You are on page 1of 14

Progress in Organic Coatings 170 (2022) 107002

Contents lists available at ScienceDirect

Progress in Organic Coatings


journal homepage: www.elsevier.com/locate/porgcoat

Corrosion and tribocorrosion protection of AZ31B Mg alloy by a


hydrothermally treated PEO/chitosan composite coating
Luyao Xu a, Xiaojing Fu a, Huijuan Su b, Huilai Sun a, Ruichuan Li a, Yong Wan a, c, *
a
School of Mechanical Engineering, Qilu University of Technology, Jinan 250353, PR China
b
School of Chemistry and Chemical Engineering, Yantai University, Yantai 264005, PR China
c
State Key laboratory of Solid Lubrication, Lanzhou Institute of Chemical Physics, Chinese Academy of Sciences, Lanzhou 730000, PR China

A R T I C L E I N F O A B S T R A C T

Keywords: This study is aimed at the enhancement of the corrosion and tribocorrosion performance of AZ31B magnesium
Tribocorrosion (Mg) alloys after plasma electrolytic oxidization (PEO) treatment. Some PEO-treated Mg samples were coated
Corrosion with chitosan (CS) film by dip-coated method, while others were subjected to hydrothermal treatment (HT)
Plasma electrolytic oxidation
followed by CS film. The microstructure and chemical composition of coatings were investigated via X-ray
Magnesium alloy
Hydrothermal treatment
diffraction, scanning electron microscopy, and Fourier transform infrared spectroscopy. Corrosion resistance of
Chitosan the specimens was evaluated via electrochemical measurements (open circuit potential and electrochemical
impedance spectroscopy) in simulated body fluid. The tribocorrosion performance was determined by simulta­
neously recording the variation of the friction coefficient and the open circuit potential with the sliding time. The
results indicated that better corrosion and tribocorrosion performance were observed when CS layer is applied on
hydrothermally treated PEO samples than on the bare PEO samples. Thus, fabrication of hydrothermally treated
PEO surface coatings with chitosan is a significant approach to protect the Mg from corrosion and tribocorrosion,
thereby increasing the potential application of Mg alloy as biological implants.

1. Introduction [8–10]. Shi et al. [11] successfully fabricated a PEO coating containing
Ca-P on AZ31B alloy in near-neutral pH solutions and evaluated the
Compared with traditional biological implants, such as corrosion resistance and biocompatibility in Hank’s solution. Experi­
cobalt‑chromium alloys, titanium alloys, stainless steels, bio-ceramics, mental results obtained by Xia et al. [12] indicated that anti-corrosion
and polymers, the density and Young’s modulus of magnesium (Mg) properties of PEO treated samples was increased by 1 order of magni­
alloys are closer to those of natural bones, and the stress shielding effect tude in the SBF, and PEO specimen had excellent mechanical properties.
can therefore be significantly reduced at the interface between the The HA containing PEO coating fabricated by suitable oxidation times
implant material and human bones [1]. Good physical and mechanical can effectively protect AZ31 alloy substrate from corrosion by Hank’s
properties are suitable for bone repair and the promotion of human bone solution [13]. However, PEO coatings are typical in the metallurgical
growth. Mg and its alloys also have excellent biocompatibility. Mg ions bonding of their double-layer microstructure, namely a thick porous
produced by degradation can participate in many metabolic reactions in outer layer and a fairly compact and thin inner layer, to the substrate
the human body, and are also cofactors of many human enzymes [2]. [5,10]. In this case, surface pores and cracks could be free paths for
However, the major challenges of Mg-based implants include their poor corrosive media to diffuse to the interface between PEO coatings and the
corrosion and wear resistance in the human environment, which easily substrate. Thus, these pores and cracks should be sealed to achieve
results in local alkalization and a reduction in mechanical properties better corrosion protection.
[3,4]. The plasma electrolytic oxidation (PEO) surface modification Creating a thin organic/inorganic layer on a PEO coating is one of the
technique has been widely utilized to effectively overcome these ob­ most attractive methods which has been considered by many re­
stacles via the rapid formation of a ceramic coating with high hardness searchers. A variety of composite coatings were prepared on Mg alloy.
[5–7]. Moreover, the PEO technique has been reported to prominently The properties of these typical composite coatings along with the
reduce the corrosion rate of Mg alloy in simulated body fluid (SBF) method of preparation are presented in Table 1 [14–23]. Toorani et al.

* Corresponding author at: School of Mechanical Engineering, Qilu University of Technology, Jinan 250353, PR China.
E-mail address: wanyong@qlu.edu.cn (Y. Wan).

https://doi.org/10.1016/j.porgcoat.2022.107002
Received 23 March 2022; Received in revised form 23 June 2022; Accepted 23 June 2022
Available online 2 July 2022
0300-9440/© 2022 Elsevier B.V. All rights reserved.
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

Table 1
Various composite coatings and results of research conducted on hybrid coatings system.
Substrate PEO Electrolytes Hybrid coatings Coating Results Ref
preparation

AZ31 Na3PO4 + KOH + Ce PEO/silane/ Dip coating The optimal protective performance and great adhesive strength via covalent [14,15]
(NO3)3 epoxy bonding between layers
AZ31 Na3PO4 + KOH + CeO2 PEO/Epoxy-Ce Dip coating Better protective behavior of the epoxy layer due to the presence of CeO2 [16]
AZ31 Na3PO4 + KOH + La PEO/Epoxy-La Dip coating Enhanced protective property of double layer coating due to the deposit of La(OH)3 [17]
(NO3)3
MA8 Na4SiO4 + NaF PEO/PTFE Spray High corrosion-resistant performance and high hydrophobicity [18]
MA8 Na2SiO3 + NaF PEO/PA Spin coating High corrosion resistance and hydrophobicity, excellent wear resistance [19]
MA8 Na4SiO4 + NaF PEO/TFE Dip coating The best anti-corrosion property and wear resistance, lowest wettability [20]
AM50 Na3PO4 + KF + NaOH PEO/sol-gel or Dip coating Enhanced corrosion performance of PEO coating due to the sealing effect of the sol- [21]
epoxy gel or epoxy layer
AZ31 Not specifically PEO/primer Spray The best corrosion resistance due to the synergistic effect of PEO and primer layer [22]
provided
AZ31 NaAlO2 + NaOH PEO/LDHs Hydrothermal Superior corrosion protection due to the synergistic effect of corrosion inhibitors and [23]
hydrothermal treatment

