1. Magnet system:
□ Magnet Weight: 3200Kg with cryogens
□ Operating field strength: 1.5 Tesla.
□ Magnet length: 1570mm
□ Field orientation: Horizontal
□ Vessel capacity: 1650L
□ Liquid Helium boil-off rate : 0.03 L/hr
□ Refilling interval : 36 months
□ Dimension: 1880mm×1570mm×2400mm
(Width ×Depth ×Height)
□ Patient aperture: 600mm
□ Shim method: Passive + Active shimming
□ 5 Gauss Line: 3.8m(axial)×2.4m(radial)
□ Homogeneity:(V-RMS) (24planes/20points )
≤0.01 ppm on 10cm DSV;
≤0.04 ppm on 20cm DSV;
≤0.12 ppm on 30cm DSV;
≤0.35 ppm on 40cm DSV;
≤0.94 ppm on 45cm DSV;
2. Gradient system:
□ Noise reduction technology comfortable scanning experience, unnecessary uses of special ear
protection.
□ Powerful gradient system with a high performance in scanning speed and image quality. Minimum slice
thickness is 0.1mm in 2D and 0.05mm in 3D.
□ Eddy current compensation technology in hardware and software can free the eddy current to improve
image quality.
□ Gradient Strength (amplitude) is 35mT/m.
□ Slew Rate up to 136mT/m/sec.
□ Rise Time 257us.
□ Active shielding in XYZ directions.
□ Noise reduction technology
□ FOV from 5mm to 500mm.
Slice thickness: 0.1 mm (2D) / 0.05mm (3D),
□ Cooling method to support water cooling
3. Digital RF System
□ 16-ch phase array transmitting and receiving system support higher SNR, excellent image quality.
□ SIMEX (simultaneous excitation) technique
To support multi-slices excited simultaneously, increase scan slice number, reduce scan time
meanwhile save exam time, increase patient throughput.
□ Phase array multi-channel coils
To support image acquisition with higher signal to noise ratio and better quality thereby keeping
image acquisition time to a minimum. PA coils available are optimized for applications.
□ Sense technique-- Parallel acquisition technique
The advanced Sense technique can decrease scanning time, avoid motion artifact, increase scanning coverage,
increase image resolution.
Receiving Coil
□ Automatic laser locating system ensures the area of interest is close to the centre of the magnet, hence
improve the image quality.
□ Fast and convenient for patient positioning.
□ Patient Space Size is 600 mm in width and 157mm in length, offering less claustrophobia and max FOV
□ Communication with patient using two way intercom
5. Computer system:
□ Windows XP Operating System.
Neuro
Neurological applications consist of brain and central nerve
system down to the th-spine. There are several standard
methods to get anatomical/morphologic images and
methods for cervical, angio, and diffusions weighted
brain imaging.
NSM-S15P can provide advanced image technology for Neuro, such as SWI(Susceptibility weighted
imaging)、PWI(Perfusion-weighted imaging)、fMRI(functional MRI)、ROKAR(remove motion artifacts),
Providing a convenient way for hospitals to carry out clinical research in emerging fields
ROKAR
(ROtate K-space Acquisition
and Reconstruction)
is used to correct for phase ,
rotation and translation and
is used to correct for motion
artifacts ,mainly rigid
motion.
ROKAR is mainly applied
to the brain and neck scan.
T2 TSE with regular motion ROKAR with regular
motion
SWI is extremely sensitive
SWI for bleeding or blood
components deoxy and is
able to provide hemorrhage,
arteriovenous malformations,
exact information iron
deposition in order to achieve
faster and more accurate
diagnoses, very small lesions
can be diagnosed quickly.
SWI bleeding is 10 times
more sensitive than
traditional methods on T2 *
fMRI
Functional MRI
analysis could supply
the high resolution
functional
images which are very
imoportant in clinical
diagnosis and treatment
such as The
preoperative evaluation
of brain surgery.
IRT2TSE T2*FFE
IR-T2TSE T2WI:TSE
T2WI:TSE
Body
THRIVE
Coil selection:
(T1w High Resolution Isotropic Volume Examination)
□16-ch flex body coil THRIVE is a variety of K-space scanning acceleration
technology, in the new 16-channel flexible body coil
and SENSE technology support, it uses a unique
CENTRA stands for Contrast Enhanced Timing Robust SENSE, CENTRA, Keyhole to accelerate the
Angiography. In this method, a true two-dimensional integration of technology, to achieve a high spatial
phase profile order is applied in the kykz-space. resolution, high temporal resolution and high contrast
It reduces the effect of timing errors and motion resolution of liver multiphase dynamic enhanced
artifacts. scanning.
Body Imaging------MRCP/MRM
Musculoskeletal (MSK)
MR mammography (MRM)
Coils selection
□ breast coil
Breast imaging ----- breast cancer (left side) and multiple adenoma (right side).
Cardiac imaging: