Professional Documents
Culture Documents
Trainee Name:__________________
Functional description
One high flow rate pump and divert valve, used
for sample infusions.
One low flow rate pump without divert valve, used
for LockSpray.
Three shared 30-mL sample vials and plumbing
for wash and waste vials.
Stepper motors use 24V nominal supply.
2 1 2 1 2
Reference Valve Analyte Valve
3 7 6 3 7 6 1 3
4 5 4 5 4
Acquity
Pump
Pump
Source
Wash
Waste
1 Wash 3 Wash
Purge Cycles
1 Sample 1 Sample
Gas fail.
Source pressure test failure.
Source pressure trip.
Source door open, probe withdrawn or probe
connector disconnected.
Probe thermal trip.
Source temperature failed to regulate.
Probe temperature failed to regulate.
Fluidics leak detected or leak detector
unplugged.
There is a reset button in the console
Flashing Flashing ½
LED Symbol Steady Off Steady On
1/12 sec sec
Insufficient
Power off Pump off
Green power Pump ON
speed < 1 Hz speed > 1 Hz
supply
Yellow No Warning - - Warning
No
Red - - Malfunction
malfunction
Why does the ToF region take the longest to reach a good vacuum?
1.
2.
Task
Locate the pump override switch and confirm the software control (Auto)
and Override positions.
Goes into Standby when any of the turbo pump speeds fall below:
Environmental requirements
Operating temperature 15 to 28 °C
Safe operating temperature 5 to 40 °C
Storage temperature -20 to 60 °C
Mass temperature coefficient <60 ppm/°C
Relative humidity 20 to 80%, non-condensing
Weight ~266 kg or 586 lbs
Source Options
Source MassLynx UNIFI
1.7/1.8/1.9
ESI
APCI (IonSABRE 2)
ESCi
ASAP
APPI
APPI/APCI
NanoLockSpray
IonKey Supported on
SCN 939 and 937
APGC
Third Party Sources
DESI (Prosolia) Supported in
SCN 937 only
DART (IonSense)
LDTD (Phytronix)
Nanomate (Advion)
UNIFI
Instrument Type Acquisition Functions MassLynx
1.7/ 1.8/ 1.9
MS
MS/MS
Fast DDA
MSE Continuum
MSE Centroid
PID Product
PID Neutral
MRM
Additional summary
information is available:
Select View >
Summary Diagnostics
This brings up the
“Diag” tab
Instrument Tab
Enables access to additional experimental parameters
when in System View.
Quad Profile
The quad profile determines which m/z or
range of m/z ions are transmitted through
the quadrupole to reach the T-Wave region.
Auto Profile
Manual Profile
Manual Fixed
The voltage applied to the final lens before the pusher stack.
Tube Lens 30 25 Tune in Resolution Mode only between 25 and 38 - at each step
re-optimise using the Reflectron Grid and Pusher Offset.
Pusher 1900 1900 Fixed. The voltage height of the pusher pulse.
The offset voltage applied to the pusher plate Tune for
maximum resolution on the TOF.
Pusher Offset 0 0
Expected to tune between +3 and –3 V, however this is normally
negative.
Puller 1400 1400 Fixed. The voltage height of the puller pulse.
Resolution &
Parameter Notes
Sensitivity Modes
Puller Offset 0 Fixed. The offset voltage applied to the puller plate.
Stress to a customer:
System settings affect the
performance of their
instrument. Therefore it is
not recommended to export
any system settings unless
advised to do so.
https://connect.waters.com/files/app#/fold
er/61554384-7863-48d0-8e2e-
490a04ce47aa
IntelliStart workflow
Accessing IntelliStart
Via Sample List
Via MS Console
IntelliStart Functionality
Overview
Initial Pumpdown
Detector Conditioning
Standby States
Configuration Mode
Creating a Calibration
LockSpray Setup
Detector Setup
Detector Setup Theory
Automatic Resolution Optimization (ARO)
Normal Mode
Use a Calibration Profile
Calibration Check
LockSpray Check
Detector Check
LCMS System Check
Source Pressure Test
MS Console:
Source Standby
MS Operate
Launch the MS Tune Page
Pause/Run MS Console Plots
Start IntelliStart
Configuration Mode.
Normal Mode.
Standby buttons
Source Standby - Yellow State
Only turns off the capillary voltage, the gasses
and diverts the LC pump flow to waste.
All other voltages throughout the ion path of
the instrument (including ToF voltages) are
still present.
Full Standby - Red State
In system view this is available as a button.
In basic “Tune Pages” need to select:
Setup > “Instrument Standby”.
This does not turn off the gasses.
Turns off all voltages along the instrument ion
path.
For engineers and advanced customers.
Notes
Create Calibration
Can do an “Automatic” calibration where
IntelliStart set’s everything itself.
Can do an “Assisted” calibration where
IntelliStart runs the calibration and peaks can
be deselected prior to accepting the calibration:
Can also do “Manual Calibration”, but this is not
recommended due to lower reproducibility and
should not be encouraged with customers,
however this is necessary for some source options
that cannot use the fluidics i.e. APGC.
We recommend that this should only be
accessed by the person responsible for
calibrating and maintaining the instrument.
LockSpray Set-up:
A profile needs to be set-up and saved.
Optimisation of the LS source conditions is
then set-up automatically.
Please note the capillary voltage is not saved
during this set-up procedure – this is
different from the Xevo QToF MS instrument.
Perform this setup only when:
A new LockMass sample is to be used.
The LockMass check fails in Normal mode.
We recommend that this should only be
accessed by the person responsible for
calibrating and maintaining the instrument.
Detector Set-up
Only Leucine Enkephalin can be used as the
reference compound.
Choose how to control the fluidics:
For Nano-LockSpray select ‘Tune Page’.
Ensure correct mass is selected and user can
specify the polarity in which to run.
Detector voltage setup progress bar is displayed.
An Ion Area check starts once Detector voltage
has been set. This must be manually checked by
the user to be between 15 and 65 to confirm that
IntelliStart successfully completed a valid setup.
A green tick is displayed once setup has completed
successfully.
A non-interactive report will be displayed in a web
browser once the setup is complete.
Peak shape/asymmetry.
Sensitivity changes.
Questions Observe
Open source door, disconnect the gas exhaust pipe from the
waste bottle.
Close the source door and wait for source pressure test to
commence.
NOTES
MassLynx Setup
Creating a Project
Instrument Tuning
Default Parameters
Tuning Resolution Mode
Tuning Sensitivity Mode
Tuning Negative Ion Modes
Xevo G2 Tuning
Instrument Setup
Installation Process Overview
ADC Setup
Veff Check
Detector Setup
Quad Checks
Quad Tuning
Static Mode
Scanning Mode
LockSpray Setup
Creating Calibrations
Once
Step Basic Instructions
complete
Off = 6
Capillary 3.0 (0.3-3.2) Collision Energy Entrance 2.0
On = 15
Desolvation
280 (150-650) Pre-Filter 2.0 Exit 5.0
Temperature
Target
Cone Gas 0 (0-50) Pos Modes = 0.2 4
Enhancement Exit
Neg Modes = 0.8
Ion Energy Target
(Tuned in factory
Desolvation gas 600 (0-1000) 0-1 V) Enhancement 5
Extraction
Collision
Acceleration 1 5 (2-25) Collector* 60 6
Energy
SW 2
Transport 1 50 (35-60) Stopper Pulse* 20 25.0
Offset
0 (Tuneable as required,
Steering expected between -2 to Puller Offset* 0 Diff App 2 0.0
+2)
30 (25-38)
1.710 RF Settings
Tube Lens Reflectron Grid (Tuneable as
required) StepWave 300
Off = 6
Capillary 3.0 (0.3-3.2) Collision Energy Entrance 2.0
On = 15
Desolvation
280 (150-650) Pre-Filter 2.0 Exit 5.0
Temperature
Target
Cone Gas 0 (0-50) Pos Modes = 0.2 4
Enhancement Exit
Neg Modes = 0.8
Ion Energy Target
(Tuned in factory
Desolvation gas 600 (0-1000) 0-1 V) Enhancement 5
Extraction
Collision
Acceleration 1 10 (10-30) Collector* 60 6
Energy
SW 2
Transport 1 30 Stopper Pulse* 20 25.0
Offset
0 (Tuneable as required,
Steering expected between -2 to Puller Offset* 0 Diff App 2 0.0
+2)
25
1.710 RF Settings
Tube Lens Reflectron Grid (Tuneable as
required)
StepWave 300
Pusher* 1900 Flight Tube* 9.00
Compare it to your result in Step 3. The high intensity signal should show
a significantly higher resolution – this difference is the artificial
resolution enhancement.
Tune Reflectron Grid and Pusher Offset so that the instrument is tuned
to a resolution of >35,000 with GFP. (Check that your GFP signal
2
intensity is ~0.1 IPP in TDC centroid mode as you get close to the spec
resolution value).
Note: The 0.1 s Scan Time ensures that we collect data points at all Steering
values as we Tune.
Once you begin the acquisition, open the Chromatogram (from the top of the
2
Sample list) and ensure that the “Realtime” option is selected.
Use the “Right” arrow key on the keyboard to sweep the Steering value from
-2.00 to +2.00.until you locate where the maximum TIC is achieved.
3
Note: The chromatogram intensity should rise and then fall – examples of
what you should see are shown on the next page.
Then use the “Left” arrow key to sweep the Steering value back towards “more
4 negative” until the same level of intensity is observed on the chromatogram.
This will be your tuned value.
Resolution Mode
+0.40
-2.00
+2.00
Sensitivity Mode
+0.80
-2.00
+2.00
Once
Steps Basic Instructions
Complete
Combine the acquired data. Zoom in so that the GFP peak at m/z 785 is
the base peak in the window, and open the ResCalc program from the
2
desktop to measure the resolution (your trainer can assist you if needed).
A resolution will be displayed in the bottom left of the Rescalc screen.
Ensure the tune page window is focused on the GFP signal so that the
peak shape can be observed. Print the tune page to your training folder
4
using Windows XPS Writer or a PDF writer (name the file similar to
“Sensitivity Pos Tune Page”).
Expected Resolution
Mode
Result (GFP m/z 785.8)
Positive Sensitivity
Positive Resolution
Negative Sensitivity
Negative Resolution
Amplitude Threshold
time
ADC Setup
Amplitude Threshold
time
Amplitude threshold
Question
Are the “before” and “ after” values for your ADC setup the same?
Can two different sets of ADC setup values give the same final noise? (Discuss with trainer)
Detector Voltage
Pos Sensitivity
Pos Resolution
Neg Sensitivity
Neg Resolution
Close the Calibration Profile Editor, and select your Calibration Profile from
4
the drop-down list. Click Next.
Select the check-boxes for Pos Resolution and Pos Sensitivity Modes only
(we will perform the Neg calibrations later).
Ensure the two Options check-boxes are selected. Click Next.
NOTE: You cannot perform all four calibrations sequentially when using
5 two different calibrants. i.e. Pos Mode is using MSMS Sodium Formate
and Neg Mode is using (MS) Sodium Formate, which IntelliStart
determines to be to be different compounds due to the different
Reference files. However, you can do the two Pos Mode calibrations
sequentially, followed by the two neg Mode ones after.
Once
Steps Basic Instructions
complete
Leave the Calibration Setup wizard open. Open the Tune Page and switch
1
to Positive Ion Resolution Mode.
On the Tune Page infuse 0.5 mM Sodium Formate from the sample sprayer
2
at 5 μL/min.
Once you see a stable beam, enter MSMS Mode.
3 Mass range: 100-1500
Set mass: 770.8
4 Tune the parameters on the Source tab, to maximise signal intensity.
5 On the instrument tab, turn the collision energy ON.
Tune the collision energy between 15.0 and 20.0 V to optimise the
6
intensity of the high masses (>m/z 1100).
Save your tuning parameters. Note: You may want to make a note your
7
tuned parameters for the next step.
Switch to Positive Ion Sensitivity Mode. Input your tuned parameters from
8 Resolution Mode for the Source tab and collision energy.
Save your tuning parameters.
Return to the open Calibration Profile Setup.
For the Data Acquisition Options:
9 In Positive polarity select MSMS. Input the set mass and tuned collision
energy you setup on the Tune Page.
Click Next.
Select the check-boxes for Pos Resolution and Sensitivity Modes, and De-
select the two options for Negative Mode.
10
Ensure the two Options check-boxes are selected.
Click Next.
11 Select the “Tune Page” options for cone and capillary voltages. Click Next.
12 For the Fluidics, select the “Tune Page” option. Click Next.
On the Tune Page, ensure enough sample is available in the sample syringe
for several minutes (refill the syringe from Vial C and wait for a stable
13
beam if necessary).
Click Start on the Create Calibration wizard.
Pos Sensitivity
Pos Resolution
Neg Sensitivity
Neg Resolution
Installation Specifications
Applying LockSpray Profiles
Running the Standalone MS
Specification Tests
High Mass Resolution Specs (all Modes)
Sensitivity Specs (Positive Modes)
MMA Specification (Positive Sensitivity
Mode only)
Additional MMA Tests (for training only)
Sensitivity Specifications (Negative
Modes)
Saving System Settings
ToF Specifications
QToF vs. ToF
Activating the Quadrupole
Specifications
ToF Tuning for Sensitivity Increase
IntelliStart Functionality For Customers
Overview of Configuration Mode
Manual Calibrations
Resolution Optimization
Overview of Normal Mode
System Checks
Once
Steps Basic Instructions
complete
Ensure you are in Positive Ion Resolution Mode and that you
1
can see the Sample signal.
In each of the four Modes, set the sensitivity of the 1st isotope
of GFP (m/z 785.84 or 783.84) to ~0.1IPP (2e4 ADC Continuum
counts) using the parameters on the source tab, and/or
collision energy, or the pDRE test on the System 2 tab.
2 NOTE: If using collision energy, you need to update the MS
Method file with the desired collision energy value.
NOTE: If using the pDRE, the setting value is carried through
all Modes so you will need to setup and then run each Mode
individually.
On the Sample List, highlight the following experiment rows:
Commission_001_Resolution_PosResMode
3 Commission_002_Resolution_PosSensMode
Commission_003_Resolution_NegResMode
Commission_004_Resolution_NegSensMode
Click .
4 Select Acquire Sample Data and Auto Process Sample.
Click OK.
Once the acquisitions are complete print the SpecProc reports
5
and corresponding Tune Pages to your Training folder.
Once
Steps Basic Instructions
complete
On the Tune Page, ensure that you can see the GFP signal in each
1 positive ion mode, and that you have >20 mins of sample in the sample
syringe.
Optimise (maximise) the parent GFP signal (m/z 785.8) in each mode
2 using the parameters on the source tab (the same settings will be
required for both Modes).
Click .
4 Select Acquire Sample Data and Auto Process Sample.
Click OK.
Once the acquisitions are complete print the SpecProc reports and
5
corresponding Tune Pages to your Training folder.
Once
Steps Basic Instructions
complete
Set the tune page to Sensitivity Mode and purge the sample fluidics
1
with 500 pg/μL Raffinose from reservoir A (m/z 527).
Infuse 500 pg/μL Raffinose through the sample probe and 200 pg/μL
2 Leu-Enk through the Reference probe (at 0.5 or 5 μL/min depending
on the source you are using).
Click .
6 Select Acquire Sample Data and Auto Process Sample.
Click OK.
Once the acquisitions are complete print the SpecProc report and a
7
Tune Page to your Training folder.
Once
Steps Basic Instructions
complete
On the Tune Page, switch to Neg Resolution mode and ensure that
you can see the Raffinose signal (m/z 503) .
1
Be sure that use the correct sample that does not contain Formic
acid – check with your trainer.
Click .
5 Select Acquire Sample Data and Auto Process Sample.
Click OK.
Once the acquisitions are complete print the SpecProc reports and
corresponding Tune Pages to your Training folder.
Once
Steps Processing to be completed
completed
Check that your training instrument has the most up-to-date
1 firmware level for it’s components (there are seven firmware
files to check).
Questions
Assuming the instrument is on and all components are functioning correctly, what reasons could
cause MS SWAT to not be able to connect to the instrument?
What are the possible reasons for MS SWAT to not correctly match firmware to a PCB?
