Professional Documents
Culture Documents
Table of
Contents
Safety
3 Operation
4 Periodic
Maintenance
5 Calibration
& Adjustments
6 Component
Replacement
7 Trouble-
shooting
8 Electrical
Schematics
P/N 8000-ANTHEM-U
Revision F, October 10 2016 9 Illustrated
Parts List
Copyright © 2016, Del Medical Inc. All rights reserved.
This document is the property of Del Medical Inc. and contains confidential and proprietary information
owned by Del Medical Inc. Any unauthorized copying, use or disclosure of this document without the prior
written permission of Del Medical Inc. is strictly prohibited.
Revision History
Chapter 1 — Introduction
1.1 Document Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1.2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
1.3 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
1.3.1 General Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
1.3.2 Anthem Generator Models and Options . . . . . . . . . . . . . . . . . . . . . . 1–3
1.3.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–3
1.3.4 Architecture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–3
1.3.5 X-Ray Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
1.3.6 Power Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
1.3.7 In-Table Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
1.4 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–5
1.5 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–7
1.5.1 Power Specifications, by Anthem Model. . . . . . . . . . . . . . . . . . . . . . 1–7
1.5.2 General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–7
1.5.3 AEC Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–8
1.5.4 System Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–9
1.6 X-Ray Tube Compatibility Listing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–9
1.7 Identification Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–10
1.7.1 Generator Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–10
1.7.2 Equipment Label Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–11
1.8 Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–13
1.9 Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–14
1.9.1 Prep/Prep Ready Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–14
1.9.2 Exposure Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–14
1.9.3 kV and mA Control Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–14
1.9.4 Rotor Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–15
iii
Anthem Generator Installation, Operation & Service Manual
Chapter 2 — Installation
2.1 Pre-Installation Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
2.1.1 X-ray Room Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–1
2.1.2 Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–2
2.1.3 Power Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–3
2.1.4 Wiring Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–4
2.1.5 Shipment Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–6
2.1.6 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–6
2.1.7 In-Table Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–7
2.1.8 Installation Tools Required: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–7
2.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–8
2.2.1 Cable Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–8
2.2.2 Connection of Incoming Main Power Cable . . . . . . . . . . . . . . . . . . 2–10
2.2.3 Connecting Components to the Generator . . . . . . . . . . . . . . . . . . . 2–14
2.2.4 Control Console Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–25
2.2.5 High Tension (HT) Cable Connections . . . . . . . . . . . . . . . . . . . . . . 2–26
2.3 Initial Testing and Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–29
2.3.1 Test Equipment Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–30
2.3.2 Inspection of High Tension Transformer. . . . . . . . . . . . . . . . . . . . . 2–30
2.3.3 Inspection of Power Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–30
2.3.4 Inspection of Control Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–31
2.3.5 Operational Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–32
2.3.6 Verify the Door Interlock Safety Function. . . . . . . . . . . . . . . . . . . . 2–37
2.3.7 The Service Mode Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–37
2.3.8 Installing the Del APR Utility Software . . . . . . . . . . . . . . . . . . . . . . 2–44
2.3.9 Initial Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–45
Chapter 3 — Operation
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3.2 Safety Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
3.3 Operator Control Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–2
3.3.1 Control Console Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–2
3.3.2 Multifunction Buttons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3
3.3.3 Data Knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3
3.3.4 Prep and Expose buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3
3.3.5 Display Parameters and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3
3.4 Power On/Off Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–6
3.4.1 Powering Up The Anthem Generator . . . . . . . . . . . . . . . . . . . . . . . . 3–6
3.4.2 Powering Down the Anthem Generator . . . . . . . . . . . . . . . . . . . . . . 3–6
iv
Anthem Generator Installation, Operation & Service Manual
v
Anthem Generator Installation, Operation & Service Manual
Chapter 7 — Troubleshooting
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
7.2 Discharge Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
7.3 Troubleshooting Index & Charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–1
vi
Safety Information
Introduction
Del Medical X-ray equipment meets high standards of performance and
reliability. We enforce strict quality control techniques to eliminate the
potential for defects and hazards in our products.
This equipment provides an adjustable X-ray source for acquiring images of
parts of a patient’s anatomy. Use of this equipment in any other fashion may
lead to serious personal injury.
These safety guidelines are intended to educate the operator on safety issues
to operate and maintain the Anthem Generator in a safe manner.
Warning
Read and fully understand this manual before operating this
equipment. Failure to follow operating instructions can result in
serious injury or death.
Statement of Liability
To prevent excess radiation exposure (primary or secondary) to patients or
staff, this equipment must be operated and serviced by trained personnel
knowledgeable in the required safety precautions. While this equipment is
designed for safe use, improper operation can cause serious injury or
equipment damage. The manufacturer or its agents and representatives
assume no responsibility for the following:
1. Injury or danger to any person from X-ray exposure.
2. Overexposure of patients or staff to X-radiation due to improper
techniques or procedures.
3. Injury or danger from improper use of the adjustable column or tube head
assembly.
4. Problems or hazards resulting from failure to maintain the equipment as
specified in this manual.
5. Damage or injury caused by equipment that has been modified or used in
any application other than that intended by the manufacturer.
vii
Anthem Generator Installation, Operation, & Service Manual
Warning Statements
• indicate hazards or unsafe practices that can cause severe injury or death.
• appear in bold type.
• have a triangular symbol enclosing an exclamation point, with the word
Warning.
• precede the step or piece of information to which they refer.
• appear as the following example:
Warning
This text describes special safety precautions to avoid unsafe practices
that can result in severe injury or death.
Caution Statements
• indicate hazards or unsafe practices that can cause minor injury, or
product or property damage.
• appear in bold type.
• have a triangular symbol enclosing an exclamation point, with the word
Caution.
• precede the step or piece of information to which they refer.
• appear as the following example:
Caution
This text describes special safety precautions to avoid unsafe practices
that can result in personal injury or product or property damage.
viii
Anthem Generator Installation, Operation, & Service Manual
ix
Anthem Generator Installation, Operation, & Service Manual
Warning
To avoid serious electric shock, disconnect all generator power and
follow the discharge procedure before servicing, disconnecting high-
tension cables, or removing access panels. Avoid dropping any tools
or objects in the unit.
Radiation Protection
Serious detrimental health effects can result from exposure to both short-
term/high-level and long-term/low-level X-rays. Become familiar with the
effects of both types of radiation exposure, and take appropriate measures to
minimize exposure. Some effects of X-radiation are cumulative over a period
of months or years. Most importantly, avoid exposure to the primary beam at
all times.
X-ray facilities typically use lead to shield the radiologist's workstation. The
typical workstation partition is generally effective in reducing residual
radiation to a safe level. Windows in the partition are typically glazed with
lead glass and provide effective protection. Do not leave the protected area
during exposure.
Any object in the path of the primary beam produces scattered radiation,
which can result in a substantial radiation dose to unprotected persons. To
minimize exposure to scattered radiation, use movable lead screens, and
lead-impregnated gloves and aprons. These protective devices must contain a
lead (or equivalent) thickness of at least 0.25 mm. Use such protective devices
for all operators, observers, and/or servicing personnel exposed to radiation
fields of five or more milli-Roentgens per hour.
The X-ray system produces X-rays only when high voltage is applied to the
X-ray tube. When the high voltage is removed, X-ray emission ceases
immediately.
When the distance between the X-ray tube focal spot and the patient’s skin is
decreased, the radiation dose to the patient increases.
Ensure audio and visual communication between operator and patient even
during exposure. This may require video cameras or an intercom.
x
Anthem Generator Installation, Operation, & Service Manual
To minimize risk, ensure that all operating personnel are educated in the
hazards of radiation and understand the safety requirements for X-ray
operation. Review the manuals for each component in the system.
Note: This system may include an anti-scatter grid. For pediactric applications, the
grid can be removed by a qualified service technician. Refer to the service manual for
the pertinent table or wall stand.
Warning
Ensure exposure parameters are properly adjusted to within safety
limits.
Caution
Incorrectly positioning the X-ray tube and collimator can cause the
X-ray field to be misaligned, resulting in an unacceptable image and
the need to re-expose.
xi
Anthem Generator Installation, Operation, & Service Manual
Hazardous Materials
The X-ray tube, PCBs, and collimator used with this equipment contain lead
and oil. Refer to the equipment manuals or contact Del Medical for additional
information regarding use and disposal of hazardous materials.
Restrictions on Use
The hardware specified for intended applications of the Anthem Generator
has been selected, tested, and verified by Del Medical to meet applicable
regulatory agency requirements for those countries where it is offered for sale
with respect to its intended applications.
Warning
Do not install components or accessories not intended for use with
this system. Failure to comply can cause injury or equipment damage.
You are responsible for ensuring that the application and use of this
equipment does not compromise the patient contact rating of any equipment
used in the vicinity of or in conjunction with the system.
xii
Anthem Generator Installation, Operation, & Service Manual
Grounding
Provide protective earth grounding for all components. All associated
components must be grounded to a common bus according to national and
local electrical codes. Ensure that the earth grounding connections between
this equipment and the host system are maintained at all times.
xiii
Anthem Generator Installation, Operation, & Service Manual
xiv
1 Introduction
Introduction
1
1.1 Document Information
This manual provides installation, operation, calibration and service
information for the Anthem Generator. For specific information regarding the
operation and service of the tube, collimator, stands, tables, and other
components, see their respective manuals.
1.2 Introduction
The Anthem generator is a high-frequency, microprocessor-based X-ray
generator designed for general radiographic applications. Features include a
user-friendly operator’s Control Console, programmable APR operation, and
optional support of Digital Radiography and Automatic Exposure Control.
All persons operating or servicing this equipment must read this manual
before performing tasks. You must have a thorough understanding of the
Anthem generator and its proper use before you make any radiographic
exposures.
1 Introduction 1-1
Anthem Generator Installation, Operation, & Service Manual
1.3 Description
The DR-ready Anthem generator is intended for use as part of radiographic
stationary X-ray systems. As part of a typical system, the generator provides
power and control for other parts of the system, as shown in Figure 1-2.
