Professional Documents
Culture Documents
1
of Terms
Glossary of Terms Used in ICP-OES
Array Detectors: An array detector is a silicon chip with a one or two-
dimensional array of photosensitive material on its surface. The photons
strike the photosensitive regions called pixels and are converted into
electrons which are collected and read out as a charge. Examples of array
detectors include Photodiode Arrays (PDAs), Charge-Coupled Devices (CCDs),
Charge-Injection Devices (CIDs), and Segmented-array Charge-Coupled
Devices (SCDs).
Axial View: Axial viewing is the positioning of an ICP torch such that the
spectrometer views the plasma end-on. Usually, the torch is positioned in
a horizontal orientation for axial viewing. When configured properly, axial
viewing provides an increase in sensitivity, typically 5-20x greater than radial
viewing.
2
Background Noise: The background noise is the precision of the
measurement of the background signal at the analyte wavelength, and is
normally determined by taking the square root of the spectral background of
the instrument (in emission intensity). Refer to “Detection Limit”.
Blurring: The effect known as blurring occurs when the image quality on the
edge of a CTD detector is worse than the image quality in the center where
the optimum (blaze) wavelength falls. Blurring affects the practical resolution
at the detector and will vary from the theoretical resolution, if blurring is
severe. Blurring is measured by comparing theoretical resolution to practical
resolution. The ratio should be as close to 1:1, as possible.
3
Charge-Injection Device (CID): A charge-injection device is a two-
dimensional array detector that reads the charge accumulated on the
photosensitive surface and then stores it by injecting it into the silicon
substrate. The benefit of CID technology is that the charge is not destroyed
after being read-out, allowing for interrogation of the data at any time.
4
Diffraction Grating: A diffraction grating is an optical component with many
finely spaced, etched or ruled, grooves on its reflecting surface. Gratings are
used to disperse light according to wavelength and/or spectral order.
5
Echelle Grating: Echelle gratings are precisely-ruled diffraction gratings used
in ICP spectrometry that utilize very high angles of incidence, very low ruling
density (grooves/mm), and provide the use of high orders of diffraction,
typically in the range of 20-120th order, using a prism as the 2nd dispersion
optic. The main benefits of Echelle gratings for ICP-OES is that they provide an
optimum combination of resolution, dispersion, and high light throughput for
multi-element detection on a small CTD detector.
Entrance Slit: A narrow slit through which light enters a spectrometer. The
size of the entrance slit will determine the resolution of the spectrometer.
Exit Slit: A narrow slit through which light exits the spectrometer before it is
converted into an electric current by the detection system. The size of the exit
slit will dictate the resolution of the spectrometer.
Fixed Optics: Fixed optics do not move, such as those found in all direct-
reading echelle sequential and simultaneous spectrometers. Moving optics,
such as those found in all peak search (i.e. Czerny-Turner) spectrometers,
are inherently unstable and require integration at several points to acquire
correct peak wavelength data. This leads to extended analysis times and
incorrect peak wavelength identification. Fixed optics integrate directly on-
peak, and therefore are not subject to the pitfalls associated with moving
optics.
6
Free-Running (RF) Generator: This is a class of radio frequency (RF) generator
that uses the electronic constants of their circuit components to determine
and control the frequency of operation, without the use of external triggers
such as an oscillating crystal unit.
Focal Length: An optical element (e.g. mirror, prism, lens, or curved grating)
that focuses light on exit slits. The focal length is the distance from the
optical element to the point of focus. In the early days of spectrometer
development, increasing focal length (and consequently the length and size
of the instrument) was the only way to improve resolution. Improved grating
production methods have eliminated this requirement, allowing for the
compact instruments we enjoy today.
Grating: The grating is the optical element that disperses light. Echelle
gratings provide exceptionally high resolution, dispersion, and light
throughput. Conventional gratings are manufactured by the mechanical
ruling of concave blanks or via holographic reproduction techniques.
Concave ruled and holographic gratings typically do not compare in optical
performance with Echelle gratings.
