CALIBRATiON CERTIFICATE
Certificate Number: VCS2022-7 708
VC-RF-E-04, REV.N0:01,REV.DATE: MARCH 2022
q .lg('
Calibrated By: Approved By:
~ ), ~--- · ~----
~
.l'.+ t:.~' 1't't ;,,.,. . 1~ . c:_,.• ""
r. :·~• . · ~ '"i
ti t; ---':--~ . .~.:.;:;/J
Nimsi ldiculla J
c
't'o-f
•
•
~·~· .) ~:..>.J.!'I-'l ~~
c .R.No: 1254 85, Doha . Qatar J Ferdinand de Leon
Calibra tion En9ineer QA/QC Engineer
Cal ibration cycle are resu lting due dates were submitted/approved by the customer.Any number of factors may cause an instrument to drift out of to lerance before the next scheduled of
calibration.Reca li bration cycles should be based of frequency of used, environmental condition and customer's established systmatic accuracy. The information on this report, pertains only to the
instrument identified.
This certificate may not be reproduced other than in fu ll, except w ith the prior written approval of issuing laboratory. Th is report shall not be valid if not accompanied with Vertex Ca libration official seal.
\IERTEX CALl BRATION
- "'----~'.i
VERTEX CALIBRATION W.L.L
P.O Box: 1291, Office No 9, Second Floor,Building No 242,
C-Ring Road, Doha- Qatar
Tel: +974 44 44 11 26 I Fax: +974 40 17 2179 1Email : info@vertexcal.com
www.vertexcalibra t ion.com
. .~~''0'~'~;
tm
~~~i
l,,f, l n\1~'
m
ACCREDITED"
Calibration laboratory
Cl-226
CALIBRATION RESULT
Certificate Number: VCS2022-7 708
VC-RF-E-04, REV.N0:01,REV.DATE: MARCH 2022
Description: TRM S 600A AC Cla m p Meter Inw ard I CSR Number: CSR/271/06
Calibration Procedu re: VC/C M/E/01 -05 Calibration Date: 16-Mar-2022
Serial Num ber: R151 201415 Suggested Due Date: 15-Mar-2023
Eq uipment ID Num ber: R1860 Temperature: 22.3°C
Brand I M ake: EXTECH INSTRU M ENTS Hum idity: 49.5 %RH
M odel / Type EX650 Location: : Vertex Lab
Size I Ra nge M ulti-range Page Num ber: : 2 of 3
DC Voltag e Accuracy
Set Point Given Average Observed
Rang e Calibrated Deviation Tolerance Result Estimated Uncertainty
by Calibrator Reading by UUC
600mV 590.0mV 589.0mV -1.0mV ±6.4 mV
-· ±0.000150 mV
-- PA~~- _,.
6V 5.900 v 5.878 v -0.022 v ±0.0 v PASS - -r- ±0.000130 v
60V 59.00 v 58.89 v -0.11 v ±0.5 v PASS ±0.000100 v
--;
600V 590.0 v 588.6 v - 1.4 v ±S.OV PASS ±0.000100 v
-- 4
1000V 1000.0 v 998 v -2 v ±15.0 v PASS ±0.000100 v
AC Voltage Accuracy
Set Point Given Average Observed Estimated
Function Calibrated Deviation Tolerance Result
by Calibrator Read ing by UUC Uncertatinty _ __ J
@ 50Hz 5. 900 v 5.897 v -0.003 v ±0.1 v PASS ±0.000270- v I
@ 50 Hz 59.00 v 58.87 v -0.13 v ±0.8 v PASS ±0.000709 v
@ 50Hz 590.0 v 588.2 v -1 .8 v ±7.6 v PASS ±0.000150 v
-- 1 -
@50Hz 750 V 747V -3 v ±16.3 v PASS ±0.000649 v
DC Current Accuracy
Set Point Given Average Observed Estimated
Range Calibrated Deviation Tolerance Result
by Calibrator Reading by UUC Uncertatinty
30.00 A 30.21 A 0.21 A ±0.80 A PASS ±0.000760- A
60.00 A
60.00 A 60.35 A 0.35 A ± 1.55 A PASS ± 0.000300 A
300.0 A 302.6 A 2.60 A ±8.00 A PASS ±0.000290 A
600 .0 !
