Professional Documents
Culture Documents
!"#$
%
"
& '
' ( ($
$
)*
+),-.+/$
!
#
0 1
0 -.-) %
"
& '
features.4 (Fig 2)
ͳ
Introduction
variations available.5
initial placement of the implant into the bone and the subsequent
newer connections.
external — connection.
ʹ
Introduction
restoration
threaded shaft of a dental implant 10,11 with the same taper. Other
their use of joint designs (e.g. bevel, butt), or the numbers of 'hexes'
8,9,10 present for the restorative phase.
by using angulated friction fit internal wall into which an abutment with
͵
Introduction
welded.
because the intimate contact between the implant and the abutment
Ͷ
Introduction
implant system may cause not only frequent screw loosening and
chronic fracture of the screws but can also cause the accumulation of
integration,
Several published studies have shown that crestal bone loss occurs
the IAJ is situated relative to the original level of the bony crest.20
associated.
The gap between implant and abutment is an ideal place for bacterial
proliferation and fluid micro leakage what can lead to peri-implantitis .It
ͷ
Introduction
is important to say that the force applied in the tightening torque is only
implant were proper because high levels of tightening torque would not
perspectives.
Review of Literature
each group were further divided into 2 groups, 1 group was tightened
device was used, the magnitude of the torque force that was
Review of Literature
element models.
Within the limitations of this study, it was suggested that the platform
ͺ
Review of Literature
The distance of the point of force application from the implant platform
was 8 mm; the gradation of the force was 0.3 N/m s. The interface of
per second).
ͻ
Review of Literature
connection.
preload. This study did not look into other types of mechanical
ͳͲ
Review of Literature
(11.1%) in each group. This implied that both premachined and cast
ͳͳ
Review of Literature
Axial load were applied on the top center of the prosthesis and on the
fractured.
ͳʹ
Review of Literature
The initial removal torque for Ta screws required higher force than
WC/CTa screws, whereas post load removal torque for Ta screws was
was higher than for WC/CTa screws, indicating that the WC/CTa screw
Select
ͳ͵
Review of Literature
Artificial teeth were arranged on the framework, and the prosthesis was
conditions (4-point load; 2-point anterior load; and 2-point lateral load).
connection at the connection level and at the apical level. With the 2-
lower stresses at the connection level both in the loaded and non-
ͳͶ
Review of Literature
joint with a conical tapered neck (EJ), internal joint with a straight neck
(IJ), and conical joint with a reverse conical neck (CJ). 3D FEA was
micro movement were analyzed. It was concluded that the shear stress
was concentrated on the mesiodistal side of the cortical bone for EJ.
EJ had the largest amount of abutment micro movement. While the von
Mises and shear stresses around the implant neck were concentrated
on the labial bone for IJ, they were distributed on the mesiodistal side
of the cortical bone for CJ. CJ had the least amount of abutment micro
implant collar and the marginal bone. These findings should help guide
ͳͷ
Review of Literature
implants. The first phase of this study used nine EH implants and
abutments that were divided into three groups with different amounts of
assemblies that were similar to those used in the first phase. These
turbidity of the broth every 24 hours for 14 days, and the bacteria
viability was tested after that period. The statistical evaluation was
samples from group T10 and one from T20 presented positive results
The tightening torques did not statistically affect the micro leakage in
ͳ
Review of Literature
torque did not show any bacterial contamination. It was concluded that
for groups SWT-EH (1.31), REG-EH (1.55), SWT-IH (1.83) and REG-
IH (1.82) indicated that fatigue accelerated the failure of all groups. The
in agreement with the lower reliability observed for the external hex
ͳ
Review of Literature
showed the highest strain values, and the internal hexagonal implant-
used for single tooth implant and concluded that with a conical implant–
with retention elements at the implant neck, and with suitable values of
interface between implant and abutment and the use of the torque
ͳͺ
Review of Literature
bone levels.