Off = 6
Capillary 3.0 (0.3-3.2) Collision Energy Entrance 2.0
On = 15
Desolvation
280 (150-650) Pre-Filter 2.0 Exit 5.0
Temperature
Target
Cone Gas 0 (0-50) Pos Modes = 0.2 4
Enhancement Exit
Neg Modes = 0.8
Ion Energy Target
(Tuned in factory
Desolvation gas 600 (0-1000) 0-1 V) Enhancement 5
Extraction
Collision
Acceleration 1 5 (2-25) Collector* 60 6
Energy
SW 2
Transport 1 50 (35-60) Stopper Pulse* 20 25.0
Offset
0 (Tuneable as required,
Steering expected between -2 to Puller Offset* 0 Diff App 2 0.0
+2)
30 (25-38)
1.710 RF Settings
Tube Lens Reflectron Grid (Tuneable as
required) StepWave 300
Off = 6
Capillary 3.0 (0.3-3.2) Collision Energy Entrance 2.0
On = 15
Desolvation
280 (150-650) Pre-Filter 2.0 Exit 5.0
Temperature
Target
Cone Gas 0 (0-50) Pos Modes = 0.2 4
Enhancement Exit
Neg Modes = 0.8
Ion Energy Target
(Tuned in factory
Desolvation gas 600 (0-1000) 0-1 V) Enhancement 5
Extraction
Collision
Acceleration 1 10 (10-30) Collector* 60 6
Energy
SW 2
Transport 1 30 Stopper Pulse* 20 25.0
Offset
0 (Tuneable as required,
Steering expected between -2 to Puller Offset* 0 Diff App 2 0.0
+2)
25
1.710 RF Settings
Tube Lens Reflectron Grid (Tuneable as
required)
StepWave 300
Pusher* 1900 Flight Tube* 9.00
Once
Steps Basic Instructions
complete
Pick a Mode and what peak characteristic(s) you want to to try to
1 improve. Take a Tune Page Printout of your currently tuned
parameters in that Mode (for reference later).
In Pos Ion Resolution Mode, retune the instrument using the
appropriate additional parameters previously discussed.
•Maximise sensitivity by observing the Tune page readback.
2
•Ensure that you still meet the resolution specifications.
•Ensure that you retain a symmetrical peak shape.
Save your new settings.
Switch to Pos Ion Sensitivity Mode, retune the instrument using the
appropriate additional parameters previously discussed.
•Maximise sensitivity by observing the Tune page readback.
3
•Ensure that you still meet the resolution specifications.
•Ensure that you retain a symmetrical peak shape.
Save your new settings.
On the Sample List, copy the relevant experiment rows (depending
on what peak characteristic you are investigating) and rename the
file as appropriate, e.g.:
4
Commission_002_Resolution_PosSensMode
or
Commission_008_MSMS_PosSensMode
Once the acquisitions are complete print the SpecProc report and
6
corresponding Tune Pages to your Training folder.
Compare the report you have just generated to the reports you
created previously for Resolution and Sensitivity Modes.
7
Fill in the table below, as appropriate. You can compare these
results of the specifications you ran earlier.
Remember!
These parameters should only be
used when necessary, not as regular
practice during install.
Configuration Mode
This area should be the responsibility of the
system administrators. Normal users will
likely not routinely access this Mode.
At this point you should have completed an
Assisted calibration in Positive Ion mode.
Complete an automatic calibration with Na
Formate and fill in the table below:
Save a copy of the Automatic Calibration
report to your training folder.
Run and Print
Function Description
Report
Done with
Detector Setup
specs
Done with
LockSpray Setup
specs
Instrument Automatic
Configuration
Calibration
Done with
Setup Assisted
specs
Manual
Resolution Optimisation
Note: 0.1 IPP is very important for running detector setup, calibration and LS Setup.
IntelliStart uses the DRE lenses to achieve this, but if the signal intensity is not set to a
reasonable level (~1e5 ADC cps) before the Setup is started then it may not be possible
for the software to achieve 0.1 IPP with the DRE lenses, which may then affect the
ability of the system to achieve its goals.
Once
Steps Basic Instructions
complete
You only really need to calibrate one Mode here (you can ask your
Instructor if you have more time to run more Modes), but we do
want to use this for the calibration check later.
You should be able to work your way through the wizard without
instructions, but consider the following:
Check there is a suitable sample beam on the Tune Page in the Mode(s)
you plan to calibrate.
Type: ToF MS
Data Format: Continuum
Dynamic Range: Normal
Run Duration: 1 min
2 Scan Time: 1 s
Mass Range: 50 – 1200
Precursor mass: None
LockSpray: No LockSpray.
Once
Steps Basic Instructions
complete
Take a tune page printout to record your current tune settings
and peak shape in Sensitivity Mode. Save it to your training
folder with an appropriate name.
Begin by running ARO for Sensitivity Mode, starting from your
1 tuned settings that you have used to run the specifications.
Infuse 200 pg/μL Leu-Enk at 5 μL/min from the LockSpray sprayer.
Ensure you have a stable Leu-Enk signal, and zoom in on the 1st
isotope. Set the signal intensity to ~0.1 IPP (TDC centroid Mode)
using the capillary voltage.
In the MS console select:
Configure > Configuration mode > Resolution
Optimization
In the wizard:
Select a valid LockSpray profile that you have created.
2
On the Options page, select the check-box for Sensitivity Mode
only, and ensure that “Display Report” is selected.
Click Next.
Click Start.
Once the acquisition is complete print the report and another Tune
3
Page to your Training folder.
Using the report and the two Tune Page printouts, describe what settings the Resolution
Optimization process may have changed and any effect on the resolution, sensitivity and peak
shape. NOTE: the optimization may not be able to improve the resolution from your tuned
settings – if this is so, the report will report the exact same resolution for before and after the
optimization.
Resolution Resolution Sensitivity Sensitivity Peak shape Peak Shape
Before After Before After Before After
Once
Steps Basic Instructions
complete
Now detune the instrument settings of Sensitivity Mode
4 (Reflectron grid and Pusher Offset) so that the Leu-Enk resolution
is <20,000. Try to retain a symmetrical peak shape as you do this.
Take a tune page printout to record your current tune settings and
5 peak shape. Save it to your training folder with an appropriate
name.
In the MS console select:
Configure > Configuration mode > Resolution
Optimization
In the wizard:
Select a valid LockSpray profile that you have created.
6
On the Options page, select the check-box for Resolution Mode
only, and ensure that “Display Report” is selected.
Click Next.
Click Start.
Once the acquisition is complete print the report and another
7
Tune Page to your Training folder.
Using the report and the two Tune Page printouts, describe what settings the Resolution
Optimization process has changed and any effect on the resolution, sensitivity and peak shape.
NOTE: The algorithm is limited to how far it can change the settings in one cycle – if the
resolution has not increased (returned) to a similar level that you had tuned yourself, run the
optimization again.
Resolution Resolution Sensitivity Sensitivity Peak shape Peak Shape
Before After Before After Before After
Normal Mode
This area is routinely used by normal users.
It allows selection of profiles for use and
checks on instrument/system readiness.
Complete each of the instrument checks on
the profiles/setups that you are currently
using.
They should all pass, but discuss with your
Trainer what should be done if they fail.
The System Check will be covered later.
Run and
Function Description Print
Report
Calibration Profile
Detector
Check
LockMass
Check
Detector Check
This is a quick experiment that measures the
ion area of the detector.
This is compared to the value saved during
the last Detector Setup.
A change of >20% causes a fail.
If this fails – run Detector setup.
LockSpray Check
This looks at the LockSpray beam using the
settings in the profile.
If the signal is not an acceptable intensity for
the instrument to use as a Lockmass, it will
cause a fail.
The simple fix is to rerun the LockSpray
Setup, however, the reason for the fail may
need to be investigated:
Different (wrong) capillary voltage used?
Is there a flow/spray issue?
Is it due to instrument sensitivity decrease?
Is the sample preparation consistent?
Etc.
Calibration Check
This will collect data on the calibrant (in the
same way as Setup does) and compare the
results to a completed Calibration Setup.
Acceptance criteria are required as a pass/fail
threshold.
Once
Steps Basic Instructions
complete
What determines if this test passes or fails? What part should be changed if this test fails and
what is it’s part number?
Once
Steps Basic Instructions
complete
LockSpray Crosstalk Test Pass/fail
Read the instructions for the :
LockSprayCrosstalk_XX001
7
Note: It is best to run this test with no flow from the sample sprayer
as any contaminant around the Leu-Enk m/z in the sample spray will
factor in to the calculation.
What should you do if this test fails?
Input your results into the DRE.xls document to calculate the r2 value
for the linearity.
What should you do if this test fails?
Candidate:
This is an intermediate level for Waters personnel
where an FSE Consumer receives training and
mentoring to attain Author status.
FSEs can apply to become a Candidate by
emailing KCSFeedback@waters.com.
Author:
This advanced level allows the user to create and
update articles in real-time.
Mentor:
The expert level which involves the user in
mentoring Candidates.
Only attained by experienced users that
demonstrate understanding and consistent use of
best practices for authoring articles.
Once
Step Setting Up On an Instrument PC
complete
Once
Step Basic Instructions
complete
Access KCS.
