You are on page 1of 14

WARNINGS AND CAUTIONS

TO AVOID POTENTIAL SERIOUS INJURY TO THE USER


AND THE PATIENT AND/OR DAMAGE TO THIS DEVICE, Please read this manual and follow its instructions
THE USER MUST: carefully. The words WARNING, CAUTION, and NOTE
carry special meanings and should be carefully re
1. Read this operating manual thoroughly and be familiar viewed:
with its contents prior to using this equipment.
WARNING The personal safety of the patient may be
2. Carefully unpack the unit and check if any damage involved. Disregarding this information
occurred during shipment. If damage is detected, could result in injury to the patient.
please refer to the Service and Claims section in this
manual. CAUTION Special service procedures or precautions
must be followed to avoid damaging the
3. Be a qualified physician, having complete knowledge of instrument.
the use of this equipment.
NOTE Special information to make maintenance
4. Test this equipment prior to a surgical procedure. This easier or important information more clear.
medical video camera was fully tested at the factory
before shipment. An exclamation mark within a triangle is
intended to alert the user to the presence
5. Avoid removing covers on control unit to avoid electric
shock and breaking of camera head seals. A of important operating and maintenance
instructions in the literature accompanying
the product.
6. Attempt no internal repairs or adjustments not
specifically detailed in this operating manual. A lightning bolt within a triangle is intended
to warn of the presence of hazardous
7. Pay close attention to the care, cleaning, sterilization,
and disinfection instructions in this manual. A deviation
A voltage. Refer all service to authorized
personnel.
may cause damage.

8. DO NOT STERILIZE CAMERA CONTROL CONSOLE.

9. Disconnect the control unit from the electrical outlet


when inspecting fuses. NOTE Stryker Endoscopy reserves the right to make
improvements in the product(s) described
10. Read the entire instruction manual before assembling or herein. Product(s) therefore may not agree in
connecting the camera. detail to the published design or specifications.
All specifications are subject to change without
The camera warranty is void if any of these warnings are notice. Please contact the local Stryker
disregarded. Endoscopy Distributor listed in the Other
Service section or phone your local Stryker
Stryker Endoscopy accepts full responsibility for the effects Endoscopy sales representative or agent for
on safety, reliability, and performance of the equipment only information on changes and new products.
if:

• Re-adjustments, modifications, and/or repairs are Ambient temperature range


carried out exclusively by Stryker Endoscopy.

• The electrical installation of the relevant operating room


complies with the applicable IEC, CEC, and NEC Relative humidity range
requirements.

WARNING Federal law (United States of America)


restricts this device to use by or on order of a —*•'•— Atmospheric pressure range
physician.

Denotes compliance to IEC 601-1 with


amendments 1 & 2, CSA Std. C22.2 No. 601.1-
ui,-^, M90, and UL Std. No. 2601-1.

Type BF equipment as labeled on the product

QL

Type CF equipment as labeled on the product


PRODUCT DESCRIPTION & USE

Description of Functions continued

© © © © © © 14 ©

striker'
ENOdSCXMt/ so&tui
\
HGS wc TALOUWUTS

f<* mpg r-r3 sjfi &s

ANAlOf. OUTPUTS HfRUCS

9mm
fc»

(20) fill f?.?1

B REAR PANEL

8. LIGHT/ZOOM- Selects either LIGHT or ZOOM function for the arrow buttons on camera head.

9. Button Function - Enables/Disables functions on camera head right ('W') button.

10. Flex Filter - Enables/Disables Flexible Scope Filter.

11. AG Filter - Enables/Disables Anti-Glistening Filter.

12. Shutter - Enables/Disables Auto Shutter.

13. SDI Out - Serial Digital Interface (SDI) Video Output.

14. DVI Out 1 - Digital Visual Interface (DVI) Video Output.

15. DVI Out 2 - Digital Visual Interface (DVI) Video Output.

16. AC Power Input/Fuse - Input for AC Power and Fuse

17. Remote Out 1 - Output to Video Accessory Remote Switch.

18. Remote Out 2 - Output to Video Accessory Remote Switch.

19. RGB Out - RGB/Component Video Output.

20. S-VIDEO Out 1 - S-VIDEO Component Video Output.

21. S-VIDEO Out 2 - S-VIDEO Component Video Output.

22. Video Out - NTSC Composite Video Output.

23. Hermes - Hermes Operating Room Control Center Interface Connection.

24. Potential Equalization Connection -


PRODUCT DESCRIPTION & USE

ATTACHING AND USING TURNING ON THE CAMERA


THE COUPLER
Connect the power cord supplied with the camera to the AC
inlet on the rear panel (16) and ensure that it is fully inserted.
Connect the other end to a hospital grade AC receptacle.

