Slurry sampling

Slurry sampling for on-line


analysis and accounting
Dependable on-line analyzer measurement An integral part of sampling, often
Outokumpu Technology of slurry particle size and composition is overlooked, is the sample transport and
sampling systems provide
not possible directly from a large process delivery system to and from an analyzer.
• High availability
flow. Analyzers require a representative Outokumpu Technology has over 50
• Low maintenance
sample from most process streams. years of experience in sampling and can
Composite samples are collected for offer an extensive range of slurry samplers
Advantages
laboratory analysis, whose results are used for practically all applications. The company
• Representative sample for on-line
analysis
in metallurgical balance accounting and offers design and engineering of the
• Representative sample for
product quality assurance. complete sampling system and consulting
composite analysis for its customers.
Representative sampling

Homogeneity and mixing of


process flows
All material in a process flow must have the
same probability of entering the sample.
Turbulent flows are required in typical
process streams to maintain the slurry in
suspension (speed >1.5 m/s). Turbulence
keeps the slurry well mixed in contrast to
dry solids on a conveyer belt. The finer
and lighter the solids in the slurry, the
more ideal is the mixing. Gravity begins to
play a significant role, when particles are
above 100 µm in size and/or are of high
SG material. Mixing is still very effective
in the horizontal direction. Such mixing
enables the use of samplers with a simple
robust fixed cutter to acquire a representative
sample.

Sampling for on-line


analysis
An acquired sample must be representative restricted sample flow, cannot produce a See the big picture
of the characteristic of the process stream sample suitable for accounting purposes. • For what purpose is the sample required?
being measured. Therefore, broad-range • What is the source of the slurry to be
particle size analysis requires a representative Isokinetic sampling sampled and what is happening there?
sample in respect to the particle size, while For the correct representation of all particle • What is the destination of the process
stream and sampled material?
elemental analysis has to be representative sizes, the flow velocity of slurry into a cutter
• Are any future changes expected to be
in respect to the measured elemental or nozzle must be about the same as the in the process e.g. a significant increase
composition. velocity of the bulk flow around the cutter or decrease in process flows?
In vertical pressure pipes, a round or nozzle. If the sample flow into the cutter • Does the material being sampled contain
nozzle in the middle of the pipe can be or nozzle is too high, the sample will be significant amounts of trash, rocks,
grinding balls or acid?
used to extract a sample, the flow being biased with fine material. Differences of
• Pay attention to the sampler location
thoroughly mixed after a pump or a ±1 m/s are permissible for slurries of normal and correct installation, since they can
suitable length of vertical pipe. (i.e. longer particle size range. have a significant effect on sample
representativity.
than 10 x pipe’s diameter).
In horizontal gravity pipes or launders, Sampling design
vertical cutters are used to take an equal First select a location from which to take a
slice of all horizontal layers, which might sample based either on gravity or pressure
have different contents due to segregation flow, although sometimes both options are
by gravity. available. Closeness to the analyzer location
is an advantage. Information on the process
Metallurgical accounting pipe, flow rates and the amount of sample
Accounting samples must represent the required, determines the sampler type.
variations that occur in both process Sample line routing and sizing are essential
flow and solids content, in addition to for proper operation and high availability of
requirements for on-line analysis. Strictly the analyzer system.
speaking, a pressure pipe sampler with

2
Primary samplers for pressure flow

When the process stream to be sampled is Advantages


under pressure, the primary sample flow is • Gives representative sample from well
determined by the balance between sampler mixed process flow
pressure head and sample pipe friction. • In most cases, no extra pump is required
to move sample to the analyzer or
Standard size nozzles are used in samplers. composite sampler
Isokinetic samples are obtained by correct Limitations
sample pipe diameter, length and routing. • Sample flow rate is not directly
proportional to process flow for
composite sampling

Representative sampling of process flows


under pressure is possible when the sample
pipe diameter, length and routing are correctly
designed.

A PSA pressure pipe sampler with an optional ASA sector samplers are used to divide
automatic VSA valve set for flushing. PSA small pressure flows. A sector sampler can
samplers are typically used at pump outlets. be delivered to take 25, 50 or 75 % of the
process flow.