Grating Drive: The grating drive is the mechanism used to move the
grating by sequential spectrometers without fixed optics. Common grating
drives include stepper motor-driven worm gears, sine-bar drives, as well as
galvanometer systems.
7
Hydride Generation (HG): A very sensitive analytical technique for determining
trace levels of volatile elements such as As, Bi, Sb, Se, and Te. Generation of the
elemental hydride is carried out in a closed vessel by the addition of a reducing
agent, such as sodium borohydride, to the acidic sample. The resulting gaseous
hydride is swept into the plasma where atomization occurs. Multi-element
quantitation is then carried out in the conventional way, by comparing the
emission intensity of unknown samples against known calibration or reference
standards. Hydride generation can also be used with AA using a special heated
quartz cell in place of the traditional flame burner head.
8
Instrument Detection Limit (IDL): The concentration equivalent to the
analyte signal which is equal to three times the standard deviation of a series
of 10 replicate measurements of the calibration blank signal at the same
wavelength.
Integration Time: This is the length of time the detector is allowed to acquire
an intensity measurement. The most precise measurements are taken at
the wavelength peak intensity. Some instruments that do a peak-search
do not integrate the actual peak signal, but instead calculate an intensity
measurement by fitting a curve to many measurements made at wavelength
intervals across the width of the peak.
9
Internal Standardization (IS): A quantitation technique used to correct for
changes in analyte sensitivity caused by variations in the concentration and
type of matrix components found in the sample. An internal standard is a
non-analyte element that is added to the blank solution, standards, and
samples before analysis. It is typical to add two or three internal standard
elements to the samples to cover all the analyte elements of interest
across the wavelength range. To work properly, it is critical that the internal
standards are of a similar excitation potential to the analytes and that they are
monitored at exactly the same time (simultaneously). The software adjusts the
analyte concentration in the unknown samples by comparing the intensity
values of the internal standard elements in the unknown sample to those
in the calibration standards. Because ICP-OES is prone to many matrix and
sample-transport-based interferences, internal standardization is considered
necessary to analyze most sample types.
Linear Dynamic Range: Linear dynamic range is the concentration range that
an element may be calibrated for at a particular wavelength. Linear dynamic
range is dependent on source conditions and detector, the latter of which is
often the limiting factor. For example, PMTs routinely accommodate a linear
dynamic range of 106 or better, but solid-state detectors, such as SCCD’s and
CID’s can be limited to 104. Lower linear dynamic range forces the use of
smaller integration times, sample dilutions, or use of alternative wavelengths
when exceeding the limits of the range.
Load Coil: Another name for the RF coil used to generate a plasma discharge.
Refer to “Matching Network (RF)”.
10
Matrix Interferences: There are basically two types of matrix-induced
interferences in ICP-OES. The first, and simplest to overcome, is often called a
“sample transport effect” and is a physical suppression of the analyte signal
brought on by the level of dissolved solids or acid concentration in the sample.
The second type of matrix suppression is caused when the sample matrix
affects the excitation conditions of the plasma discharge, which results in
varying amounts of signal suppression depending on the concentration of the
matrix components. The classical way to compensate for matrix interferences in
ICP-OES is to use internal standardization. Refer to “Internal Standardization”.
11
Nebulizer: A nebulizer is a sample introduction device that creates a fine
aerosol of sample solution for delivery to the plasma source. Various types of
nebulizers include:
12
Ultrasonic Nebulizer: Ultrasonic nebulizers are configured with a transducer
assembly containing a piezoelectric ceramic material bonded to a thin quartz
blank (often referred to as a “crystal”). While sample solution is allowed to
drip over the crystal, the ceramic material is caused to oscillate by a small RF
generator (usually operating at about 1.4 MHz). The energy provided to the
crystal results in the sample bursting into an ultra-fine cloud or mist. This mist is
passed first through a heated tube, which is hot enough to effectively drive off
much of the solvent from the analyte present. Subsequently, the mist is passed
through a cooling tube, which condenses and removes the separated solvent
via gravity drain. The end-product remaining is a dry aerosol, which is then
passed directly to the torch. The ultrasonic nebulizer excels over conventional
nebulizers in two ways: First, it is much more efficient at delivering sample to
the torch - 10% efficiency, as opposed to 3% for typical pneumatic nebulizers.
