You are on page 1of 5

A CASE OF IMPLEMENTING ON-LINE ANALYSIS OF IRON ORE SLURRIES

Daniel Michaud1, Guy Chevalier2

1
COREM, 1180, rue de la Minéralogie,
Québec (Québec) G1N 1X7, CANADA
2
Québec Cartier Mining Company, 24, boul. des Îles,
Port-Cartier (Québec) G5B 2H3, CANADA

Abstract: An account is presented of the development and implementation of an on-line


analyzer in an iron ore pelletizing plant. The analyzer uses Laser-Induced Breakdown
Spectroscopy (LIBS) technology to interrogate the target material and determine the
concentration of key elements. Following primary slurry sampling, prior conditioning is
necessary to ensure reliable determinations. The instrument has proved to be a very
sensitive detector of process perturbations of either chemical or physical origin and has
enabled plant operators to exert tighter control over the chemistry of the final product.
Copyright © 2007 IFAC

Keywords: Process control, Control applications, Technology transfer, Spectroscopy,


Calibration

1. INTRODUCTION large acceptance in the industry due to reasons


related to prohibitive costs, lack of robustness,
For cost and quality control purposes, the ore demanding installations and security issues. Laser-
dressing industry needs practicable means of Induced Breakdown Spectroscopy (LIBS) is a
monitoring the chemistry of the ore as it is being relatively recent technology that has demonstrated
processed. At an early step in the process, the ore is good potential for real-time applications (Sabsabi et
ground to some minute size and from then on is al., 2004).
transported as solids in slurry flows. The next steps
aim to concentrate the ore by pulling out the valuable The paper reports on the development of a LIBS-
mineral and discarding the gangue material. based on-line analyzer and on its implementation in
Knowledge of the composition of the ore at some an iron ore pelletizing plant. The work stemmed from
given time or location along the process line enables an initial and intensive endeavour encompassing both
plant operators to maintain specified chemical laboratory and field tests (Turmel et al., 1999;
targets. This control becomes especially beneficial Barrette et al., 1999; Michaud et al., 2003; Barrette et
and timely when performed in real time. The most al., 2004). The objective was to monitor the
widely used method to accomplish this task is X-ray concentrations of carbon (occurring essentially as
fluorescence but when applied to the measurement of added coke), silicon (occurring as native silica and as
slurries most light elements fall out of reach because added silicates), magnesium, calcium and manganese
water absorbs their characteristic emission (Braden in order to gain higher control over the chemistry of
et al., 2002). Among these elements are carbon and the iron ore pellets. Substantial savings were foreseen
silicon, which are of paramount importance to iron from the use of such an on-line analyzer, the benefits
ore pelletizing plant operators. Other methods to spawning from a stabilized process, an increased
measure these elements exist but have not gained
product quality, and a better fuel partitioning The discrete components of the analyzer include an
(between coke and oil) in the induration step, etc. industrial Nd-YAG laser (model CFR200 from Big
Sky Laser Technologies) set to emit pulses of 150 mJ
at the fundamental wavelength of 1064 nm. The laser
2. EXPERIMENTAL beam is focussed on a free-falling section of slurry
using a lens of 40 cm nominal wavelength. The firing
LIBS is a growing technology offering the notable repetition rate is set at 2 Hz. The emission of the
advantages of speed (analyses produced in a matter plasma formed on the target is collected with a pair
of seconds), sensitivity (in the 10-100 ppm range), of lenses and transported to the spectrometer via a
and the capacity for multi-elemental analyses of cable of optical fibres. The spectrometer is of the
virtually all elements (Radziemski and Cremers, Czerny-Turner type (model Triax 550 from Jobin-
1989). LIBS uses energetic laser pulses to excite the Yvon) and the detector is an ICCD camera (model
sample and promote characteristic optical emission. iStar from Andor Technologies) with a grating of
This emission is collected with suitable optics and 2400 lines per mm. Using this combination, spectral
analyzed with an optical emission spectrometer. A windows of 12 nm in wavelength span can be
spectrum is acquired from which an interpretation in recorded. To cover all elements of interest in this
terms of elemental species and their intensities can application, three windows are necessary.
be completed. A calibration curve can then be
established to determine the concentration of each Following common practice, the intensities are
component in the sample. averaged over many consecutive laser shots (120
here) to reduce standard deviation of signals. In
A LIBS instrument was thus designed and addition, intensities are ratioed over a reference
assembled. The system is made up of three modules element (here an iron peak) in order to further
as shown in Figure 1. The first module is a slurry stabilize the measurements. Calibration curves and
conditioner; it receives the primary sample from the precision are calculated using these element ratios.
main process (about 150 litres of slurry per minute), All control and data software for the LIBS instrument
homogenizes the material and controls the density of was developed in-house.
the slurry (solids content) prior to extracting a final
sample (about 30 L/min) for the LIBS analyzer. The Particular efforts were invested to build in the
second module is the measurement “head” – an instrument the properties of robustness, reliability
enclosure holding the laser and the optical and security. Robustness was required on account of
components. The module is divided into three using scientific instrumentation in a tough industrial
compartments, one of which is the sample chamber environment. The spectrometer in particular is a
through which the slurry column falls; this chamber sensitive tool and was consequently seated on a
is kept under positive pressure with flowing air and vibration-damping stage. Reliability refers to the
periodically water-cleaned with automatic sprinklers. quality of the data acquired, in terms of precision and
The last module is the command module, which accuracy. Security measures were needed to confine
contains the analytical instrumentation and all the laser radiation inside its enclosure and to halt the
control and communications means. The command analyses in response to different external failures.
module is Nema4 compliant and includes a cooling The absence of flowing slurry in front of the laser
unit. beam, the shortage in water feed to the cooling unit
or to the laser head and the lack of air pressure in the
pneumatic valve lines are examples of detected
failures.

3. RESULTS

Before delivering the newly built LIBS analyzer to


the pellet plant, tests were conducted in COREM’s
pilot plant using a closed loop of circulating iron ore
slurry. Slurry flow (at 30 L/min) and homogeneity
were checked to satisfaction; control over the percent
solids in the slurry reached ±1% (around the 70%
level). Laser module toughness and watertightness
were successfully checked. Software and safety
devices were challenged over extended periods of
time. Applying the method of standard additions
(using compatible ores), calibration curves were built
to determine the measurement precision. Precision is
Fig. 1. Schematic of the application including the expressed as the smallest detectable variation in
components of the on-line analyzer. absolute concentration (SDVAC). This indicator is
computed as the ratio of twice the standard deviation 4.5
of the measurements over the regression slope and
thus holds to a confidence level of 95%. 4

Results of pilot plant preliminary tests appear in 3.5

Concentration (% w/w)
Table 1. Linear responses (R²) are excellent over the
concentration spans of each element measured. The 3
relative slope expresses the percentage of change of 2.5
the signal as a reaction to a 1% change in
concentration; it is an indication of the amplitude of 2
the response. It is thus seen that the manganese
response is strong while that of silicon is not. As can 1.5
be expected, this indicator is anti-correlated with the
precision obtained (SDVAC). The relative SDVAC 1
Si LIBS
on the other hand weighs the response of each 0.5
Si Chem
C LIBS
element against its concentration in the ore; on this C Chem
basis, the quality of the silicon measurements is very 0
good compared to the other elements. 9:30 10:30 11:30 12:30 13:30 14:30

Time
Table 1 Results of pilot plant preliminary tests Fig. 2. Tracking of silicon and carbon concentrations
in the iron ore slurry on September 7, 2005.
C Si Mg Ca Mn
Concentration span 0.32 − 2.62 − 0.25 − 0.54 − 0.11 −
(%) 1.12 3.31 0.55 0.87 0.32
One month after start-up, the tracking graph
reproduced in Fig. 2 was obtained. During the trial,
Relative stdev (%) 7.0 2.1 2.2 1.7 1.5 some perturbation of the process promoted the
R² 0.9986 0.9906 0.9981 0.9970 0.9766 variation of the concentrations of carbon and silicon.
The correlation between the LIBS and laboratory
Relative slope 119 19 70 36 273
(% / %)
results scores 0.91 for silicon and 0.96 for carbon.

