Professional Documents
Culture Documents
VM-742B & VM-742 Network Communication Module Manual
VM-742B & VM-742 Network Communication Module Manual
1
Published in Nov. 2011
Revised in Jan. 2012
● This manual is solely designed for technicians to set, maintain, check for the
VM-7 Monitoring System and deal with trouble on the monitoring system.
● Read this manual thoroughly and understand the contents before setting,
maintaining, checking and trouble-shooting for VM-7 Monitoring System.
● Keep this manual so that the operator can refer to it when needed.
Thank you very much for purchasing the VM-742B & VM-742 Network Communication Module (hereafter
“the product”).
Name VM-703B & VM-703 Tachometer & Eccentricity Monitor Module Manual
Intended users Personnel who perform the detail settings and maintenance of the product
of the manual The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Contents • How to set the product
• How to adjust the product
• Troubleshooting
1
VM-742B & VM-742 Preface
Name VM-741B & VM-741 Local Communication & Phase Marker Module Manual
Intended users Personnel who perform the detail settings and maintenance of the product
of the manual The use of this manual is restricted to those who have completed the maintenance training
course run by our company.
Contents • How to set the product
• Description of communication specifications
• Troubleshooting
2
VM-742B & VM-742 Preface
3
VM-742B & VM-742 Preface
<Important Notices>
● Be sure to read this manual thoroughly and understand the contents before using this product. Keep this
manual so that the operator can refer to it when needed.
● If you have any inquiry about this product or manual, contact the sales office where you purchased the
product.
● For the use of the product, adequate knowledge about other instruments connected to the product and
transducers is also required. In addition, not only the contents of this manual, but also common sense
need to be applied for the safe use of the product.
● This product has had its EMC (Electro-magnetic compatibility) evaluated by various noise tests.
However, since the performance of the product varies depending on the environmental conditions such
as below, follow instructions in this manual to stabilize the product.
– Intensity of noise
– Temperature and humidity
– Layout of the product
● When reselling the product as an assembly or transferring the product (to a third party), please provide
the end users with this manual.
● Company and product names contained herein are trademarks or registered trademarks of their
respective owners.
● Copyright is retained by Shinkawa Sensor Technology, Inc. No part of this document may be disclosed
or reproduced in any form without prior consent of SHINKAWA Sensor Technology, Inc.
4
VM-742B & VM-742 Preface
<Disclaimer>
● The company assumes no liability for any damage or injury caused by the user's fault including not
following the instructions of this manual or not installing or handling the product in an appropriate
manner.
● The company shall not be liable for any direct or indirect damage, injury, or lost income resulting from
misuse, abuse, modification, or disassembly of all or any part of the product by the user.
● Unless expressly stated herein, the company makes no other warranties, express or implied, including
without limitation, any warranty of merchantability, fitness for a particular purpose, or infringement of
intellectual property rights.
● A Warranty Provision is included at the end of this manual. In the event that the product failure occurs,
the company shall repair the product in accordance with the Warranty Provision.
● The company shall not be liable for any incidents resulting from the use of parts other than those that
are recommended by the company.
5
VM-742B & VM-742 Table of Contents
<Table of Contents>
Regarding the Manuals for the VM-7 Monitoring System ................................................1
Important Notices ............................................................................................................4
Disclaimer........................................................................................................................5
Table of Contents.............................................................................................................6
6
VM-742B & VM-742 Table of Contents
Warranty Provision
7
VM-742B & VM-742 Chapter 1 For Safe Use of the Product
Point In this manual, Point is used to highlight some important issues as well as to provide useful
information.
8
VM-742B & VM-742 Chapter 1 For Safe Use of the Product
Never disassemble or modify the product. Do not use the product under conditions that exceed its
specifications. Otherwise, it may result in malfunction of the product, fire or injury.
Do not use radio equipment such as transceivers, or mobile phones around the product.
If electromagnetic waves are radiated to the product, signal cables or the transducer, the product
may not provide satisfactory precision described in the specifications.
Avoid any contact with the component or soldered parts of the instrument rack or the module.
Otherwise, it may result in damage to or malfunction of the rack or the module.
The instrument rack and the module are precision mechanical equipment. Avoid any excessive
shock or vibration.
Otherwise, it may result in damage to or malfunction of the product.
Avoid a voltage resistance test or a withstand voltage test in any places other than specified.
Otherwise, it may result in damage to the product.
9
VM-742B & VM-742 Chapter 1 For Safe Use of the Product
1.3.2 Compliances
Be sure to follow the instructions in this manual when installing the product.
Otherwise, it may result in malfunction of the product, fire or injury.
Before installation, be sure to read the manuals for this product as well as other instruments that
are to be connected. The use of this product requires expert knowledge of instruments, and the
knowledge of any associated equipment.
Before any contact with the instrument rack or module, be sure to touch the metal objects, etc. so
that the static electricity is discharged.
Static electricity accumulated in the human body may damage the product.
When carrying out a megger test, ensure that all the cables are disconnected from the driver
(transducer) as well as the converter.
After carrying out a megger test, ensure that any accumulated charge is discharged by
short-circuiting prior to reconnection.
Reconnecting the driver or the converter while the cables still carry a charge may damage the
product.
Be sure to connect the power input terminal to the power supply with the voltage and frequency
advised in the specifications.
Otherwise, it may result in damage to or malfunction of the product.
Make sure to place the module in the instructed position when installing the module onto the
instrument rack.
Otherwise, it may result in damage to the product or malfunction of the turbine monitoring system.
Ensure that all the screws on the instrument rack are tightly fastened and the instrument rack is
securely fixed.
Also, be sure to use module fasteners to fix each module properly.
10
VM-742B & VM-742 Chapter 1 For Safe Use of the Product
- Place surge killers around the product, the related instrumentation equipment, and the relay
around the signal cables to which these are connected.
- Ensure that the screws are tightly fastened and the product is securely fixed.
- Place the product in a metal cabinet, etc. and ensure that the lid is closed while the product is
operated.
- Use of ferrite cores (ZCAT3050-1330/TDK) can expect to reduce the noise induced by the
cables to a certain extent. It is recommended to use anti-noise devices in the case that the
conditions surrounding the product are considered to be unfavorable.
Contact the sales office you purchased the product from if any of the following occurs.
Continued use of the product under any of the conditions below may result in fire or malfunction.
Ensure that an interlock function is added to the product in order to prevent the product from
making mechanical errors in the case it is damaged.
Store the product in a place without the direct rays of the sun where the following conditions are
satisfied. Do not store it in a hot and humid location or any location where corrosive gases are
present.
11
VM-742B & VM-742 Chapter 2 Overview of Network Communication Module
2.1 Functions
This section provides basic information on the communication function, display function and control function
provided by this module.
Two types of the modules are available for the Temperature Monitor Module; VM-742B usable with VM-7B
monitor and VM-742 usable with VM-7 monitor.
There are restrictions as described below on the racks and slots to which each module can be installed.
VM-742B VM-742
Available rack VM-761B VM-761
VM-762B VM-762
Available slot Slot C1 or C2 Slot 1, 2
12
VM-742B & VM-742 Chapter 2 Overview of Network Communication Module
13
VM-742B & VM-742 Chapter 2 Overview of Network Communication Module
14
VM-742B & VM-742 Chapter 3 Installation/Uninstallation of a Module
Do not install the module in slots other than the one specified above. Otherwise, the module failure
could occur.
Before any contact with the instrument rack or module, be sure to touch the metal objects, etc. so
that the static electricity is discharged.
Avoid any contact with the component or soldered parts of the instrument rack or the module.
Otherwise, it may result in damage to or malfunction of the rack or the module.
The instrument rack and the module are precision mechanical equipment. Avoid any excessive
shock or vibration.
Do not install the module in slots other than the one specified above. Otherwise module failure
could occur.
15
VM-742B & VM-742 Chapter 3 Installation/Uninstallation of a Module
Slot C1
Slot C2
Slot 2
Slot 1
16
VM-742B & VM-742 Chapter 3 Installation/Uninstallation of a Module
1 Insert the circuit board of the module by sliding it straight along the slits
on the top and bottom of the instrument rack.
2 Turn the upper and lower module fasteners evenly with your hands until
the module shoulder hits the turned edge of the instrument rack.
(Do not use a tool to tighten the fasteners.
The module is inserted into the connector located inside the rack.)
Shoulder of the
module
Module fastener
3 Turn on the power of the module. Once the module is activated
successfully, the LED on the front panel will either light up or Turned edge
flash. of the
instrument
(Tighten the fasteners until the shoulder of the module comes in rack
contact with the turned edge of the instrument rack. If the LED
does not light up within 10 seconds, the module may be faulty.
In such a case, immediately uninstall the module to prevent a
heating hazard etc.)
Make sure to install the module in the specified slot on the instrument rack.
Ensure that all the fasteners on the instrument rack are tightly fastened and the instrument rack is
securely fixed.
4 If MCL View and Device Config are connected, update rack configuration data.
For details about how to update the software, see "Updating the Information Displayed on MCL View
P.18" and "Updating the Information Displayed on Device Config P.18".
17
VM-742B & VM-742 Chapter 3 Installation/Uninstallation of a Module
If you have changed the rack configuration during installation or uninstallation of the module, you need to
update the rack configuration information in the Device Config.
Update the information according to the steps described below.
1 Connect the PC and device with a USB cable or LAN 2
cable.
3
2 From the Device Config menu, select [Communication] -
[Connect].
Communication between Device Config and the device
starts.
3 In the [Communication] menu, click [Get Configuration].
4 Click [OK].
The latest information is reflected on the window.
18
VM-742B & VM-742 Chapter 4 Edit the Module Setting
4.1 Workflow
You can set this module as follows:
To daisy chain an external device with the VM-7B monitor via Modbus/RTU, you need to mount a terminating
resistor on the device located at the end of the connection. The VM-7B monitor allows you to set the
terminating resistor by using the switch located on the main board of the VM-742B Network Communication
module.
For details, see "Appendix B Setting the Terminating Resistor P.53" at the end of this manual.
19
VM-742B & VM-742 Chapter 4 Edit the Module Setting
4 Click [OK].
Loading of the setting starts.
When the channel configuration and settings are changed, an alarm may be output.
Open the circuit between each relay and the connection system, as necessary.
