You are on page 1of 11

Chapter

Transcranial Doppler
19 Dean B. Andropoulos

Introduction sometimes, the filter needs to be turned off completely in cir-


cumstances of very low CBFV.
Transcranial Doppler (TCD) ultrasound is a sensitive, real-time
monitor of cerebral blood flow velocity and emboli that has
been used for neurologic monitoring during adult open cardiac Cerebral blood flow velocity measurement
surgery, vascular surgery, intracranial aneurysm surgery, and As discussed in Chapter 12, erythrocytes moving toward
congenital heart surgery. This chapter will review the techni- or away from the ultrasonic transducer are insonated; the
cal concepts of TCD, parameters monitored and clinical uses of
intraoperative TCD, evidence of utility for TCD, use to improve
neurological outcomes, risks and complications of TCD, and
finally, training and certification for TCD.

Technical considerations for transcranial


Doppler ultrasound
Currently available instruments use pulsed-wave ultrasound of
2 MHz frequency with a power of 100 mW, which is range-
gated with a sample volume length of up to 15 mm. As in
cardiac Doppler examination, TCD measures cerebral blood
flow velocity (CBFV), not absolute cerebral blood flow. Several
transducer probes are available, ranging from very small disc
Figure 19.1. 2-MHz probes for intraoperative monitoring, capable of attach-
probes suitable for infants and children, to larger, heavier disk ment to several external fixation devices. Courtesy of Compumedics DWL
probes with holders that attach to fixation devices for adoles- Doppler, Compumedics Ltd, www. dwl.de/.
cents and adults (Figure 19.1). Microvascular probes are also
available for the assessment of cerebral hemodynamics during
cerebral aneurysm surgery.1,2 These small probes have a diam-
eter of 1.5 mm or less and use frequencies of 16 or 20 MHz.
The probes are placed on the vessel of interest by the neuro-
surgeon using either a malleable wand or a rigid suction can-
nula. Intraoperative TCD monitoring usually requires fixation
of the ultrasonic probe with a specialized holder to prevent
probe movement, avoid compression of cranial or facial tissues,
and permit the surgeon to have access to cranial or neck regions
(Figures 19.2–19.5). These TCD and microvascular probes are
small in size and their design must incorporate shielding from
radiofrequency artifact from electrosurgical units.
High-pass filtering of TCD signals is necessary to mini-
mize vessel wall motion artifact. During surgery, extremely
low CBFV signals are often encountered, especially during
cardiopulmonary bypass with deep hypothermia. Filter set-
tings as low as 50 Hz may be required to visualize these low
Figure 19.2. Padded adjustable Doppler probe holder, with probe secured
CBFVs. TCD devices used during cardiac surgery must have over temporal window. Courtesy of Compumedics DWL Doppler, Com-
appropriate filter settings to accurately monitor low flow states; pumedics Ltd, www. dwl.de/.

226
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Chapter 19 – Transcranial Doppler

Figure 19.3. Padded probe fixation device in a child with probe secured
over the temporal window.

difference in frequency between the transmitted ultrasound and


its reflected waves (echoes) enables the calculation of erythro- Figure 19.5. Alternative monitoring site with hand-held probe secured over
cyte velocity using the Doppler equation. Because of laminar anterior fontanelle in a neonate, with sample volume directed at right middle
cerebral artery–anterior cerebral artery junction.
flow, these echoes create a family of distinct frequencies, and
spectral analysis permits presentation of the blood flow veloc-
ity as a function of time through the sample volume. The sig- be selected, which insonates only the arteries of interest, with-
nal power of each velocity component is determined and is out contamination from other sources. The peak CBFV can be
displayed as a color- or gray-scaled spectral profile. Through trended easily because of a favorable signal-to-noise ratio; how-
fast Fourier transformation (FFT), the spectral analyzer auto- ever, during surgery with bypass and nonpulsatile flow, mean
matically calculates CBFV parameters such as peak systolic, CBFV will be the only velocity parameter displayed. The direc-
end-diastolic, and intensity-weighted mean CBFVs (cm/sec), as tion of blood flow should be clearly labeled; the usual con-
well as pulsatility indices (PIs), which describe the downstream vention is to have flow toward the transducer displayed above
resistance. The PI is equal to the peak velocity minus the end- the zero velocity baseline, and flows away from the transducer
diastolic velocity, divided by the mean velocity (Figure 19.6). below the baseline.
The information can be stored digitally and analyzed offline at a
later time. As with cardiac ultrasound, the advantage of pulsed-
wave Doppler ultrasound is that a precise sample volume can

Figure 19.6. Transcranial Doppler cerebral blood flow velocity (CBFV) spec-
trum in an adult patient. Depth is depth of sample volume in mm, HITS is the
high-intensity transient signal (embolus) counter, sample is width of sample
volume in mm, mean is mean CBFV in cm/sec, PI is pulsatility index, equal
Figure 19.4. Neonatal/pediatric disk probe secured with clear adhesive to the peak velocity minus the end-diastolic velocity, divided by the mean
dressing over temporal window. velocity. Sys is systolic CBFV.

227
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Monitoring in Anesthesia and Perioperative Care

Posterior communicating artery Anterior choroidal artery Figure 19.7. Transcranial Doppler anatomy in an
Middle cerebral artery infant. The usual sample volume for intraoperative
Mesencephalic artery
Anterior communicating artery monitoring is the junction of the middle and anterior
Posterior choroidal artery
cerebral arteries. Reproduced with permission from
Anterior cerebral artery Diencephalic artery ref. 7.
Central retinal artery Posterior cerebellar artery
(remnant of hyaloid artery)
Superior cerebellar artery
Supraorbital artery Basilar artery
Frontal artery Internal auditory artery

Lacrimal artery Anterior inferior cerebellar artery

Anterior ethmoidal artery Posterior inferior cerebellar artery

Ophthalmic artery Anterior spinal artery


HYP Posterior auricular artery
Internal carotid artery
Posterior spinal artery
Infroorbital artery
Occipital artery
Middle meningeal artery Internal and external carotid arterial trunk
Superficial temporal artery Superior thyroid artery
Vertebral artery
Internal maxillary artery
Thyrocervical trunk
Inferior alveolar artery
Subclavian artery
External maxillary artery
Innominate Internal mammary artery
Lingual artery artery Aortic trunk