X-ray
tube
high collimator
voltage
supply voltage ion image
Anthem chamber receptor
generator (AEC)
table/stand,
interlock functions, or
Control DR prep/expose
Console interface
PC or DR
Workstation
1-2 1 Introduction
Anthem Generator Installation, Operation, & Service Manual
1.3.3 Operation
The Anthem generator operates in one of four modes:
• APR (Anatomically Programmed Radiography). Select from pre-
programmed techniques based on body region, examination type, and
patient size. You can then modify mA, kV, and/or time parameters and
save these customized settings.
• Manual 2 point. Set mAs and kV parameters. The Anthem software
automatically chooses the highest mA and shortest time to reduce motion
and deliver high quality images.
• Manual 3 point. The traditional 3 point selection of kV, mA and time
accommodates the needs of specialized views and the preference of some
technologists.
• DR (Digital Radiography). This optional mode replaces the APR mode.
Use the DR mode when interfacing a DR workstation and DR detector
devices. Technique parameters on the DR workstation also display at the
Control Console in real time.
The Anthem generator Control Console can also be placed in the following
support modes:
• Administrator Mode. Use this mode to set default parameter values.
• Service Mode. Use this mode to set options, configure the system, and
perform calibration.
1.3.4 Architecture
The X-ray generator consists of an upper main power cabinet, a lower cabinet,
and a control console.
1 Introduction 1-3
Anthem Generator Installation, Operation, & Service Manual
The upper main power cabinet contains the inverter, the filament drivers, the
power supply, the main control board, and interface connections to the room
equipment.
The lower cabinet contains the HT tank, and the optional line match
transformer.
The Control console allows the operator to select/program technique factors,
configure features, and prep/initiate an X-ray exposure.
1-4 1 Introduction
Anthem Generator Installation, Operation, & Service Manual
1.4 Dimensions
17.5”
444 mm
28”
711 mm
19.5”
495 mm
27”
686 mm
Caution
Do not drill equipment mounting holes in concrete floors before
determining the location of tensioned wire ropes. Consult with
customer or responsible project engineer to locate and avoid drilling
through wire rope. Cutting through tensioning cables can cause severe
structural damage.
1 Introduction 1-5
Anthem Generator Installation, Operation, & Service Manual
Ø 0.281” (x4)
Ø 7 mm (x4)
2.75”
70 mm
4.0”
102 mm
Figure 1-4. Optional control console wall bracket mounting hole dimensions
36.5”
930 mm
8.3”
211 mm
Anti-Tip
11.0” Bracket
230 mm
1-6 1 Introduction
Anthem Generator Installation, Operation, & Service Manual
1.5 Specifications
1 Introduction 1-7
Anthem Generator Installation, Operation, & Service Manual
D.H.H.S. Compliance
The Anthem High Frequency x ray generator and its associated accessories
are certified components conforming to all applicable radiation performance
standards as outlined by the Food and Drug Administration, Code of Federal
Regulations, Title 21, Subchapter J, at the time of manufacture.
Regulatory Conformance
Complies with FDA radiation performance standards under 21 CFR
subchapter J, in effect at the date of manufacture.
1-8 1 Introduction
Anthem Generator Installation, Operation, & Service Manual
1 Introduction 1-9
Anthem Generator Installation, Operation, & Service Manual
1-10 1 Introduction
Anthem Generator Installation, Operation, & Service Manual
Alternating Current
1 Introduction 1-11
Anthem Generator Installation, Operation, & Service Manual
The product is RoHS compliant. RoHS marks can vary greatly from
one product to the next.
1-12 1 Introduction
Anthem Generator Installation, Operation, & Service Manual
1 Introduction 1-13
Anthem Generator Installation, Operation, & Service Manual
1-14 1 Introduction
Anthem Generator Installation, Operation, & Service Manual
configurations of the capacitors are used to vary the frequency of the high
voltage primary switching circuit, providing optimum operation for various
power output levels.
The Main Control board sends the kV drive signal to the IPM Driver board to
generate the high frequency AC input to the H.V. tank.
On the Main Control board, the feedback voltages from H.V. tank are
summed, filtered, and used to adjust the pulse width per tube kV feedback or
to detect under voltage or over voltage conditions that may require an error
shutdown.
The mA or mAs settings provide the mA and Filament reference voltage to
the mA and Filament control circuit.
1.10 Abbreviations
AEC Automatic Exposure Control
AP Anteroposterior
APR Anatomically Programmable Radiography
AWG American Wire Gauge
BTU British Thermal Unit
CAN Controller Area Network
CDRH Center for Devices and Radiological Health
CE Communautés Européennes
DAC Digital to Analog Converter
DFMTS Digital Floor Mounted Tubestand
DR Digital Radiography
ga Gauge
hPa Hecto Pascal
HT High Tension
ICRP International Commission on Radiological Protection
inHg Inches Mercury
IPM Intelligent Power Module
kN kiloNewton
kV kilovolt
kVp peak kilovoltage
kW kilowatt
1 Introduction 1-15
Anthem Generator Installation, Operation, & Service Manual
LAT lateral
mA milliampere
mAs milliampere second
MeV mega-electron Volt
NCRP National Council on Radiation Protection and Measurements
NIST National Institute for Standards and Technology
OBL oblique
PA posteroanterior
PBL Positive Beam Limitation
PID Patient ID
PN Patient Name
RAO Right Anterior Oblique
UL Underwriters Laboratories
1.11 Glossary
Automatic Exposure Control (AEC) — AEC uses an ionization chamber or a
solid-state detector within the Bucky to command the X-ray generator to stop
the exposure when the correct image density is reached.
AEC field — Commonly known as ionizing vector, the AEC field measures the
dose of X-ray in the interested region.
Angle of tube rotation — This is the angle of deviation from the vertical
direction when the tube rotates, and ranges from -180 to 180 degrees.
Bucky/Grid Cabinet — The component that includes a film cassette, AEC,
moving or stationary grid, and related components. In a Digital Radiology
system, the Bucky contains a Digital Radiology detector instead of a film
cassette.
Collimator — Device that regulates the size and shape of the X-ray beam to
accurately localize the area of interest on the patient, while reducing overall
patient irradiation exposure.
Free Mode — Exposing position independent of a table or wallstand.
Generator — Device that supplies power to and controls the X-ray tube.
N — Newton, the amount of the net force required to accelerate a mass of one
kilogram at a rate of one meter per second.
kVp — Peak kilo-volts, the highest energy of X-rays emitted by an X-ray tube
(equal to the peak applied tube voltage).
LAT — Lateral view position for X-ray exposure.
Longitudinal — Head-to-foot positioning of focal spot.
mAs — Milliampere-seconds. Combined with kVp, it indicates the dose of
X-rays.
1-16 1 Introduction
Anthem Generator Installation, Operation, & Service Manual
1 Introduction 1-17
Anthem Generator Installation, Operation, & Service Manual
1-18 1 Introduction
2 Installation
Installation
2
2.1 Pre-Installation Planning
The Anthem generator is shipped partially assembled. Review all installation,
assembly and calibration instructions before beginning the installation. Save
all packing materials until after the installation is complete. Check that the
physical location, conduit arrangement and junction boxes are in readiness for
permanent installation of this equipment.
Before you begin installation, perform the following tasks:
• Consult with the client and/or client physical plant superintendent to
verify that the site has sufficient power, utility, and network access.
• Determine site-specific requirements, client requests, and room workflow
considerations, by reviewing the following documents, as available:
- the shipping order
- statement of work
- job orders
- client room drawings
- special written or unwritten requirements
2 Installation 2-1
Anthem Generator Installation, Operation, & Service Manual
Caution
If the floor condition is unsuitable for a plumb and secure
installation, take corrective action with floor repair or modification.
Note: To fully open the front of the power module, an additional 12 3/4” of space is
required in front of the generator at the height of the upper cabinet.
2-2 2 Installation
Anthem Generator Installation, Operation, & Service Manual
21 1/2”
12 3/4”
27 1/2”
FP10-0005.00
23”
27 1/2”
31 1/2”
FP10-0006.00
2 Installation 2-3
Anthem Generator Installation, Operation, & Service Manual
Warning
Lethal voltages may be present!
Caution
Attempts to attach accessory items other than those described in this manual
may damage circuitry. Do not use aluminum wire. All wiring MUST be
copper.
2-4 2 Installation
Anthem Generator Installation, Operation, & Service Manual
2 Installation 2-5
Anthem Generator Installation, Operation, & Service Manual
Note: The minimum wire sizes listed above have been selected for the maximum wire
resistance allowed at maximum momentary full load current. To meet National
Electric Code Specifications, select a fuse or circuit breaker with an amperage rating of
at least 50% of the generator’s peak current demand.
2.1.6 Unpacking
1. Move the pallet to the approximate position where generator will be
installed.
2. Cut the plastic straps. Remove the straps and the top cover.
3. Lift off the outer box shell.
4. Remove shrink wrapping from around the generator.
2-6 2 Installation
Anthem Generator Installation, Operation, & Service Manual
5. Slightly tilt the generator and gently “walk” the generator off of the pallet
onto the floor at its final location.
Power Module
HT Tank
For 3-phase systems, you can install the line transformer enclosure behind the
HT Tank.
Input power connections (“Connection of Incoming Main Power Cable” on
page 2-10) are made at the rear of the Power Module. Make these connections
before sliding the Power Module into the table.
2 Installation 2-7
Anthem Generator Installation, Operation, & Service Manual
2.2 Installation
Before starting the installation, study the cable index (Figure 2-4 on page 2-8)
and the accompanying table (Table 2-1) to get familiar with the
interconnections between the Anthem and other equipment.
