Specifications
system
Function
Sample
Reagent
Probe
‘& Mixer
Reaction
System
Optical
System
calaQc
Others
‘Category Rondomiaccess: uly automatic chemisty analyzer
Throughput 150 testsfhour
Methodology End point, Fixed Time: Kinetic:
Reagent ‘Open System
Reagent Position 40
Sample Position 40
Minimum Reaction Volume! 1500L
‘Canyover <0,005%
Water Consumption Low tol at/Hour
‘Maximum Reaction Time 18 minutes (Single reagent), 12 minuies(Dual reagent)
Maximum Reaction Volume 500.UL
Principle ‘Colorimetty, Turbidimetry
Testing Mode. Regular mode (singles.dual reagents}; fas! mode|single reagent)
‘Sample Mode Random access, SIAT sample priority
‘Sample Volume! 2-50UL, step by 0.1/UL
‘Sample Disk 40 posiions
reer poet (One probe for somple'ard reagent with liquid level detection: vertical ang
horizontal colision protection and reagent volume real fime monitoring function
Sample Tube Semel ea 12x100 mii 12x75 mm), Vacuum
Reagent Volume £87:150°450 UL. R2:10-300UL, step by 1ul
Keoger Ose “01683 postions 24ers Molen DSHD 2
Reagent Bolties Compatibie with Olympus and Hitachi reagent bolies
Mixer High poised nonomatgriel miner ‘
Reaction Cuvettes 50 cuvetles, semi:permanent rigid UV special plaslc ewveltes
Wash Station éstep ago wash station — i
Reaction Temperature” 3720.1
Heater Metal thifmostatt forreaction cisk aN
Lamp Halogen Tungsten lamp, 12V20W, about 2000hours, water cooling system
Filter FMSS(Fiisealed Matrix Spectiomelric System)
Wavelength SB Wavelengths: 340/408 430,510,546 578.630.6708
Resolution = TQ) \ap.0o01Abs a_i
Linearity Range: 0:3 SAbs
‘Accuracy ‘0.5A:< + O.02Abs. | 0A: <* 0.04Abs E
stray Uoht 245
Stability “<0.01Abs/hour aa |
cv <15%
Half Band Accuracy <22nm ye
Detector Photodiode detector onay
LL SS "= ie ae
‘Calibration Curve Coltbration curve avto check, auto curve fiting
QC Rule Wastgaittmultivule.COmmtive sum check, [win Pior
ac Curve Weastgard mult-rvle. Cumulotive sum check, Twin Plot
‘QC Warning Out of contol auto waming, date Bomatic record and analysis
‘Operating System Windows 7/10
Date Storage ‘Decided by computer hard dsc capacity
Us interface BiUSinterioce
Print Mode Mult defauit formats, user-defined formats
Real fire monitoring for sample disk, reagent disk; reaction disk
Real ime monitoring for QC status. Real fime monitoring for reaction cuvettes
System Monitor staius, lamp status ond temperature. Real time monitoring for reagent volume,
feaction curve. calbralion curve and QC curve. linearly range imitation,
substrate exhaustion judgement and prozone detecting. Abnormal status waming,
User authority setting, test panel function, calculated/manual parameters
programming, carryover setting, sample and reagent blank auto calculation,
gether Funcien, ‘automatic failure recovery, aulomatic print, data statistic, auto/manval