HCS sector samplers are used to divide small


horizontal flows. One primary sample flow
can be divided between two analyzers using a
HCS sampler.

SPSA straight pressure pipe sampler. Where A SPA suction pipe sampler for sampling a
it is inconvenient to use a PSA sampler in flotation cell discharge box, with an automatic
conventional vertical position. VSA valve set for flushing.

Process location Flow-rate Pipe size Sampler type Note Sample use

After pump or > 600 l/min >100 mm PSA pressure pipe sampler Analysis
long vertical pipe (or SPSA) Composite
After pump or < 600 l/min ≤100 mm ASA sector sampler Can take 25, 50 or 75 % Analysis
long vertical pipe of flow Composite
Horizontal pipe, > 500 l/min >100 mm SPSA straight pressure pipe Can also be used for Analysis
no froth sampler vertical pipes
Horizontal pipe < 500 l/min ≤100 mm HCS sector sampler Can split sample Analysis
between analyzers Composite
Flotation cell All N/A SPA suction pipe sampler Analysis
discharge box

3
Primary samplers for gravity flow

When the process stream is essentially at Advantages


ambient pressure, a vertical sample cutter • Takes a representative sample from a
can be used. The cutter opening must vertically segregated process flow
be many times larger than the maximum • The sample flow rate of single stage
samplers is directly proportional to
particle size of sampled slurry, at least 8 mm, the process flow for composite and
but preferably >20 mm. The gravity primary accounting sampling
sample flow is not restricted, ensuring Limitations
isokinetics at the sampling point. The • A pump may be required to transport A LSA vertical cutter sampler with an optional
sample flow can be adjusted using an the sample to the analyzer or composite VSA automatic flushing valve set for process
sampler flows at ambient pressure. A cyclone overflow
adjustable cutter opening for pipe sizes up
is a typical LSA sampling point.
to 400 mm.
The sampler box design improves
process flow mixing prior to the sample
cut. When representative isokinetic
sampling is applied, high process flows
produce large primary sample flows,
which requires many sampling steps or
innovative sampling designs.

NLA non plug vertical cutter sampler with


automatic mechanical cutter cleaner for
ambient pressure process flow containing
trash. Victaulic process pipe connections.

Process location Flow-rate m³/h Pipe size Sampler Note Sample use

Nearly horizontal 25–1000–(1600) ≤600 mm LSA or NLA vertical cutter Flow >1000 m³/h Analysis
gravity pipe sampler requires a partly Composite
blanked cutter Accounting
Gravity flow in <1000–2000 N/A SKA vertical sample cutter Analysis
launder Composite
Accounting
Nearly horizontal >500 N/A TSC two stage vertical Two Stage Cutter Analysis
gravity flow cutter sampler allows broader cutter Composite
opening than LSA/NLA
Nearly horizontal >350 ≥250 mm CPS primary sampler Two stage sampler. Analysis
gravity pipe or Requires control from Composite
launder, no froth analyzer. Restricted
sample flow.
Horizontal pipe < 30 ≤100 mm HCS sector sampler Can split sample Analysis
or LSA 80 between analyzers Composite
Accounting
ASD 50 adjustable sample
Sample line 6–18 50 mm Reduces primary Analysis
divider
sample flow and splits Composite
sample Accounting

4
Sample flow from vertical cutters should not
be restricted.

A SKA vertical cutter sampler for process An ASD adjustable sample divider located
flows under ambient pressure in a launder. on top of a PSI 500™ analyzer multiplexer,
allowing a large primary sample flow to be
used for easily settling slurry.

Two stage samplers for large process flows


Single stage sampler produces too high sample flow-rate from a large process
flow. Two stage samplers are able to provide a suitable sample flow-rate for
on-stream analysis.

A TSC two stage vertical cutter sampler for A CPS two stage sampler for high speed
large process flows under ambient pressure. process flows. The sample entering the tube
The sampler can be installed at a launder is re-sampled using a controlled syphon
bottom or in a LSA style sampling box. sampler.