[14–17] fabricated the multiple PEO/Epoxy systems to effectively composition of 3.20 wt% Al, 1.30 wt% Zn, 0.32 wt% Mn, 0.02 wt% Si,
enhance the anti-corrosion performance of Mg alloys in corrosive media. 0.003 wt% Fe, 0.002 wt% Ca, 0.002 wt% Cu, 0.001 wt% Ni and balance
Additionally, some other polymer materials, such as fluoroparaffin (PA), Mg. All Mg alloys were purchased from Lianwei Technology (Dongguan)
polytetrafluoroethylene (PTFE), and tetrafluoroethylene (TFE), were Co., Ltd. (China). The alloy sheets were cut into regular pieces with a
also widely used in the sealing of PEO coatings [18–20]. Especially, dimension of 35 mm × 35 mm × 2 mm. The specimens were ground with
evaluation of the corrosion and wear resistances of composite coatings SiC sandpaper to Ra = 0.1 μm.
reflected a great decrease in corrosion current density (6.3 × 10− 9 A/
cm− 2) and wear rate (4.6 × 10− 5 mm3 N− 1 m− 1) of composite coatings 2.2. Preparation of PEO coatings
by 2 orders of magnitude in comparison with the PEO coating [20]. Sol-
gel process is also a proven and simple technique to seal the pores of PEO All chemical agents were purchased from Sinopharm Chemical Re­
coating [21,22]. Moreover, layered double hydroxides (LDHs) were agent Co., Ltd. (China). The alkaline electrolytes solution for the PEO
prepared to improve the anti-corrosion property of PEO coating [23]. process was composed of 10 g/L Na2SiO3⋅9H2O, 5 g/L Na2B4O7⋅10H2O
More information could also be found in a recently published review and 1 g/L NaOH. The pH and conductivity values of the electrolytes
[24]. solution were 13.2 and 17.3 mS/cm, respectively. The PEO procedure
Given their biological applications, biomedical polymer materials was performed at a constant current of 1.7 A for 20 min. The aqueous
are employed as an ideal choice to seal PEO-treated samples [25]. Chi­ solution temperature was controlled at around 30 ◦ C. Other parameters
tosan (CS) has been universally investigated for use in human clinical on the PEO instrument panel were also set: 25 % duty cycle, 300 Hz
materials due to its superb biodegradability, biocompatibility, film- frequency. The 304 annular tube and clamped Mg alloy sheet were used
forming ability, and non-toxicity, among other characteristics [26]. as the cathode and the anode respectively. After PEO treatment, the
Dou et al. [27] reported that CS coating could seal the pores of PEO specimens were thoroughly rinsed with deionized water and dried in air
coatings, and exhibited a lower corrosion rate in SBF. Bai et al. [28] at room temperature. The samples are denoted as PEO.
found that a composite PEO/CS coating significantly increased the bio-
corrosion resistance of Mg alloy. Yu et al. [29] investigated the in vitro 2.3. Hydrothermal treatment of PEO coatings
degradation behavior of a PEO coating on Mg alloy after being sealed by
a CS coating. The homogeneous hydrothermal solution consisted of 1.5 mol/L
Hydrothermal treatment (HT) is an effective, inexpensive and envi­ NaOH. PEO sample was placed in a stainless steel container with built-in
ronmentally friendly sealing technique that enhances the corrosion polytetrafluoroethylene reactor, and then completely immersed in the
resistance of PEO coatings. Dey et al. [30] revealed that the anti- above 80 mL hydrothermal solution. Subsequently, the reaction
corrosion property of a PEO sample sealed in boiled water was container was heated to 125 ◦ C with a muffle furnace at a heating rate of
evidently improved as compared to that of an unsealed PEO sample. An 5 ◦ C/min, and kept for 18 h. After the hydrothermal treatment, the
experimental analysis conducted by Liu et al. [31] revealed that a PEO sample was slightly rinsed with deionized water several times, and
coating treated with steam sealing at 453 K for 2 h exhibited the most finally dried in air at room temperature. These samples are marked as
remarkable anti-corrosion property. Thus, the corrosion resistance of PEO/HT.
PEO-treated Mg is expected to be extended via the combination of HT
treatment with a CS coating. In this study, a CS layer was coated on a
2.4. Fabrication of CS coatings
PEO-treated Mg alloy after HT treatment, and the anti-corrosion per­
formance of the resulting composite PEO/HT/CS coating was compared
PEO and PEO/HT samples were subjected to CS coating. 1 g chitosan
with that of a PEO/CS coating. Moreover, many joint parts in bio-
powder (Mw = 270 kDa, >90 % deacetylated, Aladdin) was dissolved
implants are often affected by both corrosion and wear. However,
per 100 mL 0.4 vol% acetic acid solution. The CS solution was
there has been no research on the tribocorrosion behavior of CS coat­
magnetically stirred for 12 h until the chitosan was completely dis­
ings. Thus, the second aim of this work is to investigate the tribocorro­
solved. 1 mol/L NaOH was dropping in the CS solution to change the pH
sion protection performance of the composite coatings.
from about 4.0 to 6.5–7.0 during the process of the next three-hour
stirring. Immediately afterwards, the coated specimen was immersed
2. Experimental details
perpendicularly in the CS solution for 5 min, then the sample was drawn
out at a constant rate of 2 mm/s with the lifting coating machine. The
2.1. Materials
above treatment was repeated five times to ensure that micro-pores on
the sample surface were completely sealed. Finally, the specimen was
The used substrate materials are AZ31B Mg alloys with a chemical
naturally dried at 25 ◦ C for 120 h. The PEO and PEO/HT coated with CS

2
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

Table 2
Chemical composition of the SBF (pH 7.4, 1 L).
Reagent NaCl NaHCO3 KCl K2HPO4⋅3H2O MgCl2⋅6H2O CaCl2 Na2SO4 1 mol/L HCl

Amount (g) 8.035 0.355 0.225 0.231 0.311 0.292 0.072 39 mL

coatings are denoted as PEO/CS and PEO/HT/CS, respectively. goniometer (JC2000D1, China) at the room temperature. The volume of
water droplet was 5 μL.
According to ASTM D3359-09, the adhesion tester (QFH-HD600,
2.5. Characterization of samples China) was utilized to examine the adhesion strength between the
coating and Mg alloy substrate [32]. The surface roughness of coatings
The phase structures of the coatings were analyzed by X-ray were detected using portable surface roughness tester (SJ-210, Mitu­
diffraction (XRD; D8 ADVANCE, Bruker, Germany) with Cu Kα radia­ toyo, Japan) with an accuracy of 0.001 μm at a probe sliding rate of 0.25
tion, 40 kV tube voltage and 40 mA output current. The chemical mm/s. In order to ensure the reliability of the micro-hardness testing,
bonding of the specimens was probed using Fourier transform infrared the rough surface of the PEO-coated specimen was slightly ground with
spectroscopy (FT-IR, INVENIO-R, Bruker, Germany). The scanning 1200# SiC sandpaper to obtain an average surface roughness of Ra ≈
electron microscope (SEM, Merlin-Compact, Germany) was utilized to 0.5 μm, and then the hardness testing was carried out by digital Vickers
investigate the surface and cross-section morphologies of the coatings. tester (FUTURE-TECHFM-700, Japan). In the tests of hardness and
The chemical compositions of the specimens were examined using en­ roughness, 10 places were randomly investigated to obtain the average
ergy dispersive X-ray spectrometry (EDS) with 15 kV accelerating value and standard deviation.
voltage. The water contact angle (CA) was determined utilizing the CA

Fig. 1. The surface morphologies of the (a) PEO, (b) PEO/HT, (d) PEO/CS, and (e) PEO/HT/CS samples with individual enlarged parts.

3
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

Table 3
The EDS analysis of the surfaces of different samples (wt%).
Sample (region) O Mg Al Zn Si Na B C N

PEO (#1) 46.41 25.73 0.76 0.46 16.21 1.48 8.95 – –


PEO/HT (#2) 47.38 26.58 0.70 0.34 17.12 1.14 6.74 – –
PEO/CS (#3) 50.25 0.15 0.02 0.05 0.13 7.42 0.06 31.59 10.33
PEO/HT/CS (#4) 50.75 0.12 0.04 0.03 0.10 6.91 0.04 30.88 11.13

2.6. Electrochemical measurements testing was specifically set: normal load of 1 N, track length of 5 mm,
vibration frequency of 1 Hz. The three-electrode cell containing the SBF
The open circuit potential (OCP) and electrochemical impedance was heated to a temperature of 36.5 ± 0.5 ◦ C by the built-in heating
spectra (EIS) measurements were utilized as an effective and convenient system. The OCP value was detected by the tribo-electrochemical sys­
method to analyze the corrosion behavior of the coatings on Mg alloys in tem, where the testing specimen, an Ag/AgCl electrode and a carbon
the simulated body fluid (SBF) with the chemical components listed in electrode as the working electrode, reference electrode and auxiliary
Table 2 [33]. These electrochemical tests were conducted using a stan­ electrode, respectively. The OCP value was recorded 10 min before, 30
dardized corrosion testing system (CHI660E, China). The SBF was min during and 10 min after the sliding. Topographies and elements
buffered at pH 7.4 using tris(hydroxymethyl) aminomethane from the wear tracks were analyzed by SEM and EDS. Wear rate of the
((CH2OH)3CNH2) and 1.0 mol/L hydrochloric acid (HCl). specimens was attained by dividing the wear volume by the normal load
The corrosion testing system consists of a conventional three- and the distance moved [34]. Each testing was repeated three times to
electrode cell with a testing specimen with an exposure area of 4.80 minimize accidental errors.
cm2 as the working electrode, an Ag/AgCl electrode as the reference
electrode, a carbon electrode as the auxiliary electrode, respectively. 3. Results and discussion
The testing cell was filled with the SBF of 40 mL in each measurement.
Before EIS measurements, OCP measurements were performed firstly, 3.1. Surface and cross-sectional morphologies
and EIS measurements started only when the OCP values kept stable. EIS
measurements were carried out at the OCP in a frequency range of Fig. 1 displays the surface morphologies of the samples after different
10− 2–105 Hz with 10 points/10 frequency multiplication, and the signal treatments. The surface of the PEO-treated sample (Fig. 1(a)) reveals the
amplitude of voltage sinusoidal disturbance was 10 mV RMS. The three- existence of randomly distributed volcano-like micro-pores of various
electrode testing cell was used and the temperature of the SBF was sizes and micro-cracks penetrating the pores. The discharge channels,
remained at 36.5 ± 0.5 ◦ C. All the tests were repeated three times and namely micro-pores, were developed in large quantities due to the
the value was averaged. partial spark discharge and gas overflow during the coating growth [31].
In addition, the micro-cracks are supposed to be caused by inhomoge­
2.7. Tribocorrosion tests neous thermal stress distribution during the molten substance cooling
process [35].
A reciprocating ball-on-disk tribometer (MFT-EC4000, China) was The surface morphological features of the PEO samples after HT, as
utilized to evaluate the tribocorrosion performance of the specimens illustrated in Fig. 1(b), notably indicate the appearance of hydration as
against a 9.525 mm diameter Al2O3 ball. The working condition of wear compared with the PEO samples. Dense, short, hexagonal columnar