TCD measurements reflect the CBFV in the insonated ves- or asymmetric carotid stenosis or previous cerebral infarction,
sel in cm/sec, which is not synonymous with cerebral blood vascular reactivity in the two hemispheres may differ.5 In such
flow in cm3 /sec. Flow velocity is a function of pressure gra- cases, unilateral insonation may result in a misperception about
dient across the cerebral vascular bed, cerebral vascular resis- brain tissue at risk. This is also a significant issue in patients with
tance to flow, and diameter of the vessel being insonated. If intracranial vascular disease. Knowledge of the status of the
a vessel becomes narrowed in the region of insonation, the patient’s intracranial vasculature preoperatively may be helpful
CBFV through the vessel will increase, although total blood in guiding intraoperative monitoring.
flow decreases. In normal patients, cerebral autoregulation is The most consistent and reproducible technique for clini-
intact, so small decreases or increases in blood pressure cause cal use in patients of all ages is to monitor the middle cere-
compensatory vasodilation and vasoconstriction in the periph- bral artery (MCA) through the temporal window, which can
eral arterioles to maintain the total cerebral blood flow constant. usually be found just above the zygoma and just anterior to
CBFV depends on the diameter of the blood vessel, whereas the tragus of the ear (Figures 19.2–19.4).6 Several transducer
cerebral blood flow depends on cerebral vascular resistance, probes are available, ranging from very small disk probes suit-
which changes in response to changes in CO2 , temperature, able for infants and children, to larger, heavier probes for ado-
cerebral perfusion pressure, and pump flow. Thus, CBFV often lescents and adults. The depth of the sample volume and angle
correlates well with cerebral blood flow in the individual of insonation are adjusted until the bifurcation of the MCA
patient, particularly at deep hypothermia, when autoregulation and the anterior cerebral artery (ACA) are detected, ideally at a
is lost and the caliber of the blood vessels is unchanged. How- zero degree angle of insonation, which yields the most accurate
ever, the clinician must always be mindful and estimate the state CBFV information, just as in cardiac ultrasound (Figure 19.7).
of the patient’s cerebral vascular resistance to translate TCD into Correct position and sample volume adjustment are heralded
meaningful information for clinical decision making. by a maximal antegrade signal (positive deflection, toward the
Interpretation of CBFV changes occurring during surgery transducer) from the MCA, accompanied by retrograde flow as
or critical care is heavily influenced by the patient’s underly- the ACA is insonated (negative deflection, away from the trans-
ing pathology and the use of anesthetics and other vasoac- ducer). This spectrum should be the same or very similar veloc-
tive agents. Chronic hypertension, diabetes mellitus, cerebral ity and waveform as the MCA flow (Figure 19.8). The same loca-
atherosclerosis, and nicotine use may diminish or abolish cere- tion should be monitored for an individual patient.
bral pressure autoregulation,3 making both flow and flow veloc- Insonation at the MCA–ACA bifurcation also offers the
ity dependent on systemic arterial pressure. Cerebral pressure advantage of minimizing interpatient variability. In addition,
flow autoregulation is also blunted at moderate hypothermia the MCA supplies the largest volume of tissue of any of the
and abolished at deep hypothermia during cardiopulmonary basal cerebral arteries.7 After obtaining an optimal signal, the
bypass (CPB).4 These patients may also lack cerebral arterial probe must be secured, usually by adhesive tape or clear adhe-
reactivity to changes in PaCO2 . With these progressive dis- sive dressing for the small disk probe, or by adjustments to
orders, changes in blood flow may not be reflected in trends a padded head ring, or adhesive disk in larger patients. Care
in the same direction as CBFV. In addition, with unilateral must be taken with the padded head ring to thoroughly pad

228
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Chapter 19 – Transcranial Doppler

Figure 19.9. High-intensity transient signals (HITS). Note two deep signals
with a velocity well above the systolic velocity in an infant before cardiopul-
monary bypass (arrows).
Figure 19.8. Sample volume near the junction of the middle cerebral
artery (positive deflection, toward the transducer) and anterior cerebral artery
(negative deflection, away from the transducer) in an infant.
such as a large patent ductus arteriosus, will have an effect on
diastolic blood flow to the brain. These normal velocities were
all pressure points and to pay particular care to the orbits. For obtained in awake children and adults under perfect exami-
smaller patients, securing the probe by wrapping the head with nation conditions. Hemodynamic instability, less-than-optimal
an elastic bandage is discouraged because pressure sores may probe positioning, and general anesthesia may reduce these
develop under the area of the transducer. Also, adjustment to velocities in clinical practice. Often the clinician must accept
the probe position is often necessary during the case, so access a stable baseline for the individual patient and use it as the
to the transducer is important. In infants, an alternative site for basis for comparison, rather than expect a perfect signal. Typ-
monitoring is through the anterior fontanelle, using a hand- ically, the skull attenuates about 80 percent of the ultrasonic
held pencil-type probe, placing the probe over the lateral edge energy prior cerebral vessel insonation.8 Because of temporal
of the fontanelle, and aiming caudally, at a larger depth than for bone hyperostosis related to age, sex, and race, inadequate tem-
the temporal window, at the internal carotid artery (Figure 19.5) poral ultrasonic windows occur in a significant subpopulation
The depth of measurement and normal flow velocities for of patients.9,10 Identification of this situation prior to surgery
the MCA for infants, children, and adults are listed in Table 19.1. often enables the sonographer to use an alternative insonation
These normal velocities were determined in patients without site (i.e. foramenal or submandibular).
cerebrovascular or cardiovascular disease. In addition to the
problems in adults with cerebrovascular disease noted earlier,
congenital cardiac defects producing large diastolic runoffs, Cerebral emboli detection and counting
Cerebral emboli are very common during open-heart surgery
in adults and children. Emboli are easily detected by TCD,
Table 19.1. Normal transcranial Doppler values for infants, children, although artifacts, such as electrocautery and physical con-
and adults; temporal window
tact with the ultrasound transducer, may be observed (Figure
Peak systolic End-diastolic 19.9).11 True emboli have characteristic audio and visual signals
Mean velocity velocity velocity and are designated as high-intensity transient signals (HITS)
Age Depth (cm/sec) (cm/sec) (cm/sec) that can actually be counted by the TCD software. The filtering
0–3 mo 25 24–42 ± 10 46–75 ± 15 12–24 ± 8 criteria must be set to exclude artifacts, and the HITS counter
3–12 mo 30 74 ± 14 114 ± 20 46 ± 9
1–3 yr 35–45 85 ±10 124 ± 10 65 ± 11
can be a gauge of the number of emboli detected in the artery
3–6 yr 40–45 94 ± 10 147 ± 17 65 ± 9 being monitored.
6–10 yr 45–50 97 ± 9 143 ± 13 72 ± 9
10–18 yr 45–50 81 ± 11 129 ± 17 60 ± 8
⬎18 yr 30–65 62 ± 12 100 ± 20 45 ± 15 Uses of transcranial Doppler monitoring
Normal transcranial Doppler parameters in infants, children, and adults. Val-
ues were obtained in awake, normal patients without cardio- or cerebral
in the intraoperative setting
vascular pathology, breathing room air. For adults, a mean velocity ⬎80– TCD has widespread application in a number of clinical settings
100 cm/sec may signify significant stenosis or hyperemia. in the operating room, including assessment of intracranial
Reproduced with permission from refs. 6 and 54. hemodynamics, circle of Willis anatomy and function, ischemic