WARNING Always use a hospital grade power cord with


this camera.

Press the power switch (1).

WARNING Test the video camera before each procedure


begins. Ensure that a video image appears on
all video monitors before beginning each
procedure.

Figure 1

If a coupler is not already attached to the camera head when CONNECTING VIDEO OUTPUT
received, screw the coupler into the front of the camera head.
Turn the coupler clockwise until it stops.

To attach an endoscope to the coupler, push down on the


eyepiece clamp (as shown above) and remove the red dust
cap if present. Push the eyepiece into the coupler and
release the eyepiece clamp. (See Figure 1).

CAUTION Always treat the camera head with care.


Dropping the camera head could cause
-"LJLZ————-"1 !• ••"
inioocn/niw5 • :. i I
damage.

C-MOUNT CAMERA SYSTEM


There are two camera heads: BF and CF applied part. While
both the BF and CF camera head may be used with the 988
control unit, The CF model (Indentified by the heart logo) is
Figure 3

specially designed for procedures with direct cardiac Connect the S-VIDEO or BNC cable provided with the camera
applications from VIDEO OUT to the VIDEO IN on a video monitor (Figure
3) and turn on the monitor.
WARNING: When using more than one applied part
simultaneously, patient leakage current NOTE When the camera head is not plugged into the
may be additive. When using the 988 type CAMERA connector (7), a color bar pattern will
CF camera head for cardiac applications, appear on the video monitor.
all other applied parts should be type CF
rated to minimize total patient leakage
current. WARNING Do not touch the internal pin of the SDI and
VIDEO OUT BNC jacks (13,22) and the patient
simultaneously.

ELECTROMAGNETIC INTERFERENCE

The camera has been designed and tested to comply with EN


60601-1-2: 1993 requirements for electromagnetic compatibil
ity with other devices. If interference is experienced, refer to
the Troubleshooting section in this manual.

Figure 2
A direct-coupled endoscope may be attached to a C-mount
camera head by threading the endoscope directly into the
camera head. A coupler is not necessary in this case. (See
figure 2).

CAUTION: Do not over tighten endoscope, as this may


damage the front window of the camera.
PRODUCT DESCRIPTION & USE

CONNECTING THE CAMERA HEAD GAIN

The GAIN switch (3) increases the brightness of the video


picture. There are four levels of gain: OFF, LOW, MEDIUM,
and HIGH. Selecting a particular level establishes the GAIN
range. OFF is the setting you should normally use.

If the picture is too dark, first attempt to raise the light level as
described on page 6. If the picture is still too dark, adjust the
light source. As a last resort, GAIN can be used to illuminate
an area that does not have enough light. The picture will get
brighter but may also be more grainy. Text displayed on the
screen indicates current gain level.

AUTO FOCUS

Figure 4
The Auto Focus button (4) turn the Auto Focus feature on or
off. This feature maintains picture clarity throughout a larger
Unscrew (counter-clockwise) the soaking cap from the depth of field than the coupler manual focusing ring alone.
camera cable. Holding the connector by the knobbed portion
only, plug the connector into the control unit by aligning the Press the Auto Focus button to activate/deactivate the
red dot on the cable connector with the red dot on the control feature. Text displayed on the screen indicates the current
unit CAMERA connector (7). Push in until it clicks (Figure 4). state of the feature. After activating the feature, establish a
A picture should now be visible on the video monitor. good baseline focus using the coupler manual focusing ring,
and then press a short pulse on the camera head 'W' button
NOTE To unplug the camera from the control unit, to set the baseline.
grasp the knobbed portion of the connector
only and pull straight out. This releases the NOTE When the Auto Focus feature is activated, the
connector latching mechanism. enhance feature will be inoperative.