Fig. 2. The cross-sectional morphologies and element mappings of Mg, O, Si, and C for (a) PEO, (b) PEO/HT, (c) PEO/CS, and (d) PEO/HT/CS samples.

4
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

Si-O-Si
5 Mg(AlH4)2 7 Mg3(BO3)2

Mg-OH
3 Al2O3 9 chitosan

N-H
1 Mg

C-O-C
N-H
C-H

C-N
C=O
C-H

Mg-O
2 MgO 4 Mg2SiO4 6 Mg2B2O5 8 Mg(OH)2
5
1
1 PEO
4 6 2 5
4 4 4 7 7 PEO/HT
4

Transmittance (%)
4 44 43 1 4

3695
PEO/CS

O-H

1010
890
Intensity(a.u.)

PEO 1
PEO/HT/CS
1 6 7 7
4 4 84 2 5

515
84 4 4
43 1 4 8 4 5

C-N
PEO/HT 1
chitosan powder
9 4 4 4 1 6 7 7
4 2 1 4 4
5 5
1

1163
1660
PEO/CS

3366

2924

1418
1572
2863
89 4 4 4 18 6 7 7 5

1064
4 2 4 5

1026
1 8 4
PEO/HT/CS
4000 3600 3200 2800 2400 2000 1600 1200 800 400
10 20 30 40 50 60 70 80
-1
2-Theta/degree Wavenumber (cm )

Fig. 3. The XRD patterns of the PEO, PEO/HT, PEO/CS, and PEO/HT/ Fig. 4. The FT-IR spectra of the PEO, PEO/HT, PEO/CS, and PEO/HT/
CS samples. CS samples.

crystals were found to be distributed on the surface of the PEO/HT 3.2. Chemical compositions
coating under high-magnification SEM (see Fig. 1(c)). The hexagonal
crystals precipitated around or inside the micro-pores were character­ As is evident from the XRD spectra of the samples obtained using
ized as Mg(OH)2 crystals [36], which was further confirmed by XRD and different treatments shown in Fig. 3, all samples had clear Mg peaks,
FT-IR analyses, as presented in Section 3.2. which is ascribed to X-rays penetrating the coating to reach the under­
Fig. 1(d) reveals that the porous microstructures of the PEO-treated lying Mg alloy substrate [37]. The diffraction patterns of all specimens
samples were completely sealed by the CS layer. In particular, no demonstrate the common existence of MgO, Mg2SiO4, Mg(AlH4)2,
obvious defects in the appearance of the samples, such as micro-pores or Mg2B2O5, and Mg3(BO3)2 crystalline phases, excluding the Mg peaks
micro-cracks, were found. This was also observed for the PEO-HT- from the alloy substrate, which were due to the complicated plasma
treated samples after being covered with the CS layer (Fig. 1(e)). chemical oxidation reactions during the spark discharge process
The EDS analysis of the different specimens, as presented in Table 3, [7,38,39]. In the case of the PEO/HT and PEO/HT/CS coatings, the
reveals that the PEO coatings were mainly composed of O, Mg, Si, and B, typical diffraction peaks of Mg(OH)2 at approximately 2θ = 18◦ , 38◦ ,
which is not a significantly different composition from that of the PEO/ and 59◦ confirm the formation of Mg(OH)2 crystals during the HT pro­
HT coating. The result also illustrates that the formation of the PEO cess. As can be seen, the XRD spectra of the PEO/CS and PEO/HT/CS
coating was associated with both the deposition of electrolyte com­ coatings each display a broad peak at about 2θ = 20◦ corresponding to
pounds and the self-oxidation of the alloy substrate [31]. Primarily O, C, CS [27,40].
and N were found to exist in the CS layer. Some Na was also observed in Fig. 4 presents the IR spectra of the samples obtained using different
the CS layer, which is ascribed to the addition of NaOH in the CS treatments. For the PEO and PEO/HT coatings, the adsorption peaks at
solution. around 1010 and 890 cm− 1 are respectively ascribed to the asymmetric
Fig. 2 exhibits the cross-sectional SEM images and element distri­ and symmetric stretching vibration of Si–O–Si [41,42], which strongly
bution in the cross-sections of the different samples. The thickness of the confirms the existence of Mg2SiO4. A weak absorption band located at
PEO coating was found to be approximately 38 μm, which is the same as about 515 cm− 1 was due to Mg–O stretching vibration, implying the
that of the PEO/HT coating, demonstrating that the HT process did not growth of MgO in the PEO-treated coating [43]. Additionally, an
change the thickness of the PEO coating. A hierarchical structure was adsorption peak at 3695 cm− 1 was detected for the PEO/HT coating,
obvious in the cross-sectional images of the PEO/CS and PEO/HT/CS which corresponds to the high basicity of the O–H groups in Mg(OH)2
coatings. The CS layer was approximately 15 μm thick. The EDS map­ [31,44].
ping analysis presented in Fig. 2(a) and (b) reveals that Mg, O, and Si Accordingly, the HT process can be described by the following re­
were uniformly dispersed in the PEO and PEO/HT coatings. This result is actions [31,43].
consistent with the EDS analysis (see Table 3). However, the B element Mg2+ (aq) + 2OH− (aq)→Mg(OH)2 (s) (1)
was not found due to the unclear EDS mapping image, which may be
related to the relatively low content of B. As clearly shown in Fig. 2(c) MgO (s) + H2 O (g)→Mg(OH)2 (s) (2)
and (d), the discernible and even distribution of C and O provides evi­
2+
dence of the existence of the CS layer on the PEO-treated samples. On the one hand, the Mg derived from compounds in the coating
The roughness of different samples was also measured. The charac­ combines with OH− in the solution to produce Mg(OH)2. On the other
teristic surface topography of the PEO specimen led to a higher surface hand, boiled water can transform MgO to Mg(OH)2 due to the hydration
roughness (Ra ≈ 1.66 μm) as compared to that of the uncoated substrate reaction between MgO and H2O.
(Ra ≈ 0.12 μm). The PEO/HT sample had a relatively low roughness (Ra CS is a linear polysaccharide consisting of β-(1 → 4)-linked N-acetyl-
≈ 1.32 μm) as compared to the PEO sample, which may have been due to D-glucosamine (the deacetylated part) and N-acetyl-D-glucosamine (the

the surface sealing effect of Mg(OH)2 crystals. The roughness testing acetylated part) [45,46]. The characteristic wide absorption bands
results also reveal that the mean roughness values (Ra) of the PEO/CS located at 3600–3300 cm− 1 represent the stretching vibration of non-
and PEO/HT/CS samples were respectively 0.72 and 0.64 μm. associated O–H and N–H [47,48]. The shoulder peaks at 2863 and
2924 cm− 1 are associated with the stretching vibration of C–H in
methyl or methenyl [49]. The adsorption band at 1660 cm− 1 is ascribed
to the stretching vibration mode of C– – O (amide I) [50], and the ab­
sorption bands at 1572 (amide II) and 1418 cm− 1 correspond to the

5
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

Fig. 5. The static CA for the (a) Mg substrate, (b) PEO, (c) PEO/HT, (d) PEO/CS and (e) PEO/HT/CS samples.

Fig. 6. The optical images of (a) PEO, (b) PEO/HT, (c) PEO/CS, and (d) PEO/HT/CS samples after the adhesion test.

unique N–H [51]. The characteristic absorption peaks at 1163 and 3.3. Surface wettability
1026 cm− 1 are associated with the stretching vibration of C–N, and the
absorption band located at 1064 cm− 1 represents the typical anti- Fig. 5 exhibits optical images of water droplets on the surfaces of
symmetric stretching of the C–O–C bridge [48]. The XRD and FT-IR different samples. The surface of the bare Mg substrate had a CA of 56.5◦
analysis results consistently indicate that the CS layer was successfully ± 1.5◦ (Fig. 5(a)), revealing a hydrophilic nature. As can be seen, the
deposited on the PEO-treated coating. PEO coating reflects an evidently hydrophilic feature with a static CA of
35◦ ± 0.5◦ (Fig. 5(b)), which is in accordance with that reported in the
existing literature [52,53]. The static CA on the surface of the PEO/HT
sample was 10◦ ± 0.5◦ (Fig. 5(c)), which caused the water droplet to

6
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

-1.25 120000
Substrate
PEO (a) PEO
-1.30 PEO/HT 105000 300
PEO/HT
Potential (V vs.Ag/AgCl)

PEO/CS 250
PEO
PEO/HT PEO/CS
PEO/HT/CS PEO/CS
-1.35 90000 200
PEO/HT/CS PEO/HT/CS

-Z /(ohm cm )
Substrate

2
150
Fit Result Substrate

-Z /(ohm cm )
2
100

-1.40
75000 Fit Result
50

0
60000
-1.45 -50

-100
0 50 100 150 200 250 300 350 400
45000 Z /(ohm cm )
2

-1.50

30000
-1.55
0 200 400 600 800 1000 1200 1400 1600 1800
15000
Time (s)

Fig. 7. OCP variation with time of immersion in SBF for substrate and 0
coated specimens. 0 20000 40000 60000 80000 100000 120000
2
Z /(ohm cm )
diffuse on the surface and rapidly penetrate the coating. As reported by
Jin et al. [54], a super-hydrophilic Mg(OH)2 film prepared on AZ31B Mg 5 (b) PEO
alloy via HT treatment was found to have a static CA of ≤8◦ . It can be 10 PEO/HT
considered that staggered flaked Mg(OH)2 crystals were generated on PEO/CS
PEO/HT/CS
the surface of the PEO coating during HT, giving rise to a more hydro­ 4 (a)
Substrate
10
philic surface with a large number of hydroxyl groups [54]. Regarding 2 Fit Result
Z /ohm cm
the PEO/CS and PEO/HT/CS coatings, the water CA values were
respectively 72◦ ± 1◦ and 72.5◦ ± 0.5◦ , thereby reflecting a hydrophilic 10
3

characteristic.

3.4. Mechanical and adhesion performances 10


2

The micro-hardness values of the coating surfaces were measured.


The micro-hardness of the PEO-treated specimens was significantly
1
10
-2 -1 0 1 2 3 4 5
higher than that of the alloy substrate (55.3 HV0.1). The hardness of the 10 10 10 10 10 10 10 10
Frequency/Hz
PEO/HT coating was about 526.7 HV0.1, which was slightly lower than
that of the PEO coating (552.4 HV0.1). This result also corresponds to the
fact that the staggered growth of Mg(OH)2 crystals in the micro-pores 60 (c)
will cause the overall decrease of the hardness of the PEO coating
[30]. The micro-hardness measurements revealed mean values of 31.8 40
and 32.5 HV0.1 for the PEO/CS and PEO/HT/CS samples, respectively.
-Theta /degree

The adhesion strength of coatings is a critical index of their practical


usage. Fig. 6 presents the surface topographies of all the coated samples
20

after adhesion testing according to the ASTM D3359-09 standard. The


results indicate that the PEO and PEO/HT coatings exhibited no distinct 0 PEO
chipping off at the edge of the square lattice, so the rank of these coat­ PEO/HT
PEO/CS
ings was identified as 5B. The region encircled by the red lines signifies -20 PEO/HT/CS
the peeling areas of the coating. It can be observed in Fig. 6(c) and (d) Substrate
that the peeling ratio of the PEO/CS coating was about 41 % (class 1B), Fit Result
-40
whereas that of the PEO/HT/CS coating was merely about 12 % (class 10
-2
10
-1
10
0
10
1
10
2 3
10 10
4
10
5

3B). This is because Mg(OH)2 grew on the surface of the PEO/HT coating Frequency/Hz
to form a super-hydrophilic surface with plenty of hydroxyl groups
(Fig. 5(c)), which is beneficial to the diffusion and grafting of the CS Fig. 8. The EIS and fitting results of the substrate and coated samples: (a)
solution into the interior of the coating to form a CS layer with high Nyquist plots; (b) Bode plots of |Z| vs. frequency; (c) bode plots of phase angle
adhesion strength. Secondly, due to the high viscosity and poor fluidity vs. frequency.
of the CS solution, some micro-pores were observed at the junction be­
tween the CS layer and the PEO-treated coating. Moreover, Mg(OH)2 can be used to monitor the chemical stability and corrosion process of
crystals filled most of the holes to a certain extent, which increased the specimens during immersion time. Fig. 7 exhibits the evolution of the
contact area between the base material and the CS layer, thereby OCP during 30 min exposure in SBF. The OCP of bare substrate
improving the adhesion strength. approached to the stable value of − 1.51 V vs. Ag/AgCl after showing a
slight increase. The stabilization of the potential demonstrates the for­
3.5. Electrochemical measurements mation of corrosion products acting as a barrier layer [55].
If this variation was compared once bare substrate was oxidized, the
3.5.1. Open circuit potential (OCP) measurements PEO coating shifted the potential to the positive direction reflecting an
The variation of open circuit potential during the immersion time irregular fluctuation. The OCP value initially rose rapidly and then

7
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

Fig. 9. Equivalent circuit diagrams of the EIS plots: (a) substrate, (b) PEO, (c) PEO/HT, (d) PEO/CS, and (e) PEO/HT/CS samples in SBF.

decreased sharply, and started to oscillate violently between − 1.36 and until potentials remained stable at the values of − 1.49 and − 1.41 V vs.
− 1.42 V vs. Ag/AgCl. This may be due to that corrosive ions penetrated Ag/AgCl after 10 min, respectively. It could suggest that the composite
through the loose layer reaching the interface between the inner dense coating adsorbed electrolyte, thereby reaching a stable and saturated
layer and the outer porous layer [56]. Subsequently, the OCP value state [56]. The above results and corrosion behavior can be testified
gradually decreased and then stabilized at approximately − 1.41 V vs. further by the following EIS analysis.
Ag/AgCl. In term of PEO/HT sample, the OCP with lower amplitude
dropped gradually and regularly from around − 1.28 to − 1.37 V vs. Ag/ 3.5.2. Electrochemical impedance spectroscopy (EIS) analysis
AgCl. In addition, for PEO and PEO/HT specimens, the slight fluctuation EIS was carried out to further analyze the anti-corrosion properties of
of stable OCP values occurred during soaking stage, indicating the het­ the coated specimens. Fig. 8 exhibits the Nyquist and Bode plots of the
erogeneous reactions, which may be related to the non-uniform Mg substrate and the PEO, PEO/HT, PEO/CS, and PEO/HT/CS speci­
composition and structure of the outer layer of the PEO-treated mens immersed in SBF for 30 min. As presented in the Nyquist plots
coating [57]. (Fig. 8(a)), two visible capacitive loops were found at the high- and low-
In contrast, both PEO/CS and PEO/HT/CS coatings showed a smooth frequency regions for the coated specimens, which respectively denote
evolution of the OCP with immersion time. The OCPs gradually dropped the impedance characteristics of the outer layer and the inner dense

8
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

Table 4
EIS fitting parameters of equivalent circuits for the substrate and coated samples.
Samples Rs (Ω cm2) CPEpo-T CPEpo-P Rpo CPEdl-T CPEdl-P Rct RL L
(10− 6 Ω− 1
cm− 2 n
s ) (Ω cm2) (10− 5 Ω− 1
cm− 2 n
s ) (Ω cm2) (Ω cm2) (H cm2)

Substrate 76 – – – 4.85 0.93 236 99 174


PEO 53 1.54 0.74 5702 2.45 0.71 48,632 – –
PEO/HT 41 1.27 0.71 5863 2.03 0.76 65,527 – –
PEO/CS 36 0.82 0.66 5936 1.62 0.73 96,507 – –
PEO/HT/CS 51 0.34 0.69 7863 0.99 0.74 117,550 – –

Fig. 10. Schematic illustrations of the proposed corrosion resistance mechanisms of (a) PEO-treated samples and (b) CS-coated samples.

layer/substrate interface. The two time constants can be observed in the passivation coating formed due to corrosive pitting, which is linked in
Bode plots of the coated samples (Fig. 8(c)), which also corroborate with series with the inductance L. In particular, CPE belongs to the interface
the characteristics of the Nyquist plots in Fig. 8(a). It is universally diffusion constant phase element, which replaces the conventional
agreed that the anticorrosive properties of the tested specimens can be capacitance due to the roughness and inhomogeneity of the specimen
evaluated by the axial curvature radius of the low-frequency capacitive surface [63,64].
loop [31,58]. The maximum curvature radius of the sample reflects the The corresponding fitting parameters of the EIS spectra based on
most excellent corrosion resistance [58]. As is evident, the PEO/HT/CS equivalent circuits are reported in Table 4. After the PEO samples were
samples had the most remarkable anti-corrosion performance. subjected to HT and CS coating, the Rpo value increased slightly, but the
The Nyquist plot of the uncoated substrate was magnified for clarity Rct value was much higher than the corresponding Rpo value; thus, the
(see Fig. 8(a)), and a similar corrosion behavior in Mg alloys has pre­ anti-corrosion effect of the coating was found to be primarily dominated
viously been reported [59]. A capacitive loop was found in the high- and by the Rct value [8,62]. As exhibited in Table 4, the increase of Rct from
medium-frequency region, which is associated with the charge transfer 236 to 48,632, 65,527, 96,507, and 117,550 Ω⋅cm2 corroborates that the
behavior of the electric double layer formed between the Mg alloy and PEO/HT/CS specimens most effectively improved the anticorrosive ef­
SBF. In addition, a low-frequency inductive loop was distinctively found fect of the Mg alloy substrate. These experimental results are also veri­
to take shape, which is supposed to be caused by the pitting corrosion of fied in other similar reports [29,38].
the Mg alloy [31,59,60]. The phase angle of the uncoated Mg alloy was According to the testing results, the PEO/HT/CS coating can mark­
also found to have a negative value in the low-frequency region, indi­ edly improve the corrosion resistance of the Mg substrate. Fig. 10
cating the occurrence of corrosive pitting on the Mg alloy substrate [59]. schematically illustrates the protective mechanisms of the PEO-treated
The value of |Z|f→0 in the Bode plots (Fig. 8(b)) can be used as an and CS-coated coatings based on the preceding analyses. For the PEO
effective indicator to examine the corrosion rate of the coating; the coating, surface pores and cracks could be free paths for a corrosive
higher the value of |Z|f→0, the lower the corrosion rate [61]. The order of medium like Cl− to diffuse to the interface between the PEO coating and
the |Z| values of different specimens at 0.01 Hz, from greatest to least, the substrate, as exhibited in Fig. 10(a). The local corrosion of Mg
was found to be as follows: |Z|PEO/HT/CS (108,580 Ω⋅cm2) > |Z| PEO/CS occurred in accordance with the following reaction [65].
(76,424 Ω⋅cm2) > |Z| PEO/HT (64,286 Ω⋅cm2) > |Z| PEO (50,162 Ω⋅cm2)
Mg (s) + 2H2 O (g)→Mg(OH)2 (s) + H2 (g)↑ (3)
> |Z| substrate (147 Ω⋅cm2). This signifies that the PEO/HT/CS specimens
had the lowest corrosion rate. These results are in good agreement with Because Mg(OH)2 has a larger molar volume than MgO, the contin­
the Nyquist plots. uous generation of Mg(OH)2 will constantly damage the PEO coating
According to the EIS results and the related literature [8,62], the [66]. Regarding the PEO/HT coating, the inherent defects such as pores
electric circuit fitting results are shown in Fig. 9, where Rs represents the and cracks can be sealed due to the superior expansion and sealing effect
SBF resistance between the workpiece and the reference electrode. Rpo of the Mg(OH)2 growth process [43,67]. Thus, the anti-corrosion effect
and CPEpo are connected in parallel, and respectively characterize the of the PEO coating was further enhanced by HT. The PEO/HT/CS
outer layer resistance and the dispersed capacitance behavior of the coating can effectively prevent the arrival of Cl− ions on the Mg sub­
outer layer within the micro-pores. Moreover, CPEdl corresponds to the strate due to the collaborative protection of the PEO/HT coating and the
electric double-layer capacitance, and Rct illustrates the charge transfer compact CS layer, as exhibited in Fig. 10(b).
resistance of the inner dense layer/substrate interface. For the Mg alloy
substrate, Rct characterizes the charge transfer resistance of the SBF/
substrate interface, which is parallel to CPEdl. RL is the resistance of the

9
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

-1.15 rupture of the passivation film, which led to the exposure of the fresh
-1.20 (a) PEO substrate to the SBF [68]. Then, the potential stabilized at around − 1.63
-1.25
PEO/HT
V vs. Ag/AgCl and oscillated with the rubbing movement, which is
PEO/CS
-1.30 PEO/HT/CS associated with the dynamic equilibrium of OCP between the mechan­
OCP/(V vs. Ag/AgCl)

-1.35 Untreated ical destruction (de-passivation) and the passive film formation (re-
-1.40 passivation) in the rubbing interface region. After sliding, the OCPs
-1.45
began to increase, but the final potential was slightly lower than that
-1.50
recorded before sliding.
The drop in potential also occurred for the PEO and PEO/HT samples
-1.55
at the onset of rubbing, suggesting the local rupture or wear of the
-1.60
rubbed zone [69]. During the sliding phase, the OCPs of the PEO and
-1.65
PEO/HT samples fluctuated violently at approximately − 1.44 and
-1.70
− 1.39 V vs. Ag/AgCl, respectively. However, unique OCP variation was
-1.75 Soaking Reciprocating sliding Soaking detected for the PEO/CS and PEO/HT/CS specimens. The OCPs did not
decline sharply at the onset of the rubbing motion, but rather progres­
-1.80
0 10 20 30 40 50
Time/min sively decayed with the continuation of the sliding motion. After rub­
bing for ~10 min, the OCPs of the PEO/CS and PEO/HT/CS samples
0.60
PEO
respectively stabilized at − 1.53 and − 1.48 V vs. Ag/AgCl, but that of the
0.55 (b) PEO/HT PEO/CS sample exhibited a slight decreasing trend in the later period.
0.50 PEO/CS Moreover, no intense oscillation of OCP occurred during the sliding,
0.45 PEO/HT/CS suggesting the excellent stability between the mechanical de-passivation
and electrochemical re-passivation [70]. After the sliding ended, the
Untreated
0.40
0.35 OCPs of both samples returned to the initial values.
Fig. 11(b) displays the variation of the COFs. The COFs of the PEO-
COF

0.30
treated specimens were markedly higher than that of the untreated
0.25
substrate, which is ascribed to the holey surface structure obtained by
0.20
the PEO technique. During the entire rubbing process, the COF of the
0.15
PEO/HT sample was more stable and lower than that of the PEO sample,
0.10 but the wear rate was slightly higher than that of the PEO sample (see
0.05
Soaking Reciprocating sliding Soaking
Fig. 11(c)). This is attributed to the slight decrease in the surface
0.00 roughness and hardness of the PEO specimen after HT. The application
0 10 20 30 40 50 of the CS layer clearly decreased the friction and wear rate of the PEO
Time/min and PEO/HT samples. The most remarkable case was that of the PEO/
HT/CS specimen, for which the average COF value was the lowest
18
0.35 (~0.11) and the wear rate was also reduced to the lowest value (~1.83
16
(c) × 10− 5 mm3 N− 1 m− 1) as compared with the untreated substrate (~1.13
× 10− 4 mm3 N− 1 m− 1).
0.30
14
Wear rate/ [10 mm /(N m)]

0.25
Fig. 12 exhibits the SEM morphologies of wear tracks after tribo­
12 corrosion testing. Regarding the uncoated substrate (Fig. 12(a)), a
Average COF
3

10 0.20 typical topography with numerous abrasive furrows at the worn zone
was distinctly observed, which is recognized as a typical abrasive wear
-5

mechanism [71]. For the PEO and PEO/HT samples, a smooth wear
8
0.15
6 track was observed, suggesting that the pores and cracks at the wear
0.10 marks were filled and compacted by the delaminated debris [72]. In the
4
case of the PEO/CS and PEO/HT/CS samples, plenty of micro-cracks and
scratches were clearly detected at the wear tracks. However, the plastic
0.05
2
deformation was more serious for the PEO/CS sample than for the PEO/
0 0.00
Untreated PEO PEO/HT PEO/CS PEO/HT/CS HT/CS sample, and the width of the wear track (662 μm) of the PEO/CS
Samples sample was evidently wider than that of the PEO/HT/CS sample. When
the Al2O3 ball was rubbed against the CS layer, the surface of the CS
Fig. 11. The evolution of the (a) open circuit potential, (b) coefficient of fric­ layer was prone to propagating crack generation. The high adhesion
tion (COF) with the sliding time, and (c) average COF and wear rate for the strength between the CS layer and PEO/HT sample could limit and
uncoated and coated samples. hinder the initiation and propagation of micro-cracks to a great extent,
which also effectively reduced the contact area of the wear scar region.
3.6. Tribocorrosion tests Accordingly, the COF and wear rate of the PEO/HT/CS sample were
significantly lower than those of the PEO/CS sample.
The tribocorrosion behavior of all samples was tested by recording Fig. 13 exhibits the EDS mapping of the main elements at the various
the variation of the OCP in the SBF before, during, and after sliding, and wear scars. For the Mg alloy, O was visibly detected in the worn zone,
the results are shown in Fig. 11(a). Before sliding, the samples were illustrating that the oxidation reaction occurred during the test [72]. In
soaked in SBF for 10 min to achieve a steady electrochemical state. The the case of the PEO and PEO/HT samples, Si was visibly observed at the
OCP values of the PEO, PEO/HT, PEO/CS, and PEO/HT/CS samples wear track regions, which implies that the PEO-treated coatings still
remained stable at the values of − 1.41, − 1.37, − 1.49, and − 1.46 V vs. existed after the tribocorrosion tests. In addition, abundant C elements
Ag/AgCl, respectively, and that of the uncoated sample was − 1.51 V vs. were also observed at positions #4 and #5, thus directly revealing that
Ag/AgCl. the CS layers did not detach from the substrate.
When the sliding motion began, a sharp drop in potential was The schematic diagrams of the tribocorrosion mechanisms are shown
distinctly observed for the uncoated specimen and caused the overall in Fig. 14. Due to the abrasive wear behavior, a deeper wear groove was

10
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

Fig. 12. SEM topographies of wear tracks after tribocorrosion tests: (a) substrate, (b) PEO, (c) PEO/HT, (d) PEO/CS and (e) PEO/HT/CS samples.

11
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

O Mg Si C

#1

#2

#3

#4

#5

Fig. 13. The EDS mapping of the main elements at the different positions shown in Fig. 11.

Fig. 14. Schematic illustrations of the proposed tribocorrosion protection mechanisms of (a) PEO-treated samples and (b) CS-coated samples.

formed on the PEO-treated coating. The wear of the coating aggravated and wear rate of the PEO/HT/CS coating may be associated with the
the corrosion of the Mg substrate in the SBF, as revealed in Fig. 14(a). presence of solid lubrication at the interface, which effectively protected
However, the PEO/HT/CS coating was found to have a much smaller the PEO/HT coating from destruction and helped to prevent Cl− ions
wear scar than the PEO-treated coating due to its lower COF and higher from eroding the Mg alloy substrate during tribocorrosion.
anti-wear property, as displayed in Fig. 14(b). The CS coating could
provide convenience to the low-shear-strength film at the surface,
eventually resulting in a change in the wear mechanism. The low COF

12
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

4. Conclusion [13] H. Tang, T. Wu, H. Wang, X. Jian, Y. Wu, Corrosion behavior of HA containing
ceramic coated magnesium alloy in Hank’s solution, J. Alloys Compd. 698 (2017)
643–653.
PEO-treated AZ31B Mg alloy is characterized by porousness and [14] M. Toorani, M. Aliofkhazraei, M. Mahdavian, R. Naderi, Effective PEO/silane
contains plenty of micro-cracks. HT enables the precipitation of Mg pretreatment of epoxy coating applied on AZ31B Mg alloy for corrosion protection,
(OH)2 crystals over the PEO sample. The deposition of the CS layer on Corros. Sci. 169 (2020), 108608.
[15] M. Toorani, M. Aliofkhazraei, M. Mahdavian, R. Naderi, Superior corrosion
the PEO and PEO/HT samples was found to effectively seal the pores and protection and adhesion strength of epoxy coating applied on AZ31 magnesium
cracks of the PEO-treated specimens. Based on their anti-corrosion alloy pre-treated by PEO/silane with inorganic and organic corrosion inhibitors,
property in SBF, the anticorrosive effect of the specimens, from great­ Corros. Sci. 178 (2021), 109065.
[16] M. Toorani, M. Aliofkhazraei, R. Naderi, Ceria-embedded MAO process as
est to least, was found to be as follows: PEO/HT/CS > PEO/CS > PEO/ pretreatment for corrosion protection of epoxy films applied on AZ31-magnesium
HT > PEO > substrate. Regarding the tribocorrosion behavior, the stable alloy, J. Alloys Compd. 785 (2019) 669–683.
OCP values, from greatest to least, were found to be as follows: PEO/HT [17] M. Toorani, M. Aliofkhazraei, R. Naderi, Role of lanthanum nitrate in protective
performance of PEO/epoxy double layer on AZ31 Mg alloy: electrochemical and
> PEO > PEO/HT/CS > PEO/CS > substrate. Among them, the PEO/ thermodynamic investigations, J. Ind. Eng. Chem. 53 (2017) 213–227.
HT/CS coating exhibited the lowest COF and wear rate. It also displayed [18] D. Mashtalyar, K. Nadaraia, S. Sinebryukhov, S. Gnedenkov, Polymer-containing
the most stable tribocorrosion performance with an absence of oscilla­ layers formed by PEO and spray-coating method, Mater.TodayProc. 11 (2019)
150–154.
tion in the OCP. [19] D. Mashtalyar, S. Gnedenkov, S. Sinebryukhov, I. Imshinetskiy, A. Gnedenkov,
V. Bouznik, Composite coatings formed using plasma electrolytic oxidation and
fluoroparaffin materials, J. Alloys Compd. 767 (2018) 353–360.
CRediT authorship contribution statement
[20] S. Gnedenkov, S. Sinebryukhov, D. Mashtalyar, Composite coatings formed on the
PEO-layers with the use of solutions of tetrafluoroethylene telomers, Surf. Coat.
L. X.: Writing - original draft, Investigation. X. F.: Investigation. H. J. Technol. 346 (2018) 53–62.
S. Investigation. H. L. S: Writing - review & editing. Y. W.: Funding [21] C. Liu, X. Lu, Y. Li, Influence of post-treatment process on corrosion and wear
properties of PEO coatings on AM50 Mg alloy, J. Alloys Compd. 870 (2021),
acquisition, Writing - review & editing, Supervision. R. L.: Writing - 159462.
review & editing. [22] M. Ostapiuk, Corrosion resistance of PEO and primer coatings on magnesium alloy,
J. AsianCeram. Soc. 9 (2021) 40–52.
[23] G. Zhang, E. Jiang, L. Wu, W. Ma, H. Yang, A. Tang, F. Pan, Corrosion protection
properties of different inhibitors containing PEO/LDHs composite coating on
Declaration of competing interest magnesium alloy AZ31, Sci. Rep. 11 (2021) 1–14.
[24] A. Fattah-alhosseini, R. Chaharmahali, K. Babaei, Impressive strides in
amelioration of corrosion and wear behaviors of Mg alloys using applied polymer
The authors declare that they have no known competing financial coatings on PEO porous coatings: a review, J. Magnes. Alloy. 10 (2022)
interests or personal relationships that could have appeared to influence 1171–1190.
the work reported in this paper. [25] M.J. Anjum, J. Zhao, V.Z. Asl, G. Yasin, W. Wang, S. Wei, W.Q. Khan, In-situ
intercalation of 8-hydroxyquinoline in Mg-Al LDH coating to improve the corrosion
resistance of AZ31, Corros. Sci. 157 (2019) 1–10.
Acknowledgments [26] J.K.F. Suh, H.W. Matthew, Application of chitosan-based polysaccharide
biomaterials in cartilage tissue engineering: a review, Biomaterials 21 (2000)
2589–2598.
The authors acknowledge the financial support by National Natural [27] J. Dou, H. Yu, C. Chen, R.L.W. Ma, M.M.F. Yuen, Preparation and microstructure of
Science Foundation of China (51975304) and Natural Science Founda­ MAO/CS composite coatings on Mg alloy, Mater. Lett. 271 (2020), 127729.
tion of Shandong Province, China (ZR2019MEE016). [28] K. Bai, Y. Zhang, Z. Fu, C. Zhang, X. Cui, E. Meng, S. Guan, J. Hu, Fabrication of
chitosan/magnesium phosphate composite coating and the in vitro degradation
properties of coated magnesium alloy, Mater. Lett. 73 (2012) 59–61.
References [29] C. Yu, L.Y. Cui, Y.F. Zhou, Z.Z. Han, X.B. Chen, R.C. Zeng, S.K. Guan, Self-
degradation of micro-arc oxidation/chitosan composite coating on Mg-4Li-1Ca
alloy, Surf. Coat. Technol. 344 (2018) 1–11.
[1] J. Nagels, M. Stokdijk, P.M. Rozing, Stress shielding and bone resorption in
[30] A. Dey, R. Umarani, H.K. Thota, P. Bandyopadhyay, A. Rajendra, A.K. Sharma, A.
shoulder arthroplasty, J. Shoulder Elb. Surg. 12 (2003) 35–39.
K. Mukhopadhyay, Corrosion and nanoindentation studies of MAO coatings, Surf.
[2] A. Hartwig, Role of magnesium in genomic stability, Mutat. Res. Fundam. Mol.
Eng. 30 (2014) 913–919.
Mech. Mutagen. 475 (2001) 113–121.
[31] S. Liu, Y. Qi, Z. Peng, J. Liang, A chemical-free sealing method for micro-arc
[3] M. Ascencio, M. Pekguleryuz, S. Omanovic, An investigation of the corrosion
oxidation coatings on AZ31 Mg alloy, Surf. Coat. Technol. 406 (2021), 126655.
mechanisms of WE43Mg alloy in a modified simulated body fluid solution: the
[32] ASTM D3359, Standard Test Methods for Measuring Adhesion by Tape Test, 2009.
effect of electrolyte renewal, Corros. Sci. 91 (2015) 297–310.
[33] T. Kokubo, H. Takadama, How useful is SBF in predicting in vivo bone bioactivity?
[4] W. Yao, W. Liang, G. Huang, B. Jiang, A. Atrens, F. Pan, Superhydrophobic coatings
Biomaterials 27 (2006) 2907–2915.
for corrosion protection of magnesium alloys, J. Mater. Sci. Technol. 52 (2020)
[34] Y. Yürektürk, F. Muhaffel, M. Baydoğan, Characterization of micro arc oxidized
100–118.
6082 aluminum alloy in an electrolyte containing carbon nanotubes, Surf. Coat.
[5] T.S. Narayanan, I.S. Park, M.H. Lee, Strategies to improve the corrosion resistance
Technol. 269 (2015) 83–90.
of microarc oxidation (MAO) coated magnesium alloys for degradable implants:
[35] X. Lv, L. Cao, Y. Wan, T. Xu, Effect of different electrolytes in micro-arc oxidation
prospects and challenges, Prog. Mater. Sci. 60 (2014) 1–71.
on corrosion and tribological performance of 7075 aluminum alloy, Mater. Res.
[6] M. Golabadi, M. Aliofkhazraei, M. Toorani, A.S. Rouhaghdam, Evaluation of La
Express 6 (2019), 086421.
containing PEO pretreatment on protective performance of epoxy coating on
[36] L. Xiang, Y.C. Jin, Y. Jin, Hydrothermal modification of Mg(OH)2 particles in
magnesium, Prog. Org. Coat. 105 (2017) 258–266.
NaOH solution, <sb:contribution><sb:title>Chin. J.</sb:title> </sb:
[7] Y. Guo, L. Xu, J. Luan, Y. Wan, R. Li, Effect of carbon nanotubes additive on
contribution><sb:host><sb:issue><sb:series><sb:title>Process. Eng.</sb:
tribocorrosion performance of micro-arc oxidized coatings on Ti6Al4V alloy, Surf.
title></sb:series></sb:issue></sb:host> 3 (2003) 116–120.
Interfaces 28 (2022), 101626.
[37] A. Ghasemi, V.S. Raja, C. Blawert, W. Dietzel, K.U. Kainer, Study of the structure
[8] Y. Gu, C.F. Chen, S. Bandopadhyay, C. Ning, Y. Zhang, Y. Guo, Corrosion
and corrosion behavior of PEO coatings on AM50 magnesium alloy by
mechanism and model of pulsed DC microarc oxidation treated AZ31 alloy in
electrochemical impedance spectroscopy, Surf. Coat. Technol. 202 (2008)
simulated body fluid, Appl. Surf. Sci. 258 (2012) 6116–6126.
3513–3518.
[9] R. Chaharmahali, A. Fattah-alhosseini, K. Babaei, Surface characterization and
[38] J. Dou, J. Wang, Y. Lu, C. Chen, H. Yu, R.L.W. Ma, Bioactive MAO/CS composite
corrosion behavior of calcium phosphate (Ca-P) base composite layer on Mg and its
coatings on Mg-Zn-Ca alloy for orthopedic applications, Prog. Org. Coat. 152
alloys using plasma electrolytic oxidation (PEO): a review, J. Alloys Compd. 9
(2021), 106112.
(2021) 21–40.
[39] L.Y. Cui, R.C. Zeng, S.K. Guan, W.C. Qi, F. Zhang, S.Q. Li, E.H. Han, Degradation
[10] R. Chaharmahali, A. Fattah-alhosseini, M. Nouri, K. Babaei, Improving surface
mechanism of micro-arc oxidation coatings on biodegradable Mg-Ca alloys: the
characteristics of PEO coatings of mg and its alloys with zirconia nanoparticles: a
influence of porosity, J. Alloys Compd. 695 (2017) 2464–2476.
review, Appl.Surf.Sci.Adv. 6 (2021), 100131.
[40] B.D. Hahn, D.S. Park, J.J. Choi, J. Ryu, W.H. Yoon, J.H. Choi, S.G. Kim, Aerosol
[11] X. Shi, Y. Wang, H. Li, S. Zhang, R. Zhao, G. Li, Corrosion resistance and
deposition of hydroxyapatite-chitosan composite coatings on biodegradable
biocompatibility of calcium-containing coatings developed in near-neutral
magnesium alloy, Surf. Coat. Technol. 205 (2011) 3112–3118.
solutions containing phytic acid and phosphoric acid on AZ31B alloy, J. Alloys
[41] M. Toorani, M. Aliofkhazraei, A.S. Rouhaghdam, Microstructural, protective,
Compd. 823 (2020), 153721.
inhibitory and semiconducting properties of PEO coatings containing CeO2
[12] Y.H. Xia, B.P. Zhang, C.X. Lu, L. Geng, Improving the corrosion resistance of Mg-
nanoparticles formed on AZ31 Mg alloy, Surf. Coat. Technol. 352 (2018) 561–580.
4.0Zn-0.2 Ca alloy by micro-arc oxidation, Mater. Sci. Eng. C-Mater. 33 (2013)
5044–5050.

