You are on page 1of 2

SWITCH-OVER

SSB-800-06

AUTOMATIC
Page 1 of 2
12/31/2007

Automatic Switchover Medical Liquid Manifold with Automatic Reset


(LifeLine)
SPECIFICATION

Automatic Switchover Medical Liquid Manifold latches for component accessibility. The enclosure may
The BeaconMedaes Lifeline automatic switchover be secured from unauthorized access by locking the
manifold accommodates multiple cryogenic containers draw latches (locks provided by others). The manifold
equally divided into two banks while providing an includes gas specific pigtail-to-header high-flow check
uninterrupted, reliable supply of gas. The manifold valves to permit changing of cryogenic container(s)
system referred to as “liquid x liquid” or “LQ x LQ”, without gas leakage. Thermoplastic hose assemblies,
consists of a manifold control panel with its wall designed for cryogenic service, are provided for each
mounting bracket and a high-pressure reserve header container gas connection. A separate power supply is
assembly (separately ordered, see SSB-800-08). The furnished and converts 120 VAC to 24 VAC output
manifold is cleaned, tested, and prepared for the power. The power supply provides dry normally closed
indicated gas service and constructed in accordance contacts for (3) alarm conditions: secondary supply,
with requirements of the latest edition of NFPA 99 and reserve in use, and reserve low. The power supply is
CGA. housed in a NEMA 3R enclosure with electrical
requirements of (0.125) amp at 120 VAC, 60Hz, single
Manifold Design phase. External LED visual indicators for “RESERVE
A bank regulator (one for each cryogenic container IN USE” (red) and “RESERVE LOW” (red) are
bank) is used to initially reduce the container pressure included within the power supply assembly. The
to the two line regulators which control the final line manifold is supplied with ¾” FNPT (manifold outlet)
pressure. Both line pressure regulators are in service at and ½” FNPT (high-preserve reserve header inlet) o-
all times to maximize flow rates. The manifold ring sealed “zero clearance” unions. A ¾” full port,
automatically changes from the depleted primary supply three piece, ball-type source shut-off valve with a 1/8”
bank to the secondary supply bank without fluctuation FNPT port is also included. Source valve has a ¾” NPT
in line pressure utilizing dome-bias loading and attachment to the outlet union and a ¾” nominal copper
unloading of the bank regulators. Economizer gas (type k) tube for brazing to main supply line.
circuits are provided to conserve evaporation of the
cryogenic liquid from the secondary supply bank, thus NOTE:
preventing the exhausting of useful gas into the The flow capacity of a nitrous oxide and carbon
atmosphere. This mechanism discharges the conserved dioxide manifold depends upon environmental
gas upstream of the final two line regulators. After conditions at the installation site and the number of
replacement of the depleted cryogenic container(s), the containers in service. Installing a nitrous oxide or a
manifold automatically indicates the container bank carbon dioxide manifold in a location that exposes it
recently replaced as the new secondary supply. Manual to an ambient temperature below 32O F (0O C) is not
resetting of the control panel is not necessary. If the recommended.
primary and secondary supply banks of container(s) are
both depleted, the manifold system will automatically Environmental Considerations
switch to the external high-pressure reserve header Manifolds are to be installed in accordance with
assembly. A pressure switch is provided to signal the requirements stated by NFPA 99, CGA, and all
master gas alarm system just before change over from applicable local codes. Manifold components are
the secondary supply to the high-pressure reserve. The designed to work best over a temperature range of 32O F
manifold includes a line pressure gauge, two bank through 130O F. Wider temperature variations may
contents gauges (left-bank and right-bank), and LED cause manifold malfunctions to occur. BeaconMedaes
visual indicators for “IN USE” (green), “READY” recommends the control cabinet be located with its
(yellow), and “EMPTY” (red) for each bank. The power supply at an installation site protected from
manifold has bank regulator, intermediate and line moisture, continuous exposure to direct sun rays, ice
pressure relief valves internally connected to a common and snow.
vent port, terminating into a ½” FNPT o-ring sealed
“zero clearance” union. Master shutoff valves (one for
each bank) are located within the manifold cabinet and
are fabricated with metallic seating surfaces. The
cabinet enclosure is easily removable by releasing draw

BeaconMedæs Š 1800 Overview Drive Rock Hill, SC 29730 Š Phone: (803) 817-5600 Fax: (803) 817-5750
SSB-800-06
Page 2 of 2
12/31/2007

Flow Characteristics Gas Type Flow per Container


Manifold system flow is limited by maximum flow capacity of Oxygen 5.8 SCFM (350 SCFH)
liquid container(s). Approximate maximum continuous flows Nitrous Oxide 1.8 SCFM (110 SCFH)
of one liquid container are shown in the following chart. Flow Carbon Dioxide 2.5 SCFM (150 SCFH)
capacity is increased with the addition of liquid container(s). Nitrogen 5.8 SCFM (350 SCFH)
An external vaporizer (sourced by others) is necessary for high Argon 5.8 SCFM (350 SCFH)
flow requirements.

ORDERING INFORMATION
The Lifeline automatic switchover medical liquid manifold may be configured in 1 container (1 x 1) or 2 container (2 x 2)
arrangements.
Lifeline Automatic Medical Liquid Manifold
Note:
Gas Service Service Pressure Part Number 1 1
Enter the number of containers required for each bank, 01 or 02.
55 psig 6-108010-XX Example: Oxygen manifold for 55 psig service with
Oxygen
100 psig 6-108110-XX 2 x 2 arrangement (6-108010-02).
Nitrous Oxide 55 psig 6-108011-XX
Nitrogen 180 psig 6-108014-XX
Carbon Dioxide 55 psig 6-108020-XX
Argon 55 psig 6-108026-XX
Accessories:
6-290840-48 Liquid Vent Cylinder Hose Assembly, Oxygen
6-290841-48 Liquid Vent Cylinder Hose Assembly, N2O, N2, CO2, & AR
6-135000-KT Control Panel Stand Kit (for floor mounting)
6-515725-00 90 Degree Fitting (turn source valve from vertical to horizontal orientation)

BeaconMedæs Š 1800 Overview Drive Rock Hill, SC 29730 Š Phone: (803) 817-5600 Fax: (803) 817-5750

You might also like