229
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Monitoring in Anesthesia and Perioperative Care

threshold and CBFV monitoring during carotid endarterec- In unanesthetized patients, clinical signs of cerebral hypo-
tomy, intracranial aneurysm repair, CPB, detection of cerebral perfusion appear with reductions in middle cerebral artery
emboli, and use in congenital heart surgery. flow velocity of greater than 60 percent.16 During carotid
endarterectomy, an exponential relationship between carotid
clamp-related decreases in middle cerebral artery mean CBFV
Assessment of intracranial hemodynamics and carotid artery stump pressure has been described. The
The specific objectives of intraoperative TCD monitoring need CBFV fell to undetectable levels at a stump pressure of 15
to be understood and discussed with the surgical team preop- mmHg.17 The authors concluded that TCD “provided an excel-
eratively. Again, TCD measures CBFV, and the use of CBFV lent indicator as to the necessity of shunting.” An additional
measurements as an estimate of cerebral blood flow is neither study found that patients with stump pressures below 30
appropriate nor recommended. In contrast, continuous CBFV mmHg had significantly lower CBFV than those with higher
recordings from a single insonation site can provide reliable pressures.18
and clinically valuable trend information. TCD monitoring can A multicenter retrospective study of carotid endarterectomy
identify changes in flow direction or detect sudden dramatic patients suggested that severe ischemia (i.e. high probability
CBFV change (i.e. relative hypo- or hyperperfusion). Change of new neurological deficit) was associated with a greater than
in the ratio of velocities at peak systole and end-diastole or PIs 85 percent reduction in CBFV, and moderate ischemia rep-
also can be used to estimate cerebral vascular resistance (i.e. resented a 60 percent to 85 percent decrease.19,20 Jorgensen
venous obstruction or intracranial hypertension).12 In certain and colleagues found that during carotid endarterectomy under
clinical situations, each of these pieces of information, which general anesthesia with combined TCD, cerebral blood flow,
are often unobtainable by other means, may prevent a neuro- and EEG monitoring, a greater than 60 percent decrease in
logic injury.13 Interpretation of sudden cerebral hemodynamic mean CBFV below the preocclusion reference resulted in flow of
changes requires the anesthesiologist to pay attention to maneu- approximately 20 mL/100 g/min and pathologic EEG suppres-
vers by the surgeon or perfusionist that can cause these changes. sion.21 Spencer22 suggested that CBFV reductions of greater
Thus, in the surgical environment, the anesthesiologist should than 70 percent persisting for more than 5 minutes signified
have immediate visual access to the surgical field to facilitate cerebral ischemia in need of immediate corrective action. In
rapid integration of surgical information. the absence of definitive criteria for the ischemic threshold, a
70 percent reduction in mean CBFV appears to be a reasonable
Circle of Willis function temporary guide.23 However, some patients have adequate lep-
Circle of Willis CBFV measurements are based on established tomeningeal collateral flow, and EEG activity occasionally may
criteria including cranial window, sample volume depth and remain unchanged in the presence of a severely decreased or
extension, spatial relationships of probe angle to intracra- absent MCA CBFV spectrum.
nial vessels, relative flow velocity, and response of oscillation
maneuvers.9,14 The most proximal segment of the middle cere- Intracranial aneurysm repair
bral artery (M1) is the most frequently used for TCD moni- Occlusion of an intracranial artery is often necessary for
toring applications because it carries up to 40 percent of the catheter or surgical treatment of a giant cerebral aneurysm, and
hemispheric blood flow. This segment generally has the high- testing the integrity of the circle of Willis and collateral arterial
est peak and intensity-weighted mean CBFV of any intracranial supply is important to plan the approach. TCD monitoring dur-
vessel, although there is a large variation (i.e. 35–90 cm/sec) in ing manual24 or intravascular balloon25 carotid artery occlusion
the awake adult patient.15 Correct identification of this segment can identify the lack of adequate alternative arterial supply to
is especially important when assessing collateral flow during the hemisphere of interest. The appearance of neurologic seque-
carotid occlusion. With carotid occlusion, the TCD signal from lae correlated with the magnitude of the relative CBFV decrease
the ipsilateral internal carotid artery disappears, but with nor- from individualized preocclusion reference, but not with the
mal intact circle of Willis function, CBFV is maintained in the absolute CBFV. In these studies, neurologic signs of transient
ipsilateral MCA segment. Typically, the MCA can be insonated focal deficit consistently occurred with CBFV decreases greater
at a depth of 50 mm in adults and can be traced back to the than 65 percent.
bifurcation to the carotid artery as the depth increases. With
bifurcation of the MCA into the ACA, flow direction is toward
the transducer in the MCA and away from the transducer for
Cardiopulmonary bypass
the ACA. TCD has been used to continuously assess changes in cerebral
hemodynamics during CPB. TCD indicates CBF presence and
direction and is able to detect an evolving ischemic process aris-
Carotid endarterectomy and the ischemic threshold ing from a malpositioned bypass cannula or inadvertent occlu-
An absolute CBFV threshold for ischemia has not been sion of the great vessels (Figure 19.10). During surgical repair of
established, and so common practice relies on changes rela- the aortic arch, TCD documents CBF direction during attempts
tive to the patient’s baseline to determine risk of ischemia. at selective antegrade or retrograde cerebral perfusion in both

