Professional Documents
Culture Documents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev.March2004
contents
General Information Important safety information
The design of Amersham Biosciences instruments is subject to continuous
development and improvement. Consequently, this instrument may incorporate
minor changes in detail from the information contained in this Service Manual.
Suggestions about improving or correcting this manual should be sent to:
Label located at the back of the instrument.
Amersham Biosciences
Instrument Support Department
S-751 82 UPPSALA
SWEDEN
This document is our property and shall not without our written permission be Meaning: The instrument meets the requirement of the EMC Direc-
altered, copied, used for manufacturing or communicated to any other person or tive 89/336/EEC through the harmonized standards EN
company. 50081-1 (emission) and EN 50082-1 (immunity) and
requirements of the Low Voltage Directive (LVD) 73/23/EEC
This document is written for trained Amersham Biosiensces Service personnel. through the harmonized standard EN 61010-1.
Warning: This instrument is designed according to EC regulations. Always use
original spare parts and always assemble the instrument to it's original condition,
or else the regulatory demands may not be fulfilled after servicing.
Trade marks
UNICORNTM Controller, ÄKTATMexplorer are exclusive trade marks of Amersham
Biosciences. In view of the risk of trade mark degeneration, it is respectfully
suggested that authors wishing to use these designations refer to the trade mark
status at least once in each article.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev.March2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev.March2004
contents
Chapters
1. Theory of operation
2. Installation
3. Interfacing
4. Troubleshooting
6. Dissassembling
7. Spare parts
9. Appendix
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev.March2004
contents
Chapter 1
Chapter Contents
1. Theory of operation
Theory of operation 1.1 Maldi Time of flight MS
1.2 The basic principles of Maldi TOF
1.3 Description of Ettan Maldi TOF system
1.3.1 Features
1.3.2 Specifications
1.3.3 The principle of operation
1.3.4 Vacuum system
1.3.5 Mechanics
1.3.6 Glossary Abreviations
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1. Theory of operation Schematics of a Maldi
Drift tube
A time of flight mass spectrometer works on the principle Ionization Reflectron
chamber
of ion physics.
The instruments consists of the following:
Target
- a vacuum system with sample
- a pulsed laser
Detector
- a sample compartment
- a time of flight tube
- a beam dump
- an ion gate
- a reflectron - ion mirror
- a linear detector
- a micro chanel plate detector for reflectron mode
- a software to control and evaluate
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1.2 The basic principles of Maldi TOF
Linear detector
Sample slide
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1.3 Description of Ettan Maldi Tof system
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1.3.1 Features
- Vacuum System as described in the Vacuum Section.
Laser Bench with UV nitrogen laser (337nm. wavelength).
Laser Optics control with beam detection.
Air Outlet - Fans
Pulsed extraction ion optics (time delay).
Low Mass Rejection controller for matrix ion suppression.
Flight Tube vertically mounted.
Timed Ion Gate control and beam deflector.
- Linear and Reflectron Mode of operation
Reflectron (Harmonic field).
Linear Detector (ETP).
Reflectron Detector (MCP).
- Linear mode for high molecular weight peptides and pro
teins 4000 - 130000 M/Z
- Reflectron mode for increased resolution and peptide fin-
gerprinting with PSD capabilities 500 - 4000 M/Z Indicator LED´s
- 1 GHz Transient recorder board
- Optional automatic slide loader, 12 slides per tray
- 24 position sample slide.
- Integrated instrument control and data base search rou-
tines.
- Integrated data analysis software from Proteometrics
- Bench top fingerprint.
Sample Inlet
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1.3.2 Specifications
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1.3.2.2 System specifications
Resolution: 10000
Sample
Mass range: 500 Da - 500.000 Da entry
door
Sample
Quadratic field reflectron focus: 18 kV door
switch
(75% of mass range)
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1.3.3 The principle of operation
The Ettan Maldi ToF uses a pulsed UV Nitrogen laser at Ettan Maldi Tof MS design
337nm, which is steered by two mirrors, one of which is
adjustable, and focused by a convex lens onto the sample, to
desorb the target analyte from a solid to gaseous state.
The ions in this state form a cloud or plume in a temporary
field free region of the source between the sample slide and
the extraction lens.
A time delay that is mass specific is applied to the extraction
lens before it is fired. This delay is measured with respect to a
pulse generated by the laser beam detector.
Once the pulsed extraction lens is fired the ions pass into the
flight tube.
All the ions have the same kinetic energy and hence lower
mass ions reach the detector faster than higher mass ions.
Consequently a time to mass conversion can be made which
enables a mass analysis of the whole spectrum to be per-
formed.
The ion’s travel up the flight tube into the low mass reject cell.
This can have a timed deflection voltage applied to reject the
lower mass ions which are generally of no interest and are
formed predominantly from the matrix.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
There are two detectors. An electron multiplier (ETP) .
which is used to detect ions in linear mode, figure 1: Linear Mode
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
If the time taken by these molecules is known it is then possi-
ble to apply a timed voltage to the gate which only allows the figure 4: PSD
precursor and fragments of the molecule of interest to reach
the detectors.
After the gate region the ions pass into the Harmonic field Post-Source Decay
reflectron.
P a re nt ve lo city is ke pt by fragm e nts after d eca y
In linear mode no voltage is applied to the reflectron lens so
they then travel to the linear detector. M
G ate
In reflectron mode a voltage is applied to the reflectron lens
which creates an electrical field which increases quadratically
re flec tro n
as the ions pass further into the lens. Ion sourc e
At a given point in the lens, the ions are reversed and re-
pelled towards the reflectron detector. Lig hte r frag m ent has
sm alle r en ergy m v 2/2
The reflectron mirror effectively increases the flight path by a an d tu rns ba c k at an
ea rlier tim e
factor of three thus giving greater mass separations T im e se paration of fra gm en t ions in reflectron
at the detector and hence higher resolution.
t U
Z ap th e m olec u le s
in duc e s po st
sour ce dec ay
D etecto r
reflec tro n
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1.3.4 Vacuum system
The function of a vacuum system is easy to understand, but
troubleshooting a vacuum system can be time consuming. Do
not vent the system and start to replace parts without thinking
twice (pump down time is long, days can be lost).
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
3) The gas is transported through the pump to the outlet by the molecules collide with the wire and cooles it. As gas molecules
rotation of the vanes. The rotation of the vane creates a are pumped away there are less molecules and therefore less
smooth flow of the gas through the pumping chamber. cooling and the wire heats up. As it heats up its electrical re-
sistance increases. A simple circuit utilising the wire detects
The turbo pump is a blade turbine that compresses gases. By the change in resistance and once calibrated it can directly
the momentum transfer from the rapidly rotating rotor blades correlate the relationship between pressure and resistance.
to the gas molecules the non-directed thermal motion of the The Pirani gauge works down to 10e-4 mbar. Other forms of
molecules is changed to a directed motion. Hence, the pump- gauge have to be used to measure pressures lower than this.
ing process in a turbomolecular pump results from the directed
motion of the gas molecules from the inlet flange to the fore- The Magnetron is a so called Cold-Cathode Gauge. All CCGs
vacuum port. To create the directed motion of the gas mol- utilize crossed electric and magnetic fields to trap electrons.
ecules the tips of the rotor blades have to move with super- The electron plasma, responsible for ionization, originates
sonic velocity. Hence, a high rotational speed of the rotor is from the random release of an electron at the cathode. A dis-
required. charge slowly builds inside the ionization volume to the point
where the entry of new electrons into the plasma is limited by
In the molecular flow range - i.e. at pressures below 10-3 mbar space charge repulsion. At pressures below 10-4 mbar, the
(0.75 x 10-3 Torr) - the mean free path of the gas molecules is discharge is practically a pure electron plasma. The electrons
larger than the spacing between rotor and stator blades (some move in cycloidal jumps, circling about the anode, and during
tenths of a millimeter). Consequently, the molecules collide part of each jump they have sufficient energy to ionize gas
primarily with the rotor blades with the result that the pumping molecules through electron impact ionization. The probability
process is highly efficient. of collision is proportional to the gas density. The slow ions
In the laminar flow range - i.e. at pressures higher than 10 -1
generated, are quickly captured by the cathode. The current
mbar (0.75 x 10 Torr) - the situation is completely different.
-1
generated by this ion collection process is measured and used
The action of the rotor is diminished by the frequent collisions as an indirect indication of gas pressure. This ion-induced
between the molecules. Therefore, a turbo pump is not capa- current is not linearly related to the pressure in the chamber. A
ble of pumping gases against atmospheric pressure and has typical range for a magnetron is 10e-2 - 10e-9 mbar.
to be backed by an adequate fore-vacuum pump. See next page for a schematic off the vacuum system and a
The Pirani gauge head is based around a heated wire placed description of its function.
in a vacuum system, the electrical resistance of the wire being
proportional to its temperature. At atmospheric pressure, gas
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1.3.4.3 Basic function of the vacuum system
A rotary pump is connected to the source turbo pump and the
reflectron turbo pump by stainless steel vacuum tubing to back
them up. A Backing valve is fitted between the rotary pump
and the two turbo pumps, this valve is used to isolate the high
vacuum side of the vacuum system from atmosphere pressure
Turbo pump during the sample exchange sequence (see below for a de-
tailed description). A Pirani gauge is fitted on top of the rotary
pump to measure the vacuum on the low vacuum side. A for
line trap is fitted between the rotary pump and the backing
valve to protect the vacuum system from back streaming oil
fumes.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
the gauge from poor vacuum. The trip point for the magnetron 1.3.4.5 Sample Analysis
is set to 1e-6 mbar. The trip point is set on the I/O pcb. After a sample spot is selected for analysis (see the software
The readings from the gauges can be checked in the Ettan manual for setting up files, etc.). The sample motor will drive
Maldi Control software. the sample slide to the appropriate position so that the laser
Typically value for these gauges will be between 1e-2 and 1e–3 hits the required sample.
mbar for the Pirani gauge and from 1e–6 to 1.0e–7 mbar for the
Magnetron gauge. 1.3.5 Mechanics
The “OPEN DOOR” command drives the sample probe to the The Ettan Maldi Tof has only three moving parts (pumps and
sample exchange position and seals and isolates the high fans excluded)
vacuum side of the instrument from the door interlock.
- The sample probe mechanism
The vent valve opens for a short time venting the sample ex-
change lock and allowing the sample door to open. The magn- - The sample door.
etron reading normally improves slightly during this sequence. - The filter wheel assembly.
If it gets worse then the seal between the sample holder and All three mechanism are motordriven and their limits of travel
sample lock is leaking. Normally this requires a small adjust- are controlled by micro switches. The motors are controlled by
ment to the micro-switch, which controls the maximum motion the I/O pcb, it`s the second pcb from left in the electronic rack.
of the sample holder into the sample exchange lock. The sample probe- and filter wheel motor are stepper motors
The “CLOSE DOOR” command closes the door. The isolation and the door is operated by a dc motor.
or backing valve to the turbo pumps closes. The inlet valve
opens allowing the sample door interlock to be evacuated to
roughing pressure by the rotary pump. 1.3.5.1 Sample probe mechanism
Once the backing line to the sample door reaches the Pirani The sample probe mechanism moves the sample probe with
trip point the backing valve opens, the inlet valve closes and the sample slides in and out of the Maldi. The movement of
the sample probe moves to its reference position. A dip in the the probe is done by a long drive screw made of peek. The
Magnetron reading will be seen whilst the rough pressure from drive screw is turned by a stepper motor and a gearbox. This
the sample lock is pumped away by the turbo pumps. unit is called Sample probe mechanism. The inner and outer
end positions for the sample probe is detected by two micro-
switches. The outer position is the sample exchange position
and the inner position is the start position. The start position is
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
where the sample probe stepper motor starts to count the 1.3.5.3 Laser and optics
number of steps it is moving.
The laser used in the Maldi is a nitrogen laser class 3A whit a
The rotation of the drive screw feedthrough is sealed with two wavelength of 337 nm. There are no parts to be replaced in-
o-rings and lubricated with vacuumgrease. The volume side the laser box, all parts is embedded in to a solid piece of
between the o-rings is pumped out by the rotary pump via the rubber. The laser use 24V from the electronic rack as supply
pump out line (see drawing of the vacuum system). This will voltage. The trigger signal for the laser comes in a coaxial
compensate for leaks in the o-rings when the sample probe cable from the T/R pcb. The pulse frequence is 1 or 8 Hz and
feedthrough is turned by the stepper motor and maintain the is set up in the method. At the exit of the laser is a manual
high vacuum in the system. operated shutter which should be open in normal operation.
The upper micro switch on the sample probe mechanism After the laser comes the filter wheel which is used to set re-
determine the start point. It is used to find the first sample quested laser power.
on the slide by counting the number of steps from the micro
switch to the first sample spot on the sample slide. The lower
switch is for the stop position of the sample probe in the 1.3.5.4 The filter wheel assembly
sample exchange position. In this position the sample probe The min. and max position for the filter wheel is controlled by
acts like a isolation valve between the ion source and the two micro switches. The switches is activated by an arm fitted
sample lock. It is important to find the right position to on the filter wheel. The distance between min. and max is
maintain a good vacuum in the Maldi during the sample approximately 2700 and is set to 850 steps by the stepper
exchange procedure. The movement of the sample probe is motor parameter in the Maldi .def file.
controlled by the I/O pcb.
The filter wheel is fitted in a 450 angel to reflect some light
from the laser to the laser detect sensor. The filter is turned by
1.3.5.2 The sample door assy a 12 V stepper motor via a drive belt and its positioning is
controlled by the I/O pcb. After the filter is a transparent mirror
The sample door is moved up and down by a dc motor. The the directs the laser puls to the entrance lens of the ion
open and closed positions for the door is detected by micro source. Inside in the ion source is a wobbulator that reflects
switches. The lower micro switch for the closed position is the laser beam to the sample.
connected in series with the HV interlocks on the I/O pcb, this
means that the sample door has to be closed before the HV is
turned on.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
1.3.6 Abreviations calculated using the average atomic weights for each
element. An average mass is obtained in MALDI-TOF-MS
when a peak is not isotopically resolved.
10% valley
For defining resolution. Two peaks of equal height in a mass Background subtraction
spectrum at masses m and m-m are separated by a valley The process in which the background (chemical noise and
that at its lowestpoint is just 10% of the height of either peak. detector noise) is subtracted, thus leaving the peaks above
the noise at the base level.
Ablation step
Molecular components of a solid are put into a gas phase and Base peak
ionized. The most intense peak in a mass spectrum. A mass spectrum
is usually normalized so that this peak has an intensity of
Adaptive backgroundcorrection (ABC) 100%. See also parent peak.
ABC filters are usually also applied to low resolution data to
enhance the high-frequency signal and remove the low Calibrant
frequency background: I’j = Ij-d/(2n+1).Ij+k k = -n,...,n, where A compound used for the calibration of an instrument.
0 < d < 1, n is the width of the filter, I j+k is the intensity in
channel j+k before the filter has been applied to the data, and Calibration
I’ j is the intensity in channel j after the filter has been applied A process where well-known masses are assigned to certain
to the data. peaks. The purpose is to improve the mass accuracy of an
By applying an ABC filter in combination with a smoothing MS instrument. For ESI-TOF, the calibrant is introduced into
filter the quality of peaks can be improved. the chamber via a second electrospray needle. For MALDI-
TOF-MS, the calibrant is either mixed with the sample
Adduct peak (internal calibration) or run as a separate sample (external
Results from the photochemical breakdown of the matrix into calibration).
more reactive species, which can add to the polypeptide. Can
also result from salt ions, Na + etc., that are embedded in the Centroided mass peak
matrix. The centroided mass peak is located at the weighted centre of
mass of the profile peak.
Average molecular weight
The mass of a molecule of a given empirical formula
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Collision induced dissociation (CID) created from a flowing liquid sample, the extraction of gas
A process whereby a mass-selected ion is excited and caused phase ions from these droplets under a high electric field, and
to fragment by collision with a target gas. Used in some ESI transportation of the ions into a vacuum suitable for MS.
and MALDI instruments. The fragment ions need to be
charged to be mass separated, so many of the MALDI Electrospray ionization (ESI)
fragments may not be detected since they are neutral. An ionization method that allows production of molecular ions
directly from samples in solution. A sample solution is sprayed
C-terminal through a needle, which is kept at a potential (typically 3.5 to 6
The carboxyl residue at the end of a polypeptide chain. Proper kV). The voltage on the needle causes the formed droplets to
nomenclature for indicating product ions containing a C- be charged. As the solvent evaporates, the droplets disappear
terminal is v, w, x, y, and z. leaving highly charged analyte molecules. The resultant ions
often carry multiple charges, which reduces their mass-to-
Daughter ion charge ratio compared to a singly charged species. It is a
See fragment ion continuous method suitable for interfacing with HPLC or
capillary electrophoresis. It is also a very sensitive
Desorption ionization (DI) method well suited for the analysis of small amounts of large
An ionization method used to ionize non-volatile solid samples biological molecules such as peptides and proteins that are
by impact of energetic particles or photon beams. difficult to vaporize or ionize, especially thermally labile
compounds. Generates ions carrying many charges.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Full width at half height (FWHM) optics (open), or that can deflect the ions out of the way of the
A measure of the peak width at 50% of its maximum intensity. later stages of the mass spectrometer.
Tuning
The process of optimizing the MALDI instrument’s laser power
and target position to obtain the best possible sensitivity and
signal-to-noise ratio for a specific type of experiment.
Unit resolution
Distinguishes between ions separated by 1 m/z unit.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Chapter 2
Chapter Contents
2. Installation
Installation
2.1 Hardware Installation
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.1 Hardware installation 2.1.5 Lifting the instrument
Only lift the instrument if you are four strong people, the weight
2.1.1 Lab preparation is 150 kg. Remove the transportation screw for the rotary
Make shore that the lab is prepared and ready for the installa- pump at this stage, it is fitted from the under side of the Maldi.
tion by checking that the lab manager has filled in and re- This will reduce the vibration in the instrument and make it
turned the site prep check list (see page 14 in Site prep guide easier to replace the rotary pump in the future. Check the oil
18-1144-63.) level in the rotary pump fill up if necessary.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.1.3 Prefixes 2.2.1.4 Examples
The units described above are often used together with pre-
fixes. The prefixes are described in the table below: 2.2.1.4.1 Preparation of a mixture of liquids with a defined
volume percentage
Prefix Symbol Factor
exa E 1 0 18 1. Prepare 25 ml of 10 vol% TFA in water.
Use the formula for vol% and put in the information we
peta P 1 0 15 know (vol% and total volume).
tera T 1 0 12 10/100 = (volume of TFA)/25 à Volume of TFA = 2.5 ml, vol-
ume of water = 25 – 2.5 = 22.5 ml.
gi ga G 109
mega M 106 2. Prepare 50 ml of 50 vol% Acetonitrile and 0.9 vol% TFA
in water (use a 10 vol% TFA solution).
kilo K 103 Use the formula for vol% and together with known information
hecto H 102 and calculate the volume for each substance one at a time.
Acetonitrile: 50/100 = (volume of Acetonitrile)/50 à Volume of
d e ka Da 10 Acetonitrile = 25 ml.
d e ci D 10-1 TFA: Volume of pure TFA à 0.9/100 = (Volume of TFA)/50 à
Volume of TFA = 0.45 ml.
centi C 10-2 Volume of 10 vol% TFA à 10/100 = 0.45/(total volume of solu-
milli m 10-3 tion) à Total volume = 4.5 ml.
micro µ 10-6
2.2.1.4.2 How to prepare a solution of defined concentra-
nano N 10-9
tion from a vial with dried peptide
p i co P 10-12 Prepare a solution with concentration 1000 pmol/ml (equiva-
femto F 10-15 lent to 1*10-3 mol/l) from 1 mg of a dried peptide with molecu-
lar mass 897.531.
atto A 10-18 1. First calculate the molar content of dried peptide.
zepto Z 10-21 Molar content = mass/molecular weight = 1*10-3/897.531 mol
yocto Y 10-24 = 1.114*10-6 mol = 1.114 mmol.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2. Then calculate the volume needed to get the desired con- Neurotensin and Hexatyrosine shall be prepared once every
centration. year (and in case of problems).
Molar content = concentration*volume à Volume = 1.114*10-6/ · Working solutions (20 pmol/ml) of each peptide shall be pre-
1*10-3 l = 1.114*10-3 l = 1114 ml pared every 6 months. 6 vials shall be prepared for each pep-
tide. 5 vials of each peptide shall be stored in a common box
2.2.2 Sample handling routines in the freezer. This box shall never be brought out of the
freezer (except for exchanging samples). 1 vial of each pep-
2.2.2.1 Buffer tide shall be stored in a second box in the freezer. This box
· New bottles of Acetonitrile (ACN) and TFA shall be bought shall be used for preparing samples for spotting (i.e. can be
once every year (and in case of problems). brought in and out of the freezer).
· A new batch of Buffer A and Buffer B shall be prepared once · The peptides stored in the box used for spotting shall be
every month. The buffer shall be stored in a common box in exchanged once every month.
the freezer.
· Each time buffer is needed, an appropriate number of 2.2.2.4 Protein digests
Eppendorf tubes are drawn from the box. However, the box · A new vial with dried digest should be dissolved to 1 pmol/ml
itself shall stay in the freezer at all times until a new batch is each month. The vial shall then be transferred to the box used
prepared. for spotting and be stored in the freezer.
· Vials containing dried digest shall be stored in the freezer.
2.2.2.2 Matrix This box shall never be brought out of the freezer (except for
· New bottles of matrix (Sinapinic acid and alpha-cyano) shall exchanging samples).
be bought and used with a two-month interval.
· Alpha-cyano matrix shall be stored in an exicator in the 2.2.2.5 BSA.
freezer. · A new vial of BSA powder should be bought once every
· Sinapinic acid matrix shall be wrapped in aluminium foil and year. The powder shall be stored in the fridge.
stored in an exicator. · A new batch (6 vials with 4 mg/ml BSA containing 1 ml each)
· Prepared matrix solutions shall be stored dark and are only of working solution shall be prepared every 6 months. 5 vials
allowed to be used for 12 hours after preparation. shall be stored together with working solutions of peptides and
digests in a common box in the freezer. This box shall never
2.2.2.3 Standard peptides be brought out of the freezer (except for exchanging samples).
· New stock solutions of Ile7-Ang III, hACTH18-39, 1 vial shall be stored in a second box in the freezer. This box
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
shall be used for preparing samples for spotting (i.e. can be Acetonitrile, TFA, digests, BSA, pipette tips, eppendorf tubes,
brought in and out of the freezer). slides …and more). 2. Description of the preparation of all
· The BSA vial stored in the box for spotting shall be sample solutions (with references to 1.). 3. Notes describing
exchanged once every month. the exchange of samples for spotting (with references to 2).
4. Notes describing the spotting of samples (with references to
2.2.2.6 Dilution series 1-3). 5. References between prepared slides and MALDI
· Dilution series shall be prepared fresh from the working results/delivery test result protocol.
solutions every time samples are to be spotted.
