CL-226
m
ACCREDITED"
Calibralionl.llbora:01y

CALIBRATiON CERTIFICATE
Certificate Number: VCS2022-7 708
VC-RF-E-04, REV.N0:01,REV.DATE: MARCH 2022

Customer Name :ADVANCED CONSTRUCTION TECHNOLOGY SERVICES (ACTS)


Address : Doha Qatar
CSR Number : CSR/271/06 Delivery Date 23-Mar-2022
Date Received : 16-Mar-2022 Calibration Date 16-Mar-2022
Description : TRMS 600A AC Clamp Meter Suggested Due Date 15-Mar-2023
BrandI Make : EXTECH INSTRUMENTS Validity 1 year
Model I Type : EX650 Received Condition In Tolerance
Serial Number : R151201415 Returned Condition In Tolerance
Equipment ID Number : R1860 Calibration Status Pass
Size I Range : Multi-range Location : Vertex Lab
Resolution/Least Count : as per data sheet Temperature : 22.3 oc
Calibration Procedure : VC/CM/E/01-05 Humidity : 49.5 %RH
Page Number : 1 of 3
DETAILS OF MASTER IN STRUMENTS USED FOR CALIBRATION
DESCRIPTION ASSET NUMBER BRAND/MAKE MODEL/TYPE DUE DATE TRACEABILITY
Multi-Product Calibrator VER-CAL- MPC-001 Fluke 5522A 2-Mar-22 F7945016
50 Turn Current Coil VER-CAL-COIL-001 Fluke SSOOA 3-Sep-22 551220084290480
Barometric Pressure VER-CAL -BAR-001 Extech SD 700 15-Aug-22 551220084995741
Remarks:
1. Averages of minimum five readings are reported.
2. Suggested due date is given based on customer requirements.
3. These results are obtained at the time of calibration.
4. The user should determine the suitability of the instrument for its intended use.
5. Uncertainty of measurement at 95% confidence level by multiplying coverage factor k = 2.
6. Condition of instrument found satisfactory during receipt.
7. Our masters are directly calibrated through ISO/IEC 17025 accredited calibration lab having direct traceability with national 1 International
standard.
8. Vertex Calibration Labo ratory is accredited in accordance with the recognized International Standard as per ISO/IEC 17025:2017.
9. UUC- Unit Under Calibration
1O.Pass I Fail criteria is decided by comparing the error of the unit under calibration and its associated by measurement uncertainty with the
tolerance specified by the manufacturer I customer.
11.Accuracy, tolearance and acceptance criteia stated as per customer requierments.
12.Any hand written corrections (except @) or photocopies of the report invalidate this certificate.
13.Results reported outside the scope af accreditation are mark with asterisk(*).

q .lg('
Calibrated By: Approved By:
~ ), ~--- · ~----

~
.l'.+ t:.~' 1't't ;,,.,. . 1~ . c:_,.• ""
r. :·~• . · ~ '"i
ti t; ---':--~ . .~.:.;:;/J

