PACKAGING

CODE SIZE
E2100650 6x50 ml
E2100630 6x30 ml
C2100620A 6x20 ml
C2101020D 10x20 ml
Chloride measurements are used in the diagnosis and treatment of disorders
characterized by chlorides plasmatic variation like metabolic acidosis,
hypercorticalsurrenalism and renal insufficiency .

Chlorides determination method is based on the formation of a coloured


compound between trivalent iron and thiocyanate ions released by reaction of
chloride with mercurious thiocyanate. The resulting colour intensity is
proportional to chlorine concentration in the sample

Control sera and calibrators are recommended to monitor the performance of assay
procedures: SERACAL Calibrator,SERACONTROL N, SERACONTROL P

Mercurious
method thiocyanate
working reagent READY TO USE
R 1000µL+S 10 µL reagent volume unit 1000
A Sample volume unit 10
505nm
wavelength 480-520
reaction END POINT
0 300’’
incubation 37°C 5 min
AFTER
reading time INCUBATION
ANALYTICAL
PERFORMANCES
LINEARITY From 22 up to 150 mmol/l

REAGENT STABILITY Until expire date if stored at the correct


temperature and conditions
CALIBRATION STABILITY 7 days

REFERENCE VALUES Serum-plasma 90-110 mmol/l


Urine 150-250 mmol/l