You are on page 1of 12

Hearing Research 430 (2023) 108709

Contents lists available at ScienceDirect

Hearing Research
journal homepage: www.elsevier.com/locate/heares

Effects of preloads on middle-ear transfer function and acoustic reflex


in ossiculoplasty with PORP
Merlin Schär∗, Ivo Dobrev, Christof Röösli, Alexander M. Huber, Jae Hoon Sim
Department of Otorhinolaryngology, Head & Neck Surgery, University Hospital Zurich, University of Zurich, Switzerland

a r t i c l e i n f o a b s t r a c t

Article history: Introduction: Surgical outcomes in ossiculoplasty with partial ossicular replacement prostheses (PORPs)
Received 1 August 2022 are greatly influenced by the amount of preload imposed on the PORP. In this study, the attenuation of
Revised 10 December 2022
the middle-ear transfer function (METF) was experimentally investigated for prosthesis-related preloads
Accepted 30 January 2023
in different directions, with and without concurrent application of stapedial muscle tension. Different
Available online 1 February 2023
PORP designs were assessed to determine functional benefits of specific design features under preload
Keywords: conditions.
Ossiculoplasty Methods: The experiments were performed on fresh-frozen human cadaveric temporal bones. The effect
Partial-ossicular replacement prosthesis of preloads along different directions were experimentally assessed by simulating anatomical variance
Preload and postoperative position changes in a controlled setup. The assessments were performed for three dif-
Annular ligament ferent PORP designs featuring either a fixed shaft or ball joint and a Bell-type or Clip-interface. Further,
Stapedial muscle
the combined effect of the preloads towards the medial direction with tensional forces of the stapedial
Acoustic reflex
muscle was assessed. The METF was obtained via laser-Doppler vibrometry for each measurement condi-
tion.
Results: The preloads as well as the stapedial muscle tension primarily attenuated the METF between 0.5
and 4 kHz. The largest attenuations resulted from the preload towards the medial direction. The attenua-
tion of the METF with stapedial muscle tension was reduced with concurrent PORP preloads. PORPs with
a ball joint resulted in reduced attenuation only for preloads along the long axis of the stapes footplate.
In contrast to the clip interface, the Bell-type interface was prone to lose coupling with the stapes head
for preloads in the medial direction.
Conclusions: The experimental study of the preload effects indicates a direction-dependent attenuation
of the METF, with the most pronounced effects resulting from preloads towards the medial direction.
Based on the obtained results, the ball joint offers tolerance for angular positioning while the clip in-
terface prevents PORP dislocations for preloads in lateral direction. At high preloads, the attenuation of
the METF with stapedial muscle tension is reduced, which should be considered for the interpretation of
postoperative acoustic reflex tests.
© 2023 Published by Elsevier B.V.

1. Introduction The ossiculoplasty with a partial ossicular replacement prosthe-


sis (PORP) establishes a linkage between the tympanic membrane
The functional integrity of the middle-ear ossicular chain can (TM) and the stapes head. While the ossiculoplasty with PORPs
be compromised by a variety of congenital and non-congenital may provide adequate closure of the air-bone gap in some patients,
pathologies and damages of traumatic origin. The resultant con- the procedure suffers from variable outcomes with comparably low
ductive hearing loss may be surgically treated by otologic proce- success rates (Kamrava and Roehm, 2017). One of the major under-
dures which aim to rearrange or replace the affected middle-ear lying causes is that the surgical outcome is highly sensitive even to
structures. Ossiculoplasty is a common otologic procedure which small changes of the prosthesis position and prosthesis length. The
employs different types of implantable prostheses for the surgical position and length of the prosthesis affect the pretension of the
reconstruction of the middle-ear ossicular chain. stapedial annular ligament (Neudert et al., 2016), which has been
shown to exhibit nonlinear stiffening properties under quasi-static
loads (Lauxmann et al., 2014). While excessive pretension causes
significant attenuation of the middle-ear sound transmission due

Corresponding author. to the stiffening of the stapedial annular ligament, the lack of suf-
E-mail address: merlin.schaer@usz.ch (M. Schär).

https://doi.org/10.1016/j.heares.2023.108709
0378-5955/© 2023 Published by Elsevier B.V.
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Fig. 1. Schematic illustrations of the partial-ossicular replacement prostheses (PORPs) included in the experimental assessments: TTB (TTP- Tübingen Bell), CF (Clip Partial
FlexiBAL), and BF (Bell Partial FlexiBal), manufactured by Heinz Kurz GmbH (Dußlingen, Germany).

