You are on page 1of 4

iGC SEA

Measuring Surface Energy for Viscous Fluids Technical


Using Inverse Gas Chromatography
Note 803
Armando R. Garcia
Surface Measurement Systems Ltd.

The measurement of surface energies of solids is a well proven technique using Inverse Gas
Chromatography (IGC). There is a great deal of interest however in extending this valuable technique to
liquids. This paper explores this possibility and provides data on techniques and results.
concentrations. A flame ionisation detector was
Introduction used to determine the chromatogram and
resulting retention times.
The analysis of the dispersive component of the
IGC has been shown to be a valuable research surface energy was performed by measuring the
tool for the measurement of surface energies of net retention volume VN (measured retention
solids [1,2,3]. A wide range of materials have volume – dead volume) for a series of alkane
been analyzed with IGC, yielding repeatable elutants (undecane, decane, nonane, octane and
values of surface energies, where other methods heptane). The dead volume was determined by
have been less reliable and more complex to injecting methane, a probe molecule that does
perform. However, IGC is not limited to solid not interact significantly with low energy
surfaces. A substantial amount of work on liquid materials. A minimum of four alkanes were used
systems has been reported in the literature [4,5]. in all experiments.
This technical note explores different methods by By plotting RTln(VN) versus a(γLD)1/2 , where a is
which viscous liquids may be analyzed using iGC- the molecular area of the probe molecule and γLD
SEA to measure surface energetics, in a manner is the dispersive component of the surface energy
that is reproducible and straight forward of the liquid elutant (surface tension), a straight
line is produced. The slope equals to 2NA(γSD)1/2
from which γSD, the dispersive component of the
solid surface energy, can easily be deduced [6].
Method Three column preparation methods have been
All analyses were carried out using the SMS-iGC explored. One method uses a support material,
SEA instrument and the SMS-iGC v1.3 analysis such as glass wool to hold the liquid in place and
suite of macros. All columns were initially to provide a surface upon which the liquid can
conditioned at the experiment temperatures spread as a thin film. Another method which has
listed below, for 2 hours and 0% relative humidity been explored is the simple coating of the viscous
(RH) and experiments were carried out with liquid on the inside wall of the IGC SEA column.
injections of 0.03 P/P0 to 0.06 P/P0 elutant vapour For this method to work, the film deposited on

P-SCI-319 v1.0 12NOV2013


the inside wall must be thin and therefore the
liquid should be amenable to spreading on the As can be seen in Table 1, the initial experiment
glass. A non-silanised glass column is preferred on day 1, results in a value of 35.85 mJ/m2. As
for this procedure. Finally, the third method time progresses, the value increases significantly
explored is one where the liquid is dispersed on until it reaches 49.95 mJ/m2 on the sixth day.
glass beads. Excess liquid is allowed to drain from Upon close inspection of the column, the sample
the beads prior to loading the beads into the appears to have formed regions where it has
column. pooled and is no longer dispersed on the glass
Data has been taken using the above three wool. This method therefore is not a good one,
methods with two resin materials, A and B, and as the data is not reliable.
also with a very viscous pitch sample C.
Column Internal Coating

Results In the second method, samples are coated on the


walls of the column. This is done by wetting a
Glass Wool Support sterile swab made of foam, and inserting it into
The results obtained with silanised glass wool as the column to wet the inside walls. Fairly good
the sample support, initially appear to give results are obtained with this method, but usually
reasonable data. Data on a 500 mg sample of the sample has to be kept at a low temperature
resin dispersed on glass wool is shown in Table 1. to reduce the possibility of the sample flowing to
However, as time passes and additional the bottom. The low temperatures (from 5 to 10
experiments are performed, the dispersive °C) maintain a film on the walls of the column for
surface energy values appear to increase. This a reasonable amount of time, rather than
effect may be explained as a bulk effect during allowing the sample to form droplets. This
the experiment, where solvent is absorbed into requires the use of non-silanised columns in order
the bulk of the fluid and is retained for a longer to create a film on the surface of the glass.
amount of time than it would be, if the Eventually however, the sample will flow and
interaction were simply with the surface of the begin to form droplets, unless the temperature is
fluid. In doing so, the solvent molecules spend below the sample freezing point. The data shown
time dissolved in the sample rather than simply below is for the second sample B.
interacting with the surface. This occurs because
Table 2: Dispersive component of the surface
the fluid forms droplets within the glass wool
energy for Sample B coated onto column
once it begins to coalesce. These droplets
Experiment Time Column Dispersive Surface
provide an area for the solvent to be adsorbed (Days after sampling) Temperature (°C) Energy (mJ/m2
into the bulk of the sample. This lengthening of 1 10 34.45
the retention time is incorrectly interpreted as an 2 10 37.05
increase in γSD by the analysis software. 3 10 35.53

