Professional Documents
Culture Documents
VegaSys Service Manual Rev.01
VegaSys Service Manual Rev.01
Service Manual
TABLE OF CONTENTS
Chapter 01 INSTALLATION
Chapter 06 MAINTENANCE
CHAPTER 01
- INSTALLATION -
INDEX
1.1 UNPACKING……........................................................................................................... 2
1.2 INSTRUMENT INSTALLATION... ............................................................................... 4
1.2.1 INSTALLATION SITE……............................................................................................ 4
1.2.2 ELECTRICAL POWER SUPPLY ................................................................................... 4
1.2.3 ACCESSORIES CONNECTIONS .................................................................................. 5
1.2.3.1 Power Supply........................................................................................................... ..... 5
1.2.3.2 Instrument-Computer connection ................................................................................. .5
1.2.3.3 Liquid connections…………….............................…………………………………....5
1.2.4 LABELS AND SYMBOLS..............................................................................................6
1.2.5 WARRANTY...…….…………….……………………………………………………...7
1.2.6 REGULATORY COMPLIANCE...………………...……..………..……………..….…7
1.2.7 LIMITATION OF USE..........................................................................................……....7
1.1 UNPACKING
The Vegasys is packed and delivered in one wooden crate/box containing the instrument and
the its accessories.
The packing modalities have been expressly designed to insure maximum protection of the
contents during shipping and handling. Nevertheless it is important to carefully check the
crate/box upon delivery in order to ascertain its integrity.
Special attention should be dedicated to examining the color of the ShockWatch ® glued to
the crate: it must be white. A ‘red’ ShockWatch ® indicates that the crate has experienced some
sort of mishandling (shock) during transport and/or delivery. This fact must be recorded by the
courier on the delivery note, as must any and all visible external damage to the packing crate
(e.g.: holes, dents, rips or tears, water marks, etc.) evident at the delivery moment. This serves to
simplify matters in the event of any future claims for damages.
NOTE: THE CARRIER ASSUMES TOTAL RESPONSIBILITY FOR THE CRATE AND/OR
BOX AND THEIR CONTENTS ONCE THE GOODS HAVE LEFT THE FACTORY
UNTIL THE MOMENT THEY ARE DELIVERED AT DESTINATION. THE CARRIER
IS RESPONSIBLE FOR ANY AND ALL DAMAGES ARISING DURING HANDLING,
STORAGE AND TRANSPORTATION. ANY AND ALL DAMAGE CLAIMS MUST BE
FILED AGAINST CARRIER IMMEDIATELY UPON DISCOVERY.
Do not discard the delivery crate/box or the packing material until the correct functioning of
the instrument has been checked.
Before connecting the "Vegasys", remove the protective packing material of the sampling and
aspiration arms.
Warning: in case it is necessary to repack the delivered unit, the following procedure must be
carefully followed:
Reposition the protective packing material of the sampling and aspiration arm.
Tape down (using masking tape if possible) the main cover of the instrument.
Fill the empty spaces around the accessories packed in the crate using “pluriball” or other
suitable packing material.
In case the repacking is done after clinical diagnostic use of the instrument, it is necessary to
first perform the decontamination procedure as outlined in Chapter 6 of this manual.
⇒ ELECTRONIC STABILIZER
Used to stabilize the power source voltage in the laboratory. Any stabilizer available on
the market can be used (power not less than 0.5 kW).
The following WARNING label can be found on the left side of the instrument:
PLEASE NOTE: NEVER OPEN THE INSTRUMENT WITHOUT HAVING FIRST TURNED IT OFF AND
DISCONNECTED IT FROM ITS POWER SOURCE (UNPLUG THE POWER SUPPLY
CABLE). CAREFULLY FOLLOW THE CLEANING AND MAINTENANCE PROCEDURES
AS OUTLINED IN CHAPTER 6 OF THIS MANUAL.
EARTH GROUND
BIOLOGICAL HAZARD
(Placed next to the Samples Plate and the waste discharge red fitting)
Wear proper protective clothing/equipment and handle all biological
samples, consumables and waste liquids produced during analysis with
extreme care. The disposal of the used consumables and the waste liquids
produced must conform to all applicable government laws and regulations.
Furthermore, to prevent overflow of the discharge tanks, make sure to
periodically check the level of the waste liquids collected.
MANUFACTURING COMPANY
This symbol is placed on the S/N plate next to the name of the
manufacturer.
DISPOSABLE WARNING
This symbol is placed on the S/N plate and means that the instrument and
the parts are not to be disposed as ordinary household waste/garbage. They
must be disposed according to applicable government regulations regarding
electrical and electronic equipment (Directive 2002/96/EC WEEE).
1.2.5 WARRANTY
AMS guarantees the replacement of all defective components and/or materials for a period up
to 14 (fourteen) months from the date of invoice. Warranty, as for all Technical Assistance in
general, is intended as provided ex works, Rome, Italy.
Warranty does not cover consumables or those parts of the instrument in direct contact with
liquids. A list of the parts excluded from warranty is provided in the table below.
Moreover, the warranty is not valid in case damages to the VegaSys arise from:
¾ improper use of the instrument (or use in any manner not in conformance with
instructions provided by the Manufacturer or the Seller);
¾ mishandling and/or transportation errors;
¾ insufficient (or lack of) ordinary preventive maintenance by the User.
Any damage arising from mishandling or caused during transport must be immediately
reported to the Shipping Company upon delivery of the goods.
Description
Reagents containers
Samples cups
Reaction Racks
Liquid bottles
Tubes
Probes
Lamp
Interferential filters
CHAPTER 02
- GENERAL DESCRIPTION -
INDEX
2.2.7 INCUBATOR....................................................................................................................5
The maximum volume capacity of the reaction cells, sample cups and reagent containers for
the analytical plate are as follows:
¾ Volume of reaction cell: 1.2 ml;
¾ Volume of sample cup: 0.8 ml;
¾ Volume of reagent containers: 25 ml.
The recommended minimum sampling volume for analysis is 5 µl and for reagent volume is
400 µl. There is also an automatic washing cycle in between tests to avoid the risk of cross-
contamination.
Washing well
Sampling arm
2.2.7 INCUBATOR
The incubation temperature of the reaction cell is fixed at 37°C ( ± 0,2°C ) to ensure proper
incubation for all samples.
Fig. 5 – Photometer
NOTE: even though most computers declare the same technical and operative characteristics,
some of these could have different hardware installed.
This could cause problems to the VegaSys software while running tests. Therefore, if
the PC is bought separately/locally, it is strongly recommended to test the system
before proceeding with the installation at customer laboratories.
Consequently, AMS denies any responsibility for software problems that are due to
computer not previously tested with the instruments.
⇒ Dimensions:
Height: 32 cm
Depth: 45 cm
Length: 50 cm
⇒ Weight: 22 Kg
⇒ Sampling Arm: provides to aspirate reagent and sample; automatic probe washing.
⇒ Aspiration Arm: transfers the reaction mixture from the reaction cell to the micro-
flow cell of the photometric system of measurement.
⇒ Reaction Plate: six sectors of 19 reaction cuvettes for a total of 114 cuvettes.
⇒ Optic System:
Photometer: multi wavelength photometer with narrow band interferential filters;
Wavelength: 340 nm, 405 nm, 492 nm, 510 nm, 546 nm, 578 nm and 620 nm;
Light source: halogen lamp 6V/10 W;
Linear range: 0.001 ÷ 2.500 Abs;
Resolution: 0.001 Abs;
Detector: silicon photodiode.
CHAPTER 03
- ELECTRICAL SCHEMES AND DRAWINGS -
INDEX
CHAPTER 04
- DIAGNOSTIC PROGRAM -
INDEX
4 DIAGNOSTIC PROGRAM
The diagnostic program enables the operator to perform a complete check for each of the Vegasys
module functions.
This program has a folder structure, with each folder containing functions pertaining to the specific
module. To launch the program, the operator has to click with the left side of the mouse, on the
Diagnostic button located on the lower right side of the screen of the System Monitor (Fig. 1).
The Vegasys must be in the stand-by state to access this area.
“Diagnostic” button
Click on the “Diagnostic” button (Fig. 1), type the password “1234” in the “Password request”
dialog (Fig. 2) and click on its “OK” button to enter in the Diagnostic program.