ficient pretension bears the risk of a postoperative prosthesis dislo- nence and inserts at the posterior part of the stapes neck. Ten-
cation. Unfortunately, prosthesis-related preloads on the stapedial sional forces acting on the stapes neck via the stapedial tendon
annular ligament are difficult to control objectively during middle- upon contraction of the stapedial muscle are therefore approxi-
ear surgery (Neudert et al., 2016). mately parallel to the long axis of the stapes footplate. The stape-
Conventional PORPs share a common design blueprint con- dial muscle tension increases the middle-ear impedance via stiff-
sisting of 1) a head plate interfacing with the TM and/or the ening of the stapedial annular ligament, resulting in a reduction
manubrium of the malleus, 2) an interface for coupling with the of the middle-ear sound transmission. The contractile force of the
stapes head, and 3) a shaft establishing a connection between the stapedial muscle becomes larger with increasing stimulus intensi-
headplate and the interface for the stapes head. To mitigate the ties (Kobrak, 1959), exhibiting fine-tuned contraction responses for
challenges which are commonly encountered in the ossiculoplasty incremental changes of the stimulus intensity of down to 1 dB SPL
with PORPs, prosthesis designs with novel features have been de- (Popelka and Hunter, 2013). The intensity of the eliciting stimu-
veloped. These new designs were created to 1) improve the tol- lus further affects the latency of the mechanical response, i.e., the
erance of prosthesis length and angulation, thereby accounting for latency of the stapedial muscle contraction with subsequent stiff-
high anatomical variance encountered across the patient popula- ening of the middle-ear ossicular chain. The latency is ∼100 ms
tion, 2) improve intraoperative handling for the otologist, and 3) around the threshold of the acoustic reflex but may become as
increase robustness against postoperative position changes and po- small as 25 ms for higher-intensity stimuli (Yost, 20 0 0). While the
tential postoperative dislocation. acoustic reflex in humans only involves the stapedial muscle, the
The first objective of this investigation was to experimentally tensor tympani is part of the acoustic reflex in a variety of other
assess the functional roles of two novel design features which have mammals (Gelfand, 2017).
been added to the conventional PORP design, i.e., 1) a clip inter- The second objective of this study was to explore effects of
face for the stapes head (Hüttenbrink et al., 2004), and 2) a flexible stapedial-muscle tension in correlation with different amounts of
ball joint between the head plate and shaft (Beutner et al., 2011; preload between the stapes and the PORP and between the TM
Gostian et al., 2013; Stoppe et al., 2018). The focus of the experi- and the PORP, which may be imposed intraoperatively and/or post-
mental assessment was to determine how the novel features help operatively. These effects were investigated to identify potential
to retain functional outcomes despite the high anatomical variation changes of acoustic reflex function via contractions of the stape-
across different patients and the possibility of postoperative posi- dial muscle in the surgically reconstructed ears.
tion changes of the prosthesis. In the reconstructed ears, the pri-
mary purpose of the ossicular replacement prosthesis is to restore
2. Materials and methods
middle-ear sound transmission akin to the sound transmission of
a normal middle ear. The experimental design of this investigation
2.1. Prostheses
was therefore based on the simulation of possible anatomical vari-
ance and postoperative position changes of the PORP. For each ex-
Three different designs of PORPs were included in this investi-
perimental condition, the middle-ear transfer function (METF), i.e.,
gation, i.e., prostheses with 1) a Bell-interface to the stapes head
the ratio of the stapes velocity to the sound pressure level (SPL) in
and a rigid connection between the head plate and the prosthesis
the ear canal, was obtained. The measurements with the different
shaft (TTP-Tübingen Bell, Heinz Kurz GmbH, Dußlingen, Germany),
types of prostheses were performed in fresh-frozen human cadav-
2) a Clip-interface with a ball joint between the head plate and the
eric temporal bones. The results of the measurements with the dif-
prosthesis shaft (Clip Partial FlexiBAL, Heinz Kurz GmbH), and 3)
ferent prostheses were then compared to assess potential benefits
a Bell-interface with a ball joint between the head plate and the
of the new design features.
prosthesis shaft (Bell Partial FlexiBal, Heinz Kurz GmbH) (Fig. 1).
The stapedial tendon may be preserved in the ossiculoplasty
The non-commercial ball-joint prosthesis with a Bell-interface (Bell
with PORPs, thereby maintaining the functionality of the acous-
Partial FlexiBal) was manufactured by Heinz Kurz GmbhH for the
tic reflex in the reconstructed middle ears. The acoustic reflex is
purpose of this study. The prostheses are hereafter referred to as
assumed to provide an active protection mechanism for inner-ear
TTB (TTP- Tübingen Bell), CF (Clip Partial FlexiBAL), and BF (Bell
structures against high-intensity stimuli and/or enhancement of
Partial FlexiBal).
speech discrimination in noisy environments (Djupesland, 1964;
Borg et al., 1984; Gelfand, 2017). The acoustic reflex is based on 2.2. Temporal bone preparation and surgical reconstruction
a neural feedback loop, resulting in contractions of the stapedial
muscle in the presence of loud sound above a threshold of 70– The investigation was performed on eight fresh-frozen human
100 dB SPL via innervation by the facial nerve (Djupesland, 1964; cadaveric temporal bones (TBs) under the ethical approval by the
Laukli and Mair, 1980). The stapedial muscle is enclosed in a Ethic Committee of the Canton of Zurich (KEK-ZH-Nr. 2014–0544).
bony canal which protrudes into a small tip, the pyramidal em- The specimens were acquired from Life Legacy Foundation (Tucson,
inence. The stapedial tendon originates from the pyramidal emi- Arizona, USA) and Science Care (Phoenix, Arizona, USA). The TBs

2
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Fig. 2. Experimental setup with the assembly of the artificial ear canal (AEC), artificial tympanic membrane (ATM), and PORP on a stack of three micro-stages.

