3.2.5 Insulation monitoring devices for 24 V
SELV circuits
SELVTESTER-24 is an earth insulation tester for
SELV24 V AC/DC circuits, particularly recom-
mended for installation in medical locations in
whiten ee v and ¢su vines exist. ine systemauic
and continuous monitoring of the low voltage line
in these environments is recommended by IEC
60364-7-710 precisely because a failure o” a short
circuit could transfer a potential of morethan 250
Vwith consequent damage to persons ard
‘things.
Characteristics
SELVTESTER-24 monitors the insulation rests-
tance of dedicated 24 V AC/DC circuits for power-
Ing a zelalyticiamp.
It is important to monitor the insulation of the
scialyticlamp because conductors couldbecome
detached and enter into contact with the metal
structure while the lamp is being manoeuvred.
SELVTESTER-26 measures the variation of the po-
tential ofthe two network polarities in relation to
‘ne earth, ana signals a decrease in resistance De-
low a predefined value, allowing immediate inter-
‘ception of the malfunction. Ina direct current sit-
uation itis also possible to discern the polarity
{involved in the failure. The output signal can be
carried remotely in rooms with the greatest medi-
cal staff presence thanks to remote signalling
panels called QSD-DIG 230/24.
‘The front of the SELVTESTER-24 Includes a test
pushbutton, status indicator and two LEDs for
alarms caused by low i
allow the activation threchold to be varied
(o....50 ka). The TEST pushbutton performs the
periodic correct operation test.
ation. Mieroswitehes
“echnical charactor ietce of SELVTESTER
Network voltage and auxiliary power supply
Max lose
Internal impedance
‘Activation threshold setting
‘Activation delay
Pav 50.60H/OC# 20%
BVA
50k0
‘adjustable 10+ $0 kA (levels by means of miroswitehes)
approx ts
‘Signale
‘output
Operating temperature
Storage temperature
Operating temperature
‘cross section of terminals
Protection deoree
Weight
Reference standards
SELVTESTER-24 Ordering Codes
For max. 2 @50-016230/24 remote panels Max. 24 V1.A
20+ 60°C
20+ 70°C
595%
2SkV 608 / AKV imp. 12/50 8
Ip front panel with cover /1P20 container
DIN modules
200 9 approx.
Monitored network
SFLv insulated tine 28 VACTOO