· A diluted sample must be spotted within an hour after dilution 2.2.3 Sample Handling Protocols
(as quick as possible). If a slide with all samples for the acceptance tests is not
available start the acceptance test by preparing buffers, matrix
2.2.2.7 Slides and samples as described below.
· Unused slides shall be used each time samples are spotted.
Clean the slides, according to instructions in kit, before use. 2.2.3.1 Preparation of buffers and matrix solutions
· Opened sample slide kits shall be stored in a dust free
environment, wrapped in aluminum foil. 2.2.3.1.1 Preparation of 25 ml 10 vol% TFA
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.3.1.2 Preparations of 50 ml BUFFER A (50 vol% ACN, Water (HPLC grade) 22.5 ml
0.5 vol% TFA)
The solution should be split into smaller aliquots (1 ml 10 vol% TrifuoroAcetic Acid (TFA) 2.5 ml
Eppendorf tubes), wrapped in “parafilm”, stored at -20°C, and
100 vol% Acetonitrile (ACN) 25 ml
used one at a time. Use glass pipettes when handling TFA ith
concentrations in the range 5-100 vol% (plastic pipettes re- Total volume 50 ml
lease softeners that can affect the MALDI spectrum).
Use HPLC grade of all chemicals.
Water (HPLC grade) 22.5 ml
2.2.3.1.4 Preparation of ACHCA matrix
10 vol% TrifuoroAcetic Acid (TFA) 2.5 ml
The alpha-cyano-4-hydroxy cinnamic acid (ACHCA) matrix
100 vol% Acetonitrile (ACN) 25 ml should be prepared fresh everyday it is needed. When pre-
pared it should be stored in an Eppendorf tube wrapped in
Total volume 50 ml aluminum foil at room temperature (it can be stored like this for
12 hours).
Use HPLC grade of all chemicals.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
5. Spin down in a centrifuge 10 s at 14000 g. 2.2.3.2. Preparation of peptide standard stock solutions
6. Transfer 600 ml of the supernatant (the fluid portion) to a new
Eppendorf tube.
7. The matrix is ready to use! (Store as described above). AmBio Name Water(HP- Final Termed
Order # LC)* Conc.
2.2.3.1.5 Preparation of Sinapic matrix
17-6000-84 hACTH(18-39) 145.7 l 1000 S1
The Sinapic acid matrix should be prepared fresh everyday it is
µ pmol/ l
needed. When prepared it should be stored in an Eppendorf tube µ
wrapped in aluminum foil at room temperature (it can be stored 17-6002-83 Ile7AngIII 470.7 l 2000 S3
like this for 12 hours). µ pmol/ l
µ
Sigma
AmBio Order # Order #
17-6002-82 Sinapic acid Powder 20 mg N6383 Neurotensin 247.1 l 2000 S4
(1mg) µ pmol/ l
BUFFER B 1 ml
µ
T1780 Hexa-L-Tyrosine 392.4 l 2000 S5
(1mg) µ pmol/ l
1. Mix as described above. µ
2. Heat the mixture to 37°C for 1 minute
3. Vortex the tube with BUFFER B and matrix powder for 30 Mix according to the table above and vortex for 30 seconds.
seconds. Transfer the solutions to eppendorf tubes and wrap with
4. Let the tube sit on the bench for a few minutes. “parafilm”. Store at –20 °C until use.
5. Spin down in a centrifuge shortly (if no one is available, just let *The amount of water (HPLC grade) to add to the delivered
the tube sit longer on the bench until all material has settled at the vial (see order #) to produce the described Final concentra-
bottom). tion.
6. Transfer the supernatant (the fluid portion) to a new Eppendorf
NB! The calculations are based on the following values for
tube.
purity and peptide content:
7. The matrix is ready to use! (Store as described above).
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Name Peptide purity Peptide amount of Name From previous Buffer A Final conc. Termed
content peptide in vial dilution
hACTH18-39 0.97 0.74 0.359 mg hACTH(18- 10 l S1 490 µ l 20 pmol/ µ l WS1
ILE7
AngIII 0.97 0.87 0.844 mg 39) (2.11.3.2)
Neurotensin 6 l S4 594 µ l 20 pmol/ µ l WS3
Neurotensin 0.96 0.86 0.826 mg
(2.11.3.2)
Hexa-L-Tyrosine 0.99 0.79 0.782 mg Neurotensin 6 l S4 594 µ l 20 pmol/ µ l WS3
(2.11.3.2)
For each new peptide bought, be sure to check the peptide
purity and peptide content for the corresponding batch (this Hexa-L-Tyr- 6 l S5 594 µ l 20 pmol/ µ l WS4
information should be collected from the supplier). If (peptide osine (2.11.3.2)
purity)*(peptide content) deviates more than 10% from the
values above, the preparation of the stock solution has to be 2.2.3.3.2 Preparation of full length BSA working solution
adjusted accordingly.
AmBio Order # Components Amount Final Conc. Termed
2.2.3.3 Preparation of working solutions 17-6002-86 BSA powder 24 mg
2.2.3.3.1 Preparation of peptide working solutions Water 6 ml 4 mg/ml WS6
Follow the instruction below for each vial of working solution that
should be prepared. Wrap eppendorf tubes with “parafilm”. 1. Weigh the BSA sample according to above and put it into a 15
Store at –20°C until use. ml “falcon tube”.
2. Add water according to above.
3. Vortex until the entire sample is dissolved.
4. Distribute the sample into 1 ml Eppendorf tubes (e.g. 1
ml/tube) and wrap with “parafilm”.
5. Store at –20°C until use.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.3.3.3 Preparation of Phosphorylase B working solution 2.2.3.5 Preparation of test slides
Michrom Components Amount Final conc. Termed 2.2.3.5.1 1.5 pmol Ile7-AngIII and 1.5 pmol hACTH18-
Bioresour- 39Slide A spot1-6 and Slide B spot 1-11
ces order# 1. Add 6 ml WS1 (2.11.3.3), 6 ml WS2 (2.11.3.3) and 12 ml
910-0002-- 1 vial 500 pmol ACHCA matrix solution (prepared according to 2.11.3.1.4) to
29 Phosforylase an Eppendorf tube.
B digest. 2. Mix by pipetting up and down 5 times
3. Add 0.3 ml to each appropriate spot (see 2.11.4).
Buffer A 500 µl 1 pmol/µl WS7 4. A maximum of 12 spots can be spotted before a new prepa-
ration must be made.
1. Mix according to table
2. Vortex for 30 seconds 2.2.3.5.2 5 fmol Neurotensin; Slide A spot 9-10 and Slide
3. Store at –20°C until use. B 16-19
2.2.3.4 Preparation of dilution series 1. Add 10 ml D6 (2.11.3.4) and 10 ml ACHCA matrix solution
(2.11.3.1.4) to an Eppendorf tube.
A diluted sample must be spotted within 1 hour after prepara- 2. Mix by pipetting up and down 5 times.
tion (as quick as possible).
3. Add 0.3 ml to each appropriate spot (see 2.11.4).
Name Amount Buffer A Final conc. Termed 4. A maximum of 6 spots can be spotted before a new prepa-
from ration must be made.
previus
dilution
2.2.3.5.3 100 fmol Phosforylase B digest with internal
Neurotensin 5 l WS3 145 µl 33 fmol/ µl D6 calibrants; Slide A spot 11-15 and Slide C spot 1-4
Hexa-L-Tyr- 5 l WS4 495 µl 200 fmol/ µl D 16 1. Add 6 ml WS2 (2.11.3.3), 4 ml WS1 (11.3.3) and 90 ml
osine ACHCA matrix solution (2.11.3.1.4) to an Eppendorf tube.
2. Mix by pipetting up and down 5 times.
Phosforyla- 10 l WS7 10 µl 500 fmol/µl D 17 3. Add 6 ml of the solution above and 12 ml of D17 (2.11.3.4)
se B to a new Eppendorf tube.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
4. Mix by pipetting up and down 5 times. 2.2.3.5.7 50 fmol fmol Phosforylase B digest with internal
5. Add 0.3 ml to each appropriate spot (see 2.11.4). calibrants; Slide C spot 5-8
6. A maximum of 10 spots can be spotted before a new prepa- 1. Add 10 ml from the100 fmol solution made in step
ration must be made. 2.11.3.5.4 to an Eppendorf tube and 10 ml ACHCA matrix solu-
tion (2.11.3.1.4).
2.2.3.5.4 3 pmol Hexatyrosine; Slide A spot 16-19 2. Mix by pipetting up and down 5 times.
1. Add 6 ml WS4 (2.11.3.3.2) and 6 ml ACHCA matrix solution 3. Add 0.3 ml to each appropriate spot (see 2.11.4).
(2.11.3.1.4) to an Eppendorf tube.
2. Mix by pipetting up and down 5 times. 2.2.3.5.8 5 fmol Phosforylase B digest with internal
3. Add 0.3 ml to each appropriate spot (see 2.11.4). calibrants; Slide C spot 9-12
4. A maximum of 10 spots can be spotted before a new prepa- 1. Add 10 ml from the 50 fmol solution made in step 11.3.5.8
ration must be made. to an Eppendorf tube and 10 ml ACHCA matrix solution
(11.3.1.4).
2.2.3.5.5 30 fmol Hexatyrosine; Slide A spot 20-24 2. Mix by pipetting up and down 5 times.
1. Add 6 ml D16 (2.11.3.4) and 6 ml ACHCA matrix solution 3. Add 0.3 ml to each appropriate spot (see 2.11.4).
(2.11.3.1.4) to an Eppendorf tube.
2. Mix by pipetting up and down 5 times.
3. Add 0.3 ml to each appropriate spot (see 2.11.4).
4. A maximum of 5 spots can be spotted before a new prepa-
ration must be made.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.4 Slide layout 2.2.4.2 Slide B
2.2.4.1 Slide A
Slide Pos Sample Amount on Expected mono-
sp o t isotopic m/z [M+H]
Slide Sample Amount on spot Expected mono-
P o s. isotopic m/z 1-11 Ile7-AngIII and 1.5 pmol Ile7-AngIII
[M+H] hACTH18-39 each =897.53117
1-6 Ile7-AngIII and 1.5 pmol each Ile7-AngIII HACTH18-39
hACTH18-39 =897.53117 = 2465.19897
HACTH18-39 12-15 Ile7-AngIII 5 fmol 897.53117
= 2465.19897
16-19 Neurotensin 5 fmol 1672.89
7-10 Neurotensin 5 fmol 1672.89
20-24 BSA 3.6 fmol Monomer = 66423
Dimer = 132846
11-15 Ile7-AngIII,hACTH 180 fmol Ile7- 800<m/z< 2600
trimeer = 199269
18-39 and Iodo- AngIII
acetamide 120 fmol
modified hACTH18-39
Phosforylase 100 fmol 2.2.4.3 Slide C
B digest. Phosforylase B
16-19 Hexatyrosine 3 pmol 997.396 slide pos Sample Amount on Expected
(Adducts: sp o t monoisotopic m/z
(Na+)= 1019.396 1-8 Phosforylase B 100 fmol 800<m/z<2600
(K+)=1035.396)
9-16 Phosforylase B 50 fmol 800<m/z<2600
20-24 Hexatyrosine 3 pmol 997.396
Adducts: 17-24 Phosforylase B 25 fmol 800<m/z<2600
1019.396
Each spot Include AngIII (180 fmol) and hACTH (120 fmol) as
1035.396
internal calibrants.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.5 Default method parameters
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.5.2 PSD PRO acquisition method
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.5.3 Full lenght protein PRO acqusition method
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.5.4 Protein Digest spectrum processing method
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.5.5 PSD spectrum processing method
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.5.6 Full lenght protein spectrum processing method
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.5.7 Protein digest identification method
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.6 Verification cases 4 Reflectron sensitivity
Insert test slide A. One spectrum out Resolution
Run positions con- of five should ca n b e
taining 5 fmol Neu- contain a peak in sacrificed to
Item Test method Acceptance limits comments rotensin with the the right mass obtain the
Expected resulte following method: range with S/N>3 required S/N.
1 Linear sensitivity Acquisition method or above.
Insert test slide B (if For at least one Resolution = The default met-
not already inserted). out of five spectra ca n b e hod Protein Digest
Run two positions must contain a sacrificed PRO with the num-
containing 5 fmol ile7 peak in the right to obtain ber of spectra chan-
AngIII and two mass range with the ged to 5. Manually
positions containing 5 S/N of 3 or higher. required adjust laser power
fmol Neurotensin with S/N. and target position
the following method: to obtain optimal
Acquisition method = S/N.
Use the default method 5 Reflectron resolution
Protein Digest PRO, Insert test slide A. One of five Same as for
but change detection Run positions spectra must fulfil linear
mode to linear and containing 1.5 pmol all the following resolution.
number of spectra to Ile7-AngIII and 1.5 demands: If the
5. pmol hACTH18-39 (1) A resolution monoisotopic
Manually adjust laser using the following above 10 000 peak is out
power and target methods: FWHM. of mass
position to obtain Acquisition method (2) The signal scale do a
optimal S/N. = The default met- height (see 10.5) mass
hod Protein Digest of the monoisoto- calibration
PRO with the num- pic peak must be and save it
ber of spectra chan- 300 or higher. as a system
ged to 15. (3) The mass of calibration
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
3 Linear high mass 4 Reflectron sensitivity
testI Insert test slide A. One spectrum out Resolution
nsert test slide B. The monomer Run positions con- of five should ca n b e
Run positions con- (m/z = 66 423), taining 5 fmol Neu- contain a peak in sacrificed to
taining 5 pmol BSA the dimer (m/z = rotensin with the the right mass obtain the
with the following 132 846) and the following method: range with S/N>3 required S/N.
methods: Acquisi- trimer (m/z = 199 Acquisition method or above.
tion method = The 269) should all be = The default met-
default method Full present in one hod Protein Digest
length Protein PRO spectrum with PRO with the num-
with the number of S/N>3. ber of spectra chan-
spectra changed to ged to 5. Manually
5 and the lower adjust laser power
intensity for and target position
selective to obtain optimal
accumulation S/N.
changed to 2. 5 Reflectron resolution
Spectrum Insert test slide A. One of five Same as for
processing method Run positions spectra must fulfil linear
= The default containing 1.5 pmol all the following resolution.
method Full length Ile7-AngIII and 1.5 demands: If the
Protein pmol hACTH18-39 (1) A resolution monoisotopic
Manually change the using the following above 10 000 peak is out
laser power and methods: FWHM. of mass
sample position in Acquisition method (2) The signal scale do a
order to obtain the = The default met- height (see 10.5) mass
best possible S/N hod Protein Digest of the monoisoto- calibration
on the m/z = 66423 PRO with the num- pic peak must be and save it
p e a k. ber of spectra chan- 300 or higher. as a system
ged to 15. (3) The mass of calibration
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
5 Spectrum Process- the monoisotopic and rerun the tolerance (A0) to resolution on
cont ing method = The peak must be sample. 0.02%. Identification ile7AngIII must be
default method within 2465.19897 method = The above 3000.
Protein Digest. +/- 1 Da. default method
6 Mass accuracy Protein digest but
reflectron mode with add the
internal calibration. The modification
Insert test slide A . Two of five calibration Iodoacetamide in
Run a position spectra must fulfil should be the "all appropriate
containing 180 fmol the following performed residues modified"
ile7AngIII, 120 fmol criteria. automatically field.
hACTH18-39 and 1. The protein on ile7AngIII Manually change the
225 fmol Phosforylase B and laser power and
Iodoacetamide must be identified hACTH18-3- position to get an
modified as the top rank. 9. The mass optimal trade off
Phosforylase B 2. At least 15 errors can be between resolution
digest. Use the masses must be read directly and number of
following methods: assigned to in the clearly visible
Acquisition method Phosforylase B identification p e a ks.
= The default and more than result
7 Mass accuracy The mass
method Protein 40% of the (protein
reflectron mode with errors can be
digest but change masses must details).
external calibration read directly
the number of have a mass error
Use the same Same as in 6 but in the
spectra to 5. below 20 ppm in
results as in item 6. more than 30% of identification
Spectrum the mass range
Select a good the masses must result
processing method m/z 900-2500.
spectrum from the have an mass (protein
= The default 3. The resolution
first position and error less than details).
method Protein on hACTH18-39
save the calibration 100 ppm.
digest but change must be above
on this spectrum as
monoisotopic mass 6000 and the
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
7 system calibration 8 reflectron = The and the signal
cont (both calibrants cont default method height should drop
must have been Protein Digest PRO more then tenfold.
found and have with number of 3. With the gate
clear monoisotopic spectra changed to set to 1980, the
resolution). 1. Spectrum resolution of the
Then load the processing method 2007 peak should
system calibration reflectron = The disappear and the
on all other spectra default method signal height
on the other 3 Protein digest should drop more
positions and add Acquisition method then tenfold.
the identification PSD = The default
method again. method PSD PRO
but change the num-
8 Ion gate calibration
ber of spectra to 5.
Insert test slide A. 1. With the gate See 10.1
In "waiting for PSD
Run a position con- set on 2007, a page 3 for
input" view, set the
taining 180 fmol peak with ion gate
gate on the follow-
ile7AngIII, 120 fmol resolution above calibration.
ing masses;
hACTH18-39 and 2000 FWHM and
2007.114 (true pep-
225 fmol Iodoace- signal height
tide mass), 2040
tamide modified above 200 should
(noise peak) and
Phosforylase B be present at a
1980 (noise peak).
digest. Add mass of 2007 +/-
Adjust the gate
methods for both 1 D a.
width to 2 for all
reflectron and PSD 2. With the gate
masses. Adjust
to the position. Use set to 2040, the
laser power and
the following resolution of the
position manually to
methods: 2007 peak
achieve good
Acquisition method should disappear
resolution and S/N.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
9 PSD performance, 9 with the number of
PSD mass cont spectra changed to
accuracy test 5.
Insert test slide A. Two out of five Spectrum proces-
Run positions with 3 spectra must fulfil sing method PSD =
pmol Hexatyrosine. the following The default method
In the "waiting for demands. PSD with a smoot-
PSD input" view; (1) Peaks with hing filter of width 8
calibrate the S/N>3 added. Manually
reflectron spectrum representing all adjust position and
using the mono- the following laser power to ob-
isotopic masses masses must be tain good resolution
997.396 and present; 816.888, and S/N in reflect-
1019.396. Use the 653.712, 490.536, ron mode and best
following methods 327.360. possible S/N in
for the run: (2) The mean PSD mode.
Acquisition method value of the errors Open the results in
reflectron = The de- (calculated from Evaluation. Use a
fault method Protein the absolute value spectrum on the first
Digest PRO with of measured position to calibrate
the number of spec- mass - true mass) the PSD mode
tra changed to 1. must be lower (save the calibration
Spectrum proces- than 1,0 Da. as system calibra-
sing method tion). Use the exact
reflectron = The average masses
default method Pro- 998.07 and 327.36
tein digest. for this purpose.
Acquisition method Finally, add the
PSD = The default PSD spectrum
method PSD PRO processing method
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
9 described above 10 Acquisition method
cont and load the system cont PSD = The default
calibration on each method PSD PRO
position. with the number of
spectra changed to
10 PSD sensitivity test
5.
Insert test slide A. Two out of five
Spectrum proces-
Run positions with spectra must fulfil
sing method PSD =
60 fmol Hexatyro- the following
The default method
sine. In the "waiting demands:
PSD with a smoot-
for PSD input" view; 1. Peaks with
hing filter of width 8
calibrate the reflec- S/N>3
added.
tron spectrum using representing all
Manually adjust pos-
the monoisotopic the following
ition and laser pow-
masses 997.396 masses must be
er to obtain good
and 1019.396. Use present; 816.888,
resolution and S/N
the following met- 653.712, 490.536,
in reflectron mode
hods for the run: 327.360.
and best pos-sible
Acquisition method 2. The mean value
S/N in PSD mode.
reflectron = The de- of the errors
Open the results in
fault method Protein (calculated from
Evaluation. Use a
Digest PRO with the absolute value
spectrum on the first
the number of spec- of measured
position to calibrate
tra changed to 1. mass - true mass)
the PSD mode
Spectrum must be lower
(save the calibration
processing method than 1,0 Da
as system calibrat-
reflectron = The
ion). Use the exact
default method
average masses
Protein digest.
998.07 and 327.36
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
characteristics, MCP´s etc). The method settings therefore
10 for this purpose.
need to be regularly checked. The same is true after service or
cont Finally, add the
transport operations.
PSD spectrum
processing method
described above 2.2.8 Description of laser Characteristics in Ettan MALDI-
and load the system ToF Pro
calibration on each
position. 2.2.8.1 Sample slide dimensions.
2.2.7 Auto tuning setup.
The sample slide consist of 24 different sample positions
An important function in order to perform high-throughput ex-
separated a distance of 2.25 mm (center to center). The diam-
periments with MALDI-TOF MS is automatic data acquisition.
eter of each sample position is 1.55 mm (see Figure 1).
The Ettan MALDI-TOF auto-tuning algorithm is designed to
support this function. Laser power and sample position is
changed, according to the rules of the algorithm, in order to 1.55 mm
obtain peaks with high signal to noise ratio as well as
monoisotopic resolution. However, for the automatic acquisi- 2.25 mm
tion to function properly settings for the auto tuning as well as
the acquisition methods have to be optimized. In this memo
the function of the auto-tuning algorithm is described, followed
by optimization procedures in order to find start settings for Figure 2. Sample slide
Figure 1. Sample slide dimensions.
different instrument modes. It should be noted that these set- movement.
tings most likely need tailored optimization for each individual
instrument on the market (depending on laser characteristics). 2.2.8.2 Spot size and Laser intensity profile
Optimal settings is also influenced by which type of matrix that
In Ettan MALDI-ToF Pro the laser beam is focused at a con-
is used as well as sample preparation, (matrix type, relative
stant position (in the plane of the sample slide) that is aligned
matrix quantity, solvent type, salt content, spotting technique
with the extraction lens of the accelerating system. The laser
etc.).
spot is moved over the sample slide by movement of the sam-
It is important to stress that after a certain period of instrument
usage the characteristics of the instrument will change (laser
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
ple slide itself. Movement is only performed in one direction
(see Figure 2).
The focused laser spot is shaped like an ellipse with the ap-
proximate dimensions of 0.8x1.6 mm (1.6 mm in the direction
of slide movement). However, the laser intensity varies over
the spot area. The maximum intensity and intensity profile
varies between laser individuals. See Figure 3 for intensity
profiles from 5 different laser individuals (white = highest inten-
sity). The size of the area with intensity between 80-100% of
the maximum is normally between 0.08-0.15x0.25-0.5 mm
(0.25-0.5 mm in the direction of movement).
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.8.3 Control of laser power
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
total tuning time for the position has reached the time-out limit.
The selective accumulation box set the criteria for accept-
able data for accumulation. For each 1- or 8-shot bundle, the
highest and the lowest intensity value in the defined mass
range (as defined by the low and high values in Figure 5) are
extracted. The difference between the highest and the lowest
value is calculated and divided by the laser frequency (1 or 8).