were extracted within 72 h postmortem and subsequently stored 2.3. Measurement of middle-ear transfer function (METF) with
at a temperature of −20 °C in temperature-monitored freezers un- preloads on prosthesis
til preparation.
The TB preparation involved a wide mastoidectomy and poste- In place of the natural TM, powder-free exam gloves (KC300,
rior tympanotomy, followed by a visual inspection to detect ev- Kimberly-Clark, Irving, Texas, USA) were used to create artificial
ident pathological conditions of the middle-ear ossicular chain. tympanic membranes (ATMs). These ATMs were attached to a
Subsequently, the vibrational motion of the stapes under acous- small plastic tube with an inner diameter of 9 mm, which served
tic stimulation was measured and the METF was calculated based as an artificial ear canal (AEC). The assembly of the ATM and AEC
on the monitored sound pressure levels (SPL) in the artificial ear enabled efficient and precise control of preloads in four direc-
canal (see Section 2.3 for more details). The results were checked tions. Based on tympanometry measurements (ZODIAC 901, Mad-
for compliance with the criteria for normal middle-ear behavior, sen Corp., USA), the compliance of the ATM was adjusted to yield
as established by the American Society for Testing and Materi- volume changes of 0.4–0.6 ml for a pressure of 2 kPa. These vol-
als (American Society for Testing and Materials (ASTM) F2504-05, ume changes resemble values observed in normal-hearing patients.
2005). Only when the METF of the intact TBs complied with the The PORP was attached to the ATM by cyanoacrylate glue (Denseal
ASTM standard, the preparation of the specimen was continued, Superior, Prevest Denpro GmbH, Heidelberg, Germany), with the
by separating the intact TB into two parts: 1) An upper part con- prosthesis shaft aligned perpendicular to the ATM surface. Great
sisting of the ear canal, eardrum, malleus-incus complex, and sur- care was taken to apply small amounts of glue to avoid harden-
rounding bone, and 2) a lower part with the intact stapes and ing of the ATM or inadvertent fixation of the ball joint. An acoustic
cochlea. chamber with a volume of 0.5 ml was created via insertion of a
For the experimental assessments with stapedial muscle ten- yellow-foam ear-tip (ER1–14, Etymotic Research, Elk Grove Village,
sion, the preparation involved additional steps to fully expose Illinois, USA) into the AEC. To provide acoustic stimulation and
the belly of the stapedial muscle. The top parts of the bony monitor the sound pressure levels in the AEC, probes of a loud-
canal harboring the belly of the stapedial muscle were carefully speaker (ER-2, Etymotic Research) and a microphone (ER-7C, Ety-
removed using diamond burrs and micro-surgical tools. Subse- motic Research) were introduced into the AEC through the yellow-
quently, the tissue adhesions attaching the muscle belly to the foam ear-tip.
bony canal wall were meticulously detached along the full ex- The assembly consisting of the AEC, ATM, and the PORP was
tent of the muscle, including the portions close to the pyramidal subsequently mounted on a stack of three mutually orthogonal lin-
process. Hüttenbrink (1988) proposed such a thorough removal of ear translation micro-stages. With this setup, the position of the
soft-tissue adhesions to ensure an unimpeded load transfer to the assembly could be precisely translated with micron-level accuracy
ossicles. along three mutually orthogonal directions. The orientation of the
The lower part was mounted on an experimental bench setup, assembly was adjusted such that the longitudinal direction of the
and the short and long axes of the stapes footplate (SF) were PORP, i.e., the shaft of the prosthesis, was parallel to the Z-axis of
aligned with the X- and Y-axes of the reference frame of the micro- the reference frame of the micro-stages (as shown in Fig. 2).
stages (Fig. 2). A high-resolution camera providing a perpendicular Via the three translation micros-stages, the upper assembly
view onto the stapes footplate was used to visually assist the align- with the PORP was moved towards the stapes head, to establish
ment. a connection with the PORP interface. A surgical microscope and
For each prosthesis design and all preload conditions, the mea- a high-resolution camera were used to assist in the positioning of
surements of the METF were performed within approximately the upper assembly. After the upper assembly with the PORP was
20 min. The samples were kept moist by using a humidifier and moved into proximity of the stapes head, subsequent adjustments
by applying saline spray after each change of the preload condi- via the micro-stages were performed in small motion increments
tion. In case of suspected drying and after changing the prosthesis, of 2 μm. While incrementally moving the assembly with the PORP
the sample was rehydrated in saline solution for at least 30 min towards the stapes head, the stapes motion in response to acoustic
and the measurements were repeated, based on the results from a stimulation was measured via laser-Doppler vibrometry. The first
previous study by our group (Gerig et al., 2015). position which resulted in detectable stapes motion was used as

3
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Fig. 3. Outline of the experimental protocol for the investigation of preload effects. Position changes of the artificial tympanic membrane (ATM) relative to the stapes
provided incremental amounts of preloads between the stapes and the PORP and between the ATM and the PORP. Preload effects were assessed for preloads applied along
the long axis of the stapes footplate, the short axis of the footplate, towards the lateral direction (i.e., away from the stapes footplate), and towards the medial direction (i.e.,
towards the stapes footplate).

an indicator for an established coupling between the stapes head 2.4. Measurement of middle-ear transfer function (METF) with
and the PORP without preload, thereby representing the baseline stapedial muscle tension
position for the measurements.
The APx525 system (Audio Precision Inc., Beaverton, Oregon, The effect of stapedial muscle tension on the METF was only
USA) was used for data acquisition and control of stimulation and assessed for the application of PORP preloads in the medial direc-
measurement parameters. For acoustic stimulation, harmonic sig- tion (i.e., towards the stapes footplate), which allowed for use of
nals in the range of 0.5–8 kHz were fed to the loudspeaker via an the natural TM with the malleus instead of an ATM. As an addi-
audio amplifier (RMX 850a, QSC Audio Products, Costa Mesa, Cal- tional change to the experimental protocol, only two PORPs of the
ifornia, USA), with AEC sound pressure levels above 90 dB up to TTB type (see Fig. 1) were used for these measurements. As de-
8 kHz. The vibrational motion of the stapes was measured via a scribed in Section 2.3, small quantities of cyanoacrylate glue were
laser Doppler vibrometry (LDV) system (OFV-3001, Polytec GmbH, used to attach the head plate of the PORP to the malleus handle,
Waldbronn, Germany), which was mounted on a robot arm (KR16, under careful avoidance of glue reaching the natural TM.
Kuka AG, Augsburg, Germany) with six degrees of freedom. The To mimic stapedial muscle tension in the temporal bone spec-
laser beam of the LDV system was aimed near the center of the imen, constant quasi-static forces were applied to the stape-
stapes footplate by adjusting the position and orientation of the dial muscle via a 3-axis force sensor (PalpEar, Sensoptic, Losone,
robot arm accordingly. To establish a sufficiently high surface re- Switzerland). The force sensor was connected to the stapedial
flectivity for the laser beam of the LDV system, barium titanate muscle via a custom-made interface based on a micro-pipette tip
beads with retroreflective aluminum coating (∼50-μm diameter, (Fig. 4). To obtain the core component of the custom-made inter-
Cospheric LLC, Goleta, California, USA) were carefully placed on the face, the narrow front region of a plastic micro-pipette tip was ex-
stapes footplate. tracted. The shape of the removed segment resembled a frustum
Different preload conditions were imposed by incrementally of a cone with two different opening diameters. The belly of the
changing the position of the upper assembly with the PORP rela- stapedial muscle was gently pushed into the larger opening, while
tive to the stapes 1) along the long axis of the stapes footplate (to a surgical suture (4–0) was inserted into the smaller opening. Prior
the anterior direction), 2) along the short axis of the stapes foot- to insertion of the thread and the muscle through the two aper-
plate (to the superior direction), 3) to the lateral direction (away tures, the tube had been filled with cyanoacrylate glue. After the
from the stapes footplate), and 4) to the medial direction (towards insertion of the muscle belly and the suture, UV-curable adhesive
the stapes footplate) (Fig. 3). (Blufixx GmbH, Wesseling, Germany) was additionally applied to
Starting from the baseline position, the position changes were the region around the apertures to reinforce the attachment. The
performed in increments of 50 or 100 μm. For each position of 3-axis force sensor was mounted on a motorized micro-stage for
the upper assembly and each prosthesis design, the measurements precise control of the muscle-pulling forces. In the last step, the
were repeated 5 times and subsequently averaged. After acquiring free end of the surgical suture, which was maintained in a tension-
the sound pressure in the AEC and the velocity of the center of free state, was connected to the tip of the 3-axis force sensor with
the stapes footplate, the middle-ear transfer function (METF) was UV-curable adhesive.
calculated by After determining the baseline position of the upper assembly
consisting of the AEC, TM, and PORP, increasing preloads were ap-
plied by incremental movements towards the medial direction (i.e.,
vSF towards the stapes footplate). For each position of the PORP, the
MET F = ,
PAEC measurements were first performed without muscle-pulling forces,
followed by measurements with application of constant muscle-
pulling forces with a magnitude of 10, 20, and 40 mN. The muscle-
where vSF denotes the velocity of the stapes footplate center, and pulling forces were applied towards the posterior direction, i.e., in
PAEC the sound pressure in the AEC. While the exact angular posi- negative X-direction of the reference frame of the micro-stages,
tion of the laser beam in the measurement frame was not deter- parallel to the orientation of the stapedial tendon. For each of the
mined, it was held constant for all measurement positions of the preload conditions, i.e., each position of the PORP, the magnitude
upper assembly (AEC, ATM, and PORP) and for all prosthesis de- of the muscle-pulling force was continuously monitored and main-
signs. tained within approximately ± 1 mN of the targeted force magni-

