You are on page 1of 10

Preface

Thank you for purchasing our products. In order that you can
use it correctly,please read the instruction first and make sure
the accessories are complete. Any questions please contact us
or your local agentcy.
LianYi SH-X Attention
1.Use the instrument with the instruction. When we upgrade our
product change the instruction, sorry about not telling.
Instruction Manual 2.The user's understanding of the manual has been expressed in
Multichannel temperature recorder the simplest way possible. Please contact our company or agent
if you find any inaccuracy or unclear explanation.

S H -X P
Version: V1.0

Warning
For your personal safety and proper use of the instrument, please
SH-XP
follow the instructions. And strictly observe the following safety
regulations.
USB
5寸 高 清 REC
STOP 1. Power Supply and earthing protection, this product work power
触控显示屏 supply for AC86-265V power supply, before turning on the power
Trig
854*480dbi SETUP ESC

设置

LIST FILE
退出

ALARM
supply should ensure that the power supply is matched with the
列表

CURVE
文件

BAR
报警

MUTE rated voltage, and ensure that the power supply has been connected
曲线 柱图 静音
东莞市联仪仪器仪表有限公司
LianYi SH-X Dongguan Lianyi Instrument Co., Ltd.
to the protective ground to prevent electric shock, this instrument shell
has been connected to the ground terminal of the power socket.
2. Do not operate in explosive environment to avoid personal injury caused by explosion.
3. Please do not open the instrument shell by yourself. There is high voltage in
some parts of the instrument to prevent electric shock.
4. It is not allowed to plug and unplug the terminal in case of electric shock.
5. The company shall not be liable for instrument damage due to violation
of safety regulations.
Dongguan Lianyi Instrument Co., Ltd.

-1-
2.
.Fundamental
1.Description
1.Thermocouple keyboard

Single Chip Microcomputer


The machine use 32-bit cpu to collect data, 5 inch screen to display. Support K-type 2.Thermocouple
J-type E –type T-type N-type S-type R-type and B-type thermocouple,Multiple display

Light relay
modes, perfect functions, and simple operation suitable for production and R&D needs. Display Unit
Signal amplifier A/D
Widely used in lighting appliances, power tools, household appliances, motors, commu
electric heating appliances, medical, petroleum, chemical, metallurgical and electric. nicate
THERMOCOUPLE N
We can customize more functions for more applications according to your memory
requirements. cold junction
compensation

SH-XP Multiplex temperature recorder has the following characteristics:

▲ HD 5 inch IPS touch screen, 854X480 resolution. . Basic Block Diagram


▲ 32-bit high-speed MCU, 24-bit high-speed AD, fast response, high precision
and stability. As the picture shows, the instrument consists of thermocouple, Light relay, Signal
amplifier, A/D, Single Chip Microcomputer, keyboard, screen, communication port,
▲ Contains the file list, real-time list, bar graph, real-time recording, curve , data storage, cold junction compensation.
alarm function,etc.
▲ Support 4/8/16/32 multi-interface real-time value display.
The corresponding channel signal is selected by the photoelectric switch selector, amplified
▲ Support a variety of thermocouples : K-type J-type E –type T-type N-type S-type
by the signal amplifier, and then converted into data signal by the analog signal through the
R-type and B-type AD converter to the single chip microcomputer for data processing, the temperature of the
▲ Each channel can be set with different types of thermocouples cold end is measured by the compensation circuit of the Cold End, and the measured signal
and the temperature of the cold end are processed by the single chip microcomputer.
▲ The name of each channel can be changed. Can be exported to EXCEL sheet.
The voltage between each channel can be as high as AC/DC 350V, Strong
The keyboard, communication and data storage can set and store the data on the display screen.
anti-interference ability.
Data analysis can also be performed directly from the computer through a communication interface
▲ Each channel can set high and low alarm value with sound and light alarm function. to the computer.
▲ Each channel can modify the error value according to the formula” Y=KX+B”.
▲ .
8G memory, file list display, up to 64 temperature data can be recorded at one
time, each file can store 130,000 sets of data, record every 1 second can record
up to 97 days.
▲ Record files can be single-or multi-select delete and export, U Disk and PC directly
export EXCEL files.
▲ With USB, 485 interface, relay output, alarm function.
▲ With communication address code setting, machines can be connected to each
other, we can provide communication protocol.
▲ Standard configuration is USB, you can choose RS232 or RS485.
▲ 8 channels per module, maximum support 64, identified by connecting.
▲ Multi-level alarm instruction, HH, H, L, LL alarm setting and alarm instruction
(with relay output) , and can view historical alarm record function.