peri-implant bone level before and after the occlusal loading of dental
implants. First, the level of peri-implant crestal bone does not differ
(healing phase, loading phase 1, and loading phase 2), with it being
finite element model, with the maximum von Mises stress set as output
ͳͻ
Review of Literature
the fixtures was higher than the stress value in bone in all of the
samples. Stress values in sample B were the lowest amongst all of the
models. Any alterations in the amount and direction of the 100-N axial
the connection results in a minimal micro gap between the implant and
ʹͲ
Review of Literature
is important to say that the force applied in the tightening torque is only
implant were proper because high levels of tightening torque would not
the screw joint stability. The results of this study showed that the
external butt joint was more advantageous than the internal cone in
terms of the post load removal torque loss. For the difference in the
ʹͳ
Review of Literature
observed that placing the smooth part of the implant below the alveolar
bone loss.
and after the leakage tests. Three different abutment types (Internal
different geometries were connected to its own implant fixture. All the
hex zirconium abutment. Observed mean torque value loss was also
ʹʹ
Review of Literature
atmosphere inside a box, where they were inoculated with 0.3 μ l of the
controls were bred in blood agar to analyze the colonies formed. The
ʹ͵
Review of Literature
were observed between the different hex systems from the same
The finite element analysis found that the Ankylos implant system has
implant system.
Lower stress levels in the periimplant bone with a more uniform stress
distribution were found for the Ankylos implant system with a platform-
ʹͶ
Review of Literature
dental implants divided into two equal groups were compared; Group 1
for 14 days. The specimens were disassembled and the inner surfaces
of the implants were sampled by sterile paper points. Then the paper
points were immersed in test tubes containing sterile BHI broth. From
the broth, culture was done on blood agar plates and incubated at 37
Kai Blum et al (2015)45, evaluated the micro gap formation and wear
ʹͷ
Review of Literature
connection (IAC) after 200,000 cycles and 1 million cycles were further
characterized using SEM. All implants exhibit a micro gap between the
implant and abutment prior to loading. The gap size increased with
cyclic loading with its changes being significantly higher within the first
around external hexagon (EH) and Morse taper (MT) implants with
casting technique (CT), and laser welding (LW)) by using photo elastic
The samples were divided into six groups. G1: EH implants with DB
and acrylic resin; G2: EH implants with titanium infrastructure CT; G3:
ʹ
Review of Literature
(except for CT). Different bar types submitted to immediate loading not
pieces.
for forty-eight hours and then bacteria amount was measured inside
tested implants, bacteria were found in the inner side, with a median
ʹ
Review of Literature
significant at each time point. Ditron Implant IAC (MPI, Ditron Dental,
and the screw coating on the loosening torque of screws used for
and clinical parameters on the removal force for implant systems that
workbench for removal force. Four levels were chosen for the preload,
FP, and the taper mismatch D q; 3 levels for the wait time t; and 2
ʹͺ
Review of Literature
The results of this study can be used for decision making in the design
and use of TIF type systems. The study supports the use of artificial
designs.
ʹͻ
Review of Literature
all IACs tested with undetectable leakage after 7 days. Each of these
explore this.
͵Ͳ
Review of Literature
the thicker the tissue, the lower the reverse torque values noted.
the potential benefits with the use of Morse taper dental implant
inflammation and possible bone loss with the peri-implant soft tissues.
were tested.
Within the limitations of this study, the results indicate the possibility
͵ͳ
Review of Literature
internal part of the dental implant. These findings suggest that a one-
yeast Candida albicans. Within the limitations of this study, the results
between the abutment and the internal part of the dental implant These
in-tube interfaces.
med data base. The search was focused on ability of different implant
͵ʹ
Review of Literature
implant interface. Micro leakage was very less in Morse taper implants
promising in case of static loading and also showed less micro leakage
studies indicate that the internal connection, together with the Morse
tooth implants.
͵͵
Review of Literature
factors upon the loosening of abutment screw. The current study has
implant-supported restorations.
influence.