SDVAC (%) 0.07 0.21 0.09 0.07 0.01 During the first year of operation, the up-time for the
Relative SDVAC 9.7 7.0 21.7 10.0 6.5 instrument was over 90% (excluding those periods
(%) when the system was intentionally stopped). The
most time-consuming problem in the first months
was related to maintaining a steady flow of slurry;
The instrument was shipped and installed at QCM’s pipe clogging and pipe wear were especially
pelletizing plant in August 2005. QCM took charge troublesome; these problems were solved over time.
of the primary sampling of the process; and of the An optics alignment problem and a security failure
transportation and disposal of the slurry after use. occurred which necessitated corrective actions. Apart
The material to be measured was the output of the from these discrete events, the system only needed
ball mills, just prior to the final filtration and regular maintenance (to change laser flashlamps bi-
pelletizing steps, and occurred as a slurry holding yearly among other things). It is concluded that the
some 72% solids. A primary sampling device was instrument needs only minimal supervision – less
installed on the process line to extract slurry at a rate than 10 hours a week on the part of QCM personnel –
of about 150 litres per minute. A secondary sampler in line with prospective specifications.
was inserted after the LIBS measurement head to
extract samples for calibration purposes. On-site performance evaluation of the analyzer was
done once the instrument and sample manipulation
The analyzer was started and set to provide refreshed were mastered. This consisted in determining the
concentration values every three minutes (each one accuracy of the measurements with respect to the
of the three acquisition windows is analyzed for one analyses produced by QCM’s chemistry laboratory.
minute). A connection was established between the The precision (smallest detectable variations in
instrument’s command unit and the plant’s local concentration) determined for the measurements is
control system so that the LIBS data could be comparable to what was obtained in the preliminary
exported, formatted and displayed on the operator’s tests in COREM’s pilot plant. The accuracy though,
control screen. Local personnel was trained to look which was anticipated to hold within the precision
over the instrument on a daily basis; COREM span, was at times not met. The most problematic
remained as backup for servicing instrumental and case was that of silica for which absolute biases
technical issues. exceeding 0.25% were noted when comparing the
LIBS results with daily check tests.
4. DISCUSSION measurements add up statistically to represent tons of
actual material as it flows by, compared to the minute
Calibration of the LIBS instrument is needed in order grams of slurry bleeded off the process line every 15
to maintain the analyzer in a proper working state. minutes and averaged over some 2 to 4 hours.
To this purpose, periodic calibration campaigns are
to be performed (a few times a year) by intentionally Any steady variation of the LIBS signal can readily
upsetting the process. This is done by slowing down be interpreted as a deviation from a prior state. In this
slurry pumps, altering mill throughputs, temporarily manner, the process operator can bring down his
modifying additives feed, or by any other means that reaction time from hours to some 15 minutes or so,
will induce a sizable variation of the concentrations thereby gaining timely and tighter control over the
of the components in the slurry solids. Taking a process. This advantage, in turn, translates in
number of samples during the event provides data to products of higher and consistent quality.
compute a new calibration curve for each element
measured. Afterwards, daily checks serve to follow Furthermore, the on-line analyzer turned out to be a
the instrument’s routine level of performance; in this very sensitive detector of various process upsets, like
case, only one sample is taken to establish the tuning mill failures, occasional use of recycled water, etc.
between the LIBS and the lab. Also, the instrument is a powerful tool to study the
process itself. For example, establishing the grinding
The accuracy problem observed (i.e. the fit to the kinetics of the ore (around 20 minutes) with respect
laboratory results) is limited to the comparison of to the additives coke (15 minutes) and dolomite (18
LIBS results with routine checks performed on a minutes) was merely a straightforward task with the
daily basis. Indeed, when specific calibration on-line instrument.
campaigns are run by intentionally perturbing the
process, the quality of the calibration curves and the
precisions obtained are very good (similar to results 5. CONCLUSION
from preliminary tests). It was observed though, that
daily check results line up on a slightly different An on-line analyzer based on Laser-Induced
calibration slope. It is thus concluded that provoked Breakdown Spectroscopy was designed, assembled
process upsets are not truly representative of real and installed in an iron ore pellet plant. The field
process dynamics. This situation may arise from the application consists in measuring elements C, Si, Mg,
fact that over the short period during which a Ca and Mn in the solids embedded in a slurry; the
perturbation occurs, long term variables such as objective is to check compliance with prescribed
particle grain size and ore matrix composition remain concentration levels. Tracking of these elements in
essentially constant; conversely, significant intervals of 3 minutes is a considerable enhancement
excursions may be experienced on longer time relative to the pre-existing situation (hours). The
scales. Based on this reasoning, it was proposed that precision achieved is sufficient to enable plant
practicable calibration curves should be constructed operators to detect any process upsets and apply
by accumulating daily check results rather than corrective measures in a timely manner. The on-line
relying on calibration curves derived from analyzer can also be used to study the process itself
intentionally perturbed process situations. by performing tests to determine process dynamics or
the response to some deliberate action. The LIBS
In addition, it was found that calibration (for silica analyzer is up and running in the pellet plant since
especially) must be tailored to the current pellet type September 2005.
in production in order to correctly account for the
different yields observed as a function of the pellet
composition. (Iron ore pellet plants produce pellets ACKNOWLEDGEMENTS
of different “recipes”, varying in silicate content,
notably, to meet customer specifications.) It is A commitment for quality and close collaboration
argued that the dual nature of the silicon-bearing between the research and industrial partners are two
minerals in the slurry (silica in the native ore and factors that have significantly contributed to make
silicates in the additives) is responsible for this this project a success. The authors wish to thank
situation. Indeed, indications to the effect that silica Steeven Dubord and Stéphane Morin from QCM for
yields less than silicates under LIBS excitation have their technical implication in looking over the
been accumulated but remain to be verified. installed instrument. The devoted involvement in the
project of COREM’s Éric Proulx, Luc Tardif,
In spite of the slight bias in accuracy, the instrument Frédéric Michaud, Pierre Laprade, Rémi Leclerc,
remains a very capable trend monitor for following Daniel Mainville and Jean-Guy Chartrand is also
the chemistry of the ore in quasi real-time (three- gratefully acknowledged.
minute intervals between measurements). The on-
line measurements display a clear advantage over
conventional average-sampling and delayed
chemical assaying. Indeed, the continuous
REFERENCES