Second, by removing the solvent, less quenching of the plasma occurs,
providing better effective excitation within the ICP itself. In this way, ultrasonic
nebulizers typically provide a factor of 10 enhancement over standard
pneumatic nebulizers.
Normal Analytical Zone (NAZ): This is the region of the plasma where the most
stable and useful analyte emission originates.
13
Paschen-Rünge Spectrometer: This is an optical design used in emission
spectroscopy, where the diffraction grating mount and exit slits are stationary
and fixed in position in relation to the detector. Traditional mounting for an
ICP-OES is in the Rowland circle configuration.
PhotoMultiplier Tubes (PMT): PMT’s are one type of detector used in ICP
instruments. When struck by incident light, a PMT generates a current
proportional to the intensity of the input light. This current can be multiplied
by several orders of magnitude (typically 6-8), and processed as an output
signal by a readout device. Simultaneous ICPs use an array of fixed PMT’s
optimized for particular wavelengths. Sequential ICPs use one or two broad-
range PMT’s to cover the entire ICP spectrum. The signal from each PMT has
an Analog/Digital (A/D) converter that changes current to a digital signal,
registering intensity as counts integrated over a given time period.
Plasma Source: Refers to the RF hardware components that create the plasma
discharge, including the RF generator, matching network, plasma torch, and
argon gas pneumatics.
14
Plasma Torch: Another name for the quartz torch that is used to generate the
plasma discharge. The plasma torch consists of three concentric tubes: an outer
tube, middle tube, and sample injector. The torch can either be one piece,
where all three tubes are connected, or it can have a demountable design, in
which the tubes and the sample injector are separate. The gas (usually argon)
that is used to form the plasma (plasma gas) is passed between the outer and
middle tubes at a flow rate of 12–17 L/min. A second gas flow (auxiliary gas)
passes between the middle tube and the sample injector at 1 L/min, and is
used to change the position of the base of the plasma relative to the tube and
the injector. A third gas flow (nebulizer gas), also at 1 L/min delivers the sample,
in the form of a fine-droplet aerosol, from the sample introduction system and
physically punches a channel through the center of the plasma. The sample
injector is often made from other materials besides quartz, such as alumina,
platinum, and sapphire, if highly corrosive materials need to be analyzed.
Rapid Sampling Systems: These are automated delivery systems that minimize
delivery time of the sample to the sample introduction system, as well as
optimizing the washout time after the analytical measurement has been
carried out. They offer quite a significant increase in sample throughput over
traditional autosamplers.
15
RF Generator: The RF generator provides a radio frequency alternating
current to sustain the ionized argon plasma at a temperature of
approximately 8,000 °C. Two popular RF generator designs include the “tuned
oscillator” (or “free-running”) design, and the older “crystal-controlled” design.
The tuned oscillator is smaller and more tolerant of sample variations (organic
and aqueous). The crystal-controlled design is much larger and requires a
costly electro-mechanical impedance-matching device to control power
levels. The inherent design of the tuned oscillator demands much less strain
on the power tube, increasing its lifespan over those in a crystal-controlled
system. Two frequencies are commonly used for RF power, 40.68 and 27.12
MHz. While both provide similar performance, there is conclusive evidence
that 40.68 provides higher tolerance to variation in sample matrices.
Ruled Gratings: Ruled gratings are diffraction gratings where the grooves are
ruled or scribed into the surface of the mirror with a diamond cutting tool.
