S P E C I F I C AT I O N PERFORMANCE
Geometry Fan-beam
Dose per MVCT image (typical) 0.5 - 3 cGy depending on resolution and body thickness
Image resolution (xy) 512 x 512 (0.78 mm pixels)
Slice spacing available 2 mm, 4 mm, 6 mm
Scan time Typically 2 minutes per 10 cm length at 4 mm slice spacing
Field of view (FOV) 40 cm diameter
Source to detector distance 145 cm
Spatial resolution Nominal 0.5 lp/mm at 10% MTF
Contrast resolution 2% density for 2 cm object (typical)
Image guidance mode Daily 3D MVCT matched with 3D kVCT
* Actual treatment volumes are variable depending upon Couch height. Volume measures based upon single set-up, without field matching.
** Performance based on 5 cm field size at SSD=85 cm.
TomoTherapy Inc. reserves the right to make changes in specifications and/or to discontinue any product at any time
without notice or obligation and will not be liable for any consequences resulting from the use of this publication.
©2011 TomoTherapy Incorporated. All rights reserved. The following marks as used herein in italicized or stylized
text are among trademarks, service marks or registered trademarks of TomoTherapy Incorporated in the United
States and other countries: TomoTherapy, Tomo, TomoDirect, TomoHelical, TomoHD, TomoMobile, TQA, Hi·Art, Cert no. BV-COC-963674