This operation can be canceled by clicking on “Exit” button of the “Password request” dialog (Fig. 2).
Once the “Diagnostic” program starts, the following two buttons are available:
The following information for the options currently being used is given on this screen:
⇒ Icon Symbol for the Language Selected
⇒ Master Hardware Version
⇒ Serial Port
⇒ Firmware Version of the Analytical Control Board
⇒ Master Hardware DataBase Version
⇒ Serial Number of the Analytical Control Board
⇒ Master Hardware Ini Version
⇒ Instrument link
To enter in the “Sampling/Arm Plate Settings” folder, click on “Arm/Plate Setting” button (Fig. 4);
then, the “Input Operator Name and Password” dialog appears (Fig. 4):
If type the password “69231” in the “Password” field of the “Input Operator Name and Password”
dialog (Fig. 4), any name in its “Operator Name” field and give the confirm the “Sampling/Arm Plate
Settings” folder appears (Fig. 5):
Status table
Otherwise, to enter in the “Sampling/Arm Plate Settings” folder and have the chance to enter in the
“Program Download”, “Cuvette Temperature” and “Plate Temperature” folders, click on “Arm/Plate
Setting” button (Fig. 4); then, the “Input Operator Name and Password” dialog appears (Fig. 4). If type
the password “28761” in its “Password” field, any name in its “Operator Name” field and give the
confirm the “Sampling/Arm Plate Settings” folder appears and “Download”, “Cuvette Temperature”,
“Plate Temperature” buttons appear (Fig. 6):
The Diagnostic program has several buttons on the left side of the screen (Fig. 5 and Fig. 6). These
buttons refer to the following functions of the instrument:
Arm/Plate Setting: checks if the Sampling Arm, the Aspiration Arm and the Reactions Plate
are functioning correctly and performs mechanical calibration.
Diluter: checks the performance of the Diluter module.
Change Photometer Lamp: checks the performance of the Photometer Lamp.
Optical Density: measures the absorbance of the solution in the Micro-flow Cell.
Peristaltic Pump: to calibrate the aspiration of the Peristaltic Pump.
Miscellaneous: to reset the Reactions sectors, print the parameters set for the various modules
and to select a serial port from those identified by the program.
Download: to download and update the firmware of the Analytical Control Board.
Cuvette Temperature: to calibrate the temperature of the Micro-flow Cell.
Plate Temperature: to calibrate the temperature of the Reactions Plate.
The Status table (Fig. 5) shows - near every folder - the results (under the heading “Status”) of
various actions which have been run by the technician/operator in the Diagnostic program.
To save the new setted parameters exit from the pertinent folder, then select “Yes” in the
“MHVega” form.
⇒ Cuvette-Aspiration
⇒ NO3
⇒ Free Cover
⇒ All Positions
⇒ System Settings
: disengages the motors to enable the Sampling Arm and Reactions Plate
to be manually adjusted.
Note: to engage the motors again just select “General Reset”.
Once the number of steps for the descent are defined, by selecting the command “Down” the
Sampling Probe goes down to the Washing Well using the number of steps programmed (in the Fig. 8,
60 steps).
The four small buttons for the “Down” command are the following:
: increases the number of steps down of “1”;
: increases the number of steps down of “5”;
: decreases the number of steps down of “1”;
: decreases the number of steps down of “5”.
The relative “Up” button rises up the Sampling Probe from the Reagent container of a number of
steps specified in the small window under its arrow (Fig. 9).
Modify the above mentioned number of steps up by opportunely selecting the four small buttons
placed under the arrow of the “Up” button; enable them (Fig. 9) by selecting the command “Go -
Reagent”.
The four small buttons for the “Up” command are the following:
: increases the number of steps up of “1”;
: increases the number of steps up of “5”;
: decreases the number of steps up of “1”;
: decreases the number of steps up of “5”.
It is possible to improve the centering of the Sampling Arm (Arm) with respect to the Reagents
Plate (Plate) position #1, if the option “1” is selected (else, with respect to the Reagents Plate position
#10 if the option “10” is selected), by selecting:
¾ the four small buttons placed above the “Go - Reagent” button to move the Sampling Arm
in a clockwise/counter-clockwise movement;
¾ the four small buttons placed under the “Go - Reagent” button to move the Plate in a
clockwise/counter-clockwise movement.
The four small buttons for the “Arm” are the following:
: moves the Sampling Arm clockwise 1 step at a time;
: moves the Sampling Arm clockwise 5 steps at a time;
: moves the Sampling Arm counter-clockwise 1 step at a time;
: moves the Sampling Arm counter-clockwise 5 steps at a time.
The same rules apply to the four small buttons of the “Plate” for moving the Reagents Plate with
respect to the Sampling Arm.
The same rules given in the “Reagents” area apply to the four small buttons of this “Down” button.
The relative “Up” button rises up the Sampling Probe from the Sample cup of a number of steps
specified in the small window under its arrow (Fig. 10).
Modify the above mentioned number of steps up by opportunely selecting the four small buttons
placed under the arrow of the “Up” button; enable them (Fig. 10) by selecting the command “Go –
Sample”.
The same rules given in the “Reagents” area apply to the four small buttons of this “Up” button.
It is possible to improve the centering of the Sampling Arm (Arm) with respect to the Samples Plate
(Plate) position #1 by selecting:
¾ the four small buttons placed above the “Go - Sample” button to move the Sampling Arm in
a clockwise/counter-clockwise movement;
¾ the four small buttons placed under the “Go - Sample” button to move the Plate in a
clockwise/counter-clockwise movement.
The same rules given in the “Reagents” area apply to the eight small buttons of the “Go - Sample”
button.
The same rules given in the “Reagents” area apply to the eight small buttons of the “Go / Cuvette-
Sampling” button.
The relative second “Down” button lowers the Aspiration Probe down for the successive second
descent into the cuvette of a number of steps specified in the small window under its arrow (Fig. 12).
Modify the above mentioned number of steps down by opportunely selecting the four small buttons
placed under the arrow of the second “Down” button; enable them (Fig. 12) by selecting the command
“Go / Cuvette-Aspiration” button.
This successive second descent should aspirate the minimum quantity of the reaction liquid
necessary for a reading.
The same rules given in the “Reagents” area apply to the four small buttons of this two “Down”
buttons.
It is possible to improve the centering of the cuvette #1 of the Reactions Plate (Plate), if the option
“1” is selected (else, the centering of the cuvette #58 of the Reactions Plate if the option “58” is
selected), with respect to the Aspiration Arm (Arm) by selecting the four small buttons placed under
the “Go / Cuvette-Aspiration” button to move the Plate in a clockwise/counter-clockwise movement.
The same rules given in the “Reagents” area apply to the four small buttons of the “Go / Cuvette-
Aspiration” button.
: allows to execute endlessly a specific cycle by moving the two Arms and
the Analytical Plate. Note: to stop, click on the “General Reset” button.
Status table
The field “Cycles (1..9)” enables the operator to set the number of cycles to be run (Priming). The
minimum number of cycles to be run is 1 and the maximum is 9. To confirm the number of cycles
programmed, push “Enter” (from the keyboard).
The command “STOP” enables the operator to stop the “PRIME” procedure.
Status table
The command “STOP” enables the operator to stop the “AUTOZERO” procedure.
By selecting the “Start Reading” button the following appears on the screen (Fig. 15):
¾ a graph showing the time trend of the voltage read [“mv”] for the filter selected;
¾ a table which shows the voltage read [“mv”] for the filter selected.
: this command automatically performs the reading of the liquid present in the
Micro-flow cell for each filter. It is important to make sure that the Micro-
flow cell is full of liquid.
The command “STOP” enables the operator to stop the “Filters Reading” procedure.
Status table
By selecting the “Start Reading” button the following appears on the screen (Fig. 17):
¾ a graph showing the time trend of the O.D. read for the filter selected;
¾ a table which shows the O.D. and “mV” read for the filter selected.
Status table
: these fields enables the operator to program the number of aspiration steps
for the Peristaltic Pump during calibration (“Final Aspirat.”) and the
number of calibration cycles to run (“Cycles (1..9)”). To confirm the two
numbers programmed, push “Enter” (from the keyboard).