230
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Chapter 19 – Transcranial Doppler

Two different approaches have been used in adult patients


for TCD ischemia detection during CPB. The first is simply
a modification of the relative ischemia criteria used for non-
CPB surgery, based on the view that larger decreases in CBFV
are needed to reliably identify hypoperfusion with extracorpo-
real circulatory support. In a retrospective study of coronary
artery bypass patients, Edmonds26 defined cerebral ischemia as
a mean CBFV decline of 80 percent below the preincision base-
line. Decreases of this magnitude occurred in 13 percent of the
cases and were associated with EEG suppression and/or cerebral
oxygen desaturation. Nearly all the decreases were correctable
by increasing mean arterial pressure, cardiopulmonary bypass
pump outflow, or arterial CO2 tension. The 3 percent neuro-
logic deficit incidence in this patient cohort was half that of a
historical control group that received no neuromonitoring.
The other approach to TCD monitoring during CPB is qual-
itative, with goals of maintaining a measurable CBFV spectral
Figure 19.10. Transcranial Doppler signal during cardiopulmonary bypass
in an adult patient. Note minimal pulsatility. Sudden decrease or cessation of
signal and verifying that the flow direction is as expected,27
flow, as may be seen with aortic cannula malposition, is rapidly detected. to prevent the infrequent but devastating injury that may be
associated with cannula malposition or other technical errors
with the conduct of CPB. This approach seems especially well
adults and children (Figure 19.11). Because peak CBFV changes suited for procedures that involve multimodality neuromoni-
in large basal cerebral arteries are usually related to peak CBF toring and the use of deep hypothermic circulatory arrest with
changes, TCD may aid in the determination of safe upper and selective cerebral perfusion. The effectiveness of the qualitative
lower limits for pump flow and perfusion pressure. In addi- method has been described for both adult28,29 and pediatric30
tion, end-diastolic CBFV change is inversely related to change cardiac surgery.
in cerebrovascular resistance. A sudden increase in resistance TCD has been used extensively in pediatric cardiac surgi-
during cardiopulmonary bypass may indicate impaired venous cal research to examine cerebral physiology in response to car-
return, possibly caused by a malpositioned or partially occluded diopulmonary bypass, hypothermia, low-flow bypass, regional
venous cannula. Persistence of the pattern following termina- low-flow perfusion to the brain, and circulatory arrest. Hillier
tion of bypass may suggest developing cerebral edema and the and associates31 used TCD to study cerebrovascular hemo-
need for its treatment. dynamics during hypothermic bypass with deep hypothermic
circulatory arrest (DHCA) in 10 infants. Cerebral blood flow
velocity did not return to baseline levels after DHCA. Calcu-
lated cerebral vascular resistance (mean arterial pressure – cen-
tral venous pressure/CBFV) was increased immediately after
DHCA and remained so until the end of bypass. The observed
decrease in CBFV during cooling was thought to be the result
of decreased metabolic demand by the brain and thus less blood
flow, although ␣-stat strategy was used. This could be explained
by relative cerebral vasoconstriction during cooling in smaller
arterioles downstream to the MCA and ACA, as these large
arteries do not change their caliber in response to changes in
PaCO2 .32 TCD of the MCA through the temporal window was
used to describe the cerebral pressure–flow velocity relation-
ship during hypothermic bypass in 25 neonates and infants
less than nine months old. CBFV was examined over a wide
range of cerebral perfusion pressure varying from 6 to 90 mm
Hg, at three temperatures – normothermia (36–37◦ C), moder-
ate hypothermia (23–25◦ C), and profound hypothermia (14–
20◦ C). Cerebral pressure flow autoregulation was preserved at
normothermia, partially affected at moderate hypothermia, and
totally lost at profound hypothermia, results that agree with pre-
Figure 19.11. Documentation of CBFV during antegrade cerebral perfu-
sion in a neonate. The very low flow rate is indicated by variation in CBFV in vious research done using xenon to quantitate cerebral blood
parallel with the roller head rate of rotation by the CPB pump. flow.4

231
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Monitoring in Anesthesia and Perioperative Care

