Professional Documents
Culture Documents
SS/AX4CO_5
Single and Dual Input Analyzers
for Low-level Conductivity
■ Cost effective
– select one or two conductivity inputs or combine
conductivity and pH/Redox (ORP) in one analyzer
– integral PID controller (AX410)
■ Energy saving
– high visibility, backlit display with auto switch-off
function
7.00 pH
7.00 pH
Monitoring pH
3 Standard
Relay Outputs
7.00 pH 2 Additional
7.00 pH
Monitoring pH Relay Outputs
AX400 Outputs
2
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
Cell 1
Ratio = Cell 1
Cell 2
Inlet = Typically 1.5 to 1
Cation Column
Cell 2
Cation
Cell 1 Exchanger
Cell 2
Outlet
Boiler or
Boiler Feed-Water
Outlet
Reverse Osmosis
3
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
The single-point calibration facility enables local adjustment Bi-linear and logarithmic ranges effectively provide a split output
where on-site validation is required. facility. General monitoring is covered with the lower range while
still supplying excellent discrimination during high level
excursions, such as during an alarm state.
Combined Conductivity and pH
The AX416 version measures conductivity and pH with the same
Span
analyzer. The option board provides the capability to retransmit Value
pH, conductivity and both sample temperatures. Five 20
programmable alarms can be assigned to suit the needs of the 18
application, enabling even the most demanding requirements to 16
be met. 14
Analog Output (mA)
12
10
8
6 Breakpoint
4
2
0
0 10 20 30 40 50 60 70 80 90 100
2nd Conductivity Measurement
Value
4
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
Revalidation Facility
ABB provides a revalidation facility for the 2278 conductivity cell
where the cell can be returned periodically for re-certification of
the cell constant. This quick turnaround facility fulfills the
stringent certification requirements of today’s pharmaceutical
industry.
Recording USP<645>
Two standard current outputs, or four (optional), enable both
conductivity and temperature to be recorded externally as a
permanent record of compliance.
Process Temperature
0 to 1µS/cm to 0 to 10µS/cm,
Measurement Range 0.2 to 4µS/cm ✓ fully user configurable.
5
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
Specification
6
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
Configuration
Relay Outputs – On/Off Can be assigned to either measured variable or
Number of relays either sample temperature
Three supplied as standard or five with option board fitted
Controller action
Direct or reverse
Proportional band
0.1 to 999.9%, programmable in increments of 0.1%
Derivative
0.1 to 999.9s in increments of 0.1s –
only available for single set point control
Auto/Manual
User-programmable
7
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
…Specification
Mechanical Data
Wall-/Pipe-mount versions Safety
IP66/NEMA4X General safety
Dimensions 192mm high x 230mm wide x 94mm deep EN61010-1
(7.56 in. high x 9.06 in. wide x 3.7 in. deep) Overvoltage Class II on inputs and outputs
Weight 1kg (2.2 lb) Pollution category 2
Panel-mount versions
IP66/NEMA4X (front only)
Languages
Dimensions 96mm x 96mm x 162mm deep Languages configurable:
(3.78 in. x 3.78 in. x 6.38 in. deep)
English
Weight 0.6kg (1.32 lb)
French
Cable Entry Types
German
Standard 5 or 7 x M20 cable glands
Italian
North American 7 x knockouts suitable for 1/2 in. Hubble gland
Spanish
Power Supply
Voltage requirements
85 to 265V AC 50/60 Hz
24V AC or 12 to 30V DC (optional)
Power consumption
<10VA
Insulation
Mains to earth (line to ground) 2kV RMS
8
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
Overall Dimensions
Dimension in mm (in.)
210 (8.27)
192 (7.56)
R1 96 (3.78)
(0. 0
4)
61mm (23/8in.) OD
Vertical or Horizontal
Post
150 (5.9)
175 (6.9)
192 (7.56)
.2 0
(0 6.5
6)
25 (1) Ø
23 (0.9)
94 (3.7)
25 (1)
22.5 22.5
(0.88) (0.88)
45 (1.77) 45 (1.77)
Dimensions in mm (in.)
91.60 (3.60)
91.60 (3.60)
96 (3.77)
5.40 (0.21)
Panel-mount Version
9
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
Electrical Connections
Terminal Block A
Terminal Block C
(Analog Option Board)
Terminal Block B
Terminal Block C
(Analog Option Board)
Common
Third Lead
Temperature
Compensator
10
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
Ordering Information
Single and Dual Input Analyzers for Conductivity and pH
AX4 X X X X X 0 X
Models AX410, AX411 & AX416
First Process Variable (PV1)
Conductivity 0 to 10,000µS/cm 1
Conductivity 0 to 1,999mS/cm 3
Conductivity USP<645> 5
pH/Redox (ORP) 6
Dissolved Oxygen 8
Second Process Variable (PV2)*
No second process variable – select for PID control of PV1 0
Conductivity 0 to 10,000µS/cm 1
Conductivity 0 to 1,999mS/cm** 3
Conductivity USP<645> 5
pH/Redox (ORP) 6
Dissolved Oxygen 8
Enclosure Types
Wall-mount IP66/NEMA 4X, General
Wall-mount – cable glands fitted 1
Pipe-mount 2
Wall-mount NEMA 4X/IP66, North American
Wall-mount 6
Pipe-mount 7
Panel-mount, Universal
Panel-mount** 5
Advanced Functions & Communications
Basic (2 current outputs + 3 relays) 0
Advanced (4 current outputs + 5 relays + logbook) 1
Profibus DP, basic (2 current outputs + 3 relays)** 2
Profibus DP, advanced (4 current outputs + 5 relays + logbook)** 3
Power Supply
85 to 265V AC 45 to 65Hz 0
24V AC/12 to 30V DC (auto-selection) 1
85 to 265V AC 45 to 65Hz (tropicalized)† 2
24V AC/12 to 30V DC (auto-selection, tropicalized)† 3
Reserved 0
Manual
English 1
French 2
Italian 3
German 4
Spanish 5
* When ordering units with a second process variable (PV2) the code digit for PV2 in the order code number must be equal to, or greater than,
the code digit for PV1, e.g. AX416 permissible, AX461 not permissible.
** Profibus DP not available in panel-mount housing when 0 to 1,999mS/cm is selected as Second Process Variable (PV2).
† Check factory for availability of tropicalization.
11
Single and Dual Input Analyzers for Low-level Conductivity
Models AX410, AX411, AX416, AX450 and AX455 SS/AX4CO_5
ABB has Sales & Customer Support The Company’s policy is one of continuous product
expertise in over 100 countries worldwide improvement and the right is reserved to modify the
information contained herein without notice.
www.abb.com Printed in UK (03.04)
© ABB 2004
Issue 5
SS/AX4CO