If the obtained value falls within the defined intensity range,
the signal from the bundle (1 or 8 shots) is accumulated. If not,
the laser power or target position is changed according to the
auto-tuning algorithm (see below). For each change in laser or
position performed during the auto tuning the first shots (the
exact number specified by the parameter ”No. Of initial shots
to ignore”) are neither accumulated nor considered by the
auto-tuning algorithm.
During fast scan the laser is kept at its initial value while the
whole sample position is searched in the number of steps
specified by the Consecutive Adjustments parameter (figure 6
and 7). If an acceptable shot is found the algorithm continues
to refine mode. If no acceptable shot is found during fast scan,
the algorithm continues to initial mode.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Fast scan Initial (or refine) mode
B
Signal
No signal Signal
*The optimization procedure includes deactivation of the laser power check value
(see 2.2.10.1.1) since there are other embedded features of the software correcting
for the same thing. 6 4 4
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.10.1.2 Acquisition method
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.10.2 Post source decay with 4-hydroxy-alpha-cyano cinnamic acid as matrix
Except for the parameters in the acquisition method, the same parameters as in peptide mass fingerprinting can be used.
2.2.10.2.1 Acquisition method
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.10.3 Full length protein analysis with sinapinic acid as
matrix
The greyed out boxes contain default values and should not
be changed (see 2.2.11.1).
toSet
according
to 2.2.11.1
6 4 4
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.10.3.2 Acquisition method
32
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.10.4 Acquisition method for tuning of initial laser
power
Low Mass rejection and auto tuning is disabled and that the 3
first shots should be ignored.
100
3
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.10.5 Acquisition method for tuning of intensity range
for selective accumulation (PMF)
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.11 Guidelines for optimization of parameters “auto adjust laser power” as well as “sample position” and
successively increase laser power (in quite large steps) until
Depending on the type of samples analysed, matrix type, laser you reach the point just above threshold where signals are
characteristics of the instrument etc., different parameter set- accumulated.
tings are needed. Guidelines for starting optimizing the param- 4. Decrease laser with 80 units, move the slide 20 subposi-
eters are found below. tions and then successively increase laser power, 20 units in
each step. If an increment of laser power does not results in
any signal accumulation after approximately 5 seconds (or
2.2.11.1 Initial laser power and laser power adjustment less), increase the laser power by 20 units. It is important to
parameters perform the different iterations swiftly in order to avoid unnec-
essary loss of material. Threshold value is reached when (at
Desorption and ionization in MALDI is a threshold phenomena. least) one spectrum (80 shots) is automatically accumulated.
The laser threshold varies largely with the type of matrix used, Then move the slide another 20 subpositions and perform the
the specific laser of the instrument and (to a lesser extent) the described laser adjustment procedure once again. Record
nature of the matrix crystals (not homogeneous matrix layers). each threshold laser value for 10 sample positions.
For good results it is important that the laser is operating in a 5. Remove the laser threshold values those obvious differ
power regime ranging from the lowest threshold value to from the data set and then calculate the average threshold
slightly to moderately above the highest threshold value for a laser power and set the initial laser power to that value. For
certain matrix. To set appropriate values for laser parameters PSD investigations, set the initial laser power to 1.4 x average
we recommend you to start with the following procedure. threshold value.
1. Prepare a matrix solution of the matrix you want to optimize 6. Remove the laser threshold values those obvious differ
the laser settings for. Spot the solution to all 24 positions on a from the data set (as above). Calculate the value 1.2 x (maxi-
sample slide using the same spotting method you are using for mum threshold value – minimum threshold value)/4 and set
real samples. the laser power maximum adjustments value for both initial
2. Investigate the threshold value for the different positions by and refine mode to the closest upper multiple of 10. If the
using the acquisition method described in 2.2.10.4. maximum threshold value is 240 and the minimum threshold
value is 150 this yields a value of 30.
3. Start acquisition. The first step is to approximately find the
range where the threshold value is. In processing view, disable 7. Set the laser power consecutive adjustments to 4 in initial
mode and 2 in refine mode.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.11.2 Mass range for selective accumulation concentration, presence of other molecules, ease of ionization,
The mass range used for selective accumulation should cover incorporation in the matrix, type of matrix and heterogeneity of
the m/z range in which most of the analytes are likely to ap- the sample.
pear. If possible, m/z areas often displaying contaminating
peaks should be avoided. 4-hydroxy-alpha-cyano cinnamic Resolution Signal intensity
acid generates cluster ions in the range of m/z = 190-1000.
Therefore, for PMF, it is recommended to set the lower mass
range value to 1000 when this matrix is used. The recom-
mended upper mass range value for PMF using 4-hydroxy-
alpha-cyano cinnamic acid is 2200 in order to tune on analyte Laser power
signal instead of calibrants.
For PSD, the mass range for selective accumulation should Threshold value
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2. Use the same acquisition method as described in 4.4 but 2. Create a new PSD acquisition method using the default
use a constant laser of 1.2 x average threshold level (closest PSD Pro method but with a start laser power of 1.5*minimum
lower multiple of 10), a low ion reject value of 500, a mass threshold. Acquire PSD data from different positions. Set the
range of 1000-2200 (m/z). Set the lower limit for selective gate to m/z =900, 1500 and 2000 (by selecting a noise peak at
accumulation different values for different positions (from 1 to that mass in “Waiting for PSD input” view). For each gate
10). Acquire data from the different positions and find out the value, try different values of the lower intensity value to find
background noise level (chemical and electrical), i.e. a lower out the background noise level.
limit that just barely yields accumulation.
3. Set the lower intensity limit to a value that avoids accumula-
3. Set the lower intensity limit for selective accumulation to a tion of background noise regardless of gate value.
value that avoids accumulation of background noise.
4. Set the upper range for selective accumulation to 10 x
4. Set the upper limit for selective accumulation to 10 x lower lower limit for selective accumulation.
limit for selective accumulation.
5. Preferentially verify the your settings on a CAF-derivatized
5. Preferentially verify your settings on a digested gel sample sample lacking proteins (a blank). The lower intensity limit
lacking proteins (a blank). The lower intensity limit for selective might need some fine-tuning.
accumulation might need some fine-tuning, due to higher salt
content leading to higher chemical noise (matrix clusters etc). 2.2.12 The test procedure of automatic acquisition
If there are problems with matrix clusters in the mass range >
500 (m/z) the low mass reject can be changed to 750 (m/z)).
2.2.12.1 Introduction
2.2.11.5 Optimisation procedure for PSD analysis One of the most important steps during a installation of a Maldi
Pro is to optimize the auto tuning set up and the start settings
1. Prepare a matrix solution of the matrix you want to optimize for different run modes. The procedures for this are described
the settings for. Spot the solution to all 24 positions on a sam- in chapter 2.2.7-2.2.11.5.
ple slide using the same spotting method you are using for
This section describes a test that is aimed to verify the per-
real samples.
formance of automatic acquisition (after method optimization).
This is a part of the acceptance test for Ettan MALDI ToF Pro.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.12.2 Software methods 2.2.12.4 Spectrum processing method (spm)
32
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.2.12.5 Identification method clude AngIII (180 fmol) and hACTH (120 fmol) as internal
calibrants for all dilutions.
Table 1. Sample composition of the test slide.
1-8 100
9-16 50
17-24 25
1 100 19 25
2 100 20 25
3 100 21 25
4 100 22 25
5 100 23 25
6 100 24 25
9 50
· The internal calibrants of slide C are in large excess result-
10 50 ing in the fact that these 2 peptides will be the dominating
ones in the spectra. This is particularly true for the digests of
11 50 low concentration. The quality of the acquisition is best viewed
if a mass window that is narrower than the mass range for the
12 50
calibrants are used.
13 50 · To assure high-quality identification of the material of test
slide C, assure that a high quality system calibration recently
14 50
has been performed.
15 50 · Assure that the sample slide was prepared in less than 48
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
hours ago. The matrix is degraded with time leading to lower 1 +…Probability N*isotopic mass N). The probability for pep-
sensitivity, something that is critical for samples with low tides containing one or several C13 atoms increases with in-
amount. creasing peptide mass (Increasing number of carbon atoms).
· If the instrument does not pass test for automatic acquisi- The difference between an average and a monoisotopic mass
tion, please contact the person responsible for the acquisition is approximately 0.06%. See Figure 10 and Figure 11 for ex-
optimization procedure if the instrument in question and amples of monoisotopic and average masses for two peptides
specify the problem (to long acquisition time, noise is accumu- of different mass. In the verification, monoisotopic masses are
lated instead of analyte signal, etc). used for mass accuracy measurements in reflectron and aver-
age masses are used for mass accuracy measurements in
PSD (and for the masses reported in high mass sensitivity).
2.2.14 Measurement definitions The software will automatically calculate these masses if the
correct method settings are used (see Ettan MALDI-ToF Pro
2.2.14.1 Monoisotopic and average masses user manual).
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Baseline
Figure 12
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Signal height
Peak signal R = (m/z)/FWHM
m/z
FWHM
10 Da 10 Da
W
M
H
F
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
This is a theoretical definition of noise and is fully correct but
difficult to use. To make it user-friendlier we calculate the noise
as follow:
Signal height Noisep-p/2
Peak signal
So the calculation off s/n is done as follows:
signalp-p∗2/noisep-p
2.2.14.6 Calculation of mass errors in PSD
10 Da 10 Da
Mass errors for investigation of PSD mass accuracy should be
calculated in the following way.
(1)Average mass error = (1/n)∗∑√((measured massi-true
mass)2); where j = 1 to n and n = number of fragments.
(2)Maximum fragment error = maximum of all √((measured
mass-true mass)2).
Figure 15
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.3 Acceptance test document
Technical Specification
lation of a Maldi ToF Pro. When the installation is finished and Minimum retention time: 10 years after termination of the product(s) concerned document no.
edition
219223509
AA
title
customer the final page of the acceptance test has to be Maldi Pro acceptance test
project name date
signed by the customer. If the customer does not sign the Protfly
author(s)
Mats Källgren
2003-03-03
project no
986148
acceptance test the installation has not been completed.
approved by
Anders Tangen
Customer:
Department:
Street:
City:
Post code:
Telephone:
Fax:
Email:
User:
Instrument:____________________________
Serial number:________________________________
Version:_____230V ,_____110V
Computer type: ___________________________ Serial number:___________________
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Technical Specification
Technical Specification Maldi Pro acceptance test
document no. 219223509
Maldi Pro acceptance test edition AA
document no. 219223509 page 3 (6)
edition AA
page 2 (7)
Reflectron resolution test
Acceptance limit: Resolution> 10000, FWHM (R) for the monoisotopic peak in the isotopic
cluster of 1.5 pmol hACTH 18-39. The result should be obtained accumulating 200 shots per
spectrum in 8-shots mode.
All methods used in these tests are described in the installation guide. The preparation of hACTH 18-39 Resolution
samples for the tests is described in the installation guide. 2465,199 m/z
Acceptance limit: S/N > 3 for the m/z 1672.89 peak of 5 fmol Neurotensin, accumulating 200 Compare measured masses with theoretical, using 100-fmol iodoacetic acid modified
shots per spectrum in 8-shots mode. phosphorylase b tryptic digest.
Acceptance limit: More than 15 masses must be assigned to Phosphorylase B and for 40% of
the assigned masses, the error should be equal or less than 20 ppm, in the mass range 900-
Neurotensin 5 fmol Signal to noise
2500 m/z.
1672,89 m/z
Mass errors
Linear Resolution test total matched <20ppm
First approved
Acceptance limit: A resolution t1000, FWHM (R) should be achieved for the monoisotopic spectrum
peak in the isotopic cluster of 1.5 pmol Ile 7 Ang III, m/z 897,531. The result should be Second approved
obtained accumulating 200 shots per spectrum in 8-shots mode. spectrum
The monomer, m/z 66431, the dimer, m/z 132859 and the trimer, m/z 199287 of 5 pmol BSA Compare measured masses with theoretical, using 100-fmol iodoacetic acid modified
should be detected in the spectra. The result should be obtained accumulating 200 shots per phosphorylase B.
spectrum in 8-shots mode. Acceptance limit: More than 15 masses must be assigned to Phosphorylase B tryptic digest
and for 30% of the masses, the error should be equal or less than 100 ppm.
BSA full-length 5 pmol Detected
66431 m/z
BSA dimer at 132859m/z Mass errors
total matched <100ppm
BSA trimer at 199287m/z First approved
spectrum
Reflectron mode performance Second approved
spectrum
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Technical Specification Technical Specification
For 24 positions of a slide, containing 100, 50 and 25 fmol iodoacetic acid modified
Phosphorylase B and after accumulation of 1 spectra on each position, following criteria must
be fulfilled:
Acquisition time. Total acquisition time for all 24 spectra should not exceed 40 minutes.
Quality of identification result; Phosphorylase B must be identified as the highest ranked
candidate on all positions. 20 of 24 spectra must have an expectation below 0.01.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Technical Specification
ACCEPTANCE
Amersham Biosciences personnel have given an introduction on handling and work with the
instrument and the software.
Amersham Biosciences will not be responsible for loss of stored or generated data.
I here by fully accept that the Ettan MALDI-ToF Pro has passed all test and works according
specifications. I agree that installation of the instrument has been completed and warranty
begins.
Signature:
Name:
Date: Location/Place:
Customer:
Signature:
Name:
Date: Location/Place:
2002-01-01 70-7611-09 / AH
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.4 Software Installation Oracle Client: Oracle9i Client Release 1 (9.0.1.1)
Installation of Windows operating system will not be coverd Shorthand Installation Guide.doc.
in this manual. If you are attempting to install a client server installation
2.4.1 Oracle installation you also need to install Oracle client runtime
components. Please refer to document below: Oracle9i
Client Runtime Release 1 (9.0.1.1) Shorthand Installation
Guide.doc.
Database Backup: Amersham Biosciences Oracle9i
Before You Begin Database Backup Shorthand Installation Guide.doc.
This guide is shorthand installation requirement information
for Oracle9i on MS Windows. For a detail system 2.4.1.1 Oracle9i Database Installation Guide Release 1
requirements information refer to Oracle documentation (9.0.1.1) for Windows 2000
located on your Oracle CD (1 of 3) at: \Doc\ a90162.pdf
This guide is shorthand installation requirement information for
Oracle is a registered trademark of Oracle Corporation. Oracle9i on Windows 2000.
Oracle9i is trademark of Oracle Corporation. For a detail system requirements information refer to Oracle
documentation located on your Oracle CD.
You should have some Oracle DBA experiences or have
some knowledge about databases to install Oracle Data- Oracle is a registered trademark of Oracle Corporation.
base or Oracle Client and you should have some knowledge System Requirements for FAT and NTFS File Systems
about Network. Oracle Corporation recommends installing on NTFS for
Following shorthand documents can be used to install: Windows 2000.
Oracle9i Database: Oracle9i Database Release 1 (9.0.1.1)
Shorthand Installation Guide.doc Oracle9i Database System Requirements
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Enterprise Manager Client on the same computer, the
minimal processor requirement is a Pentium III 866, the
Requirement Standard Edition
recommended processor requirement is a Pentium III 1
GHz, the minimal RAM requirement is 128 MB and the
Operating System Windows 2000 Note 2 recommended RAM requirement is 256 MB.
Windows 2000 Service Pack Not required; certified with 1 or higher
5
You cannot run Oracle Universal Installer and Oracle Data
Migration Assistant or Oracle Database Configuration
Minimal Processor Note 4 Pentium 166 or Pentium 200 Assistant during the same installation session on a 128 MB
computer. To run these assistants, answer No if prompted to
Recommended Processor Pentium 233 or Pentium 266
migrate or create a database. After installation is complete
RAM Note 5 128 MB (256 MB recommended) and Oracle Universal Installer has exited, run these
assistants.
NTFS file system 6
If your default temp space is less than 400 MB, then
· Oracle home drive 2.7 Gb updates the TEMP user environment variable to a location
with the required amount of space. You must reboot your
· System drive 140 Mb computer for this change to take effect.
Temp Space Note 6 400 Mb
7
Depending on how many applications are running on the
computer, you may need to further increase the paging file
Virtual Memory Note 7 Initial Size 200 MB , Maximum Size size or reduce the size of the System Global Area (SGA) if
400 MB you run out of virtual memory. Note that if temporary files
V i deo 256 Color and the paging file are both stored on the same physical
drive, a situation can occur where the space requirements
for one can limit the size of another. If your system has
limited free space, then first install the oracle9i software.
2
Windows 2000 includes: Windows 2000 Professional, After the installation is finished, create a database with the
Windows 2000 Server, Windows 2000-Advanced Server, Oracle Database Configuration Assistant.
and Windows 2000 Datacenter.
4
If you run the Oracle Intelligent Agent, Oracle Management 2.4.1.2 Install Products on Windows 2000
Server (includes Oracle HTTP Server), and Oracle
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Click Cancel in any Oracle Installer dialog box to exit.
To install Oracle9i Products:
1. Start your operating system. If you are running Windows
2000, log on as a member of the Administrators group.
2. Stop all Oracle applications (if any are running).
3. If installing on Windows 2000, Stop all Oracle services (if
any are running). To stop services
a. From the Control Panel, select Administrative Tools and
then select the Services icon.
b. If any Oracle services exist and have the status Started,
select the service, and click stop.
c. Click Close to exit the Services windows.
4. Insert the Oracle9i for Windows 2000 CD-ROM. The operating
system automatically detects your Oracle9i for Windows 2000
CD-ROM. The Oracle9i Autorun dialog box appears and prompts
you to make a selection:
· Install/Deinstall Products
· Explore CD
· Browse Documentation
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Oracle Installation Setting Destination dialog box
appears:
File Locations
Enter the stage location of your products and the location
where you want to install your product.
Please specify full paths with up to 32 alphanumeric
characters.
6. Click Next.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Source The Select Installation Options dialog box appears:
Path: Enter the full path of the products.jar file from which
the product(s) will be installed. You may also use the
Browse button to find the products.jar file.
Note: Do not leave any fields empty.
7. Accept the default values and then click Next
(Recommended).
The Loading Product Information process bar appears:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Select Installation Types dialog box appears: The Loading Product Information process bar appears:
11. Select Custom and then Click Next. Please wait until loading product information is finished.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Select Product Components dialog box appears: See Checked Components:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
See Checked Components: See Checked Components:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Loading Product Information process bar appears: The Select Alternate Location dialog box appears:
Please wait until loading product information is finished. 13. Accept the default location and then click Next.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Loading Product Information process bar appears: The Create Database dialog box appears:
Please wait until loading product information is finished. 14. Select No and then click Next.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Loading Product Information process bar appears: The Summary dialog box appears:
Please wait until loading product information is finished. 15. Click Install.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Install dialog box appears: The Disk Location dialog box appears:
Installation process continues… Insert Oracle9I disk 2 into your disk drive. Installation
process continues…
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Disk Location dialog box appears: The Configuration Tools dialog box appears:
Insert Oracle9I disk 3 into your disk drive. Installation Installation process continues…
process continues…
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Net Configuration Assistant dialog box appears: The Directory Service Access dialog box appears:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Listener Configuration dialog box appears: The Select Protocols dialog box appears:
18. Click Next. 19. Select TCP(ask your network administrator for help)
and then Click Next.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The TCP/IP Protocol dialog box appears: The Listener Configuration dialog box appears:
20. Accept the default standard port number (ask your 21. Select No and then click Next.
network administrator for help) and then Click Next.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Net Configuration Assistant dialog box appears: The Naming Methods Configuration dialog box appears:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Net Configuration Assistant dialog box appears: The Configuration Tools dialog box appears:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The End of Installation dialog box appears: The dialog box appears:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
2.4.1.2 Amersham Biosciences Products Database 1. Choose Start > Programs > Oracle – OraHome90 >
Installation for Oracle9i on Windows 2000 Configuration and Migration Tools >
Before you begin: Database Configuration Assistant.
Copy file AmershamBiosciencesRDBMS.dbt to:
(ORACLE_HOME)\oracle\ora90\assistants\dbca\templates
Installation Process
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Database Operation Dialog Box appears:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Database Template Dialog Bx appears: The Database Identification Dialog Box appears:
4. Select AmershamBiosciencesRDBMS and then click 5. Accept the default values and then click Finish
Next. (Recommended).
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Database Summary Dialog Box appears: The Database Creation Progress dialog box appears:
6. Click OK.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The Database Creation Progress dialog box appears: When Database creation is complete … then the following
dialog box appears:
The installation process takes some time, please wait! 7. Click Password Management…
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The following page appears: Do not change the password for SYS and SYSTEM account
until the entire installation is complete.
8. Unlock Lock Accounts and then see the next page. All database accounts except SYS and SYSTEM have been
locked. Unlock the accounts and change the default
passwords.
Please Note the information in dialog box and keep it save.
9. Click OK.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
following page appears: Create Database Schema
Copy the Product database installation files from your media
into the HD where Oracle Database is installed, se example
below for Maldi installation:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
11. Open an DOS Command Prompt. 12. Go to the Product database directory.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
The installation process will begin:
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Open file AmershamBiosciencesMaldi.log and search for
ORA-.
fI
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
If installation is successful then you will see:
Instalation is complete.
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Chapter 3
Chapter Contents
3. Interfacing
Interfacing
3.1 Introduction
3.3 Cables
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
3. Interfacing figure 2: interconnections
Mains outlet
Figure 1 beside shows the mains connection and a fuse table PCB MS 1 Network LPT1
Parallel Network
Figure 2 shows the interconnections between Maldi ToF Ms Mains outlet Optional
and computer. Network terminal Mains outlet Network terminal Mains outlet
Mains
witch WARNING!
Mains
inlet For continued protection against risk
of fire, replace only with fuse of the
Mains
fuse
specified type and current ratings.
100 - 120 V~ T8A
220 - 240 V~ T4A
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
3.2.2 TCU connectors
3.2 Table of connectors
The connections of the TCU, Tof Control Unit, internal and
external are described on the following pages.
3.2.1 Instrument connections
Connections between instrument and PC and for the PC inter-
nal connections, see figure 2 on previous page.
picture of the TCU from the back
Mains
PC 1
PC 2
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
TCU Backpanel conectors
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
TCU Connector/Pin layout
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
TCU Connector/Pin layout
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
TCU Connector/ Pin layout
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
TCU Cable Interconnections Block Diagram
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
TCU Interconnections Block diagram
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
3.3 Cables Connects between the TCU (U01) and the Reflectron HV sup-
List of all cables ply (U06).
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
15. Valves and covers cable. 23. Pulser trigger cable.
Cable connects between the TCU (U01) and (C0245A) which Cable connects between TCU and the pulser (U13).
consists of Vent valve, Inlet valve and covers loop.
24. Computer interface cable.
16. Pirani gauge cable. Cable connects between the computer and the TCU (U01).
Cable connects between the TCU (U01) and the Pirani gauge
(U17). 25. Gate trigger cable.
Cable connects between the TCU ( U01 ) and the gate driver
17. Magnetron gauge cable. (U21).
Cable connects between the TCU (U01) and the Magnetron
Gauge (U18). 26. Laser detect cable.
Cable connects between the TCU (U01) and laser detect assy
18. Laser power cable. (U10).
Cable connects between the TCU (U01) and the Laser (U19).
