The major cation of the extracellular fluid is sodium. The typical daily diet contains 130-280
mmol (8-15 g) sodium chloride. The body requirement is for 1-2 mmol per day, so the excess is
excreted by the kidneys in the urine.
Hyponatraemia (lowered plasma [Na+]) and hypernatraemia (raised plasma [Na+]) are
associated with a variety of diseases and illnesses and the accurate measurement of [Na +] in body
fluids is an important diagnostic aid.
Potassium is the major cation found intracellularly. The average cell has 140 mM K+ inside but
only about 10 mM Na+. K+ slowly diffuses out of cells so a membrane pump (the Na +/K+-
ATPase) continually transports K+ into cells against a concentration gradient. The human body
requires about 50-150 mmol/day.
Flame photometry is a traditional and simple method for determining sodium and potassium
concentration in biological fluids. It relies on the principle that an alkali metal salt drawn into a
non-luminous flame will ionise, absorb energy from the flame and then emit light of a
characteristic wavelength as the excited atoms decay to an unexcited ground state. The intensity
of emission is proportional to the concentration of the element in the solution. Quantitative
analysis of these species is performed by measuring the flame emission of solutions containing
the metal salts. Solutions are aspirated into the flame. The hot flame evaporates the solvent,
atomizes the metal, and excites a valence electron to an upper state(higher orbitals). Light is
emitted (emission of a photon of radiation) at characteristic wavelengths for each metal as the
electron returns to the ground state. Optical filters are used to select the emission wavelength
monitored for the analyte species. A photocell detects the emitted light and converts it to a
voltage, which can be recorded. Since Na+ and K+ emit light of different wavelengths, by using
appropriate coloured filters (monochromator), the emission due to Na+ and K+ (and hence their
concentrations) can be specifically measured in the same sample.
1. They respond linearly to ion concentrations over a rather narrow concentration range, so
suitable dilutions usually have to be prepared.
2. It employs relatively expensive machines.
3. Many different experimental variables affect the intensity of light emitted from the flame.
Therefore, careful and frequent calibration is necessary for good results. This makes the
method a bit complex and tedious.
Glass industry
Clinical applications
APPARATUS USED
REAGEANTS USED.
1000 ppm (mgl-1) Na stock solution (from NaCl), 1000 ppm K+ stock solution from KCl and an
unknown sample of fertilizer. Distilled water.
1. From the prepared 1000ppm (mgl-1) Na+ stock solution (from NaCl) we made 0, 5, 10, 15
and 20ppm working standard solutions.
2. These solutions were then aspirated into the flame photometer and the galvanometer
reading at each reading (abscienssae) against the concentration, (ordinates) noted and
recorded in table 1 (Sodium in Drinking water) on the result sheet attached onto this
report.
3. A calibration curve for Na was prepared as indicated on graph 1 attached to this report.
4. A sample of ordinary drinking tap water was sprayed into the flame of the flame
photometer. The galvanometer deflection was noted, and then the concentration of
sodium in ordinary drinking tap water was evaluated from the calibration curve.
I. STANDARD K+ SOLUTION.
From 1000ppm K+ stock solution, ( made by dissolving 0.47 gm of A.R KCl in deionized water
and diluting 250ml in a Volumetric flask). We prepared a stock solution of 100ppm K+. From
this solution, we prepared 1,2,4,8 and 10ppm solution for calibration of the photometer.
14g of finely ground sample of fertilizer was accurately weighed then dissolved in deionized
water in a 250ml volumetric flask. 25ml of this solution was dissolved in deionized water in a
250ml volumetric solution.
The solution was sprayed into the flame of the flame photometer. The emission was compared to
the calibration curve of potassium.
ANALYSIS OF THE RESULTS.
= 10.15*5
=52.75ppm
=5.5*2.5
=13.75ppm
The straight line graphs indicate that the quantity of emission is directly propoetional to
concentration.
The following are possible sources of errors that could have caused deviation in the results
obtained from:
CONCLUSION.
The experiment was a huge success since the objectives of the experiment were all met.
QUESTIONS.
1. Why is it difficult to determine group II elements than group I elements using flame
photometry?
It is because group II elements have the two valence electrons in the s-orbitals thus
requires more energy to excite the to a higher state.
Ensure that the photometer drain is leading into a sink and that the instrument is
connected to gas, air and electricity supplies. Ensure the mains supply gas tap is off.
Turn the "Sensitivity" and instrument "Gas" controls control fully counterclockwise
Turn on the air supply control and adjust the air pressure to 10 lb/in 2. Leave for 1-2
Place a beaker of distilled water into position at the left hand side of the instrument and
minutes to stabilize.
insert the narrow draw tube into it to allow water to pass through the photometer.
(NOTE: once set up, the photometer must have water running through it at all times
when a salt solution is not being measured. The rate of uptake is fast, so make sure
Adjust the gas control to give a flame with a large central blue cone then, with water
there is always enough water in the beaker).
passing through the instrument, slowly close the gas control until ten separate blue cones
Replace the distilled water with the 5 mM NaCl standard and adjust the "Sensitivity"
Quickly but carefully, replace the 5 mM NaCl standard with standards of decreasing
control till the galvanometer reads 100.
Run water through the instrument again for 1-2 min then place the draw tube into a
concentration from 4 mM to 0.25 mM and note the readings in the Table below.
beaker containing the 1 in 50 diluted rehydration sachet solution and note the
Run water through the instrument again and replace the sodium with the potassium filter.
galvanometer reading.
Repeat the above procedure with the KCl standards, setting to 100 with 2.0 mM KCl,
then reading the others in reverse order. Then read the 1 in 50 diluted rehydration sachet
Finally, run water through the instrument until the flame appears free of colour again.
solution.
When the instrument is no longer required, switch off in the following sequence:
i. Turn off the gas control and the mains gas supply
ii. Wait for the flame to die out.
iii. Turn off the air supply.
iv. Switch off the electricity
v. Clamp the galvanometer.
CAUTION Although the flame is quite small, it has so high a temperature that contact of the
flesh with even the outer edge of the flame will instantly produce a thirddegree burn. Except for
lighting the flame, the hands should be kept completely out of the housing whenever the flame is
burning, even if it has been turned very low between analyses
REFERENCES
D. C. Harris Quantitative Chemical Analysis 4th Ed., W. H. Freeman and Company, New
York 1995 Chapter 21