The field “Aspiration Steps” enables the operator to program the number of steps of the Peristaltic
Pump in order to improve the aspiration of the liquid (Minimum number of steps =1; Max= 4000). To
confirm the number of steps programmed, push “Enter” (from the keyboard).
Status table
Section heading
ÆImportant: it is possible to change the status of each single cuvette by clicking on the same.
The field “First Available Cuvette” indicates the first reaction cuvette available for sampling
(cuvette #54 in the Fig. 19).
S3/C16 - see Fig. 19 - indicates the sector and location available for sampling.
ÆImportant: the symbol “?” indicates that there are no reaction cuvettes available for the
sampling process.
The command “Serial Port” allows the user to select a serial port
from a pull down menu list.
Status table
Settings table
Fig. 20 – “Program Download” folder
The command “FDT Download” (Fig. 20) allows the technician/operator to execute the download
procedure of the firmware in the instrument (using the FDT program).
The command “Self Download” (Fig. 20) allows the technician/operator to execute the download of
the firmware in the instrument when the firmware version is ≥ 1.14 or instrument link = USB.
The Settings table (Fig. 20) shows all the System Settings (under the heading “Value”).
Status table
The command “Start Reading” (Fig. 21) allows to start the temperature reading into the Micro-flow
cell. It turns to “STOP Reading” when selected (Fig. 22), to stop the reading.
By selecting the “Start Reading” button the following appears on the screen (Fig. 22):
¾ a graph showing the time trend of the cuvette temperature [°C];
¾ a table which shows the cuvette temperature [°C] and its value in counts.
The command “Auto - Regulation” (Fig. 21) allows to start the auto-regulation procedure of the
temperature into the Micro-flow cell. This regulation procedure stops automatically.
Status table
The command “Start Reading” (Fig. 23) allows to start the temperature reading of the Reactions
Plate. It turns to “STOP Reading” when selected (Fig. 24), to stop the reading.
By selecting the “Start Reading” button the following appears on the screen (Fig. 24):
¾ a graph showing the time trend of the plate temperature [°C];
¾ a table which shows the plate temperature [°C] and its value in counts.
It is possible to modify the Set Point of the plate temperature by selecting the four small buttons
placed under the “STOP Reading” button; enable them (Fig. 24) by selecting the command “Start
Reading” button.
The same above-mentioned rules apply to the four small buttons of the Set Point.
Æ Pay attention: the recommended Set Point is “1000” (as showed in Fig. 23).
CHAPTER 05
- SETTINGS AND ADJUSTMENTS -
INDEX
5.2 Diluter………………………………………………………………... 7
5.2.1 Replacement of the Diluter Assy……….……………………………. 7
5.2.2 Replacement of the Diluter Home Sensor……………………….…… 9
5.2.3 Replacement of the Diluter Electrovalve…………………………….. 10
5.2.4 Replacement of the Reel………………………………...…………… 10
5.3 Photometer…………………………………………………………... 12
5.3.1 Electronic adjustments of the Photometer……………………….…… 12
5.3.2 Replacement of the Photometer Assy………………………………... 14
5.3.3 Replacement of the Filters Drum Motor……………………………... 15
5.3.4 Replacement of the Filters Drum Home Sensor…………….……….. 16
5.3.5 Replacement of the Photometer Halogen Lamp……………………... 17
5.3.6 Replacement of the “A/D C Board” of the Photometer……………….. 19
5.3.7 Replacement of the Interferential Filters Kit…………………………. 20
5.3.8 Replacement of the Micro-flow Cell…………………………………. 21
Knurled knob
Cover
Fastening
screws
Arm screw
Sampling Probe
Fastening
Cover screw
(a) (b)
Fig. 1 [(a), (b)] – Sampling Arm
To replace the Sampling Arm (Fig. 3 – P/N: 1-15-0004-01) perform the following steps:
6. Replace the new Sampling Arm on its axis and screw slightly the Arm screw [Fig. 1 (b)].
7. Plug carefully the motor connector in the Sampling Arm board.
8. Replace the two parts of the cover on the arm inserting carefully the tube of the probe into its
small canal on the cover, then screw its four fastening screws (Fig. 1).
9. Replace the Sampling Probe in its housing and tighten its knurled knob [Fig. 1 (a)].
10. Rotate the Sampling Arm in the right toward in order to obtain the centering with respect to the
Washing Well (Fig. 3), then tighten the Arm screw [Fig. 1 (b)].
11. Turn on the VegaSys system (instrument and computer) and launch the “Diagnostic” program.
12. Select “Sampling Arm/Plate Settings” folder and request an “General Reset” (Fig. 2). Wait
until the reset instrument procedure has been completed.
After having mechanically centered the Sampling Probe over the Washing Well (compare the above
step 10), regulate all the other positions by using the buttons included in the “Sampling Arm/Plate
Settings” folder (Fig. 2) as described in the following instructions:
13. Select “Go Æ Washing Well” button. Make sure that the number placed below the relative
“Down” button is “60” (as showed in Fig. 2). If it not be so, insert the number “60” by using
the relative shift buttons.
14. Select the “Go Æ Reagent” button. The Sampling Probe will automatically position itself over
the Reagent in position # 1.
15. Select the relative “Down” button. If necessary, adjust the centering of the Sampling Probe
with respect to the reagent container opening by using the buttons placed above and below the
“Go Æ Reagent” button (“Arm - shift buttons” used to adjust the position of the Sampling
Arm with respect to the Analytical Plate and “Plate - shift buttons” used to adjust the position
of the Analytical Plate with respect to the Sampling Arm).
16. Adjust the descent of the Sampling Probe using the relative “Down” button. Make sure that the
distance between the bottom of the reagent container and the tip of the Sampling Probe is about
2 mm. If it not be so, adjust by using the relative shift buttons to change the number of steps.
17. Adjust the upward movement of the Sampling Probe by using the relative “Up” button. Make
sure that the tip of the Sampling Probe comes out of the cleaning pad by about 1 mm. If it not
be so, adjust by using the relative shift buttons to change the number of steps.
18. Next to the “Go Æ Reagent” button, select option “10”.
19. Click on “Go Æ Reagent” button. The Sampling Probe will automatically place itself over the
Reagent in position # 10.
20. Repeat the above steps in the following order: 15, 16 and 17 (for the position # 10 of the
Reagent).
21. Select the “Go Æ Sample” button. The Sampling Probe will automatically position itself over
the Sample in position # 1.
22. Select the relative “Down” button. If necessary, adjust the centering of the Sampling Probe
with respect to the sample container opening by using the buttons placed above and below the
“Go Æ Sample” button (“Arm - shift buttons” used to adjust the position of the Sampling
Arm with respect to the Analytical Plate and “Plate - shift buttons” used to adjust the position
of the Analytical Plate with respect to the Sampling Arm).
23. Adjust the descent of the Sampling Probe by using the relative “Down” button. Make sure that
the distance between the bottom of the sample container and the tip of the Sampling Probe is
about 2 mm. If it not be so, adjust by using the relative shift buttons to change the number of
steps.
24. Adjust the upward movement of the Sampling Probe by using the relative “Up” button. Make
sure that the tip of the Sampling Probe comes out of the cleaning pad by about 1 mm. If it not
be so, adjust by using the relative shift buttons to change the number of steps.
25. Select the “Go Æ Cuvette-Sampling” button. The Sampling Probe will automatically position
itself over the cuvette # 1.
26. If necessary, adjust the centering of the Sampling Probe with respect to the cuvette opening by
using the buttons placed above and below the “Go Æ Cuvette-Sampling” button (“Arm - shift
buttons” used to adjust the position of the Sampling Arm with respect to the Analytical Plate
and “Plate - shift buttons” used to adjust the position of the Analytical Plate with respect to the
Sampling Arm).
27. Make sure that the number placed below the relative “Down” button is “160” (as showed in
Fig. 2). If it not be so, insert the number “160” by using the relative shift buttons.
28. Select the “General Reset” button (Fig. 2). Wait until the reset instrument procedure has been
completed.
29. Exit from “Sampling Arm/Plate Settings” folder, then select “Yes” in the “MHVega” form to
save the new setting parameters.
Considering that the Aspiration Arm is alike at the Sampling Arm, make reference
Note
to the Fig. 1 for its replacement procedure.