TCD has also been used to determine the threshold bidirectionality and higher intensities (⬎25 dB above back-
of detectable cerebral perfusion during low-flow cardiopul- ground). In contrast, acoustic artifacts, although also bidirec-
monary bypass. Zimmerman and coworkers33 studied 28 tional, are predominantly of low frequency (⬍400 Hz). They
neonates undergoing the arterial switch operation with ␣-stat may arise from movement of the probe or probe cable, elec-
pH management. At 14◦ C to 15◦ C the pump flow was sequen- trical interference, and patient actions such as coughing, swal-
tially reduced to 0 mL/kg/min. All patients had detectable cere- lowing, talking, and facial movement. Recent development of
bral blood flow down to 20 mL/kg/min, whereas 1 had no per- multifrequency and multigated Doppler instruments has led to
fusion at 20 mL/kg/min, and 8 had none at 10 mL/kg/min, claims of improved sensitivity and specificity of emboli detec-
leading the authors to conclude that 30 mL/kg/min was tion. For example, Brucher and Russell41 reported 99 percent
the minimum acceptable flow in this population. Finally, correct classification of gaseous and particulate emboli using an
Andropoulos and associates34 used TCD of the MCA to deter- in vitro test system. However, the US Food and Drug Adminis-
mine the level of bypass flow necessary during regional low- tration has yet to clear for clinical use ultrasonographs equipped
flow perfusion for neonatal aortic arch reconstruction (Figure with this technology, and in clinical practice it is difficult to
19.11) They studied 34 neonates undergoing the Norwood oper- distinguish differences between solid and gaseous emboli using
ation or aortic arch advancement and established a baseline standard commercially available TCD systems.
mean CBFV under hypothermic full flow bypass (17–22◦ C, 150 An important limitation of the current generation of emboli
mL/kg/min) using pH stat management. During these condi- measurement systems is the inability of the automated features
tions, the mean CBFV was 22 cm/sec. The researchers then of the system to detect the most ominous situation, a mas-
used TCD to determine how much bypass flow was neces- sive gaseous embolization called an embolic shower or curtain.
sary to match this value, finding that a mean of 63 mL/kg/min Because all current emboli detection algorithms focus on the
was necessary. Interestingly, this necessary level of bypass flow discrete nature of individual transient embolic events, sustained
did not correlate with mean arterial pressure in the radial high-intensity signals are unrecognized; the anesthesiologist
artery or cerebral saturation measured by near-infrared spec- needs to be vigilant to recognize these events.
troscopy (NIRS). The necessary flow as determined by TCD HITS may be detected in as many as 60 percent of carotid
varied widely, leading the authors to conclude that TCD was a endarterectomies.42 In this setting, the emboliform ultrasonic
valuable monitor to ensure adequate but not excessive cerebral signals will be primarily of particulate origin, either thrombus
blood flow during this complicated technique. or atheroma. These emboli appear to be responsible for approx-
imately half the cerebrovascular complications associated with
this procedure.22 Payne and colleagues43 found that in the first
Detection and quantification of cerebral emboli hours after endarterectomy, neurologic injury often accom-
TCD is very sensitive to the presence of emboli, regardless panied HITS formation at a rate exceeding two per minute.
of their size or composition.35 Detection of embolization pro- Aggressive thrombolysis subsequently eliminated both HITS
vided by TCD can improve surgical and perfusion technique26 and the neurologic complications.
and facilitate correction of technical problems such as an air HITS composition during cardiac surgery is less certain
leak.36 Echogenic substances with acoustic impedance greater because of the frequent unintentional introduction of air into
than that of erythrocytes have been referred to as microemboli, the cerebral circulation during cardiopulmonary bypass and
microbubbles, particulate emboli, formed-element emboli, and the production of cavitation bubbles by mechanical prosthetic
Doppler microembolic signals (MES),37 whereas the term high- devices. Some studies have implicated HITS in the etiology
intensity transient signals (HITS) appears currently to be the of both neurologic injury and cognitive decline after cardiac
most widely used.38 surgery,16,44 whereas others have not.45 Although some inves-
The basic features of Doppler embolic signals were initially tigators have found the severity of the injury or decline to
defined by Spencer and colleages.17 Subsequently, an interna- be related to aggregate HITS,44,46 others found no relation-
tional consensus committee39 characterized the basic ultrasonic ship.47,48 No study has proposed a critical HITS injury thresh-
criteria of particulate (i.e. formed-element) HITS: old. There is retrospective evidence that HITS reduction is asso-
1. Transient, with a duration ⬍300 ms; ciated with a decreased incidence of cerebral injury.49 Despite
2. Duration dependent on passage time through the sample the association of aggregate HITS in some studies with neu-
volume; rocognitive injury, however, the relative impact of particulate
3. High-intensity, with an amplitude ⬎3 dB above and gaseous emboli is unclear.
background flow signal;
4. Unidirectional; and Outcome studies using transcranial Doppler
5. Acoustically resemble “snaps, tonal chirps, or moans.”
The majority of outcome studies assessing the use of TCD have
Gavrilescu and associates40 compared the ultrasonic signatures focused on the association between the number of detected
of particulate HITS to those presumably representing uncoated microemboli and later cognitive dysfunction or brain injury. In
bubbles. Distinguishing characteristics of the latter were their a review of the occurrence and clinical impact of microembolic

232
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Chapter 19 – Transcranial Doppler

signals during cardiac surgery, Dittrich and Ringelstein50 studies (or other study designs) performed that TCD monitor-
reviewed more than 50 studies in more than 1500 patients with ing will improve acute or longer-term neurologic outcomes.
TCD microemboli during or after coronary artery bypass graft In addition to limited evidence of utility for routine use in
(CABG), valve surgery, left ventricular assist device (LVAD) intraoperative monitoring, TCD is difficult to perform for long
placement, and catheter interventions. Prevalence of HITS in periods by the anesthesiologist, because of the frequent adjust-
these studies ranged from 0 to 100 percent of patients, with ments to the probe that are often necessary to maintain an opti-
most studies having a prevalence of 40 percent to 80 percent, mal signal. There is a significant learning curve for placement of
and number of HITS from one to several hundred per patient, the probe and fixation devices. These practical considerations
with most experiencing ⬍20 HITS. There were a few studies make this technique difficult to implement in multiple oper-
demonstrating a difference in number of HITS in those patients ating rooms simultaneously. A monitoring technician can be
with postoperative neurocognitive dysfunction or thromboem- used; however, this staffing may not be reimbursible by third-
bolic events, but the great majority of studies could detect no party payors, making TCD monitoring by a separate person
association. The authors’ conclusions were that the number of besides the anesthesiologist financially unsustainable in many
HITS can provide clues to successfully modify surgical inter- settings.
ventions, but TCD detection of HITS is not an accepted sur-
rogate parameter for the prediction or prevention of neuronal Complications of transcranial Doppler
damage with cognitive decline. They cited the limitations of
different monitoring devices, protocols, and monitoring time ultrasound
period as hampering the interpretation of the results of these Prolonged high-intensity ultrasound has potential risks to
trials. There is still no reliable method to distinguish between patients primarily from overheating of tissues in the path of the
gaseous and solid emboli, and there are no large prospective, ultrasound energy.53 The highest energy absorption coefficients
randomized trials with long-term neurological outcome stud- occur in bone. Because all the energy studies have been per-
ies to determine the utility of HITS detection by TCD. formed in vitro, the potential intracranial cooling effect asso-
In a recent systematic review of 14 cardiac surgery articles, 5 ciated with tissue perfusion is unknown. The maximum out-
carotid endarterectomy articles, and 2 orthopedic surgery arti- put of 100 mW/cm2 recommended by the American Institute
cles after screening for adequate study design and methods, of Ultrasound in Medicine has been adopted by all manufac-
Martin and associates51 determined that HITS occurred dur- turers of TCD devices cleared for use by the US Food and Drug
ing nearly all cardiac surgical procedures. Off-pump CABG was Administration, which should be well below any threshold for
associated with significantly fewer HITS (i.e. 30 or less off pump significant heating. The effects of continuous long-term expo-
vs several hundred with CPB). In both orthopedic surgery sure, such as that which may occur in the surgical/critical care
(range 0–40 HITS) and carotid endarterectomy (range 0–700 environments, are not fully understood. Therefore, the acous-
HITS), many patients had HITS; the numbers were lower than tic power and pulse amplitude should be kept to a minimum
with cardiac surgery with CPB (0–5260 HITS). The authors con- commensurate with the production of waveforms that can be
cluded that no consistent association between HITS and neu- accurately measured. This is particularly important with trans-
rocognitive outcome could be determined, and that the stud- orbital insonation, which is not recommended for intraopera-
ies were limited by differences in monitoring techniques, and tive or continuous critical care monitoring. Pressure sores or
in timing and type of neurocognitive tests. pressure injury to scalp, forehead, or orbits are possible if probes
In a careful study of HITS in pediatric cardiac surgery of and fixation devices are applied too tightly, without adequate
25 patients, O’Brien and associates11 noted that all patients had padding, or improperly, without checking for impingement on
HITS (range 2–2664), 42 percent of all emboli occurred within vital structures.
three minutes of the release of the aortic crossclamp, and there Sources of TCD artifact that give erroneous signals and
was no difference in the number of HITS for patients with and could result in inappropriate interventions include radiofre-
without obligate intracardiac shunting. The number of HITS quency interference from electrosurgical units from the imme-
was not associated with gross neurologic deficits postopera- diate or neighboring operating rooms; inadequate probe fix-
tively (Figure 19.12). ation, resulting in ambiguous, obscured, or evanescent CBFV
In retrospective reviews of adult and pediatric cardiac spectra; or mechanical contact of the probe by surgical instru-
surgery cases with CPB, Edmonds and co-workers26,52 deter- ments. No monitoring at all is preferable to potentially mislead-
mined that a multimodality neurologic monitoring strategy for ing information.
including NIRS, processed EEG, and TCD reduced acute post-
operative neurological changes (25% to 6% in children, 6% to Education, credentialing, and certification
3% in adults). However, TCD abnormalities were involved in
only 5 percent to 10 percent of these patients, and often the in transcranial Doppler ultrasound
other modalities, such as NIRS, were abnormal during abnor- There is currently no credentialing by any professional anes-
mal TCD conditions as well. Taken together, there is insufficient thesiology or critical care organization. The American Soci-
evidence from the few randomized, prospective, controlled ety of Neurophysiologic Monitoring (ASNM, www.ASNM.org)

233
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Monitoring in Anesthesia and Perioperative Care

A Figure 19.12 (a) Number of emboli detected dur-


ing 13 phases of cardiac surgery with bypass in 25
pediatric patients. Each symbol denotes one patient.
1000 Numbers in parentheses are numbers of patients
experiencing no emboli at each phase. The circled
solid diamond symbol indicates median number
of emboli at each time period. (b) The time interval
associated with the greatest number of emboli in
Number of emboli

100
each of 25 patients. The greatest number of emboli
were apparent either after skin incision or ⬎3 min
after the aortic clamp was released. Reproduced
with permission from ref. 11.
10

(4) (14) (11) (12) (17) (17) (11) (9) (3) (16) (17) (20) (18)
0

Total emboli
Postincision

Cannulation

Postcannulation

On CPB

Post on CPB

Aorta clamped

Misc. bypass

Aorta unclamped

Post unclamping

Off CPB

Post off CPB

Decannulation

Postdecannulation
B
12
11
10
9
Number of patients

8
7
6
5
4
3
2
1
Postcannula

Post decannula
Aorta clamped

Aorta unclamp

Decannulation
Post unclamp
Post on CPB

Misc. bypass

Post off CPB


Postincision

Cannula

On CPB

Off CPB

recommends certification by the American Board of Neuro- Training for physician proficiency in interpretation of cere-
physiologic Monitoring (ABNM), or its equivalent, as a measure brovascular ultrasound studies, with particular emphasis on
of professional level qualification in multimodality neuromon- applied principles of ultrasound physics, fluid dynamics, and
itoring. Criteria for ABNM certification include1 an advanced various aspects of TCD examination, is offered by the Ameri-
degree – master’s, Ph.D., M.D., or D.O.;2 documented clinical can Society of Neuroimaging (ASN; www.asnweb.org). Never-
experience, with the requirement of at least 300 monitored cases theless, certification of technical sonographic expertise in TCD
over a minimum of three years;3 surgeon attestations regarding evaluation of intracranial vessels is currently not available in
monitoring experience;4 the passing of two examinations, one the United States. At the beginning of 2006, the ASN board of
written and the other oral. These examinations focus on elec- directors approved creation of the first national examination to
trophysiological techniques, and neither covers TCD monitor- assess a technologist’s ability to apply knowledge, concepts, and
ing in any detail nor certifies competence in the production or principles of neurovascular ultrasound that constitute the basis
interpretation of ultrasonic waveforms. of safe and effective patient care. It is designed to measure the