1
https://support.waters.com
Discuss as a group what you feel is useful about each of these tabs
4
(make any notes in the space below this table).
On the “Tag Directory”, review the Service Alerts for the Xevo G2-XS
(you can also consider Xevo G2 and G2-S).
5 We do not have time to review all of these Service Alerts in detail,
but ensure you review the current titles and access some that
interest you.
Access KCS.
1
https://support.waters.com
4 Discuss as a group what you feel is useful about each of these tabs.
On the “Tag Directory”, review the list of articles for the Xevo G2-XS
You could also consider Xevo G2 and G2-S, however many of the
5
articles will be common to all these instruments due to their high
similarity in terms of their components and operation.
https://success.mindtouch.com/Support/Content_Man
agement/Search/Perform_advanced_searches
Once
Step Basic Instructions
complete
In the main “How can we help you?” search bar, type:
1
aperture plate
What instruments or system modules do you get results for?
2
Approximately how many results do you get now?
Return to the main “How can we help you?” search bar, and type:
3
aperture
What instruments or system modules do you get results for?
4
Approximately how many results do you get now?
Return to the main “How can we help you?” search bar, and type:
5
“aperture plate”
What instruments or system modules do you get results for?
6
Approximately how many results do you get now?
Return to the main “How can we help you?” search bar, and type:
7
title: aperture AND title: “xevo g2-xs”
What instruments or system modules do you get results for?
8
Approximately how many results do you get now?
Once
Step Basic Instructions
complete
In the main “How can we help you?” search bar, type:
1
1350V on quad pcb
Return to the main “How can we help you?” search bar, and type:
3
1350v
4
Approximately how many results do you get now?
Return to the main “How can we help you?” search bar, and type:
5
1350
Once
Step Basic Instructions
complete
In the main “How can we help you?” search bar, type:
1
detector
What instruments or system modules do you get results for?
Return to the main “How can we help you?” search bar, and type:
3
magnetof
What instruments or system modules do you get results for?
4
Approximately how many results do you get now?
Once
Point Basic Instructions
complete
Review your question and pick one or two keywords that represent
the entire question.
2 Type these into the search bar (consider inputting them individually
using the Boolean operators)
Review the results you get.
Consider refining further using “Search Constraints” (a list of possible
search constraints is on the next page). They should be written in the
format:
3
<constraint>: <text>
Review the results you get for each constraint you choose to try.
Notes:
Lists all pages that are tagged with the specific tag you
are searching. (You can also click the tag in Page
tag:<tag> Settings on any page to trigger a search for all pages
associated with that tag.)
System
List All parts of the System If no solution has been found thus far,
potentially related to your think about the instrument from the
problem: front to back (or back to front).
Source
Quad
Collision Cell
Transfer Optics
Pusher Stack or
Unit
Reflectron
Detector
Signal Line
Should be performed or
Calibration checked by Customers on a
(4-5 d.p. mass acc.) regular basis (e.g. weekly)
LockSpray.
Provides a mass position correction for changes
in temperature over time.
A LockSpray profile should only need setting up
when new LockSpray samples are made or if LS
sensitivity dramatically changes (if sample
concentration and sensitivity is consistent the
profile will remain valid for a long period of time).
However, it may be simpler for Customers SOPs
to redo LS setup when they recalibrate (weekly).
It is beneficial for the selected LockSpray
compound to have m/z in the region of analytes
of interest.
Dual point LS profiles may be beneficial if
analytes of interest span a wide mass range.
NOTE: LockSpray corrections are only applied to
data acquired from the sample list. Customers
should always run with a LockSpray. They should
select the “Acquire and apply correction” option,
unless they plan to export the data into another
processing software, then “Acquire but do not
apply” should be selected.
ON/OFF ON/OFF
Command Control Status
Red variables
To create a threshold:
Use the “+” button to add a threshold.
Select the parameter to monitor from the
drop-down menu on the left.
The threshold can then be set in the drop-
down menu on the right.
The threshold can be set as:
A specific value.
A value range.
Data
Used to apply filters to the parameter list on
this tab.
Charting
Selects the parameters to record from your
filtered list.
Record
Begins to record the parameters selected in
“Charting”. Switches to a Stop recording button
when recording.
Delete Data
Clears the data collected in the current chart.
Zoom/ Reset zoom
You can zoom-in on areas of the chart by click-
and-drag with the mouse. The “Reset Zoom”
button zooms out to the full chart.
Chart Colours
Used to set the colours of the data lines/points
of the recorded parameters.
Selecting reset in here will apply different
colours to the various selected parameters.
What other parameters would you suggest this charting function could be useful
for? Please list any successful tests you perform.
DESOLVATION_GAS_FLOW_RB
SOURCE_PRESSURE_STATUES
Open the source door – close it and monitor with engineer’s dashboard
What is observed?
Open the source door – unplug the exhaust from the back of the instrument –
close the door
What is observed?
Readback
Any comments or observations
Measurement
Reflectron Voltage
Reflectron Current
NOTES
Completed
TASKS
Remove the Ion Block:
If sensitivity on the instrument is low or if significant
1
ion burns can be seen, clean the ion block. Your
trainer will provide details on this.
Remove the StepWave Ion Guide:
Follow the procedure and correct tools detailed in the
StepWave Technical Reference.
2
Discuss with your trainer the StepWave cleaning
procedure.
If necessary clean the StepWave.
Remove the HT unit:
3 The HT unit is heavy and should be moved by two
people.
1. Aperture Plate
A very common issue often encountered at PM.
Some Customers may need more than one
Aperture Plate per year depending on their
use/application of the instrument.
SYMPTOMS
Poor resolution (untunable).
Resolution decrease over time.
May be evident between start and finish of a long LC
run.
CAUSES
Contamination (charging).
Frequent use of Nano-applications, Higher Resolution
Modes, or long LC runs can make this occur more
quickly.
Additionally, poorly tuned Steering (System 1 tab) in
any mode will increase the rate of contamination.
TESTING
Compare resolution over a period of time.
Re-tuning. In particular, the need for significant Tube
Lens tuning to achieve acceptable resolution suggests
contamination.
FIX
Regularly running resolution optimisation in
IntelliStart is recommended for Customers who are
susceptible to this problem.
If retuning the instrument by IntelliStart or FSE
cannot resolve the issue, aperture plate must be
replaced.
It cannot be cleaned adequately in the field.
2. Protection PCB
Part of the purpose of this component is to
fail to protect more expensive components
from voltage spikes or surges.
Damage to the SMA connectors is likely
only caused by FSE actions.
SYMPTOMS
Excessive “ringing” (See next page for further
explanation)
Low sensitivity or no beam.
CAUSES
Over tightening or physical damage to SMA
connectors.
High voltage spikes from ToF region damage
PCB.
TESTING
Test by DVM to check SMA connections.
Substitution with new PCB.
Having a “test” PCB in an FSEs personal stock
for this is useful.
FIX
The only fix is replacement of the part.
3. Pusher Unit
This is a hard working unit, producing 2kV pulses
at up to 30,000 Hz almost all the time.
SYMPTOMS
Poor resolution.
Unstable resolution.
Unstable mass accuracy.
ADC not acquiring (no trigger signal).
No beam (unit not turning on).
CAUSES
Voltages low or inconsistent.
No trigger being sent to ADC (check Trigger LED on
ADC). No electronic noise on Tune page.
Not turning on will be caused by issue internal to the
unit (assuming all other systems have power).
TESTING
Voltages - Check readbacks (Pusher tab, System 1).
Observing the pusher artefact on the Tune Page may
give further confirmation if voltages are inconsistent.
Check LEDs on the unit to look for issues (these can
give false positives depending on Firmware).
Trigger – Reduce ADC baseline threshold to confirm
no electronic noise seen. Check trigger and signal
cables at the ADC are in the correct ports. Check
continuity of the trigger cable.
FIX
Rebooting the system can clear errors. Give it some
time to allow capacitors to properly discharge.
Internal faults to the unit require replacement of the
unit, as there are no serviceable parts inside.
4. ADC PCB
NOTE: This part is commonly misdiagnosed as a
problem and replaced without proper
troubleshooting.
SYMPTOMS
No communication.
No data being acquired.
Peak splitting (all peaks in spectrum are doublets
usually <0.5 Da apart).
CAUSES
No Comms - EPC-ADC communication not occurring.
No Data - Data input signal interrupted at PCB. No
SIP signal or Trigger signal. Is there power to PCB?
Peak splitting – caused by the two alternate data
collection “halves” of the ADC not booting up in sync.