It has been concluded that the optimal choice and number of implant
long-term complications.
͵Ͷ
Review of Literature
cost and technique sensitivity incurred with powdered screws may not
applied torque levels based on the screw elastic limit. The static
the abutment screw was inserted tightly This study concluded that the
applied more tightly on the screw. Less torque loss was also found
͵ͷ
Review of Literature
and abutment. The implants that deformed in this test were examined
studied from the viewpoint of microleakage from the gap between the
abutment screw.
͵
Review of Literature
systems.
Within the limitation of the study, CC was stable even after the loading
microgap between the implant body and the abutment, among the
͵
Review of Literature
and the bone when subjected to various loading conditions. The result
showed that vertical loads closer to the implant long axis produced
much lower Von Mises stresses than when loads applied away to the
the least amount of stress to the surrounding bone and most of the
most of these stresses were located at the 1st thread of the implant.
All the four implant abutment designs studied showed the least
surrounding bone, when the loads applied were close to the long axis
favorable connections, and the parallel hex connection was the least
stable on loads applied along and closer to the implant axis, but
͵ͺ
Review of Literature
long screws with more number of threads for effective preload retention
by the screws.
finite element analysis (3D-FEA The regions with the highest von Mises
stress results were at the bottom of the initial two threads of both
Under the simulated chewing loads, the von Mises stresses for both
͵ͻ
Review of Literature
the materials analyzed which thus supports the clinical use of both
prosthetic abutments.
switching helps preserve crestal bone around the implants and this
placement permit.
ͶͲ
Discussion
What Is An Implant?
A dental implant is an artificial root that replaces the natural tooth root.
(Fig. 3)
the implant.
IAC is the point of contact between the surgical and prosthetic phase
force within the screw called the preload. It is defined as the tension
Ͷͳ
Discussion
input torque to friction. The head of the screw is wider than the thread
diameter and for an abutment most often is flat. Tapered head design
reduces the clamping effect and reduces the tensile force in the
threads of the screw. The tapered screw head distorts and aligns
flat-head screw distributes forces more evenly within the threads and
the head of the screw and is less likely to distort a nonpassive casting.
As a result, the dentist can identify and correct the nonpassive casting.
As such the abutment head also should be flat on top to increase the
clamping force in the screw head and the tensile force in the threads.
The thread design and number of threads are also primary factors
Ͷʹ
Discussion
shaped 30 degree angle .The fixture design allows the preload torque
degree angle of the female component of the screw to help fixate the
Screw Diameter
The diameter of the screw may affect the amount of preload applied to
the implant system before deformation. The greater the diameter, the
higher the preload that may be applied and the greater the clamping
less often and can take a higher preload compared with coping screws.
the length of the screw, the thread and groove shape, the number of
Ͷ͵
Discussion
types.69
ͶͶ
Discussion
Slip fit: slight space exists between the connecting parts, and the
connection is passive.(Fig. 9)
Friction Fit: no space exists between the components and the parts
externally.
a. Octagonal,
b. Hexagonal,
c. Conical,
e. Spline, etc
Ͷͷ
Discussion
because implants were used to treat fully edentulous patients and were
The external hex portion of the implant was added to the design to
device (fixture mount) during the initial placement of the implant into
extension.
Ͷ
Discussion
Ͷ
Discussion
The abutment modifications that have occurred are vast and complex.
height and width. Besides altering the size, other modifications were
also made in an effort to improve upon the original external hex design.
types of connections.
The initial relationship between the abutment and implant body was
mainly associated with external connections. Over time the simple butt
Prosthetic success remains high with the external connection but the
Screw loosening can create serious challenges for the clinician and
Ͷͺ
Discussion
This is because with this connection type the screw alone secures the
abutment.
