Table 37 Common laboratory tests for soils (2 of 10)
Category of test Name of test or Where details Remarks
parameter can be found measured Classification Particle size Coarse grained - BS 1377-2 Sieving methods give the grading of soil tests analysis sieving (BS EN ISO particles coarser than silt (<0.063 mm). (continued) 17892-4) When the sample contains silt or clay the test is done by wet sieving. The relative proportions of silt and clay are determined by means of sedimentation tests. Fine-grained - BS 1377-2 Hydrometer or pipette analysis. sedimentation (BS EN ISO 17892-4) Automated Equipment Automated methods including x-ray methods primarily manufacturer sedigraph, laser or granulometer for fine grade methods methods are only to be used after they have been shown to give equivalent results as the standard methods. It is important to ensure that the sample sizes used are relevant for the particle sizes tested. Chemical tests Chemical tests for civil engineering purposes might be carried out in geotechnical laboratories. Most of the methods are described in BS 1377-3. These methods do not generally include modern analytical techniques. Analytical methods may be used if they are shown to give equivalent results to standard methods. The sample preparation should follow the methods in BS 1377-3 and the results reported as specified in this British Standard. pH value BS 1377-3 Measures the acidity or alkalinity of the soil or water. Needs to be performed as soon as possible after sampling. Values of pH might change during storage when pyrite oxidizes as the disturbed materials are exposed to the air. These materials are generally grey and dark grey clays and mudstones. It is often carried out in conjunction with sulfate tests. Excessive acidity or alkalinity of the groundwater in soils can have detrimental effects on buried concrete, cause corrosion of metals and use of resinous materials is unsuitable in alkaline soils.