27. Laser trigger cable.
19. Beam blanking cable. Cable connects between the TCU (U01) and the Laser (U19).
Cable connects between the TCU (U01) and the beam blank-
ing (U20). 28. Linear detector signal cable.
Cable connects between the TCU (U01) and the linear de-
20. Gate power and controll cable. tector amplifier (U02).
Cable connects between the TCU (U01) and the gate driver
(U21). 29. Reflectron detector signal cable.
Cable connects between the TCU (U01) and the flight tube
21. Pulser power supply input cable. (U2).
Cable connects between the TCU (U01) and the pulser supply
(U22). 30. Linear detector amp int power.
Cable connects internally in linear detector amplifier (U02A).
22. Sample motor cable.
Cable connects between the TCU (U01) sample probe con- 31. Sample inlet interconnect.
nection and the sample probe (U23). Cable is located in Sample Inlet assy (U03).
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
32. Pulser control internal cable. ac pcb.
Cable is located internally in the pulser assy (U13A).
42. Scope trigger ttl TCU.
33. Valves & covers interconnect. Cable connects between TCU backplane (U01) and the ac
Cable connects to vent valve, inlet valve and and instrument pcb C0257.
covers.
43. Switched mains TCU.
34. Beam blanking control internal. Cable connects from rear of the TCU backplane and finishes
Located internally in beam blanking unit (U20A). on the front of the backplane.
39. Computer interface cable TCU. 49. Switched power supply unit.
Cable is located on TCU (U01) backplane pcb. Located in the TCU (U01).
40. Pulser trigger cable TCU. 50. Laser detect internal power cable.
Cable is located in the TCU (U01) between the backplane and Located in the laser detect assy (U10A).
ac pcb.
51. Mains distribution cable.
41. Gate trigger cable TCU. Cable connects from mains inlert to the mains distribution.
Cable is located in the TCU (U01) between the backplane and
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
52. Reflectron turbo mains cable.
Cable connects from the mains block to the reflectron turbo
controller (U09) and the reflectron turbo fan.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Chapter 4
Chapter Contents
4. Troubleshooting
Troubleshooting
4.1 Vacuum system
4.2 Mechanics
4.3 Electronics
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
Vacuum problem
No vacuum
Vacuum reading goes up Yes Sample probe does not
No with probe in forward seal in forward position
Pumps on? Start/repair pump position
Yes No
No
Pirani reading ok Clean/replace Pirani. Reading Vacuum reading unstable Yes Replace drive shaft seal
OK? during probe movment?
Yes No No
Vacuum OK
Fill up oil in rotary pump/replace
pump. Replace molecular sieve
No
Magnetron reading ok Clean/replace door o-ring
Clean/replace Magnetron
Yes Repair/replace leaking valve
Clean/replace leaking o-ring
Leakcheck air admit valve
Leakcheck the sample inlet
valve assembly
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
No peaks detected:
Some peaks detected, low sensitivity, low resolution
No peaks A big unresolved peak Yes Pulser problem
at low mass?
No
Is interface and interface No Fit interface and interface No
cable firmly fitted? cable firmly. Is the sample good? Use a good known sample
Yes (Bic ink).
Yes
Is the laser shutter open? No Open the shutter. Disconnect and ground beamdump and gate.
Yes
Does the laser fire? No Connect M77, Change Acq,T/R Adjust sample position and laser mirror.
pcb.
Yes
Does the laser hit No Adjust sample position
No
the sample? and laser mirror. Correctly set laser trig level? Check laser trig level
Yes Yes
Are the HV on? No Check HV readbaks. No
Correct acq methode? Adjust focus mass/delay
Yes time. Beam dump set correctly?
Yes
No
Is the sample good? Use a well known fresh sample No high mass peaks in No Low laser energy,
(Bic Ink). linear mode? adjust/replace laser.
Yes
Does the Maldi work No Check the other mode. Ground
in booth modes? beam dump, gate and kicker.
Yes
Is there a zero burst No Replace T/R
in reflectron mode? pcb
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
4.1 Vacuum system
4.1.1 Trouble shooting the vacuum system
Typical values for the vacuum are on the Pirani 5e-03 mbar and
on the Magnetron 5e–07 mbar. This is not any absolute figures
that has to be reached, the main thing is that the vacuum
gauges reaches its trip point so the HV can be turned on.
Turbo pump If the instrument can`t reach this values check the following:
-Is the rotary pump on and is the oil level normal.
-Are the turbo pumps on, is the led on each controller green.
The pumps are switched on by the mains switch on the Maldi.
-Does the sample door close, clean the back of the door and
the o-ring, replace the o-ring if it is worn.
-Is the pirani reading OK, but not the magnetron? Check the
cable to the magnetron. Tap on the gauge with a screwdriver
to see if it will turn on. If it turns on by tapping on the magnet-
N2 or dry air ron housing, it needs to be cleaned. See Edwards manual for
a proper cleaning procedure of the magnetron. If the reading
Turbo vent valve is 1e-8 mbar, it is likely that the electronics in the magnetron is
faulty or the magnetron has not been turned on and the magn-
Pump out line etron needs to be cleaned or replaced.
for probe bearing Backing valve -Is only the pirani gauge faulty and the magnetron OK. Check
Pirani
Turbo pump
the cable, remove the gauge and rinse it inside with methanol.
If this does not help, replace it.
Forline trap To turn on the magnetron four events must take place: sample
Door Magnetron
(penning gauge)
door must be closed, both turbo pump up to speed, the panel
interlock switch must be closed, the vacuum reading on the
N2 or pirani gauge must be lower than its set point. All this events
dry air Vent valve Inlet valve
Rotary pump
are represented by a closed contact (activated relay) con-
nected in series on the I/O pcb. When the magnetron is turned
on and the reading has passed its set point it will be possible
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
to turn on the HT and start a run. probe in its forward position, loosen the four allen screws that
holds the sample lock compartment in place. Remove the vent
4.1.2 Sample probe seal leak valve and inlet valve assembly by losen the Swagelock nut.
If the vacuum is good with the sample door closed but it is to Remove the sample lock by pulling it outwards. Replace the
low when the door is opened. This indicates a leak between o-ring and inspect the sealing surface of the sample probe.
the sample probe and the lock. The first thing to try is to adjust Reassemble the sample lock compartment in reverse order.
the micro switch for the sample exchange position. Move it The lock will be guided to its correct position by the sample
approximately 0.5 mm clockwise to move the probe further in probe when it is in its forward position. Close the sample door
to the lock. This will improve the seal. The correct position for manualy, reconnect all contacts and switch on the instruments
the micro switch is when the magnetron reading improve a to start up the pump down sequence. The Maldi has to have
little bit when the sample probe reaches the sample exchange good vacuum before it can be tested again.
position and the sample door opens. Make sure that the motor The first thing to look for if the motor does not move is that all
does not have to work to hard to activate the micro switch. If the contacts are in place. Check the 24 V power supply for the
the leak is still there the seal has to be cleaned or replaced. stepper motor. If the voltage is OK, check the micro switches
To get access to the seal in the sample lock the instrument for the motor. If the micro switches is OK, make sure you get
has to be vented. If possible vent the instrument with Nitrogen 24 V pulses to the stepper motor from the I/O pcb. The motor
(to minimize pump down time). When the instrument has been is controlled by IC 34, 35, and 40. If the pulses is OK replace
vented, open the sample door manually. It is important to dis- the motor, if the pulses are missing replace the I/O pcb.
connect the door motor before it is moved manually. This will
prevent back emf from the motor to damaging the driver circuit 4.1.3 Leaking seals at the probe drive shaft
on the I/O pcb. When the sample door is opened and the If there is a fluctuation in the magnetron reading during the
probe is in its inner position it is possible to see the o-ring in movement of the sample probe there is a leaking seal at the
the sample lock. probe drive shaft. To replace the probe drive shaft seals is a
Try first to clean the o-ring and the sealing surface of the lengthy procedure, so make sure that you have necessary o-
sample probe with some methanol on a cotton bud before the rings pn: 18114588 and vacuum grease before you start on
o-ring is replaced. If the o-ring is damaged it has to be re- this work.
placed. To replace the o-ring, the sample probe has to be in It is also recommended to update the probe drive shaft to its
its forward position. If the system is vented the probe has to latest version at the same time, see SI no 10 for part numbers
be moved manually. Disconnect the motors electric contact to and more information. If the drive shaft is the latest version
prevent the I/O pcb to be damaged by back emf. With the only the o-rings should be replaced.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
Vent the instrument. Remove the Laser Optics Bench. Undo do this but you can usually feel when the drive screw has re-
the cables to the Switch Gear assembly and remove the assy engaged. Now very gently push the drive shaft back into it’s
from the drive housing. To do this, undo the grub screw fas- housing, rotating it clockwise at the same time. If you force the
tening the drive shaft coupling to the motor shaft. Undo the shaft in without turning simultaneously you will almost cer-
four allen screws fastening the gear assembly to the drive tainly damage the o-rings. You can feel the resistance as each
housing. Undo the bearing housing and remove the outer o-ring starts to enter it’s proper location. When it is fully in-
bearing with it. Now the drive shaft can be gently pulled and serted simply refit everything as you took them apart. This
rotated anticlockwise. The rotation unscrews the drive screw procedure can result in the sample offset being altered.
from the sample holder. It should slowly come out from its seal
housing. When free ensure the drive screw has fully disen- 4.1.4 Sample probe positioning and sample offset
gaged. If not continue to turn anticlockwise until it has. Re-
move the drive shaft. Two seals are fitted. The seal nearest to The start position is set by the upper (large) micro switch on
the peek drive screw is the high vacuum seal and the other is sample exchange mechanism. The distance the sample probe
the low vacuum seal. These are a tight fit and are difficult to has to move from the start position to where sample no.1 can
remove and refit. Do not scratch the sealing faces on the drive be hit by the laser is called sample offset and is set in the
shaft. Use tweezers and something nonmetallic to jam under Maldi def file. A typical value is about 18000-20000. This is
the seal. Remove the low vacuum seal first. You can slide this actually the number of steps the sample probe has to move to
over the high vacuum seal to remove it. get the first sample in place. It is 320 steps/mm. The proce-
When you have removed both seals make sure you have not dure to set up sample probe offset is described below.
scratched the o-ring groove faces. If there are light scratches
you can remove them with very fine grade emery paper. If 4.1.5 Sample probe offset adjustment and laser beam
there are deep scratches then the drive shaft will need to be targeting
replaced. ALWAYS USE LASER SAFETY GLASSES WHEN WORKING
Ensure you clean the shaft before reassemble. Apply a small WITH THE LASER.
amount of the vacuum grease to the o-ring grooves. Fit the Prepare a slide with marker pen so it easy to identify the sam-
new high vacuum seal and then the new low vacuum seal, ple spots inside the source. Mark for instance spot 1-3 and
which will slide over the other one. Again apply a little more 22-24, it easier to identify the spots if different colours are
grease to the outer circumference of the two o-rings. Now refit used. The source can be illuminated by a torch. Shine the
the drive shaft by first screwing the drive screw back into the light through the laser mirror and lens to illuminate the sample
sample holder. You cannot see what you are doing when you slide inside the source. Load the sample slide and start a run
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
at position 1. Block the laser beam with a business card or until the peaks are back. Make a not off the knobs position.
similar after the filter wheel assembly. Observe the sample Continue to turn the knob clockwise until the peaks are gone.
slide through the laser mirror, adjust the sample motor offset Make a note off the position. Set the knob to the centre off the
value in the .SED file if you are not on target. It is 320 steps / two positions. Repeat the procedure above for the right knob.
mm. Reduce the laser power if the peaks are higher than 150.
Higher motor offset value => probe moves towards the Repeat the procedure a few times. Small adjustment (a 1/4 of
door a turn) might be needed to get the best sensitivity. The laser is
Lower motor offset value => probe moves towards the now focused.
home position
When a sample is run set the sample target region to 0% and 4.1.5 Leaking valve
100%. Peaks should be visible in both position.
The laser optics is focused during the manufacturing. It needs There are three valves in the Maldi: Air admit valve fitted on
normally only some small adjustments upon installation and the small turbo pump. The function of this valve is to vent the
service. The only time this focusing has to be done is when system when the turbo pumps are of. The valve is opened
the laser is replaced or when the laser bench has been re- when the turbo speed is approximately 50% off max speed.
moved. The valve is controlled by the turbo pump controller. To find
To get easy access to the laser bench and its adjustment the out if the valve is leaking remove the tubing from the valve.
front corner panel with the three leds and the side panel to the Block the inlet of the valve where the tubing was connected
left of the panel with the three leds has to be removed. To with your finger or a swagelok plugg. If the vacuum reading
reach the adjustments of the transparent mirror it is enough to impoves on the magnetron the valve is leaking and needs to
remove the front corner panel with the three leds. Place a be replaced.
sample slide with BIC ink (or some ink from a marker pen that Door
is known to give good peaks) on some off the sample spots
and or matrix on some of the other spots.
Start a run with the Ink method for acquisition. Make sure the N2 or
laser beam hit the lens in the center by fitting the laser focus- dry air Vent valve Inlet valve
ing tool in front of the lens. Adjust the knurled nuts on the
mirror to until the laser spot in the centre hole of the laser Sample inlet valve assembly
focusing tool. Remove the focusing tool. Increase laser power
until ink peaks are visible. Turn the left knurled knob counter
clockwise until the peaks are gone. Turn the knob clockwise
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
it can be seen as the pirani reading goes up when the door is
opened. Normaly the reading should improve (drop) or be the
N2 or dry air same when the sample door is in the opened possition.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
part of the covers loop, it has to be closed if the HV should be on top of the rotary pump and one Active Inverted Magnetron
enabled. If the HV can`t be turned on check this micro switch. (Penning) gauge fitted next to the pulser box. The only re-
It has to be closed by the door mechanism. maining active device is the linear detector.
needed on the knurled nuts to get highest sensitivity. Start a The Linear detector (ETP) is fitted on the inside of the top
run and check if further adjustment is necessary. Make sure flange of the reflectron.
that peaks are present when the adjust target region is set to
0 and 100% (5 and 95%). This is to make sure that the laser The linear amplifier box sits on top of the reflectron chamber
is able to fire on the hole sample. directly connected to the linear detector output.
The reflectron detector that is fitted inside the vacuum system
4.3 Electronics just under the reflectron stack.
This instrument contains dangerous high voltages.
The electronics is controlled by a pc with pcimca pcb. There
Maximum voltage is 20000 V.
are three switches on the instrument. Two main switches (one
Be careful when the panels are removed.
on the back of the instrument and one on the back of the elec-
The electronics in the Maldi is fitted in to a case that contains tronic case). The third switch is a safety interlock switch for
six pcb. The pcb`s are: the high voltages. It has to be pulled out if the high volt-
Low voltage power supply pcb ages should be on when the panels are removed.
Acquisition control pcb On next page is a block diagram of the electronics showing
Transient recorder pcb how the different electronic units are connected with each
24 V power supply pcb other.
High voltage control pcb
4.3.1 HV to the source
On top of the case are five HT power supply fitted. They are The function of the source voltage is to accelerate the ions
for the source, reflectron stack, pulser, reflectron detector and from the source in to the flight tube.
linear detector. The HV to the source comes from one of HV supply on top of
To the right hand side of the source is the pulser located. Just the electronic case and it is controlled by the high voltage
above the laser is a small aluminium box that contains the control pcb. The source voltage is set by the software. The HV
beam blank drive. On the right side of the reflectron chamber could be checked on pin 3 on contact M45 and ground with a
is an aluminium box for the gate pulse circuit and the kicker. DMM (0-10 V where 10V = 20 kV). The source voltage could
There are also two turbo pump controllers fitted on the back of also be checked at the HV lead that goes in to the pulser box
the instrument. Finally there are also one Pirani gauge fitted with a HV probe and a DMM. It should be 18 kV dc in the 20
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
COMPUTER
ACQUISITION CONTROL WITH
IO * GAS CELL CONTROL (AC3) COMMS PCB
* SAMPLE MOTOR (IO1) * GATE CONTROL (AC4)
POWER SUPPLY * IRIS MOTOR (IO9) HIGH VOLTAGE CONTROL
* DATA ACCUMULATION
BOARD * SAMPLE PORT MOTOR (IO10) * HP20R - SOURCE (HVC2)
* EXTERNAL COMMS (AC5)
* VALVES (IO5,IO6,IO7) TRANSIENT RECORDER 24V SUPPLY * HP20R - REFLECTRON END (HVC1)
* LASER TRIGGER (AC2)
+5V LOGIC * TURBO PUMP CONTROLLER INTERFACE (IO4) 1GSPS SWITCH OFF FOR ZERO VOLTS
* PULSER TRIGGER (AC1) OPTIONAL
+24V IO * PIRANI GAUGE CONTROL (IO3) INPUT SIGNALS * LINEAR OPERATION * H5P - DETECTOR (HVC7)
* LASER DETECT (AC7) 2GSPS
+/- 5V T.R. * MAGNETRON GAUGE CONTROL (IO2) * REQUIRED CURRNET 2.4A * HP5R - PULSER (HVC4)
* FAST TRANSIENT RECORDER. TRANSIENT
+/- 5V HEAD AMP * HIGH VOLTAGE CONTROL INTERFACE * LINEAR DETECTOR (TR2) * HP2.5P - REFLECTRON DET 2KV (HV10)
TRIGGER (AC8) RECORDER
* COVERS LOOP * REFLECTRON DETECTOR (TR1) * PULSED EXTRACTION UNIT (HVC4)
* BEAM BLANKING CONTROL * GATE DRIVER (HVC6)
* AUTO-SAMPLER / Y AXIS ON ADDTIONAL BOARD * LASER POWER SUPPLY (HVC5)
* GAS CELL DRIVE * BEAM BLANKING (HVC3) TR2 AC8 TR1
IO11 AC1
PS1 PS2 IO2 IO3 IO4 IO5 IO6 IO7 AC8 AC6 TR2 HVC2 HVC4 HVC5 HVC6 HVC3
HVC1 HV7
HV10
IO9 AC2
AUTO
SAMPLER AC7 LIN:0V (OFF)
OPTION SIG?? REFLECTRON
DETECTOR REFLECTRON DETECTOR
IO1 COVERS
UV LASER, OPTICS, IRIS AC4
HP2.5P 4mA HP20R HP5P
(plus port switch)
HVC4
MOTOR & LIMIT SWITCHES
AC3
+2.5KV REF :+/-20KV +2KV
IO10
FIBRE OPTIC & PULSE DETECT
HP5R
SAMPLE MOTOR +/-4KV AC1
HVC2
HVC6
PULSED
SOURCE EXTRACTION UNIT
HP20R -/+4KV HV3 2KV
CONICAL DUAL
ANODE MCP
REFLECTRON AMPLIFIER
60 PLATES ELECTRON 5nF
WITH
INCLUDING ENDS MULTIPLIER OFFSET
INLET CHAMBER
ADJUST
GATE ION FLIGHT TUBE
SAMPLE PULSED BEAM GAS +/-500V KICKER
LENS TUBE TUBE
+/-20KV LENS BLANK CELL OR 0V +/-90V
RESISTIVE DIVIDER. (67M)
EDWARDS
MAGNETRON EDWARDS E DWAR DS MAINS BLOCK
IO2
GAUGE TURBO TUR BO (6 OUTPUTS)
AIM-S-MW25 PUMP EXT70 P U MP E X T 70 * TURBO PUMP CONTROLLER(s)
BACKING VALVE (24V,N.O.) IO7 * TCU
* FAN TRAY
* ROTARY PUMP
INLET * BACK PANEL FANS
VALVE
24V N.C.
IO5
PORT ON
EDWARDS
SAMPLE
PIRAINI GAUGE IO4 IO4
PROBE
APG-M-NW16 ROTARY LAST CHANGE : 09/07/99 17:35
VENT PUMP
VALVE IO3
SJ OR IG INAL XX/98 A
24V N.C. EDWARDS TURBO PUMP EDWARDS TURBO PUMP BY MODIF IC AT ION DAT E IS S
CONTROLLER CONTROLLER DR G NO : S AI-02B D S HE E T 1 O F 1
IO6 EXC100L EXC100L
LT3-P 1 E LE C TR IC AL S YS TE M DIAG .
S C IE NT IF IC ANAL Y S IS INS T R UME NT S L T D
S T R IC T LY C ONF IDE NT IAL C OP Y R IG HT S AI LT D 1997
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
kV reflectron mode, 9 kV in the 10 kV reflectron mode and 4 more than this it will cause variation in flight time from one
kV in the 5 kV reflectron mode. In linear mode the set voltage shot to another which will result in broader peaks i.e. bad
is equal to the source voltage i.e. 20 kV linear mode sets 20 resolution.
kV to the source. The HV is connected to the extraction lens
and to the sample probe via the pulser box. 4.3.3 Pulse extraction signal
The pulser box is located on the right hand side of the ion
4.3.2 Laser detect signal source in an aluminium box.
The laser is used to increase to energy level of the sample !!!! The lid is floating at 20 kV if it is not grounded when
molecules, so they vaporize and form ions. A fraction of the the box is open and the HV is on. The HV might be on
laser beam is reflected down to a photo transistor by the filter even if the instrument is not in operate. Switch of the
wheel. This is the laser detect signal. The laser detect pcb is
A B C
fitted in a aluminium box under the laser. The laser detect
signal goes in to the electronic case via M34. The trigger level
is adjusted on the acq control pcb by rv1 and measured on
TP6, TP7 is ground. It should be set to trigger on the steepest
flank 0.8-1.2V (A). Avoid an area where it is some ringing on
the signal (B) because that could cause some jitter on the
peaks. On the oscilloscope picture to the right where Ch1 is
laser detect signal on M34. Ch3 is the clock pulse on M35.
The clock pulse is generated on the acquisition pcb by the
laser detect signal and IC 29, 30, 31. This pulse is the start
signal for the hole system. It trigger the extraction pulse, start
the delay for time for the beam dump and the gate. On the
oscilloscope picture below C indicates the time different be-
tween A: laser detect signal and B: clock signal. Since the
clock signal is time zero i.e. starts the acceleration pulse this
is the time deferents between the laser shot and the extraction
puls. This is the delay time that also is called focus mass in
the acquisition method. The variation in time between the two
pulses should not be more that a few nano seconds, if it is
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
electronic rack before open the pulser box. Make sure the
ground wire between the pulser lid and the chassie
ground is connected .
The function of the pulser is to extract the ions from the ion
source and get the ions of the same mass focused when they
reach the detectors. This is achieved by keeping the ions in a
constant electric field between the source block and the ex-
traction lens after the laser has been fired. After a well con-
trolled time delay (focus mass) the extraction lens voltage is
dropped and the ions leaves the source and enter the flight
tube. In 20 kV reflectron mode the extractor drops about 3kV
from 18 kV to 15 kV. To check the pulser do as follows. Switch
of the electronics. Open the pulser box, connect a ground
wire between the lid and chassis ground. Connect a HV probe
to the junction of R2 and C1 in the top left corner of the pcb.