During communication between the device and Device Config, do not turn off the device power or
disconnect the communication cable.
Otherwise, the setting values changed using Device Config may not be reflected in the device.
As the setting is written, the measurement values in MCL View and host network may disappear.
20
VM-742B & VM-742 Chapter 4 Edit the Module Setting
Two modes are available for writing the setting from Device Config to the module.
- With [Set Configuration], the setting is written to all modules on the instrument rack. It may take some time to
transfer the setting to all modules.
- With [Set Slot Configuration], the setting is written to the modules in the selected slots.
3 2
1
1 From [Device Tree] on the left of the window, select the [Communication] node.
2 Select the [Common] button on the right of the window to display the setting items.
3 Set or check the items below.
21
VM-742B & VM-742 Chapter 4 Edit the Module Setting
Information
Serial Number The serial number is displayed.
Firmware ID (VM-742B only) The firmware ID is displayed.
Function
Control Signal Accept This item displays whether the control function provided by
(Under development) the host network is enabled. This is an optional function for
VM-742B.
• Off: Reject the control signals from the host network.
• On: Permit the control signals from the host network.
Communication
MAC Address The MAC address is displayed.
IP Address Enter the IP address for TCP/IP communication.
Make sure to set an address that is different from the
addresses set for other devices connected to the same network.
Subnet Mask Enter the subnet mask for TCP/IP communication.
Analysis (VM-742B only)
Communication Enabled Displays whether the infiSYS Analysis View communication
function is enabled or disabled.
Port Number Enter the port number for infiSYS Analysis View
communication.
(This setting is applicable only if [Communication Enabled]
under [Analysis] is enabled.)
Timeout Enter the timeout period for infiSYS Analysis View
communication.
(This setting is applicable only if [Communication Enabled]
under [Analysis] is enabled.)
Setting value: From 5 to 60 seconds
* Deselecting this checkbox disables the timeout function.
Modbus/TCP
Communication Enabled Displays whether the Modbus/TCP communication function is
enabled or disabled.
Port Number Enter the port number for Modbus/TCP communication.
(This setting is applicable only if [Communication Enabled]
under [Modbus/TCP] is enabled.)
Timeout Enter the timeout period for Modbus/TCP communication.
(This setting is applicable only if [Communication Enabled]
under [Modbus/TCP] is enabled.)
Setting value: From 5 to 60 seconds
* Deselecting this checkbox disables the timeout function.
22
VM-742B & VM-742 Chapter 4 Edit the Module Setting
23
VM-742B & VM-742 Chapter 4 Edit the Module Setting
Point Display colors and background colors of the setting value on Device Config
The font colors and background colors indicate the current state of the setting items in Device Config. Examples
are shown below.
Black bold font: The value has been changed but not yet specified. This value has not been set to the monitor
or saved in a file.
Red bold font: This value can be set. However, some sort of constraint/limit exists or the value is not
recommendable. (Warning value)
Yellow background & red bold font: The value does not operate the monitor normally. This value cannot be
set or saved. (Error value)
24
VM-742B & VM-742 Chapter 5 Maintenance and Service
Before replacing this module, stop any communication that may be going on with the devices connected via the
module.
Before any contact with the instrument rack or module, be sure to touch the metal objects, etc. so
that the static electricity is discharged.
Avoid any contact with the component or soldered parts of the instrument rack or the module.
Otherwise, it may result in damage to or malfunction of the rack or the module.
The instrument rack and the module are precision mechanical equipment. Avoid any excessive
shock or vibration.
Ensure that all the fasteners on the instrument rack are tightly fastened and the instrument rack is
securely fixed.
25
VM-742B & VM-742 Chapter 5 Maintenance and Service
The service life of the module also becomes shortened depending on the surrounding conditions of the module.
Ten years as a rough guide, it is preferable to replace the product at an earlier stage so that the stability of the
system can be ensured.
This module uses an aluminum electrolytic capacitor with the limited service life depending on the timing when
it is manufactured. Please see "Appendix C Replacement of Aluminum Electrolytic Capacitor" about the target
board and replacement parts.
To remove and attach the module, follow the steps described in "5.1 Replacing the Module".
26
VM-742B & VM-742 Chapter 6 Trouble Shooting
27
VM-742B & VM-742 Appendix A
By using the Network Communication Module VM-742B, VM-7 can transmit measurement data and other data
to a computer connected to the host network.
VM-742B supports Modbus/TCP communication and Modbus/RTU communication.
VM-742B always works as a slave (server). The connection of which the VM-742B can establish by
Modbus/TCP communication is only one.
The Modbus protocol has been developed by Modicon Inc. (AEG Schneider Automation International S.A.S.)
as the communication protocol for PLC. For details, refer to the following specifications. These specifications
can be obtained at http://www.modbus.org/.
MODBUS APPLICATION PROTOCOL SPECIFICATION V1.1b
MODBUS MESSAGING ON TCP/IP IMPLEMENTATION GUIDE V1.0b
Communication Overview
Multiple VM-7 Monitor Racks can be connected to the VM-7 Monitoring System with Modbus communication.
In this case, it is necessary to allocate individual IP address in each VM-7 Rack to distinguish individual VM-7.
The IP address can be changed by using VM-772B Device Config via the communication port used for
maintenance purposes located on the front of VM-741B Local Communication & Phase Marker Module.
In this case, you need to assign separate IP addresses to these VM-742B modules.
To connect several VM-7 modules to the network at a time, make sure to set a unique IP address to
each one of them beforehand.
By installing VM-742B, you can change the internal clock of VM-7 via VM-742B. The time data in
the VM-7 internal clock can be changed by using Device Config or infiSYS Analysis View as well.
Thus, if time change made from VM-742B is to be used as a standard, be careful not to change
time data from other software.
If you are not using SlaveID with Modbus/TCP communication, set SlaveID to "1".
Overview
VM-7 has the following three items as data for Modbus communications:
• Input status No. = 10001 - 19999
• Holding register No. = 40001 - 49999
•Coil No. = 00001 - 09999
The various status and measurement values for the VM-7 are assigned to the input status and holding register.
They may be accessed via Modbus communications.
This chapter shows the assignment of input status and holding register.
In the Modbus communications messages, the target status register is specified by 0 - 65535 addresses (data
reference numbers). In VM-7, assignment is as follows:
• Input status 10001 - 19999 becomes 0 - 9998
• Holding register 40001 - 49999 becomes 0 - 9998
* The term "Slot 0" in the manual refers to Local Communication & Phase Marker Module.
29
VM-742B & VM-742 Appendix A
The alarm settings specified for the individual module channels as well as the OK alarm settings are assigned to
this domain.
The alarm is identified by comparing the value against the value specified in "Measure Value in 32 Bits Floating
Point Format". The alarm identification is not affected by the value specified in "Scaling Constant for Measure
Value in 16 Bits Integer Format".
To retrieve each setting value, the slot numbers must be specified in advance in the following registers:
Status Name Holding Register No. Notes
Slot No. 40998 Specify the slot to be retrieved (from 0 to 11)
Block No. 40999 Fixed to 0 (If a value other than 0 is specified, no value can be retrieved.)
Specify the slot number, and then proceed to retrieve the individual setting values, according to the table below.
Slot N (N=0, 1, 2...)
Status Name Channel 1 Channel 2 Channel 3 Channel 4 Channel 5 Channel 6
30
VM-742B & VM-742 Appendix A
The measurement values of the individual module channels are assigned to this domain. They are in 32 bit
floating point format.
This is not effected by “Scaling Constant for Measure Value in 16 Bits Integer Format”.
"Measure" is the channel's primary measurement value. The overall value is output for the vibration monitor,
and the rotation speed is output for the tachometer monitor.
"Gap" is the sensor voltage.
If the module is not inserted into the slot, "0" will be output.
Measure - 42101 - 42102 42201 - 42202 42301 - 42302 42401 - 42402 42501 - 42502
Channel 1
Gap 42003 - 42004 42103 - 42104 42203 - 42204 42303 - 42304 42403 - 42404 42503 - 42504
Measure - 42105 - 42106 42205 - 42206 42305 - 42306 42405 - 42406 42505 - 42506
Channel 2
Gap 42007 - 42008 42107 - 42108 42207 - 42208 42307 - 42308 42407 - 42408 42507 - 42508
Measure - 42109 - 42110 42209 - 42210 42309 - 42310 42409 - 42410 42509 - 42510
Channel 3
Gap 42011- 42012 42111 - 42112 42211 - 42212 42311 - 42312 42411 - 42412 42511 - 42512
Measure - 42113 - 42114 42213 - 42214 42313 - 42314 42413 - 42414 42513 - 42514
Channel 4
Gap 42015 - 42016 42115 - 42116 42215 - 42216 42315 - 42316 42415 - 42416 42515 - 42516
Measure - 42117 - 42118 42217 - 42218 42317 - 42318 42417 - 42418 42517 - 42518
Channel 5
Gap - - - - - -
Measure - 42121 - 42122 42221 - 42222 42321 - 42322 42421 - 42422 42521 - 42522
Channel 6
Gap - - - - - -
Measure 42601 - 42602 42701 - 42702 42801 - 42802 42901 - 42902 43001 - 43002 43101 - 43102
Channel 1
Gap 42603 - 42604 42703 - 42704 42803 - 42804 42903 - 42904 43003 - 43004 43103 - 43104
Measure 42605 - 42606 42705 - 42706 42805 - 42806 42905 - 42906 43005 - 43006 43105 - 43106
Channel 2
Gap 42607 - 42608 42707 - 42708 42807 - 42808 42907 - 42908 43007 - 43008 43107 - 43108
Measure 42609 - 42610 42709 - 42710 42809 - 42810 42909 - 42910 43009 - 43010 43109 - 43110
Channel 3
Gap 42611 - 42612 42711 - 42712 42811 - 42812 42911 - 42912 43011 - 43012 43111 - 43112
Measure 42613 - 42614 42713 - 42714 42813 - 42814 42913 - 42914 43013 - 43014 43113 - 43114
Channel 4
Gap 42615 - 42616 42715 - 42716 42815 - 42816 42915 - 42916 43015 - 43016 43115 - 43116
Measure 42617 - 42618 42717 - 42718 42817 - 42818 42917 - 42918 43017 - 43018 43117 - 43118
Channel 5
Gap - - - - - -
Measure 42621 - 42622 42721 - 42722 42821 - 42822 42921 - 42922 43021 - 43022 43121 - 43122
Channel 6
Gap - - - - - -
31
VM-742B & VM-742 Appendix A
The measurement values (analysis data) of the individual module channels are assigned to this domain. They are
in 32 bit floating point format.