14)
Ͷͻ
Discussion
Currently available external hex heights range from 0.7 to 1.2 mm and
ͷͲ
Discussion
tapered hexagon, external octagon and the spline dental implant are
now available
hexed implants was designed with a 1.7 mm-deep hex below a 0.5 mm
wide, 45° bevel. This was the corevent implant developed by Nickzick
The design has been proven to distribute intraoral forces deep within
engagements that create a stiff, unified body that resists joint opening,
wall engagement with the implant that buffers vibration, the potential for
ͷͳ
Discussion
between the two which provides superb microbial seal and joint
stability.
Advantages:
loosening
x Extensive flexibilty
components)
ͷʹ
Discussion
is, a hexagon recessed into the body of the implant. As the internal
geometry is a hexagon, the abutment can fit over the implant at every
60 degree rotation
of the implant over the abutment, but not at any other intermediate
length of the internal connection. This implant has evolved from the
original core vent implant, with a hollow basket design to the tapered
screw vent
implant.. The design has been proven to distribute intraoral forces deep
within the implant and thus improve the implant-abutment joint stability
ͷ͵
Discussion
abutment on the implant for every 30 degree rotation, thus useful when
implant.
for positioning of the abutment over the implant at only 120 degree of
Nobel Bio care, which was the replace select system. It represents the
ͷͶ
Discussion
connecting the implant and the abutment allowing for positioning of the
includes a tapered projection from the implant abutment, which fits into
a tapered recess in the implant. There is a friction fit and cold welding
depends on this friction fit for elimination for rotation at the implant-
18)
ͷͷ
Discussion
related industries.
precise fit between the implant and the abutment, it did not allow for
and also precise transfer of the implant position to the master cast. As
a result, only one transfer system and one analog are required.
This implant is marketed by Astra Tech. The fixture and abutment are
The conical design seals off the connection and decreases micro
ͷ
Discussion
conical neck and TiO blast surface. Micro threads on the fixture top
prevent concentration of the stress around the alveolar ridge crest and
This is a true morse taper implant with an angle of taper: 1.5 degree is
available from Bicon implants. The Bicon locking taper abutment has
into the matching socket in the implant. A high clamping force between
The high friction force is the result of relative slip between the two
surface oxide layers breaking down and causing cold welding at the
ͷ
Discussion
led to the development of the internal hexagon design. This has further
maximum for the 12-point internal hex and minimum for the 3-point
internal tripod.
ͷͺ
Discussion
MAKING)
extends above the coronal surface of the implant. The connection can
components and the parts are literally forced together. The joined
Dr. Gerald A Niznick was the first one to suggest modification to the
the implant body which was believed to help in better stress distribution
ͷͻ
Discussion
and provide better and more prosthetic options. Almost all vitro studies,
believed to give better marginal seal and reduce the micro movements
also being tried in newer implant systems. Some screw less implant
Ͳ
Discussion
Most dental implant systems consist of two main parts: the abutment
hard tissues.78-80
therapy can be used to treat peri-implant disease, bone loss that has
interface.81
Three main factors have been identified as possible causes for the
ͳ
Discussion
abutment connections are the conical and the butt-joint, the latter type
Micro motion and stress are believed to play pivotal roles in micro gap
of micro motion.
ʹ
Discussion
Unlike the internal conical abutment, the trilobe abutment produced the
polygonal profile.
micro fracture, and therefore for micro gap formation. Micro motion is
components.85,86
of dental implants.76,82
bacterial leakage from the IAI was same in both the methods 89
͵
Discussion
the study showed that less micro leakage was shown by Morse taper
was more than Morse taper implants with gap of 2-3 μ m. When 30 N
microleakage.90
implants91.
3. Screw Mechanics
fixate an implant into bone and load the bone after hard tissue healing.
Ͷ
Discussion
problems.
bone and is several millimeters below the margin of the tissue. It may
ͷ
Discussion
1. Preload
2. Component fit
3. Platform dimension
4. Hex height
internal implant threads and the prosthetic screw threads. Unlike screw
Discussion
(Winkler et al. 2003; Cantwell and Hobirk 2004). Torque loss will
the prosthetic screw at each recall visit. This has not been shown to
al. 2011).
