ISOLAB
Excellence In Calibration
|
CERTIFICATE OF CALIBRATION
No: CS191035
Date of Issue: 13" March 2019
sue By: ISOLAB (SINGAPORE) PTE LTD
Requested by Barghest Building Performance Pte Lid
150 Cantonment Road, #01-06
yeapore 089762
‘Acez Sensing Job No EIS 9472
Job Order No 190421
Ambient Temperature (22=2)°C
Relative Humidity 10) % Relative Humidity
Page
Descriptions
Instrument 2 10K Thermistors
Brand AS
Model NIA
Serial No EP 108202
Immersion 200 mm
Calibration Range 2 001°C to 29.765°C
Calibration Start Date 11" March 2019
Calibration End Date 11" March 2019
Isolab ($) Pte Ltd and its practices are in compliance with ISO/IEC 17025: 2005. The Quality System practice
is in accordance with the Quality Standard ISO 9001
Method of Calibration
The 10K Thermistor has been calibrated at ISOLAB’s laboratory under the ambient conditions stated above
according to in-house calibration procedure STCP-003-R4 and STCP-016-R3. The references used are traceable
to national measurement standards maintained at National Metrology Centre (NMC-Singapore), National
Physical Laboratory (NPL-UK) and Northern Temperature Primary Laboratory (NTPL, UKAS). The
calibration was performed in accordance to International Temperature Scale of 1990 (ITS-90),
Reference Standards used for Calibration:
SIN: Reference Standards Serial No, Reference Report No: Next Calibration Due Date
1. Reference Sensor 3698172 C118083-S 20 August 2019
2. Reference Sensor 36482/2 C118081-s, 20 August 2019
3. Reference Indicator My44037443 EL006857 26" Suly 2019
4. Reference Indicator 09-P123 291092/1 18-08-41 14% August 2019
5. Gallium Melting Point Cell GASOS 12-03-50 12" March, 2022
Calibrated By: Approved By:
Approved Signatory
* “The results reported herein have been performed in accordance with the laboratory's terms of accreditation under the
Singapore Accreditation Council”.
‘This certicaydmgyiot be reproduced other than in fall except with prior writen approval of SOLAB,
jaenex “ccrcorco ISOLAB (Singapore) Pte Ltd wuemec/iasca b (OEE
MSIL AS 2-400 Koon Circle Singapore 629031 3
‘extn Tal (65) 626.6155. Fax: (65) 62666824 Email: service@isolabcom sg ‘wwwisolab.comsgCERTIFICATE OF CALIBRATION
No: CS191035
Date of Issue: 13" March 2019
10K Thermistor
Serial Number: EP 108202 /
Page 20f3
Calibration Start Date 11" March 2019
Calibration End Date 11 March 2019
Results of Calibration
The results of the calibration are shown in the table below. The expanded uncertainty is at a level of confidence
of approximately 95% with a coverage factor.
Calibration And Measurement Capability (CMC*):
(). Gallium Melting Point Q) @0°C 10 30°C: 0.011°C
(29.7646°C) :0.003°C
Calibration Results (As Found)
‘Actual’ Temperature TOK Thermistor Uncertainty ieFactor
co) Indicated Values (2) (CCormK)
0.01 32.6321 12 mK 2.00
15.00 15.7129 13 mK 2.00
29.165 8.1365 3 mK 2.00
* “The results reported herein have been performed in accordance with the laboratory's terms of accreditation
under the Singapore Accreditation Council”
Remarks
The interpolation Equation for the “Thermistor” is:
1
=A+B(inR)+C(nR)
T
Where; = T = Temperature in Kelvin
Resistance values in Ohms at Temperature t
In, = Natural logarithm of Resistance values in Ohms
A,B, C, Coefficients Derived From Measurement
A= | 1.126774E-03
B 2.344833E-04
C__= | 8.653168E-08CERTIFICATE OF CALIBRATION
No: CS191035
Date of Issue: 13 March 2019
10K Thermistor
Page
Calibration Start Date
Calibration End Date
Serial Number: EP 108202
30f3
11" March 2019
11" March 2019
Appendix
Temperature Resistance Table for 10K Thermistors
Temperature | Resistance | Temperature | Resistance
eC) Value (A) eo) Value (0)
0 32642.0973 16 14997.5098
1 310223778 7 143213472
2 294923898 18 13679.4490
3 28046.6822 19 13069,9022
4 26680.1624 20 12490.9096
3 25388.0713 21 11940.7827
6 24165.9595 2 11417.9345
7 23009.6660 2 10920.8727
8 219152979 24 10448. 1941
9 20879.2120 23 998.5783
10 19897.9973 26 9570.7830
i 18968.4597 2 9163.6385
2 18087.6066 28 876.0436
B 17252.6335 29 8406.9608
14 164609113 30 8055.4128
15 15709.9744 : =
Calibrayion OF