Incorrect firmware – however this will likely be due to
FSE action.
TESTING
Check firmware revision from ADC Diagnostics Tab
and compare to ADC Firmware Service Note (this is
unlikely to be wrong, but is worth checking).
Comms – establish comms with the ADC via the
debug port bypassing the EPC (See ADC Tech Ref for
instructions). This will confirm ADC vs. EPC problem.
Also confirms whether ADC is ready to acquire data.
No data – Confirm no electronic noise can be seen by
reducing the ADC baseline threshold (ADC tab). If
there is noise, the ADC is not the problem.
No Trigger, SIP or power – Check LED on board is on.
If not, troubleshoot Trigger signal from Pusher unit,
SIP from Quad PCB or power from power module.
FIX
Firmware - Reflash correct firmware.
Peak splitting – reboot the electronics. It may take
more than one attempt to clear.
Only when all other possibilities have been
eliminated, replace the ADC PCB.
CAUSES
No spray - Blocked capillary. Leaks. Blocked
filter.
Unstable spray – Bent tip. Check the
protrusion of capillary from nebuliser sheath,
>3mm can cause beam instability (excessive
protrusion can be caused by use of high flow
rates >20 μL/min).
TESTING
Visual inspection of capillary tip.
Work backwards to ensure flow is present at
each section.
FIX
Reseat capillary if protruding excessively.
Discuss flow rate with Customer.
Replace Capillary or filter if blocked.
If capillary blocked, install a filter if not already
present.
6. StepWave Assembly
Be aware that charging issues happen throughout the
instrument. As the StepWave is near the front, it does
get dirtier, quicker. However, other components can be
the problem part.
SYMPTOMS
No Beam.
Low sensitivity.
Decreasing sensitivity over time (minutes).
CAUSES
Voltages low or missing.
Physical damage.
Improper cleaning.
Dirty from samples (charging).
TESTING
Readbacks on StepWave diagnostics tab.
Check voltage settings are correct.
If altering Dif Ap1 and 2 voltages from default brings a
sensitivity increases, this hints at charging.
If lowering Static Offset to 0 restores a beam, voltages
are missing somewhere, but no necessarily the
StepWave.
Eliminate delivery of voltage to the assembly -
Continuity of cables. Actual output with DVM at cable
ends or PCB port. Continuity of StepWave flange.
Visual inspection of pogo pins on StepWave flange.
Visual inspection for damage.
Continuity of PCBs on StepWave Assembly.
Review your cleaning methodology – if using MS
cleaning solution, take extra steps to remove this.
Also consider cleaning the pads that connect to the
pogo pins with methanol and a cotton swab.
FIX
Any damage requires replacement of part.
If charging, ensure a thorough and proper cleaning
procedure is performed. Clean ion block also.
Any stuck pogo pins require a new StepWave Flange.
7. MagneToF Detector
Detectors should last a few years (~5), but this is
dependent on instrument usage.
SYMPTOMS
No Beam.
Low sensitivity.
Excessive ringing.
Fast detector aging.
CAUSES
No beam or low sensitivity - Physical damage, failure of
detector or too low a voltage. Detector age (i.e. required
voltage exceeds maximum allowable voltage).
Excessive ringing can occur from too high a detector voltage
setting.
Detector aging will be increased by the number of ions hitting
it (sample conc., long LC experiments, nano-applications).
Significant increases are normal early in the detector lifetime,
which should then slow once it beds in.
TESTING
Readbacks for voltage and current on Diag tab. Wrong
readbacks could be due to the voltage cable from HT not being
seated correctly, or the HT unit is faulty.
Manually change Detector voltage (Auto values) to observe
changes.
Check voltage settings are correct.
IntelliStart Detector Check and Setup.
Visual inspection of detector for damage and that voltage
cables are properly seated..
DVM and Megger tests (ToF Optics Tech. Ref.).
FIX
Detector setup. Discussion with Customer on regularity and
running Detector Check.
Faulty or damaged unit requires replacement.
Detector aging should be controlled by running Detector Check
weekly, and only running detector setup when this fails.
Customers should reasonably limit sample concentrations.
Event Control should be enabled in the System 2 Tab, and TIC
Control if the Customer runs nano-applications.
Discuss with Customers appropriate intensities they should
see (more intensity is not better!). Advise on best practices to
limit ion current at the detector (sample concentration,
Extended dynamic range experiments).
8. HT Unit
This component should last a long time if
properly used. May fail in positive or negative
polarity only.
SYMPTOMS
No Beam.
Poor resolution.
Fluctuating Resolution.
CAUSES
Voltages low or missing.
Voltages unstable.
TESTING
Readbacks on Diag tab (Detector Reflectron,
Reflectron Grid, Flight Tube) and HT I/O tab.
Ensure cables are seated correctly.
Cable output can be measured with DVM and HV
Probe, however, this is not normal practice and
should only be done if absolutely necessary
(Danger HV!).
Engineers dashboard recommended to get fine
detail on voltage stability at a high sampling rate.
Large irregular dropouts hint at voltage
breakdown inside the ToF region, which is not a
faulty HT unit.
FIX
Customers should only use Source Standby
regularly. The high voltage should be left on for
stability and lower stress to the hardware.
Faulty units (or cables) must be replaced as a full
assembly.
Voltage outputs
Read the Power Module PCB technical reference.
Perform the following procedure:
Ensure that the instrument electronics are
switched on.
Using a digital voltmeter, measure the following
voltages on the diagnostic connector.
Useful for troubleshooting if one of the
following doesn’t appear to have power:
Vacuum (turbo pumps)
RF power (RF generator, collision cell, StepWave)
Electronics (main electronics unit, all PCBS,
pusher unit)
Heater (Source or desolvation heaters)
Vacuum_supply
RF_supply
Electronics_supply
Heater_supply
Thermocouple 1 2 <5Ω
Desolvation
7 > 10 kΩ
outer nozzle
RF Frequency 1 RF Frequency 2
Standby
Operate
What could be the reason for a much higher RF 1/2 frequency readback in
Operate mode?
On a Xevo G2-S (or XS), where does in-source fragmentation occur and what
setting can be used to control it?
Sample
Cone
DC Offset 2
Diff Ap 1
Diff Ap 2
Once
Steps Basic Instructions
complete
Create a new experiment with the File name “Quad Transmission Test”.
2
Create a new MS Method from the sample list (right click in an empty slot
in the “MS File” column).
Add two functions, one MS and one MSMS (Note which type of
3
experiment is labelled as function 1 and 2).
ToF MS Tab:
Low Mass: 500
High Mass: 600
Scan Time: 1 second
4 Data Format: Continuum
Ensure the Tune page is set to Resolution Mode, and infuse Leu-Enk from
the Reference sprayer @ 5μL/min.
In MSMS Mode, set the LM Resolution slider on the instrument tab to 15.
6 Run the experiment from the Sample List.
Once completed, open the chromatograms for the 2 functions. Combine
the entire experiment for each function. (ensure that all of the scans for
7
each function are included). The results should look similar to those shown
below.
Zoom in on the 557.2771 isotope in each function. Perform an intensity
percentage calculation:
(MSMS intensity / MS intensity ) x 100
Pass/Fail
Ratio of 557.27 peak heights in MS
mode and MS/MS mode (must be
>45%)
Pass/Fail
Pass/Fail
1. Find all four quad peaks. – Start with a wide quad window
and reduce LM Resolution and Resolution slider values, until
definite peaks can be seen for all four masses.
2. Achieve resolution on the lowest mass peak, 172.88 –
tune LM Resolution until you achieve 1.0 Da ±0.1. Correct
mass position is not important at this time unless you need to
move the peak into the quad window as you reduce its span.
3. Achieve reasonable resolution (see also point 4.) on the
highest mass peak, 1971.61 – tune Resolution until you
achieve 1.0 Da ±0.1. Again mass position is not important at
this time unless you need to move the peak into the quad
window as you reduce its span.
4. Adjust the resolution of the middle mass peaks, 622.57
and 1072.25. – tune the Linearity. Increase value to increase
resolution. Note: this will have significant effect on the 1971.61
peak, as this is in the middle of the mass scale for a 4K quad!,
so some further adjustment (balancing) of the Resolution
slider may be necessary.
5. Fine adjustment of the resolution to ensure that all peak
widths are 1.0 Da ±0.1.
6. Set mass positions. – Finally, centre the peaks in their
windows as best you can by balancing the Low Mass Scale
Adjust and High Mass Scale Adjust.