Instrument cover
Washing Well
Aspiration Arm
Photometer
cover
Sampling Arm
Analytical Plate
Chassis
Quarter-turn
locks
Fig. 3 – Instrument
To replace the Aspiration Arm (Fig. 3 – P/N: 1-15-0005-01) perform the following steps:
6. Replace the new Aspiration Arm on its axis and screw slightly the Arm screw [Fig. 1 (b)].
7. Plug carefully the motor connector in the Aspiration Arm board.
8. Replace the two parts of the cover on the arm inserting carefully the tube of the probe into its
small canal on the cover, then screw its four fastening screws (Fig. 1).
9. Center the Aspiration Arm with respect to the relative opening on the chassis, then tighten the
Arm screw [Fig. 1 (b)].
10. Replace the Sampling Probe in its housing and tighten its knurled knob [Fig. 1 (a)].
11. Turn on the VegaSys system (instrument and computer) and launch the “Diagnostic” program.
12. Select “Sampling Arm/Plate Settings” folder and request an “General Reset” (Fig. 2). Wait
until the reset instrument procedure has been completed.
After having mechanically centered the Aspiration Arm with respect to the relative opening on the
chassis (compare the above step 9), regulate the other positions by using the buttons included in the
“Sampling Arm/Plate Settings” folder (Fig. 2) as described in the following instructions:
13. Select “Go Æ Cuvette-Aspiration” button, then select the first “Down” button. If necessary,
adjust the centering of the Sampling Probe with respect to the cuvette opening by using the
buttons placed below the “Go Æ Cuvette-Aspiration” button (“Plate - shift buttons” used to
adjust the position of the Analytical Plate with respect to the Aspiration Arm).
14. Make sure that the number placed below the first “Down” button is “94” (as showed in Fig. 2).
If it not be so, insert the number “94” by using the relative shift buttons.
15. Adjust the second descent of the Sampling Probe by using the second relative “Down” button.
Make sure that the distance between the bottom of the cuvette and the tip of the Sampling
Probe is about 2 mm. If it not be so, adjust by using the relative shift buttons to change the
number of steps.
16. Next to the “Go Æ Cuvette-Aspiration” button, select option “58”.
17. Click on “Go Æ Cuvette-Aspiration” button. The Sampling Probe will automatically place
itself over the cuvette # 58.
18. Repeat the above steps in the following order: 13, 14 and 15 (for the cuvette # 58).
19. Select the “General Reset” button (Fig. 2). Wait until the reset instrument procedure has been
completed.
20. Exit from “Sampling Arm/Plate Settings” folder, then select “Yes” in the “MHVega” form to
save the new setting parameters.
5.2 DILUTER
5.2.1 Replacement of the Diluter Assy
1. Make sure that the VegaSys instrument is turned off.
2. Center the Sampling Arm with respect to the relative opening on the chassis, unscrew the four
quarter-turn locks (Fig. 3) and remove the chassis from the instrument.
3. Unscrew the two fittings from the head of the Diluter (Fig. 4), paying attention to the spillage of
liquid from the tubes.
4. Unscrew the four fastening screws of the cover of the Analytical Control Board and remove the
cover from its housing.
5. Unplug the J5 and J9 connectors from the Analytical Control Board (compare the scheme
SC-30-02314-00).
6. Unscrew the two fastening screws of the Diluter mechanical support (Fig. 4) and take out it
from the instrument.
7. Unscrew the four fastening screws of the Diluter (Fig. 4) and take out it from its mechanical
support.
Ref. 4 tube
Diluter mechanical support
Fastening screw
Fastening screw
of the Diluter
mechanical support
Reel
Diluter
Diluter Electrovalve
Diluter Electrovalve fastening screw
To replace the new Diluter Assy (Fig. 5) perform the following steps:
8. Repeat the above steps in inverse order: from 7 to 3.
9. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
10. Select “PRIME - Priming” button in the “Diluter” folder. Wait until the priming procedure has
been completed (Note: perform at least five successive cycles of the priming procedure, if
necessary, to fill the hydraulic circuit with the liquid).
11. Select the “Exit” button to exit from the Diagnostic program.
12. Turn off the VegaSys instrument.
13. Replace the chassis on the instrument and screw its four quarter-turn locks (Fig. 3).
To replace the new Diluter Home Sensor (Fig. 7) perform the following steps:
6. Replace the home sensor in its housing in the Diluter as showed in Fig. 6 and screw its two
fastening screws (Fig. 6).
7. Repeat the above steps in inverse order: from 4 to 2.
To replace the new Diluter Electrovalve (Fig. 4 – P/N: 05-01539-50) perform the following steps:
7. Replace the electrovalve in its housing on the Diluter mechanical support as showed in Fig. 4
and screw its two fastening screws (Fig. 4).
8. Repeat the above steps in inverse order: from 5 to 3.
9. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
10. Select “EV Test” button in the “Diluter” folder. Make sure that the Diluter Electrovalve is
correctly activated.
11. Select “PRIME - Priming” button in the “Diluter” folder. Wait until the priming procedure has
been completed (Note: perform at least five successive cycles of the priming procedure, if
necessary, to fill the hydraulic circuit with the liquid).
12. Select the “Exit” button to exit from the Diagnostic program.
13. Turn off the VegaSys instrument.
14. Replace the chassis on the instrument and screw its four quarter-turn locks (Fig. 3).
To replace the new Reel (Fig. 8 – P/N: 041032/1) perform the following steps:
6. Replace the Reel in its position on the instrument as showed in Fig. 8 and screw its fastening
screw (Fig. 8).
7. Replace the tube of the Reel in correct mode as showed in Fig. 4 and Fig. 8.
8. Repeat the above steps in inverse order: from 4 to 3.
Ref. 1 tube
Ref. 4 tube
Fitting Reel
Fastening screw
Diluter Electrovalve
9. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
10. Select “PRIME - Priming” button in the “Diluter” folder. Wait until the priming procedure
has been completed (Note: perform at least five successive cycles of the priming procedure, if
necessary, to fill the hydraulic circuit with the liquid).
11. Verify the absence of liquid leakage from the hydraulic circuit.
12. Verify the absence of air bubbles in the hydraulic circuit.
13. Select the “Exit” button to exit from the Diagnostic program.
14. Turn off the VegaSys instrument.
15. Replace the chassis on the instrument and screw its four quarter-turn locks (Fig. 3).
5.3 PHOTOMETER
5.3.1 Electronic adjustments of the Photometer
1. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
2. Unscrew the fastening screw of the Photometer cover (Fig. 9) and remove it from the
instrument.
3. Make sure that the voltage on the Lamp connector of the Photometer Lamp (Fig. 20) is equal to
6.0 Volts ± 0.1. If it not be so, turn the TR1 trimmer of the Analytical Control Board (compare
the scheme SC-30-02314-00) for adjusting the voltage.
4. Select “AUTOZERO” button in the “Photometer Test” folder (Fig. 10). Wait until the requested
operation has been completed.
Photometer cover
Fastening screw
5. Make sure that the mV value of the Dark, visuliazed in the “mV (Last)” column of the reading
table, is within voltage range indicated in the “Range (mV)” column and about 2 mV
(recommended value).
Ref. 6 tube
Ref. 5 tube
Photometer protection
6. Make sure that the mV value of the filter which transmits the highest signal, visuliazed in the
“mV (Last)” column of the reading table, is within voltage range indicated in the “Range
(mV)” column and about 500 mV (recommended value). If it not be so, performing the
following steps:
a. select the filter which transmits the highest signal in the “Select Filter” area (Fig. 10);
b. select the “Start Reading” button (Fig. 10);
c. read its voltage value showed in the diagnostic folder and turn the GAIN trimmer of
the “A/D C Board” (Fig. 12) in order to set the correct voltage value (recommended
value: 500 mV);
d. click the ”STOP Reading" button;
e. select the “Filters Reading” button (Fig. 10) and wait until the reading operation
has been completed;
f. repeat the above step 6.
7. Make sure that the mV value of each other’s filters, visuliazed in the “mV (Last)” column of
the reading table, is within relative voltage range indicated in the “Range (mV)” column and
shorter than voltage value of the filter which transmits the highest signal.