5
Metallurgical accounting sampling

MSA Multi-stage
Metallurgical Accounting
Sampler
Operation of the sampler is based on a
combination of static multiple cutter
stages (depending on the process flow),
followed by a moving crosscut sampler
stage, which produces the sample. The
MSA samplers can handle process flows
from 60–18,000 m³/h. A MSA 1/40 metallurgical sampler.
The MSA sampler can be configured
to deliver a composite sample or a
continuous flow of about 150 l/min, Model Nominal Stages Number Main Sample use
which is suitable for Courier® and PSI™ capacity m³/h of static launder
cutters width mm
analyzers. Analyzer controlled sampling by
MSA 1/40 60–180 1 none 400 Analysis
the MSA samplers is also possible. Composite
Accounting
MSA 2/50 180–600 2 3 500 Analysis
Composite
Accounting

MSA 2/80 600–1800 2 4 800 Analysis


Composite
Accounting
MSA 3/140 1800–6000 3 5/3 1400 Analysis
Composite
Accounting
MSA 3/200 6000–18000 3 5/4 2000 Analysis
Composite
Accounting

Composite sampling
and filtering
The final composite sample is taken from
the 150 l/min primary sample flow by the
secondary sampler in a Courier® on-stream
analyzer system (see figure on left) or a
PSI™ particle size analyzer. A SSA
composite sampler can be used to collect
composite sample after primary sampler
instead of an on-stream analyzer.
The composite sample can be filtered
using an optional VFU vacuum sample
filter unit. The VFU is self-contained,
equipped with a vacuum blower and
uses standard filter papers.
A composite sampler SSA 50 with sample filter
Composite sampling by a Courier® analyzer and vacuum blower to sample 50–400 l/min
multiplexer. flow.

6
Controlled sampling

Controlled primary sampling


On-stream analyzers have traditionally used
a continuously flowing primary sample,
where the sample is always available for
filling the multiplexer tank prior to assay.
In this case the analyzer has no control over
the primary sample flow.
Controlled sampling is a built-in
feature in new Outokumpu Technology
analyzers. Samples can be taken when
required for analysis or composite sampling.
This reduces the amount of material to
be returned to process and improves the
availability of the sample lines. Controlled
primary sampling entails that the analyzer
plays an active role in controlling the
primary sample acquisition and delivery
to the analyzer. Remote controlled sampler
valve sets are required and the desired
sampling sequence has to be configured.
A combination of continuous and
controlled sample lines can be used in the
same analyzer system.

Options
On-demand sampling
Flanges: DIN 2576, AS 2129–TABLE D, SABS 1123–TABLE 10,
The primary sample is controlled so that it ANSI B16.5–150 lbs, Victaulic
only flows when required for measurement
Lining: Natural rubber as standard other materials as an option
and/or composite sampling. When the
Cutters: Wear areas rubber lined, hard metals and ceramics as options
sample flow is stopped after measurement
Sample valves: Manual and automatic flush valves are available
or composite sampling, the delivery line is
flushed with water. This radically improves
the availability of samples and reduces
return pump volumes. Sampling, sample presentation and
patent applications are covered by
several patents owned by Outokumpu
Automatically flushed
Technology Oy.
sample lines
The primary sample usually flows Courier® is a registered trademark of
continuously but is interrupted at Outokumpu Technology Oy.

programmed intervals to flush the sampler


Specifications subject to change
nozzle or cutter and the sample line. The without notice. Please see sampler
flushing sequence is synchronized so as data sheets and installation drawings
not to interrupt normal measurements or for dimensions and details.

composite sampling.
© 2006 Outokumpu Technology Oy.
All rights reserved.

7
3020EN. Helsinki, Finland. July 2006.

Outokumpu Technology is a worldwide


technology leader in minerals and metals
processing, providing innovative tailored
solutions for a wide variety of customer
needs in iron and steel, aluminum and
non-ferrous metals industries. With a
global network of sales and service centers,
research facilities and 1800 experts the
company generates annual sales of some E-mail: technology.automation@outokumpu.com

600 million euro.

You might also like