Switch on the electronics and start an acquisition. Ch1 is the clock /trigger pulse at M35 on the back of the electronic box,
It is very important to have no time different from one shot to connected to the scope with a coaxial cable. Ch2 uses a HV probe
the next, if the time difference is more than one nano second (1000:1) and show the extraction pulse. It is 3 kV in the 20 kV mode.
it will have negative influence on the resolution.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
correct then the pulser box is faulty or the cable from J8 to the
pulser box.
Make sure SI21 has been performed. This will prevent that R3
burns and damage the pcb.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
TP1 is the testpoint for the 24 V. The 24 V is converted on the
pcb to 5 V, 15 V, -15V, 500 V and -500 V. The 500 V and the -
500 V can be checked on TP 4 and TP 5. The 5V can be
monitored at IC 7 pins 3 & 6 (middle pins on the 3 pins
grouped together). The 15 V can be checked on ic1 pin 3 and
the -15 V can be checked on ic9 pin 3. Use TP 2 as ground .
The kicker tilts the ion beam a little bit in the reflectron mode
so the beam will hit the MCP detector at the end of the reflec-
tron. It applies approximately +/-50 volts to the kicker elec-
trodes to give the ions a parabolic curve in the reflector. This Ch 2 is the negative output (TP 7) ch 3 is the positve gate
is to make sure the low mass ions hits the reflectron detector puls (TP 6), use TP 8 as ground.
in psd mod. The adjustment of the kicker should only be done To verify the function of the Gate and kicker electronics the
in psd mod to optimize the intensity of low mass fragment. first thing to measure is the voltage on the pcb as described
The supply voltage is 24 V and comes in on J1 pin 1,3 and 6. above.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
To test that the gate is opening and closing as it should an
oscilloscope is necessary. See picture below.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
If the gate puls is on all the time (constant 400 V to gate elec-
trode) no signal at all will pass the gate electrodes. This will
results in no peaks at all in reflectron mode, the linear peaks
will be there as normal since the gate and kicker electronics is
disabled in linear mode.
If all the signal is there as described above but it is still prob-
lem with the gate or kicker make sure the electrode inside the
Maldi is not shorted to ground or has come lose. To find out if
the electrodes are shorted do as follow: Disconnect the four
pins on the gate and kicker feedthrough, make a drawing of
how they are connected first. Check the resistance between
each pin and ground and between each other, there should
not be any continuity anywhere. To verify if the electrodes has
been disconnected by some reason the system has to be
vented, the refletron has to be removed and a inspection must
be done from the inside. See Chapter 6 paragraph 6.4 for
instructions
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
supply, 0,8*2.5kV= 2kV. If there is no hv, check the 24V on pin electrodes. This is to make sure nothing will block the ions
X in the connector for the HV supply. If the 24V is missing it is from hitting the detector. Check the detector supply voltage is
a problem with the cable or the backplane pcb since the 2000 V.
peaks are normal in reflectron mode all HV for source, both
detectors are enabled at the same time. Replace/repair any
faulty cables, pcb or power supply that is found.
The linear amplifier has a gain off 1 and works like a buffer to
protect the T/R pcb. The system can be used without the lin-
ear amplifier for a test run. Remove the linear amp and con-
nect the bnc cable direct to the output from the linear detector.
If the peaks are back fit a new amplifier.
If the peaks still are missing use a oscilloscope to see if there
is any signal coming out from the linear detector. See picture
on next side. Replace the T/R pcb. If there still are no peaks in
linear mode replace the linear detector and the linear detector
pcb. If there still are no peaks in linear mode contact your
local Maldi service specialist for advice.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
Remove the grid holder by loosen the three screws (B). Re-
move the mcp by loosen the three nylon screws (C) that holds
the mcp clamp in place. Replace the MCP assy. Make sure B
the index mark on the MCPs are turned 180 degrees in rela-
tion to each other. Do not forget the spacer between the
MCPs. Tighten the nylon screws in a criscross pattern. Do not
over tighten the nylon screws, the MCP will brake very easily.
Verify that the height of the assy is correct at each set screw
according to picture to the right. Adjust the set screws to the
correct height if needed. This should be done with adjustment
pins (F) in the middle position.
Refit the grid holder. Make shure the grid is grounded by a
wire connected to on of the screws.
Reassemble the reflectron in reverse order. When the Maldi
has been reassembled, start the vacuum pumps. Wait until C F
the vacuum reaches 1e-6 mbar or better on the magnetron,
start to ramp the HT for the MCP from 0V up to 2000 V by
following the procedure below.
Apply voltage in 100V steps, wait 2 minutes. Repeat until the
voltage is 1.0 kV. Wait 2 minutes.
Adjust the voltage in 100V steps, wait 5 minutes. Repeat until 14.4 mm +/-0.1
the voltage is 1.5 kV. Wait 5 minutes.
Adjust the voltage at 50V steps, wait 10 minutes. Repeat until
the voltage is 2 kV.
The adjustments of HT is done on the trim pot Vi in on the
MCP HT supply. The voltage is checked at the readback at pin
13.8 mm +/-0.1 14.4 mm +/-0.1
three in the contact to the MCP HT supply.
Link to replacement of old detector
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
4.3.11 High Voltage problem/Arcing this. A dmm can also be used but in some cases the dmm will
not be fast enough to detect the voltage dip.
Arcing in the Maldi is when the high voltage (HV) discharges
to ground. This can be caused by dust, hair or other particles
4.3.11.1 Measure reflectron resistance and all HV feed
that has come in to the vacuum system and shorts the high
through
voltage to ground. Dust and similar things can come into the
Remove the HV cable on the top of the reflector housing for
system with the sample slides or when the system has been
the reflector HV. Measure the resistance between the center
vented or opened for some repair. If the system is suspected
pin in the feedthrough (M131) and ground with a megger. It
to be dusty it has to be vented and the dust, hair has to be
should be between 67,2 MΩ Ω - 68,4 MΩ. If the resistance is not
removed. If the source is dirty the source housing has to be
within this range, the Maldi has to be opened and the fault
opened and the dirt can be blown away with nitrogen or wiped
must be corrected. Most dmm only works up to 40 MΩ and
of with some methanol on a lint free tissue. If the problem is in
does not use 500V as a test voltage as a megger does.
the reflectron stack it has to be blown of with nitrogen or some
Check the resistance in the bnc feedthrough with a DMM for
other pure gas.
the linear and reflectron detectors. The resistance between HT
The arcing can also take place in the feedthrough for the HV.
(M130) for the linear detector and signal output (M132) should
It is caused by some pore in the plastic that the feedthrough is
be > 40 MΩ, between HT (M130) and ground >40 MΩ, be-
made of. The feedthrough can be OK for quit long times and
tween output (M132) and ground > 40 MΩ.
then suddenly the high voltage brakes through the plastic and
It should be >40 MΩ between HT (M80) and output (M129), 11
shorts to ground. The feed through has to be replaced if it is
MΩ between HT (M80) and ground. If the resistance is lower
causing arcing.
the fault has to be corrected.
The effect off arcing can be noticed as reduced resolution
since the arcing will cause the HV to vary from one shot to
4.3.11.2 HV Power supplies
another. This will give variation in flight time and a broadening
The arcing problem can also be with the HV power supplies.
of the peaks i.e. bad resolution when they are accumulated. In
The source power supply and the reflector power supply are
real bad cases no peaks at all will be present due to that the
identical and can swapped with each other for trouble shoot-
HV is shorted to ground. ing. What seems to be a problem with the HV supply can also
In some cases the arcing problem can be found by measuring be a problem with the 24V supply for the HV or the connection
the readback voltage on pin 3 in the connector of each HV to the HV supply. The 24V must be stable and free from ripple.
connector. Pin 3 is connected to the HV output by a internal If the there is ripple on the 24 V supply, it will give the same
resistor network that divides the voltage to a value from 0-10 kind of problem as arcing. See next page for a faulty 24 V
V where 10 V is full output. It is best to use an oscilloscope for power supply.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
4.3.12 Electronics measurement and calibration
I/O Board
Test point
TP7 & TP9
TP14 & TP12
TP14 & TP13
TP14 & TP15
TP14 & TP10
TP16 & TP17
This is a scope picture of a the ripple on a faulty 24 V power TP14 & TP12 24 24
supply. When the voltage was measured with a dmm the dc
level was 24.0 V and the ac 0. The 24 V is connected to ch1 TP14 & TP15 12 12
with a 10:1 probe. The input is ac coupled. The scope is set to
trigger on ch1. The ripple is about 700mv p-p. The normal TP14 & TP13 5 5
value is about 20mV.
TP14 & RN6 20 20 RN6 pin 1 (mitered corner)
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
4.5 Fault finding -Is it somenoise or random events in the spectrum? If so,
most likely the detector, amplifierand TR pcb is ok.
4.5.1 No peaks or low sesitivity -Run the instrument in linear mode 5kV. Ground the extraction
lens with a pice of wire, the source (sample slide assembly)
-The first thing to check is your method: Acceleration voltage should be connected directly to the source HT supply. This
= 20kV, Low mass reject = 500, single shot, focus mass = will bypass the pulser box. If younow see some peaks at a
900 or 1500 depending on what sample you are running, higher mass and whit lower resolution than normal it is most
accumulate=20 shots, manualy tune the laser power and likely that the pulser or pulser HT is faulty.
sample possitioning to achive max signal. -If it is still no peaks, disconnect the beam dump bnc contact
-Is the vacuum ok, oil level in rotary pump, rotary pump is on the flight tube. Make a run, if the peaks is back replace
runninng, turbo speed ok? the beam dump pcb.
-Is there a sample slide white a sample in the source? -The next thing to check is the gate box. Disconnect gate +/-
-Check all the voltages in the electronic box? in the box, is the peak back? The gate is on all the time and
-Is the HV for the source, pulser, reflector and detector ok? blocking the beam. Replace the gate pcb and tune it up (se
-Do you get HV to the sample probe? procedure on page 109-110.
-Does the laser fire? Check the 24 V and trigger pulse to the -If the peaks is still missing it could be a detector problem: In
laser. The Acqusition control pcb controlles the laser trigger linear mode: Check the bnc conntacts and coaxcial cables.
puls. Check the resistance in the linear detector. It should be >999
-Is the shutter on the laser open? Mohm between HT and output, 35 Mohm betweenHT and
-Is the laser power enough?Turn up the settings in software ground. If the recistance is lower and the peaks is missing
and on the laser it self if nessesary. It is a trim pot in a small the detector needs to be replaced.
hole on top of the laser. Is the laser focused on the correct In reflectron mode: Check the bnc conntacts and coaxcial
sample spot ( se page 120)? cables. Check resistance between HT M121 and output, it
- Is the sample in the right place? Check offset for sample should be > 40 Mohm, >40 Mohm between output M80 and
slide and the focus for the laser. ground, >40 Mohm between HT M121 and ground. If the
-If the peaksare missing in reflectron mode, are they ok in recistance is low and the signal weak, replace the detectore.
linear mode or? If so, this indicates that the ion source is ok
and it could be a detector, amplifier or TR pcb problem.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
resistance in the reflector with a megger. It should be 67.3
4.5.3 Low resolution Mohm between HV input for the reflector and ground. The
The first thing to check is the focus mass in your method, it megger normaly uses 500 V to check resistance which is
should be set to the same mass as you check the resolution not so much compared with the 20 kV the reflectron could
at. be operated at. The reflectron H.T could be checked by
The low resolution could also be caused by the sample. meassure the voltage at pin 3 on M46. If the voltage is not
Prepare a new slide whit new matrix and sample. Make constant it´s an indication of arcing in the reflectron. This
shure all the solvets, pipetts and other things used for sample has to be corrected to get a good resolution. So it might
preparation is clean. Check the resolution according to be nessesary to open the reflectron chamber and look for
the specifications using hACT whit 3 pmol concentration at some dark spots which are sign of arcing. The first place to
mass 2465. look for this is around the spring contact in the lid.
Do not trust unknown samples from customer. Low resolution could also be caused by the detector in
Do not use to much laser energy. It is no exact figur for this booth linear mode and reflectron mode. In linear mode
but try to keep the laser energy down so you get a signal of check the possition of the detector and that it is not lose.
about 60 -100 in each scan. Also check the resistance of the detector (page 18). The
Check that the laser detect signal is triggering at the right mcp detector for the reflectron mode should be firmly fit
level, se 4.3.2 for a correct set up procedure. but not overtighend. Lose mcp ore aligned the wrong way
Check the pulser extraction signal, the ratio between the will give low resolution (se page 27 for correct orientation
source voltage and the extraction signal should be set up and fitting). Check the signal cable from the detector to the
correct for a good resolution. If the peaks are very broad bnc feed through, if the cable and the contacts are not
and in the wrong place it could be that the pulser is not correctly
working at all. Se 4.3.3 for a correct setting of the extraction fitted it could be a impedance missmatch and that will
puls. ruin the resolution. This is also some thing to check in the
Check the settings in the DefEpt.SED file according to coaxial cable from the detector to the electronic box and
procedure on page 20. the transient recorder pcb.
Low resolution could also be caused by incorrect reflectron Resolution problems could also be caused by the electronics.
voltage or arcing in the reflectron. Check the voltage on the It could bea problem on thetransient recorder
reflectron power supply (pin 3 on the contact M46) it´s the pcb,aquisition control pcb. The electronics on these pcb
readback voltage from reflectron H.T. and is 0-10 V where are very fast and difficult to trouble shoot so it is recomended
10 V is the max output from the power supply. Check the to replace them.
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
4.4 Sample probe offset adjustment and laser beam remove the front corner panel with the three leds. Place a
targeting sample slide with BIC ink (or some ink from a marker pen that
ALWAYS USE LASER SAFETY GLASSES WHEN WORK- is known to give good peaks) on some off the sample spots
ING WITH THE LASER. and or matrix on some of the other spots.
Prepare a slide with marker pen so it easy to identify the sam- Start a run with the Ink method for acquisition. Make sure the
ple spots inside the source. Mark for instance spot 1-3 and laser beam hit the lens in the center by fitting the laser focus-
22-24, it easier to identify the spots if different colours are ing tool in front of the lens. Adjust the knurled nuts on the
used. The source can be illuminated by a torch. Shine the mirror to until the laser spot in the centre hole of the laser
light through the laser mirror and lens to illuminate the sample focusing tool. Remove the focusing tool. Increase laser power
slide inside the source. Load the sample side and start a run until ink peaks are visible. Turn the left knurled knob counter
at position 1. Block the laser beam with a business card or clockwise until the peaks are gone. Turn the knob clockwise
similar after the filter wheel assembly. Observe the sample until the peaks are back. Make a not off the knobs position.
slide through the laser mirror, adjust the sample motor offset Continue to turn the knob clockwise until the peaks are gone.
value in the .SED file if you are not on target. It is 320 steps / Make a note off the position. Set the knob to the centre off the
mm. two positions. Repeat the procedure above for the right knob.
Higher motor offset value => probe moves towards the Reduce the laser power if the peaks are higher than 150.
door Repeat the procedure above a few times. The laser is now
Lower motor offset value => probe moves towards the focused. Some small adjustment (a 1/4 of a turn) might be
home position needed on the knurled nuts to get highest sensitivity. Start a
When a sample is run set the sample target region to 0% and run and check if further adjustment is necessary. Make sure
100%. Peaks should be visible in both position. that peaks are present when the adjust target region is set to
The laser optics is focused during the manufacturing. It needs 0 and 100% (5 and 95%). This is to make sure that the laser
normally only some small adjustments upon installation and is able to fire on the hole sample.
service. The only time this focusing has to be done is when
the laser is replaced or when the laser bench has been re-
moved.
To get easy access to the laser bench and its adjustment the
front corner panel with the three leds and the side panel to the
left of the panel with the three leds has to be removed. To
reach the adjustments of the transparent mirror it is enough to
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
Sequence of pulses and signals 4.5 Software and PC
There is no service information about software and pc for
the moment.
4.6 Error code list
Laser 70ns (when delay is set to 0. Will increase the same amount as delay)
There is no error code list for the moment.
4.7 Service equipment
Oscill. M35 starts at the same time as Pulser signal below Multimeter
High voltage probe 20kV for multimeter
Megger
TTL to pulser Depends on selected mass in ion reject box Oscilliscope 500 MHz (portable from: Tektronix TDS3052 or
simmilar)
High voltage probe for oscilloscope 100:1 or 1000:1. 4kV
Beam dump or more (Tektronix P6015A)
Special tool kit supplied at service training.
Gate reflectron mode. Gate drops to zero the laser puls is detected.
UV light protection glasses.
Gate psd mode. Max 300ns delay, depend on selected mass for psd
Contents Chapter Service Manual Ettan Maldi ToF PRO MS 18-1160-04, Rev. March 2004
contents
Chapter 5
Chapter Contents
5. Maintenance
Maintenance
5.1 Vacuum system
5.2 Electronics
5.3 Mechanics
5.4.1 SOP
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
5.1 Vacuum system 5.2 Electronics
From time to time the vacuum lines and connections should Verify that all fans are working; five on the tray under the
be checked. electronic crate, one at the reflectron turbo pump and one to
the left of the H.T supply. Replace not working fans. Check
Once a year, the oil of the rotary pump should be changed, all power supply pcb (first from left): +5V, +24V, +/-(7V to
according the recommendations of the supplier. 12V) unregulated linear low noise supply for Transient Re-
The pirani gauge should be cleaned once a year. corder analog components,+/-(7V to 12V) unregulated linear
The two turbo pumps should only be changed in case of failur. low noise supply for Head amplifier. Also check the 24V on
the 24V power supply pcb (the fifth from left). Make sure
there is no ripple on any voltages.
Verify that the readback on the H.T supplies are ok. This is
done by meassure the readback voltage on pin three in the
contact to each H.T supply. The readback will be 0-10 V and
10V is full output.
Clean the fan filter under the ion source.
Adjustment of electronics.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
5.3 Mechanics 5.4 Preventive Maintenance
Maintenance for the mechanical system. Check the Motor Routine Service and Maintenance
offset by looking through the transparent mirror for the laser The Ettan Maldi ToF has been designed to require a minimum
beam. If the sample spot is out of alignment adjust the motor amount of routine service. Standard service schedules have
offset value in the .sed file until they all are in the right posi- been defined and are shown below along with the necessary
tion. instructions.
If the sample probe motor is noisy the bearing and o-rings for
the drive screw feed through need to be lubricated.
5.4.1 Standard operation procedure for the PM
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
PREVENTIVE MAINTENANCE VISIT RECORD
Customer: Date: ........... ............... ..............
Serial No: ..................................
Engineer:...................................................
Modification status: Accessories:
Measure the s/n for the m/z 1672.89 peak of 5 fmol Neurotensin .The result should be obtained
accumulating 200 shots per spectrum in 8-shots mode.
The resolution, FWHM (R) should be determed for the monoisotopic peak in the isotopic cluster
of 1.5 pmol Ile 7 Ang III, m/z 897,531. The result should be obtained accumulating 200 shots
per spectrum in 8-shots mode.
Run 5 pmol BSA accumulating 200 shots per spectrum in 8-shots mode. The monomer, m/z
66431, the dimer, m/z 132859 and the trimer, m/z 199287 should be detected in the spectra.
Measure the s/n for the m/z 1672.89 peak of 5 fmol Neurotensin .The result should be obtained
accumulating 200 shots per spectrum in 8-shots mode.
The resolution FWHM (R) for the monoisotopic peak in the isotopic cluster of 1.5 pmol hACTH
18-39 should be determend. The result should be obtained accumulating 200 shots per
spectrum in 8-shots mode.
Mass accuracy
Compare measured masses with theoretical, using 100-fmol iodoacetic acid modified
phosphorylase b tryptic digest.
Mass errors
total matched
<20ppm
First approved
spectrum
Second approved
spectrum
Compare measured masses with theoretical, using 100-fmol iodoacetic acid modified
phosphorylase B.
Mass errors
total matched
<100ppm
First approved
spectrum
Second approved
spectrum
The fragment masses of 60 fmol Hexatyrosine (m/z 997.396) should be detected and
annotated. In a spectra the following fragment ions S/N should be determed, b2=327.36;
b3=490.536; b4=653.712; b5=816.888.
The average mass error of the fragment masses, m/z 327.36, 490.536, 653.712 and 816.888 of
3 pmol Hexatyrosine should be determed.
Hexatyrosine B2=327.36 B3=490.536 B4=653.712 B5=816.888
S/N
FRAGMENT M/Z
ERROR
For 24 positions of a slide, containing 100, 50 and 25 fmol iodoacetic acid modified
Phosphorylase B and after accumulation of 1 spectra on each position, following criteria must
be fulfilled:
Acquisition time. Total acquisition time for all 24 spectra should not exceed 40 minutes.
Quality of identification result; Phosphorylase B must be identified as the highest ranked
candidate on all positions. 20 of 24 spectra must have an expectation below 0.01.
Table 2.
Spectrum Amount on target Expectation
(fmol) value
1 100
2
3
4
5
6
7
8
9 50
10
11
12
13
14
15
16
17 25
18
19
20
21
22
23
24
Total run tim:
Post pm test
This test should be performed after the pm. This test is the same as the installation test and the
specification is the same. This is to verify that Maldi fulfill all specification after the pm.
All methods used in these tests are described in the installation guide. The preparation of
samples for the tests is described in the installation guide.
Acceptance limit: S/N > 3 for the m/z 1672.89 peak of 5 fmol Neurotensin, accumulating 200
shots per spectrum in 8-shots mode.
Acceptance limit: A resolution ≥1000, FWHM (R) should be achieved for the monoisotopic peak
in the isotopic cluster of 1.5 pmol Ile 7 Ang III, m/z 897,531. The result should be obtained
accumulating 200 shots per spectrum in 8-shots mode.
The monomer, m/z 66431, the dimer, m/z 132859 and the trimer, m/z 199287 of 5 pmol BSA
should be detected in the spectra. The result should be obtained accumulating 200 shots per
spectrum in 8-shots mode.
Acceptance limit: S/N> 3 for the m/z 1672.89 peak of 5 fmol Neurotensin .The result should be
obtained accumulating 200 shots per spectrum in 8-shots mode.
Acceptance limit: Resolution> 10000, FWHM (R) for the monoisotopic peak in the isotopic
cluster of 1.5 pmol hACTH 18-39. The result should be obtained accumulating 200 shots per
spectrum in 8-shots mode.
Mass accuracy
Compare measured masses with theoretical, using 100-fmol iodoacetic acid modified
phosphorylase b tryptic digest.
Acceptance limit: More than 15 masses must be assigned to Phosphorylase B and for 40% of
the assigned masses, the error should be equal or less than 20 ppm, in the mass range 900-
2500 m/z.
Mass errors
total matched
<20ppm
First approved
spectrum
Second approved
spectrum
Compare measured masses with theoretical, using 100-fmol iodoacetic acid modified
phosphorylase B.
Acceptance limit: More than 15 masses must be assigned to Phosphorylase B tryptic digest
and for 30% of the masses, the error should be equal or less than 100 ppm.
Mass errors
total matched
<100ppm
First approved
spectrum
Second approved
spectrum
The fragment masses of 60 fmol Hexatyrosine (m/z 997.396) should be detected and
annotated. Acceptance limit: In a spectra the following fragment ions must be detected with
S/N≥ 3, b2=327.36; b3=490.536; b4=653.712; b5=816.888.