D CAUTION To avoid camera head cable damage do not


severely bend, crush or cut the camera head
cable.
ENHANCE (VI BUTTONS)
The Enhance button (5) feature increase or decrease the
WHITE BALANCE picture sharpness. There are 8 levels of enhancement. Text
displayed on the screen indicates current enhancement level.
The WHITE BALANCE switch (2) is used to correct slight
Level 1 No Enhancement
color differences caused by using different light sources and
Level 8 Maximum Enhancement
endoscopes. With the light source and scope attached, point
the scope at several stacked 4x4 white gauze pads, a white
laparoscopic sponge, or any clean white surface. Look at the
monitor and make sure that no glare is visible off of the white
surface.

Press and hold the WHITE BALANCE button until "WB"


begins flashing on the video monitor. Continue pointing the
scope at a clean, white surface until the video monitor
indicates that white balance is "OK." The video picture will
change color. If the video monitor indicates that white
balance is "NOT GOOD," repeat the process. If you cannot
achieve a White Balance that is "OK", refer to the Trouble
shooting section in this manual.
PRODUCT DESCRIPTION & USE

SURGICAL SPECIALTY LIGHT/ZOOM

The Surgical Specialty selector switch (6) is used to select The camera control unit has the ability to change the function
one of four surgical type settings: of the arrow buttons on the camera head. To activate the
zoom feature from the arrow buttons, place the LIGHT/ZOOM
Arthroscopy - Shutter programming optimized for small switch (8) in the ZOOM position. To activate the light level
scopes, bright conditions, high contrast and low color. feature from the arrow buttons, place this switch in the LIGHT
position.
Cystology - Shutter programming, color and lighting
optimized for urology procedures.
BUTTON FUNCTION
Laparoscopy- Shutter programming optimized for large
scopes and high color differentiation. Place the BUTTON FUNCTION switch (9) ON to activate the
short pulse function (either zoom or gain) of the right ('W')
Thoracoscopy - Shutter programming, color and lighting button on the camera head. Place this switch OFF to
optimized for Thoracic procedures. deactivate the short pulse functions of the right camera head
button.
Manufacturer presets will optimize the video performance for
the surgery type selected. A green light will highlight the
current selection (see front panel). FLEX FILTER

The FLEX FILTER can be used to reduce the grid pattern


caused by a flexible scope. The Flex Filter status can be
REAR PANEL CONNECTIONS
toggled by changing the position of the Flex Filter switch (10)
or by pressing both enhancement buttons (5) for at least one
There are four analog and three digital video outputs on second. Graphics will indicate if the Flex Filter is ON. Ensure
the rear panel. Analog outputs include one RGB (19) the Flex Filter is OFF when using rigid scopes.
output, one NTSC Composite output, and two S-VHS
outputs (20,21). Digital outputs include one SDI output
(13) and two DVI outputs (14,15). The DVI outputs can AG FILTER
also display analog XGA outputs through a DVI-I to VGA
adapter. These are separate outputs and can be used Position the AG FILTER switch (11) ON to reduce the
together or independently. glistening or bright points of light caused by moisture

Additionally, there are two Remote outputs (17,18)


which can be used to control documentation equipment.
For more information about connecting the camera to
reflection. Position this switch OFF to deactivate the anti-
glistening (AG) filter.

SHUTTER SWITCH
B
video peripherals, refer to the INTERCONNECT section
in this manual.
Position the SHUTTER switch (12) ON/OFF to activate/
There is also one interface connection (23) for the deactivate the automatic shutter.
Hermes Operating Room Control Center. For more
information about using the Hermes System refer to the
ON: The camera automatically adjusts the brightness
Hermes Operating Room Control Center Operating
of the video picture in response to varying light
Manual.
levels without using an automatic light source.

REMOTE ACCESSORY CONTROL OFF: The camera does not adjust to varying light
levels. You must use an automatic light
The camera supports two remote outputs (17,18) for source to adjust picture brightness.
controlling a Stryker SDC Pro, VCR, or other video
accessory. The camera head button can activate one or NOTE You may use an automatic light source and
two peripherals. the camera auto shutter at the same time,
but the camera auto shutter should normally
In order to use the remote outputs to interface with be used alone.
different peripherals, one six-foot cable with 2.5mm
diameter ends and one six foot cable with a 2.5mm and
a 3.5mm diameter end have been provided.