8. Select the “Exit” button to exit from the Diagnostic program.
9. Replace the Photometer cover on the instrument and screw its fastening screw (Fig. 9).
To replace the Photometer Assy (Fig. 14) perform the following steps:
8. Repeat the above steps in inverse order: from 7 to 2.
9. Perform the procedure “Electronic adjustments of the Photometer” described into the previous
Section 5.3.1.
Photometer
Micro-flow Cell
Fastening screw
Lamp wires
Fastening screw
Photometer protection
Fastening screw
Photometer
plane
Photometer
plane
1.5 mm
Filters Drum
Motor
Motor fastening screws
Fig. 15 – Filters Drum Motor pulley Fig. 16 – Drum Motor fastening screws
Filters Drum
Motor
Photometer plane
(bottom side)
Filters Drum
Home Sensor
To replace the new Filters Drum Motor (Fig. 17 – P/N: 05-02402-40) perform the following steps:
5. Insert the motor shaft in its housing of the Photometer plane as showed in Fig. 17, then screw
its two fastening screws as showed in Fig. 16.
6. Position the Filters Drum Motor pulley to 1.5 mm with respect to the upper side of the
Photometer plane (as showed in Fig.15), then tighten the Pulley set screw (Fig. 15).
7. Replace the Photometer protection on the Photometer and screw its fastening screw (Fig. 14).
8. Replace the Photometer Assy (Fig. 14) in its housing performing the relative procedure
described into the previous Section 5.3.2, from step 8 to step 9 (recommended step).
Filters Drum
Filters Drum
Fastening screw
Photometer plane
Filters Drum (upper side)
Home Sensor
Fig. 19 – Filters Drum Home Sensor
4. Unscrew the fastening screw of the Photometer protection (Fig. 11) and remove it from the
photometer.
5. Unplug the J6 connector from the Analytical Control Board (compare the scheme
SC-30-02314-00).
6. Position the Filters Drum as showed in Fig. 19.
7. Unscrew the fastening screw of the Filters Drum Home Sensor (Fig. 19) and take out it
from the upper side of the Photometer plane (Fig.19).
To replace the new Filters Drum Home Sensor (Fig. 19 – P/N: 9-10-0024-52) perform the
following steps:
8. Insert the home sensor in its housing as showed in Fig. 19, then screw its fastening screw
(Fig.19).
9. Repeat the above steps in inverse order: from 4 to 2.
10. Perform the procedure “Electronic adjustments of the Photometer” described into the
previous Section 5.3.1 (recommended procedure).
4. Unscrew the Lamp screw (Fig. 20) and take off the Photometer Lamp (Fig. 20) from its housing.
Photometer Lamp
Lamp connector
Lamp screw
Lamp wires
Fig. 20 – Photometer Lamp connector
To replace the new Photometer Halogen Lamp (Fig. 21 – P/N: 900154) perform the following
steps:
5. Insert the lamp in its housing as showed in Fig. 20, then screw the Lamp screw (Fig.20).
6. Repeat the above step 3 in inverse mode.
7. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
8. Perform the procedure “Electronic adjustments of the Photometer” described into the previous
Section 5.3.1, from step 3 to step 9.
Lamp wires
Halogen Lamp
Terminal wires
Case
fastening
screw
Board
fastening screw
A/D C Board
Board
fastening screw
To replace the new “A/D C Board” (Fig. 24 – P/N: 1-30-0001-00) perform the following steps:
7. Replace the board on the Micro-flow Cell case (as showed in Fig. 24) inserting the photodiode
in its hole of the case, then screw its two fastening screw (Fig. 24).
8. Repeat the above steps in inverse order: from 5 to 3.
9. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
10. Perform the procedure “Electronic adjustments of the Photometer” described into the previous
Section 5.3.1, from step 3 to step 9.
Dark
(position #1)
Filter n°2
(position #3)
Retaining ring
To replace the new Interferential Filters Kit (P/N: 65-02619-00), perform the following steps:
Do not touch the filters with your hands. Make use of tweezers to replace them.
5. For each filter position on the Filters Drum (Fig. 18 and Fig. 25) perform the following steps:
a. make sure that the filter wavelenght is right;
b. replace the filter in its seating, positioning the smooth side of the small filter box
on the outer side of the Filters Drum (Fig. 25);
c. insert (right to the end) the retaining ring in its seating on the small filter box
(as showed in Fig. 25), using the straight tweezers.
6. Replace the Photometer protection on the Photometer and screw its fastening screw (Fig. 11).
7. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
8. Perform the procedure “Electronic adjustments of the Photometer” described into the previous
Section 5.3.1, from step 3 to step 9.
Micro-flow Cell
connector
To replace the new Micro-flow Cell (Fig. 26 – P/N: 041008), perform the following steps:
6. Repeat the above steps in inverse order: from 5 to 3.
7. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
8. Perform the procedure “Electronic adjustments of the Photometer” described into the previous
Section 5.3.1, from step 3 to step 7.
9. Select “Cuvette Temperature” folder, then click on “START Reading” button. Make sure that
the temperature in the Micro-flow Cell is 37,0 °C ± 0,2 °C [note: the cooling and heating cell are
respectively indicated by the lighting of the LD2 (green Led) and LD3 (red Led) of the
Analytical Control Board as showed in the scheme SC-30-02314-00].
10. Click on “STOP Reading” button, then select the “Exit” button to exit from the Diagnostic
program.
11. Replace the Photometer cover on the instrument and screw its fastening screw (Fig. 9).
Before applying the following adjustment procedure, make sure that the Sampling and
Aspiration Arms are setted correctly (compare the above Sections 5.1.1 and 5.1.2).
1. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
2. Unscrew the fastening screw of the Photometer cover (Fig. 9) and remove it from the
instrument.
3. Put a drop of lubricating oil - with the P. P. sliding guide lowered - on the Ref. 8 tube, between
the P. P. Rotor and the P. P. sliding guide (Fig. 27).
Peristaltic Pump
Ref. 9 tube
Rotor Fitting
Ref. 8 tube
Peristaltic Pump
Electrovalve
Lock nut
Ref. 10 tube
Peristaltic Pump
Lock screw
sliding guide
Peristaltic Pump
Rotor
Peristaltic Pump
Electrovalve
Peristaltic Pump
Lock screw
sliding guide
4. Make sure that the stop plate nut is loosed and the stop plate (Fig. 30) is high-placed.
5. Draw up the Peristaltic Pump sliding guide to Ref. 8 tube (compare the Fig. 28), then screw its
lock screw (Fig. 28).
6. Insert a reagent bottle with Probe Rinse solution in the position #R1 of the Reagents Plate.
7. Make sure that a reaction rack is correctly inserted in the position #SR1 of the Analytical Plate.
8. Select “Aspiration” button in the “Peristaltic Pump” folder. Verify the free rotation of the
Peristaltic Pump Rotor during the execution of the requested aspiration procedure. If it not be
so, unscrew the lock screw (Fig. 28), draw up the Peristaltic Pump sliding guide to Ref. 8 tube
with less pressure, screw its lock screw, then repeat the step 8.
9. Insert the value “700” in the “Final Aspirat.” field, then select 5 cycles in the “Cycles (1..9)”
field of the “Peristaltic Pump” folder.
10. Select “Start Calibration” button in the “Peristaltic Pump” folder. Wait until the five calibration
procedures has been completed.
11. Make sure that the liquid (Probe Rinse solution) is situated - in every calibration procedure -
approximately between the point “A” of the Ref. 5 tube and the point “B” of the Ref. 6 tube – as
showed in Fig. 29 – in order to assure the Micro-flow Cell filling. If it not be so, perform the
following steps:
a. change the value in the “Final Aspirat.” field (note: increase the aspiration volume
increasing the value and vice versa);
b. repeat from the above step 10.
Liquid
A
Ref. 5 tube
Aspiration Arm
Ref. 6 tube
B Liquid
Liquid
Micro-flow Cell
Peristaltic Pump
Motor
Lock nut
Mechanical support
fastening screw
To replace the Peristaltic Pump (Fig. 32), perform the following steps:
11. Repeat the above steps in inverse order: from 10 to 2.
12. Perform the procedure “Adjustment of the Peristaltic Pump” described into the previous
Section 5.4.1.