PSD mass accuracy test
Acceptance limit: The average mass error of the fragment masses, m/z 327.36, 490.536,
653.712 and 816.888 of 3 pmol Hexatyrosine must be below +/- 1 Da and the mass error of one
individual fragment peak must not exceed1.5 Da.
For 24 positions of a slide, containing 100, 50 and 25 fmol iodoacetic acid modified
Phosphorylase B and after accumulation of 1 spectra on each position, following criteria must
be fulfilled:
Acquisition time. Total acquisition time for all 24 spectra should not exceed 40 minutes.
Quality of identification result; Phosphorylase B must be identified as the highest ranked
candidate on all positions. 20 of 24 spectra must have an expectation below 0.01.
Table 2.
Spectrum Amount on target Expectation
(fmol) value
1 100
2
3
4
5
6
7
8
9 50
10
11
12
13
14
15
16
17 25
18
19
20
21
22
23
24
Total run tim:
Chapter 6
Chapter Contents
6. Disassembling
Disassembling
6.1 Removing panels
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.1 Removing panels
The first panel to remove is the curved front panel with the
blue LED. It is hold in place by a screw in the lower left
corner.
The second panel is the front of the ”chimney”. It is secured
by the curved panel with the LED, two screws at the top.
Also the lid of the chimney has to be removed before the
front of the chimney can be removed. When this has been
removed, it is only to lift the front panel up and remove it.
The panel is hanging on some screw heads that is fitted in
the chassie of the maldi.
The third panel to remove is the right front panel. It is hold
in place by some screws that is hidden under the front of
the ”chimney”.
The fourth panel to remove is the back of the ”chimney”.
Losen the vent tube hold in place by some tie wrap then it is
only to lift the panel upwards.
The fifth panel is the top that cover all HT supplies. This
panel is just to lift it of the maldi.
The sixth panel to remove is the back panel. It`s secured by
al allen screw from the top. When the allen screw has ben
removed, lift the panel up and forward to remov it.
To remove the last panel which covers the sides, losen
three screws at each side and slide the wrap a round
panels off.
To refit the panels the order is reversed from above and it
requires a black belt in origami to get them in place.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.1 Vacuum pumps If a separate 24 V power supply is not avilabel the system
has to be vented by switching of the main switch on the
When doing service on a vacuum system you should
right side of the instrument. The system should be vented to
always use clean dustfree rubber gloves so you do not
dry N2 and not to air. This is to reduce the pump down time
leave finger prints inside the vacuum system. When the
after venting the Maldi. When the rotary pump is back in
system is opend keep all parts and openings coverd whit
place it should be switched on for one minute before the 24
alumina foil to keep the inside as clean as possible. Small
V is removed from the isolation valve. It could be a good
dust particles and hair could cause discharges and arcing
idea to replace/service the pirani gauge at the same time as
inside a vacuum camber when voltages up to 20 kV is
the rotary pump. It is fitted on top of the rotary pump and
applied to the different parts inside.
hold in place by a clamp ring. The only service that can be
All connections between the pumps are made by stainless done is to replace the oil inthe rotary pump. In the turbo
steel bellow tubing. The bellow tubing is connected to the pumps there are no service to be done.
pump by a stainless steel clamp ring and the sealing is
The system has to be vented if the turbo pumps should be
done whit a viton o-ring.
removed. Each turbo pump is hold in place by four clamps
When reassembly a vacuum system, make shure that and four screws. Remove all contacts and valves from the
all o-rings and sealing surfaces are clean and dust free. turbo pump, losen the screws that holds the turbo pump in
They could be cleaned whit metanol and a lint free place a 1/4 of a turn in a crisscross pattern, repeat until the
tissue. pump is lose. Remove the pump. When putting back the
The first thing to do when removing the rotary vacuum pump it is important to make shure to get the o-ring in the
pump is to separate the pump from the rest of the vacuum right place and thighten the screws that holds the pump in
system by activating the isolation valve with a external 24 V place in a crisscross pattern and 1/4 of a turn each time.
supply. It is done by disconnecting the normal contact and This is done to achive a good seal between the pump and
connect the 24 V supply to the pins on the valve. When the the vacuum chamber and not to bend the flange of the
valve is activated the rotary pump can be switched of and pump. To get acces to the source turbo pump it might be
removed from the system by losen the clamping ring and nessesary to remove the laser and the pulser from the
the screw under the Maldi (this screw is a transport screw instrument.
and should be removed when the system is installed). The The magnetron (penning) gauge should only be removed
system will keep the vacuum for about 15 minutes whitout when the system is vented. See in the magnetron manual
the rotary pump. The rotary vacuum pump can now be from Edwards on how to dissassembly the magnetron
removed and replaced/serviced whitout venting the system. gauge.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.3 Flight tube
There is no reason to open up the flight, because that is
only a empty tube. If you still has to open it contact your
local service specialist for advice.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.4 Detector system
- Went the system with pure N2.
-Disconnect Reflectron HT, linear amplifier and HT to the
linear detector (ETP) from the top of the instrument.
-LINEAR DETECTOR REPLACEMENT:Remove then top of
the Maldi by losen 6 allen screws. The ETP detector is fitted
inside the lid. Make shure you have the right crimp butts for
the connection of the detector. Remove the detector by losen
three small screws that holds the detector in place.
Reacemble the instrument in the reverse order. Be careful
with the H.T spring contact for the reflectron stack.
-REFLECTRON DETECTOR REPLACEMENT Old style
detector: Remove the turbopump from the reflectron chamber.
Disconnect M79 and M80 from the left hand side of the A
refletor. It is the connectors for the MCP (reflectron detector).
Loosen the four nuts (A) and lift the reflector gently uppwards
to remove it from the Maldi. Place the top part of the Maldi on
a clean bench. Cover the openings on the instrument with
alumina foil. Loosen the four allen screw (B) and remove the
alumina plate that holds the MCP, Gate electrode and Kicker
electrodes.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
Be careful with the eight ceramic distances when you lose
the (B) allen screw. Turn the alumina pice up side down.
Losen the three nylon screw (C). Remove (D). Make a note
of the mark on the left side of the MCP. Do not touch the
MCP with your fingers, use tweezers one the edges of the
MCP when the plates has to be moved. The mark on the
top one should be to the right and the bottom one to the left.
Be careful and keep the parts as clean and dustfree as
possible. The reassemble is in the reverse order. Be careful
when thigtening the nylon screews (C), thigtnen the
screews a 1/4 of a turn each time to achive a uniform
pressure on the detector plates. Do not over thigten the
screews, the MCP are very fragile and brake easily. Do not
expose the plates to air for a long time since the coating on
the detectors will oxidice, they will also absorbe humidity
from the air which have to evaporate from the plates before
the HT is applied to to the detectors in the vacuum
chamber. If not, the water in the tiny holes of the plate will
explode and ruin the MCP.
C D MCP
Link to replacement of the new detector.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.5 Slide Loader
See Slide Loader service manual.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.6 Laser and optics
The laser is fitted on a bench to the left of the ion source. It
is mounted with three allen screws (a) from the underside of
the bench. To remove the laser it is just to disconect the
contacts for 24V and laser trigger at the back of the laser
box and remove the two allen screw at the bottom.
The filter thats regulate the laser power is fitted just after the
laser and is hold in place by two screws (b) on the
underside of the bench were also the Laser detect box is
fitted. To remove the Filter assy a contact has to be
disconnected. There are no parts to service inside the laser,
it is replaced as a units.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.7 Pcb´s and power supply
The electronics are fitted in a alumina box on the back of
the instrument. To get acces to the box the back pannel has
to be removed. To get acces to the pcb´s a panel on the
alumina box has to be removed by lose 20 screws from it
and three bnc contacts has to be disconnected. Four nuts
and washer has to be removed from the bnc contacts
before the alumina panel can be removed.
The five H.T supplies are fitted on two rods on the top of the
electronic box. Each H.T power supply is fitted whit four
screws. All pcb´s and power supplies are not supposed to
be repaird on seit, they are replaced when needed.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
Chapter 7
Chapter Contents
7. Spare Parts
Spare Parts
7.1 Spare Parts
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
7.1 Spare part list
Instrument
18-1156-53
18-1156-54
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Manuals
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Computer and Software
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB 1
PSU 20KV HP20R/211 Refl.
18-1157-19
PSU 2.5KV HP 2.5N Refl.det.
18-1146-55
PSU 2.5KV HP 2.5P Linear detector
18-1172-39
PSU 2.5KV HP 2.5N Linear detector, used on early Maldi Pro
18-1146-55
PSU 5kV HP 5R/1M Pulser
18114653
PSU 20KV HP20R/212 Source
18-1157-18 REFLECTRON
PULSER
MCP
ETP
SOURCE
24 V Switched Power Supply board (PSU)
18-1146-57
Power supply PCB
18-1159-98
Transient recorder board
18-1159-94
Acqusition control board
18-1159-96
Main I/O board
18-1146-18
HV control board
C
NO
NC
18-1146-17
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB 2
Linear head Amplifier
18-1146-52
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Laser and optical system
LSI
M52
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Vacuum system main parts
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Vacuum system equipment
NW 10 elbow
18-1146-15
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Vacuum system equipment
Reflectron Assembly
Clamp bulkhead NW 16
18-1146-32
Pumping Kit Assembly
NW 25/10 reducer aluminium
18-1146-35
NW 10 x 750 tubing flexible
18-1146-36
Elbow NW 25
18-1146-37
Tee piece NW 25
18-1146-38
Clamp craw full ISO 63
18-1146-39
Blank NW 25
18-1146-40
NW 25 clamp ring ST ST
18-1146-42
NW 25 viton on AL carrier
18-1146-43
NW 16 viton on AL carrier
18-1146-49
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
O-Ring 44.2x3 FPM
O-rings in the area drive lens HSG Assy
18-1150-37 door o-ring
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
O-rings in the area reflectron assembly
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
O-Rings in the area Pumping Kit assembly
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Parts Drive/Lens HSg Assembly
Brass bushing
18-1159-99
The bushing is located under this cover, it is the bearing
and seal for the sample slide transport meccanism.
Drive shaft complete
18-1172-40 Drive shaft new design. See Si no 10 for more
info.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Detectors
ETP
18-1157-16
linear detector pcb
11-0002-79
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
High Voltage System
PSU HP 5R/1M
18-1146-53 E.U. 19-9001-02
PSU HP 2.5/1M
18-1146-54 E.U. 19-9001-03
PSU HP 2.5 N
18-1146-55 E.U. 19-9001-04
PSU HP20R/211 REFL
18-2257-19
PSU HP20R/212 SOURCE
18-2257-18
Feedthrough assy
18-1145-80
Feedthrough assy short pulser box
18-1145-69
Slim fan 230 V
18-1149-92
Slim fan 120 V
18-1160-00
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Switch Gear Assembly
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Electrical Parts
Fan 110V
18-1146-22
Fan 240V
18-1146-21
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Cables
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Valves
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PC Boards Switched PSU 24 V (TCU)
18-1146-57
E.U.
19-9001-06
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Power Supply
18-1159-98
E.U.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Backplane Assy
Backplane Board
18-1145-98
E.U.
19-9000-89
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Acqusition Control
18-1159-96
E.U.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB T/R Assy
18-1159-94
E.U.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB I/O Assy
Input-/Output - Board
18-1146-18
E.U.
19-9000-95
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB HV Assy
18-1146-17
E.U.
19-9000-94
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Pulser Assy IT 2-3
18-1145-71
E.U.
19-9000-84
Assembly contains:
Pulser board
Capacitor 5000 pF 20 kV
Front Cover
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Beam Blanking Assy
18-1146-50
E.U.
19-9000-99
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Gate Driver Assy
18-1146-51
E.U.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Linear Amplifier Assy
18-1146-52
E.U.
19-9001-00
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Assy PCI I/F
PCI Card
18-1146-56
E.U.
19-9001-05
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Laser Detect Assy
18-1145-72
E.U.
19-9000-85
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Sample slides and equipment
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Automatic Slide Loader
18-1147-63
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Instrument Service Miscellenous
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Instrument Cover
Top Cover
56-3104-14
Tube Panel
the numbers underneath are internal Umea Factory 56-3104-16
numbers
Fan Cover
56-3104-17
Upper Cover
56-3104-18
Side Panel
56-3104-20
Sample Inlet
56-3104-12
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Chapter 6
Chapter Contents
6. Disassembling
Disassembling
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.1 Removing panels
The first panel to remove is the curved front panel with the
blue LED. It is hold in place by a screw in the lower left
corner.
The second panel is the front of the ”chimney”. It is
secured by the curved panel with the LED, two screws at
the top. Also the lid of the chimney has to be removed
before the front of the chimney can be removed. When this
has been removed, it is only to lift the front panel up and
remove it. The panel is hanging on some screw heads that
is fitted in the chassie of the maldi.
The third panel to remove is the right front panel. It is hold
in place by some screws that is hidden under the front of
the ”chimney”.
The fourth panel to remove is the back of the ”chimney”.
Losen the vent tube hold in place by some tie wrap then it
is only to lift the panel upwards.
The fifth panel is the top that cover all HT supplies. This
panel is just to lift it of the maldi.
The sixth panel to remove is the back panel. It`s secured
by al allen screw from the top. When the allen screw has
ben removed, lift the panel up and forward to remov it.
To remove the last panel which covers the sides, losen
three screws at each side and slide the wrap a round
panels off.
To refit the panels the order is reversed from above and it
requires a black belt in origami to get them in place.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.1 Vacuum pumps If a separate 24 V power supply is not avilabel the system
has to be vented by switching of the main switch on the
When doing service on a vacuum system you should
right side of the instrument. The system should be vented
always use clean dustfree rubber gloves so you do not
to dry N2 and not to air. This is to reduce the pump down
leave finger prints inside the vacuum system. When the
time after venting the Maldi. When the rotary pump is back
system is opend keep all parts and openings coverd whit
in place it should be switched on for one minute before the
alumina foil to keep the inside as clean as possible. Small
24 V is removed from the isolation valve. It could be a
dust particles and hair could cause discharges and arcing
good idea to replace/service the pirani gauge at the same
inside a vacuum camber when voltages up to 20 kV is
time as the rotary pump. It is fitted on top of the rotary
applied to the different parts inside.
pump and hold in place by a clamp ring. The only service
All connections between the pumps are made by stainless that can be done is to replace the oil inthe rotary pump. In
steel bellow tubing. The bellow tubing is connected to the the turbo pumps there are no service to be done.
pump by a stainless steel clamp ring and the sealing is
The system has to be vented if the turbo pumps should be
done whit a viton o-ring.
removed. Each turbo pump is hold in place by four clamps
When reassembly a vacuum system, make shure that and four screws. Remove all contacts and valves from the
all o-rings and sealing surfaces are clean and dust turbo pump, losen the screws that holds the turbo pump in
free. They could be cleaned whit metanol and a lint place a 1/4 of a turn in a crisscross pattern, repeat until the
free tissue. pump is lose. Remove the pump. When putting back the
The first thing to do when removing the rotary vacuum pump it is important to make shure to get the o-ring in the
pump is to separate the pump from the rest of the vacuum right place and thighten the screws that holds the pump in
system by activating the isolation valve with a external 24 V place in a crisscross pattern and 1/4 of a turn each time.
supply. It is done by disconnecting the normal contact and This is done to achive a good seal between the pump and
connect the 24 V supply to the pins on the valve. When the the vacuum chamber and not to bend the flange of the
valve is activated the rotary pump can be switched of and pump. To get acces to the source turbo pump it might be
removed from the system by losen the clamping ring and nessesary to remove the laser and the pulser from the
the screw under the Maldi (this screw is a transport screw instrument.
and should be removed when the system is installed). The The magnetron (penning) gauge should only be removed
system will keep the vacuum for about 15 minutes whitout when the system is vented. See in the magnetron manual
the rotary pump. The rotary vacuum pump can now be from Edwards on how to dissassembly the magnetron
removed and replaced/serviced whitout venting the system. gauge.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.3 Flight tube
There is no reason to open up the flight, because that is
only a empty tube. If you still has to open it contact your
local service specialist for advice.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.4 Detector system
- Went the system with pure N2.
-Disconnect Reflectron HT, linear amplifier and HT to the
linear detector (ETP) from the top of the instrument.
-LINEAR DETECTOR REPLACEMENT:Remove then top of
the Maldi by losen 6 allen screws. The ETP detector is fitted
inside the lid. Make shure you have the right crimp butts for
the connection of the detector. Remove the detector by losen
three small screws that holds the detector in place.
Reacemble the instrument in the reverse order. Be careful
with the H.T spring contact for the reflectron stack.
-REFLECTRON DETECTOR REPLACEMENT Old style
detector: Remove the turbopump from the reflectron
chamber. Disconnect M79 and M80 from the left hand side of A
the refletor. It is the connectors for the MCP (reflectron
detector). Loosen the four nuts (A) and lift the reflector gently
uppwards to remove it from the Maldi. Place the top part of
the Maldi on a clean bench. Cover the openings on the
instrument with alumina foil. Loosen the four allen screw (B)
and remove the alumina plate that holds the MCP, Gate
electrode and Kicker electrodes.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
Be careful with the eight ceramic distances when you lose
the (B) allen screw. Turn the alumina pice up side down.
Losen the three nylon screw (C). Remove (D). Make a note
of the mark on the left side of the MCP. Do not touch the
MCP with your fingers, use tweezers one the edges of the
MCP when the plates has to be moved. The mark on the
top one should be to the right and the bottom one to the
left. Be careful and keep the parts as clean and dustfree as
possible. The reassemble is in the reverse order. Be
careful when thigtening the nylon screews (C), thigtnen the
screews a 1/4 of a turn each time to achive a uniform
pressure on the detector plates. Do not over thigten the
screews, the MCP are very fragile and brake easily. Do not
expose the plates to air for a long time since the coating on
the detectors will oxidice, they will also absorbe humidity
from the air which have to evaporate from the plates
before the HT is applied to to the detectors in the vacuum
chamber. If not, the water in the tiny holes of the plate will
explode and ruin the MCP.
C D MCP
Link to replacement of the new detector.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.5 Slide Loader
See Slide Loader service manual.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.6 Laser and optics
The laser is fitted on a bench to the left of the ion source.
It is mounted with three allen screws (a) from the underside
of the bench. To remove the laser it is just to disconect the
contacts for 24V and laser trigger at the back of the laser
box and remove the two allen screw at the bottom.
The filter thats regulate the laser power is fitted just after
the laser and is hold in place by two screws (b) on the
underside of the bench were also the Laser detect box is
fitted. To remove the Filter assy a contact has to be
disconnected. There are no parts to service inside the
laser, it is replaced as a units.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
6.7 Pcb´s and power supply
The electronics are fitted in a alumina box on the back of
the instrument. To get acces to the box the back pannel has
to be removed. To get acces to the pcb´s a panel on the
alumina box has to be removed by lose 20 screws from it
and three bnc contacts has to be disconnected. Four nuts
and washer has to be removed from the bnc contacts
before the alumina panel can be removed.
The five H.T supplies are fitted on two rods on the top of
the electronic box. Each H.T power supply is fitted whit
four screws. All pcb´s and power supplies are not
supposed to be repaird on seit, they are replaced when
needed.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev. March 2004
contents
Chapter 7
Chapter Contents
7. Spare Parts
Spare Parts
7.1 Spare Parts
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
7.1 Spare part list
Instrument
18-1156-53
18-1156-54
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Manuals
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Computer and Software
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB 1
PSU 20KV HP20R/211 Refl.
18-1157-19
PSU 2.5KV HP 2.5N Refl.det.
18-1146-55
PSU 2.5KV HP 2.5P Linear detector
18-1172-39
PSU 2.5KV HP 2.5N Linear detector, used on early Maldi Pro
18-1146-55
PSU 5kV HP 5R/1M Pulser
18114653
PSU 20KV HP20R/212 Source
18-1157-18 REFLECTRON
PULSER
MCP
ETP
SOURCE
24 V Switched Power Supply board (PSU)
18-1146-57
Power supply PCB
18-1159-98
Transient recorder board
18-1159-94
Acqusition control board
18-1159-96
Main I/O board
18-1146-18
HV control board
C
NO
NC
18-1146-17
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB 2
Linear head Amplifier
18-1146-52
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Laser and optical system
LSI
M52
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Vacuum system main parts
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Vacuum system equipment
NW 10 elbow
18-1146-15
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Vacuum system equipment
Reflectron Assembly
Clamp bulkhead NW 16
18-1146-32
Pump Kit Assembly
NW 25/10 reducer aluminium
18-1146-35
NW 10 x 750 tubing flexible
18-1146-36
Elbow NW 25
18-1146-37
Tee piece NW 25
18-1146-38
Clamp craw full ISO 63
18-1146-39
Blank NW 25
18-1146-40
NW 25 clamp ring ST ST
18-1146-42
NW 25 viton on AL carrier
18-1146-43
NW 16 viton on AL carrier
18-1146-49
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
O-Ring 44.2x3 FPM
O-rings in the drive lens HSG Assy
18-1150-37 door o-ring
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
O-rings in the reflectron assembly
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
O-Rings in the Pump Kit area
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Parts Drive/Lens HSg Assembly
Brass bushing
18-1159-99
The bushing is located under this cover, it is the bearing
and seal for the sample slide transport meccanism.
Drive shaft complete
18-1172-40 Drive shaft new design. See Si no 10 for more
info.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Detectors
ETP
18-1157-16
linear detector pcb
11-0002-79
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
High Voltage System
PSU HP 5R/1M
18-1146-53 E.U. 19-9001-02
PSU HP 2.5/1M
18-1146-54 E.U. 19-9001-03
PSU HP 2.5 N
18-1146-55 E.U. 19-9001-04
PSU HP20R/211 REFL
18-2257-19
PSU HP20R/212 SOURCE
18-2257-18
Feedthrough assy
18-1145-80
Feedthrough assy short pulser box
18-1145-69
Slim fan 230 V
18-1149-92
Slim fan 120 V
18-1160-00
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Switch Gear Assembly
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Electrical Parts
Fan 110V
18-1146-22
Fan 240V
18-1146-21
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Cables
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Valves
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PC Boards Switched PSU 24 V (TCU)
18-1146-57
E.U.
19-9001-06
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Power Supply
18-1159-98
E.U.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Backplane Assy
Backplane Board
18-1145-98
E.U.
19-9000-89
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Acqusition Control
18-1159-96
E.U.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB T/R Assy
18-1159-94
E.U.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB I/O Assy
Input-/Output - Board
18-1146-18
E.U.
19-9000-95
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB HV Assy
18-1146-17
E.U.
19-9000-94
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Pulser Assy IT 2-3
18-1145-71
E.U.
19-9000-84
Assembly contains:
Pulser board
Capacitor 5000 pF 20 kV
Front Cover
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Beam Blanking Assy
18-1146-50
E.U.
19-9000-99
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Gate Driver Assy
18-1146-51
E.U.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Linear Amplifier Assy
18-1146-52
E.U.
19-9001-00
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Assy PCI I/F
PCI Card
18-1146-56
E.U.
