p 1 aK -4 S| OPERATING MANUAL CARBON DISPERSION TESTER ADVANCE EQUIPMENTS 3&9, VASANT VAZE INDUSTRIAL COMPOUND, BEHIND MAHINORA SHOW ROOM, KALYAN SHIL RD, MHATRE NAGAR, MANPADA, DOMBIVLI-EAST, THANE-#21201, ‘TAL: 491.251.2871696 ‘MMosILe: +91-9820351166/9969168109 EMAIL: advanceequioments@gmall WEBSITE: www advanceequipments.in INTRODUCTION: 1 sction Microscope --~ 2, 100X, 450X lenses ~ 3. Hot Plate -- 01 Qty 4, Glass Slides 01 Qty. 5. Tweazer- o1aty PROCEDURE: LypPlace one clean Microscope slide on a hot plate. Keep the energy ._ tegulator knob oh point 8 so that temperature is maintained between Wy 170°C to 210°C. Place three specimen of pin head size having approximate weight of 5 mg on slides of 20 mm apart. 2. Place the second slide on the top of first slide, press both slide with help of tweezer applying equal pressure & keep it on hot plate for 2/3 minutes. 3, Switch “OFF” the hot plate, allow the slides to cool down & then place them under microscope for examination. REPORT: The carbon black dispersion in the material under test shall be considered to be satisfactory if the specimens show a uniform background free from white streaks,

You might also like