Peristaltic Pump
sliding guide
Peristaltic Pump Electrovalve
Fig. 32 – Peristaltic Pump mechanical support
Peristaltic Pump
mechanical support
P. P. Motor screws
To replace the new Peristaltic Pump Motor (Fig. 31 – P/N: 05-02401-25), perform the following
steps:
5. Repeat the above steps in inverse order: from 4 to 3.
6. Replace the Peristaltic Pump in the instrument performing the relative procedure described into
the previous Section 5.4.2, from step 11 to step 12.
Ref. 8 tube
Fastening screws
Ref. 9 tube
Lock nut
Ref. 10 tube
To replace the new Peristaltic Pump Electrovalve (Fig. 34 – P/N: 05-01541-60), perform the
following steps:
9. Repeat the above steps in inverse order: from 8 to 3.
10. Turn on the VegaSys system (instrument and computer) and launch the Diagnostic program.
11. Select “EV Test” button in the “Peristaltic Pump” folder. Make sure that the P. P. Electrovalve
is correctly activated.
12. Select “Start Calibration” button in the “Peristaltic Pump” folder. Verify visually that the
hydraulic circuit functioning pertinent to the Micro-flow Cell is correct.
13. Wait until all calibration procedures has been completed.
14. Select “PRIME - Priming” button in the “Diluter” folder. Verify visually that the hydraulic
circuit functioning pertinent to the Washing Well is correct.
15. Wait until the priming procedure has been completed.
16. Select the “Exit” button to exit from the Diagnostic program.
17. Turn off the VegaSys instrument.
18. Replace the chassis on the instrument and screw its four quarter-turn locks (Fig. 3).
Peristaltic Pump
P. P. Motor screw Motor
To replace the Peristaltic Pump Rotor (Fig. 28 – P/N: 645006), perform the following steps:
5. Repeat the above steps in inverse order: from 4 to 2.
6. Perform the procedure “Adjustment of the Peristaltic Pump” described into the previous
Section 5.4.1.
Take-down the P. P. sliding guide (Fig. 27), take out the Peristaltic Pump Tube (Ref. 8 tube)
from the two fittings situated on its both terminal points, then take off it from the P. P. sliding
guide.
To replace the Peristaltic Pump Tube, repeat the above steps in inverse order and then
perform the procedure “Adjustment of the Peristaltic Pump” described into the previous
Section 5.4.1.
Support screw
5. Unscrew the four fastening screws of the cover of the Analytical Control Board and remove the
cover from its housing.
6. Unplug the J3 connector from the Analytical Control Board (compare the sch. SC-30-02314-00).
7. Unscrew the three screws of the Central Mechanical Support (Fig. 36), then – carefully – skew it.
8. Unscrew the two fastening screws of the Analytical Plate Stepper Motor (Fig. 38 and Fig. 46).
9. Remove the Analytical Plate Belt from the pulley of the Analytical Plate Stepper Motor
(Fig. 39) and take out the motor from the Central Mechanical Support (Fig. 39).
10. Loosen the two set screws of the Analytical Plate Motor pulley and extract the pulley from the
motor shaft (Fig. 40).
Reagents Plate
Analytical Plate
support
Special threaded
locking nut
Central Mechanical
Support
Fig. 37 – Analytical Plate (exploded view)
Analytical Plate
axis
Sampling Arm
axis
Motor
connectors
Central Mechanical
Support Analytical Plate
axis
Circular position
indicator
Analytical Plate
pulley Analytical Plate
Belt
Analytical Plate
Motor pulley Home Sensor
set screw Analytical Plate
Motor pulley
To replace the new Analytical Plate Stepper Motor (Fig. 40 – P/N: 05-02401-45) perform the
following steps:
11. Insert the pulley of the Analytical Plate Stepper Motor on the motor shaft (Fig. 40).
12. Position the pulley to 5.00 mm with respect to upper motor plane (Fig. 41), then tighten the two
set screws of the pulley (Fig. 41).
Analytical Plate
Stepper Motor
Fig. 40 – Analytical Plate Stepper Motor Fig. 41 – Analytical Plate Motor pulley
Flywheel
Sampling Arm
Home Sensor
Sampling Arm Stepper Motor
Central Mechanical
Support Motor connector
Flywheel
Set screw
Set screw
Central Mechanical
Support Sampling Arm Motor pulley
Motor connector
Set screw
Sampling Arm Motor pulley
Upper motor plane
Sampling Arm
Stepper Motor
Fig. 44 – Sampling Arm Stepper Motor Fig. 45 – Sampling Arm Motor pulley
7. Unscrew the four fastening screws of the cover of the Analytical Control Board and remove the
cover from its housing.
8. Unplug the J2 connector from the Analytical Control Board (compare the sch. SC-30-02314-00).
9. Unscrew the two fastening screws of the pulley protection (Fig. 42) and remove it from the
Central Mechanical Support.
10. Loosen the two set screws of the flywheel of the Sampling Arm Stepper Motor (Fig. 43).
11. Unscrew the two fastening screws of the Sampling Arm Stepper Motor (Fig. 38 and Fig. 43).
12. Remove the Sampling Arm Belt from the pulley of the Sampling Arm Stepper Motor (Fig. 42).
13. Extract the flywheel from the pulley of the Sampling Arm Stepper Motor (Fig. 42).
14. Take out the Sampling Arm Stepper Motor from the Central Mechanical Support (Fig. 43).
15. Loosen the two set screws of the Sampling Arm Motor pulley and extract the pulley from the
motor shaft (Fig. 44).
To replace the new Sampling Arm Stepper Motor (Fig. 44 – P/N: 05-02402-40) perform the
following steps:
16. Insert the pulley of the Sampling Arm Stepper Motor on the motor shaft (Fig. 44).
17. Position the pulley to 6.00 mm with respect to upper motor plane (Fig. 45), then tighten the two
set screws of the pulley (Fig. 45).
18. Repeat the above steps in inverse order: from 14 to 7.
19. Turn the Sampling Arm axis (Fig. 42) and make sure that the thoots of the belt are correctly
inserted on the Sampling Arm Motor pulley and on the Sampling Arm pulley (Fig. 42).
20. Replace the chassis on the instrument and screw its four quarter-turn locks (Fig. 3).
21. Replace the Sampling Arm on its axis and adjust it performing the previous Section 5.1.1, from
step 6 to step 29.
To replace the new Analytical Plate Belt (Fig. 39 – P/N: 1-35-0005-00) perform the following steps:
5. Wrap the new belt around the pulley of the Analytical Plate Stepper Motor and around the
Analytical Plate pulley (Fig. 39).
6. Repeat the above steps in inverse order: from 3 to 2.
7. Perform the replacement procedure of the Analytical Plate Stepper Motor described into the
previous Section 5.5.1, from step 14 to step 17.
To replace the new Sampling Arm Belt (Fig. 42 – P/N: 1-35-0005-00) perform the following steps:
6. Wrap the new belt around the pulley of the Sampling Arm Stepper Motor and around the
Sampling Arm pulley (Fig. 42).
7. Repeat the above steps in inverse order: from 4 to 2.
8. Perform the replacement procedure of the Sampling Arm Stepper Motor described into the
previous Section 5.5.2, from step 19 to step 21.
Analytical Plate
Stepper Motor
Fastening screw
Fastening screw
3. Unscrew the four fastening screws of the cover of the Analytical Control Board and remove the
cover from its housing.
4. Unplug the J8 connector from the Analytical Control Board (compare the sch. SC-30-02314-00).
5. Unscrew the three screws of the Central Mechanical Support (Fig. 36), then – carefully – skew it.
6. Unscrew the two fastening screws of the Analytical Plate Home Sensor (Fig. 46) placed on the
interior side of the Central Mechanical Support (Fig. 46), then take out the sensor from its
housing.
To replace the new Analytical Plate Home Sensor (Fig. 46 – P/N: 9-10-0024-52) perform the
following steps:
7. Position the new home sensor in its housing from interior side of the Central Mechanical
Support - as showed in Fig. 46 - in order to obtain that the circular position indicator is centered
with respect to the internal sides of the home sensor (Fig. 39).
8. Screw the two fastening screws of the Analytical Plate Home Sensor placed on the interior side
of the Central Mechanical Support (Fig. 46).