Stopper
Pin 2 Pin 2
50% 10%
Record the Tune Page readbacks and measure the actual voltages out-of-line on
the Fischer connection following the instructions in the technical reference.
Expected Measured Correct
Parameter Setting RB Value
Value value
Acceleration 1 40
Acceleration 2 200
Aperture 2 80
Transport 1 90
Transport 2 50
Steering 5
Pusher 1900
Puller 1400
Entrance 17
Observe
Questions
What does observing the Pusher Artefact prove is functional (list as many
components as possible)?
750 ±50
Core lead across single AC
coupler
Detector
Follow the ToF Optics Technical Reference to:
Inspect parts for damage.
Perform measurements on either a spare
detector or the detector from the instrument
and fill in the table below:
Expected Resistance
Component Device
MagneTof A MagneTof B MagneTof C
R1 100 ±7 Ω 100 ±7 Ω 10 kΩ
R2 47 ±7 Ω 47 ±7 Ω 100 Ω
R3 <0.5 Ω <0.5 Ω 10 kΩ
DVM (set to
SMA outer resistance)
threads and
100 ±10 kΩ 100 ±10 kΩ N/A
top of the
detector
Securing lug
(+) and gold 5.5 ±1.25 MΩ 3.8 to 6.2 MΩ 10 ±2.5 MΩ
pin (-)
Megger
Strike plate (+)
and detector N/A N/A 10 ±2.5 MΩ
bias (-)
1 Stop the sample flow and turn the API and Collision gas off
From the Auto Values window, note down the detector voltage
3 set by IntelliStart. Set the detector voltage 200 V below the
detector setup value.
If > 200 on the TIC is observed generally over the entire experiment, first check that the ADC
setup is correct then repeat the test. If the test shows large spike(s) during the experiment, this
could indicate breakdown at the detector, a faulty detector or possibly electrical breakdown in
the Reflectron.
Once you have completed the test(s), return the detector
voltage to the value previously determined by the Detector
6 Setup, the Ion Energy to its initial value (~0.2) and the Entrance
value to 17.
6.97e4 2.43e4
1: TOF MS ES+
2.42e4
391.1998_
Symptom:
MS 2.07e5 MS 2.07e5
MS/MS MS/MS
55 1.72e5
Symptom:
!
556.2896
! 64758 278.1085 397.1841
!
278.1373 10424 16870
397.2197
51476
96760 425.1734
6561 556.2593
! 5301
425.2174
35657
Symptom:
Symptom:
Symptom:
Mobile phase A1, A2, SM Water + 0.01% formic acid and 0.05% Ammonia solution (pH9)
Purge & Seal Wash Add 100 µL formic acid and 500 µL Ammonia solution to 1 L water
Once
Steps Basic Instructions
Complete
“Commission_ACQUITY_QTof.SPL”
Click on the ACQUITY Console Icon to launch the Console
4
interface.
From the Console, select: Control > Startup System.
Prime each solvent line and seal wash line for 5 minutes.
Prime the Sample Manager Strong wash, Weak wash and
5 sample syringes for 5 cycles.
Characterise the Sample manager seal.
Characterise the Sample manager needle and loop
volumes.
From the Equilibrate to Method tab , enter a flow rate of 1.0
6 mL/min, and select a 50:50 flush for solvents A1 and B1.
Perform the flush for at least 2 minutes
Once the system flush has completed set the flow rate to 0.8
mL/min and 90% A1 and 10% B1. Set the column temperature
7
to 40ᵒC and the Sample temperature to 15ᵒC. Ensure the system
has stabilized and equilibrated before proceeding.
Any Comments:
Calibration Results
Mode RMS Matched Peaks
Pos Sensitivity
Neg Sensitivity
Once
Steps Basic Instructions
Complete
Infuse 200 pg/µL Leucine Enkephalin solution into the LockSpray probe at
1 20 µL/min. Ensure that a spectrum of Leucine Enkephalin is observed on
the MS Tune page.
For LCMS acquisitions, we need to perform a Dual Point LS setup to ensure
that low m/z signals (>500 Da) are properly LockMass corrected.
In Pos Ion Sensitivity mode, set the LockSpray Collision Energy to 21 and
2 ensure the fragment ion at m/z 120.08 and the parent at m/z 556.28 are
observed. Tune the collision energy so that the two signals are of
approximately similar signal height. Use the capillary voltage to control the
signal intensities to ~2e4 cps. (this is ~equivalent to 0.1 IPP).
Switch to Neg ion Sensitivity mode, set the LockSpray Collision Energy to
32 and ensure the fragment ion at m/z 179.1 and the parent at m/z 554.26
3 are observed. Tune the collision energy so that the two signals are of
approximately similar signal height. Use the capillary voltage to control the
signal intensities to ~2e4 cps. (this is ~equivalent to 0.1 IPP).
With the ACQUITY UPLC flowing through the sample probe, perform a Lock
Mass setup through the MS console.
4 Note: The LC flow can cause interference with the LockSpray signal. LS
setups are recommended to be performed with the LC flow infusing into
the source. This is also why it is recommended that the LS uses a higher
flow rate (20 µL/min).
Open the MS console, and click Configure > Configuration Mode.
5
Select LockSpray Setup, then Start.
Create a new LockSpray Profile and name it appropriately. In the LockSpray
Source Setup dialog box, select the Leucine Enkephalin (system Lock mass
compound) from the drop-down list.
6 Ensure that the dual point LockMass peaks are selected:
Positive ion: 120.0813 & 556.2771 (CE = 21 or tuned value)
Negative ion: 179.0821 & 554.2615 (CE = 32 or tuned value)
Click Next.
From the Setup Type window, select the Tune Page check box to use the
7 current source conditions.
Click Next. Click Start.
Once completed, include these Profiles within the Pos and Neg MSE
8
Acquity methods in your sample list.
Once complete, print the LockSpray reports to XPS or PDF file and save
9 them to your training folder (named appropriately).
Print a corresponding tune pages, and save them to your desktop folder.
Parameter Setting
Capillary 0.5 kV
Sampling Cone 40 V
Source Offset 80 V
Source Temperature 140 °C
Desolvation Temperature 600 °C
Cone Gas 100 L/hr
Desolvation Gas 1000 L/hr
Once
Steps Basic Instructions
Complete
Ensure that the following peaks ([M+H]) can be seen in the tune page spectrum in
Positive Ion Sensitivity mode:
Acetaminophen m/z 152.0712
2 Caffeine m/z 195.0882
Sulphadimethoxine m/z 311.0814
Verapamil m/z 455.2910
This is to confirm that all of these chemical components are present in our samples.
Observe the Sulphadimethoxine signal at m/z 311.0882. Tune the sample probe
position to ensure good signal.
Note: The optimum sprayer position at high flow rates will be further away from
the sample cone than the position for low flow rates (MS only). This may not give
maximum absolute sensitivity on m/z 311, but should improve S:N.
3
Once the probe position is tuned, observe the intensities of the other sample
components. If the 311.0814 signal is >2e5 the other signals should show intensities
of >2e4 ADC counts. The Verapamil signal is often the most difficult as this
compound is sensitive to degradation.
Switch to Negative Ion Sensitivity mode and observe the following peaks ([M-H]):
Sulphadimethoxine m/z 309.0658
Chloramphenicol m/z 321.0045
4 These signals should have sensitivities >2e3.
Note: If sensitivity is particularly poor in Neg Ion mode and the probe position
needs re-optimising to fix this. Perform the following setup instructions and LCMS
tests for Pos Ion and Neg Ion modes separately.
Interval (sec) 15 15
Scans to Average 3 3
Method Events
Enable Method Events
Time Event Action System
Initial Conditions Flow State LC Sample
Once
Steps Basic Instructions
Complete
On the Tune Page. Ste the capillary voltage back to 0.5 KV in both Pos and
Neg Sensitivity modes.
Capillary voltage: 0.5 kV (or a tuned value from the tuning
optimisation earlier)
1 Sampling Cone: 40
Source Offset: 80
Source Temp.: 140
Desolvation Temp.: 600
Cone Gas flow: 100
Desolvation Gas flow: 1000
Highlight all samples Commission_ACQUITY_001 to
2 Commission_ACQUITY_009 in the “Commission_Acquity” sample list, and
click Run > Start.
Select the Acquire Sample Data check box only.
Click OK to perform the analysis.
3 Once the acquisitions are complete, select Commission_ACQUITY_004 to
Commission_ACQUITY_009, click Run > Start.
Select the Auto process Samples check box only.
Once complete, print the SpecProc Summary report to XPS or PDF file
4 and save it to your desktop folder.