Aux. Power 6
Disconnect Table
Main Power
7
Disconnect Wall Stand
1
8 Wall 10
Receptor/Bucky
Line Match
Transformer
9 Table 11
2 3 4 Receptor/Bucky
Power
12
Module
Tube 1
13
Collimator
Wall ION
14
Chamber
15 Table ION
Chamber
16
17
HV 20
X-Ray Tube 1
18 Transformer
19
Operator
Control
2-8 2 Installation
Anthem Generator Installation, Operation, & Service Manual
2 Installation 2-9
Anthem Generator Installation, Operation, & Service Manual
Warning
Make sure that the power to the main power cable is turned off.
1. Route the main power cable to the location where the generator is being
installed.
2. Remove the rear panel of the generator power module. Retain all
hardware.
3. Locate the power module ground terminal (Figure 2-5). Remove and
retain the ground nut and temporarily disconnect the ground wires
currently connected.
Ground Terminal
Route Main
Power Cable
Through This
Clamp
Figure 2-5. Terminal TB1 and ground terminal on rear of power module
4. Route the main power cable through the cable clamp (Figure 2-5). Rotate
the cable so that the ground wire runs flat along the power module panel.
5. Connect the ground wire from the main power cable to the ground
terminal (Figure 2-6), and reconnect the other previously disconnected
2-10 2 Installation
Anthem Generator Installation, Operation, & Service Manual
ground wires to the ground terminal. Secure the ground wires with the
nut.
Neutral
(White)
Ground
(Green)
Hot (Red or
Black)
6. Connect the white wire from the main power cable to the 0 VOLTS
terminal.
7. Verify the power rating of the main power supply and connect the third
(hot) wire from the main power cable to the matching 208 or 240 Volt
terminal (4). Refer to Table 2-2 for correct voltage.
Warning
Make sure that the power to the main power cable is turned off.
2 Installation 2-11
Anthem Generator Installation, Operation, & Service Manual
1. On the lower cabinet, unscrew the four front panel screws (Figure 2-7) and
remove the front panel.
2. Route the main power cable to the generator and through the cable clamp
on the back of the lower cabinet. See Figure 2-8.
Cable Clamp
Figure 2-8. Main power cable routed through the back of the generator.
3. Connect the ground wire (Figure 2-9) to the ground stud on the bottom of
the lower cabinet. There are two nuts on the ground bolt. If there are 3 or
more ground wires, separate them between nuts.
2-12 2 Installation
Anthem Generator Installation, Operation, & Service Manual
Note: For 220 VAC, use the 208V terminal. For 230 VAC, use the 240V terminal.
2 Installation 2-13
Anthem Generator Installation, Operation, & Service Manual
Side Screws
Front Screws
2. At the side of the power module, route the cables from the tube and image
receptor(s) through the cable routing bracket holes (Figure 2-12).
Figure 2-12. Cable routing through the power module’s cable routing bracket
2-14 2 Installation
Anthem Generator Installation, Operation, & Service Manual
3. At the front of the power module, route the cables through the cable
clamps (Figure 2-13).
Cable Clamps
Figure 2-13. Cable routing through the power module’s front cable clamps
4. Connect the following cables to the room interface board per Table 2-3.
2 Installation 2-15
Anthem Generator Installation, Operation, & Service Manual
TB1
TB6
TB3
TB4
TB5
TB2 TB8
Warning
Be sure to connect ground cables to the ground screw above the board
(Figure 2-15).
Do not connect any loads to this supply that require more than 4A.
Heavier loads can damage the power supply.
2-16 2 Installation
Anthem Generator Installation, Operation, & Service Manual
Ground Connection
Progeny Progeny
Del L/F 8000 L/F 8000 Villa (see
Connection Signal Bucky L/F 9000 (earlier) Note below)
TB4-1 Wall Select 3 3 B3 7
117VAC (switched)
TB4-2 Wall Feedback 2 B2 12
TB4-3 117 VAC L L B8 3
TB4-4 Signal GND 1 1 B1 11
TB4-5 117 VAC Return N N B4 1
Chassis GND Chassis GND GND GND GND Chassis GND
Table 2-4: Wall Bucky connections to the room interface board.
2 Installation 2-17
Anthem Generator Installation, Operation, & Service Manual
Note: For the Villa Bucky, 120VAC operation, install a jumper between terminals 1
and 6 at the Bucky.
Note: If you are not installing a reciprocating grid, install a jumper between
TB4-2 and TB4-4. This jumper satisfies the Bucky interlock circuit, and if not
connected, generates a Bucky error.
Progeny Progeny
Del L/F 8000 L/F 8000 Villa (see
Connection Signal Bucky L/F 9000 (earlier) Note below)
TB5-1 Table Select 3 3 B3 7
117VAC (switched)
TB5-2 Table Feedback 2 B2 12
TB5-3 117 VAC L L B8 3
TB5-4 Signal GND 1 1 B1 11
TB5-5 117 VAC Return N N B4 1
Chassis GND Chassis GND GND GND GND Chassis GND
Table 2-5: Table Bucky connections to the room interface board
Note: For the Villa Bucky, 120VAC operation, install a jumper between terminals 1
and 6 at the Bucky.
Note: If you are not installing a reciprocating grid, install a jumper between
TB5-2 and TB5-4. This jumper satisfies the Bucky interlock circuit, and if not
connected, generates a Bucky error.
2-18 2 Installation
Anthem Generator Installation, Operation, & Service Manual
Figure 2-18. Tube rotor connections to terminal block TB3 (colors may vary)
Connection Signal
TB3-1 PHASE Stator Phase or AUX
TB3-2 MAIN Stator Main
TB3-3 COMMON Stator Common
TB3-4 THERMAL SW2 Thermal Switch *
TB3-5 THERMAL SW1 Thermal Switch Common *
*An isolated, normally closed (NC) set of contacts, mounted on the tube
housing, which opens if the housing exceeds normal operating temperatures.
If no housing thermal switch is present, install a jumper across TB3-4 and
TB3-5.
2 Installation 2-19
Anthem Generator Installation, Operation, & Service Manual
Connection Signal
TB6-1 24 VAC return
TB6-2 24 VAC
Ground Green/Yellow
Table 2-6. Collimator power supply connections.
2-20 2 Installation
Anthem Generator Installation, Operation, & Service Manual
Connection Signal
TB6-3 Marked +24
TB6-4 Marked -24
Ground Green/Yellow
Table 2-7. Table/wallstand lock 24 VDC power supply connections
Note: If the required jumpers are not installed, exposures are inhibited.
Connect the collimator interlock and/or door open interlock to terminal strip
TB1 on the room interface board as shown in Figure 2-21.
2 Installation 2-21
Anthem Generator Installation, Operation, & Service Manual
J2 J1
Figure 2-21. Jumper settings for system with no collimator interlock and/or door open
interlocks.
Connect the door interlock cable to terminals TB1-1 and TB1-3 on the room
interface board (Figure 2-22).
2-22 2 Installation
Anthem Generator Installation, Operation, & Service Manual
Connect digital imaging receptors to terminal strips TB2 and TB8 on the room
interface board as shown in Figure 2-23.
TB2 TB8
Figure 2-23. DR ready and exposure signal connections
Connection Signal
TB2-1 TABLE READY +
TB2-2 TABLE READY -
TB2-3 WALL READY +
TB2-4 WALL READY -
TB8-1 WALL EXPOSURE REQUEST +
TB8-2 WALL EXPOSURE REQUEST -
TB8-3 TABLE EXPOSURE REQUEST +
TB8-4 TABLE EXPOSURE REQUEST -
TB8-5 PREP +
TB8-6 PREP -
Table 2-9. DR Bucky signal connections
2 Installation 2-23
Anthem Generator Installation, Operation, & Service Manual
DR
Anthem detector
generator
room interface
sync box
DB9
Control
Console
DR workstation
2-24 2 Installation
Anthem Generator Installation, Operation, & Service Manual
TB1-6
15 A
24 VAC 120 VAC
TB1-7
Anthem Customer
Contactor X-Ray Warning
Room Supplied,
Light
Interface Not Included
Board With Anthem
Note: Always refer to local electrical codes for correct wiring methods.
2 Installation 2-25
Anthem Generator Installation, Operation, & Service Manual
bracket enough to prevent the cable from sliding through it. Do not over-
tighten.
Communication
Cable
Connector
Foot/Hand
Switch Cable
Connector
Ground
Connections
Communication
Cable Foot/Hand Strain Relief
Switch Cable Brackets
2-26 2 Installation
Anthem Generator Installation, Operation, & Service Manual
4. Slide the threaded ring (Figure 2-27) onto the end of each cable.
Caution
Failure to apply silicone paste can cause tube damage due to arcing.
8. Remove the protective plugs from the high-tension cable sockets in the
X-ray tube.
9. Important. Ensure the set screw on the threaded ring (Figure 2-28) is
backed out several turns so that it does not interfere with the tightening of
the cable or cause damage.
2 Installation 2-27
Anthem Generator Installation, Operation, & Service Manual
10. Connect the HT cable connectors to the tube sockets. Tighten the ring
firmly by hand.
Set Screw
Caution
Connecting a high tension cable from the X-ray tube to the wrong
socket on the generator will destroy the tube. The cables marked “+”
must go to the socket marked “+”, and the cable marked “-” must go to
the socket marked “-.”
11. Remove the protective plugs from the HT tank inside the generator.
12. Connect the HT cable going to the cathode (-) socket on the tube to the
CATHODE (-) socket on the HT tank (Figure 2-29).
2-28 2 Installation
Anthem Generator Installation, Operation, & Service Manual
13. Connect the HT cable going to the anode (+) socket on the tube to the
ANODE (+) socket on the HT tank.
14. Tighten the rings (Figure 2-30) to secure the connections.
Rings
Caution
Connecting an HT cable to the wrong socket on the generator will
destroy the tube. Verify that Cathode (-) on the generator connects to
cathode (-) on tube and that Anode (+) on the generator connects to
anode (+) on tube.
Warning
Lethal voltages are present on the DC bus capacitors in this unit.
Exercise extreme caution when working near the long copper bus bars
attached to the two large capacitors in the base of the power module.