4
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Fig. 4. Experimental setup for the mimicry of stapedial muscle tension. For each preload condition imposed on the prosthesis towards the medial direction (i.e., towards
the stapes footplate), different amounts of stapedial muscle-pulling forces were simulated via a 3-axis force sensor. Insert: Custom-made interface connecting the stapedial
muscle and the 3-axis force sensor via the extracted front segment of a micro-pipette tip.

Fig. 5. Middle-ear transfer functions (METFs) for preloads due to changes of the artificial tympanic membrane (ATM) position along the long axis of the stapes footplate
(towards the anterior direction) in TB1.

tude. A schematic representation of the measurement setup with tion from measurements in n = 3 temporal bones are provided in
the mimicry of stapedial muscle tension and the micro-pipette in- Fig. 8. Again, the overall trend was similar to the individual results
terface between the stapedial muscle and the 3-axis force sensor observed in TB1, and no clear difference between the three differ-
is provided in Fig. 4. ent PORP designs was detected. Maximum attenuation levels did
not exceed ∼5 dB relative to the baseline position.
3. Results

3.1. METFs with preloads on prosthesis


3.1.3. Movement of the TM to the lateral direction
The METFs with change of the ATM position to the lateral direc-
3.1.1. Movement of the TM along the long axis of the stapes footplate
tion (away from the stapes footplate) are outlined in Fig. 9, based
The effects on the METF resulting from the incremental changes
on data from TB1. With the TTB and BF prostheses, the coupling
of the ATM position along the long axis of the stapes footplate (to
between the PORP and the stapes head was lost for ATM position
the anterior direction) are shown in Fig. 5, based on data obtained
changes of 0.1 and 0.2 mm. In the results with the CF prosthesis,
in the sample TB1. In measurements with the TTB prosthesis, a 0.2-
there was no considerable change of the METF for any of the tested
mm movement of the ATM resulted in considerable attenuations of
ATM positions. The observation of a loss of coupling between the
the METF. In contrast, the results obtained for the CF and BF pros-
prosthesis and the stapes head with the TTB and BF prostheses
theses with a ball joint exhibited only minor attenuations. The av-
was consistent across samples for position changes of the ATM of
erage attenuation of the METF across measurements in n = 3 tem-
only 0.1 or 0.2 mm, which was reflected in the averaged data pre-
poral bones is provided in Fig. 6. The data in Fig. 6 shows the same
sented in Fig. 10. The CF prosthesis preserved a stable coupling to
trends as the results in Fig. 5, with attenuations of up to 12 dB rel-
the stapes head in the tested range of ATM positions, with peak
ative to the baseline position for a 0.2-mm movement of the upper
attenuations below 5 dB relative to the baseline position.
assembly with the TTB prosthesis.

3.1.2. Movement of the TM along the short axis of the stapes


footplate 3.1.3.1. Movement of the TM to the medial direction. The METFs with
The METFs with preloads imposed along the short axis of the change of the ATM position to the medial direction (towards the
stapes footplate due to position changes to the superior direction footplate, pushing the stapes into the cochlea) are presented in
are illustrated in Fig. 7, based on data obtained in the sample TB1. Fig. 11, for individual measurements in TB1. For all tested PORP
For all the PORP designs, the data indicated a gradual attenuation designs, considerable attenuations of the METF were observed. The
of the METF with incremental movement of the ATM. However, no averaged data follows this trend, with maximum average attenua-
major attenuation of the METF was found. The average attenua- tions of ∼9 dB (Fig. 12).

5
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Fig. 6. Averaged attenuation of METFs (n = 3) for preloads due to changes of the artificial tympanic membrane (ATM) position along the long axis of the stapes footplate
(towards the anterior direction).

Fig. 7. Middle-ear transfer functions (METFs) for preloads due to changes of the artificial tympanic membrane (ATM) position along the short axis of the stapes footplate
(towards the superior direction) in TB1.

Fig. 8. Averaged attenuation of METFs (n = 3) for preloads due to changes of the artificial tympanic membrane (ATM) position along the short axis of the stapes footplate
(towards the superior direction).

3.2. METFs with preloads on prosthesis and stapedial muscle tension muscle tension were present even for TM position changes above
50 μm. However, the averaged results followed a clear pattern
Changes of the METF due to stapedial muscle tension at differ- with reduced attenuations of the METF for larger position changes
ent positions of the upper assembly (AEC, TM, and PORP) along the of the TM.
latero-medial direction are presented in Fig. 13, for measurements
obtained in sample TB4. At the baseline condition without preload 4. Discussion and conclusion
(i.e., preload = 0), increasing force magnitudes of the stapedial
muscle tension resulted in larger attenuations of the METF. For 4.1. Effects of preloads on middle-ear sound transmission
preloads induced by position changes of the upper assembly with
the PORP which were equal to or greater than 50 μm, no clear The surgical outcomes of the ossiculoplasty with PORPs are
attenuations of the METF by the stapedial muscle tension could substantially affected by the angular alignment of the PORP and
be observed. In comparison, while the muscle tension was able to preloads between 1) the PORP and the stapes and 2) between
produce attenuations of up to ∼10 dB in the low-frequency range the PORP and the TM (Bance et al., 2004; Merchant et al., 1998;
without preload on the prosthesis, the maximal attenuation for the Morris et al., 2004; Murugasu et al., 2005; Puria et al., 2005;
0.2-mm position of the TM was smaller than 5 dB. The average Yung and Vowler, 2006; Zhao et al., 2005). The angular position
attenuations of the METF by stapedial muscle tension at differ- of the PORP and the preloads are determined by several factors.
ent preloads (i.e., TM position changes from baseline in the latero- First, the middle-ear anatomy shows a large inter-individual vari-
medial direction) are provided in Fig. 14. In these averaged results ance across different patients. Ideally, the angle between the PORP
from 5 measurements, attenuations of the METF with stapedial shaft and the stapes footplate as well as between the PORP shaft

6
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Fig. 9. Middle-ear transfer functions (METFs) for preloads due to changes of the artificial tympanic membrane (ATM) position to the lateral direction (away from the stapes
footplate) in TB1.

Fig. 10. Averaged attenuation of METFs (n = 3) for preloads due to changes of the artificial tympanic membrane (ATM) position to the lateral direction (away from the
stapes footplate).

Fig. 11. Middle-ear transfer functions (METFs) for preloads due to changes of the artificial tympanic membrane (ATM) position to the medial direction (towards the stapes
footplate) in TB1.

Fig. 12. Averaged attenuation of METFs (n = 3) for preloads due to changes of the artificial tympanic membrane (ATM) position to the medial direction (towards the stapes
footplate).

7
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Fig. 13. Middle-ear transfer functions (METFs) for different pulling-forces of the stapedial muscle, with concurrent preloads imposed via position changes of the upper
assembly with the AEC, TM, and PORP along the latero-medial direction (towards the stapes footplate) in TB4.

and the TM would be 90° However, previous studies indicate that rior and posterior sides of the stapes (Fig. 15). This configuration
the angle varies within a range of 60–90° (Beutner et al., 2011; of the prosthesis legs is compatible with preservation of the stape-
Gostian et al., 2013). Second, deviations from the optimal pros- dial tendon. Based on the results of the current study, the prosthe-
thesis length are detrimental to the surgical outcomes. Prostheses sis with the clip interface (CF) resulted in minor attenuations of
which exceed the optimal length result in considerable preloads the METF when tensional preloads were applied on the PORP (i.e.,
and resultant pretension and stiffening of the stapedial annular lig- when the upper assembly consisting of the AEC, ATM, and PORP
ament, while prostheses which are shorter than the optimal length was moved away from the stapes in lateral direction). In striking
increase the risk of postoperative prosthesis dislocations. Third, the contrast, a complete loss of coupling to the stapes head was ob-
ultimate position of the prosthesis and the resulting preloads may served for the PORPs with the Bell-type interface (TTB and BF) un-
change postoperatively, even if the optimal prosthesis length was der tensional preloads towards the lateral direction. Postoperative
selected during surgery. Underlying causes for these postoperative changes may increase the required or optimal prosthesis length,
changes include tissue growth, scar contraction, and TM move- such that the length of the implanted prosthesis becomes too short
ments due to large variations of ambient static pressure (De Vos in the process. The obtained data indicates that under these cir-
et al., 2007; Gostian et al., 2013; Yung and Vowler, 2006). Prosthe- cumstances, PORPs with the Bell-type interface (TTB and BF) bear
ses with novel design characteristics have been developed to miti- a high risk to lose coupling and to be rendered non-functional for
gate the potentially detrimental effect of these factors. middle-ear sound transmission.
The clip interface to the stapes head was designed to en- In efforts to mimic the flexibility of the human incudo-malleal
sure a secure connection of the PORP to the stapes head joint (IMJ) in reconstructed ears, the flexible ball joint between the
(Hüttenbrink et al., 2004). The prosthesis is attached to the stapes head plate and the prosthesis shaft was developed (Beutner et al.,
head such that the long legs of the clip are placed on the supe- 2011; Gostian et al., 2013; Stoppe et al., 2018). The functional role
rior and inferior sides, and the short legs are placed on the ante- of the flexible IMJ in the intact human middle ear has been in-

8
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Fig. 14. Averaged attenuation of METFs (n = 5) for different pulling-forces of the stapedial muscle, with concurrent preloads imposed via position changes of the upper
assembly with the AEC, TM, and PORP along the latero-medial direction (towards the stapes footplate).

cated that the energy loss with a flexible IMJ reduces the piston-
like motion component of the stapes above 2 kHz. These observa-
tions on the functional implications of the flexible IMJ in humans
are distinct from the effects of the ball joint, which were evaluated
in the development phase of total-ossicular replacement prosthe-
ses (TORPs) and PORPs. These evaluations showed that the design
of the ball joint resulted in a low frictional energy loss through
the ball joint under acoustic stimulation (Beutner et al., 2011;
Gostian et al., 2013). In the intact human middle ear, the flexible
IMJ serves a protective role by decoupling the incus and the stapes
from large displacements of the TM and the malleus under static
pressure differences across the TM (Hüttenbrink, 1988, 1996; Ihrle
et al., 2013). Hüttenbrink (1988, 1996) and Warnholtz (2021) ob-
served that the flexible IMJ reduces inward movements of the
Fig. 15. Schematic illustration of the long and short legs of the clip interface fea- stapes under positive external pressure and outward movements
tured in the CF prosthesis.
of the stapes under negative external pressure. The IMJ there-
fore establishes a protection mechanism with an immediate re-
sponse against large ambient static pressure changes. Further-
vestigated in previous studies. According to Willi et al. (2002), the more, the flexibility of the IMJ influences the dynamic motion
flexible IMJ results in a loss of sound-transmission energy at mid- of the middle-ear ossicular chain in response to acoustic stimuli.
dle and high frequencies. Investigations of Gerig et al. (2015) indi-