Please refer to the Interconnect section in this manual.


PRODUCT DESCRIPTION & USE

CAMERA HEAD BUTTONS 'T' AND 'I' BUTTONS


The 'f' and '4.' buttons work together to increase or decrease
the light level or zoom settings.

ZOOM: With the LIGHT/ZOOM switch in the ZOOM position,


the arrow buttons will raise or lower the magnification setting
between 1.0 and 1.5 in 44 steps. Press the button once for
individual steps or hold down the button for a quicker,
smoother transition.

LIGHT LEVEL: With the LIGHT/ZOOM switch in the LIGHT


position, the arrow buttons will raise or lower the automatic
shutter light level setting in 64 steps. Press the button once
for individual steps or hold down the button for a quicker,
smoother transition.

Figure 5

As shown in Figure 5, the camera head contains a diamond-


shaped, four-button pad, labeled 'T' to the north, 'P' to the
west, 'W' to the east, and 'I' to the south. These buttons
provide for remote control of numerous functions. Appropri
ate audible feedback will accompany the "P" and "W" buttons.

'P' BUTTON

The 'P' button controls up to two remote outputs. When


depressed less than one second, the button will activate

B Remote 1.

When depressed longer than one second, the button will


activate Remote 2.

¥ BUTTON

The 'W' button activates several different functions, which are


selectable through external switches on the front and rear
panel.

(1) Depressing this button for more than one second will
activate the white balance function.
(2) If AUTO FOCUS is ON, depressing this button for less
than one second will establish a baseline for the
automatic focus.
(3) If AUTO FOCUS is OFF:
(a) LIGHT/ZOOM switch to ZOOM - depressing this
button for less than one second will raise the gain
setting in four increments: OdB, 5dB, 10dB and
15dB. When the gain setting reaches 15dB, the
next short push will cycle the setting back to OdB.
(b) LIGHT/ZOOM switchto LIGHT - depressing this
button for less than one second will expand the
magnification setting in four increments from 1.0
to 1.5. When the magnification setting reaches
1.5, the next short push will cycle the setting
back to 1.0.
(4) The above short pulse functions may be
disabled by placing the BUTTON FUNCTION
switch in the OFF position.
STERILIZATION AND DISINFECTION

Sterility of the camera head may be achieved by use of sterile DISINFECTION WITH
drapes, ethylene oxide gas, the Steris® system and
Sterrad™. GLUTARALDEHYDE SOLUTION

WARNING Camera head, coupler and camera cable must


be cleaned and sterilized prior to first and NOTE Never soak the camera head in the same tray
every subsequent use. with scalpels or other sharp instruments. Avoid
soaking dissimilar metals in close proximity in
CAUTION The following procedures do not guarantee order to minimize galvanic corrosion. Camera
sterilization or disinfection. Sterility or malfunction caused by damaged cables is not
disinfection can only be achieved if properly covered by your warranty.
trained personnel follow these recommended
procedures and hospital practices. CAUTION Before soaking, inspect the camera head cable
for breaks or cuts. A break or cut in the cable
NOTE Soaking the coupler (adapter) separate from will allow the solution to seep into the camera
the camera head is not recommended. head and cause damage. Please return any
Moisture may get trapped between the camera with a damaged cable to Stryker
coupler and camera head and cause fogging. Endoscopy Repair Department.

1. Clean and prepare as recommended in the Cleaning


CLEANING section. Ensure the soaking cap is installed.

2. Camera head, adapter, and camera head cable should


1. Immediately after use, unplug the camera head
be immersed in Cidex Plus* glutaraldehyde solution or
connector and protect it with the soaking cap. Place the
Cidex OPA* solution for disinfecting. Specifications for
cap over the connector and turn clockwise until secure.
each is as follows:

CAUTION Install the soaking cap before rinsing camera


Cidex Plus® disinfection:
to prevent corrosion of the connector pins.
soak time: minimum of ten minutes as
Ensure cap is properly tightened.
recommended by the sterilant
manufacturer.
2. Rinse the camera head and camera head cable in
temperature: 25° C(77°F)
lukewarm tap water. Use mild enzymatic detergent and
a soft brush to remove resistant debris. Ensure

B
Cidex OPA® disinfection:
removal of bioburden from all surfaces.
soak time: minimum of 12 minutes as
recommended by the sterilant
3. Thoroughly rinse off all soap residue with water.
manufacturer.
temperature: 20°C (68°F)
4. Apply alcohol to glass surfaces with soft cotton
applicator to aid in cleaning and drying without leaving
3. After soaking, the camera head, coupler, and cable
spots or streaks.
must be rinsed with sterile water and thoroughly
agitated to remove all traces of glutaraldehyde solution.
5. Dry thoroughly with soft towel or gauze surgical
Before use, dry thoroughly with a sterile towel. Ensure
sponge.
the soaking cap and connector are completely dry
before removing the cap.
6. Before disinfection or sterilization, coil the camera head
cable into a loop about ten inches in diameter, avoiding
kinks and twists in the cable. During disinfection or
sterilization, ensure that the camera head and cable are
not placed on top of other pieces of equipment.

If the camera control unit needs to be cleaned, wipe it down


with a damp cloth or sponge and towel dry.
STERILIZATION AND DISINFECTION

STERILIZATION: USE OF STERILE DRAPES

100% ETHYLENE OXIDE GAS: Use of sterile drapes will ensure maximum longevity of your
Stryker Endoscopy Model 988 Medical Video Camera. Follow
1. Clean and prepare as recommended on page 7. instructions provided by drape manufacturer when using
Ensure the soaking cap is installed. sterile drapes. Sterile camera drapes for use with the Model
988 Medical Video Camera can be ordered under Stryker
2. Sterility has been validated using an ethylene oxide product number 240-010-200.
sterilization unit for a cycle with the following
parameters: CAUTION Maximum performance and camera life is
extended by the consistent use of a single
temperature: 55°C sterilization method. Also, avoid leaving the
humidity: 70% RH±5% camera in the sterilization solutions longer than
vacuum: 21 in Hg necessary as this will accelerate normal
preconditioning time: 60 minutes product aging.
sterilant mixture: 100% ethylene oxide
sterilant concentration: 600±5 mg/L ethylene
oxide
exposure time: 120 minutes
aeration: 12 hours, 55°C DISPOSITION OF THE PRODUCT

STERIS® SYSTEM: The device must be disposed of according to local laws and
hospital practices. The device does not contain any hazard
1. Clean and prepare as recommended on page 7. ous materials.
Ensure the soaking cap is installed.

2. Sterility has been validated using Steris® System 1 with


Steris® Sterilant 20.

STERRAD™ SYSTEM:

1. Clean and prepare as recommended on page 7.

B
Ensure the soaking cap is installed.

2. Sterility has been validated using Sterrad™ Sterilization


System.

NOTE To increase the longevity of the camera head,


observe the following process limits:

maximum ethylene oxide exposure time: 3 hours


maximum temperature: 60°C (140°F)
maximum Cidex® Glutaraldahyde soak time: 24 hours

NOTE If coupler and camera head are sterilized or


disinfected as a single unit, disconnecting the
coupler will compromise the sterility or
disinfection.
INTERCONNECT SECTION

NOTES:
Four types of cables are used in wiring the 988 Video Output: • If you are using a non-Stryker light source with an
• NTSC cables which have a BNC connector unterminated input, you must connect a NTSC cable
• S-VHS cables which have a 4 pin Mini-Din connector from the VIDEO OUT of the light source to the VIDEO
IN on the monitor.
• VGA cables which have a 15-pin mini D-SUB connector

• DVI cables which have a 24-pin connector • An additional monitor may be connected using any open
camera output.
BNC connectors are push and turn type connectors while S-
VHS cables are push only type connectors. On some • You may also connect an S-VHS or BNC cable between
monitors, S-VHS inputs may be labeled Y/C, S-VHS and
the camera and monitor and select the appropriate
describe the same input. Described in this section are three
video source.
typical set-ups. Match your equipment and wire them
according to the appropriate diagram.

Camera
Camera

::::::::::::

Monitor

Lightsource

Monitor

SYSTEM 1 SYSTEM 2

Equipment Used: Camera, Light Source and Monitor Equipment Used: Camera, Light Source, Video Peripheral
and Monitor. The video peripheral may be a video printer,
digital capture, Hermes or VCR.
INTERCONNECT SECTION

DVI-I/VGA
Camera ADAPTER

A—

=fe o„o DVI FLAT


O B H PANEL
A *
DVI

CD
m z
3 o

3m

7)
lllllllllll m
lllllllllll 3
lllllllllll
lllllllllll -Q 3
m
BNC
VCR

BNC © © © ©

B S-VIDEO
© © © © o •

BNC Printer

'••••••••••••••a ••••£•••
SSSSBSS3 SBIBESSS&J
8SSS3SB
ISBSBSESBSBSSSBB BSSIpBBS
XSS, 'V T
71
::es::::::::kbbebbb
SsSSSSSSSEsssSSSSSs
C€ G- O A

IN
SDI OUT
A-..

Lightsource
SDI Monitor
EH ft •••••••••••••
BBS H•HHMHMKHHMMMfiB

SYSTEM 3

Equipment Used: Camera, Light Source, Stryker Printer,


Stryker VCR, SDI Monitor, DVI Flat Panel.
TROUBLESHOOTING

PROBLEM SOLUTION

NO COLOR BAR Ensure video out from camera is connected to the video monitor and all video
systems are powered on

White balance against clean white surface, see WHITE BALANCE section
PICTURE COLOR INCORRECT Check color settings on monitor

Adjust light level from camera head


Adjust light source for higher output
PICTURE TOO DARK
Check fiber optic light cable for excessive broken fibers
Press GAIN switch to increase illumination
Set auto SHUTTER to OFF

Adjust light level from camera head


PICTURE TOO BRIGHT Adjust light source for lower output
Select auto SHUTTER mode
Ensure GAIN is OFF

Ensure GAIN is OFF


NOISE OR SNOW ON PICTURE Reduce Enhancement
(no electro-cautery) Select auto SHUTTER mode to OFF
Check and replace faulty video cables

Plug electro-cautery into separate electrical outlet and separate power cords
NOISE OR SNOW ON PICTURE
Separate camera cable and electro-cautery cable
WHEN USING ELECTRO-CAUTERY
Reposition electro-cautery grounding pad on patient

NO VIDEO PICTURE WHEN


Check to ensure that all components in the video systems are plugged in and
CAMERA HEAD IS PLUGGED IN power is on
Check connector on camera head cable for broken pins

IMAGE NOT CENTERED Release scope from the coupler and re-connect. Make sure the scope is seated
correctly in the coupler. B
WHITE BALANCE the camera
VARIABILITY IN COLOR Check settings on video peripherals
REPRODUCTION WITH VARIOUS Ensure light source has proper infrared filter (check with manufacturer
LIGHT SOURCES OR PERIPHERALS specifications)

Check camera focus and refocus


FOGGY PICTURE
Optics are dirty. Clean and dry both the scope and the coupler glass face.
(loss of definition & clarity)
Moisture may exist between the camera head and the coupler. Remove
coupler from the camera head and clean this area thoroughly.

Rotate the scope. If dust particles in the picture rotate, dust is located on the
CLEANING DIRTY OPTICS scope tself. Follow manufacturer's instructions for cleaning the eyepiece and
negative lens.
If particles do not move when rotating, particles are behind the scope. Remove
the scope and clean the window on the front of the coupler with a dry or alcohol
tipped cotton swab.
If dust particles lie between the coupler and camera, remove the coupler and
clean the window on the coupler and the window on the camera head.
BE SURE entire area is completely dry before reassembling or fogging may result.
BLURRY PICTURE
Check that coupler is in focus.
Check that the Flex Filter switch is turned off.