This domain is not affected by "Scaling Constant for Measure Value in 16 Bits Integer Format (Analysis)".
"RPM" is the rotation speed of the phase marker assigned to the channel.
The 0.5X, 1X, 2X or Not1X amplitude ("Amplitude") and phase ("Phase") are the amplitude and phase of the
component synchronous to the rotation frequency.
"Smax Amplitude" is the S(p-p)max amplitude.
If the analysis board is not inserted into the slot, "0" will be output.
32
VM-742B & VM-742 Appendix A
Measure Value in 32 Bits Floating Point Format (Analysis) (Holding Register) * From Slot 0 to
Slot 5
Status Name Slot 0 Slot 1 Slot 2 Slot 3 Slot 4 Slot 5
Channel 1 RPM - 43601 - 43602 43701 - 43702 43801 - 43802 43901 - 43902 44001 - 44002
0.5X Amplitude - 43603 - 43604 43703 - 43704 43803 - 43804 43903 - 43904 44003 - 44004
0.5X Phase - 43605 - 43606 43705 - 43706 43805 - 43806 43905 - 43906 44005 - 44006
1X Amplitude - 43607 - 43608 43707 - 43708 43807 - 43808 43907 - 43908 44007 - 44008
1X Phase - 43609 - 43610 43709 - 43710 43809 - 43810 43909 - 43910 44009 - 44010
2X Amplitude - 43611 - 43612 43711 - 43712 43811 - 43812 43911 - 43912 44011 - 44012
2X Phase - 43613 - 43614 43713 - 43714 43813 - 43814 43913 - 43914 44013 - 44014
Not1X Amplitude - 43615 - 43616 43715 - 43716 43815 - 43816 43915 - 43916 44015 - 44016
Smax Amplitude - 43617 - 43618 43717 - 43718 43817 - 43818 43917 - 43918 44017 - 44018
(Reserved) - - - - - -
Channel 2 RPM - 43625 - 43626 43725 - 43726 43825 - 43826 43925 - 43926 44025 - 44026
0.5X Amplitude - 43627 - 43628 43727 - 43728 43827 - 43828 43927 - 43928 44027 - 44028
0.5X Phase - 43629 - 43630 43729 - 43730 43829 - 43830 43929 - 43930 44029 - 44030
1X Amplitude - 43631 - 43632 43731 - 43732 43831 - 43832 43931 - 43932 44031 - 44032
1X Phase - 43633 - 43634 43733 - 43734 43833 - 43834 43933 - 43934 44033 - 44034
2X Amplitude - 43635 - 43636 43735 - 43736 43835 - 43836 43935 - 43936 44035 - 44036
2X Phase - 43637 - 43638 43737 - 43738 43837 - 43838 43937 - 43938 44037 - 44038
Not1X Amplitude - 43639 - 43640 43739 - 43740 43839 - 43840 43939 - 43940 44039 - 44040
Smax Amplitude - 43641 - 43642 43741 - 43742 43841 - 43842 43941 - 43942 44041 - 44042
(Reserved) - - - - - -
Channel 3 RPM - 43649 - 43650 43749 - 43750 43849 - 43850 43949 - 43950 44049 - 44050
0.5X Amplitude - 43651 - 43652 43751 - 43752 43851 - 43852 43951 - 43952 44051 - 44052
0.5X Phase - 43653 - 43654 43753 - 43754 43853 - 43854 43953 - 43954 44053 - 44054
1X Amplitude - 43655 - 43656 43755 - 43756 43855 - 43856 43955 - 43956 44055 - 44056
1X Phase - 43657 - 43658 43757 - 43758 43857 - 43858 43957 - 43958 44057 - 44058
2X Amplitude - 43659 - 43660 43759 - 43760 43859 - 43860 43959 - 43960 44059 - 44060
2X Phase - 43661 - 43662 43761 - 43762 43861 - 43862 43961 - 43962 44061 - 44062
Not1X Amplitude - 43663 - 43664 43763 - 43764 43863 - 43864 43963 - 43964 44063 - 44064
Smax Amplitude - 43665 - 43666 43765 - 43766 43865 - 43866 43965 - 43966 44065 - 44066
(Reserved) - - - - - -
Channel 4 RPM - 43673 - 43674 43773 - 43774 43873 - 43874 43973 - 43974 44073 - 44074
0.5X Amplitude - 43675 - 43676 43775 - 43776 43875 - 43876 43975 - 43976 44075 - 44076
0.5X Phase - 43677 - 43678 43777 - 43778 43877 - 43878 43977 - 43978 44077 - 44078
1X Amplitude - 43679 - 43680 43779 - 43780 43879 - 43880 43979 - 43980 44079 - 44080
1X Phase - 43681 - 43682 43781 - 43782 43881 - 43882 43981 - 43982 44081 - 44082
2X Amplitude - 43683 - 43684 43783 - 43784 43883 - 43884 43983 - 43984 44083 - 44084
2X Phase - 43685 - 43686 43785 - 43786 43885 - 43886 43985 - 43986 44085 - 44086
Not1X Amplitude - 43687 - 43688 43787 - 43788 43887 - 43888 43987 - 43988 44087 - 44088
Smax Amplitude - 43689 - 43690 43789 - 43790 43889 - 43890 43989 - 43990 44089 - 44090
(Reserved) - - - - - -
33
VM-742B & VM-742 Appendix A
Measure Value in 32 Bits Floating Point Format (Analysis) (Holding Register) * From Slot 6 to
Slot 11
Status Name Slot 6 Slot 7 Slot 8 Slot 9 Slot 10 Slot 11
Channel 1 RPM 44101 - 44102 44201- 44202 44301 - 44302 44401 - 44402 44501 - 44502 44601 - 44602
0.5X Amplitude 44103 - 44104 44203 - 44204 44303 - 44304 44403 - 44404 44503 - 44504 44603 - 44604
0.5X Phase 44105 - 44106 44205 - 44206 44305 - 44306 44405 - 44406 44505 - 44506 44605 - 44606
1X Amplitude 44107 - 44108 44207 - 44208 44307 - 44308 44407 - 44408 44507 - 44508 44607 - 44608
1X Phase 44109 - 44110 44209 - 44210 44309 - 44310 44409 - 44410 44509 - 44510 44609 - 44610
2X Amplitude 44111 - 44112 44211 - 44212 44311 - 44312 44411 - 44412 44511 - 44512 44611 - 44612
2X Phase 44113 - 44114 44213 - 44214 44313 - 44314 44413 - 44414 44513 - 44514 44613 - 44614
Not1X Amplitude 44115 - 44116 44215 - 44216 44315 - 44316 44415 - 44416 44515 - 44516 44615 - 44616
Smax 44117 - 44118 44217 - 44218 44317 - 44318 44417 - 44418 44517 - 44518 44617 - 44618
(Reserved) - - - - - -
Channel 2 RPM 44125 - 44126 44225 - 44226 44325 - 44326 44425 - 44426 44525 - 44526 44625 - 44626
0.5X Amplitude 44127 - 44128 44227 - 44228 44327 - 44328 44427 - 44428 44527 - 44528 44627 - 44628
0.5X Phase 44129 - 44130 44229 - 44230 44329 - 44330 44429 - 44430 44529 - 44530 44629 - 44630
1X Amplitude 44131 - 44132 44231 - 44232 44331 - 44332 44431 - 44432 44531 - 44532 44631 - 44632
1X Phase 44133 - 44134 44233 - 44234 44333 - 44334 44433 - 44434 44533 - 44534 44633 - 44634
2X Amplitude 44135 - 44136 44235 - 44236 44335 - 44336 44435 - 44436 44535 - 44536 44635 - 44636
2X Phase 44137 - 44138 44237 - 44238 44337 - 44338 44437 - 44438 44537 - 44538 44637 - 44638
Not1X Amplitude 44139 - 44140 44239 - 44240 44339 - 44340 44439 - 44440 44539 - 44540 44639 - 44640
Smax 44141 - 44142 44241 - 44242 44341 - 44342 44441 - 44442 44541 - 44542 44641 - 44642
(Reserved) - - - - - -
Channel 3 RPM 44149 - 44150 44249 - 44250 44349 - 44350 44449 - 44450 44549 - 44550 44649 - 44650
0.5X Amplitude 44151 - 44152 44251 - 44252 44351 - 44352 44451 - 44452 44551 - 44552 44651 - 44652
0.5X Phase 44153 - 44154 44253 - 44254 44353 - 44354 44453 - 44454 44553 - 44554 44653 - 44654
1X Amplitude 44155 - 44156 44255 - 44256 44355 - 44356 44455 - 44456 44555 - 44556 44655 - 44656
1X Phase 44157 - 44158 44257 - 44258 44357 - 44358 44457 - 44458 44557 - 44558 44657 - 44658
2X Amplitude 44159 - 44160 44259 - 44260 44359 - 44360 44459 - 44460 44559 - 44560 44659 - 44660
2X Phase 44161 - 44162 44261 - 44262 44361 - 44362 44461 - 44462 44561 - 44562 44661 - 44662
Not1X Amplitude 44163 - 44164 44263 - 44264 44363 - 44364 44463 - 44464 44563 - 44564 44663 - 44664
Smax 44165 - 44166 44265 - 44266 44365 - 44366 44465 - 44466 44565 - 44566 44665 - 44666
(Reserved) - - - - - -
Channel 4 RPM 44173 - 44174 44273 - 44274 44373 - 44374 44473 - 44474 44573 - 44574 44673 - 44674
0.5X Amplitude 44175 - 44176 44275 - 44276 44375 - 44376 44475 - 44476 44575 - 44576 44675 - 44676
0.5X Phase 44177 - 44178 44277 - 44278 44377 - 44378 44477 - 44478 44577 - 44578 44677 - 44678
1X Amplitude 44179 - 44180 44279 - 44280 44379 - 44380 44479 - 44480 44579 - 44580 44679 - 44680
1X Phase 44181 - 44182 44281 - 44282 44381 - 44382 44481 - 44482 44581 - 44582 44681 - 44682
2X Amplitude 44183 - 44184 44283 - 44284 44383 - 44384 44483 - 44484 44583 - 44584 44683 - 44684
2X Phase 44185 - 44186 44285 - 44286 44385 - 44386 44485 - 44486 44585 - 44586 44685 - 44686
Not1X Amplitude 44187 - 44188 44287 - 44288 44387 - 44388 44487 - 44488 44587 - 44588 44687 - 44688
Smax 44189 - 44190 44289 - 44290 44389 - 44390 44489 - 44490 44589 - 44590 44689 - 44690
(Reserved) - - - - - -
34
VM-742B & VM-742 Appendix A
The addresses in this domain can be customized by assigning the existing address data to a desired address.
When this is done, the existing address data will be copied to the specified address.
Data can be assigned to any addresses within the ranges 10001 to 14999, 40001 to 44999 and 46001 to 49999. If
an undefined address is specified, "0" will be output.
In this way, the required data can be mapped to a contiguous address, enabling the data to be retrieved more
efficiently.
You can customize address assignment by writing "VZ-77_config.txt" of "VZ-77 Configuration File" (optional)
to VM-742B.
(Point CONFIGURATION FILE FORMAT)
(Point TRANSMITTING THE CONFIGURATION FILE)
Scaling Constant for Measure Value in 16 Bits Integer Format [Read Only]
Coefficients used to convert the measurement values in 32-bit floating point format (Measure Value in 32 Bits
Floating Point Format) to the measurement values in 16-bit integer format (Measure Value in 16 Bits Integer
Format) are assigned to this domain.
You can customize the conversion coefficients by writing "VZ-77_config.txt" of "VZ-77 Configuration File"
(optional) to VM742B.
(Point CONFIGURATION FILE FORMAT)
(Point TRANSMITTING THE CONFIGURATION FILE)
If you expand the original measurement value prior to converting it to a measurement value in
16-bit integer format, the data may change stepwise due to problems related to resolution.
To retrieve each setting value, the slot numbers must be specified in advance in the following registers:
Status Name Holding Register No. Notes
Slot No. 46998 Specify the slot to be retrieved (from 0 to 11)
Block No. 46999 Fixed to 0 (If a value other than 0 is specified, no value can be retrieved.)
35
VM-742B & VM-742 Appendix A
Specify the slot number, and then proceed to retrieve the individual setting values, according to the table below.
Slot N (N=0, 1, 2...)
Status Name Channel 1 Channel 2 Channel 3 Channel 4 Channel 5 Channel 6
Measure Input Min 46001 - 46002 46101 - 46102 46201 - 46202 46301 - 46302 46401 - 46402 46501 - 45002
Measure Input Max 46003 - 46004 46103 - 46104 46203 - 46204 46303 - 46304 46403 - 46404 46503 - 46504
Measure Output Min 46005 - 46006 46105 - 46106 46205 - 46206 46305 - 46306 46405 - 46406 46505 - 46506
Measure Output Max 46007 - 46008 46107 - 46108 46207 - 46208 46307 - 46308 46407 - 46408 46507 - 46508
Gap Input Min 46009 - 46010 46109 - 46110 46209 - 46210 46309 - 46310 - -
Gap Input Max 46011 - 46012 46111 - 46112 46211 - 46212 46311 - 46312 - -
Gap Output Min 46013 - 46014 46113 - 46114 46213 - 46214 46313 - 46314 - -
Gap Output Max 46015 - 46016 46115 - 46116 46215 - 46216 46315 - 46316 - -
RPM Input Min 46017 - 46018 46117 - 46118 46217 - 46218 46317 - 46318 - -
RPM Input Max 46019 - 46020 46119 - 46120 46219 - 46220 46319 - 46320 - -
RPM Output Min 46021 - 46022 46121 - 46122 46221 - 46222 46321 - 46322 - -
RPM Output Max 46023 - 46024 46123 - 46124 46223 - 46224 46323 - 46324 - -
0.5X Amplitude Input Min 46025 - 46026 46125 - 46126 46225 - 46226 46325 - 46326 - -
0.5X Amplitude Input Max 46027 - 46028 46127 - 46128 46227 - 46228 46327 - 46328 - -
0.5X Amplitude Output Min 46029 - 46030 46129 - 46130 46229 - 46230 46329 - 46330 - -
0.5X Amplitude Output Max 46031 - 46032 46131 - 46132 46231 - 46232 46331 - 46332 - -
0.5X Phase Input Min 46033 - 46034 46133 - 46134 46233 - 46234 46333 - 46334 - -
0.5X Phase Input Max 46035 - 46036 46135 - 46136 46235 - 46236 46335 - 46336 - -
0.5X Phase Output Min 46037 - 46038 46137 - 46138 46237 - 46238 46337 - 46338 - -
0.5X Phase Output Max 46039 - 46040 46139 - 46140 46239 - 46240 46339 - 46340 - -
1X Amplitude Input Min 46041 - 46042 46141 - 46142 46241 - 46242 46341 - 46342 - -
1X Amplitude Input Max 46043 - 46044 46143 - 46144 46243 - 46244 46343 - 46344 - -
1X Amplitude Output Min 46045 - 46046 46145 - 46146 46245 - 46246 46345 - 46346 - -
1X Amplitude Output Max 46047 - 46048 46147 - 46148 46247 - 46248 46347 - 46348 - -
1X Phase Input Min 46049 - 46050 46149 - 46150 46249 - 46250 46349 - 46350 - -
1X Phase Input Max 46051 - 46052 46151 - 46152 46251 - 46252 46351 - 46352 - -
1X Phase Output Min 46053 - 46054 46153 - 46154 46253 - 46254 46353 - 46354 - -
1X Phase Output Max 46055 - 46056 46155 - 46156 46255 - 46256 46355 - 46356 - -
2X Amplitude Input Min 46057 - 46058 46157 - 46158 46257 - 46258 46357 - 46358 - -
2X Amplitude Input Max 46059 - 46060 46159 - 46160 46259 - 46260 46359 - 46360 - -
2X Amplitude Output Min 46061 - 46062 46161 - 46162 46261 - 46262 46361 - 46362 - -
2X Amplitude Output Max 46063 - 46064 46163 - 46164 46263 - 46264 46363 - 46364 - -
2X Phase Input Min 46065 - 46066 46165 - 46166 46265 - 46266 46365 - 46366 - -
2X Phase Input Max 46067 - 46068 46167 - 46168 46267 - 46268 46367 - 46368 - -
2X Phase Output Min 46069 - 46070 46169 - 46170 46269 - 46270 46369 - 46370 - -
2X Phase Output Max 46071 - 46072 46171 - 46172 46271 - 46272 46371 - 46372 - -
Not1X Amplitude Input Min 46073 - 46074 46173 - 46174 46273 - 46274 46373 - 46374 - -
Not1X Amplitude Input Max 46075 - 46076 46175 - 46176 46275 - 46276 46375 - 46376 - -
Not1X Amplitude Output Min 46077 - 46078 46177 - 46178 46277 - 46278 46377 - 46378 - -
Not1X Amplitude Output Max 46079 - 46080 46179 - 46180 46279 - 46280 46379 - 46380 - -
Smax Amplitude Input Min 46081 - 46082 46181 - 46182 46281 - 46282 46381 - 46382 - -
Smax Amplitude Input Max 46083 - 46084 46183 - 46184 46283 - 46284 46383 - 46384 - -
Smax Amplitude Output Min 46085 - 46086 46185 - 46186 46285 - 46286 46385 - 46386 - -
Smax Amplitude Output Max 46087 - 46088 46187 - 46188 46287 - 46288 46387 - 46388 - -
36
VM-742B & VM-742 Appendix A
37
VM-742B & VM-742 Appendix A
The measurement values of the individual module channels are assigned to this domain. They are in 16 bit
floating point format.
The measurement values in 32-bit floating point format (Measure Value in 32 Bits Floating Point Format) are
converted to the coefficients specified in "Scaling Constant for Measure Value in 16 Bits Integer Format" and
then rounded off to 16-bit integers.
The values after the decimal point will be rounded off during rounding. In addition, values larger than 32767
will be limited to 32767, and values smaller than -32768 will be limited to -32768.
"Measure" is the channel's primary measurement value. The overall value is output for the vibration monitor,
and the rotation speed is output for the tachometer monitor.
"Gap" is the sensor voltage.
If the module is not inserted into the slot, "0" will be output.
38
VM-742B & VM-742 Appendix A
The measurement values (analysis data) of the individual module channels are assigned to this domain. They are
in 16 bit floating point format.
The measurement values (analysis data) in 32-bit floating point format (Measure Value in 32 Bits Floating Point
Format (Analysis)) are converted to the coefficients specified in "Scaling Constant for Measure Value in 16 Bits
Integer Format" and then rounded off to 16-bit integers.
The values after the decimal point will be rounded off during rounding. In addition, values larger than 32767
will be limited to 32767, and values smaller than -32768 will be limited to -32768.
"RPM" is the rotation speed of the phase marker assigned to the channel.
The 0.5X, 1X, 2X or Not1X amplitude ("Amplitude") and phase ("Phase") are the amplitude and phase of the
component synchronous to the rotation frequency.
"Smax Amplitude" is the S(p-p)max amplitude.
If the analysis board is not inserted into the slot, "0" will be output.
39
VM-742B & VM-742 Appendix A
Measure Value in 16 Bits Integer Format (Analysis) (Holding Register) * From Slot 0 to Slot 5
Status Name Slot 0 Slot 1 Slot 2 Slot 3 Slot 4 Slot 5
40
VM-742B & VM-742 Appendix A
Measure Value in 16 Bits Integer Format (Analysis) (Holding Register) * From Slot 6 to Slot 11
Status Name Slot 6 Slot 7 Slot 8 Slot 9 Slot 10 Slot 11
41
VM-742B & VM-742 Appendix A
42
VM-742B & VM-742 Appendix A
43
VM-742B & VM-742 Appendix A
44
VM-742B & VM-742 Appendix A
Module Status (Holding Register & Input Status) * From Slot 0 to Slot 5
Slot 0 Slot 1 Slot 2
Holding Input Holding Input Holding Input
Status Name bit bit bit
Register No. Status No. Register No. Status No. Register No. Status No.