Nimsi ldiculla J
c
't'o-f



~·~· .) ~:..>.J.!'I-'l ~~
c .R.No: 1254 85, Doha . Qatar J Ferdinand de Leon
Calibra tion En9ineer QA/QC Engineer
Cal ibration cycle are resu lting due dates were submitted/approved by the customer.Any number of factors may cause an instrument to drift out of to lerance before the next scheduled of
calibration.Reca li bration cycles should be based of frequency of used, environmental condition and customer's established systmatic accuracy. The information on this report, pertains only to the
instrument identified.
This certificate may not be reproduced other than in fu ll, except w ith the prior written approval of issuing laboratory. Th is report shall not be valid if not accompanied with Vertex Ca libration official seal.
\IERTEX CALl BRATION
- "'----~'.i
VERTEX CALIBRATION W.L.L
P.O Box: 1291, Office No 9, Second Floor,Building No 242,
C-Ring Road, Doha- Qatar
Tel: +974 44 44 11 26 I Fax: +974 40 17 2179 1Email : info@vertexcal.com
www.vertexcalibra t ion.com
. .~~''0'~'~;
tm
~~~i
l,,f, l n\1~'
m
ACCREDITED"
Calibration laboratory
Cl-226

CALIBRATION RESULT
Certificate Number: VCS2022-7 708
VC-RF-E-04, REV.N0:01,REV.DATE: MARCH 2022
Description: TRM S 600A AC Cla m p Meter Inw ard I CSR Number: CSR/271/06
Calibration Procedu re: VC/C M/E/01 -05 Calibration Date: 16-Mar-2022
Serial Num ber: R151 201415 Suggested Due Date: 15-Mar-2023
Eq uipment ID Num ber: R1860 Temperature: 22.3°C
Brand I M ake: EXTECH INSTRU M ENTS Hum idity: 49.5 %RH
M odel / Type EX650 Location: : Vertex Lab
Size I Ra nge M ulti-range Page Num ber: : 2 of 3
DC Voltag e Accuracy
Set Point Given Average Observed
Rang e Calibrated Deviation Tolerance Result Estimated Uncertainty
by Calibrator Reading by UUC
600mV 590.0mV 589.0mV -1.0mV ±6.4 mV
-· ±0.000150 mV
-- PA~~- _,.
6V 5.900 v 5.878 v -0.022 v ±0.0 v PASS - -r- ±0.000130 v
60V 59.00 v 58.89 v -0.11 v ±0.5 v PASS ±0.000100 v
--;
600V 590.0 v 588.6 v - 1.4 v ±S.OV PASS ±0.000100 v
-- 4
1000V 1000.0 v 998 v -2 v ±15.0 v PASS ±0.000100 v
AC Voltage Accuracy
Set Point Given Average Observed Estimated
Function Calibrated Deviation Tolerance Result
by Calibrator Read ing by UUC Uncertatinty _ __ J
@ 50Hz 5. 900 v 5.897 v -0.003 v ±0.1 v PASS ±0.000270- v I
@ 50 Hz 59.00 v 58.87 v -0.13 v ±0.8 v PASS ±0.000709 v
@ 50Hz 590.0 v 588.2 v -1 .8 v ±7.6 v PASS ±0.000150 v
-- 1 -
@50Hz 750 V 747V -3 v ±16.3 v PASS ±0.000649 v
DC Current Accuracy
Set Point Given Average Observed Estimated
Range Calibrated Deviation Tolerance Result
by Calibrator Reading by UUC Uncertatinty
30.00 A 30.21 A 0.21 A ±0.80 A PASS ±0.000760- A
60.00 A
60.00 A 60.35 A 0.35 A ± 1.55 A PASS ± 0.000300 A
300.0 A 302.6 A 2.60 A ±8.00 A PASS ±0.000290 A
600 .0 !
500.0 A 504.1 A 4.10A ±13.00 A PASS ±O.OOQ645 A
AC Current Accuracy
Set Point Given Average Observed Estimated
Function Calibrated Deviation Tolerance Result
by Calibrator Reading by UUC Uncertatinty
6 A@ 50Hz 6.00 A 6.058 A 0.06 A ±0.18 A PASS ±0.006940 A
60 A@ 5CHz 60.00 A 60.42 A 0.42 A ± 1.55 A PASS ±0.002310 A
300.0 A 303 .1 A 3.10 A ±8.00 A PASS ±0.007430 A
600 A@ 50 >-JZ
500.0 A 504.3 A 4.30 A ±13.00 A PASS ±0.002760 A
Capacitance Acmn::cy
Set Point Given Average Observed Estimated
Range Calibrated Deviation Tolerance Result
by Calibrator Reading by UUC Uncertatinty
999.9 nF 900.0 nF 901.5 nF 1.5 nF ±36.50 nF PASS ±0.005580 nF
99.99 11 F 90.00 11F 92.70 [J F 2.70 11F ±3.65 [JF PASS ±0.012750 11 F
999.9 [J F 900.0 [J F 917.6 [JF 17.6 IJF ±36.50 [JF PASS ±0.005240 [JF
9.999 mF 9.000 m F 9.138 mF 0.138 mF ±0.90 mF PASS ±0.005240 mF

l : .---- ---- · . '


;\ p , H~,; . ~.";( •.,t~
j 'o/ ·~·I '
~. 14'
,--i-· 1
L:..
. , •j . • .
JA.:l.J...'
_. .. ·: ·";fl
ht.~·- ...L·...a&:.~.id'

(;' f.~)~~;'-~£~~
" C.R.No : 125485, Doha- Qatar

Th is certifi cate ma; not be rep rodu ced othe r tha:1 in full, except with t he prior written appro... al of iss uing laboratory. Thi s report shall not be valid if not accompanied with Vertex Calibration officia l seal.
VERTEX CALIBRATION I
VERTEX CALIBRATION W.L.L
P.O Box: 1291, Office No 9, Second Floor,Building No 242,
C-Ring Road, Doha - Qatar
~
~~~f
"l"l"l· m
......~,,~,,~;,.