Always verify with a DC voltmeter that the capacitors are discharged
before attempting any service in this area.
2 Installation 2-29
Anthem Generator Installation, Operation, & Service Manual
Warning
The large commutating capacitors mounted to the right on the rear
panel of the power module remain charged even when the power to
the unit is turned off.
Oil Level
Window
2-30 2 Installation
Anthem Generator Installation, Operation, & Service Manual
2 Installation 2-31
Anthem Generator Installation, Operation, & Service Manual
Filament Board
Figure 2-32. Board locations
2-32 2 Installation
Anthem Generator Installation, Operation, & Service Manual
TP 1-3
TP 4-10
2 Installation 2-33
Anthem Generator Installation, Operation, & Service Manual
TP10B
TP11
TP10
2-34 2 Installation
Anthem Generator Installation, Operation, & Service Manual
TP 4, 5, 5A
TP 1-3
TP 24-26
TP 6-13
TP 6A
TP 22A
TP 20-22
TP 47-52
TP 22B
TP 47A
TP 305
TP 306
TP 15-17
TP 23
TP53-55
TP 300
TP 18,19 TP 301
TP 304
TP 38-42
TP 31-37
TP 22C
TP 26A TP 302
TP 43-46 TP 303
2 Installation 2-35
Anthem Generator Installation, Operation, & Service Manual
2-36 2 Installation
Anthem Generator Installation, Operation, & Service Manual
If the control console does not display DOOR OPEN, check the following:
- at the room interface board, the interlock connections
- at the door, the switch wiring
3. While DOOR OPEN is still displayed, press the prep button. Verify that the
green READY light does not turn on. If the READY light does turn on,
contact Technical Support.
2 Installation 2-37
Anthem Generator Installation, Operation, & Service Manual
Caution
Write down your password and store it in a safe place. If you forget
your password, you must send the Control Console back to the factory
for reprogramming.
Power
Switch
When the Control Console boots up, you are prompted for a password. It
shows the default password of 0000.
2-38 2 Installation
Anthem Generator Installation, Operation, & Service Manual
1 2
3
Figure 2-38. Service Mode password
2 Installation 2-39
Anthem Generator Installation, Operation, & Service Manual
- IMPORT CAL PTS. (YES or NO). Load all calibration data from an SD card.
- EXPORT CAL PTS. (YES or NO). Save all calibration data to an SD card.
- BURN FAC DFTS. (YES or NO). Revert all calibration settings to their
factory default values. This selection also resets the flash memory and
clears flash memory fault messages. This selection does not affect APR
user settings.
- DISCHARGE CAPS. (YES or NO). Discharge the DC bus capacitors. Before
servicing circuits in the upper cabinet, see “Discharge Procedure” on
page 4-2.
• WALL BUCKY/TABLE BUCKY. (Yes or No). Determine which work stations
are available in User Mode.
Note: If you are not using a reciprocating grid or grid cabinets, you must install
a jumper on the Room Interface Board TB5. See “Wall Bucky Connections” on
page 2-17 or “Table Bucky Connections” on page 2-18.
• TUBE SELECT. Select from tubes available in the Anthem generator’s
library. Select this option, then use the data knob to scroll through tube
selections. See “X-Ray Tube Alternatives” on page 2-41 for available
alternatives.
• FLASH OPTIONS. Manage and update the APR database. This option is not
displayed if your Anthem generator has the DR option enabled. Select this and
you see the following options:
- EXPORT APR. (YES or NO). Save the APR database (user settings) to an
SD card.
- IMPORT APR. (YES or NO). Load an APR database (user settings) from an
SD card.
- DFLT APR DATA. (MED or VET). Set the APR regions and view names to
reflect either a medical or veterinarian nomenclature.
- RESET APR VALUES. (YES or NO). Revert all APR user settings to factory
default values. This command overwrites all APR user settings.
- LAPTOP TRANSFER. (YES or NO). Transfer the APR database to a
computer. see “Managing the APR database” on page 5-9.
• MISC OPTIONS. Go to KEY OPTIONS, ERROR LOG, TROUBLESHOOT, DEMO
MODE, and BUCKY START.
- KEY OPTIONS. Use this selection to upgrade feature options or change
the console power level (model).
- ERROR LOG. (YES or NO). When saving APR data to an SD card and this
option is set to Yes, also export an error log file in CSV (comma
separated variable) format, which you can open with Microsoft Excel.
Note: If you enable this option while an SD card is not present, you see SD
CARD FAULT.
- TROUBLESHOOT. (YES or NO). In User Mode, display calibration DAC
values.
- DEMO MODE. (YES or NO). Disable the communication link between the
console and the Main Control board. You can use this mode as a
2-40 2 Installation
Anthem Generator Installation, Operation, & Service Manual
2 Installation 2-41
Anthem Generator Installation, Operation, & Service Manual
2-42 2 Installation
Anthem Generator Installation, Operation, & Service Manual
2 Installation 2-43
Anthem Generator Installation, Operation, & Service Manual
Required Hardware
• Laptop or desktop PC running Windows XP or later.
• Internet connection.
Installation
1. Navigate a web browser to http://support.delmedical.com and log in to your
Del Medical customer support account.
2. Go to Product Information > Generators > Anthem. Under Software,
click on ANTHEM APR Utility v. 4.5 and download the file for this link,
which is filename APR_Utility45.zip.
3. Unzip the Del APR Utility files to a temporary folder on your hard drive
(e.g., ...\MyDocuments\APRtemp)
4. Exit all other applications. Go into folder APR_Utility_4.5 and run (double
click) setup.exe.
5. Verify that the default destination of the Del APR Utility Software is as
shown below (C:\Program files\DelAprUtility).
2-44 2 Installation
Anthem Generator Installation, Operation, & Service Manual
2 Installation 2-45
Anthem Generator Installation, Operation, & Service Manual
2-46 2 Installation
3 Operation
Operation
3
3.1 Introduction
The Anthem generator operator’s Control Console features an easy-to-use
menu-driven, multifunction button user interface.
Caution
Warning
Do not allow foreign objects that can attenuate or scatter the X-ray
beam between the X-ray tube and table top during exposure. Failure to
comply may result in serious injury.
3 Operation 3-1
Anthem Generator Installation, Operation, & Service Manual
9
1
1. Power Switch
2. Left Side Multifunction Buttons
3. Top Multifunction Buttons
4. Display
5. Right Side Multifunction Buttons
6. Data Knob
7. Prep Button
8. Exposure Button
9. Bottom Multifunction Buttons
3-2 3 Operation
Anthem Generator Installation, Operation, & Service Manual
READY EXPOSE
FP10-0014.00 FP10-0013.00
prep expose
3 Operation 3-3
Anthem Generator Installation, Operation, & Service Manual
Symbol/Text Description
kV Tube voltage value. This parameter typically controls con-
trast. Use the data knob to change the kV value in single
digit increments.
mAs (2-Point only) mAs value. mAs is an optimized composite
of tube current and exposure time values. Use the data
knob to change the mAs value in pre-determined steps.
mA (3-Point only) Tube current value. This parameter typically
controls image intensity, or darkness. Use the data knob
to change the mA value from 25 mA to 650 mA.
ms (3-Point only) Exposure time. Use the data knob to
change the ms value. Although you can select time val-
ues, the actual exposure duration is controlled by the mAs
timer. You can select from the following values:.010, .012,
.015, .020, .025, .030, .035, .040, .050, .060, .070, .080,
.09, 0.1, 0.12, 0.15, 0.25, 0.3, 0.35, 0.4, 0.5, 0.6, 0.75,
0.9, 1.0, 1.25, 1.5, 1.75, 2.0, 2.5, 3.0, 4.0, 5.0, and 6.0.
Toggle between 2 point and 3 point technique modes.
In 2-point mode, you select only the kV and mAs of an
2 POINT exposure. The generator calculates the mA value to pro-
duce the shortest possible exposure duration for the
3 POINT selected technique, while protecting the X-ray tube from
overloads.
In the 3-point mode, you select the kV, mA, and ms of the
exposure. Selections are limited by the X-ray tube over-
load circuit. If you see an Exceeds Tube Limits mes-
2 POINT sage, readjust either the kV, mA, or ms of the technique
3 POINT before making an exposure.
3-4 3 Operation
Anthem Generator Installation, Operation, & Service Manual
Symbol/Text Description
Image Receptor
Toggle between the international symbols for Table, Table
Top (DR only), Wall, or No Bucky.
NO BUCKY
Focal Spot Size
The focal spot size is pre-programmed to optimize the
image quality. The left symbol indicates large focal spot
and the right symbol indicates small focal spot.
Thickness
(APR only) For patient size, use either this button or the
__ cm
25cm Patient Size button. When you the select the thickness in
FP10-0022.0 centimeters (with the data knob), each cm value sets a
pre-programmed kV/mA/time combination. There are 20
combinations to select from, and the set of combinations
can vary with each Region/View. You program each of
these combinations using the APR Utility program (“Edit-
ing the APR Database” on page 5-12).
Patient Size
(APR only) For patient size, use either this button or the
Thickness button. This button toggles between small,
medium, and large patient size. Each size is pre-pro-
FP 0 002 00
grammed to select one of the Thickness values above,
and the selections can vary with each Region/View.
Symbol/Text Description
APR Operating Mode
When in Manual Mode, pressing this button switches to
APR mode. If already in APR mode, this button takes you
back to a previous screen.
Manual Operating Mode
Displayed only in APR Mode, this button switches to APR
mode.
P
Go back
This symbol appears when an error message occurs.
This button takes you back to the previous screen.
AEC
Toggle on or off the AEC (Automatic Exposure Control)
feature.
3 Operation 3-5
Anthem Generator Installation, Operation, & Service Manual
When self test finishes, the display defaults to the mode of operation selected
at the time of installation. The system is now ready for use.