4 10 42.73
Table 1: Dispersive component of the surface
energy for Sample A on glass wool
Experiment Time Column Dispersive Surface
(Days after sampling) Temperature (°C) Energy (mJ/m2 The data above shows that the initial experiment
1 30 35.85 on Day 1 yields a value of 34.45 mJ/m2, not very
2 30 43.06 different from the initial value of Sample A above.
5 30 46.97
This value increases with time, though not as
dramatically as with the previous method. On the
6 30 49.95
fourth day the value has increased to 42.73
mJ/m2. Clearly, this method is also unreliable at that this value at a temperature of 30 C is slightly
longer time intervals but good for short higher than the value obtained earlier for Sample
experimental times. B using the internal coating method at 10 °C
Coated Glass Beads (Table 2, line 1). The average of the above
measurements in Table 3 is 37.57 mJ/m2. The
Samples are prepared by mixing a small amount
average deviation is 0.35 mJ/m2. This is a very
of liquid with enough beads to fill a glass column.
reasonable deviation for a small set of
This may be done in a shallow container such as a
experiments. These data were obtained on the
weighing boat. The beads are stirred in the liquid
same column, run consecutively.
in order to completely wet their surface. The
excess liquid may be decanted, or the beads may Table 4: Dispersive component of the surface
be placed on a sieve to allow excess fluid to drain. energy for Sample A on glass beads
The beads are then loaded into the column. The Experiment Time Column Dispersive Surface
objective is to have only a thin coating of liquid (Days after sampling) Temperature (°C) Energy (mJ/m2
on the beads. The column should be plugged at 1 30 35.38
both ends with glass wool to prevent the beads 2 30 35.37
from falling out. The column is then ready for 2 30 34.97
experiments.
3 30 36.69
The column should not be handled by hand more
3 30 37.40
than necessary in order to avoid sample heating,
which will speed up the pooling of liquid between Resin sample A was also analysed using the glass
the beads. Instead, use gloves or tongs to keep bead method at 30 °C. This data is shown above
the column at room temperature. Enough glass in Table 4. The average of these measurements is
wool should be used at both ends of the column 35.96 mJ/m2, and the standard deviation is 0.866
so that if the fluid does accumulate at the bottom mJ/m2.
of the column, it will not enter the system. This is These results show a clear difference between
very important to prevent contamination of the the two samples A and B. As a check of the
IGC SEA lines. It is advisable to let the column procedure, a column of clean beads was analyzed
stand upright for a few hours outside the for surface energy. The result of those
instrument. This is to make sure that the film on experiments, six in all, was an average of 31.72
the beads is thin enough that it will not flow to mJ/m2, with a standard deviation of 0.57 mJ/m2.
the bottom in a length of time comparable to the This shows that the data obtained above is really
experiments being performed. This method a result of the interaction between the probe
yielded the most reliable results at room molecules and the resins, not the beads.
temperature.
If the columns are stored in an upright position
Table 3: Dispersive component of the surface over an extended period of time, the liquid will
energy for Sample B on glass beads eventually flow to the bottom, therefore if the
Experiment Time Column Dispersive Surface
columns will be used again, they should not be
(Days after sampling) Temperature (°C) Energy (mJ/m2 stored vertically, but horizontally.
1 30 37.10
Pitch Sample C on Beads
1 30 38.07
Experiments were also performed on a Pitch
2 30 37.34
sample C, coated on glass beads. This sample
2 30 37.77
does not flow readily at room temperature and
The data in Table 3 for Sample B show that on requires heating to 100 °C in order to be able to
Day 1, the value obtained is 37.10 mJ/m2. Notice coat the beads. A small amount of pitch was
heated in a shallow container in a lab oven and
iGC SEA
Technical
Note 803

glass beads were then introduced into the


container. Once the beads were coated with
pitch, following manual stirring, the excess pitch
was decanted. The beads were allowed to cool
Conclusions
and then loaded into the column following the
above procedure. In conclusion, of the methods studied, the most
Data were taken for the pitch sample at 30 C as promising method is the sample coated on glass
shown in Table 5. There is little sample flow at beads. It appears to yield the most reproducible
this temperature. One measurement was done at data over the longest period of time. Coating the
20 °C. Given enough time however, this sample sample on the inside of the glass columns is also
too will flow and form regions where the film promising but requires that the sample be kept at
pools into droplets. There is a large difference low temperature, and the data is not as
between the early data, taken soon after sample reproducible as the glass bead method. With
preparation, and the later data taken days after either method, analysis should always be
preparation. performed on a freshly prepared sample to
minimize sample flow. With careful sample
Table 5: Dispersive component of the surface preparation and attention to detail, it should be
energy for a Pitch sample C on glass beads possible to obtain good quality data on liquids,
Experiment Time Column Dispersive thus extending the capabilities of an already
(Days after Temperature Surface Energy
sampling) (°C) (mJ/m2
versatile instrument
1

1
30

30
61.16

61.32
References
1 30 62.70
[1]Thielmann, F., J. Chrom. A 1037 (2004), 115
3 30 76.55 [2]van Astem, A. et al, J. Chrom. A 888 (2000), 175.
[3]Pawlisch, C., Macris, A. and Laurence, R., Macromolecules 20 (1987),
3 30 74.39 1564.
[4]Goss, K.U., J. Coll. Interf. Sci. 190 (1997), 241.
3 30 76.44 [5]Davis, T.C. et al, Anal. Chem. 38 (1966), 241.
[6]Thielmann, F. and Butler, B., SMS Application Note 202.
3 30 72.51

1 20 55.63

Head Office: United States Office:


Surface Measurement Systems, Ltd Surface Measurement Systems, Ltd, NA
th
5 Wharfside, Rosemont Road 2125 28 Street SW, Suite I
London HA0 4PE, UK Allentown PA, 18103, USA
Tel: +44 (0)20 8795 9400 Tel: +1 610 798 8299
Fax: +44 (0)20 8795 9401 Fax: +1 610 798 0334
Email: science@smsuk.co.uk Email: science@smsna.com

You might also like