Sample Introduction System: The portion of the instrument that delivers the
liquid sample to the plasma torch as a fine-droplet aerosol. It is comprised
of a nebulizer to generate the aerosol, a spray chamber to only allow smaller
droplets to reach the plasma discharge (rejecting larger droplets) and a drain
system to take the redundant sample to waste. Refer to “Nebulizer” and “Spray
Chamber”.
16
Sample Injector: The central tube of the plasma torch that carries the sample
aerosol and the nebulizer gas. It can be a fixed part of the quartz torch or it
can be separate (demountable). Sample injectors can be made from a variety
of materials such as alumina, platinum, and sapphire, for the analysis of highly
corrosive materials.
Scan: A scan is a plot of light intensity vs. wavelength (in an ICP), usually
centered on a particular emission wavelength. Scans are most commonly
used during method development for locating background correction points
or checking for interferences. Scans can also be useful as a quantitative
analysis tool.
17
Self-Absorption: The absorption of intense analyte emission by atoms or ions
of that same analyte element within the plasma. Self-absorption occurs at
high analyte concentrations and is generally responsible for determining the
maximum linear concentration for strong analyte emission lines.
Slits: Slits are image-forming or masking devices. Light from an ICP source
passes through an entrance slit, through the optics, and through an exit slit to
a detector. The larger the slit, the more light is allowed through (resulting in
better sensitivity at the detector), but at a cost of diminished resolution. Refer
to “Resolution” and “Dispersion”.
18
Impact Bead Spray Chamber: A type of spray chamber more commonly
used in atomic absorption spectrometers. The aerosol from the nebulizer
is directed onto a spherical bead, where the impact breaks the sample into
large and small droplets. The large droplets fall out due to gravitational
force and the smaller droplets are directed by the nebulizer gas flow into
the sample injector of the plasma torch.
Scott Spray Chamber: A sealed spray chamber with an inner tube inside a
larger tube. The sample aerosol from the nebulizer is first directed into the
inner tube. The aerosol then travels the length of the inner tube, where the
larger droplets fall out by gravity into a drain tube and the smaller droplets
return between the inner and outer tube, where they eventually exit into
the sample injector of the plasma torch.
19
Torch: A torch is a device constructed of concentric tubes (usually quartz) that
contain the flows of argon (coolant or “outer”, auxiliary or “plasma”, and sample
carrier gases) to the ICP. The two common types of torch are “demountable”,
in which all tubes can be disassembled (common for hydrofluoric acid
applications), and “one-piece”, for all other applications. Refer to “Sample
Injector” and “Plasma Torch”.
Transient Signal: A transient signal is one that lasts a finite period of time.
This type of analysis, sometimes called Time Resolved Analysis (TRA), presents
unique challenges not seen in the more common continuous nebulization
sample introduction approach. Some examples of transient signal include
when an ICP-OES is used for the analysis of elemental speciation using
chromatography separation techniques, spatially resolved laser ablation
experiments such as depth profiling or surface rastering of solid samples and
Flow Injection Analysis (FIA) involving the discrete introduction of samples
into a continuous flow of a carrier solution.
UV Region: Generally refers to the spectral region of 200-350 nm. The low
UV region is normally 165-200 nm and the deep UV region is considered
any wavelength below 165 nm where many of the halogens have sensitive
emission lines.
Viewing Height: The vertical position in the plasma where the analyte
emission is being viewed by the spectrometer. With a radial ICP, it is quoted as
the number of mm above the top of the load coil.
Wavelength Range: This is the range of usable wavelengths for ICP analysis.
Some instruments have spectrometers limited to this range, and concentrate
primarily on shorter wavelengths. Alkali metals and/or alkaline earths, which
are most sensitive in the longer wavelength region, may not be accessible
with all ICP instruments. Auxiliary spectrometers, in addition to the main
optics, are sometimes employed to measure these longer wavelengths. Not all
instruments provide access to the full ICP spectrum.
20
4736 Socialville Foster Rd • Mason, OH 45040 USA
Tel: 800-533-6267/603-886-8400
21
www.teledyneleemanlabs.com
18061_06/2019