NO WHITE BALANCE (WB)


• Picture too dark (see above for troubleshooting)
• Picture too bright (see above for troubleshooting)
• Color temperature too low - WB with light source connected and use
metal halide or xenon lighting (no fluorescent lighting).
NOTE: If this troubleshooting guide does not solve the problems that you are having with your camera, please refer to the
Service and Claims section in this manual for instructions on obtaining repair service from Stryker Endoscopy.
SPECIFICATIONS

Imaging System 1/3" Interline Transfer, EX view HAD CCD


768 (H) x 494 (V) pixels

Scanning System 2:1 Interlaced


Horizontal: 15.734 kHz
Vertical: 59.94 Hz

Video Outputs
Composite: One NTSC standard Digital: Two Digital Visual Interface (DVI)
1.0V P-P into 75 Ohms 1024X768 XGA resolution
Connector: BNC coaxial Connector: 29-pin DVI-I
Component: Two S-VHS Digital: One Serial Digital Interface (SDI)
Y-1.0VP-P C-0.29VP-P SMPTE-259M
Connector: 4-Pin S-VHS Connector: BNC Coaxial

Resolution
Horizontal: 950 lines
Vertical: 500 lines

Signal to Noise Ratio 72 dB

Mounting Endoscope Eyepiece used with coupler


C-mount camera head used with C-mount scopes
(C-mount 1"-32UN-2A)

Minimum Illumination < 1.0 Lux

Auto Shutter Range 1/60 to 1/50,000 second

Gain
OFF OdB
LOW 5dB

B MEDIUM
HIGH

Operating Temperature
10 dB
15 dB

5° to 40°

Transport & Storage


Temperature 5° to 40° C
Relative Humidity 10% to 85%
Atmospheric Pressure 700hPato1060hPa

Power Consumption approx. 30 W

Input Voltage Range 100/120 VAC 50/60 Hz

Total Shipping Weight 13 lbs (5.9 kg)

Dimensions
Camera Control Unit: 9.8"w x 4.0"h x 15"d (24.8cm w X 10.1cm h X 38.1cm d)
Camera Head Cable
to Camera Control Unit: 10 foot (3m) sealed cable or 21.3 foot (6.5m) sealed cable (sensed by CCU)

Enhancement Switchable to 8 steps

Classification & Approvals Complies with IEC 601-1 with Amendments 1 & 2, CSA Std. c22.2 No. 601.1 - M90, and UL
Std. No. 2601-1
Complies with EN 60601-1-2: 1993 (EMC compliance of medical devices)
Type BF applied part or Type CF applied part as labeled on the product
Class I equipment
Ordinary equipment
Continuous operation

Patents: This product is covered under U.S. Patent #6,224,542. Other patents pending.
MAINTENANCE

FUSE REPLACEMENT PERIODIC MAINTENANCE SCHEDULE


INSTRUCTIONS
WARNING To ensure safe operation of the Model 988
Medical Video Camera you should perform the
1. Unplug the camera control unit from the wall socket and following procedure periodically:
remove the power cord from the camera control unit.
Activity: Check earth leakage current to <500//A
2. Unlatch the fuse holder above the AC Inlet (16) and (<300//A in U.S.A.), ground impedance to <0.1
remove. You may need to press the tab on the fuse ohms, and power consumption less than or
holder with a slender screwdriver to release the latch. equal to rated power.

3. Replace the fuse with the same value and rating. When: Every 12 months.

WARNING Toavoid the risk of fire, replace only with a Tools and Equipment Required:
fuse of the value specified on the fuse label True RMS digital multimeter and safety analyzer.
(located on the rear panel of the camera
control unit). Refer calibration and operating difficulties not detailed in this
manual to your Stryker Endoscopy sales representative, or
4. Reinstall the fuse holder until the tab snaps in place. Stryker Endoscopy Customer Service (in the United States: 1-
800-624-4422), or the nearest Stryker Endoscopy subsidiary
(refer to the Other Service section in this manual).

B
SERVICE & CLAIMS

SERVICE NOTE Never open the camera head or its cable.


These units have been factory sealed to
prevent moisture from entering the electronic
If service is needed either during or after the warranty period:
components. If the camera head or the cable
seal is intentionally broken, the equipment
1. Contact Stryker Endoscopy at 1-800-624-4422 or
warranty will be void.
phone your local Stryker Endoscopy sales
representative.
NOTE The medical video camera described in this
manual is constantly being reviewed and
2. Package all the components carefully in the
improvements may be made without notice.
original shipping container if possible.

3. Ship the camera, pre-paid and insured to:


Stryker and Stryker Endoscopy are registered trademarks of Stryker Corporation.
Stryker Endoscopy Customer Service
Attention: Repair Department
5900 Optical Court
San Jose, CA 95138

fl

You might also like