234
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Chapter 19 – Transcranial Doppler

candidate’s application of medical knowledge and understand- 8. Grolimund, P. Transmission of ultrasound through the temporal
ing of biomedical and clinical sciences that are considered to bone. In Aaslid R (ed.), Transcranial Doppler Sonography. New
be essential for the unsupervised performance of neurovascu- York: Springer-Verlag, 1986, pp. 10–21.
9. Saver JL, Feldman F. Basic transcranial Doppler examination:
lar ultrasound.
technique and anatomy. In Babikian VL, Wechsler LR (eds.).
Transcranial Doppler Ultrasonography. St. Louis: Mosby, 1993,
pp. 11–28.
Practice parameters for transcranial 10. Bass A, Krupski WC, Schneider PA, et al. Intraoperative
Doppler ultrasound transcranial Doppler: limitations of the method. Vasc Surg
1989:10;549–53.
Currently there are no practice parameters establishing recom- 11. O’Brien JJ, Butterworth J, Hammon JW, et al. Cerebral emboli
mendations to use TCD for intraoperative monitoring, nor can during cardiac surgery in children. Anesthesiology
it be considered the standard of care for any type of surgical 1997;87:1063–9.
procedure. The ASNM and ASN state, “On the basis of current 12. Hassler W. Steinmetz H, Gawlowski J. Transcranial Doppler
clinical literature and scientific evidence, TCD monitoring is an ultrasonography in raised intracranial pressure and in
intracranial circulatory arrest. J Neurosurg 1988;68:745–
established monitoring modality for the: 1) assessment of cere-
50.
bral vasomotor reactivity and autoregulation; 2) documentation 13. Guerra WK, Gaab MR, Dietz H, et al. Surgical decompression for
of the circle of Willis functional status; 3) identification of rela- traumatic brain swelling: indications and results. J Neurosurg
tive cerebral hypo- and hyperperfusion; 4) detection of cerebral 1999;90:187–96.
emboli (Class II and III evidence, Type B recommendation).”55 14. Fujioika K, Douville CM. Anatomy and free-hand examination
techniques. In Newell DW, Aaslid R (eds.). Transcranial Doppler.
New York: Raven Press, 1992, pp. 9–31.
15. Adams RJ, Nichols FT, Hess DC. Normal values and
Conclusions physiological variables. In Newell DW, Aaslid R (eds.).
TCD has been an important research tool for cardiac and vas- Transcranial Doppler. New York: Raven Press, 1992, pp. 41–48.
cular surgery and may have utility for specialized CPB tech- 16. Edmonds HL, Pollock SB, Thomas MH. Transcranial Doppler
niques, such as antegrade or retrograde cerebral perfusion, in guides emboli reduction during myocardial revascularization.
both adult and pediatric patients. In addition, TCD will detect Stroke 1998;29:2237.
intraoperative cerebral emboli. However, lack of outcome bene- 17. Spencer MP, Thomas GI, Moehring MA. Relation between
middle cerebral artery blood flow-velocity and sump pressure
fit, and the technical difficulties with obtaining and maintaining during carotid endarterectomy. Stroke 1992;23:1439–45.
a valid TCD signal in the intraoperative setting limit its utility 18. Kalra M, al-Khaffaf H, Farell A, et al. Comparison of
for routine use by the anesthesiologist. measurement of stump pressure and transcranial measurement of
flow-velocity in the middle cerebral artery in carotid surgery. Ann
Vasc Surg 1994;8:225–31.
References 19. Halsey J. Risks and benefits of shunting in carotid
1. Bailes JE, Lokesh ST, Fukushima T, et al. Intraoperative endarterectomy. The International Transcranial Doppler
microvascular Doppler sonography in aneurysm surgery. Collaborators. Stroke 1992;23:1583–7.
Neurosurgery 1997;40:965–72. 20. Halsey J. Monitoring blood flow velocity in the middle cerebral
2. Edmonds HL, Isley MR, Sloan T, et al. American Society of artery during carotid endarterectomy. In Babikian VL, Wechsler
Neurophysiologic Monitoring (ASNM) and American Society of LR (eds.). Transcranial Doppler Ultrasonography. St. Louis:
Neuroimaging (ASN) Joint Guidelines for Transcranial Doppler Mosby, 1993, pp. 216–21.
(TCD) Ultrasonic Monitoring. www.ASNM.org, January 25, 21. Jorgensen LG. Transcranial Doppler ultrasound for cerebral
2009. perfusion. Acta Physiologica Scand 1995;625 [supplement]:
3. Muller M, Schimrigk K. Vasomotor reactivity and pattern of 1–44.
collateral blood flow in severe occlusive carotid artery disease. 22. Spencer MP. Transcranial Doppler monitoring and causes of
Stroke 1996:27;296–99. stroke from carotid endarterectomy. Stroke 1997;28;685–91.
4. Greeley WJ, Ungerleider RM, Kern FH, et al. Effects of 23. Babikian VL, Cantelmo NL, Wijman CA. Neurovascular
cardiopulmonary bypass on cerebral blood flow in neonates, monitoring during carotid endarterectomy. In Babikian VL,
infants, and children. Circulation 1989;80:I209–15. Wechsler LR (eds.), Transcranial Doppler Ultrasonography, 2nd
5. Eames PJ, Blake MJ, Dawson SL, et al. Dynamic cerebral ed. Boston: Butterworth-Heinemann, 1999, pp. 231–45.
autoregulation and beat to beat blood pressure control are 24. Giller CA Mathews D, Walker B, et al. Prediction of tolerance to
impaired in acute ischaemic stroke. J Neurol Neurosurg Psychiatry carotid artery occlusion using transcranial Doppler ultrasound.
2002;72:467–72. J Neurosurg 1994;81:15–19.
6. Fischer AQ, Truemper EJ. Applications in the neonate and child. 25. Schneweis S, Urbach H, Solymosi L, et al. Preoperative risk
In Babikian VL, Wechlser LR (eds.). Transcranial Doppler assessment for carotid occlusion by transcranial Doppler
Ultrasonography, 2nd ed. Oxford: Butterworth-Heineman, 1993, ultrasound. J Neurol Neurosurg Psychiatry 1997;62:485–9.
pp. 355–75. 26. Edmonds HL. Protective effect of neuromonitoring during
7. Truemper EJ, Fischer AQ. Cerebrovascular developmental cardiac surgery. Ann NY Acad Sci 2005;1053:12–19.
anatomy and physiology in the infant and child. In Babikian VL, 27. Edmonds HL, Rodriguez RA, Audenaert SM, et al. The role of
Wechlser LR (eds.). Transcranial Doppler Ultrasonography, 2nd neuromonitoring in cardiovascular surgery. J Cardiothorac Vasc
ed. Oxford: Butterworth-Heineman, 1993, pp. 281–320. Anesth 1996;10:15–23.