9. Repeat the above steps in inverse order: from 5 to 2.
10. Turn on the VegaSys system (instrument and computer) and launch the “Diagnostic” program.
11. Select “Sampling Arm/Plate Settings” folder and request an “General Reset” (Fig. 2). Make
sure that the requested reset procedure is correctly performed.
12. Wait until the reset instrument procedure has been completed.
13. Select the “Exit” button to exit from the Diagnostic program.
After applying the previous replacement procedure, make sure that the Sampling and
Aspiration Arms are setted correctly (compare the above Sections 5.1.1 and 5.1.2).
Fastening screw
3. Unscrew the four fastening screws of the cover of the Analytical Control Board and remove the
cover from its housing.
4. Unplug the J7 connector from the Analytical Control Board (compare the sch. SC-30-02314-00).
5. Unscrew the two fastening screws of the Sampling Arm Home Sensor (Fig. 42) placed on the
interior side of the Central Mechanical Support (Fig. 47), then take out the sensor from its
housing.
To replace the new Sampling Arm Home Sensor (Fig. 42 – P/N: 9-10-0024-52) perform the
following steps:
6. Position the new home sensor in its housing from interior side of the Central Mechanical
Support - as showed in Fig. 47 - in order to obtain that the position indicator sector is centered
with respect to the internal sides of the home sensor (Fig. 42).
7. Screw the two fastening screws of the Sampling Arm Home Sensor placed on the interior side
of the Central Mechanical Support (Fig. 47).
8. Repeat the above steps in inverse order: from 4 to 2.
9. Turn on the VegaSys system (instrument and computer) and launch the “Diagnostic” program.
10. Select “Sampling Arm/Plate Settings” folder and request an “General Reset” (Fig. 2). Make
sure that the requested reset procedure is correctly performed.
11. Wait until the reset instrument procedure has been completed.
12. Select the “Exit” button to exit from the Diagnostic program.
After applying the previous replacement procedure, make sure that the Sampling and
Aspiration Arms are setted correctly (compare the above Sections 5.1.1 and 5.1.2).
Keep plate
Fastening screws
Thermostat
Plate Resistor
Through hole
To replace the new Plate Resistor (Fig. 49 – P/N: 31-02410-00) perform the following steps:
13. Repeat the above steps in inverse order: from 12 to 6, then from 4 to 3.
14. Turn on the VegaSys system (instrument and computer) and launch the “Diagnostic” program.
15. Make sure that the Plate Resistor is turned on.
16. Make sure that the LD1 – yellow Led Plate Resistor on the Analytical Control Board (compare
the sch. SC-30-02314-00) is lighted (heating plate).
17. Wait at least 30 minutes to reach thermal equilibrium of the instrument.
18. Make sure that the LD1 – yellow Led Plate Resistor on the Analytical Control Board (compare
the sch. SC-30-02314-00) blinks (holding plate temperature).
19. Select “Plate Temperature” folder, then click on “START Reading” button. Make sure of the
thermal equilibrium of the instrument.
20. Click on “STOP Reading” button, then select the “Exit” button to exit from the Diagnostic
program.
21. Turn off the VegaSys instrument.
22. Replace the cover of the Analytical Control Board and screw its four fastening screws.
23. Replace the chassis on the instrument and screw its four quarter-turn locks (Fig. 3).
Plate Resistor
support
To replace the new Plate Temperature Sensor (Fig. 50 – P/N: 05-02400-40) perform the following
steps:
7. Repeat the above steps in inverse order: from 6 to 4.
8. Perform the procedure described into the previous Section 5.5.7, from step 14 to step 23.
Fastening screw
Faston Thermostat
Wire
Wire Faston
Plate Resistor
support Fastening screw
Fig. 51 – Thermostat
To replace the new Thermostat (Fig. 51 – P/N: 630077) perform the following steps:
6. Repeat the above steps in inverse order: from 5 to 3.
7. Unscrew the four fastening screws of the cover of the Analytical Control Board and remove the
cover from its housing.
8. Perform the procedure described into the previous Section 5.5.7, from step 14 to step 23.
Replacement of the:
Instrument Vain Fan Æ P/N: 05-01206-40
Board Vain Fan Æ P/N: 05-02403-30
Unscrew all fastening screws of the fan and unplug its connector from the Analytical Control
Board (compare the scheme SC-30-02314-00).
To replace the new fan, repeat the above steps in inverse order.
CHAPTER 06 - MAINTENANCE
INDEX
6. MAINTENANCE ............................................................................................................................ 2
6.1 PREVENTIVE MAINTENANCE ................................................................................................ 2
6.2 LIST OF PARTS SUBJECT TO WEAR AND USAGE .............................................................. 3
6.3 EXTERNAL SAMPLING/ASPIRATION PROBE CLEANING PROCEDURE....................... 3
6.4 WASH SOLUTION BOTTLE - CLEANING ............................................................................. 4
6.4.1 WASH SOLUTION BOTTLE CLEANING PROCEDURE ................................................ .....4
6.5 PROBE CLEANER PAD – REPLACING PROCEDURE.......................................................... 5
6.6 HYDRAULIC CIRCUIT WASHING.......................................................................................... 5
6.6.1 HYDRAULIC CIRCUIT CLEANING PROCEDURE ............................................................. 6
6.7 CHANGING THE PERISTALTIC PUMP TUBE....................................................................... 6
6.7.1 PROCEDURE FOR REPLACING THE PERISTALTIC PUMP TUBE .................................. 6
6.7.2 CALIBRATION PUMP PROCEDURE ................................................................................... 7
6.8 WASHING MICRO FLOW CELL - PROCEDURE................................................................... 8
6.9 CHANGING THE PHOTOMETER LAMP ................................................................................ 9
6.9.1 PROCEDURE FOR REPLACING THE PHOTOMETER LAMP ........................................... 9
6.10 CHANGING THE TUBES KIT................................................................................................ 11
6.10.1 PROCEDURE FoR REPLACING THE TUBES KIT ........................................................... 11
6.11 DECONTAMINATION PROCEDURE ................................................................................... 12
6. MAINTENANCE
This chapter describes the maintenance routine operations of the Vegasys. Maintenance activities,
should be carefully done in order to guarantee the proper working of the instrument over the time.
MAINTENANCE SCHEDULE
Note: the above-described maintenance schedule refers to a workload of approximately 500 tests
per day. The interval frequency may vary according to the instrument daily workload.
Description
Racks for Reaction Cells
Packs of Sample Cups of 0.8 ml
Reagent Containers
Plastic Bottles (H2O/Rinse & Waste)
Sampling Probe
Aspiration Probe
Tubing Kit Complete
Peristaltic Pump Tube
Probe Cleaner Pad
Photometer Lamp
Kit for E.V.Diluter connection
Diluter E.V.
Peristaltic Pump E.V.
Wash Solution
Diluter Head Teflon Fitting
Diluter Fitting
Note: The manufacturer suggests that once weekly the above – described cleaning procedure be
performed using, instead of only distilled water, a 5% sodium hypochlorite solution to dampen the
gauze and then be repeated using distilled water.
How often the bottle must be cleaned depends on its use and on the quality of the distilled water
used in that particular laboratory. However, the manufacturer recommends thorough washing at
least once every two weeks.
It is important not to underestimate the risks associated with mold and dust particles. They are to
be regarded as a serious hazard as they can be the cause of instrument malfunction.
Located at the base of the sampling arm, the probe cleaner consists of a number (approximately
16) of circular paper strips mounted on a holder. These clean the sampling probe before and after
the aspiration and dispensing of reagents.
How often the hydraulic circuit must be washed depends on the operating conditions/environment
of the single instrument and the quality of the distilled water used in that particular laboratory. The
manufacturer recommends thorough washing at least once a month.
It is extremely important that the user not underestimate the risks associated with mold and dust
particles. They are to be regarded as a serious hazard as they can be the cause of instrument
malfunction.
The hydraulic circuit input cannula is located inside the wash solution bottle.
1) Launch “Diagnostic” program, type the required Password to enter and wait for an
instrument general reset;
2) Press “Peristaltic Pump” button;
3) Press “Start Calibration” button; the instrument carries out a microflow cell wash cycle;
4) Make sure that the tubes leading to and from the microflow cell (see area circled) fill up of
liquid.