-2- -3-
4. Outline dimensions (cm)
3. Technical indicators
3.1 technical indicators
Display mode 5 inch TFT true color LCD industrial touch screen 854 x 480 pixels
Heads SH-XP

USB
REC

Real-time list, curve (history curve) , Bar Chart, current alarm 5 inches HD touch screen STOP

Display form 10 11.5


(history alarm) , file record, record list. Trig 854*480dbi SETUP ESC

设置 退出

Local view record curve, history alarm record, computer software LIST FILE ALARM

Record Query 列表 文件 报警

analysis query. 东莞市联仪仪器仪表有限公司


CURVE

曲线
BAR

柱图
MUTE

静音
LianYi SH-X Dongguan Lianyi Instrument Co., Ltd.

Number of channels Each module has 8 channels and supports up to 64 channels.


27

Thermocouple K J E T N S R B

Basic accuracy 2 words at 0.2℃ (excluding thermocouple error) Side face


Measuring range - 200 ~ 1820 °C (depending on the scale of thermocouple)

Cold End compensation ACCURACY: 0.5 °C

Resolution 0.1℃
30
Calibration Independent error correction for each channel Y= Kx B (x= measured value)

Number of files 64(circular record)


5. Panel description
File capacity A file can record 130,000 sets of numbers (channel numbers are not distinguished)

USB Port
Export Record file, second change U Disk Function (instrument is U Disk) , Front panel
directly view files and software.
1-second recording interval can record 97 consecutive days, Direction Operation Keyboard, record,
Record time SH-XP
total time = record interval x 97 days. stop key, around the next key, used to
Recording indicator USB
REC
switch the page, in each interface are
Sampling speed Speed per channel: 0.1 S, SLOW: 1s 5 inches HD touch screen STOP

USB Port prompted.


Trig 854*480dbi SETUP ESC

AC / DC up to 350V, high voltage live measurement, super Start recording quickly 设置 退出


Function shortcut key: Set Key, Exit Key;
Interchannel isolation
anti-interference ability
LIST FILE ALARM

列表

CURVE
文件

BAR
报警

MUTE
real-time list, File List, alarm list; curve
Two sets of independent relay outputs are (h/l) and (HH/LL) relays Power switch LianYi SH-X 东莞市联仪仪器仪表有限公司 曲线 柱图 静音
display, Column Chart Display, mute key.
Control output Dongguan Lianyi Instrument Co., Ltd.

Alarm sounds All the way buzzer sound (any alarm when sounded, can be set to mute)

Record interval 1-9999 seconds, any setting Back Panel

Communication (Below 48, a three-ply structure is used, and above 48, a four-ply structure.)
Standard USB, Rs485
interface
CH25-32 CH57-64 USB Communication Interface (standard)
Power supply AC85-265V ± 10% , frequency 50Hz/60Hz & LT; 10w
Alarm relay output (standard)
CH17-24 CH49-56 Rs485 communication interface (standard)
Thermocouple Each channel is equipped with a standard 2M K type thermocouple
CH09-16 CH41-48
Working socket
Dimensions Width 220X depth 293X high foot 106mm
CH01-08 CH33-40

Weight About 3 kg (configuration is different)


Cold End compensation probe
Environmental conditions 5 ~ 40 °C, 20% ~ 80% rh (no condensation)

-4- -5-
6.Display and instructions 6.1.3 real time parameter list display
6.1 display the interface description Real-time parameter list, can display multiple measurement parameters at the same time,
the display interface is divided into 4 channels data display, 8 channels data display,
6.1.1 interface icon description 16 channels data display, 32 channels data display, can turn pages to show the data of each
channel.
Record symbol
On the record
Use Key can switch 4/8/16/32 channel display page.Use
Stop Recording
Record the cumulative time Alarm instruction
Key to turn the page can be displayed. Where the 4-channel page can display the
Record interval maximum and minimum values.
Brand information Page name Unit Flash drive instructions

Key can enter the menu selection, press the keyboard shortcut key can
联 仪 仪器 000-00:00:00 2017-02-28
REC CURVE ℃ INT:0001 ! 12:12:12 Perpetual calendar quickly enter the corresponding interface.
1000.0
The curve moves up 900.0

800.0
Y
~ SET Curve parameter setting
Record symbol
700.0 On the record
600.0 Stop Recording
500.0
Record the cumulative time Alarm instruction
400.0
Record interval
300.0 Brand information Page name Unit Flash drive instructions
200.0

100.0
~ 联仪仪器 000-00:00:00 2017-02-28
The curve moves down 00.0 Y LIST ℃ INT:0001 ! 12:12:12 Perpetual calendar
2017-02-28 X= 30 秒/格 2017-02-28
10:12:12 12:12:12
Menu Button 01 02 03 04 05 06 07 08 Channel number 01 K ⾃定义温度名称1 05 K ⾃定义温度名称5
390.2 500.2 750.2 600.2 640.2 690.2 480.2 850.2 Real-time measurements show
Thermocouple type 32.8 32.8
The length of each gear 02 J ⾃定义温度名称2
06 K ⾃定义温度名称6
Fine Rough Fine Rough
Curve to the left Curve moves to the right Custom name 32.8 32.8
03 E ⾃定义温度名称3 07 K ⾃定义温度名称7
Alarm instruction
From top to 32.8 32.8
bottom: HH H L LL
6.1.2 BOOT interface 04 T ⾃定义温度名称4 08 K ⾃定义温度名称8

Alarm instruction
When you go over
32.8 32.8
Press to turn page; Press to switch between 4 / 8 / 16 / 32 pages; 1/8
Page number/total
the limit, turn red number of pages
Menu key Operation area

联仪 仪器
联仪仪器 000-00:00:00
LIST
列 表 ℃ INT:0001 !
2017-02-28
12:12:12

01 自定义名称01 03 自定义名称03
联仪 仪器 000-00:00:00
多路温度记录仪 K K LIST ℃ INT:0001 !
2017-02-28
12:12:12

32.8 32.8
Multi-channel temperature recorder
01 05 09 13
型号( Name ):SH-XP 联仪仪器 000-00:00:00
32.8 LIST
2017-02-28
K ℃ INT:0001 !
通道数量(CH): 08 MAX: 33.8 M I N: 31.8
K
32.8
MAX: 33.8 32.8 M I N: 31.8
K K
32.8 12:12:12

02 06 01
10 k 32.8 09
14 k 32.8 17 k 32.8 25 k 32.8
02 自定义名称02 04 自定义名称04
K K
k 32.8
k 32.8 K k
02 32.8
32.8 10
K k
32.8
32.8 18 k 32.8 26 k 32.8

When the power-on display boot interface, display the company LOGO, 32.8 03
K
32.8
32.8
07
K
32.8
03
11 k
K k
04
32.8

32.8
32.8
11 k
15

12
K k
32.8

32.8
32.8
19 k

20 k
32.8

32.8
27 k

28 k
32.8

32.8
company name, product model information. MAX: 33.8 M I N: 31.8 MAX: 33.8 M I N: 31.8
Press to turn page; Press 04 between 4 / 8 / 16 / 08
to switch 32 pages; 1/16 05
12 k 32.8 13
16 k 32.8 21 k 32.8 29 k 32.8
K K
32.8 32.8 K k
06 32.8
32.8 14
K k
32.8
32.8 22 k 32.8 30 k 32.8

Press to turn page; Press 07 between


to switch k 32.8 15pages;
4 / 8 / 16 / 32 k 32.8
1/4 23 k 32.8 31 k 32.8

08 k 32.8 16 k 32.8 24 k 32.8 32 k 32.8


Press to turn page; Press to switch between 4 / 8 / 16 / 32 pages; 1/2

-6-
-7-
6.1.4 the curve shows 6.1.5 real time bar charts
联仪仪器 000-00:00:00 2017-02-28
BAR ℃ !
联仪仪器
CURVE ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12
Real-time curve 1000.0
INT:0001 12:12:12
Real-time bar graph
1000.0 900.0

900.0

800.0
Y
~ SET Real-time curve only for real-time view display.
Only 8-channel curves can be displayed per page.
800.0

700.0
The ratio of each channel can
be compared by column graph.
700.0
And can display real-time measurement value. 600.0

600.0 500.0 Note: The Y axis shows the same


~
500.0

400.0
400.0
temperature range as the curve
300.0
setting, set only in the curve setting.
~ Y
300.0
Touch control Y The key may cause the 200.0

200.0

100.0
~ curve to move up or down.
100.0

00.0
03 04 05 06 07
00.0 Y 01
390.2
02
500.2 750.2 600.2 640.2 690.2 480.2
08
850.2
2017-02-28 X= 30 秒/格 2017-02-28
10:12:12 12:12:12 Press to turn page; Press to switch between 8 / 16 pages; 1/4
01 02 03 04 05 06 07 08 Touch control SET Key into the curve settings page.
390.2 500.2 750.2 600.2 640.2 690.2 480.2 850.2 Use Button can switch 8/16 channels
联仪仪器 000-00:00:00 2017-02-28
BAR ℃ INT:0001 ! 12:12:12
to display the page, Use The key can turn
联仪 仪器
REC CURVE ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12 Recording curve 1000.0

1000.0
900.0
the page display.
900.0 Y
~ SET Record curve is a real-time curve 800.0

700.0
800.0

700.0
that is displayed when record is 600.0

600.0 started, or you can use left and 500.0

400.0
500.0

400.0
right moving keys in touch screen 300.0

300.0 to move to see the curve change. 200.0

200.0

100.0
~ On multiple pages,Use
100.0

00.0
01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16
00.0 Y The key can turn the page display.
390.2 500.2 750.2 600.2 640.2 690.2 480.2 850.2 390.2 500.2 750.2 600.2 640.2 690.2 480.2 850.2

2017-02-28
10:12:12
X= 30 秒/格 2017-02-28
12:12:12
Press to turn page; Press to switch between 8 / 16 pages; 1/4
01 02 03 04 05 06 07 08
390.2 500.2 750.2 600.2 640.2 690.2 480.2 850.2

6.1.7 file list page


联仪 仪器

1000.0
ANALYSIS ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12 Historical curve 联仪仪器
FILE LIST ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12
File list
900.0

800.0
NO. File Start Time File Size 记录
REC
Provides Record Start and stop buttons,
700.0 The history curve is can turn 01 0307121212 2017-3-7 12:12:12 100
File History View, file export and delete
600.0 functions. Click on the corresponding
500.0 the page, but can not be set, 02 0307121212 2017-3-6 12:12:13 5523 历史曲线
Analysis
location can be single-or multi-select,
can only show the overall curve
400.0
03 0307121212 2017-3-5 12:12:14 60000
300.0 also can be the current page quickly
04 0307121212 2017-3-4 12:12:15 50000 导出
Export select all for export and delete.
200.0

100.0
2017-02-28 2017-02-28
of the entire file screen. 05 0307121212 2017-3-3 12:12:16 10000
10:12:12 12:12:12
00.0

返 回 删除
Delete
Press to turn page; 1/8
Press to turn page; 1/4

Curve setting 联仪仪器


TES FILE LIST ℃
FILE LIST 000-00:00:00
INT: !
2017-02-28
12:12:12
联仪仪器
FILE LIST ℃
FILE LIST 000-00:00:00
INT: !
2017-02-28
12:12:12

NO. File Start Time File Size REC NO. File Start Time File Size REC
联仪仪器
SET CURVE ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12 Curve setting Export File Delete selected
2 0 1 7 - 3files
-7 12:12:12 100
2017-3-6 12:12:13 5523
Analysis Analysis
Y: SET Range - 100.0 EXIT
0.0
The curve setting bounds include 记 录
USB 记录
COPY YES NO
Group CH01-08 CH09-16 CH17-24 CH25-32 33-64 the range in which the y-axis curve Export file to
USB flash disk
Save file as local
file disk
Export 0030
3 07712
1 21121
2 122 Export

CH01 CH02 CH03 CH04 CH05 CH06 CH07 CH08


is displayed, and whether the curve Delete
!
Warning: the file cannot be recovered after
deletion, please be careful!!
Delete
CH09 CH10 CH11 CH12 CH13 CH14 CH15 CH16
for each channel is displayed or masked. Press to turn page; Press to turn page;
CH27 CH28 CH29 CH20 CH21 CH22 CH23 CH24

CH25 CH26 CH27 CH28 CH29 CH30 CH31 CH32


The machine can cycle through 64 CSV files (which can be opened directly using EXCEL) ,
C lic k th e dis pl ay posi ti on to cha ng e th e se tt i n g ;
and each file can record up to 130000 pieces of data (regardless of the number of channels) .
Files can be directly exported to the U Disk (support only under 32 g/FAT32) , or transferred
to the internal U Disk space, using the "second variable U Disk" function (set page) , PC
connected to the front panel u disk interface, directly read the recorded files.

-8- -9-
6.1.6 alarm Flash drive in seconds Flash drive in seconds
联仪仪器
CURRENT ALARM ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12 Call the police now 联仪仪器
SETUP ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12

NO. CH Type Alarm start time


Enter the second variable U Disk Page, the
Alarm record : : Turn into USB
Date Time 20 instrument is equivalent to a u disk. Use USB
01 01 H 2017-3-3 11:12:12
02 03 H 2017-3-3 11:12:02 In the alarm interface can view all Turn into USB flash drive
Language cable to connect the front panel USB interface,
Addr
SET ALARM the current alarm status information, SET INPUT wait 5 seconds, the PC detected u disk. Factory
Sample Baud S
including channel number, type, default to the specification and PC software stored
occurrence time, in multi-record can Unit Backlight Time
PC A DJ in the instrument. Use this feature to copy to a PC
USB drive
turn the page to see. INT S Alarm Sound for use. It is not recommended to open it directly.
! Tip: do not close when reading files
AUTO REC About ! EXIT
Press to turn page; 1/1

联仪 仪器
ALARM RECORD ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12 History alert Input Setting
NO. CH Type Alarm start time Alarm end time
Can view all the alarm has been 联仪仪器 000-00:00:00 2017-02-28
Input Setting
01 01 H 2017-3-3 11:12:12 2017-3-3 12:12:12 SET INPUT ℃ INT:0001 ! 12:12:12
restored status information, including The input settings provide a display function for each
02 03 H 2017-3-3 11:12:02 2017-3-3 12:12:02 ON
Type
channel number, type, occurrence time,
CH OFF Custom Name channel, custom name input, and thermocouple type
01 自定义名称1 K selection.
recovery time, in more than one record 02 自定义名称2 J
1/1
can turn the page to see. 03 自定义名称3 E Click on the corresponding location, to change the
04 自定义名称4 T settings of the items.
The delete key can be used for batch 05 自定义名称5 N
deletion. 06 自定义名称6 S The custom name uses the full keyboard pinyin input
Delete EXIT
Press to turn page; 1/1 07 自定义名称7 R method, may carry on the input to the Chinese, the
08 自定义名称8 B 返 回 English, the symbol.
Press to turn page; Click the display position to change the setting;
联仪 仪器
SET ALARM ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12 Alarm value setting
ON
CH OFF
HH H L LL
01 1000.0 100.0 0.0 -100.0
In some interface can be set directly
1/1
02 1000.0 100.0 0.0 -100.0 ultra-high value (HH) too high value
03 1000.0 100.0 0.0 -100.0 (H) too low value (l) too low value (LL) .
04 1000.0 100.0 0.0 -100.0 Directly Click on the corresponding location,
05 1000.0 100.0 0.0 -100.0 pop-up number keyboard to change the value.
06 1000.0 100.0 0.0 -100.0
07 1000.0 100.0 0.0 -100.0
08 1000.0 100.0 0.0 -100.0 EXIT
Press to turn page; Click the display position to change the setting;

Full Keyboard Type selection of thermocouple


6.1.7 system settings
联仪仪器
SETUP ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12
System settings provide a rich menu of settings,
Correction Setting
Date Time 2017 02 28 12 : 12 : 12 Turn into USB clear and easy to use. Date and time, measurement 联仪仪器
SET ADJ ℃ 000-00:00:00
INT:0001 !
2017-02-28
12:12:12

Language 中文EN Addr 1


speed, display language, measurement unit, Buzzer CH Display(y=kx+b) k b (△) Cold: 30.0
Correction Setting
SET INPUT Sound, communication address, Communication Baud 01 32.8 1.000 0.0 ADJ : 0.0
Sample FAST Baud 9600 rate, backlight time (0 for long light, 1-999 seconds to This interface provides calibration settings for each
02 32.8 1.000 0.0
turn off backlight) recording interval, on-off automatic 03 32.8 1.000 0.0
channel, and can display the current temperature
Unit ℃ Backlight Time 60 S
AD J measurement for each channel. When you change
recording, on. 04 32.8 1.000 0.0
INT 1 S Alarm Sound the calibration value, you can see the change of the
05 32.8 1.000 0.0
AUTO REC About ! EXIT 06 32.8 1.000 0.0 current value in real time, which can be corrected to
07 32.8 1.000 0.0 the actual measurement value.
08 32.8 1.000 0.0 EXIT
Press to turn page; Click the display position to change the setting; x=measurements
Note: Time and date, measurement speed, unit of measurement,
and sensor can not be changed when recording is started.
Correction formula Y=kx B (x= measured value)
The configured software communication address is 1, the Baud rate
is 9600, if changed will cause with the computer software connection failure. Click on the corresponding location, to change the settings of the items.

-10- -11-
7. Software instructions

Find the file on the disc Run the installation directly as follows.

After the installation is complete, install the USB driver again

Until after the installation, on the desktop to see A ICON, Click this icon to run the software program.

专注产品 用心服务

联仪仪器 SH-X多 路 温 度 记 录 仪
LIANYI YIQI Multi-channel temperature recorder

手 机: 1 3 7 2 9 9 8 2 5 7 7
东莞 市 联仪 仪器 仪表 有限 公 司 电 话: 0 7 6 9 - 3 3 6 8 0 7 5 1
Dongguan Lianyi Instrument Co., Ltd. QQ:389068903
http://www.lianyi17.com

Go to the start screen

-12- -13-
In this to first computer and instrument connection, in the file list after the installation
of drivers, and then in the soft interface to select the correct COM port, in the lower left
corner of the main interface display communication connection success. You can customize the name for each
channel.

Computer interface provides a rich display and analysis functions, can display file list,
curve analysis display, data list, real-time temperature list, instrument operation button
function. It can also print graphs and curves.

FILE LIST: serial number, record time, file name, data number, file size. Directly Click
on the file can open the corresponding file data.

The data list can be displayed for each channel and each time interval. Users can directly in the
installation file directly with EXCEL software open. The suffix is called * . CSV.

The curve label can analyze all the data in the file in detail, can directly use the mouse
to zoom in and out of the curve and move left, right, up and down.

A real-time data list displays the current measurement in real time.

You can change the display color


for each channel of the curve.
The instrument can be operated remotely
directly by using the instrument key page.

-14- -15-
Verification Conditions
Packing list

Projects Reference value or range Reference value or range


Main engine 1PCS
Ambient temperature℃ 20 ±5
Power Cord 1PCS
Ambient humidity% rh 45~75 User Manual 1PCS
Certificate of compliance / Warranty Card 1PCS
Atmospheric pressure KPA 86~106
Thermocouple wire 1PCS/CH
AC Supply Voltage V 220 ±2%

AC Supply Voltage Hz 50 ±1%


Warranty
AC power supply waveform Sine wave ß=0.05
The instrument has a 2-year warranty from the date of purchase. If the
External electromagnetic
Should be avoided instrument is damaged during the warranty period due to improper
interference
operation of the user, the user shall bear the maintenance fee and the
Ventilate Good
expenses arising from the maintenance.The company shall be responsible
Sunlight Avoid direct fire for the lifetime paid maintenance of the instrument.

Without the written consent of the company, users are not allowed to
open the instrumentcasing, which will affect the instrument warranty.

Instrument maintenance should be authorized by our professional and


technical personnel; maintenance, please do not replace the instrument
internal devices, instrument maintenance,the need to re-measurement
calibration, so as not to affect the accuracy of testing. If the user blindly
maintenance, replacement of parts and equipment damage, not covered by
the warranty, the user should bear the maintenance costs.

The company has the right to improve the appearance and function of
the manual and theinstrument without notice.

-16-
Certificate of conformity

Product name: Multichannel temperature recorder


SH-XP
Product model:

Product Code:

Date:

Inspector:

Verification Conclusion:

Product Warranty Card


●Warranty:
1.The warranty period is within 24 months from the date of purchase.
2. During the warranty period, in the normal use and maintenance of equipment, equipment
LianYi SH-X
problems, after verification, the company will provide free repair and replacement parts.
●Maintenance is not free of charge in the following cases.
1. Products by the company's non-technical personnel repair, modification, modification, users to replace any internal parts.
2.The airframe number has been altered or is not in accordance with this certificate.
3.Damage is caused by water or other substances seeping into the machine.
●The company may also provide maintenance services for equipment exceeding the free warranty and
not covered by the Free Warranty Ordinance, subject to the charging of accessories and maintenance
fees as appropriate.
Model
Name number
Date of
Telephone purchase
Serial
Address
number Dongguan Lianyi Instrument Co., Ltd.
Maintenance
Maintenance record Maintenance man
date

For Sales Service, please contact your local distributor