3-6 3 Operation
Anthem Generator Installation, Operation, & Service Manual
2 POINT
80kV 12mAs 3 POINT
NO BUCKY
3 Operation 3-7
Anthem Generator Installation, Operation, & Service Manual
NO BUCKY
3-8 3 Operation
Anthem Generator Installation, Operation, & Service Manual
UPPER CERVICAL
EXTREMITY SPINE
AP MODE
LOWER LUMBAR
EXTREMITY SPINE
SELECT
ANATOMICAL REGION
PELVIS CHEST &
& HIP THORAX
After selecting a region, you see the available views for that region. Select a
view.
3 Operation 3-9
Anthem Generator Installation, Operation, & Service Manual
T-SPINE STERNUM
LAT LAT
After selecting a view, you see the technique parameters for that region. Edit
these parameter values as you would in 3-point manual mode.
SAVE
125 kV 350 mA 70 ms TECH
24 mAs
3-10 3 Operation
Anthem Generator Installation, Operation, & Service Manual
Anatomi-
cal
region View of region
Upper Finger Finger Hand/ Hand/ Forearm Elbow Elbow Humerus Shoulder Trans
Extremity AP/OBL LAT Wrst AP/ Wrst LAT AP/LAT AP/OBL LAT AP/LAT AP/LAT Thoractic
OBL
Lower Toe AP/ Foot AP/ Foot LAT Ankle Ankle TIB/FIB TIB/FIB Knee LAT Knee AP/ Femur
Extremity LAT LAT AP/OBL LAT AP LAT OBL AP
Femur
LAT
Pelvis & Hip Pelvis AP Hip AP Sacrum Sacrum
Coccyx/ Coccyx/
AP LT
Abdomen Abdo- Abdo- KUB AP/
KUB men AP men OBL
Upright
Cervical AP Odontoid Lateral/ Swimmer
Spine Oblique
Lumbar Spine AP Oblique Lateral L5-S1
Spot
Chest & Tho- Chest AP Chest T-Spine T-Spine Rib Rib Sternum Sternum
rax LAT AP LAT Above Below RAO LAT
Dia- Dia-
phragm phragm
Skull & Facial AP/PA or Skull Skull Skull LAT Facial Waters Facial
Caldwell Towne SMU AP/PA LAT
Base
3 Operation 3-11
Anthem Generator Installation, Operation, & Service Manual
To activate AEC, press the AEC ( ) button. You see new parameters in the
right-side column.
2 POINT
80kV 12mAs 3 POINT
375ms
BKUP
NO BUCKY 1.2
600
For exposure procedures, see “Making an Exposure” on page 3-14. When the
exposure is complete, the display shows the actual mAs and elapsed time of
the exposure.
3-12 3 Operation
Anthem Generator Installation, Operation, & Service Manual
Symbol/Text Description
Back Up
375ms This button lets you set the time backup value with the data
BKUP knob. Set this to approximately 2 to 3 times the expected mAs
for the selected kV.
Film Density
1.2 This button lets you set the optical density on the finished
radiograph. You can select changes in density, ranging from
0.8 to 2.0, with the data knob.
When you select a film speed value (below), this field shows
the value that was used during calibration, and is your “zero”
point.
Ion Chamber Fields
This button lets you select chamber field combinations for a 3-
field ionization chamber by toggling this button.
Film/Screen Selection
3 Operation 3-13
Anthem Generator Installation, Operation, & Service Manual
2PT KV
DEFAULT
40
2PT MAS IMAGE
DEFAULT DEFAULT
1.0 NONE
SET AN OPTION
DEFAULT
CONTRAST
70
FLASH SET
OPTIONS CLOCK
EXIT
2. You can now program default settings for the parameters shown in Figure
3-9.
To exit this screen and save your changes, press EXIT. The generator completes
its power-up sequence and reboots to the new default settings.
Note: There is no default setting for the FLASH OPTIONS, which also appears in the
Service Mode menu. You use this option to transfer APR user settings to or from a
PC. If APR is not installed, you do not see FLASH OPTIONS.
3-14 3 Operation
Anthem Generator Installation, Operation, & Service Manual
To make an exposure
Note: Releasing either of the two buttons during an exposure terminates the
exposure and you see a BUTTON FAULT error message.
Fault Indication
If a fault occurs in your system, you see a fault condition message on the
screen. See Chapter 7 “Troubleshooting”.
3 Operation 3-15
Anthem Generator Installation, Operation, & Service Manual
To make an exposure
Note: Releasing either of the two buttons during an exposure terminates the
exposure and you see a BUTTON FAULT error message.
Fault Indication
If a fault occurs in your system, you see a fault condition message on the
screen. See Chapter 7 “Troubleshooting”.
To make an exposure
1. Select the APR Mode by pressing the button. If the APR symbol is not
showing, APR mode is already selected.
3-16 3 Operation
Anthem Generator Installation, Operation, & Service Manual
2. Change or verify thickness by pressing the respective button and use the
data knob to adjust the thickness.
3. Change or verify patient size by repeatedly pressing the respective button
to toggle between small, medium or large patient.
4. Press and hold down the PREP button.
5. When the green READY light turns on, continue holding down the PREP
button and press and hold the EXPOSE button.
When you start the exposure, the yellow EXPOSE light goes on and you hear a
tone for the duration of the exposure.
6. Continue to hold down both buttons until the EXPOSE light goes off and
the tone stops.
Note: Releasing either of the two buttons during an exposure terminates the
exposure and you see a BUTTON FAULT error message.
Fault Indication
If a fault occurs in your system, you see a fault condition message on the
screen. See Chapter 7 “Troubleshooting”.
To make an exposure
1. Press and hold down the PREP button
2. When the green READY light turns on, continue holding down the PREP
button and press and hold the EXPOSE button.
When you start the exposure, the yellow EXPOSE light goes on and you hear a
tone for the duration of the exposure.
3. Continue to hold down both buttons until the EXPOSE light goes off and
the tone stops.
3 Operation 3-17
Anthem Generator Installation, Operation, & Service Manual
Note: Releasing either of the two buttons during an exposure terminates the
exposure and you see a BUTTON FAULT error message.
Fault Indication
If a fault occurs in your system, you see a fault condition message on the
screen. See Chapter 7 “Troubleshooting”.
3-18 3 Operation
TIME mA Selected
ms 25 50 75 100 150 200 250 300 350 400 450 500 550 600 650 700
10 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00
11 1.10 1.65 2.20 2.75 3.30 3.85 4.40 4.95 5.50 6.05 6.60 7.15 7.70
12 1.20 1.80 2.40 3.00 3.60 4.20 4.80 5.40 6.00 6.60 7.20 7.80 8.40
13 1.30 1.95 2.60 3.25 3.90 4.55 5.20 5.85 6.50 7.15 7.80 8.45 9.10
15 1.13 1.50 2.25 3.00 3.75 4.50 5.25 6.00 6.75 7.50 8.25 9.00 9.75 10.50
17 1.28 1.70 2.55 3.40 4.25 5.10 5.95 6.80 7.65 8.50 9.35 10.20 11.05 11.90
20 1.00 1.50 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00
22 1.10 1.65 2.20 3.30 4.40 5.50 6.60 7.70 8.80 9.90 11.00 12.10 13.20 14.30 15.40
25 1.25 1.88 2.50 3.75 5.00 6.25 7.50 8.75 10.00 11.25 12.50 13.75 15.00 16.25 17.50
27 1.35 2.03 2.70 4.05 5.40 6.75 8.10 9.45 10.80 12.15 13.50 14.85 16.20 17.55 18.90
30 1.50 2.25 3.00 4.50 6.00 7.50 9.00 10.50 12.00 13.50 15.00 16.50 18.00 19.50 21.00
32 1.60 2.40 3.20 4.80 6.40 8.00 9.60 11.20 12.80 14.40 16.00 17.60 19.20 20.80 22.40
35 1.75 2.63 3.50 5.25 7.00 8.75 10.50 12.25 14.00 15.75 17.50 19.25 21.00 22.75 24.50
37 1.85 2.78 3.70 5.55 7.40 9.25 11.10 12.95 14.80 16.65 18.50 20.35 22.20 24.05 25.90
40 1.00 2.00 3.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00
45 1.13 2.25 3.38 4.50 6.75 9.00 11.25 13.50 15.75 18.00 20.25 22.50 24.75 27.00 29.25 31.50
50 1.25 2.50 3.75 5.00 7.50 10.00 12.50 15.00 17.50 20.00 22.50 25.00 27.50 30.00 32.50 35.00
55 1.38 2.75 4.13 5.50 8.25 11.00 13.75 16.50 19.25 22.00 24.75 27.50 30.25 33.00 35.75 38.50
60 1.50 3.00 4.50 6.00 9.00 12.00 15.00 18.00 21.00 24.00 27.00 30.00 33.00 36.00 39.00 42.00
65 1.63 3.25 4.88 6.50 9.75 13.00 16.25 19.50 22.75 26.00 29.25 32.50 35.75 39.00 42.25 45.50
70 1.75 3.50 5.25 7.00 10.50 14.00 17.50 21.00 24.50 28.00 31.50 35.00 38.50 42.00 45.50 49.00
75 1.88 3.75 5.63 7.50 11.25 15.00 18.75 22.50 26.25 30.00 33.75 37.50 41.25 45.00 48.75 52.50
80 2.00 4.00 6.00 8.00 12.00 16.00 20.00 24.00 28.00 32.00 36.00 40.00 44.00 48.00 52.00 56.00
85 2.13 4.25 6.38 8.50 12.75 17.00 21.25 25.50 29.75 34.00 38.25 42.50 46.75 51.00 55.25 59.50
90 2.25 4.50 6.75 9.00 13.50 18.00 22.50 27.00 31.50 36.00 40.50 45.00 49.50 54.00 58.50 63.00
Table 3-2: mAs vs. mA and time selected (10 to 135 ms)
95 2.38 4.75 7.13 9.50 14.25 19.00 23.75 28.50 33.25 38.00 42.75 47.50 52.25 57.00 61.75 66.50