235
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021
Monitoring in Anesthesia and Perioperative Care

28. Ganzel BL, Edmonds HL, Pank JR, et al. Neurophysiological 42. Jansen C, Vriens EM, Eikelboom BC, et al. Carotid
monitoring to assure delivery of retrograde cerebral perfusion. endarterectomy with transcranial Doppler and
J Thorac Cardiovasc Surg 1997;113:748–57. electroencephalographic monitoring: a prospective study in 130
29. Estrera AL, Garami Z, Millar CC, et al. Determination of cerebral operations. Stroke 1993;24:665–9.
blood flow dynamics during retrograde cerebral perfusion using 43. Payne DA, Jones CI, Hayes PD, et al. Beneficial effects of
power M-mode transcranial Doppler. Ann Thorac Surg clopidogrel combined with aspirin in reducing cerebral emboli in
2003;76:704–9. patients undergoing carotid endarterectomy. Circulation
30. Andropoulos DB, Stayer SA, Diaz LK, et al. Neurological 2004;109:1476–81.
monitoring for congenital heart surgery. Anesth Analg 44. Stump DA, Kon NA, Rogers AT, et al. Emboli and
2004;99:1365–75. neuropsychological outcome following cardiopulmonary bypass.
31. Hillier SC, Burrows FA, Bissonnette B, Taylor RH. Cerebral Echocardiography 1996;13:555–8.
hemodynamics in neonates and infants undergoing 45. Georgiadis D, Kaps M, Berg J, et al. Transcranial Doppler
cardiopulmonary bypass and profound hypothermic circulatory detection of microemboli in prosthetic heart valve patients:
arrest: assessment by transcranial Doppler sonography. Anesth dependency upon valve type. Eur J Cardiothorac Surg
Analg 1991;72:723–8. 1996;10:253–7.
32. Huber P, Handa J. Effect of contrast material, hypercapnia, 46. Clark RE, Brillman J, Davis DA, et al. Microemboli during
hyperventilation, hypertonic glucose and papaverine on the coronary artery bypass grafting: genesis and effect on outcome.
diameter of the cerebral arteries: Angiographic determination in J Thorac Cardiovasc Surg 1995;109:249–58.
man. Invest Radiol 1967;2:17–32. 47. Lund C, Hol PK, Lundblad R, et al. Comparison of cerebral
33. Zimmerman AA, Burrows FA, Jonas RA, Hickey PR. The limits of embolization during off-pump and on-pump coronary artery
detectable cerebral perfusion by transcranial Doppler sonography bypass surgery. Ann Thorac Surg 2003;76:765–70.
in neonates undergoing deep hypothermic low-flow 48. Ferrari J, Baumgartner H, Tentschert S, et al. Cerebral
cardiopulmonary bypass. J Thorac Cardiovasc Surg microembolism during transcatheter closure of patent foramen
1997;114:594–600. ovale. J. Neurol 2004;251:825–9.
34. Andropoulos DB, Stayer SA, McKenzie ED, Fraser CD Jr. Novel 49. Kaposzta Z, Martin JF, Markus HS. Switching off embolization
cerebral physiologic monitoring to guide low-flow cerebral from symptomatic carotid plaque using S-nitrosoglutathione.
perfusion during neonatal aortic arch reconstruction. J Thorac Circulation 2002;105:1480–4.
Cardiovasc Surg 2003;125:491–9. 50. Dittrich R, Ringelstein B. Occurrence and clinical impact of
35. Mullges W, Berg D, Babin-Ebell J et al. Cerebral microembolus microembolic signals during or after cardiosurgical procedures.
generation in different extracorporeal circulation systems. Stroke 2008;39:503–11.
Cerebrovasc Dis 1999;9:265–9. 51. Martin KK, Wigginton JB, Babikian VL, et al. Intraoperative
36. Yeh TJ, Austin EH III, Sehic A, Edmonds HL Jr. Role of cerebral high-intensity transient signals and postoperative
neuromonitoring in the detection and correction of cerebral air cognitive function: a systematic review. Am J Surg
embolism. J Thorac Cardiovasc Surg 2003;126:589–91. 2009;197:55–63.
37. Spencer MP. Detection of embolism with Doppler ultrasound: a 52. Austin EH III, Edmonds HL Jr, Auden SM. Benefit of
review. Echocardiography 1996;13:519–27. neurophysiologic monitoring for pediatric cardiac surgery.
38. Hennerici MG. High intensity transcranial signals (HITS): a J Thorac Cardiovasc Surg 1997;114:707–17.
questionable “jackpot” for the prediction of stroke risk. J Heart 53. Nyborg WL. Optimization of exposure conditions for medical
Valve Dis 1994;3:124–5. ultrasound. Ultrasound Med Biol 1985;11:245–60.
39. Consensus Committee, 9th International Cerebral 54. Lao A, Sharma VK, Katz ML, Alexandrov AV. Diagnostic criteria
Hemodynamics Symposium. Basic identification criteria for for transcranial doppler ultrasound. In McGahan JP, Goldberg
Doppler microembolic signals. Stroke 1995;26:1123. BB (eds.). Diagnostic Ultrasound, 2nd ed. New York: Informa
40. Gavrilescu T, Babikian VL, Cantelmo NL, et al. Cerebral Healthcare, 2008, pp. 543–5.
microembolism during carotid endarterectomy. Am J Surg 55. Edmonds HL Jr, Isley MR, Sloan TB, Alexandrov AV,
1995;170:159–64. Razumovsky AY. American Society of Neurophysiologic
41. Brucher R, Russell D. Automatic online embolus detection and Monitoring and American Society of Neuroimaging Joint
artifact rejection with the first multi-frequency transcranial Guidelines for Transcranial Doppler Ultrasonic Monitoring. J
Doppler. Stroke 2002;33:1969–74. Neuroimaging. 2010 Mar 17. [Epub ahead of print].

236
Downloaded from https:/www.cambridge.org/core. The Librarian-Seeley Historical Library, on 15 Mar 2017 at 01:41:29, subject to the Cambridge Core terms of use, available at
https:/www.cambridge.org/core/terms. https://doi.org/10.1017/CBO9780511974083.021

You might also like