19-9001-05
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
PCB Laser Detect Assy
18-1145-72
E.U.
19-9000-85
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Sample slides and equipment
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Automatic Slide Loader
18-1147-63
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Instrument Service Miscellenous
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Instrument Cover
Top Cover
56-3104-14
Tube Panel
the numbers underneath are internal Umea Factory 56-3104-16
numbers
Fan Cover
56-3104-17
Upper Cover
56-3104-18
Side Panel
56-3104-20
Sample Inlet
56-3104-12
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
8.8 High voltage control board
Chapter 8
8.9 Backpanel board
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
8.1 Block Diagram
AC5
COMPUTER
ACQUISITION CONTROL WITH
IO * GAS CELL CONTROL (AC3) COMMS PCB
* SAMPLE MOTOR (IO1) * GATE CONTROL (AC4)
POWER SUPPLY * IRIS MOTOR (IO9) HIGH VOLTAGE CONTROL
* DATA ACCUMULATION
BOARD * SAMPLE PORT MOTOR (IO10) * HP20R - SOURCE (HVC2)
* EXTERNAL COMMS (AC5)
* VALVES (IO5,IO6,IO7) TRANSIENT RECORDER 24V SUPPLY * HP20R - REFLECTRON END (HVC1)
* LASER TRIGGER (AC2)
+5V LOGIC * TURBO PUMP CONTROLLER INTERFACE (IO4) 500MSPS SWITCH OFF FOR ZERO VOLTS
* PULSER TRIGGER (AC1) OPTIONAL
-5.2V ECL * PIRANI GAUGE CONTROL (IO3) INPUT SIGNALS * LINEAR OPERATION * H5P - DETECTOR (HVC7)
* LASER DETECT (AC7) 2GSPS
-2V ECL * MAGNETRON GAUGE CONTROL (IO2) * REQUIRED CURRNET 2.4A * HP5R - PULSER (HVC4)
* FAST TRANSIENT RECORDER. TRANSIENT
+24V IO * HIGH VOLTAGE CONTROL INTERFACE * LINEAR DETECTOR (TR2) * HP2.5P - REFLECTRON DET 2KV (HV10)
TRIGGER (AC8) RECORDER
+/- 5V T.R. * COVERS LOOP * REFLECTRON DETECTOR (TR1) * PULSED EXTRACTION UNIT (HVC4)
+/- 5V HEAD AMP * BEAM BLANKING CONTROL * GATE DRIVER (HVC6)
* AUTO-SAMPLER / Y AXIS ON ADDTIONAL BOARD * LASER POWER SUPPLY (HVC5)
* GAS CELL DRIVE * BEAM BLANKING (HVC3) TR2 AC8 TR1
IO11 AC1
PS1 PS2 IO2 IO3 IO4 IO5 IO6 IO7 AC8 AC6 TR2 HVC2 HVC4 HVC5 HVC6 HVC3
HVC1 HV7
HV10
IO9 AC2
AUTO
SAMPLER AC7
OPTION SIG?? REFLECTRON REFLECTRON
DETECTOR
DETECTOR HP20R
IO1 COVERS
UV LASER, OPTICS, IRIS AC4 HP5P
(plus port switch) HP2.5P 4mA REF :+/-20KV
HVC4
MOTOR & LIMIT SWITCHES
AC3 +2KV
IO10 +2.5KV LIN:0V (OFF)
FIBRE OPTIC & PULSE DETECT
HP5R
SAMPLE MOTOR +/-4KV AC1
HVC2
HVC6
PULSED
SOURCE EXTRACTION UNIT
HP20R -/+4KV HV3 2KV
CONICAL DUAL
ANODE MCP
REFLECTRON AMPLIFIER
60 PLATES ELECTRON 5nF
WITH
INCLUDING ENDS MULTIPLIER OFFSET
INLET CHAMBER
ADJUST
GATE ION FLIGHT TUBE
SAMPLE PULSED BEAM GAS
LENS TUBE TUBE +/-500V KICKER
+/-20KV LENS BLANK CELL
OR 0V +/-90V
RESISTIVE DIVIDER. (67M)
EDWARDS
MAGNETRON EDWARDS EDWARDS MAINS BLOCK
IO2
GAUGE TURBO TURBO (6 OUTPUTS)
AIM-S-MW25 PUMP EXT70 PUMP EXT70 * TURBO PUMP CONTROLLER(s)
BACKING VALVE (24V,N.O.) IO7 * TCU
* FAN TRAY
INLET * ROTARY PUMP
VALVE * BACK PANEL FANS
24V N.C.
IO5
FORE LINE
PORT ON TRAP
EDWARDS
SAMPLE
PIRAINI GAUGE IO4 IO4
PROBE
APG-M-NW16 ROTARY LAST CHANGE : 09/07/99 17:35
VENT PUMP
VALVE IO3
SJ ORIGINAL XX/98 A
24V N.C. EDWARDS TURBO PUMP EDWARDS TURBO PUMP BY MODIFICATION DATE ISS
CONTROLLER CONTROLLER DRG NO : SAI-02BD SHEET 1 OF 1
IO6 EXC100L EXC100L
LT3-P1 ELECTRICAL SYSTEM DIAG.
SCIENTIFIC ANALYSIS INSTRUMENTS LTD
STRICTLY CONFIDENTIAL COPYRIGHT SAI LTD 1997
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
TCU Interconnections Block diagram
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
8.4 Power supply pcb picture of board location
8.4.1 Description
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
picture of the board
8.4.1 Description
The power supply assy supplies most of all the low voltages in
the system.It is located in the TCU unit in slot 1.
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Testprocedures for the PCB +14V and +27V unregulated (depends on load)
Maldi Pro has only 5V and 24V. -2V and -5.2V are only used Measure the voltages on TP2 wrt TP1. Should be between -
old Maldi with 500 MSPS. 14V and -27V unregulated (depends on load)
Initial inspection. Not needed if new power supply.
Check the quality of the soldering. ””Remove the sense plates from the power supply, returning
Check the quality of the crimping by gently tugging on thw the wires to the correct positions.””
wires. Measure between J1-C7 (-) and J1-C2(+). This should read
Check that connectors J1 and J2 are sitting correctly on the +5.1V. Ensure the polarity is correct.
board. Measure between J1-C12 (-) and J1-C17(+). This should
Check that the plate (B88) is firmly fitted onto the back of J2 read +5.1V. Ensure the polarity is correct.
ensuring that the warning label is visible. Measure between J1-C25 (-) and J1-C22(+). This should
Check that the backplate (B80) is firmly fitted onto the back read +2V. Ensure the polarity is correct.
of the board ensuring that the warning label is visible. Measure between J1-C28 (-) and J1-C31(+). This should
Check that holes H2, H13, H14, H15, H16 are earthed read +24V. Ensure the polarity is correct.
together. These holes form the connection to the backplate Add a sticker ”Warning!! NOT for use with 500MSPS
of the power supply and can be checked by using the systems”, (unless no conflict will arise).
screws on the backplate. Sign and date the pretest Label.
Check that the switch’s on the power supply module, SW1-3
are set to the appropriate voltage 220 or 110.
Connect the mains to the Ground, Neutral & Live on the
power supply.
IF USING AN ELECTRONICS CRATE TO TEST THE
SUPPLY, REMOVE ALL OTHER PCBS FROM THE UNIT.
Not needed if new power supply.
””With the sense plates still connected measure the voltages
on the supply and trim until the voltages are 24V, 5.0V, 5.2V
and 2V.””
Measure the voltages across TP7 (+) wrt TP8 (-) and TP5
(+) wrt TP6 (-). The output voltages should be around 9V.
Measure the voltages on TP4 wrt TP3. Should be between
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
8.4.2 Board Layout
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Linear Detector Power
TP 1 : 0V
TP 2 : +5V
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Testpoints and Voltages to adjust
+ 5V
-5,2V
- 2V
24V
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
8.4.3 Electronical Schematics
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof PRO MS 18-1160-04, Rev March 2004
contents
picture of board location
8.5 Transient recorder board
8.5.1 Description
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
and the Reset signal
8.5.1 Description · A ‘Start’ signal is sent. This is synchronised to the sampling
This takes analog signals from the detectors, digitises them clock, then passed on to trigger the pulsed extraction unit.
and stores them in memory. It uses an 8 bit ADC at a · The ‘Start’ takes the ADC out of reset. This also leads to
maximum sampling rate of 1GSPS. It’s input bandwidth is resynchronising of the data into the De-multiplexer which
approximately 800MHz. It is controlled by the Acquisition follows the ADC in the data path, and to providing the data
Control Board. It can digitise at three speeds ; 1GSPS, enable to the FIFOs at the appropriate time.
500MSPS and 250MSPS. These represent the three mass · The ADC has demultiplexed data output; there are two
ranges of the instrument. It has a HF relay to switch channels a 500MHz.
between the linear detector signal and reflectron detector
signal. · This data passes De-multiplexer where it emerges as a
62.5MHz 128bit word.
It has split power planes, as the power for the digital circuits
is kept separate from the power for the analog circuits. · This 128 bit word is stored in the FIFOs.
It has an input offset DAC to set the baseline. The DAC is · Successive data is written into the FIFOs until they are full.
used to test the PCB. · This memory full signal is passed to the Acquisition Control
Board.
It has a PAL to provide issue information to the software.
The sampling clock is reduced in frequency when not in · The Acquisition Control Board then stops the acquisition
use, in order to provide superior EMC performance. by removing the ‘Start’ signal.
A 500MHZ oscillator is used, which is frequency doubled · The Acquisition Control Board then reads the data out of
using a PLL. This is then divided by factors of 2 and 4. One the FIFOs and adds it to the accumulator memory.
frequency is select from the 3 generated via the Acquisition See schematic
Control Board. A 4th frequency can be selected, but this is Sheet 18 is a block diagram showing data path timing &
generated on the Acquisition Control Board for test and synchronisation
control purposes. This is a high speed board, and where
signals cannot be easily routed point to point, low skew
splitter chips are used such as 100EL15 and 100EL11.
Basic Sequence Of Operation.
· The FIFO memory is reset using the test & control clock
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
picture of the board
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Measure the following voltages (Ground test point first):-
Testprocedures for the board TP3 to TP1 : +5V
T/R Board TP3 to TP13 : 0V
TP3 to TP10 : +3V
Equipment used :-
TP13 to TP12 : +5V
ENSURE THAT THE MAIN POWER SUPPLY IN THE TCU IS TP13 to TP11 : -5VA
18-1159-98 ISSUE BD OR ABOVE !!! TP3 to pin 5 (attached to copper area) of
Ensure no cable other than the mains cable are attached to IC15,IC17,IC13,IC8,IC14,IC9,IC20 & IC22 : +3V
the TCU. TP13 to C322 pin 1 (no via) : +2.5V
Ensure that the TCU is switched off at the mains. TP13 to TP5 : -1.364V (adjust with RV2)
Connect up (2),(3),(4) & (5). Program IC12 using the JTAG port on J13 (if not done
Fit extender card into slot 23. already)
Examine the device under test (D.U.T.). Fit programmed IC27 into SKT1 (if not done already)
Check the right componets are fitted in the correct orientation. Run the test program (testall) on the computer.
Not fitted :- C321 Adjust the baseline to just below zero as instructed (if
Check for solder bridges, dry joints and pads not soldered. neccessary)
Check the following resistances :-
TP3 to TP1 : 53 ohms (0VD to 5VD)
TP3 to TP13 : < 0.1 ohms (0VD to 0VA)
TP3 to TP10 : 86 ohms (0VD to 3VD)
TP13 to TP12 : 670 ohms (0VA to +5VA)
TP13 to TP11 : 6.2 Kilo-ohms (0VA to -5VA).
TP3 to pin 5 (attached to copper area) of
IC15,IC17,IC13,IC8,IC14,IC9 & IC22 : 62K
TP3 to pin 5 (attached to copper area) of IC20 & IC22 : 14K
Switch on unit prior to inserting the board in order to check
voltages.
Measure between J1-C1 (0VD) and J1-C11 (5VD). This
should read +5V. Ensure polarity is correct. Fit board on to the
extender card.
Switch on mains to the TCU.
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
8.5.2 Board Layout
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
8.5.3 Electronical schematics
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi TofPro MS 18-1160-04, Rev March 2004
contents
8.6 Acquisition control boar d picture of board location
8.6.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
· If the gate is used, then at some time during the
8.6.1 Description acquisition, determined by the gate delay, a pulse is output
This board communicates with the computer. It then to the gate unit, whose width is determined by the gate
controls the timing of data acquisition. It does this by pulse width register.
controlling when the laser is fired, when the pulsed · The acquisition control board waits for the transient
extraction is triggered, when the transient recorder recorder memory to fill up. A signal is then passed back.
acquires, when the gate is activated etc. It reads the data
· The acquisition control board reads the transient recorder
from the transient recorder, sums it to previously acquired
boards data, and adds it into the accumulator memory.
spectra (if required) and then sends the data packet to the
computer via an on board buffer. The IO board has to · If it has been requested, the accumulator memory is
arbitrate with this board in order to send status to the transferred to the output buffer, where the data can be sent
computer. to the computer. (It is possible to accumulate spectra faster
then they can be sent to the computer).
Basic Sequence Of Control
· If data is being sent to the computer this happens in
· The computer sends instructions to set up the mode of
parallel with other operations on the acquisition control
operation, the pulsed extraction, the gate delay, gate pulse
board.
width and the mass range.
· The acquisition control board stops if only a single spectra
· The computer sends an instruction to start.
was required
· The laser is fired using a TTL trigger.
· Otherwise it will wait until the laser has re-changed before
· The laser light pulse is detected by the laser detect unit acquiring the next spectra.
and passed to this board.
See schematic page 200 onward.
· The received pulse is latched and passed down a
The Acqusition control card has one Voltage adjustment.
programmable delay line.
Laser detect level: TP6 - 2 V TP7 - GRD
· The output of the delay line is passed on to the transient
Adjust RV1 till read 2V at TP6. This is a start point, it
recorder board and is also used to trigger the delay to the
should be set to trigger on a clean step flank of the laser
gate pulse (if used).
detect pulse.
· On the transient recorder board the pulse is synchronised
to the sampling clock. It is then used to start the acquisition
of data and to trigger the pulsed extraction unit.
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Testprocedures for the Board Fit extender card into slot 31.
(1) Test Software Plug the DUT onto the extender card.
(2) Computer Switch on mains to the TCU.
(3) PCI interface board PCB017B Check there is 5V between TP2 and TP5.
(4) Computer interface cable C0231A If the board has not been programmed, program it now.
(5) Fully tested TCU, populated with PCBs Run the test program (testall ver3.0 or greater) on the
(6) Test Board LV LT3 PCB035A. computer.
(10) BNC to BNC RG58 coax cable > 1m long. 2off. Use the laser trigger signal from the upper BNC on the
transient recorder board.
Ensure no cable other than the mains cable are attached to
the TCU. Take a length of coax to M34 on the back of the crate
(second from bottom BNC)
Ensure that the TCU is switched off at the mains.
Plug the laser detect cable SMB (RG197 cable) into the
Connect up (2),(3),(4) & (5). board (bottom SMB connector, J9)
Examine the device under test (D.U.T.). Set the laser detect threshold to 1V adjusting RV1
Check the right componets are fitted in the correct monitoring TP6 wrt GND (TP7).
orientation. Plug the scope trigger cable SMB (RG197 cable) into the
Not fitted :- R1, J4, J5, J6 , J8 top tranisent recorder board SMB connector (J7).
Check for solder bridges, dry joints and pads not soldered. Trigger the scope using a probe on TP10 (Laser Fired).
Check the resistance between TP2 & TP5. This should be Ground is TP7.
~36 ohms. Montior the TTL signal to the pulser. (TP9 or IC29 pin 6).
Check the resistence between TP3 & IC29 pin 1,2,3 & 4. Ground is TP3.
This should be 270 ohms. Scope setup : Trigger at left, Timebase 200nS/div for max
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
delay, 20ns/div for min delay. J2 C22.
It is useful to set the scope to ”infinite persistance” for this Plug the scope trigger into the TCU scope tigger output on
test. the backpanel (bottom BNC, M35) and use this as a TTL
Run the Pulser Delay test (G). trigger, rising edge.
If this completes successfully, the laser detect is OK. Run the Gate Pulse Width test. (I). Check the scope trigger
can be seen on the scope.
The Delay counter has 8 bits. Each bit is set in turn. Finally
all the bits are set. ** Gate pulse Width **
The first delay (all bits zero) should be ~27nS from Laser On the other channel montior the TTL signal to the Gate.
Trigger. (TP8 or IC29 pin 7).
This is zero point for pulse delay. Scope setup : Trigger at left, Timebase 50nS/div. Run the
Gate Pulse Width test (I). The width counter has 4 bits. It
The second delay (l.s. bit set high) should also be ~32nS counts fron 0 to 15.
(+5ns).
The 1st pulse should be 20nS wide. Each successive pulse
The third delay (2nd ls bit set high) should be 37nS (+5ns). grows by 20nS, to a maximum of 300nS.
See following table...
Look at the differential PECL verion of signal at IC29 pins 1
1 2 3 4 5 6 7 8 9 10 delay number & 2.
00 01 02 04 08 10 20 40 80 ff hex value loaded Set scope offset to 3.5V. Set volts/div to 500mV. A positive
27 32 37 47 67 106 187 345 658 1280 delay from laser going 500mV pulse should be seen on pin 3.
pulse in nS. A negative going complimentary pulse should be seen on
Each succussive delay should be twice the previous pin 4.
(bearing in mind the 27nS offset). ** Gate Delay **
Look at the differential PECL verion of signal at J2 C22 & J2 Scope setup : Trigger at left, Timebase 5uS/div for max
C23. delay, 50ns/div for min delay. Run the Gate Delay test (H).
Set scope offset to -3.5V. Set volts/div to 500mV. positive The Gate Delay counter has 11 bits. Each bit is set in turn.
going 500mV pulse should be seen on pin J2 C23. Finally all the bits are set.
A negative going complimentary pulse should be seen on The first delay (0) is not valid and does not produce a pulse.
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.6.2 Board Layout
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.6.3 Electronical schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.7 Main I/O Board picture of board location
8.7.1 Description
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
8.7.1 Description picture of the board
The I/O control pcb contols a list of functions , and they are as
follows.
:Sample Motor
:Iris Motor
:Sample Port Motor
:Valves ( Inlet, Vent, backing )
:Turbo pump controllers
:Pirani gauge
:Magnetron gauge
:High voltage control Interface
:Covers loop
:Beam blanking control
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
Testprocedures for the board TP14 GRD
The I/O pcb has several test points. Lets first check and be
sure that there are no shorts between: TP1 (1V adjust)
TP14 GRD
(TP7 and TP9)
(TP14 and TP12)
(TP14 and TP13)
First we need to connect Jumpers to pin 1&2 of J5 and pin1
(TP14 and TP15)
& 2 of pin of J6.
(TP14 and TP10)
Now measure at the following Test points:
(TP16 and TP17)
TP17 24V
TP16 GRD
The following are the voltages you should see at these test
points.
Adjust TP3 for 5.6V, use RV3.
TP7 +5V (+/- 0.1V )
TP6 is the GRD. ( Magnetron set point)
TP9 Grd
Adjust TP4 for 3.8V , use RV4.
TP12 +24V (+/- 0.1V)
TP6 is the GRD. (Pirani set point)
TP14 Grd
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
· Test Software. Check that there is no short between TP7 & TP9.
· Computer. Check that there is no short between TP14 & TP12.
· PCI interface board PCB017B. Check that there is no short between TP14 & TP13.
· Computer interface cable C0231A. Check that there is no short between TP14 & TP15.
· Fully tested TCU, populated with PCB’s. Check that there is no short between TP14 & TP10.
· Test Board LV LT3 PCB035A. Check that there is no short between TP16 & TP17.
· Multimeter. Switch SW2 according to the instrument type LT2/LT3.
· Extender Card. Connect PCB022A (D.U.T.) ono the extender card.
· Oscilloscope. Connect PCB035A to M8 on the backpanel of the TCU.
· 2 Jumper connectors. On PCB035A toggle SW2, SW3, SW4 & SW5 down-
· 2 off Resistor Link 2K2 with crimp ‘D’ pins. wards (over-ride)
· Sample motor rig including limit switches. Connecting ca- Ensure SW1 & SW8 are in the centre position (“OFF”)
ble. Switch on TCU at the mains.
· Iris motor rig including limit switches. Connecting Cable. Measure 5V between TP7 & TP9
· Inlet port rig including limit switches. Connecting Cable. Measure 24V between TP14 & TP12.
Measure 12V between TP14 & TP15.
Ensure no cable other than the main’s cable are at- Measure 5V between TP14 & TP13.
tached to the TCU. Measure 20V between TP14 & RN6 pin 1 (mitred cor-
Switch TCU off at mains. ner).
Connect up (2), (3), (4) & (5). Measure the voltage between TP14 & TP1. Adjust the
Fit extender card into slot 47. voltage at TP1 with RV2 to be 1V.
Examine the device under test (D.U.T.). Connect Jumpers to J5 & J6
Check the right components are fitted in the correct Measure 24V between TP16 & TP17.
orientation. Measure the voltage between TP6 & TP3. Adjust to 5.6V
Check for solder bridges, dry joints and pads not sol- using RV3.
dered. Measure the voltage between TP6 & TP4. Adjust to 3.8V
using RV4.
Carry out a “Simulated Switch on Sequence” as detailed
on a separate page of this document.
While the “Simulated Switch on Sequence” is underway,
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
monitor the pulses from the stepper motor drive on J2 at the switches is correctly displayed.
C18, C19, C20 & C21. Observe a 24V square wave. Select the iris motor ensuring the selected motor status
Set the scope to 20V/div and 2ms/div. The pulses are bits readback correctly (J).
present while LED3 is on. Once the sequence is com- Move the iris motor from limit to limit checking the op-
plete the I.O. board is initialised. eration of the limit switches (M, R and K).
Ensure that the sequence completes as described. Check the switches stop the motor and that the status of
Once the I.O. board has initialised correctly, switch off the switches is correctly displayed.
the unit. Using Test Board LV LT3 PCB035A, set signal out (J7)
Plug in a tested sample motor rig into the connector to 2.8V with RV1.
marked M27 on the backpanel. Link J7 to J6 (SIG. OUT. to VACUUM SIGNAL).
Plug a tested iris motor rig into the backpanel connector Select the Magnetron ADC channel. Read back should
marked M12. be 2.8V +/-0.2V (L).
Switch on the unit again. The sample motor should start Remove the IO board from the extender and refit nor-
to move. If it does not then switch off immediately to mally into the unit.
avoid driver chip damage. Investigate before re-applying
power. I/O PCB Simulated sample Exchange Sequence
Carry out a “simulated exchange sequence” as detailed
on a separate page of this document. When the TCU is switched on, the I.O. board (P022)
Use the test software to manipulate the board. enters an “exchange sequence”.
Test the accumulator board to ensure that it is working. It drives the sample probe into the sample inlet.
