MCDEL STER.

SERIAL NO. CIMEER SIZE ELECTRICITY


PRESS./TMP
TM-329 L.2Ks/cp121° 35t%75D c N 220 z
2700W
OPERATION INSTRUCTIONS
1.Close DRAIN. KNOB and STEAM EAIAJST KNOB. Fil1 2400cc DISTILLED or DEAINERALI ZED
WATER 1nto CHAMBER
2. Put SHELP first, load sterilized objects, then close PRESS. LID fimly.
3. STERILIZATION:
(A) Turn on STER. SWITCH. Set TEMP REGULATOR to desired sterilizig temperature
12CVihen setting every first sterilizing temperature, it is necessary to
(B)
reset
set 2MP, RECULATOR after temperature is reached for accurate sterilization.
TIMER tsdesired sterilizing time, TIMER starts at pre-set tenperature.
STER. LAMP wil1 blink on and off indicating HEATER'S
operation.
C ntend
KNOR
to
first to
open PRESS. LID during sterilization,please open STEA1 EXTIAUST
expel steam until PRESS. GAUGE at "O" Kg. for safety.
(D)
In Case of no water during sterilization, HEATER will stop, ALAR will sound
and LOW NTER LAM wi1l light up autonatically. please follow step C) to
Open PRESS. LID and add water, then repeat sterilizing process fron step 1.
(E) When sterilizatibn is completed, TIMER automatically activates ALAR, shuts
off HEATER and sTER. LAMP.
4. DRYNESS:
(A) Open DRAIN. KNOB, or slcwly open STEAM EXHAUST KVOB first, then open DRAIN.
KVOB to exhaust water/stean. Open PRESS. LID a little.
(B) Turn on DRY, SWITCH. Set TIMER on 15 or 20 minutes, DRY. LAP will blink on
and off indicating HEATER'S operation.
(C)When dryness is completed, TIMER automatically activates ALARM, shuts off
HEATER and DRY. LAMP
a
This speedy Autoclave is pressure vessel that is fabricated of stainless steel.
to use distilled or demineralized water in order to avoid the
Itisnecessary
corrosive effects of minerals, especially chlorides, normally found in available
water sources.
FAILURE TO USE DISTILLED OR DEMINERALIZED WATER MAY IMPAIR THE INTEGRITY OF THIS
PRESSURE VESSEL WITH THE RESULTANT DANGER OF EXPLOSION AND PERSONÁL INJURY OR
PROPERTY DAMAGEE.
When sterilizing saline solutions, it is IMPERATIVE that the autoclave be cleaned
after each use,
CAUTION ELECTRICAL SHOCK HAZARD. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL.