Inter-sample Carryover Test Protocol
Scope
The purpose of this study is to demonstrate that the sequence of testing does not affect the
measured values on the iSED and iSED ELITE analyzers.
Sample Requirements
• Samples must be collected in EDTA anticoagulant 13 x 75 mm closed sample tube
(lavender tube) with at least 3.5 mL minimum sample volume.
• The samples must be non-hemolyzed and without presence of clots.
• The samples must be processed within four (4) hours of withdrawal.
• The study will be performed selecting samples with the following criteria:
• one (1) sample with result between 5 and 15 mm/hr. (low ESR result)
• one (1) sample with result between 55 and 65 mm/hr. (high ESR result)
Sample Selection
Process batches of twenty (20) random samples on the iSED until one sample from each of the
two (2) ranges defined in the “Sample Requirement” section is collected.
1. Press the “Add sample” icon on the instrument touch screen and wait for the instrument
mixing wheel to position for loading the sample.
2. Load a sample tube with the barcode oriented toward the right for sample ID recognition.
112-07-005 Rev 1 1
iSED Inter-sample Carryover Test
Protocol
Sample Preparation
1. Mix each selected sample by mild inversion for at least three (3) minutes.
2. Remove the stopper from two (2) plain sample collection tubes.
3. Using the transfer pipette transfer 1ml of the high ESR result sample to each empty tube
and replace the tube stopper.
4. Mark the empty tubes with the value a1 and a2.
5. Close the primary sample tube with high ESR result.
6. Discard the transfer pipette.
7. Remove the stopper from the other two (2) empty sample collection tubes.
8. Using the transfer pipette transfer 1ml of the low ESR result sample to each empty tube
and replace the tube stopper.
9. Mark the empty tubes with the value b1 and b2.
10. Replace the primary sample tube with low ESR result.
11. Discard the transfer pipette.
112-07-005 Rev 1 2
Data Analysis
1. Record the ESR results from the iSED analyzer on the Excel Sample Carryover worksheet
provided by ALCOR. To open the spreadsheet, click on the Excel object. The % carryover
will be automatically calculated.
2. To calculate the % carryover manually, use the following formula; where k is the carryover
result expressed in percentage.
112-07-005 Rev 1 3
Approved By:
(CO-3153) iSED2.0 Performance study procedures and reports
Description
Release of relevant performance study procedures and reports prior to release of iSED2.0 to production. These are verification and validation documents for this project.
Justification
Version History:
Author Effective Date CO# Ver. Status
Proprietary
Proprietary&&Confidential
Confidential -- Page 4 of
Page 1 of 14 -- Uncontrolled
Uncontrolledif if Printed
Printed