KYNAR® BODY
ments.
HANNA Kynar® is the best plastic material that
you can use for the electro-analytical STORAGE
INSTRUMENTS instruments. • To minimize clogging and ensure a quick
ELECTRODE Kynar® gives to your electrode the MAXI- response time, the glass bulb and the
INSTRUCTION MUM PROTECTION AGAINST junction should be kept moist and not
AGGRESIVE CHEMICALS. allowed to dry-out.
GUIDE
Replace the protective cap with a few
PREPARATION
drops of HI 80300 storage solution or, in
• Remove the protective cap. DO NOT BE its absence, filling solution (HI 8071 for
ALARMED IF ANY SALT DEPOSITS ARE single junction or HI 8082 for double junc-
PRESENT. This is normal with electrodes tion electrodes) or pH 7.01 or pH 4.01
and they will disappear when rinsed with buffers (HI 8007 or HI 8004 respectively).
Code: water.
Follow Preparation above before taking
• Shake the electrode down as you would do measurements.
with a clinical thermometer to eliminate
Note: NEVER STORE THE ELECTRODE
any air bubbles inside the glass bulb.
______________ • If the bulb and/or junction are dry, soak the
IN DISTILLED WATER.
TROUBLE SHOOTING:
pH Meter: Follow attentively the meter's
operating and calibration procedures from
the instruction manual.
Electrode: Evaluate your electrode perfor-
mance based on the following possibilities. 1) Glass Bulb
• Noise (Readings fluctuate up and down) 2) Reference Junction
could be due to:
3) Reference Wire
• Clogged/Dirty Junction: Refer to Clean-
ing Procedure above. 4) Sensitive Wire
• Loss of shielding due to low electrolyte 5) Reference Fill Hole
level (in refillable electrodes only): Empty
Made in
Canada
electrolyte with a syringe and refill with 6) Fill Hole Sleeve Cover
fresh HI 8071 for single junction or HI 8082
7) Fill Hole Screw