9
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Warnholtz et al. (2021) demonstrated that a decreased mobility chanical behavior of the natural human TM, even though the ATM
of the IMJ increases the attenuation of vibrational motion under compliance was adjusted to yield volume changes of 0.4–0.6 ml
static pressure loads. Based on this observation, the flexible IMJ for a pressure of 2 kPa. However, the relative attenuation of the
may serve as an adaptation mechanism for improved hearing un- METF with preloads is presumed to be largely due to the stiffen-
der ambient static pressure variations. The results of the current ing of the stapedial annular ligament because previous studies re-
study indicate that the effects of the preloads are dependent on vealed that the stiffening of the natural TM is negligible compared
the application direction of the preload. Compared to the prosthe- to the stiffening of the stapedial annular ligament for the range
sis without the ball joint (TTB), the attenuation of the METF in of preloads applied in this study (e.g., see Neudert et al., 2016). In
measurements with the ball-joint prostheses (BF and CF) was only Neudert et al. (2016), elongations of an expandable total ossicu-
reduced for the preloads along the long axis of the stapes footplate. lar replacement prosthesis of up to 300 μm relative to the optimal
No substantial differences in the attenuation of the METF could prosthesis length resulted in no distinct stiffening of the natural
be observed between the PORPs with or without a ball joint for TM in the reconstructed middle ear. In contrast, even prosthesis
the preloads along the short of the stapes footplate as well as for elongations of 50 μm relative to the optimal prosthesis length re-
the compressive preloads (i.e., when the ATM was moved towards sulted in considerable stiffening of the stapedial annular ligament
the stapes footplate). In the intact middle ear, similar attenua- and subsequent attenuation of the METF. Similar to the natural TM,
tions for both negative and positive static pressure loads have been the ATM used in this study does not exhibit significant stiffening
reported in the literature (Homma et al., 2010; Murakami et al., in the applied range of preloads. Therefore, we presume that the
1997; Warnholtz et al., 2021). In the current study, measurements ATM does not represent a substantial limitation for the purpose of
with the prosthesis featuring the clip interface (CF) revealed larger this investigation.
attenuations for the tensional preloads than for the compressive
preloads. These findings challenge the notion that the ball joint in 4.2. Acoustic reflex with preloads on prosthesis
the PORP mimics the physiologic behavior of the flexible IMJ. More
accurately, the functional role of the ball joint is to provide adjust- The mechanical effects resulting from the contraction of the
ments of the angle variation between the headplate and the shaft middle-ear muscles on the static and dynamic middle-ear ossic-
of the PORP. Such variations may occur due to anatomical differ- ular motion have been subject to previous experimental investi-
ences between patients and/or various postoperative changes. gations. Dahmann (1929) conducted experimental investigations of
The following conclusions may be drawn from the experimental the stiffening of the middle-ear due to contractions of the middle-
results obtained in this study. First, the flexible ball joint increases ear muscles, showing that the application of strain on the middle-
the tolerance for the angular positioning of the prosthesis. Second, ear muscles reduces the displacements of the ossicles. Hüttenbrink
the clip interface for the stapes head ensures a stable coupling be- (1988; 1989) developed an experimental setup for the measure-
tween the prosthesis and the stapes head under tensional forces ment of static displacements of the middle-ear ossicles under
between the PORP and the stapes head. Such tensional forces be- static pressure loads and pulling-forces of the middle-ear muscles.
tween the PORP and the stapes and between the PORP and the TM The reported approach for mimicry of muscular contraction was
may occur in the case that the intraoperatively employed prosthe- based on the application of pulley-connected weights to the bellies
sis length is below the required length or if the required length of the tensor tympani and the stapedial muscle. For the stapedial
increases postoperatively. Third, the effect of the preloads on the muscle, a weight of 5 g was attached in one of the experimen-
METF depends on the directionality of the preload. In compari- tal studies (Hüttenbrink, 1988), while varying weights between 1
son, compressive forces (i.e., preload towards the medial direction) and 20 g were used in another investigation (Hüttenbrink, 1989).
between the prosthesis and the stapes and between the prosthe- Pau et al. (2005) and Bance et al. (2013) applied a similar approach
sis and the TM were larger than the attenuations resulting from based on weights on a pulley system to mimic the effects of mus-
preloads which were imposed in other directions. The difference in cular contraction, but the investigations solely focused on the ten-
the attenuation of the METF depending on the direction of preload sor tympani.
can be explained by considering 1) the specific geometry and re- In the current study, it was possible to employ a more compact
sultant inhomogeneous stiffness distribution of the stapedial annu- setup for the mimicry of stapedial muscle tension by employing a
lar ligament and 2) the progressive stiffening characteristics of the 3-axis force sensor. The forces which were applied on the stape-
stapedial annular ligament (e.g., see Lauxmann et al., 2014). Based dial muscle were monitored and maintained via feedback control
on the study of Lauxmann et al. (2014), the stiffness at a given within approximately ± 1 mN of the designated force magnitude.
working point as well as the progressive stiffening of the stapedial To avoid excessive strains on the stapedial annular ligament, the
annular ligament vary depending on the displacement direction of magnitude of the tensional forces was limited to a maximum of
the stapes. The observations of this study, such as 1) different at- 40 mN (which corresponds to a weight of 4.08 g). The maximum
tenuations of the METF depending on the preload direction and force magnitude resulted in an attenuation of the METF in the sur-
2) largest attenuations for the preloads in the lateral direction are gically reconstructed ears with the PORP of around 10 dB (rela-
consistent with the findings of Lauxmann et al. (2014). tive to the METF without muscle tension) in the frequency range
Otologists commonly apply moderate compressive preloads dur- of 0.8–2.5 kHz (first graph in Fig. 14). The attenuation effect de-
ing surgery, to reduce the risk of postoperative dislocations of the creased at higher frequencies.
prosthesis and to ensure reliable coupling between the TM and the The stiffening of the middle ear due to stapedial muscle ten-
head plate as well as between the coupler and the stapes head. sion under high-intensity acoustic stimuli is utilized for diagnostic
Determining the optimal amount of preload is a delicate balancing purposes in the clinical setting. The acoustic reflex depends on the
act, since only small excessive amounts of compressive preloads condition of the hearing pathway through the middle and inner
may result in considerable attenuations of the METF in the recon- ears, the signal transmission via the 7th and 8th cranial nerves
structed ears (Neudert et al., 2016). (CNVs), and the brainstem. In tympanometry, the condition of
The results presented in Section 3.2 of this study were obtained this pathway can be checked by measuring middle-ear impedance
with an ATM. The flat shape of the ATM limits a direct translation changes resulting from the contraction of the stapedial muscle. The
of the METFs to the clinical situation due to the conical shape of tests are performed with acoustic stimuli of 0.5, 1, 2, and 4 kHz.
the natural human TM. The mechanical properties and behavior of The acoustic reflex test includes the measurements of the acoustic
the ATM are not identical to the mechanical properties and me- reflex threshold (ART), defined as the minimal sound pressure level