45
VM-742B & VM-742 Appendix A
Module Status (Holding Register & Input Status) * From Slot 6 to Slot 11
Slot 6 Slot 7 Slot 8
Holding Input Holding Input Holding Input
Status Name bit bit bit
Register No. Status No. Register No. Status No. Register No. Status No.
46
VM-742B & VM-742 Appendix A
47
VM-742B & VM-742 Appendix A
Set the time to be used by "Set Time Now". At this point, the VM-7 real time clock is not yet changed.
Read it out simultaneously for the whole system (49501-49508).
This sets the time for the VM-7 real time clock.
Before setting this, use “Preload Setting Time” to transfer the time you wish to set to VM-7.
Then, when that time arrives, specify the write command in "Set Time Now".
Then, cancel the write command.
If a valid time is not set via “Preload Setting Time”, exception response 03 will be returned.
Status Name Holding Register No. Value
Set Time Now 49531 0: Cancel write command
1: Issue write command
48
VM-742B & VM-742 Appendix A
Retrieve the last time the measurement value was retrieved using either "Measure Value in 32 Bits Floating
Point Format" or "Measure Value in 16 Bits Integer Format".
Read it out simultaneously for the whole system.
This reads out the number of alarms (ALERT, DANGER, NOT-OK) held.
If the value is 0, the event is not stored.
A maximum of 100 events are stored. If the number of events exceeds 100, the old events are automatically
deleted.
Specify the number corresponding to the history to be retrieved using "Read Alarm History" ("Read Alarm
History" is described later in this manual). Be sure to set this prior to executing “Read Alarm History".
1 shows the newest event. The larger the Alarm Event Number, the older the event.
In addition, if 0 is written, the entire history is deleted.
This reads out the history of alarms (ALERT, DANGER, NOT-OK) occurring in the rack.
Read it out simultaneously for the whole system.
If the history specified at “Select Alarm History” does not exist, exception response 03 will be returned.
49
VM-742B & VM-742 Appendix A
The communication item will be added when the /CTR option is selected. Thus, VM-7 can be controlled from
the host network (host computer). The alarm bypass, channel bypass, alarm reset, etc. can be controlled by
inputting data into the specified control status and register.
If there are two VM-742B Monitor Racks, the control signals from the host network are processed by logical
OR operation.
50
VM-742B & VM-742 Appendix A
51
VM-742B & VM-742 Appendix B
If you select Modbus/RTU communication in the VM-7B monitor, you can build a network via daisy-chain
connection. The communication specification applicable to daisy-chain connections is RS-485.
Recommended device
D-Sub Connector (9-pin Socket) : XM2D-0901 (OMRON Corporation)
Hood : XM2S-0913 (OMRON Corporation)
Make sure that all devices are connected correctly. Otherwise, it may cause damage or
malfunction to the instrument rack or module as well as to other devices.
52
VM-742B & VM-742 Appendix B
Connect the monitor and the cable connectors as shown in the figure below.
The terminating resistor must be set for the module located at the end of the daisy-chain connection. As shown
in the figure below, turn ON the terminating resistor on the monitor located at the end of the network. The
terminating resistor for other devices must be turned OFF.
Communication may not work properly if the terminating resistor is not set correctly.
If communication is performed via two channels (primary and secondary channels), set the
terminating resistor on the main board of each one of them.
Before any contact with the instrument rack or module, be sure to touch the metal objects, etc. so
that the static electricity is discharged.
You can turn ON or OFF the terminating resistor by using the switch for terminating resistor located on the main
board of this module. The switch type differs by the board. Check the board number on the main board and set
the switch. For details, see "Point Checking the Board Number".
53
VM-742B & VM-742 Appendix B
54
VM-742B & VM-742 Appendix B
Parts of this module may be changed due to improvement of a board. When a board is improved, the last-digit
alphabet of the board number will be raised. If the board number is described in the text, check the number
printed on the board.
55
VM-742B & VM-742 Appendix C
Aluminum electrolytic capacitors are used on the main board, which have a limited life. Their service life varies
according to the conditions of use, but for preventive maintenance it is recommended to check the capacitors
listed below in 5-year cycles for characteristic deterioration and replace them as necessary.
To remove and attach the module, follow the steps described in "5.1 Replacing the Module".
56
VM-742B & VM-742 Appendix D
By using the Network Communication Module VM-742, VM-7 can transmit measurement data and other data to
a computer connected to the host network.
VM-742 supports Modbus/TCP communication.
VM-742 always works as a slave (server). The connection of which the VM-742 can establish by Modbus/TCP
communication is only one.
The Modbus protocol has been developed by Modicon Inc. (AEG Schneider Automation International S.A.S.)
as the communication protocol for PLC. For details, refer to the following specifications. These specifications
can be obtained at http://www.modbus.org/.
MODBUS APPLICATION PROTOCOL SPECIFICATION V1.1b
MODBUS MESSAGING ON TCP/IP IMPLEMENTATION GUIDE V1.0b
Communication Overview
Multiple VM-7 Monitor racks can be connected to the VM-7 Monitoring System with Modbus communication.
In this case, it is necessary to allocate individual IP address in each VM-7 Rack to distinguish individual VM-7.
The IP address can be changed by using VM-772B Device Config via the communication port used for
maintenance purposes located on the front of VM-741 Local Communication & Phase Marker Module.
In this case, you need to assign separate IP addresses to these VM-742 modules.
To connect several VM-7 modules to the network at a time, make sure to set a unique IP address to
each one of them beforehand.
By installing VM-742, you can change the internal clock of VM-7 via VM-742. The time data in the
VM-7 internal clock can be changed by using Device Config or infiSYS Analysis View as well.
Thus, if time change made from VM-742 is to be used as a standard, be careful not to change time
data from other software.
If you are not using SlaveID with Modbus/TCP communication, set SlaveID to "1".
57
VM-742B & VM-742 Appendix D
Overview
VM-7 has the following three items as data for Modbus communications:
• Input status No. = 10001 - 19999
• Holding register No. = 40001 - 49999
•Coil No. = 00001 - 09999
The various status and measurement values for the VM-7 are assigned to the input status and holding register.
They may be accessed via Modbus communications.
This chapter shows the assignment of input status and holding register.
In the Modbus communications messages, the target status register is specified by 0 - 65535 addresses (data
reference numbers). In VM-7, assignment is as follows:
• Input status 10001 - 19999 becomes 0 - 9998
• Holding register 40001 - 49999 becomes 0 - 9998
* The term "Slot 0" in the manual refers to Local Communication & Phase Marker Module.
58
VM-742B & VM-742 Appendix D
The alarm settings specified for the individual module channels as well as the OK alarm settings are assigned to
this domain.
The alarm is identified by comparing the value against the value specified in "Measure Value in 32 Bits Floating
Point Format". The alarm identification is not affected by the value specified in "Scaling Constant for Measure
Value in 16 Bits Integer Format".
Address Range (Holding Register)
Holding Register no.: From 40001 to 41999
ALERT High - 40101 - 40102 40201 - 40202 40301 - 40302 40401 - 40402 40501 - 40502
ALERT Low - 40103 - 40104 40203 - 40204 40303 - 40304 40403 - 40404 40503 - 40504
DANGER High - 40105 - 40106 40205 - 40206 40305 - 40306 40405 - 40406 40505 - 40506
Channel 1 DANGER Low - 40107 - 40108 40207 - 40208 40307 - 40308 40407 - 40408 40507 - 40508
OK High 40009 - 40010 40109 - 40110 40209 - 40210 40309 - 40310 40409 - 40410 40509 - 40510
OK Low 40011 - 40012 40111 - 40112 40211 - 40212 40311 - 40312 40411 - 40412 40511 - 40512
(Reserved) - - - - - -
ALERT High - 40117 - 40118 40217 - 40218 40317 - 40318 40417 - 40418 40517 - 40518
ALERT Low - 40119 - 40120 40219 - 40220 40319 - 40320 40419 - 40420 40519 - 40520
DANGER High - 40121 - 40122 40221 - 40222 40321 - 40322 40421 - 40422 40521 - 40522
Channel 2 DANGER Low - 40123 - 40124 40223 - 40224 40323 - 40324 40423 - 40424 40523 - 40524