500.0 A 504.1 A 4.10A ±13.00 A PASS ±O.OOQ645 A
AC Current Accuracy
Set Point Given Average Observed Estimated
Function Calibrated Deviation Tolerance Result
by Calibrator Reading by UUC Uncertatinty
6 A@ 50Hz 6.00 A 6.058 A 0.06 A ±0.18 A PASS ±0.006940 A
60 A@ 5CHz 60.00 A 60.42 A 0.42 A ± 1.55 A PASS ±0.002310 A
300.0 A 303 .1 A 3.10 A ±8.00 A PASS ±0.007430 A
600 A@ 50 >-JZ
500.0 A 504.3 A 4.30 A ±13.00 A PASS ±0.002760 A
Capacitance Acmn::cy
Set Point Given Average Observed Estimated
Range Calibrated Deviation Tolerance Result
by Calibrator Reading by UUC Uncertatinty
999.9 nF 900.0 nF 901.5 nF 1.5 nF ±36.50 nF PASS ±0.005580 nF
99.99 11 F 90.00 11F 92.70 [J F 2.70 11F ±3.65 [JF PASS ±0.012750 11 F
999.9 [J F 900.0 [J F 917.6 [JF 17.6 IJF ±36.50 [JF PASS ±0.005240 [JF
9.999 mF 9.000 m F 9.138 mF 0.138 mF ±0.90 mF PASS ±0.005240 mF
(;' f.~)~~;'-~£~~
" C.R.No : 125485, Doha- Qatar
Th is certifi cate ma; not be rep rodu ced othe r tha:1 in full, except with t he prior written appro... al of iss uing laboratory. Thi s report shall not be valid if not accompanied with Vertex Calibration officia l seal.
VERTEX CALIBRATION I
VERTEX CALIBRATION W.L.L
P.O Box: 1291, Office No 9, Second Floor,Building No 242,
C-Ring Road, Doha - Qatar
~
~~~f
"l"l"l· m
......~,,~,,~;,.
~'~-::,,
'-•;,,(;),~<·~
!
ACCREDITED"
Tel: +97444 4411261 Fax: +9744017 21791 Email: info@vertexcal.com Calibrat1 ~ l.abma~ory
Cl-226
www.vertexcalibration.com
CALIBRATION RESULT
Certificate Number: VCS2022-7 708
VC-RF-E-04, REV.N0:01,REV.DATE: MARCH 2022
Resistance Accuracy
Set Point Given Average Observed Estimated
Range Calibrat ed Deviation Tolerance Result
by Calibrator Read ing by UUC Uncertatinty
·- - ·- .
600.0 0 590.0 0 590.0 0 0.0 0 ±7.28 0 PASS ±0.000670 0
-- "·· ~
'Where statements of conformity are made in this report, the following decision rules are applied:
PASS - Rest.. its ± expanded uncertainty within limits I specification
FAIL - ResL:its ± expanded uncertainty exceeds limit I specification
IN DETERMINATE - Results ± expanded uncertainty overlaps limit I specification
9 5 % e)tpando d J 1 F'ai1
1
I
Foil
UI'PW s pe cifie d to-Jora nc a
~~~~~;~lt~
95 %
-l
M .oas urod va h.to __.
oxpnnd od unc e rtain ly i
I
P noo&
Pa!.\.5
FoUl
1 Uppot o cce ptunce lunlt
Nom i r11nl
Lowe r a c coptunc e hm lt
9& 'iotu~~~~c;~~ J
9 "Jard b and l-- - - - - - - - - - - - - - - - - - - - - Lower a p ocif io d to1en.'lnca
(q,.lg<'
~
?t / r:-;·;.:;:::-~-:-; ~
Nimsi ldiculla 0 ...!:.~( i i "-' ' i.-. ~
) ... ,.,_ ' • "· •. Ferdinand de Leon
~~ -~~ ~---~ ·- . -:A~::£,-:\>1
" -. ..
Calibration Engineer
c
~
"
.. • t1 ~
~.~.J 6u':.)~"~ U"t-¥..H9
C.R.No : 125485, Doha· Qatar
... • QA/QC Engineer
Th is certificate may not be reproduced other than in full, except with the prior written approva l of issuing laboratory. This report sha ll not be valid if not accompanied with Vertex Calibration official seal.