10
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

(SPL) to elicit a mechanical response, i.e., a change in the middle- Bance, M., Makki, F.M., Garland, P., Alian, W.A., van Wijhe, R.G., Savage, J., 2013.
ear impedance due to contraction of the stapedial muscle. The ART Effects of tensor tympani muscle contraction on the middle ear and markers of
a contracted muscle. Laryngoscope 123 (4), 1021–1027. doi:10.1002/lary.23711.
is determined for both ears and commonly within the range of Bance, M., Morris, D.P., VanWijhe, R.G., Kiefte, M., Funnell, W.R.J., 2004. Comparison
70–100 dB SPL in normal-hearing subjects (Laukli and Mair, 1980). of the mechanical performance of ossiculoplasty using a prosthetic malleus-
Considering the results presented in Fig. 14, the attenuation of the to-stapes head with a tympanic membrane-to-stapes head assembly in a hu-
man cadaveric middle ear model. Otol. Neurotol. 25 (6), 903–909. doi:10.1097/
middle-ear sound transmission due to stapedial-muscle tension in 0 0129492-20 04110 0 0-0 0 0 08.
the frequency range of 0.5–4 kHz is still evident in the recon- Beutner, D., Luers, J.C., Bornitz, M., Zahnert, T., Hüttenbrink, K., 2011. Titanium clip
structed ears with the PORP. This observation indicates that the ball joint: a partial ossicular reconstruction prosthesis. Otol. Neurotol. 32 (4),
646–649. doi:10.1097/MAO.0b013e318213867a.
acoustic reflex test retains its validity even for measurements after
Borg, E., Counter, S., & Rosler, G. (1984). Theories of middle-ear muscle function.
surgical reconstruction, if the preloads are sufficiently small. How- The Acoustic Reflex: Basic Principles and Clinical Applications, 63–99.
ever, considering the results provided in the last graph of Fig. 14, Dahmann, H., 1929. Zur Physiologie des Hörens. Experimentelle Untersuchungen
über die Mechanik der Gehörknöchelchenkette sowie über deren Verhalten auf
acoustic reflex tests may not detect changes of the middle-ear
Ton und Luftdruck. Z. Hals-Nasen-u Ohrenheilk (Leipzig) 24, 462–498.
impedance upon contraction of the stapedial muscle if the preloads De Vos, C., Gersdorff, M., Gérard, J.-.M., 2007. Prognostic factors in ossiculoplasty.
are excessively high. The tensional forces of the stapedial muscle Otol. Neurotol. 28 (1), 61–67. doi:10.1097/01.mao.0 0 0 0231598.33585.8f.
stiffen the stapedial annular ligament, which results in an increase Djupesland, G., 1964. Middle ear muscle reflexes elicited by acoustic and nona-
coustic stimulation. Acta Otolaryngol. 57 (sup188), 287–292. doi:10.3109/
of the middle-ear impedance (Rosowski, 2013). The application of 0 0 016486409134578.
large preloads on the prosthesis in the reconstructed middle ears Gelfand, S.A., 2017. Hearing: An introduction to Psychological and Physiological
already results in considerable stiffening of the stapedial annular Acoustics. CRC Press doi:10.1201/9781315154718.
Gerig, R., Ihrle, S., Röösli, C., Dalbert, A., Dobrev, I., Pfiffner, F., Eiber, A., Huber, A.M.,
ligament. In this case, the tensional forces of the stapedial muscle Sim, J.H., 2015. Contribution of the incudo-malleolar joint to middle-ear sound
may not be sufficient to contribute to additional stiffening of the transmission. Hear. Res. 327, 218–226. doi:10.1016/j.heares.2015.07.011.
stapedial annular ligament and may therefore not produce further Gostian, A., Pazen, D., Luers, J., Hüttenbrink, K., Beutner, D., 2013. Titanium ball joint
total ossicular replacement prosthesis–experimental evaluation and midterm
attenuation of the middle-ear sound transmission. These aspects clinical results. Hear. Res. 301, 100–104. doi:10.1016/j.heares.2012.10.009.
should be considered for the interpretation of the acoustic reflex Homma, K., Shimizu, Y., Kim, N., Du, Y., Puria, S., 2010. Effects of ear-canal pressur-
test after ossiculoplasty. ization on middle-ear bone-and air-conduction responses. Hear. Res. 263 (1–2),
204–215. doi:10.1016/j.heares.2009.11.013.
Remarks: This research article is based on a thesis chapter of: Hüttenbrink, K., 1988. The mechanics of the middle-ear at static air pressures:
the role of the ossicular joints, the function of the middle-ear muscles and
Schär, M.W. (2022). Protective and Adaptive Functions of the Human the behaviour of stapedial prostheses. Acta Otolaryngol. 105 (sup451), 1–35.
Middle Ear: Roles of the Stapedial Annular Ligament and the Stape- doi:10.3109/0 0 0164888090990 07.
dial Muscle (PhD thesis, Faculty of Science, University of Zurich, Hüttenbrink, K., 1989. Movement of the ear ossicles by middle ear muscle contrac-
tion. Laryngorhinootologie 68 (11), 614–621.
Switzerland).
Hüttenbrink, K., 1996. The middle ear as a pressure receptor. Middle ear mechanics
in research and otosurgery. Proc. Internat. Workshop, Dresden 19–22.
Conflicts of Interest Hüttenbrink, K., Zahnert, T., Wüstenberg, E., Hofmann, G., 2004. Titanium clip pros-
thesis. Otol. Neurotol. 25 (4), 436–442.
Kamrava, B., Roehm, P.C., 2017. Systematic review of ossicular chain anatomy: strate-
No conflict of interest to declare. gic planning for development of novel middle ear prostheses. Otolaryngol.–Head
Neck Surg. 157 (2), 190–200. doi:10.1177/0194599817701717.
CRediT authorship contribution statement Kobrak, H.G., 1959. The middle ear. Acad. Med. 34 (5), 554.
Laukli, E., Mair, I., 1980. Ipsilateral and contralateral acoustic reflex thresholds. Au-
diology 19 (6), 469–479. doi:10.3109/0 02060980 09070 079.
Merlin Schär: Conceptualization, Methodology, Software, Inves- Lauxmann, M., Eiber, A., Haag, F., Ihrle, S., 2014. Nonlinear stiffness characteristics
tigation, Formal analysis, Visualization, Writing – original draft, of the annular ligament. J. Acoust. Soc. Am. 136 (4), 1756–1767. doi:10.1121/1.
4895696.
Writing – review & editing. Ivo Dobrev: Methodology, Investi-
Merchant, S., McKenna, M., Rosowski, J., 1998. Current status and future challenges
gation. Christof Röösli: Methodology, Investigation. Alexander M. of tympanoplasty. Eur. Arch. Otorhinolaryngol. 255 (5), 221–228. doi:10.1007/
Huber: Conceptualization, Supervision, Project administration, s0 04050 050 047.
Funding acquisition. Jae Hoon Sim: Conceptualization, Methodol- Morris, D.P., Bance, M., van Wijhe, R.G., Kiefte, M., Smith, R., 2004. Optimum tension
for partial ossicular replacement prosthesis reconstruction in the human middle
ogy, Investigation, Formal analysis, Writing – original draft, Writing ear. Laryngoscope 114 (2), 305–308. doi:10.1097/0 0 0 05537-20 04020 0 0-0 0 024.
– review & editing, Supervision, Project administration. Murakami, S., Gyo, K., Goode, R.L., 1997. Effect of middle ear pressure change
on middle ear mechanics. Acta Otolaryngol. 117 (3), 390–395. doi:10.3109/
Data availability 0 0 016489709113411.
Murugasu, E., Puria, S., Roberson Jr., J.B., 2005. Malleus-to-footplate versus malleus-
to-stapes-head ossicular reconstruction prostheses: temporal bone pres- sure
Data will be made available on request. gain measurements and clinical audiological data. Otol. Neurotol. 26 (4), 572–
582. doi:10.1097/01.mao.0 0 0 0178151.44505.1b.
Neudert, M., Bornitz, M., Lasurashvili, N., Schmidt, U., Beleites, T., Zahnert, T., 2016.
Acknowledgments
Impact of prosthesis length on tympanic membrane’s and annular ligament’s
stiffness and the resulting middle ear sound transmission. Otol. Neurotol. 37
We would like to acknowledge the help from Dr. Daiki Tak- (9), e369–e376. doi:10.1097/MAO.0 0 0 0 0 0 0 0 0 0 0 01064.
agi (Department of Otolaryngology, Head & Neck Surgery, Ehime Pau, H.W., Punke, C., Zehlicke, T., Dressler, D., Sievert, U., 2005. Tonic contractions
of the tensor tympani muscle: a key to some non-specific middle ear symp-
University Hospital, Toon, Japan) and Dr. Mansour Alshamani (De- toms? hypothesis and data from temporal bone experiments. Acta Otolaryngol.
partment of Otolaryngology, Head & Neck Surgery, Ohud Hospital, 125 (11), 1168–1175. doi:10.1080/0 0 016480510 012408.
Medina, Saudi Arabia) who contributed to a part of the experi- Popelka, G.R., Hunter, L.L., 2013. Diagnostic measurements and imaging technologies
for the middle ear. In: Puria, S., Fay, R.R., Popper, A.N. (Eds.), The Middle Ear:
mental work. Further, we would like to thank Heinz Kurz GmbH Science, Otosurgery, and Technology. Springer, New York, pp. 211–251. doi:10.
(Dußlingen, Germany) for providing the prostheses used in this 1007/978- 1- 4614- 6591- 1_8.
study. Puria, S., Kunda, L.D., Roberson Jr, J.B., Perkins, R.C, 2005. Malleus- to-footplate ossic-
ular reconstruction prosthesis positioning: cochleovestibular pressure optimiza-
This work was supported by the SNSF (Swiss National Science tion. Otol. Neurotol. 26 (3), 368–379. doi:10.1097/01.mao.0 0 0 0169788.07460.4a.
Foundation) Grant No. No. 320030-176123. Rosowski, J.J. (2013). Comparative middle ear structure and function in vertebrates.
The Middle Ear, 31–65. 10.1007/978-1-4614-6591-1_3
References Stoppe, T., Bornitz, M., Lasurashvili, N., Sauer, K., Zahnert, T., Zaoui, K., Beleites, T.,
2018. Function, applicability, and properties of a novel flexible total ossicular re-
American Society for Testing and Materials (ASTM) F2504-05. (2005). Standard prac- placement prosthesis with a silicone coated ball and socket joint. Otol. Neurotol.
tice for describing system output of implantable middle ear hearing devices. 39 (6), 739–747. doi:10.1097/MAO.0 0 0 0 0 0 0 0 0 0 0 01797.
Phila. 10.1520/F2504-05 Warnholtz, B., 2021. Investigation of the Normal and Reconstructed Human Middle