5) Referring to the tubes within the circled area in the figure:
- if the liquid goes too much beyond tube segment I, it means that there is too much liquid
- if the liquid does not completely fill the tube segment II, it is necessary to decrease the
steps in the “Final Aspiration” field. Push the “Start Calibration” button again and
check the liquid level in the tubes. On the other hand, if the liquid in segment II is too
much and in segment I there is not enough, increase the number of steps in the “Final
Aspiration” field. Push the “Start Calibration” button again and check the liquid level
in the tubes.
Note: If the adjustment cannot be made by adjusting the software, it is necessary to adjust the
pressure between the silicon tube and the peristaltic pump wheel through the keep plate of the
peristaltic pump.
WARNING! DO NOT TOUCH THE GLASS PART OF THE LAMP WITH YOUR FINGERS! IF
NECESSARY, USE A CLEAN CLOTH TO REMOVE DUST, OR ALCOHOL TO REMOVE
MORE STUBBORN DIRT.
The Vegasys will automatically sample and aspirate wash solution twice, performing the autozero
and displaying a chart which states the read value expressed in “mV” and “Counts” for each filter.
If any filter is out of range, before adjust the Gain and Offset, check if the microflow cell is:
perfectly clean
full of liquid and that there are no air bubbles on the inside.
Then follow procedure “Electronic Adjustment of the Photometer” Chapter 05 Par.5.3.1 of this
manual.
NECESSARY MATERIAL
an ESOFENOL solution diluted to 6% (60 cc in one liter of distilled water). ESOFENOL is an
antibacterial and antiviral substance.
Rubber gloves
Mask
Lab coat
CHAPTER 07
– HOST COMMUNICATION –
INDEX
7 HOST COMMUNICATION...........................................................................................................2
7 HOST COMMUNICATION
In order to enable communication between the Vegasys and the Host computer, select the “Host
Link” field under “Options” in the “Parameters” menu.
ÆASTM uses a number of different terms to indicate the way it groups data.
• Field: an individual piece of data often referred to as a data field or a data element.
• Record: a number of logically related data fields grouped together to form one part of a
complete message.
• Repeat field: a data field of the same type as the one immediately preceding it. A delimiter
separates one instance of a repeat field from the next.
• Component field: part of data field that might contain more than one piece of data.
The default communication configuration for the program is the following: "9600, N, 8, 1".
Æ ASTM uses record types that are common and familiar to all laboratory personnel. It uses the
following record types:
• Header Record (H): contains identifying information about the sending station,
conventions that the device uses for field recognition, and the date
and time of send station transmission.
• Patient Record (P): contains patient information and identification number.
• Test Order Record (O): contains information about the assay or requests themselves and
includes other data.
• Result Record (R): contains information about the outcome of individual tests for an
individual patient and follows a sample program record. The results
contain the actual measurements derived from the test and a
comparison of the individual result to certain ranges specified as
norms for the laboratory.
• Message Terminator Record (L): This is the last record in the message. A header record
may be transmitted after this record signifying the
start of a second message.
• Request Information Record (Q): is used by either clinical instruments or computer
systems for a remote request for information from its
reciprocal system.
• Comment Record (C): Comment records shall apply to any other record except the
message trailer record. They may be free standing messages sent
from the instrument, unrelated to a particular patient or test
procedure.
The following scheme shows the “Download” and “Upload” operations between the Vegasys
system and the Host computer:
Upload
Down Up Max
Field Field Title Description and Valid Values
Load Load Len
Patient
The Analyzer System will ignore this
10 Race/Ethnic I N
field at launch.
Origin
For Analyzer, this is a four- component
field:
• Address (25 characters)
11 Patient Address O S 60
• City (25 characters)
• State (2 characters e.g.: NY, IT)
• Zip (5 characters)
12 Reserved Field I N
Patient Tel
13 I N
Number
Attending
14 I N
Physician ID
15 Special Field 1 I N
16 Special Field 2 I N
17 Patient Height I N
18 Patient Weight I N
Patient Known
19 or Suspected I N
Diagnosis
Patient Active
20 I N
Medications
21 Patient's Diet I N
Practice Field
22 I N
No. 1
Practice Field
23 I N
No. 2
Admission Date
Admission date only.
24 and Discharge O S 8
Formatted as YYYYMMDD.
Date (if desired)
Admission
25 I N
Status
26 Location O S 20
Nature of
Alternative
27 Diagnostic I N
Code and
classifiers
Alternative
28 Diagnostic I N
Code and
classification
29 Patient Religion I N
30 Marital Status I N
31 Isolation Status I N
32 Language I N
Hospital
33 I N
Service
Hospital
34 I N
Institution
Dosage
35 I N
Category
Down Up Max
Field Field Title Description and Valid Values
Load Load Length
Request
7 Ordered I N
Date/Time
Specimen
8 Collected I N
Date/Time
Collection End
9 I N
Time
Collection
10 I N
Volume/Units
11 Collector ID I N
13 Danger Code I N
Relevant
14 I N
Clinical Info.
Date/Time
15 Specimen I N
Received
This is a numeric field indicating the
type of specimen:
Specimen The Imm. System uses the following
16 R A 1
Type ASCII characters:
0= Serum
1= Urine
Ordering
17 I N
Physician
Physician Tel.
18 I N
Number
User Field
19 I N
No. 1
User Field
20 I N
No. 2
Lab Field
21 I N
No. 1
Lab Field
22 I N
No. 2
Date /Time
Result
23 I N
Reported Last
or Modified
Instrument
24 I N
Charge
Instrument
25 I N
Section ID
The field indicates the direction of
26 Record Type 1 the transmission:
I A
O - Down Loading
F - Up Loading
27 Reserved Field I N
Location or
Ward of
28 I N
Specimen
Collection
Nosocomial
29 I N
Infection Flag
Specimen
30 I N
Service
Specimen
31 I N
Institution
R Required D Down Load
O Optional U Up Load
Legend:
I Ignored N Never
A Always S Sometimes
14 Instrument ID N
Down Up Max
Field Field Title Description and Valid Values
Load Load Len
CHAPTER 08
- TROUBLESHOOTING GUIDE -
THE SAMPLING PROBE DOES NOT The probe could be obstructed; plunge the probe.
ASPIRATE. If the diluter module works, check if there are
squeezed or torn tubing.
THE SAMPLING PROBE DOES NOT Check the valve of the diluter.
DISPENSE THE WASH SOLUTION Check if there is enough wash solution in the
INTO THE WASHING WELL bottle.
Check all tubing connection.
THE WASHING WELL IS NOT The tubing from the washing well to the electro-
EMPTIED EVEN IF THE valve may be torn or not properly connected.
PERISTALTIC PUMP WORKS Check the electro-valve (of the aspiration circuit).
PROPERLY Check the peristaltic pump tubing.
ALL AUTOZERO VALUES ARE OUT The micro-flow cell may be empty. Fill it with
OF RANGE water.
The peristaltic pump may not be working
properly. Check if some tubing need to be
changed.
Check the aspiration probe and its connection
with the micro-flow cell. The may be some
leakage or air bubbles in the micro-flow cell, or
the probe may be dirty.
The wash solution bottle may be empty.
Check the peristaltic pump electro-valve.
Adjust the aspiration volume; there may be too
much or to little that has been aspirated.
The micro-flow cell is dirty or air bubbles get
into the cell.
Lamp is nearing the end of its life cycle or
premature deterioration. In both cases, change the
lamp following the procedure described in
Chapter 06 – Maintenance.
THE RESULTS OF THE CONTROLS Make sure that the controls have been analyzed
ARE OUT OF RANGE with he methods specified by the manufacturer.
Make sure that the settings are correct for the
following parameters: wavelenght, temperature,
sampling volume, reagent volume, factor.
Make sure that the water used to prepare the
reagent and controls is either bi-distilled or
deionized water.
Check reagents, controls and standards Make sure
that they have been prepared in accordance with
the instructions given by the manufacturer.
Check that the recommended wash solution was
used and not simply distilled water.
KINETICS TESTS (PARTICULARLY The micro-flow cell may be contaminated. Clean
GOT OR GPT) GIVE LOW RESULTS. the micro-flow cell according to the procedure
described in Chapter 06 – Maintenance.
The reagent container may be contaminated.
The reagent may be expired or not fresh.
Check if the correct factor is used..
Check the volume settings.
Check the filter.
Check the temperature.