UJI KINERJA ALAT

Uji Kinerja Spektrofotometer Serapan Atom (AAS)

Sensitivity, Reprodusibility & Signal


Wavelength Accuracy
Stability
Hollow Cathode Lamp (HCL) : Hg Hollow Cathode Lamp (HCL) : Cu
Band Width (Slit Width) : 0,2 nm Concentration : 2 mg/L
Acceptance Limit for Wavelength Peak : Wavelength : 324,8 nm
1) 253,65 ± 0,3 Band Width (Slit Width) : 0,7 nm
2) 365,01 ± 0,3
3) 435,84 ± 0,3 Acceptance Criteria
4) 546,08 ± 0,3 1) Sensitivity :
5) 585,25 ± 0,3 Absorption : (0,3500 ± 0,1000) Abs
6) 640,22 ± 0,3 2) Reproducibility
7) 724,52 ± 0,3 % CV (% RSD) : ≤ 1,80 %
3) Signal Stability
Time : ≤ 5,0 % for 1 minute