APTT (secs)
Citrate samples (3ml 0.109M Becton Dickinson
alert the operator.
Vacutainer Plus) which contained a high degree of We conclude that the Sysmex CS-2100i is more
In the presence of a high lipid content the analyser lipaemia visible to the naked eye were retained. tolerant of the presence of lipid than the Sysmex
automatically switches detection from 660 to 880nm Lipid in form of ClinOleic 20% intravenous fat CA1500 in respect of Prothrombin time and
where interference from lipids is reduced. emulsion (Baxter) was added in a 1:40 dilution to a especially thrombin time and Clauss fibrinogen
Influence of Lipaemia on thrombin times
series of 20 plasma samples. Samples were 6 samples where clotting times could be determined in measurements.
CA-1500 (Sysmex Corporation) presence of gross lipaemia
analysed before and after lipid addition APTTs were significantly shorter in the presence
A fully automated coagulation analyzer in production We measured the bH value during prothrombin times of lipid on both systems although the effect was
since the late nineties. on the Sysmex CA1500 as an indication of the more marked on the Sysmex CS-2100i.
Thrombin (secs)
degree of lipaemia In view of the effects of lipid the ability to alert the
All clotting endpoints detected by change in light analyser operator to the presence of lipid via a
scatter monitored at 660 nm. All samples were tested simultaneously on Sysmex
flag on the results is useful.
CS-2100i and Sysmex CA1500
Acknowledgements Sysmex Corporation for use of HIL Estimation and Lipaemia Check information