OK High 40025 - 40026 40125 - 40126 40225 - 40226 40325 - 40326 40425 - 40426 40525 - 40526
OK Low 40027 - 40028 40127 - 40128 40227 - 40228 40327 - 40328 40427 - 40428 40527 - 40528
(Reserved) - - - - - -
ALERT High - 40133 - 40134 40233 - 40234 40333 - 40334 40433 - 40434 40533 - 40534
ALERT Low - 40135 - 40136 40235 - 40236 40335 - 40336 40435 - 40436 40535 - 40536
DANGER High - 40137 - 40138 40237 - 40238 40337 - 40338 40437 - 40438 40537 - 40538
Channel 3 DANGER Low - 40139 - 40140 40239 - 40240 40339 - 40340 40439 - 40440 40539 - 40540
OK High - 40141 - 40142 40241 - 40242 40341 - 40342 40441 - 40442 40541 - 40542
OK Low - 40143 - 40144 40243 - 40244 40343 - 40344 40443 - 40444 40543 - 40544
(Reserved) - - - - - -
ALERT High - 40149 - 40150 40249 - 40250 40349 - 40350 40449 - 40450 40549 - 40550
ALERT Low - 40151 - 40152 40251 - 40252 40351 - 40352 40451 - 40452 40551 - 40552
DANGER High - 40153 - 40154 40253 - 40254 40353 - 40354 40453 - 40454 40553 - 40554
Channel 4 DANGER Low - 40155 - 40156 40255 - 40256 40355 - 40356 40455 - 40456 40555 - 40556
OK High - 40157 - 40158 40257 - 40258 40357 - 40358 40457 - 40458 40557 - 40558
OK Low - 40159 – 40160 40259 - 40260 40359 - 40360 40459 - 40460 40559 - 40560
(Reserved) - - - - - -
ALERT High - 40165 - 40166 40265 - 40266 40365 - 40366 40465 - 40466 40565 - 40566
ALERT Low - 40167 - 40168 40267 - 40268 40367 - 40368 40467 - 40468 40567 - 40568
Channel 5 DANGER High - 40169 - 40170 40269 - 40270 40369 - 40370 40469 - 40470 40569 - 40570
DANGER Low - 40171 - 40172 40271 - 40272 40371 - 40372 40471 - 40472 40571 - 40572
(Reserved) - - - - - -
ALERT High - 40181 - 40182 40281 - 40282 40381 - 40382 40481 - 40482 40581 - 40582
ALERT Low - 40183 - 40184 40283 - 40284 40383 - 40384 40483 - 40484 40583 - 40584
Channel 6 DANGER High - 40185 - 40186 40285 - 40286 40385 - 40386 40485 - 40486 40585 - 40586
DANGER Low - 40187 - 40188 40287 - 40288 40387 - 40388 40487 - 40488 40587 - 40588
(Reserved) - - - - - -
59
VM-742B & VM-742 Appendix D
ALERT High 40601 - 40602 40701 - 40702 40801 - 40802 40901 - 40902 41001 - 41002
ALERT Low 40603 - 40604 40703 - 40704 40803 - 40804 40903 - 40904 41003 - 41004
DANGER High 40605 - 40606 40705 - 40706 40805 - 40806 40905 - 40906 41005 - 41006
Channel 1 DANGER Low 40607 - 40608 40707 - 40708 40807 - 40808 40907 - 40908 41007 - 41008
OK High 40609 - 40610 40709 - 40710 40809 - 40810 40909 - 40910 41009 - 41010
OK Low 40611 - 40612 40711 - 40712 40811 - 40812 40911 - 40912 41011 - 41012
(Reserved) - - - - -
ALERT High 40617 - 40618 40717 - 40718 40817 - 40818 40917 - 40918 41017 - 41018
ALERT Low 40619 - 40620 40719 - 40720 40819 - 40820 40919 - 40920 41019 - 41020
DANGER High 40621 - 40622 40721 - 40722 40821 - 40822 40921 - 40922 41021 - 41022
Channel 2 DANGER Low 40623 - 40624 40723 - 40724 40823 - 40824 40923 - 40924 41023 - 41024
OK High 40625 - 40626 40725 - 40726 40825 - 40826 40925 - 40926 41025 - 41026
OK Low 40627 - 40628 40727 - 40728 40827 - 40828 40927 - 40928 41027 - 41028
(Reserved) - - - - -
ALERT High 40633 - 40634 40733 - 40734 40833 - 40834 40933 - 40934 41033 - 41034
ALERT Low 40635 - 40636 40735 - 40736 40835 - 40836 40935 - 40936 41035 - 41036
DANGER High 40637 - 40638 40737 - 40738 40837 - 40838 40937 - 40938 41037 - 41038
Channel 3 DANGER Low 40639 - 40640 40739 - 40740 40839 - 40840 40939 - 40940 41039 - 41040
OK High 40641 - 40642 40741 - 40742 40841 - 40842 40941 - 40942 41041 - 41042
OK Low 40643 - 40644 40743 - 40744 40843 - 40844 40943 - 40944 41043 - 41044
(Reserved) - - - - -
ALERT High 40649 - 40650 40749 - 40750 40849 - 40850 40949 - 40950 41049 - 41050
ALERT Low 40651 - 40652 40751 - 40752 40851 - 40852 40951 - 40952 41051 - 41052
DANGER High 40653 - 40654 40753 - 40754 40853 - 40854 40953 - 40954 41053 - 41054
Channel 4 DANGER Low 40655 - 40656 40755 - 40756 40855 - 40856 40955 - 40956 41055 - 41056
OK High 40657 - 40658 40757 - 40758 40857 - 40858 40957 - 40958 41057 - 41058
OK Low 40659 - 40660 40759 - 40760 40859 - 40860 40959 - 40960 41059 - 41060
(Reserved) - - - - -
ALERT High 40665 - 40666 40765 - 40766 40865 - 40866 40965 - 40966 41065 - 41066
ALERT Low 40667 - 40668 40767 - 40768 40867 - 40868 40967 - 40968 41067 - 41068
Channel 5 DANGER High 40669 - 40670 40769 - 40770 40869 - 40870 40969 - 40970 41069 - 41070
DANGER Low 40671 - 40672 40771 - 40772 40871 - 40872 40971 - 40972 41071 - 41072
(Reserved) - - - - -
ALERT High 40681 - 40682 40781 - 40782 40881 - 40882 40981 - 40982 41081 - 41082
ALERT Low 40683 - 40684 40783 - 40784 40883 - 40884 40983 - 40984 41083 - 41084
Channel 6 DANGER High 40685 - 40686 40785 - 40786 40885 - 40886 40985 - 40986 41085 - 41086
DANGER Low 40687 - 40688 40787 - 40788 40887 - 40888 40987 - 40988 41087 - 41088
(Reserved) - - - - -
60
VM-742B & VM-742 Appendix D
The measurement values of the individual module channels are assigned to this domain. They are in 32 bit
floating point format.
This is not effected by “Scaling Constant for Measure Value in 16 Bits Integer Format”.
"Measure" is the channel's primary measurement value. The overall value is output for the vibration monitor,
and the rotation speed is output for the tachometer monitor.
"Gap" is the sensor voltage.
The 0.5X, 1X, 2X or Not1X amplitude ("Amplitude") and phase ("Phase") are the amplitude and phase of the
component synchronous to the rotation frequency.
If the module is not inserted into the slot, "0" will be output.
If the analysis module has not been installed, "0" will be output.
61
VM-742B & VM-742 Appendix D
Measure Value in 32 Bits Floating Point Format (Holding Register) * From Slot 0 to Slot 5
Status Name Slot 0 Slot 1 Slot 2 Slot 3 Slot 4 Slot 5
Measure - 42201 - 42202 42401 - 42402 42601 - 42602 42801 - 42802 43001 - 43002
Gap 42003 - 42004 42203 - 42204 42403 - 42404 42603 - 42604 42803 - 42804 43003 - 43004
0.5X Amplitude - - - 42605 - 42606 42805 - 42806 43005 - 43006
0.5X Phase - - - 42607 - 42608 42807 - 42808 43007 - 43008
1X Amplitude - - - 42609 - 42610 42809 - 42810 43009 - 43010
Channel 1
1X Phase - - - 42611 - 42612 42811 - 42812 43011 - 43012
2X Amplitude - - - 42613 - 42614 42813 - 42814 43013 - 43014
2X Phase - - - 42615 - 42616 42815 - 42816 43015 - 43016
Not1X Amplitude - - - 42617 - 42618 42817 - 42818 43017 - 43018
(Reserved) - - - - - -
Measure - 42233 - 42234 42433 - 42434 42633 - 42634 42833 - 42834 43033 - 43034
Gap 42035 - 42036 42235 - 42236 42435 - 42436 42635 - 42636 42835 - 42836 43035 - 43036
0.5X Amplitude - - - 42637 - 42638 42837 - 42838 43037 - 43038
0.5X Phase - - - 42639 - 42640 42839 - 42840 43039 - 43040
1X Amplitude - - - 42641 - 42642 42841 - 42842 43041 - 43042
Channel 2
1X Phase - - - 42643 - 42644 42843 - 42844 43043 - 43044
2X Amplitude - - - 42645 - 42646 42845 - 42846 43045 - 43046
2X Phase - - - 42647 - 42648 42847 - 42848 43047 - 43048
Not1X Amplitude - - - 42649 - 42650 42849 - 42850 43049 - 43050
(Reserved) - - - - - -
Measure - 42265 - 42266 42465 - 42466 42665 - 42666 42865 - 42866 43065 - 43066
Gap - 42267 - 42268 42467 - 42468 42667 - 42668 42867 - 42868 43067 - 43068
0.5X Amplitude - - - 42669 - 42670 42869 - 42870 43069 - 43070
0.5X Phase - - - 42671 - 42672 42871 - 42872 43071 - 43072
1X Amplitude - - - 42673 - 42674 42873 - 42874 43073 - 43074
Channel 3
1X Phase - - - 42675 - 42676 42875 - 42876 43075 - 43076
2X Amplitude - - - 42677 - 42678 42877 - 42878 43077 - 43078
2X Phase - - - 42679 - 42680 42879 - 42880 43079 - 43080
Not1X Amplitude - - - 42681 - 42682 42881 - 42882 43081 - 43082
(Reserved) - - - - - -
Measure - 42297 - 42298 42497 - 42498 42697 - 42698 42897 - 42898 43097 - 43098
Gap - 42299 - 42300 42499 - 42500 42699 - 42700 42899 - 42900 43099 - 43100
0.5X Amplitude - - - 42701 - 42702 42901 - 42902 43101 - 43102
0.5X Phase - - - 42703 - 42704 42903 - 42904 43103 - 43104
1X Amplitude - - - 42705 - 42706 42905 - 42906 43105 - 43106
Channel 4
1X Phase - - - 42707 - 42708 42907 - 42908 43107 - 43108
2X Amplitude - - - 42709 - 42710 42909 - 42910 43109 - 43110
2X Phase - - - 42711 - 42712 42911 - 42912 43111 - 43112
Not1X Amplitude - - - 42713 - 42714 42913 - 42914 43113 - 43114
(Reserved) - - - - - -
Measure - 42329 - 42330 42529 - 42530 42729 - 42730 42929 - 42930 43129 - 43130
Channel 5
(Reserved) - - - - - -
Measure - 42361 - 42362 42561 - 42562 42761 - 42762 42961 - 42962 43161 - 43162
Channel 6
(Reserved) - - - - - -
62
VM-742B & VM-742 Appendix D
Measure Value in 32 Bits Floating Point Format (Holding Register) * From Slot 6 to Slot 10
Status Name Slot 6 Slot 7 Slot 8 Slot 9 Slot 10
Measure 43201 - 43202 43401 - 43402 43601 - 43602 43801 - 43802 44001 - 44002
Gap 43203 - 43204 43403 - 43404 43603 - 43604 43803 - 43804 44003 - 44004