Barrette, L., J.G. Chartrand, D. Michaud, M. Fortin


and G. Chevalier (2004). In: Proceedings of the
36th Annual Meeting of the Canadian Mineral
Processors, paper #17.
Barrette, L., S. Turmel, J.A. Boivin, M. Sabsabi, T.I.
Martinovic and G. Ouellet (1999). On-line ore
slurry monitoring using laser induced plasma
spectroscopy. In: Proceedings of the 38th Annual
Conference of Metallurgists of CIM (D. Hodouin,
C. Bazin and A. Desbiens, editors), pp. 17-27.
Braden, T.F., M. Kongas and K. Saloheimo (2002).
On-line composition analysis of mineral slurries.
In: Mineral Processing Plant Design, Practice
and Control Proceedings 2002, pp. 2020-2048.
Michaud, D., E. Proulx, J.G. Chartrand and L.
Barrette (2003). Shooting slurries with laser-
induced breakdown spectroscopy : sampling is the
name of the game. Applied Optics, 42, 6179-6183.
Radziemski, L.J. and D. Cremers (1989).
Spectrochemical analysis using laser plasma
excitation. In: Laser-Induced Plasmas and
Applications (L.J. Radziemski and D. Cremers,
editors), pp. 295-325.
Sabsabi, M., L. St-Onge and V. Detalle (2004).
Laser-Induced Breakdown Spectroscopy: A new
tool for process control. In: Proceedings of 16th
WCNDT 2004 – World Conference on NDT, paper
#679.
Turmel, S., J.A. Boivin, M. Sabsabi, P. Cielo, L.
Barrette, D. Fleury, G. Ouellet and T.I.
Martinovic (1999). On-line elemental analysis in
an iron ore concentrate slurry. In: 58th Ironmaking
Conference Proceedings, pp. 515-523.

You might also like