100 2.50 5.00 7.50 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00 70.00
110 2.75 5.50 8.25 11.00 16.50 22.00 27.50 33.00 38.50 44.00 49.50 55.00 60.50 66.00 71.50 77.00
120 3.00 6.00 9.00 12.00 18.00 24.00 30.00 36.00 42.00 48.00 54.00 60.00 66.00 72.00 78.00 84.00
125 3.13 6.25 9.38 12.50 18.75 25.00 31.25 37.50 43.75 50.00 56.25 62.50 68.75 75.00 81.25 87.50
130 3.25 6.50 9.75 13.00 19.50 26.00 32.50 39.00 45.50 52.00 58.50 65.00 71.50 78.00 84.50 91.00
135 3.38 6.75 10.13 13.50 20.25 27.00 33.75 40.50 47.25 54.00 60.75 67.50 74.25 81.00 87.75 94.50
3 Operation
3-19
Anthem Generator Installation, Operation, & Service Manual
3-20
TIME mA Selected
ms 25 50 75 100 150 200 250 300 350 400 450 500 550 600 650 700
140 3.5 7.0 10.5 14.0 21.0 28.0 35.0 42.0 49.0 56.0 63.0 70.0 77.0 84.0 91.0 98.0
150 3.8 7.5 11.3 15.0 22.5 30.0 37.5 45.0 52.5 60.0 67.5 75.0 82.5 90.0 97.5 105.0
165 4.1 8.3 12.4 16.5 24.8 33.0 41.3 49.5 57.8 66.0 74.3 82.5 90.8 99.0 107.3 115.5
3 Operation
175 4.4 8.8 13.1 17.5 26.3 35.0 43.8 52.5 61.3 70.0 78.8 87.5 96.3 105.0 113.8 122.5
180 4.5 9.0 13.5 18.0 27.0 36.0 45.0 54.0 63.0 72.0 81.0 90.0 99.0 108.0 117.0 126.0
190 4.8 9.5 14.3 19.0 28.5 38.0 47.5 57.0 66.5 76.0 85.5 95.0 104.5 114.0 123.5 133.0
200 5.0 10.0 15.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0 100.0 110.0 120.0 130.0 140.0
225 5.6 11.3 16.9 22.5 33.8 45.0 56.3 67.5 78.8 90.0 101.3 112.5 123.8 135.0 146.3 157.5
250 6.3 12.5 18.8 25.0 37.5 50.0 62.5 75.0 87.5 100.0 112.5 125.0 137.5 150.0 162.5 175.0
275 6.9 13.8 20.6 27.5 41.3 55.0 68.8 82.5 96.3 110.0 123.8 137.5 151.3 165.0 178.8 192.5
300 7.5 15.0 22.5 30.0 45.0 60.0 75.0 90.0 105.0 120.0 135.0 150.0 165.0 180.0 195.0 210.0
325 8.1 16.3 24.4 32.5 48.8 65.0 81.3 97.5 113.8 130.0 146.3 162.5 178.8 195.0 211.3 227.5
350 8.8 17.5 26.3 35.0 52.5 70.0 87.5 105.0 122.5 140.0 157.5 175.0 192.5 210.0 227.5 245.0
375 9.4 18.8 28.1 37.5 56.3 75.0 93.8 112.5 131.3 150.0 168.8 187.5 206.3 225.0 243.8 262.5
400 10.0 20.0 30.0 40.0 60.0 80.0 100.0 120.0 140.0 160.0 180.0 200.0 220.0 240.0 260.0 280.0
450 11.3 22.5 33.8 45.0 67.5 90.0 112.5 135.0 157.5 180.0 202.5 225.0 247.5 270.0 292.5 315.0
500 12.5 25.0 37.5 50.0 75.0 100.0 125.0 150.0 175.0 200.0 225.0 250.0 275.0 300.0 325.0 350.0
550 13.8 27.5 41.3 55.0 82.5 110.0 137.5 165.0 192.5 220.0 247.5 275.0 302.5 330.0 357.5 385.0
Anthem Generator Installation, Operation, & Service Manual
600 15.0 30.0 45.0 60.0 90.0 120.0 150.0 180.0 210.0 240.0 270.0 300.0 330.0 360.0 390.0 420.0
675 16.9 33.8 50.6 67.5 101.3 135.0 168.8 202.5 236.3 270.0 303.8 337.5 371.3 405.0 438.8 472.5
700 17.5 35.0 52.5 70.0 105.0 140.0 175.0 210.0 245.0 280.0 315.0 350.0 385.0 420.0 455.0 490.0
750 18.8 37.5 56.3 75.0 112.5 150.0 187.5 225.0 262.5 300.0 337.5 375.0 412.5 450.0 487.5 525.0
800 20.0 40.0 60.0 80.0 120.0 160.0 200.0 240.0 280.0 320.0 360.0 400.0 440.0 480.0 520.0 560.0
825 20.6 41.3 61.9 82.5 123.8 165.0 206.3 247.5 288.8 330.0 371.3 412.5 453.8 495.0 536.3 577.5
900 22.5 45.0 67.5 90.0 135.0 180.0 225.0 270.0 315.0 360.0 405.0 450.0 495.0 540.0 585.0
950 23.8 47.5 71.3 95.0 142.5 190.0 237.5 285.0 332.5 380.0 427.5 475.0 522.5 570.0
Table 3-3: mAs vs. mA and time selected (140 to 1660 ms)
1000 25.0 50.0 75.0 100.0 150.0 200.0 250.0 300.0 350.0 400.0 450.0 500.0 550.0 600.0
1120 28.0 56.0 84.0 112.0 168.0 224.0 280.0 336.0 392.0 448.0 504.0 560.0
1250 31.3 62.5 93.8 125.0 187.5 250.0 312.5 375.0 437.5 500.0 562.5
1370 34.3 68.5 102.8 137.0 205.5 274.0 342.5 411.0 479.5 548.0
1500 37.5 75.0 112.5 150.0 225.0 300.0 375.0 450.0 525.0 600.0
1660 41.5 83.0 124.5 166.0 249.0 332.0 415.0 498.0 581.0
Table 3-4: mAs vs. mA and time selected (1750 to 6000 ms)
3 Operation
3-21
Anthem Generator Installation, Operation, & Service Manual
Anthem Generator Installation, Operation, & Service Manual
3-22 3 Operation
4 Periodic Maintenance
Periodic Maintenance
4
4.1 Periodic Maintenance Schedule
Maintenance procedures for the Anthem X-ray generator are required within
30 days after installation and every 6 months thereafter. Service personnel
must be specifically trained and experienced with medical X-ray apparatus.
Applicable preventive maintenance or any repair service should be performed
only by these skilled individuals.
Refer to the schedule on next page for information on when to perform
periodic maintenance on the generator.
Warning
The main capacitor bank (located on the left side of the Power
Module) contains a very high charge when power is applied. This
charge is a fatal shock hazard. Never perform any maintenance, repair,
or service work inside the generator without first performing the
“Discharge Procedure” on page 4-2. Check the capacitor bank for zero
volts with a DC voltmeter before working on any internal circuitry.
Warning
Warning
Note: Due to varying operating conditions, the following procedures might need to
be performed at greater or lesser intervals. Set these intervals according to your
generator’s performance.
5. Take five exposures. (Note: This step sends voltage to the tube, but does not
recharge the capacitors.)
6. Open the upper cabinet and measure the voltage across the terminals (1 in
Figure 4-3) at capacitor C1. If voltage is 10 VDC or higher, take two more
exposures and then verify voltage across the bus again.
7. When the voltage is below 10 VDC, Continue with the maintenance/
repair/service work on the generator.
1
1
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
• Ensure the power has been disconnected and that the emergency switches
have been activated, before starting any cleaning operation.
• Ensure no liquid gets into the unit.
• Do not immerse the equipment, including any components or accessories,
in liquid.
• Do not autoclave the equipment, including any component or accessories.
• Do not use water. Water can short-circuit the electrical installation and
cause corrosion to mechanical parts.
• Do not use acid or abrasive products.
• Use only a dry cloth to clean chrome-plated parts.
• Only the surface areas of unit parts, including accessories and connection
cables, should be disinfected using a gaseous disinfectant.
• Clean painted parts with a cloth and products appropriate for cleaning
plastic materials; after cleaning wipe the surfaces with a clean, dry cloth.
• Do not spray cleaning or disinfection solution directly on the equipment.
To disinfect, moisten a cloth with a 70% Isopropyl alcohol solution or
equivalent and wipe the surface of the equipment.
• When disinfecting the examination room, ensure the unit is covered with
plastic sheets.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
1. Verify that insert of tube matches tube insert selected in the system
installation file. If it does not, call Del Medical Technical Support.
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Caution
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Oscilloscope settings
3. Set the oscilloscope as follows:
Channel #1
Vertical 2 V/div.
Horizontal 10 ms/div.
Mode Normal Sweep
Trigger Channel 1 - Positive
Channel #2
Vertical 1 V/div.
Horizontal 10 ms/div.
Mode Normal Sweep
Note: If you are performing calibration using only a scope, read the values of kV
and mA as follows:
kVp A+C
mA/mAs Display mA
Fil. Amps/Line Fil. Amps.
Trigger Source kVp
kV Delay 3.3 msec
Window Delay Off
Procedure Normal
Trigger Mode Percent
Trigger Level 75%
Printer Off
Warning
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Caution
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
1. Check the transformer oil level. Oil is at proper level when it reaches the
horizontal level pin (Figure 4-4) in the circular window. To replenish
transformer oil, unscrew the four screws holding the circular window, and
fill with Shell Diala Oil AX only. The oil level must not be above the top of
the horizontal level pin.
Oil Level
Window
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Calibration and
Adjustments 5
5.1 Introduction
This chapter provides calibration and adjustment procedures for the
generator.
Warning
All service and maintenance, including the procedures described
within this chapter, are to be performed by qualified service
personnel only.