Print a corresponding tune page, and save it to your desktop folder.
Once the Pos Ion acquisitions are complete, select
We will now discuss extracting the following
Commission_ACQUITY_010 to Commission_ACQUITY_018 and click: Run
specification data from the SpecProc
> Start.
Select the Auto Process Samples check box only.
5 reports:
Click OK to perform the processing.
Once the acquisitions are complete, select Commission_ACQUITY_013 to
Commission_ACQUITY_009, click Run > Start.
Select the Auto process Samples check box only.
Sensitivity Mode
The average signal-to-noise ratio (10 pg/μL on-column
Average signal-to-noise Sulfadimethoxine) for the six replicate injections (RMS, ignoring
zeroes and with no extra processing) is ≥2000:1
The peak areas Sulfadimethoxine) over six replicate injections have
%RSD of peak area
%RSD ≤3.0%
Mass measurement
The RMS error for the 4 components is ≤2.0 ppm
accuracy
Retention time The peak retention times over the six replicate injections have a
reproducibility standard deviation of ≤0.047 minutes (≤2.82 seconds)
Retention time The peak retention times over the six replicate injections have a
reproducibility standard deviation of ≤0.047 minutes (≤2.82 seconds)
If you have any outliers in your data, calculate the TDC cps for the
highest intensity outlier you can see.
Default values of the Peak ID Tab (left) and Mass Difference Tab (right)
of 6mix_Pos.olp viewed in SpecProc editor
However...
If outlier intensities are greater than our
peaks of interest, we will not be able to
eliminate them using intensity alone as we
will start to miss our peaks of interest.
High outlier intensities near the m/z of peaks
of interest have been noted when using low
quality water for the mobile phase of the
UPLC.
Mass 311.0814
Rt STD Dev
Click the Windows Start button and select Programs > Accessories > Calculator.
The Windows calculator appears.
Mass 321.0045
% RSD Area
Sulfadimethoxine Chloramphenicol
Mass
m/z 309.0658 m/z 321.0045
Rt STD Dev
Overview
Instrument description.
Status Panel LEDs.
The vacuum system.
Instrument communications.
Host PC Username and password
Gas and Electrical connections.
MS Console
Overview of page.
Control.
Configure.
Maintain.
Troubleshoot.
IntelliStart
Overview (Normal and Configuration Modes)
Calibration
LockMass
Detector
IntelliStart (cont.)
Resolution Optimisation
MS method editor
Tune Page
Summary diagnostics
Source options
Quad Profiles
Target Enhancement
Inlet Editor
Connecting to inlets
Source Options
APGC
DESI
REIMS
Customer Familiarization
Additional resources
More accessible
More reliable
More productive
Continuous
background system
Integrated fluidics
monitoring
system for sample
Automated tools to and calibrant delivery
enable quick system Updated electronics
start-up to enable diagnostic
functionality
Automated LC/MS
system performance
checking
— Communications
— Automatically checks
o Retention Time o Peak Width
o Peak Area o Signal / Noise
o Peak Height o Exact Mass
QuanTof combines…
— High voltage pusher
— Dual stage Reflectron
— Novel ion detection system
— …in an optimized, folded
ToF geometry
Performance
— Resolution – over 35,000
FWHM (at m/z 785)
— Mass Measurement – 1 ppm
RMS
— Dynamic Range – up to 105
— Class-leading Sensitivity
— Speed - 30 Spectra/sec
QuanTof™
High Voltage Pusher:
To reduce turn around time.
Dual Stage Reflectron:
Allows focusing of large energy spreads
imparted by the pusher.
Parallel Wire Grids:
Parallel wire grids for maximum ion
transmission and minimum aberration.
High tolerance analyser:
10 micron parallelism over entire
instrument.
Ultra stable HV power supplies:
<1 ppm ripple.
MagneTOF™ e- multiplier:
DM366i detector model has longer lifetime
than MCP detectors used in instruments
prior to the Xevo G2.
Fast 6GHz ADC.
The Result
Resolution >35,000 @ m/z 785.8 in
Resolution Mode.
Sensitivity of 750 cps (TDC) on a sample
of 100 fmol/μl GFP in Resolution Mode.
Summary of characteristics
for a MagneTOF™ electron
multiplier detector:
EMC Shielding
NOTES
Gigabyte Ethernet
ETP MagneTOF
e- multiplier
Ortec Output
6 GHz ADC
Histogram 1 Histogram 2
3.2 Gbits Gigabyte 3.2 Gbits
Ethernet
EPC
VxWorks
Histograms combined
Signal
Pre-Amp
Noise
Detector
Figure 1
Signal
Push 1 Signal Height
Height
Signal
Noise
Bins
Bins
Push 2 Signal
Signal
Height
Height
Signal
Noise
Bins Bins
Signal
Noise
Bins Bins
Figure 2
Signal
Height
Bins
TDC Detector
The signal response in Figure 1, is seen as 1 ion event
and the point where the signal increases above the signal
threshold is saved with a time stamp (edge detection)
It is combined with data from other pushes (Figures 2 and
3) to give a Gaussian shaped peak within the scan time
set by the software (as covered in the ADC Module of the
pre-course e-learning material).
Deadtime must be avoided and this is achieved by
ensuring signal intensities of < 0.1 IPP.
ADC Detector
The area under the signal observed in Figure 1 is
proportional to the number of ions hitting the detector.
This system does not rely on edge detection instead the
signal response is centred (the height of the centred data
is proportional to the area underneath the signal).
This information is then combined with data from all of
the other pushes to give a Gaussian shaped peak within a
given scan time set by the software.
The peak area/height of this Gaussian shaped peak is
proportional to how many ions are present within the
peak (proportional to the sum of all the current detected
under that peak).
Detector Saturation occurs if the maximum level of
quantitation is exceeded on the ADC detector by the
signal intensity generated (we use an 8 bit ADC so
quantitation level available is 0 – 255).
VD VD
N = Signal Response N = Signal Response
VD = Detector Voltage VD = Detector Voltage
0.1 IPP
1 2 3 3.5 4
Detector Voltage
VD = Detector Voltage
However:
The additional parameters that may be tuned
are different. These may be needed during
PM’s to optimise the instrument sensitivity.
The tuneable ranges of parameters are
different.
See the following Xevo G2 parameter tables
for full details.
A key difference is that the Xevo G2
resolution specifications are performed on a
sample of Bovine Insulin (an example is
shown later).
The (tuneable) parameters are the most important for tuning – however the
parameters showing ranges may also be adjusted if necessary for sensitivity.
Most instruments tune with a negative pusher offset value in positive
ion mode. It can be Positive or Negative in Negative Ion mode
Instrument Tuning
This instrument has two “Acq Modes”
available while tuning the instrument:
ADC Mode.
TDC Mode.
When checking resolution or peak shape
you should always use ADC mode.
When comparing/checking sensitivity
always use TDC mode.
You will be required to switch between
the two modes during the tuning
exercises.
Be aware of which mode you are using
at all times.
Once tuning is complete the instrument
must be in ADC Mode for customer use.
Thermocouple 1 2 <5Ω
Desolvation
7 > 10 kΩ
outer nozzle
2 7 <1.0 Ω
2 9 <1.0 Ω
3 6 <1.0 Ω
4 8 <1.0 Ω
5 5 <1.0 Ω
6 4 <1.0 Ω
6 12 <1.0 Ω
8 2 <1.0 Ω
9 3 <1.0 Ω
7 2 5.9 Ω ±10%
Insert
Allen
Key Here
Probe
Nebuliser Gas location
O-ring and dowel
feedthrough
Probe
recognition
LockSpray Probe
LockSpray Probe
in position
It consists of:
XYZ translation
stage.
Reference probe.
Motor assembly.
Camera.
NanoFlow Mass
flow controller.
Sample Infusion
LockSpray
Infusion
What are the three available options for control of the fluidics system?
1.
2.
3.
What are the three steps for clearing persistent errors from any component of the Fluidics?
1.
2.
3.
Once
Steps Basic Instructions
Complete the following table and fill in the complete
results
With the instrument in operate infuse 1 pmol/uL Glu Fib at 5
1
uL/min – do not start the acquisition yet.
On the tune page, enter resolution Mode, then select:
Setup > Quad Profile
2
Select Manual Fixed and set the Mass to 250, click Update
then Close.