13
L. Xu et al. Progress in Organic Coatings 170 (2022) 107002

[42] G. Zhang, L. Wu, A. Tang, B. Weng, A. Atrens, S. Ma, F. Pan, Sealing of anodized [57] H. Duan, K. Du, C. Yan, F. Wang, Electrochemical corrosion behavior of composite
magnesium alloy AZ31 with MgAl layered double hydroxides layers, RSC Adv. 8 coatings of sealed MAO film on magnesium alloy AZ91D, Electrochim. Acta 51
(2018) 2248–2259. (2006) 2898–2908.
[43] C.Y. Li, X.L. Fan, R.C. Zeng, L.Y. Cui, S.Q. Li, F. Zhang, S.K. Guan, Corrosion [58] A. Ghasemi, V.S. Raja, C. Blawert, W. Dietzel, K.U. Kainer, The role of anions in the
resistance of in-situ growth of nano-sized Mg(OH)2 on micro-arc oxidized formation and corrosion resistance of the plasma electrolytic oxidation coatings,
magnesium alloy AZ31-influence of EDTA, J. Mater. Sci. Technol. 25 (2019) Surf. Coat. Technol. 204 (2010) 1469–1478.
1088–1098. [59] X.J. Cui, X.Z. Lin, C.H. Liu, R.S. Yang, X.W. Zheng, M. Gong, Fabrication and
[44] T. Ishizaki, S. Chiba, K. Watanabe, H. Suzuki, Corrosion resistance of Mg-Al layered corrosion resistance of a hydrophobic micro-arc oxidation coating on AZ31 Mg
double hydroxide container-containing magnesium hydroxide films formed alloy, Corros. Sci. 90 (2015) 402–412.
directly on magnesium alloy by chemical-free steam coating, J. Mater. Chem. A 1 [60] Z. Li, W. Yang, Q. Yu, Y. Wu, D. Wang, J. Liang, F. Zhou, New method for the
(2013) 8968–8977. corrosion resistance of AZ31 Mg alloy with porous micro-arc oxidation membrane
[45] A. Di Martino, M. Sittinger, M.V. Risbud, Chitosan: a versatile biopolymer for as an ionic corrosion inhibitor container, Langmuir 35 (2018) 1134–1145.
orthopaedic tissue-engineering, Biomaterials 26 (2005) 5983–5990. [61] Y. Zuo, T. Li, P. Yu, Z. Zhao, X. Chen, Y. Zhang, F. Chen, Effect of graphene oxide
[46] J.I. Kadokawa, R. Shimohigoshi, K. Yamashita, K. Yamamoto, Synthesis of chitin additive on tribocorrosion behavior of MAO coatings prepared on Ti6Al4V alloy,
and chitosan stereoisomers by thermostable α-glucan phosphorylase-catalyzed Appl. Surf. Sci. 480 (2019) 26–34.
enzymatic polymerization of α-D-glucosamine 1-phosphate, Org. Biomol. Chem. 13 [62] J. Liang, P.B. Srinivasan, C. Blawert, M. Störmer, W. Dietzel, Electrochemical
(2015) 4336–4343. corrosion behaviour of plasma electrolytic oxidation coatings on AM50 magnesium
[47] C.S. Wu, Y.C. Hsu, H.T. Liao, Y.X. Cai, Antibacterial activity and in vitro evaluation alloy formed in silicate and phosphate based electrolytes, Electrochim. Acta 54
of the biocompatibility of chitosan-based polysaccharide/polyester membranes, (2009) 3842–3850.
Carbohydr. Polym. 134 (2015) 438–447. [63] S. Khorsand, K. Raeissi, M.A. Golozar, Effect of oxalate anions on zinc
[48] M.L. Zheludkevich, J. Tedim, C.S.R. Freire, S.C. Fernandes, S. Kallip, A. Lisenkov, electrodeposition from an acidic sulphate bath, J. Electrochem. Soc. 158 (2011)
M.G.S. Ferreira, Self-healing protective coatings with “green” chitosan based pre- D377.
layer reservoir of corrosion inhibitor, J. Mater. Chem. 21 (2011) 4805–4812. [64] A.I. Muñoz, J.G. Antón, J.L. Guiñón, V.P. Herranz, The effect of chromate in the
[49] T. Sun, C.L. Wu, H. Hao, Y. Dai, J.R. Li, Preparation and preservation properties of corrosion behavior of duplex stainless steel in LiBr solutions, Corros. Sci. 48 (2006)
the chitosan coatings modified with the in situ synthesized nano SiOx, Food 4127–4151.
Hydrocolloid. 54 (2016) 130–138. [65] S. Liu, G. Li, Y. Qi, Z. Peng, Y. Ye, J. Liang, Corrosion and tribocorrosion resistance
[50] W. Ye, M.F. Leung, J. Xin, T.L. Kwong, D.K.L. Lee, P. Li, Novel core-shell particles of MAO-based composite coating on AZ31 magnesium alloy, J. Magnes.Alloy.
with poly(n-butyl acrylate) cores and chitosan shells as an antibacterial coating for (2021), https://doi.org/10.1016/j.jma.2021.04.004. In press.
textiles, Polymer 46 (2005) 10538–10543. [66] J. Liang, P.B. Srinivasan, C. Blawert, W. Dietzel, Influence of chloride ion
[51] Z. Jia, P. Xiong, Y. Shi, W. Zhou, Y. Cheng, Y. Zheng, S. Wei, Inhibitor concentration on the electrochemical corrosion behaviour of plasma electrolytic
encapsulated, self-healable and cytocompatible chitosan multilayer coating on oxidation coated AM50 magnesium alloy, Electrochim. Acta 55 (2010) 6802–6811.
biodegradable mg alloy: a pH-responsive design, J. Mater. Chem. B 4 (2016) [67] F. Cao, M. Miao, P. Yan, Hydration characteristics and expansive mechanism of
2498–2511. MgO expansive agents, Constr. Build. Mater. 183 (2018) 234–242.
[52] Z. Qiu, J. Sun, R. Wang, Y. Zhang, X. Wu, Magnet-induced fabrication of a [68] M. Fazel, H.R. Salimijazi, M.A. Golozar, A comparison of corrosion, tribocorrosion
superhydrophobic surface on ZK60 magnesium alloy, Surf. Coat. Technol. 286 and electrochemical impedance properties of pure Ti and Ti6Al4V alloy treated by
(2016) 246–250. micro-arc oxidation process, Appl. Surf. Sci. 324 (2015) 751–756.
[53] L.Y. Cui, H.P. Liu, W.L. Zhang, Z.Z. Han, M.X. Deng, R.C. Zeng, Z.L. Wang, [69] M.A. Golozar, K. Raeissi, M. Fazel, Evaluation of corrosion and tribocorrosion of
Corrosion resistance of a superhydrophobic micro-arc oxidation coating on Mg-4Li- plasma electrolytic oxidation treated Ti-6Al-4V alloy, Surf. Coat. Technol. 244
1Ca alloy, J. Mater. Sci. Technol. 33 (2017) 1263–1271. (2014) 29–36.
[54] Q. Jin, G. Tian, J. Li, Y. Zhao, H. Yan, The study on corrosion resistance of [70] M. Buciumeanu, A. Araujo, O. Carvalho, G. Miranda, J.C.M. Souza, F.S. Silva,
superhydrophobic magnesium hydroxide coating on AZ31B magnesium alloy, B. Henriques, Study of the tribocorrosion behaviour of Ti6Al4V- HA biocomposites,
Colloids Surf. A 577 (2019) 8–16. Tribol. Int. 107 (2017) 77–84.
[55] L. Wang, B.P. Zhang, T. Shinohara, Corrosion behavior of AZ91 magnesium alloy in [71] E. Matykina, I. Garcia, R. Arrabal, M. Mohedano, B. Mingo, J. Sancho, A. Pardo,
dilute NaCl solutions, Mater. Des. 31 (2010) 857–863. Role of PEO coatings in long-term biodegradation of a Mg alloy, Appl. Surf. Sci.
[56] A. Castellanos, A. Altube, J.M. Vega, E. García-Lecina, J.A. Díez, H.J. Grande, 389 (2016) 810–823.
Effect of different post-treatments on the corrosion resistance and tribological [72] L. Xu, N. Liu, L. Cao, Y. Wan, Influences of electrolytes on tribocorrosion
properties of AZ91D magnesium alloy coated PEO, Surf. Coat. Technol. 278 (2015) performance of MAO coating on AZ31B magnesium alloy in simulated body fluid,
99–107. Int. J. Appl. Ceram. Technol. 18 (2021) 1657–1669.

14

You might also like