5.2.1 Preparations
Caution
The following procedures produces X-rays. Take precautions to
ensure the safety of all personnel in the area. Minimum precautions
are as follows:
- Wear lead aprons.
- Personnel remaining in the X-ray room during exposure must
be behind a lead shield.
- Minimize radiation scatter through doorways, walls and floor.
2. On the top of the H.V. tank, unscrew the 2 feedback board cover mounting
screws (Figure 5-2). Remove the cover.
Feedback Board
Cover Mounting
Screws
Binding Posts
Figure 5-3. Binding posts that hold copper jumper J2 on the feedback board.
Jumper
5. Connect the mAs meter, black wire to black plug and red wire to red plug
(Figure 5-5).
Black
Red
Warning
Make sure to remount the jumper and put the cover back on after
calibration is finished!
Oscilloscope Settings
6. Set the oscilloscope as follows:
Channel #1
Vertical 2 V/div.
Horizontal 10 ms/div.
Mode Normal Sweep
Trigger Channel 1 - Positive
Channel #2
Vertical 1 V/div.
Horizontal 10 ms/div.
Mode Normal Sweep
Note: If you are performing calibration using only a scope, read the values of kV
and mA as follows:
kVp A+C
mA/mAs Display mA
Fil. Amps/Line Fil. Amps.
Trigger Source kVp
kV Delay 3.3 msec
Window Delay Off
Procedure Normal
Trigger Mode Percent
Trigger Level 75%
Printer Off
Initial Exposure
Warning
Lethal voltages are present on the large filter capacitors in this unit.
Exercise extreme caution when working near the long copper bus bars
attached to the large capacitors in the power module. Never perform
any service work inside the generator without first performing the
“Discharge Procedure” on page 4-2. Always verify with a DC
voltmeter that all three capacitors (two mounted to the back wall and
one on the bottom of the power module) are discharged before
attempting any service in this area.
Warning
Do not touch any part of the mAs meter during exposures.
Dangerously high voltages may be present.
Mode 2 Points
kV 40 kV
mAs 1 mAs
5.2.3 mA Calibration
1. From the Service Mode menu, press CALIBRATE. You see the Main
Calibration Screen.
2. Press mA. You see STEP 1 in the upper left corner, and the following values
displayed in the center:
LARGE
100mA, 75kV, 50ms
3. Make an exposure and observe the waveform on oscilloscope channel 2.
The flat part of the wave should stabilize at 1.0 V (example in Figure 5-6).
If not, proceed to the next step.
Stabilized at 1V
4. Select MA DAC, and use the data knob to adjust the mA DAC value to
achieve the 1.0 V target level (STEP 1 in Table 5-1).
5.2.4 kV Calibration
1. From the Main Calibration screen, select KV. You see STEP 1 in the upper
left corner, and the following values displayed in the center:
200mA, 40kV, 50ms
2. Make an exposure and observe the waveform on oscilloscope channel 1.
The flat top of the wave (see Figure 5-7) should be at 2.0 V. If not, proceed
to the next step
3. Select KV DAC, and use the data knob to adjust the kV DAC value to
achieve the 2.0 V target level (per Table 5-2).
Fil DAC too high Fil DAC too low Fil DAC good
Figure 5-8. oscilloscope screens for high, low, and good Fil DAC settings.
4. Select FIL DAC, and use the data knob to adjust the Fil DAC value to
achieve the target shape.
5. If needed, repeat the previous two steps.
6. When you have achieved the target waveform shape on the oscilloscope,
press SAVE.
7. Press NEXT PREHEAT. You see 2 appear above NEXT PREHEAT.
8. Repeat 3 through 7 above for the remaining pre-heat calibration steps.
Always press SAVE for each step.
9. When you press NEXT PREHEAT and you again see 1 appear above NEXT
PREHEAT, press GLOBAL ADJUST. In the center, you see ARE YOU SURE?
Press YES.
Control Console to a PC running the Del APR Utility, edit settings, then
upload the modified data back to the Control Console.
Note: This feature is not available on Anthem generators with DR activated.
Required Equipment
• Laptop computer, running Windows 98 or later.
• SD memory card.
• SD memory card adapter (if PC or laptop has no SD card slot).
• Interface cable:
- USB to DB9 active-circuit adapter cable, (Digi-Key part number 768-
1013-ND or equivalent), or
- DB9 female to DB9 female, wired pin-to-pin (Digi-Key part AE1016-ND
or equivalent). This is NOT a Null Modem or RS232 cable.
SD card
USB-DB9 adapter
DB9 cable
card reader
Caution
Do not perform any database optimization until you have selected the
X-ray tube installed with your generator. Incorrect tube selection may
cause permanent damage to the X-ray tube and generator.
1
Figure 5-10. DB9 connector at the rear of the Control Console
4. Enter Service Mode (“Entering Service Mode” on page 5-6). You see the
Service Mode menu.
5. Press FLASH OPTIONS.
6. Press LAPTOP TRANSFER. You see WAITING FOR CMD.
(To cancel this mode and return to the previous screen, press EXIT.)
7. On the PC, start the APR Utility application (DelAprUtility)
8. On the APR Utility screen, Data Transfer, click the Get Data button. The
status window beneath the button indicates that the program is contacting
the console, waiting for data, downloading data, and verifying the
checksum.
When the download is complete, the APR utility displays CRC Valid and
the Control Console shows the number of bytes transferred.
The APR utility is now loaded with the APR configuration data from the
generator.
2. Enter Service Mode (“Entering Service Mode” on page 5-6). You see the
Service Mode menu.
3. Press FLASH OPTIONS.
4. Press EXPORT APR. If an APR file already exists on the SD card, the console
asks if you want to overwrite it. The control screen shows that it is
exporting the data. When the export to the card is complete, the console
returns to the FLASH OPTIONS screen.
5. Remove the SD memory card from the Control Console and insert it in the
laptop (or card reader) SD slot.
6. On the laptop APR Utility, select File -> Load and navigate to the SD card.
Select the file APRUSER and click OK.
The APR utility is now loaded with the APR configuration data from the
generator.
The APR Utility user interface screen consists of the following parts:
Generator Parameters. These unchangeable parameters show the Anthem
generator configuration. They must match the connected generator or else the
APR Utility cannot exchange settings with the console.
Region. Click a radio button to view/edit 1 of 8 regions. These selections
correspond to the console APR first-level screen.
View. Click a radio button to view/edit 1 of 12 views. These selections
correspond to the console APR second-level screen.
Index. This displays the individual parameter values for each of 20 indexes
(density values in cm) for the selected Region/View. (Use the scrollbar at the
bottom to view Index 11-20.) This defines each of the 20 values that are
selectable from the console APR Thickness button ( ). To change a value,
__ cm
25cm
Global Adjustments. You can modify kVp and mAs values and apply the
changes to all indexes in a View, in a Region, or for all APR techniques. First
set the global mode (All, Region, or View), then select a kVp value to add to
the existing value, or an mAs value by which to multiply the existing value.
Each selection of a value is cumulative. The Undo button can undo multiple
operations when pressed repeatedly.
Figure 5-13. New regional views can be created in SPARE View fields.
When creating a new view, you must enter exposure technique data for all 20
densities. After entering the kV, mA, and time for all available selections in
the new view, you must optimize the view for your X-ray tube, using the
View selection within the Global Adjustments section on APR Utility screen,
and selecting a mAs multiplier of 1.0.
You must enter the value for patient thickness (in cm) for all thicknesses
allowed within this view. To limit the available patient thickness of any
view, enter a value of 0 for all non-required selections.
Did the column header turn red when you changed a parameter?
Technique changes that fall outside the tube protection profile for the selected
X-ray tube are not allowed, and cause the column header to turn red.
Changing the excessive value of mA, Time, or kV clears this error.
Note: The database downloaded from an Anthem generator contains the specific
signature of that generator’s power limit, tube type, and derating status. Because of
this, you can upload a database to only be a generator equipped with the same tube,
generator power limit, and derating status.
1. Plug the adapter’s USB end or the cable’s DB9 end into the laptop.
2. Plug the other (DB9) end of the adapter or cable into the DB9 connector at
the rear of the Control Console.
3. Enter Service Mode (“Entering Service Mode” on page 5-6). You see the
Service Mode menu.
4. Press FLASH OPTIONS.
5. Press LAPTOP TRANSFER. You see WAITING FOR CMD.
(To cancel this mode and return to the previous screen, press EXIT.)
6. On the laptop APR Utility, Data Transfer, click the Send Data button.
The APR Utility connects with the Control Console, transfers the data, and
verifies the checksum. When this process is complete, the APR Utility
displays Flash Updated, and the Control Console returns to the WAITING
FOR CMD state.
7. At the Control Console, exit all utility screens by pressing EXIT at all
prompts. The Control Console then reboots into normal operating mode.
The Control Console is now loaded with the APR configuration data from the
PC.
Note: The database downloaded from any generator contains the specific signature of
that generator’s power limit, tube type, and derating status. As a result, the database
may only be uploaded to a generator equipped with the same tube, generator power
limit, and derating status.
1. When the APR data has been modified in the Del APR Utility software
chose File -> Save in the Menu and save the new modified data to the SD
memory card. Name the file APRUSER.csv. DO NOT name the file
anything else.
2. Remove the SD card from the PC/laptop. With the SD memory card
contacts facing upward, insert the card in the SD slot at the rear of the
Control Console.
3. Enter Service Mode (“Entering Service Mode” on page 5-6). You see the
Service Mode menu.
4. Press FLASH OPTIONS.
5. Press Import APR. The console screen shows that it is importing the data.
When the import from the card is complete, the console returns to the
FLASH OPTIONS screen.
The upload of APR data is now complete.
Warning
The following procedure produces X-rays. Take precautions to ensure
your safety and the safety of others in close proximity. Minimum
precautions are as follows:
a. Wear lead aprons.
b. Persons in the room during exposure must be behind a lead shield.
c. Minimize radiation scatter through doorways, walls and floor.