Preload
material, screw head and thread design and surface roughness 93-95.
Discussion
1. Screw head: The head contains the driver fitting site, which is
used to rotate the screw into position. Various driver fitting site
(square), hex, and star. By far, the most common type used in
3. Thread: Without going into too much complexity, the thread can
ͺ
Discussion
4. Connection Geometry:
influence the amount of micro motion, stress distribution and micro gap
applied torque develops a force within the screw called the pre-load. As
of the screw pulls the 2 parts together, creating a clamping force. The
ͻ
Discussion
joint are greater than the clamping forces holding the screw unit
determined by the applied torque and other factors, such as the screw
of the joint and result in screw fatigue, loosening, and failure. Too large
a torque may strip the screw threads.Increasing the torque will increase
the preload. Increasing the preload maximizes the stability of the screw
Ͳ
Discussion
Dellinges and Tebrock found that the average torque applied with a
Carr et al.101 and Byren et al102. reported that the fitting of the implant-
system. Moreover, under such conditions, the preload also reaches the
maximum value.
Jaarda et al found that test subjects with little implant experience were
Stress Distribution
ͳ
Discussion
success rates.
ʹ
Discussion
therefore causing micro gap formation. 110 Micro motion and stress are
Chewing forces of adult individuals with natural dentition and those with
N to 178N119-125.
connections since these are more realistic than using occlusal forces
alone122,126.
͵
Discussion
that the connection screw receives all static and dynamic lateral force
that such forces can cause loosening and/or fracture of the connection
screw.
connection system, and the tension in the neck of the screw shows the
maximum values.
Sutter et al129. had shown that the conical angled design could reduce
connection.
Merz et al130. considered the internal conical design and the use of 1-
Ͷ
Discussion
Nobel replace.
the four implant abutment designs showed the least amount of stress
on the various areas of the assembly and the surrounding bone, when
the loads applied were close to the long axis of the implant. Both the
the parallel hex connection was the least favorable design. The tri-
along and closer to the implant axis, but produced high amount of
stress on the bone on oblique loads applied at a distance from the long
ͷ
Discussion
5. Platform Switching
around dental implants. This term was coined by Gardner in 2005 and
the inflammatory reaction medially, away from the crestal bone and
Step, created between abutment and implant allows the biologic width
using abutments with a diameter smaller than the implant neck or body
Discussion
which involves an inward bone ring in the coronal part of the implant
platform
IAJ.
space is limited.
esthetics
Discussion
area where biologic attachment may take place, thus limiting the
crest.
ͺ
Discussion
Disadvantages
abutments are the micro gap, consecutive fatigue and wear at the
interface.
ͻ
Discussion
ͺͲ
Conclusion
minimal micro gap between the implant and the abutment, which in turn
leakage what can lead to peri-implantitis .It is important to say that the
force applied in the tightening torque is only valid if the machining and
because high levels of tightening torque would not produce the desired
ͺͳ
Conclusion
ͺʹ
References
ͺ͵
References
ͺͶ
References
624.
15. Lang LA, May KB, Wang RF. The effect of the use of a counter-
16. Song Park, Sang Yong Won, Tae Sung Bae et al. Fatigue
1994; 9:48-62
2007; 16(2):165-75.
ͺͷ
References
16(4):475-485.
21. Lang LA, May KB, Wang RF. The effect of the use of a counter-
ͺ
References
80.
29. Park JK, Choi JU, Jeon YC, Choi KS, Jeong CM. Effects of
ͺ
References
33. Silva-Neto JP, Prudente MS, Carneiro TD, Nóbilo MA, Penatti
Anchieta RB, Martini AP, Assunção WG, Silva NR, Coelho PG.
ͺͺ
References
37. Lin MI, Shen YW, Huang HL, Hsu JT, Fuh LJ. A retrospective
40. Shin HM, Huh JB, Yun MJ, Jeon YC, Chang BM, Jeong CM.
ͺͻ
References
25(11):1301-3.