11
M. Schär, I. Dobrev, C. Röösli et al. Hearing Research 430 (2023) 108709

Ear Under Static Pressure loads. Doctoral dissertation, Faculty of Science. Uni- Yost, W. (20 0 0). Chapter 14: auditory perception and sound source determination
versity of Zurich, Switzerland. in fundamentals of hearing. An introduction.
Warnholtz, B., Schär, M., Sackmann, B., Lauxmann, M., Chatzimichalis, M., Proc- Yung, M., Vowler, S.L., 2006. Long-term results in ossiculoplasty: an analysis
hazka, L., Dobrev, I., Huber, A.M., Sim, J.H., 2021. Contribution of the flexible of prognostic factors. Otol. Neurotol. 27 (6), 874–881. doi:10.1097/01.mao.
incudo-malleal joint to middle-ear sound transmission under static pressure 0 0 0 0226305.43951.13.
loads. Hear. Res. 406, 108272. doi:10.1016/j.heares.2021.108272. Zhao, S., Hato, N., Goode, R.L., 2005. Experimental study of an adjustable- length
Willi, U.B., Ferrazzini, M.A., Huber, A.M., 2002. The incudo-malleolar joint and sound titanium ossicular prosthesis in a temporal bone model. Acta Otolaryngol. 125
transmission losses. Hear. Res. 174 (1–2), 32–44. doi:10.1016/S0378-5955(02) (1), 33–37. doi:10.1080/0 0 016480410 018250.
00632-9.

12

You might also like