0.5X Amplitude 43205 - 43206 43405 - 43406 43605 - 43606 - -
0.5X Phase 43207 - 43208 43407 - 43408 43607 - 43608 - -
1X Amplitude 43209 - 43210 43409 - 43410 43609 - 43610 - -
Channel 1
1X Phase 43211 - 43212 43411 - 43412 43611 - 43612 - -
2X Amplitude 43213 - 43214 43413 - 43414 43613 - 43614 - -
2X Phase 43215 - 43216 43415 - 43416 43615 - 43616 - -
Not1X Amplitude 43217 - 43218 43417 - 43418 43617 - 43618 - -
(Reserved) - - - - -
Measure 43233 - 43234 43433 - 43434 43633 - 43634 43833 - 43834 44033 - 44034
Gap 43235 - 43236 43435 - 43436 43635 - 43636 43835 - 43836 44035 - 44036
0.5X Amplitude 43237 - 43238 43437 - 43438 43637 - 43638 - -
0.5X Phase 43239 - 43240 43439 - 43440 43639 - 43640 - -
1X Amplitude 43241 - 43242 43441 - 43442 43641 - 43642 - -
Channel 2
1X Phase 43243 - 43244 43443 - 43444 43643 - 43644 - -
2X Amplitude 43245 - 43246 43445 - 43446 43645 - 43646 - -
2X Phase 43247 - 43248 43447 - 43448 43647 - 43648 - -
Not1X Amplitude 43249 - 43250 43449 - 43450 43649 - 43650 - -
(Reserved) - - - - -
Measure 43265 - 43266 43465 - 43466 43665 - 43666 43865 - 43866 44065 - 44066
Gap 43267 - 43268 43467 - 43468 43667 - 43668 43867 - 43868 44067 - 44068
0.5X Amplitude 43269 - 43270 43469 - 43470 43669 - 43670 - -
0.5X Phase 43271 - 43272 43471 - 43472 43671 - 43672 - -
1X Amplitude 43273 - 43274 43473 - 43474 43673 - 43674 - -
Channel 3
1X Phase 43275 - 43276 43475 - 43476 43675 - 43676 - -
2X Amplitude 43277 - 43278 43477 - 43478 43677 - 43678 - -
2X Phase 43279 - 43280 43479 - 43480 43679 - 43680 - -
Not1X Amplitude 43281 - 43282 43481 - 43482 43681 - 43682 - -
(Reserved) - - - - -
Measure 43297 - 43298 43497 - 43498 43697 - 43698 43897 - 43898 44097 - 44098
Gap 43299 - 43300 43499 - 43500 43699 - 43700 43899 - 43900 44099 - 44100
0.5X Amplitude 43301 - 43302 43501 - 43502 43701 - 43702 - -
0.5X Phase 43303 - 43304 43503 - 43504 43703 - 43704 - -
1X Amplitude 43305 - 43306 43505 - 43506 43705 - 43706 - -
Channel 4
1X Phase 43307 - 43308 43507 - 43508 43707 - 43708 - -
2X Amplitude 43309 - 43310 43509 - 43510 43709 - 43710 - -
2X Phase 43311 - 43312 43511 - 43512 43711 - 43712 - -
Not1X Amplitude 43313 - 43314 43513 - 43514 43713 - 43714 - -
(Reserved) - - - - -
Measure 43329 - 43330 43529 - 43530 43729 - 43730 43929 - 43930 44129 - 44130
Channel 5
(Reserved) - - - - -
Measure 43361 - 43362 43561 - 43562 43761 - 43762 43961 - 43962 44161 - 44162
Channel 6
(Reserved) - - - - -
63
VM-742B & VM-742 Appendix D
The addresses in this domain can be customized by assigning the existing address data to a desired address.
When this is done, the existing address data will be copied to the specified address.
Data can be assigned to any addresses within the ranges 10001 to 14999, 40001 to 44999 and 46001 to 49999. If
an undefined address is specified, "0" will be output.
In this way, the required data can be mapped to a contiguous address, enabling the data to be retrieved more
efficiently.
You can customize address assignment by writing "VZ-74_config.txt" of "VZ-74 Configuration File" (optional)
to VM-742.
(Point CONFIGURATION FILE FORMAT)
(Point TRANSMITTING THE CONFIGURATION FILE)
Scaling Constant for Measure Value in 16 Bits Integer Format [Read Only]
Coefficients used to convert the measurement values in 32-bit floating point format (Measure Value in 32 Bits
Floating Point Format) to the measurement values in 16-bit integer format (Measure Value in 16 Bits Integer
Format) are assigned to this domain.
You can customize the conversion coefficients by writing "VZ-74_config.txt" of "VZ-74 Configuration File"
(optional) to VM742.
(Point CONFIGURATION FILE FORMAT)
(Point TRANSMITTING THE CONFIGURATION FILE)
If you expand the original measurement value prior to converting it to a measurement value in
16-bit integer format, the data may change stepwise due to problems related to resolution.
64
VM-742B & VM-742 Appendix D
Scaling Constant for Measure Value in 16 Bits Integer Format (Holding Register)
* From Slot 0 to Slot 5
Status Name Slot 0 Slot 1 Slot 2 Slot 3 Slot 4 Slot 5
Measure Input Min - 46101 - 46102 46201 - 46202 46301 - 46302 46401 - 46402 46501 - 46502
Measure Input Min - 46103 - 46104 46203 - 46204 46303 - 46304 46403 - 46404 46503 - 46504
Measure Output Min - 46105 - 46106 46205 - 46206 46305 - 46306 46405 - 46406 46505 - 46506
Measure Output Max - 46107 - 46108 46207 - 46208 46307 - 46308 46407 - 46408 46507 - 46508
Channel 1
Gap Input Min 46009 - 46010 46109 - 46110 46209 - 46210 46309 - 46310 46409 - 46410 46509 - 46510
Gap Input Max 46011 - 46012 46111 - 46112 46211 - 46212 46311 - 46312 46411 - 46412 46511 - 46512
Gap Output Min 46013 - 46014 46113 - 46114 46213 - 46214 46313 - 46314 46413 - 46414 46513 - 46514
Gap Output Max 46015 - 46,-16 46115 - 46116 46215 - 46216 46315 - 46316 46415 - 46416 46515 - 46516
Measure Input Min - 46117 - 46118 46217 - 46218 46317 - 46318 46417 - 46418 46517 - 46518
Measure Input Min - 46119 - 46120 46219 - 46220 46319 - 46320 46419 - 46420 46519 - 46520
Measure Output Min - 46121 - 46122 46221 - 46222 46321 - 46322 46421 - 46422 46521 - 46522
Measure Output Max - 46123 - 46124 46223 - 46224 46323 - 46324 46423 - 46424 46523 - 46524
Channel 2
Gap Input Min 46025 - 46026 46125 - 46126 46225 - 46226 46325 - 46326 46425 - 46426 46525 - 46526
Gap Input Max 46027 - 46028 46127 - 46128 46227 - 46228 46327 - 46328 46427 - 46428 46527 - 46528
Gap Output Min 46029 - 46030 46129 - 46130 46229 - 46230 46329 - 46330 46429 - 46430 46529 - 46530
Gap Output Max 46031 - 46032 46131 - 46132 46231 - 46232 46331 - 46332 46431 - 46432 46531 - 46532
Measure Input Min - 46133 - 46134 46233 - 46234 46333 - 46334 46433 - 46434 46533 - 46534
Measure Input Min - 46135 - 46136 46235 - 46236 46335 - 46336 46435 - 46436 46535 - 46536
Measure Output Min - 46137 - 46138 46237 - 46238 46337 - 46338 46437 - 46438 46537 - 46538
Measure Output Max - 46139 - 46140 46239 - 46240 46339 - 46340 46439 - 46440 46539 - 46540
Channel 3
Gap Input Min - 46141 - 46142 46241 - 46242 46341 - 46342 46441 - 46442 46541 - 46542
Gap Input Max - 46143 - 46144 46243 - 46244 46343 - 46344 46443 - 46444 46543 - 46544
Gap Output Min - 46145 - 46146 46245 - 46246 46345 - 46346 46445 - 46446 46545 - 46546
Gap Output Max - 46147 - 46148 46247 - 46248 46347 - 46348 46447 - 46448 46547 - 46548
Measure Input Min - 46149 - 46150 46249 - 46250 46349 - 46350 46449 - 46450 46549 - 46550
Measure Input Min - 46151 - 46152 46251 - 46252 46351 - 46352 46451 - 46452 46551 - 46552
Measure Output Min - 46153 - 46154 46253 - 46254 46353 - 46354 46453 - 46454 46553 - 46554
Measure Output Max - 46155 - 46156 46255 - 46256 46355 - 46356 46455 - 46456 46555 - 46556
Channel 4
Gap Input Min - 46157 - 46158 46257 - 46258 46357 - 46358 46457 - 46458 46557 - 46558
Gap Input Max - 46159 - 46160 46259 - 46260 46359 - 46360 46459 - 46460 46559 - 46560
Gap Output Min - 46161 - 46162 46261 - 46262 46361 - 46362 46461 - 46462 46561 - 46562
Gap Output Max - 46163 - 46164 46263 - 46264 46363 - 46364 46463 - 46464 46563 - 46564
Measure Input Min - 46165 - 46166 46265 - 46266 46365 - 46366 46465 - 46466 46565 - 46566
Measure Input Min - 46167 - 46168 46267 - 46268 46367 - 46368 46467 - 46468 46567 - 46568
Channel 5 Measure Output Min - 46169 - 46170 46269 - 46270 46369 - 46370 46469 - 46470 46569 - 46570
Measure Output Max - 46171 - 46172 46271 - 46272 46371 - 46372 46471 - 46472 46571 - 46572
(Reserved) - - - - - -
Measure Input Min - 46181 - 46182 46281 - 46282 46381 - 46382 46481 - 46482 46581 - 46582
Measure Input Min - 46183 - 46184 46283 - 46284 46383 - 46384 46483 - 46484 46583 - 46584
Channel 6 Measure Output Min - 46185 - 46186 46285 - 46286 46385 - 46386 46485 - 46486 46585 - 46586
Measure Output Max - 46187 - 46188 46287 - 46288 46387 - 46388 46487 - 46488 46587 - 46588
(Reserved) - - - - - -
65
VM-742B & VM-742 Appendix D
Scaling Constant for Measure Value in 16 Bits Integer Format (Holding Register)
* From Slot 6 to Slot 10
Status Name Slot 6 Slot 7 Slot 8 Slot 9 Slot 10
Measure Input Min 46601 - 46602 46701 - 46702 46801 - 46802 46901 - 46902 47001 - 47002
Measure Input Min 46603 - 46604 46703 - 46704 46803 - 46804 46903 - 46904 47003 - 47004
Measure Output Min 46605 - 46606 46705 - 46706 46805 - 46806 46905 - 46906 47005 - 47006
Measure Output Max 46607 - 46608 46707 - 46708 46807 - 46808 46907 - 46908 47007 - 47008
Channel 1
Gap Input Min 46609 - 46610 46709 - 46710 46809 - 46810 46909 - 46910 47009 - 47010
Gap Input Max 46611 - 46612 46711 - 46712 46811 - 46812 46911 - 46912 47011 - 47012
Gap Output Min 46613 - 46614 46713 - 46714 46813 - 46814 46913 - 46914 47013 - 47014
Gap Output Max 46615 - 46616 46715 - 46716 46815 - 46816 46915 - 46916 47015 - 47016
Measure Input Min 46617 - 46618 46717 - 46718 46817 - 46818 46917 - 46918 47017 - 47018
Measure Input Min 46619 - 46620 46719 - 46720 46819 - 46820 46919 - 46920 47019 - 47020
Measure Output Min 46621 - 46622 46721 - 46722 46821 - 46822 46921 - 46922 47021 - 47022
Measure Output Max 46623 - 46624 46723 - 46724 46823 - 46824 46923 - 46924 47023 - 47024
Channel 2
Gap Input Min 46625 - 46626 46725 - 46726 46825 - 46826 46925 - 46926 47025 - 47026
Gap Input Max 46627 - 46628 46727 - 46728 46827 - 46828 46927 - 46928 47027 - 47028
Gap Output Min 46629 - 46630 46729 - 46730 46829 - 46830 46929 - 46930 47029 - 47030
Gap Output Max 46631 - 46632 46731 - 46732 46831 - 46832 46931 - 46932 47031 - 47032
Measure Input Min 46633 - 46634 46733 - 46734 46833 - 46834 46933 - 46934 47033 - 47034
Measure Input Min 46635 - 46636 46735 - 46736 46835 - 46836 46935 - 46936 47035 - 47036
Measure Output Min 46637 - 46638 46737 - 46738 46837 - 46838 46937 - 46938 47037 - 47038
Measure Output Max 46639 - 46640 46739 - 46740 46839 - 46840 46939 - 46940 47039 - 47040
Channel 3
Gap Input Min 46641 - 46642 46741 - 46742 46841 - 46842 46941 - 46942 47041 - 47042
Gap Input Max 46643 - 46644 46743 - 46744 46843 - 46844 46943 - 46944 47043 - 47044
Gap Output Min 46645 - 46646 46745 - 46746 46845 - 46846 46945 - 46946 47045 - 47046
Gap Output Max 46647 - 46648 46747 - 46748 46847 - 46848 46947 - 46948 47047 - 47048
Measure Input Min 46649 - 46650 46749 - 46750 46849 - 46850 46949 - 46950 47049 - 47050
Measure Input Min 46651 - 46652 46751 - 46752 46851 - 46852 46951 - 46952 47051 - 47052
Measure Output Min 46653 - 46654 46753 - 46754 46853 - 46854 46953 - 46954 47053 - 47054
Measure Output Max 46655 - 46656 46755 - 46756 46855 - 46856 46955 - 46956 47055 - 47056
Channel 4
Gap Input Min 46657 - 46658 46757 - 46758 46857 - 46858 46957 - 46958 47057 - 47058
Gap Input Max 46659 - 46660 46759 - 46760 46859 - 46860 46959 - 46960 47059 - 47060
Gap Output Min 46661 - 46662 46761 - 46762 46861 - 46862 46961 - 46962 47061 - 47062
Gap Output Max 46663 - 46664 46763 - 46764 46863 - 46864 46963 - 46964 47063 - 47064
Measure Input Min 46665 - 46666 46765 - 46766 46865 - 46866 46965 - 46966 47065 - 47066
Measure Input Min 46667 - 46668 46767 - 46768 46867 - 46868 46967 - 46968 47067 - 47068
Channel 5 Measure Output Min 46669 - 46670 46769 - 46770 46869 - 46870 46969 - 46970 47069 - 47070
Measure Output Max 46671 - 46672 46771 - 46772 46871 - 46872 46971 - 46972 47071 - 47072
(Reserved) - - - - -
Measure Input Min 46681 - 46682 46781 - 46782 46881 - 46882 46981 - 46982 47081 - 47082
Measure Input Min 46683 - 46684 46783 - 46784 46883 - 46884 46983 - 46984 47083 - 47084
Channel 6 Measure Output Min 46685 - 46686 46785 - 46786 46885 - 46886 46985 - 46986 47085 - 47086
Measure Output Max 46687 - 46688 46787 - 46788 46887 - 46888 46987 - 46988 47087 - 47088
(Reserved) - - - - -
66
VM-742B & VM-742 Appendix D
67
VM-742B & VM-742 Appendix D
The measurement values of the individual module channels are assigned to this domain. They are in 16 bit
floating point format.
The measurement values in 32-bit floating point format (Measure Value in 32 Bits Floating Point Format) are
converted to the coefficients specified in "Scaling Constant for Measure Value in 16 Bits Integer Format" and
then rounded off to 16-bit integers.
The values after the decimal point will be rounded off during rounding. In addition, values larger than 32767
will be limited to 32767, and values smaller than -32768 will be limited to -32768.
"Measure" is the channel's primary measurement value. The overall value is output for the vibration monitor,
and the rotation speed is output for the tachometer monitor.
"Gap" is the sensor voltage.
If the module is not inserted into the slot, "0" will be output.
68
VM-742B & VM-742 Appendix D
Retrieve the last time the measurement value was retrieved using either "Measure Value in 32 Bits Floating
Point Format" or "Measure Value in 16 Bits Integer Format".
Read it out simultaneously for the whole system.
69
VM-742B & VM-742 Appendix D
70
VM-742B & VM-742 Appendix D
71
VM-742B & VM-742 Appendix D
Module Status (Holding Register & Input Status) * From Slot 0 to Slot 5
Slot 0 Slot 1 Slot 2
Holding Input Holding Input Holding Input
Status Name bit bit bit
Register No. Status No. Register No. Status No. Register No. Status No.
72
VM-742B & VM-742 Appendix D
Module Status (Holding Register & Input Status) * From Slot 6 to Slot 10
Slot 6 Slot 7 Slot 8
Holding Input Holding Input Holding Input
Status Name bit bit bit
Register No. Status No. Register No. Status No. Register No. Status No.
Slot 9 Slot 10
Holding Input Holding Input
Status Name bit bit
Register No. Status No. Register No. Status No.
73
VM-742B & VM-742 Appendix D
74
VM-742B & VM-742 Appendix D
Set the time to be used by "Set Time Now". At this point, the VM-7 real time clock is not yet changed.
Read it out simultaneously for the whole system (49521-49527).
This sets the time for the VM-7 real time clock.
Before setting this, use “Preload Setting Time” to transfer the time you wish to set to VM-7.
Then, when that time arrives, specify the write command in "Set Time Now".
Then, cancel the write command.
If a valid time is not set via “Preload Setting Time”, exception response 03 will be returned.
Status Name Holding Register No. Value
Set Time Now 49531 0: Cancel write command
1: Issue write command
75
VM-742B & VM-742 Appendix D
This reads out the number of alarms (ALERT, DANGER, NOT-OK) held.
If the value is 0, the event is not stored.
A maximum of 100 events are stored. If the number of events exceeds 100, the old events are automatically
deleted.
Specify the number corresponding to the history to be retrieved using "Read Alarm History" ("Read Alarm
History" is described later in this manual). Be sure to set this prior to executing “Read Alarm History".
1 shows the newest event. The larger the Alarm Event Number, the older the event.
In addition, if 0 is written, the entire history is deleted.
This reads out the history of alarms (ALERT, DANGER, NOT-OK) occurring in the rack.
Read it out simultaneously for the whole system.
If the history specified at “Select Alarm History” does not exist, exception response 03 will be returned.
Status Name Holding Register No. Value
Alarm Event Number 49611 1 - 100
Slot Number 49612 0 - 10
Channel Number 49613 1-6
Alarm Type 49614 0 : Alert, 1 : Danger, 2 : Not-OK
Alarm Direction 49615 0 : Enter, 1 : Exit
Year-h 49616 00 - 99
Year-l 49617 00 - 99
Month 49618 1 - 12
Day 49619 1 - 31
Hour 49620 0 - 23
Minute 49621 0 - 59
Second 49622 0 - 59
76
VM-742B & VM-742 Appendix D
The communication item will be added when the /CTR option is selected. Thus, VM-7 can be controlled from
the host network (host computer). The alarm bypass, channel bypass, alarm reset, etc. can be controlled by
inputting data into the specified control status and register.
If there are two VM-742 Monitor Racks, the control signals from the host network are processed by logical OR
operation.
77
VM-742B & VM-742 Appendix D
78
VM-742B & VM-742 Warranty Provision
Warranty Provision
The warranty period for the delivered product shall be 3 years (except for special products and software, and of
which shall be 1 year) from the date of delivery from our factory.
In the event that the delivered product develops any defects within the warranty period for which our company
is liable, the defective product will be replaced or repaired in our factory at our own expense.
In the event that an engineer needs to be dispatched for repair purposes during the warranty period, travel
expenses for the said engineer will be charged in accordance with the company's regulations. The cost of repair
will be borne by our company.
However, the following are excluded from our warranty:
Other defects are considered not to be attributable to our company. This warranty applies exclusively to the
delivered product unit, and our company will not be liable for consequential damage caused either directly or
indirectly through a defect of the delivered product.
The warranty period and scope of equipment made by manufacturers other than our company shall be in
accordance with the conditions of respective manufacturer, regardless of the provisions made above.
The warranty for repaired products shall be 6 months from the date of the service on the concerned part.
All the other conditions are as stated in "1. Warranty Conditions for Newly Delivered Products ".
79
Hiroshima Factory
4-22 Yoshikawa-Kogyodanchi, Higashihiroshima
739-0153, Japan
Phone : 81-82-429-1118 Fax : 81-82-429-0804
[Service Division] E-mail : service@sst.shinkawa.co.jp