1; 1(1):1-5.
ͻͲ
References
47. Odo CH, Pimentel MJ, Consani RL, Mesquita MF, Nóbilo MA.
Jul; 9(3):99.
51. Abraham HM, Philip JM, Jain AR, Venkatakrishnan CJ. The
ͻͳ
References
2016.
53. Zeno HA, Buitrago RL, Sternberger SS, Patt ME, Tovar N,
10(1):148.
ͻʹ
References
36(2).
11(6):ZE10.
60. Jeng MD, Liu PY, Kuo JH, Lin CL. Load Fatigue Performance
ͻ͵
References
61.
18(2):139
ͻͶ
References
18(2):161.
Jan 1; 1(1):13.
Publication 2005.
ͻͷ
References
937.
75. Ding TA, Woody RD, Higginbottom FL, Miller BH. Evaluation of
ͻ
References
76. Park IS, Won SY, Bae TS, Song KY, Park CW, Eom TG, et al.
55
ͻ
References
17:623–30
32:123–32.
ͻͺ
References
72
90. Verdugo CL, Núñez G J, Avila AA, San Martín CL. Micro
2014; 25:1078-83.
ͻͻ
References
171.
93. Lang LA, Kang B, Wang R-F, Lang BR. Finite element analysis
546.
94. Stuker RA, Teixeira ER, Beck JCP, da Costa NP. Preload and
16(1):55–58.
95. K-T Yao, H-C Kao, C-K Cheng, H-W Fang, S-W Yip, M-L Hsu
23:1–6.
96. Saidin S, Abdul Khadir MR, Sulaiman E, Abu Kasim NH. Effects
ͳͲͲ
References
2012; 40:467–474.
flat top versus conical. Clin Implant Dent Relat Res. 2000; 2:33–
41.
ͳͲͳ
References
80:184- 192
103. Geng JP, Tan KBC, Liu GR. Application of finite element
ͳͲʹ
References
108. Chun HJ, Cheong YS, Han JH, et al. Evaluation of design
109. Segundo RM, Oshima HM, da Silva IN, Burnett Jr LH, Mota
55–9.
110. Segundo RM, Oshima HM, Silva IN, Ju´nior LH, Mota EG,
112. Craig RG, O’Brien WJ, Powers JM. Dental Materials. 6th ed.
113. Craig RG, Ward ML. Restorative Dental Materials. 10th ed. St
ͳͲ͵
References
114. Gibbs CH, Mahan PE, Mauderli A, Lundeen HC, Walsh EK.
229.
116. Koolstra JH, Van Eijden TMGJ, Weijs WA, Naeije M. A three-
ͳͲͶ
References
121. Clelland NL, Lee JK, Bimbenet OC, Brantley AW. Basic
ͳͲͷ
References
570.
1997; 12:237
ͳͲ
References
15:519– 526.
133. Nebot XV, Ciurana XR, Alonso CR and Torres SM. Benefits of
1000.
ͳͲ
References
2005; 17:735-40
ͳͲͺ
7DEOHRIFRQWHQWV
,QWURGXFWLRQ
5HYLHZRI/LWHUDWXUH
'LVFXVVLRQ
&RQFOXVLRQ
5HIHUHQFHV
ͳͲͻ
Buy your books fast and straightforward online - at one of world’s
fastest growing online book stores! Environmentally sound due to
Print-on-Demand technologies.
Buy your books online at
www.morebooks.shop
Kaufen Sie Ihre Bücher schnell und unkompliziert online – auf einer
der am schnellsten wachsenden Buchhandelsplattformen weltweit!
Dank Print-On-Demand umwelt- und ressourcenschonend produzi
ert.
Bücher schneller online kaufen
www.morebooks.shop
KS OmniScriptum Publishing
Brivibas gatve 197
LV-1039 Riga, Latvia info@omniscriptum.com
Telefax: +371 686 204 55 www.omniscriptum.com