Start a ToF MS acquisition from the tune page with the
following parameters:
Acquisition Type: TOF MS
Data Format: Continuum
3 Dynamic Range: Normal
Run Duration: 60 mins
Scan Time: 1
Low Mass: 150
High Mass: 1500
Once complete open the chromatograph, it should look
similar to this :
Results
The intensity of 785.788 has dropped by < 50% (i.e. the intensity of the mass
chromatograph at 40 mins is ≥ 50% of the intensity at 3 mins).
No more then 1 dropout in 40 mins (Any dropouts are > 20% of the
chromatograph intensity level immediately before the dropout event).
If the flow sensor is replaced what has to be done with these values?
If the reference fluidics PCB is replaced what has to be done with these values?
Functional description
The gas handling system distributes and
controls the pressures of the following
gases within the system:
Nitrogen (>95 % purity) externally regulated
to ~100 psi (7 bar). It supplies:
Desolvation gas (purge gas in Nano-
LockSpray) flow 0 to 1200 L/Hr, user-defined.
Cone gas flow 0 to 300 L/Hr, user-defined.
Fixed nebulizer gas flow ~75 L/Hr.
Argon (>99.997 %) externally regulated to
0.5 bar (7 psi) for the collision cell.
The source PCB provides inputs for the
solenoid valves, gas control, and source
pressure sensor.
Base Plate
Components
Source Pressure
Sensor Component
Xevo G2 gas
handling system
* Fluidics in the
Aborted state
Nano-LockSpray
source fitted
Look at the information generated through IntelliStart in the MS Console – does the error
message generated agree with the information in the technical reference information?
2 4
3
1
8 6
7
Components
1
2
3
4
5
6
7
8
EPC
(Double slot on newer instruments due to
size of EPC hardware)
ADC Lite PCB
Enhanced Lens Control PCB
Quad PCB
Source PCB
Enhanced Source Transfer PCB
Collision T-Wave PCB
Or
Multipole Trap PCB
Arrangement of PCBs in the
main electronics unit
nSource_5V Used to indicate source fitted; from source PCB via backplane PCB
Readback and
Cable and Pin Measured Readback
Lens measurement
Measured at: Voltage Value
same ()
Transport 1
Transport 2
(Steering 1)
Steering
(Steering 2)
Tube Lens
Acceleration 1
Acceleration 2
+1.2 V ~ 1.2
+5 V ~5
+15 V ~ + 15
-15 V ~ - 15
+24 V 24 ± 1.25
+430 V ~ + 430
-430 V ~ - 430
Vent Cycles -
Board Temp °C ~ 25
Interlock Logic
Interlock logic is used to ensure that high
voltages are switched on only when the
vacuum is OK.
nStandby
Standby switch not pressed
DOperate
Software set to Operate
OP_RQST
nInterlock_5V
Standby switch not pressed
Switches on
Back up for nStandby
24Voperate_INT
24Voperate_EXT
1 x TURBO TRIP And sets signal
(>80%) 24Voperate_API_ON
TOF_VAC_OK_5V
TURBO_ON_5V
Complete
TASK
Review the 24V operate Logic as detailed in the Lens Control PCB
Technical Reference (p18).
F9 Divert Valve 2A
F7 Select Valve 2A
F F
F F F
What should the RF settings be set to for the StepWave? Where are these set?
Readback
Questions
The values you have recorded should all be different from each other with a
general pattern of voltages from the front to the back of T-Wave 1. What is the
purpose of this difference?
Where does in-source fragmentation occur on a Xevo G2-S, and which Tune Page
slider has most effect on it?
+HT
+ 320 ± 30 V
Supply
HT Supplies
-HT
-320 ± 30 V
Supply
RF Amplitude Within 1%
Amplifier Current < 0.4 A
+ 1.5 V
Within 2% of expected
Low Voltage +3.3 V
value
Supplies +5V
+ 15 V Within 3.5%
RF Frequency 2000 kHz ±7%
24 V Operate Within 2%
Out 0 (Offset C) -120 ±2 V
Out 2 (Gradient) -118 ±2 V
Out 4 (Offset B) -118 ±2 V
Out 6 (Static Offset) -120 ±2 V
What would an RF readback indicating a frequency outside the tolerance limits indicate?
Communication
The EPC communicates with the ADC via
Ethernet and it communicates with the rest
of the instrument via EPCAS:
It is possible to have control of the
instrument and not be able to acquire data.
If red LED(s) appear on the ADC
diagnostics tab in the software but a reboot
or an electronic reset does not clear them
then we need to decide if the EPC or the
ADC is at fault.
There is a procedure in the ADC PCB
technical reference for this (715003767):
Once
Read test and comment
complete
ADC Setup
The ADC Setup was performed earlier in
this course:
Review the technical reference and ensure
the setup procedure is understood.
Complete
Using the EPC Technical Reference obtain an electronic
copy of the EPC boot sequence and save it to your
training folder.
No. Answer
1
2
3
4
5
6
7
What PCB outputs this signal onto the Backplane as nSource and what is it used
for?
Questions Observe
Where are the calibration values stored for the flow sensor?
Is this different from the Fluidics/Options PCB? – Explain why this could be
significant.
RF Generator LEDs
These are located on the side of the RF
Generator that faces the instrument’s
flight tube.
D 10
D11
What are two main reasons why the instrument would trip and go into Standby
in connection with the quadrupole electronics?
Which other PCB provides the Input Voltages to the Status PCB?
D3 Green Data In
Questions
How many interlock groups are generated by relays on the User Event PCB?
If the gas supply pressure falls below 4 bar – the nAPI solenoid signal is sent to?
When using Capillary Electrophoresis with the probe is attached and the instrument in
operate, which PCB triggers the isolation circuit on the user even PCB?
Probe HV control:
There are four digital signals for
this unit.
Complete the table below using
the Source PSU Technical
Reference (715003695).
Polarity nAPI_Source
Inhibit Capillary_or_nCorona
Readback =
Always within Detector
Determined by 20 V of set NOTE: actual
Detector
detector setup test point – always voltage applied is
positive the sum of ToF +
Detector Voltage
Nano-LockSpray Source
Overview.
Safety Interlocks.
Assembling the Sample Sprayer Stage.
Camera installation.
Setting the Sample Capillary Position
(tuning for sensitivity).
Tune Page.
On Site:
If you have both a nano-LockSpray source
and an ESI Source to install, you would run:
All the IntelliStart setup procedures using the
ESI LS source.
All ESI LS source specs
Only the Pos Sens Mode specs with the nano-
LS source to confirm source functionality.
Once
Step Basic Instructions
complete
Switch the baffle position to “LockSpray” and check that flow can be
5
observed through the reference sprayer.
Once
Step Basic Instructions
complete
When flow is observed from both sprayers, set the position of the sample
1
emitter using the XYZ stage to optimise the signal intensity.
Begin by setting the initial Z axis of the sample emitter position. Retract the
emitter well away from the baffle. Set the baffle to the LockSpray position. Move
the Z axis forward until the tip of the emitter nears the baffle.
2
Note: Ensure that the emitter tip does not touch the baffle!
Switch the baffle position a couple of times to check that it will not hit the
emitter. Leave the baffle in the LockSpray position.
Set the initial Y axis of the emitter position. The emitter tip should be within the
right angle of the baffle (so that the baffle actually stops the sample spray from
3 reaching the sample cone when in the LockSpray position). You may need to look
through the source window to confirm the alignment of the capillary height with
the baffle.
Set the initial X axis of the capillary position.
Switch the baffle to the Sample position. You should be able to see a beam on
4
the Tune Page. Observe the sensitivity of the GFP beam. Adjust the X axis
position to maximise sensitivity.
Now that the initial positions of each axis is set, adjust each axis in a cyclic
manner to further increase sensitivity.
Note: Be careful with the Z and Y axes. Do not adjust them too far at one time
and be aware of moving them closer to the baffle so that the baffle does not hit
5
the capillary when in the LockSpray position.
Note: Moving the capillary closer to the sample cone does not always improve
sensitivity. When optimising capillary position, always check whether moving
further from the sample cone increases sensitivity.
Once the sample capillary position is optimised, the nano-LockSpray source
should be able to achieve sensitivities similar to the standard ESI source (even
though it has 1/10th of the flow rate)., however the specification values are 50%
of those for the regular ESI LockSpray source.
As a guideline when using the nano-source on the Xevo G2-XS, aim for the
6 following sensitivities (or better) on the peak at m/z 785.8 (in MS Mode) using
the default source parameters shown on the next page and infusing 100 fmol/μL
GFP at a flow rate of 0.5 μL/min (double these values if you are using the ESI
source running at 5 μL/min):
Resolution, MS Mode: ≥6.0e4
Sensitivity, MS Mode: ≥6.0e5