Warning
Ensure that main power circuit breaker for the generator is off and
console power switch is OFF before performing the following steps.
Otherwise, you may get severely shocked.
JP300
Figure 5-16. J300 jumper connections
Warning
Use extreme caution when working around or adjusting the generator
when it is turned on and covers are off. Do not touch any circuits or
components. Circuits can carry lethal voltages which can severely
burn or electrocute you.
4. Enter Service Mode (“Entering Service Mode” on page 5-6). You see the
Service Mode menu.
5. If AEC is not installed,
a. Press AEC INSTALLED. You then see a screen with buttons labeled
BUCKY AEC WALL and BUCKY AEC TABLE. Select YES for either or both, as
applicable.
b. Set AEC defaults (the values an operator sees when first entering AEC
mode). To do this, first set CHAMBER TYPE to either ION (ionization) or
SOLID (solid state).
c. Set the DEFAULT OD value, which is typically 1.2. Do not set the film
speed, as it is not relevant to calibration. Press EXIT. Make a note of the
selected optical density value. You will need this in step 10 below.
This is the OD value that the operator sees when first entering AEC
Mode.
Note: If you want to change the DEFAULT OD value and AEC is already
installed (AEC INSTALLED YES), then you must change AEC INSTALLED to NO
and perform step 5 above.
6. Press the CALIBRATE button. You see the Generator Calibration screen.
7. Press the AEC button.
8. Press the REMOVE CALIB button (1 in Figure 5-17) to clear all previous
calibration values.
9. Press the image receptor button (2) to select the chamber to calibrate
(either wall or table).
Note: For the wall chamber, perform calibration at 40” SID with no grid.
10. Press the DESIRED OD button (3) until you see the desired optical density.
Select the same OD that you selected in step 5, step c above. Your
calibration data will be saved under the selected OD and film speed
combination.
11. Press the CHAMBER TYPE button (4) to select the desired chamber type
(ION or SOLID STATE).
12. Press the FILM SCREEN button (5) until the desired film screen appears.
Note: The Anthem stores separate AEC calibration data for each film screen speed at
the desired OD. This lets you calibrate multiple film speeds.
3 4
2
5
1 6
19. If needed, repeat step 16 and step 17. Do not re-adjust gain for the
remainder of the calibration.
20. Press the NEXT STEP button. You see the STEP 3 screen, as shown in
Figure 5-19 with the ion chamber’s right field highlighted (1).
21. Insert an empty (no film) cassette into the table.
Note: Be sure to collimate the beam to cover the upper two cells or all three cells.
25. You see the STEP 4 screen, as shown in Figure 5-20, with the ion chamber’s
left field (1) highlighted. Repeat step 22 and step 23 to balance the left
field.
26. Press the NEXT STEP. You see the STEP 5 screen as shown in Figure 5-21..
Place 6” of acrylic or water in a bucket over the center cell. Collimate the
beam over the entire cell.
27. Place film in the cassette.
28. Take an exposure and develop the film.
29. Check the optical density of the film with a densitometer.
30. If the optical density is within 5% of DESIRED OD (1 in Figure 5-21), press
the NEXT STEP button (2) and skip to step 32 of this procedure.
31. If optical density is not within 5% of desired OD, adjust the AEC DAC value
(3) as follows:
a. Press the AEC DAC button (4) and use the data knob (5) to increase or
decrease the value. A clockwise turn increases the DAC value, thus
increasing OD and elapsed mAs.
b. When optical density is within 5% of the desired OD, press the SAVE
button (6).
c. Press the NEXT STEP button (2).
Note: You can calculate the new AEC DAC with the following formula:
4 2
6
4 2
6
4 2
6
Component
Replacement 6
6.1 Introduction
This chapter provides instructions for replacing most of the major assemblies
on the generator.
Warning
All service and maintenance including the procedures described
within this chapter, are to be performed by qualified service
personnel only.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
Turn off all electrical power to the generator. Also, make sure that
power source is locked out and tagged “Equipment Being Serviced”
before servicing the generator. Otherwise you can be seriously
injured.
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
Side Screws
Front Screws
2. Unscrew the two thumb screws (Figure 6-3) to swing open the power
module door.
Thumbscrews
Tools Required
• Standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Caution
Replace fuses only with same type and rating of fuse or machine may
get damaged.
F4
10 A, 250 V
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
• needle-nose pliers
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
4. At the open front of the power module, pull the disconnected transformer
wires through the rear panel opening.
5. Unscrew the mounting nut from the bolt that secures the line match
transformer to the mounting bracket (Figure 6-7). Remove the line match
transformer.
Line Match
Transformer
Mounting Nut
7. Route and connect the line match transformer power wires to the TB1 on
the back of the power module. Each wire from the line match transformer
is labeled with the corresponding socket number on TB1.
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
• needle-nose pliers
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
3. At the applicable terminal block, disconnect the wires from the line match
transformer(s) you are about to replace.
4. Unscrew and retain the bolt(s) holding the line match transformer(s) in
place and remove the line match transformer(s).
Note: When replacing the line match transformer in the rear, you must
remove the terminal block bracket.
5. Connect the line match transformer(s) power wires to its corresponding
terminal block on the terminal block bracket. Each wire from the line
match transformer is labeled with the corresponding socket number on
the terminal block.
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
Mounting Bolt
Isolation
Transformer
TB6 TB7
Figure 6-9. Location of isolation transformer and connections (3-phase)
5. Unscrew and retain the transformer mounting bolt, and remove the
isolation transformer.
6. Mount the replacement isolation transformer to the bottom of the power
module using the bolt retained in Step 5.
7. Connect all wires to the Power Relay Board. Each wire from the isolation
transformer is labeled with the corresponding terminal number on the
Power Relay Board TB6 and TB7 (see Figure 6-9).
Caution
Make sure there are no tools or loose hardware left inside the
generator before remounting the cover.
Filament Board
Figure 6-10. Board locations
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Tools Required
• standard hand tool kit (screwdrivers, wrenches, etc.)
Warning
To avoid serious injury, before servicing the generator, turn off all
electrical power to the generator. Lock out the power source and tag it
“Equipment Being Serviced.”
Warning
Never perform any service work inside the generator without first
performing the “Discharge Procedure” on page 4-2.
Caution
Always use an anti-static wrist strap when working on electrostatic
sensitive devices.
Troubleshooting
7
7.1 Introduction
This chapter contains troubleshooting tables and charts that guide you
through most of the problems that may be related to the generator. Included
are also status/error messages along with the possible solution(s) to the error.
Warning
7 Troubleshooting 7-1
Anthem Generator Installation, Operation, & Service Manual
Warning
7-2 7 Troubleshooting
Anthem Generator Installation, Operation, & Service Manual
7 Troubleshooting 7-3
Anthem Generator Installation, Operation, & Service Manual
7-4 7 Troubleshooting
Anthem Generator Installation, Operation, & Service Manual
7 Troubleshooting 7-5
Anthem Generator Installation, Operation, & Service Manual
7-6 7 Troubleshooting
8 Electrical Schematics
Electrical Schematics
8
8.1 List of Electrical Schematics
Current
Drawing Number Description Revision
034-5089 Wiring Diagram, single-phase 5
034-5090 Wiring diagram, 3-phase 6
110-5105 HV Tank 1
124-5162 Room Interface 0
Table 8-1. List of Schematics
8-3
Anthem Generator Installation, Operation, & Service Manual
8-4
Anthem Generator Installation, Operation, & Service Manual
8-5
Anthem Generator Installation, Operation, & Service Manual
8-6
9 Illustrated Parts List
Current
Drawing Number Description Revision
110-5105 HV tank 1
110-5115 Single phase power module 9
110-5115 Single phase power module, in table 9
110-5116 3-phase power module 9
110-5116 3-phase power module, in table 9
112-5469 Base power module 2
112-5471 Door power module 7
112-5472 Inverter power module 1
112-5484 lower cabinet 4
Table 9-1. List of assembly drawings
110-5142G1-G10
35235,(7$5<$1'&21),'(17,$/
7+(,1)250$7,21&217$,1(',17+,6
'5$:,1*,67+(62/(3523(57<
$60+97$1.
2)'(/0(',&$/$1<5(352'8&7,21,1
3$5725$6$:+2/(:,7+2877+(:5,77(1 5(9
3(50,66,212)'(/0(',&$/,6352+,%,7('
%/220,1*'$/(,/
9-5
Anthem Installation, Operation, & Service Manual
9-6
Anthem Installation, Operation, & Service Manual
9-7
Anthem Installation, Operation, & Service Manual
9-8
Anthem Installation, Operation, & Service Manual
9-9
Anthem Installation, Operation, & Service Manual
9-10
Anthem Installation, Operation, & Service Manual
9-11
Anthem Installation, Operation, & Service Manual
112-5700G1/G2/G3
9-12
Anthem Installation, Operation, & Service Manual
112-5700G1/G2/G3
$
& &
* $60/2:(5&$%,1(73+
* $60/2:(5&$%,1(73+
$
6(&7,21$$
,7(012 3$57180%(5 '(6&5,37,21 *47< *47<
3 (48,30(17&$%,1(7
187+(;.(36
3 /$%(/($57+*5281'
6&5(:++06[
3 /$%(/6(5,$//2:(5&$%,1(73+
* +$51(66/2:(5&$%,1(7
6&5(:331+066(06[
3 75$16)250(53+$6(/,1(0$7&+
3 %5$&.(77(50,1$/675,3
3 7(50,1$/%/2&.326
3 /$%(/7(50,1$/675,33+;)05
3 /$%(/7%3+;)057%
3 /$%(/7%3+;)057%
3 /$%(/7%3+;)057%
187+(;.(36
:$6+(5)/$7
:$6+(563/,7/.67'
6&5(:++06;
3 /$%(/6(5,$//2:(5&$%,1(73+
3 7$*<(//2:&$87,21
%/220,1*'$/(,/
9-13
Anthem Installation, Operation, & Service Manual
9-14