Check that the I.O. Board Type is shown as “3”. In order to run the TCU without an instrument it is nec-
Perform a “Basic IO Test” to show that communication essary to simulate inputs.
with the computer is working, then use the “IO Interac- This is done with the Test Board P035A. There are
tive Menu”. switches to simulate the inputs required.
Make sure a sample exchange is not underway. During the sequence observe the LED on the test board
Select the sample motor ensuring the selected motor and the IO board.
status bits readback correctly (J). On the IO board LED3 will be on (green) when the sam-
Move the sample motor from limit to limit checking the ple motor is being driven.
operation of the limit switches (M, R and K). On the IO board LED4 will be on (red) while the ex-
Check the switches stop the motor and that the status of change sequence still underway.
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
On the IO board LED5 will be on (red) when the direc-
off. (Motor changed direction & moving)
tion of the sample motor is towards the port.
LED14 will go off (inlet valve)
The following instructions refer to the Test Board P035A
LED15 will go off (backing valve)
unless otherwise stated: -
Switch SW1 to the centre (“OFF”) position.
Ensure SW1 & SW8 are in the centre position.
Switch SW1 to the down (DATUM) position.
Ensure SW2, SW3, SW4 & SW7 are in the down (over-
On the IO board LED3 & LED4 will go off.
ride) position.
Switch SW1 to the centre (“OFF”) position.
Ensure SW6 is in the up position.
Switch On TCU (or send “Exchange” message).
The following LED’s should be on - LED1, LED3, LED4,
LED5, LED8, LED9 & LED10.
The following LED’s should be off - LED6, LED7,
LED13, LED14 & LED15.
On the IO board LED3, LED4, LED5, LED8, LED6,
LED9 should be on. LED7 should be off.
Switch SW1 to the up position. (Simulate motor limit).
LED9 will go off. (Limit reached)
On the IO Board LED3 will go off. (Sample has reached
inlet port)
LED13 will go on (vent valve on).
Switch SW7 up (lid open).
LED13 will go off (vent valve closed).
Switch SW7 down (lid closed again).
LED15 will go on (backing valve)
LED14 will go on (inlet valve - sample inlet is being
pumped)
Wait for 20 seconds. This is the minimum pump down
time.
Contents Chapter
contents
On the IO board
Service Manual Maldi Tof Pro MS
LED3 will come on and LED5 will go
18-1160-04, Revised March 2004
schematics
8.7.2 Board Layout
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
8.7.3 Electronical schematics
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
Contents Chapter
contents Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
picture of board location
8.8 High Voltage Control Board
8.8.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.8.1 Description picture of the board
Source (HP20R)
Reflectron (HP20R)
(Switch off for zero volts)
Detector (HP5)
Pulser (HP5R)
Gate driver
Beam blanking
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Testprocedures for the board TP9-10V (+/- 50mV)
Using the test software set ” Source DAC ” to 0xFFFF
The high voltage PCB has many test points which will need to
checked and adjusted at times. You will also need to use TP14-5V (+/- 50mV)
diagnostics software to do some of the following tests. All the Using the test software set ” Reflectron DAC ” to 0x8000
following Test will use TP 15 as the ground connection.:
TP14-10V (+/- 50mV)
TP 2 -24V (+/- 50mV) Using the test software set ” Reflectron DAC ” to 0xFFFF
TP 4 -7.1V (+/- 50mV)
TP1- 1V Adjust RV1 to set Pulser Ref Check in the test software status that the Source supply
TP 5 - 8v (+/- 50mv) Adjust RV 2 to set Gate Ref Positive and the Pulser supply negative readback is asserted.
Measure the voltage at the following test point:
TP 6 -9.9V to 10.1V Adjust RV 3 to set HMD Ref
TP 7 -9.9V to 10.1V Adjust RV 4 to set Spare Ref 2 TP15 Grd & TP3 24V (+/-50mV)
TP 8 -9.9V to 10.1V Adjust RV 5 to set Dynode Ref
TP13 -9.9V to 10.1V Adjust RV 7 to set Spare Ref 1 Using the test software select ”Negative Ion” E
Check in the test software status that the ”Source supply
The following will use the test software: Negative and the pulser supply Positive ” readback is
asserted.
TP224V (+/- 50mV) Measure the voltage at the following test point:
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
• Test Software. Fit Resistor link on backpanel of the TCU on the connec-
• Computer. tor marked M26 between pins 2 & 3.
• PCI interface board PCB017B. Switch on TCU at the mains.
• Computer interface cable C0231A. Carry out a “simulated exchange sequence” as detailed
• Fully tested TCU, populated with PCB’s. below
• Test Board LV LT3 PCB035A. Using the test software (IO interactive) select “Linear
• Multimeter. Mode” (V).
• Extender Card. Using the test software select “Positive Ion” (E).
• 2 off Resistor Link 2K2 with crimp ‘D’ pins. Observe the Leeds on the IO board in slot 47.
• Oscilloscope. On the IO board the bottom 5 LED’s should all be on
except the lowest one.
Check the correct components are fitted in the correct No LED’s should be on the board under test.
orientation. Using the test software select “HV request” (F).
Check for solder bridges, dry joints and pads not sol- The bottom LED of the IO board should illuminate.
dered. Ensure no cable other than the main’s cable are Check in test software status that the “High Voltage On”
attached to the TCU. readback is asserted.
Switch TCU off at mains. On the D.U.T. the LED1, LED3 should be on.
Connect up (2), (3),(4) & (5). On the D.U.T. the LED2, LED4 should be off.
Fit extender card into slot 59. Use TP15 as a ground point.
Examine the device under test (DUT). Measure +24V at TP2.
Check that there is no short between TP15 & TP2. Measure +5V at TP17.
Connect PCB021A (D.U.T.) ono the extender card. Measure +2.5V at TP18.
Connect PCB035A to M8 on the backpanel of the TCU. Using the test software set “Source DAC” to 0x8000. (A)
On PCB035A toggle SW2, SW3, SW4, SW6 downwards Measure +5V at TP9.
(over-ride). Using the test software set “Source DAC” to 0xFFFF. (A)
On PCB035A toggle SW5 & SW7 upwards. Measure +10V at TP9.
Ensure SW1 & SW8 are in the centre position (“OFF”). Using the test software set “Reflectron DAC” to 0x8000.
Fit Resistor link on backpanel of the TCU on the connec- (B)
tor marked M5 between pins 2 & 3. Measure +5V at TP14.
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Using the test software set “Reflectron DAC” to 0xFFFF. LED2 should illuminate.
(B) Measure +24V at R8 pin 1 (mitred corner).
Measure +10V at TP14. Using the test software, deselect “HV request” (F)
At TP4 using RV1 set the “PULSER REF” to +7.1V Attach scope probe to TP18. Test trigger point to 2.4V.
At TP5 using RV2 set the “GATE REF” to +8.0V Using the test software, select “HV request” (F)
At TP6 using RV3 set the “HMD REF” to +10.0V (+/-0.1) View the 2.5V reference ramping up to full scale in
At TP7 using RV4 set the “SPARE REF” 2 to +10.0V(+/- around 0.5 second.
0.1) Using the test software, select “Blanking On”. (X)
At TP8 using RV5 set the “DYNODE REF” to +10.0V (+/ Using the test software, set “Blanking Time To 0x7F”. (Y)
-0.1) Attach a scope probe to J1 C13.
At TP13 using RV7 set the “SPARE REF 1” to +10.0V Trigger the scope using the “Scope Trigger” BNC output
(+/-0.1) on the back of the TCU (M35)
Check in test software status that the “Source Supply
Positive” readback is asserted.
Check in test software status that the “Pulser Supply
Negative” readback is asserted.
At TP3 measure +24V.
Using the test software select “Negative Ion” (E)
Check in test software status that the “Source Supply
Negative” readback is negated.
Check in test software status that the “Pulser Supply
Positive” readback is negated.
At TP3 measure +0V.
On the DUT the LED1, LED4 should be on.
On the DUT the LED2, LED3 should be off.
Check in test software status that the “Flight path - Lin-
ear” readback is asserted.
Using the test software select “Reflectron Mode”. (V)
Check in test software status that the “Flight path - Re-
flectron” readback is asserted.
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.8.2 Board Layout schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.8.3 Electronical schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
location of the board
8.9 Backpanel Board
8.9.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.9.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.9.2 Board Layout schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.9.3 Electronical schematics
8.9.3.1 Block Diagram of Backpanel Board
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.9.3.2 Power Supply board Slot
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.9.3.3 Transient Recorder Board Slot
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.9.3.4 Acquisition Control Board Slot
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.9.3.5 Spare I/O Board Slot
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.9.3.6 I/O Board Slot
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.9.3.7 High Voltage Control Board Slot
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.10 Backplane Board
8.10.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.10.4.1 Description
picture of the board
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.10.4.2 Board Layout
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.10.4.3 Electronical
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
picture of the board location
8.11 PCI Bus Interface
8.11.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.11.1 Description picture of the board
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
schematics
8.11.2 Board Layout
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.11.3 Electronical schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.12 24 V Power Supply (PSU) picture of the location of the board
8.12.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.12.1 Description of the board picture of the board
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.12.3 Electronic schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
picture of board location
8.13 Gate PCB
8.13.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.13.1 Description picture of the board
The gate pcb is located in the box at the base of the reflectron
stack. To gain access to the pcb remove all the screws and
remove the cover. There are several test points and
adjustments on this pcb. To test the pcb do the following.
Find J1, this is the 10 pin molex connector. Apply 24v to pin 1,
this is the power connection.
Pin 2 - Grd
Pin 4 - Polarity
Pin 7 - Reflectron-on
Pin 9 - Gate-Ref
Pin 10 - Gate-Ref-Rtn
Measure the following voltages.
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Measure the following.
Next apply +8V to J1 pin 9 (Gate-Ref) and check voltages and
adjust if needed.
Tp11 (Low) and Tp10 (Grd) (Low < 0.7V)
Tp 4 +400v Adjust Rv3 Tp9(High) and Tp10n (Grd) (High > 2.0V)
Tp 5-400V Adjust Rv4
Tp7(-400V) and Tp J4 (Grd) (-400V)
J7 +90V Adjust Rv1
Tp6(+400V) and Tp J4 (Grd) (+400V)
J8 -90V Adjust Rv2
Now we will set the TTL input to the input driving circuit to a
Next apply +24V to J1 pin4 (Polarity) Check the following. High state (> +2.0V)
Measure the following.
Tp J7(-90V)
Tp J8(+90V) Tp11 (High) and Tp10 (Grd) (High > 2.0V)
Tp 9 (Low) and Tp 10 (Grd) (Low < 0.7V)
Next apply a signal to J2 (SMB connector). Tp7 and Tp J4 (0)
This input requires a differential PECL pulse. Tp6 and Tp J4 (0)
The input can be generated from a TTL pulse.
The chip to use is a Synergy SY100ELT22ZC TTL- PECL Now apply a positive going 40ns wide TTL pulse to the input
driver. For an example of the input driving circuit see driving circuit and observe both pulses at the same time.
schematic # P023B05.
On the schematic, see the bottom center for the Tp7 (-400V to 0 to -400V) and Tp J4 (Grd)
SY100ELT22ZC. Rising pulse to 0 then falling back
The input driving circuit and the unit under test should share
the same ground. Tp6 (+400V to 0 to +400V) and Tp J4 (Grd)
Set the TTL input to the input driving circuit to a low state Falling pulse to 0 and rising back
(<+0.7V). Use RV5 and RV6 to make the pulses start at the same point
in time.
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
schematics
8.13.2 Board Layout
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.13.3 Electronical schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
picture of board location
8.14 Pulser Board
8.14.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
picture of the board
8.14.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Testprocedure for the board For an example of the input driving circuit see schematic
sheet P023B05.
• Multimeter. On the schematic, see bottom centre for the
• High voltage probe. SY100ELT22ZC integrated circuit.
• Oscilloscope >250Mhz bandwidth 1M ohm input im- The input driving circuit & the unit under test should
pedance. share the same ground.
• Scope probe 1.25Kv max input bandwidth >250MHz. Apply a positive going 150Us wide TTL pulse to the
100:1 input driving circuit at 8Hz.
• Input driving circuit. Using the oscilloscope between ground point at R1 pin 2
• TTL pulse generator. (No mitre on the ident) and J10 pin 2, observe a positive
• 24V power supplies. going TTL pulse 150Us wide with an amplitude > 3.5V
Switch off the TTL pulse generator.
Place the assembled PCB on mounting pillars on a Apply voltages to J9 (spade connector) as follows:
workbench. Pin 1 (HV in) -1KV
Apply the following voltages to connector J2 Use the Multimeter in conjunction with the high voltage
Pin 1 (polarity) +24V probe.
Pin 2 (ground) 0V Use J6 as a ground point.
Pin 3 (power) +24V On the spade terminals of the relays nearest to IC2
Apply the following voltages to connector J9 Test equip- (Behlke switch) measure the following: -
ment required: RLY1 and RLY2 should be 0V.
Electronic Multimeter RLY3 and RLY4 should be -1KV.
Oscilloscope > 250MHz Bandwidth On the spade terminals of the relays furthest from IC2
Pin 1 (HV in) +0V (Behkle switch) measure the following: -
Measure +5V between J10 pin 1 and 3. RLY1 and RLY4 should be -1KV.
Apply a signal to J4. RLY2 and RLY3 should be 0V.
This input requires a differential PECL pulse. Apply voltages to J9 (spade connector) as follows: -
The input can be generated from a TTL pulse. Pin 1 (HV in) 0V.
The chip to use is a Synergy SY100ELT22ZC TTL-PECL
driver.
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Apply voltages to J2 (3-way Molex connector) as fol- Pin 1 (HV in) 0V.
lows: - Apply voltages to J2 (3-way Molex connector) as fol-
Pin 1 (Polarity) 0V (negative ion). lows: -
Pin 2 (Ground) 0V. Pin 1 (Polarity) 24V (positive ion).
Pin 3 (Power) 24V. Pin 2 (Ground) 0V.
Apply voltages to J9 (spade connector) as follows: - Pin 3 (Power) 24V.
Pin 1 (HV in) +1KV. Apply voltages to J9 (spade connector) as follows: -
Use the Multimeter in conjunction with the high voltage Pin 1 (HV in) -1KV.
probe. Measure a negative going pulse with amplitude of -1KV
Use J6 as a ground point. on the scope.
On spade terminals of the relays nearest IC2 (Behkle The falling edge should be sharp, and the rising edge
switch) measure as follows: will climb slowly.
RLY1 and RLY2 should be +1KV. The pulse should remain at -1KV for 100uS.
RLY3 and RLY4 should be 0V. The falling edge should have a fall time (10% to 90%) of
On spade terminals of the relays furthest from IC2 less than 20nS.
(Behkle switch) measure as follows: -
RLY1 and RLY4 should be +1KV.
RLY2 and RLY3 should be 0V.
Apply the TTL pulse as described above.
Attach the scope probe ground to J6.
Attach the scope probe tip to the M5 ring terminal on the
wire from J5.
Measure a positive going pulse with amplitude of +1KV
on the scope.
The rising edge should be sharp, and the falling edge
will drop slowly.
The pulse should remain at +1KV for 100uS.
The rising edge should have a rise time (10% to 90%) of
less than 20nS.
Apply voltages to J9 (spade connector) as follows: -
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
schematics
8.14.2 Board Layout
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.14.3 Electronical schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.15 Beam Dump Board picture of board location
8.15.1 Description
Contents Chapter Service Manual Maldi Pro Tof MS 18-1153-59, Revised March 2004
contents
8.15.1 Description picture of the board
The beam dump assy is located on the first shelf from the
bottom of the unit and is connected to the TOF tube by a
cable.
The beam dump has only on adjustment. Listed below is the
proper procedure for this adjustment.
You need to apply 24V between pin 1 (24V) and pin 6 (Grd) of
the 9way D connector.
The beam dump is used to redirect the lower mass ions that
are mainly from the matrix so the detector does not saturate.
Contents Chapter Service Manual Maldi Pro Tof MS 18-1153-59, Revised March 2004
contents
Testprocedures for the board
Contents Chapter Service Manual Maldi Pro Tof MS 18-1153-59, Revised March 2004
contents
8.15.2 Board Layout schematics
Contents Chapter Service Manual Maldi Pro Tof MS 18-1153-59, Revised March 2004
contents
8.15.3 Electronical schematics
Contents Chapter Service Manual Maldi Pro Tof MS 18-1153-59, Revised March 2004
contents
8.16 Preamplifier Board picture of board location
8.16.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.16.1 Description picture of the board
To check or adjust pcb. Open cover of box and leave the pcb
assembled in the box, there is no need to remove it.
Plug in the power cable.
Plug output into a scope terminated in 50 ohms.
Ground the output.
Check the output for 0V and adjust the dc offset using RV1
and observe the level on the oscilloscope.
Then you are done.
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Testprocedures for the board
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.16.2 Board Layout schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.16.3 Electronical schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
picture of board location
8.17 Laser detect PCB
8.17.1 Description
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.17.1 Description picture of the board
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Testprocedures for the board
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.17.2 Board Layout
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.17.3 Electronical schematics
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.18 Reflectron detector
8.18.1 Description
8.18.2
8.18.3
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.18.1 Description
MCP
Distance washer
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.18.3 Electronical schematics
HT 2000V
10 M MCP assy
1M
Grid
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
8.19 Linear detector detector
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
Schematic of the linear detector
Contents Chapter Service Manual Maldi Tof Pro MS 18-1160-04, Revised March 2004
contents
1 volt.
LOAD REGULATION: -Less than 10ppm for load changes from 100uA to
maximum. (Measured at maximum voltage).
PRECISION HIGH VOLTAGE MODULES DRIFT: -Typically 0.01% per hour Measured at constant input voltage, 0.02% per 8
hours load current & ambient temperature after 1 hour warm up.
OUTPUT: By 0.5 metre screened cable (HP20, HP30 & HP60 1 metre). N.B.
INPUT VOLTAGE: -24 volt d.c. +10% at less than 1A. Negative input terminal Reducing the cable length may increase the ripple voltage.
common to HV earth return.
INPUT CONNECTOR: Molex 10 pin, mating connector supplied.
OUTPUT VOLTAGE: -See table above.
LINE REGULATION: -Less than 10ppm of maximum voltage for input changes of Applied Kilovolts Ltd., 38-40 Eastbrook Road
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
technician who is fully aware of the particular dangers associated with
SAFETY & INSTALLATION high voltage switch mode power supplies. Metallic or conductive tools
INSTRUCTIONS (December 1996) should not be used to adjust any of the potentiometers. The unit has no
user serviceable parts and should not be dismantled.
FOR HP & KS SERIES DO NOT HANDLE OR TOUCH THESE UNITS WHEN THE SUPPLY
IS CONNECTED. AFTER DISCONNECTION FROM THE SUPPLY,
PLEASE READ CAREFULLY BEFORE INSTALLING OR ALLOW 30 SECONDS BEFORE HANDLING SO THAT ALL THE
CAPACITORS CAN DISCHARGE. To ensure that the output is fully
OPERATING THIS POWER SUPPLY discharged short to ground before touching any high voltage circuit.
Care should be taken not to operate the unit outside the specified limits
given above, failure to do so may damage the unit.
Under worst case conditions the unit draws a current of 1A (HP) and 0.5A
GENERAL (KS) and any input supply cable must be of a suitable type and rating. The
unit is not fitted with a fuse and so should be operated from a limited
On receipt the unit should be carefully unpacked and inspected to ensure supply. Fuses may be fitted externally to the unit to protect unit and
that no transit damage has occurred. Provided that this inspection is interconnecting wiring etc. but these should be rated to prevent nuisance
satisfactory and reveals no evidence of damage then installation can failures. Care should be taken in the design of the interconnecting wiring
proceed. within the system to ensure that connectors with hazardous voltages
cannot be connected to accessible circuits.
If an electrical test is to be carried out prior to fitting the power supply, it
is essential that the person undertaking this work be a qualified electronics Ensure that the output is connected to the load prior to operation of the
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
technician who is fully aware of the particular dangers associated with unit and that a good low impedance high voltage joint is made. Sharp
high voltage switch mode power supplies. Metallic or conductive tools points on either the high voltage or return joint should be avoided as this
should not be used to adjust any of the potentiometers. The unit has no will cause corona which will make the output appear noisy. In general a
user serviceable parts and should not be dismantled. tracking distance (creepage distance) of 25mm (1 inch), per 10kV to earth
is advised as a minimum to ensure no breakdown or corona occurs, a
DO NOT HANDLE OR TOUCH THESE UNITS WHEN THE SUPPLY much greater distance will be required under adverse conditions. Care
IS CONNECTED. AFTER DISCONNECTION FROM THE SUPPLY, must be taken not to damage the cable inner when forming the
ALLOW 30 SECONDS BEFORE HANDLING SO THAT ALL THE connections.
CAPACITORS CAN DISCHARGE. To ensure that the output is fully
discharged short to ground before touching any high voltage circuit. During arcing currents exceeding 1000 Amps will flow. It is important
that these currents return to the high voltage power supply by the shortest
Care should be taken not to operate the unit outside the specified limits possible route using the screen (shield) of the output cable. Failure to
given above, failure to do so may damage the unit. observe this will result in the control terminals of the unit seeing large
voltage spikes during arcing and radiation of electromagnetic interference.
COMPLIANCE WITH SAFETY STANDARDS
Adequate ventilation should be provided to keep the unit cool and the
The unit is designed to meet Normalised European Safety Standards and ventilation inlets should not be covered in any way. The ambient air
hence installation of the power supply unit into the equipment should temperature around the inlet must not exceed 45 C (KS), 50 C (HP). The
comply with the following requirements. unit will operate in any orientation, however it is not recommended to
operate with the side fitted with the silk-screen as the lowest face.
a. A PROTECTIVE EARTH must be provided for safety in accordance
with EN61010 Part 1 : 1993 : Clause 6.5.1. The case of the units should be OPERATING NOTES
bonded to this protective earth.
1/ HIGH VOLTAGES ARE DANGEROUS. ENSURE THE OUTPUT IS
b. The output is classed as hazardous and must therefore not be accessible FULLY DISCHARGED BY SHORTING TO GROUND BEFORE
to operators. The output must be isolated from accessible circuits by TOUCHING ANY HIGH VOLTAGE CIRCUIT.
Double Insulation or a protective screen as defined in EN61010-1.
2/ The unit is short circuit proof but care should be taken that the high
c. The unit is intended to be installed in an electrical enclosure and should voltage cannot be shorted into one of the control pin connections.
not be accessible to the operator. Access should be restricted to authorised
service personnel only, with use of a tool. Care should be taken to prevent 3/ KS SERIES ONLY. R101 sets the oscillator frequency, in the event of
access to the interior of the unit and protect against items (e.g. tools or this pot being disturbed it can be reset by adjusting for minimum input
wires) inadvertently entering the interior of the unit. current at full load and full output voltage.
Ensure that the output is connected to the load prior to operation of the
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
Contents Chapter Service Manual Ettan Maldi Tof Pro MS 18-1160-04, Rev March 2004
contents
9.8 Edwards compound molecular pumps
Chapter 9
Appendix
Chapter Contents
9. Appendix