~'~-::,,

'-•;,,(;),~<·~
!

ACCREDITED"
Tel: +97444 4411261 Fax: +9744017 21791 Email: info@vertexcal.com Calibrat1 ~ l.abma~ory
Cl-226
www.vertexcalibration.com

CALIBRATION RESULT
Certificate Number: VCS2022-7 708
VC-RF-E-04, REV.N0:01,REV.DATE: MARCH 2022

Description: TRMS 600A AC Clamp Meter Inward I CSR Number: CSR/271/06


Calibration Procedure: VC/CM/E/01-05 Calibration Date: 16-Mar-2022
Serial Number: R151201415 Suggested Due Date: 15-Mar-2023
Equipment ID Number: R1860 Temperature: 22.3°C
Brand I Make: EXTECH INSTRUMENTS Humidity: 49.5%RH
Model /Type EX650 Location: Vertex Lab
Size I Range Multi-range Page Number: 3 of 3

Resistance Accuracy
Set Point Given Average Observed Estimated
Range Calibrat ed Deviation Tolerance Result
by Calibrator Read ing by UUC Uncertatinty
·- - ·- .
600.0 0 590.0 0 590.0 0 0.0 0 ±7.28 0 PASS ±0.000670 0
-- "·· ~

6.000 kO 5.900 kO 5.901 kO 0.001 kO ±0.06 kO PASS __ .L ±0.000150 kO


60.00 kO 59.00 kO 59.00 kO 0.00 kO ±0.61 kO PAS~ .. ~ ±0.000180 kO
600.0 kO 590.0 kO 590.0 kO 0.0 kO ±6.10 kO PASS ±0.000090 kO
5.900 MO 5.898 MO -0.002 MO ±0.07 MO
- .-
6.000 MO PASS ' ±0.000090 MO
-· - -- - j- ·-

60.00 MO 59.00 MO 58.99 MO -0.01 MO ±0.94 MO PASS ±0.000090 MO


***End of Calibration***

'Where statements of conformity are made in this report, the following decision rules are applied:
PASS - Rest.. its ± expanded uncertainty within limits I specification
FAIL - ResL:its ± expanded uncertainty exceeds limit I specification
IN DETERMINATE - Results ± expanded uncertainty overlaps limit I specification

9 5 % e)tpando d J 1 F'ai1
1
I
Foil
UI'PW s pe cifie d to-Jora nc a

~~~~~;~lt~

95 %
-l
M .oas urod va h.to __.
oxpnnd od unc e rtain ly i
I
P noo&
Pa!.\.5
FoUl
1 Uppot o cce ptunce lunlt

Nom i r11nl

Lowe r a c coptunc e hm lt
9& 'iotu~~~~c;~~ J
9 "Jard b and l-- - - - - - - - - - - - - - - - - - - - - Lower a p ocif io d to1en.'lnca

Calibrated By: Approved By:

(q,.lg<'
~
?t / r:-;·;.:;:::-~-:-; ~
Nimsi ldiculla 0 ...!:.~( i i "-' ' i.-. ~
) ... ,.,_ ' • "· •. Ferdinand de Leon
~~ -~~ ~---~ ·- . -:A~::£,-:\>1
" -. ..
Calibration Engineer
c
~

"
.. • t1 ~
~.~.J 6u':.)~"~ U"t-¥..H9
C.R.No : 125485, Doha· Qatar
... • QA/QC Engineer

Th is certificate may not be reproduced other than in full, except with the prior written approva l of issuing laboratory. This report sha ll not be valid if not accompanied with Vertex Calibration official seal.