Professional Documents
Culture Documents
© Hach Company, 1999. All rights reserved. Printed in the U.S.A. hm/dk 4/99 1ed
TRADEMARKS OF HACH COMPANY
2
TABLE OF CONTENTS
SPECIFICATIONS................................................................................................................. ... 5
SAFETY.................................................................................................................................... 7
3
TABLE OF CONTENTS, continued
4
SPECIFICATIONS
Specifications subject to change without notice
Concentration Range*
0.06 to 17,000 mg/L NH3
0.05 to 14,000 mg/L NH3-N
3.6 x 10-6 to 1M
Temperature Range
Operating: 0 to 50 °C
Storage: -40 to 60 °C
pH Range
> pH 11; Ionic Strength Adjustor raises pH above 11
Slope
–57 ±3 mV/decade in linear concentration range
Response Time
95% response in three minute(s) or less in linear range
Electrode Storage
Dry in air (membrane in sealed vial), or in Ammonia Storage
Solution. See Storing the Electrode.
Electrode Resistance
<500 Megohm
Electrode Connector
BNC (twist type)
Dimensions
Electrode length: 14 cm (5.5 in.)
Electrode diameter: 1.3 cm (0.5 in.)
Cable length: 91 cm (36 in.)
* In this manual, concentrations are expressed in mg/L (ppm) NH3-N and moles/L (M).
5
6
SAFETY
7
8
SECTION 1 INTRODUCTION
9
SECTION 1, continued
2. Unscrew the top cap and remove the glass electrode inner
body from the electrode outer body. See Figure 2.
5. Place the inner body into the outer body and screw on the
upper cap.
10
SECTION 1, continued
11
SECTION 1, continued
4. Peel the white membrane off the clear paper backing. Start at
one edge and lift the membrane gently off the paper.
6. Push the end cap over the membrane and electrode tip. Using
some pressure, screw the end cap onto the electrode until the
end cap is completely on. There must not be a gap between
the membrane and end cap.
12
SECTION 1, continued
13
SECTION 1, continued
3. Select two standards above 0.5 mg/L NH3-N that are a decade
(factor of 10) apart and fall within a concentration range that
is close to the sample concentration range.
Note: The electrode must be properly conditioned before checking
the slope.
11. Subtract the first mV reading from the second reading. The
potential should have changed –57 ±3 mV.
14
SECTION 1, continued
• Use a magnetic stirrer and stir bar; stir at a constant rate. This
speeds response time and permits a lower detection limit.
15
SECTION 1, continued
16
SECTION 2 APPLICATIONS (sension™2 & 4 meters)
17
18
Method 10000
NITROGEN, AMMONIA IN WATER (0.1 to 1000 mg/L NH3-N)
10 mg/L
100mg/L
1. Prepare the electrode 2. Rinse the electrode 3. For calibration, use 4. After conditioning,
as described in Section and place it in the two ammonia standards connect the Ammonia
1.2 on page 10. Ammonia ISE Storage of 10 and 100 mg/L ISE to the pH/ISE meter.
Solution with the NH3-N. For more Verify that BNC is
module on to condition accurate analysis, selected in Setup 1 of the
for at least 15 minutes. bracket the expected Setup menu.
sample concentration
Note: Before removing the Note: One BNC and one
electrode from the storage with two standards that five-pin connector are on
solution, the reading on are one decade apart in the back of the meter.
the meter should be below concentration. Choose the BNC for the
0.1 mg/L. This insures the ammonia electrode.
electrode is properly
conditioned. After every
hour of continuous use,
place the electrode in the
storage solution for
10 minutes to thoroughly
recondition. Check with a
10 mg/L NH3-N standard
for accuracy and
recalibrate if necessary.
19
NITROGEN, AMMONIA IN WATER, continued
Standard 1
5. Press ISE/mV until 6. Press CAL. The 7. The display will 8. If necessary, use the
the display shows mg/L display will show CAL, show Standard 1 ? and number keys to change
or other selected ?, mg/L, and the active _ _ _ _ or the value of the value to match the
concentration units. keys. The units will be standard 1 from the concentration of the
flashing. Use the arrow previous calibration. standard. Do not press
keys to select the desired ENTER yet.
units, then press ENTER.
Stabilizing...
9. Transfer 25 mL of 10. Remove the 11. Add the contents of 12. Stabilizing... will
the 10-mg/L NH3-N electrode from the one Ammonia Ionic be displayed until the
standard to a 50-mL storage solution. Rinse it Strength Adjustor reading is stable.
beaker. Add a stir bar to with deionized water Powder Pillow to the
the beaker. Place the and blot dry. Place it into standard and
beaker on a magnetic the 10-mg/L NH3-N immediately press
stirrer and stir at a standard. ENTER.
moderate rate.
Note: Be sure no air Note: At high pH,
bubbles are trapped under ammonia solutions lose
the tip of the electrode. ammonia to the
atmosphere, lowering the
concentration. It is
important to take
measurements as soon as
possible after the solution
is basic.
20
NITROGEN, AMMONIA IN WATER, continued
Repeat steps
8–14
Standard 2
13. The display will 14. Remove the 15. Repeat steps 8-14 16. The display will
show Standard 2 ? and electrode from the for each standard. After show Store?. Press
_ _ _ _ or the value of standard. Rinse with the last standard is ENTER to store the
standard 2 from the deionized water and blot measured, press EXIT. calibration or EXIT to
previous calibration. dry. Place in storage leave the calibration
solution for one minute. mode without storing the
calibration values.
-58.0
17. Press REVIEW. Use 18. Be sure the slope 19. Transfer 25 mL of 20. Remove the
the Up arrow key to displayed is within the the sample to a 50-mL electrode from the
scroll to the last slope range recommended in beaker. Add a stir bar to storage solution. Rinse it
value. It should be the electrode manual. the beaker. Place the with deionized water
-57 ±3 mV/decade. beaker on a magnetic and blot dry. Place it into
Press EXIT to return to stirrer and stir at a the sample.
the measurement mode. moderate rate.
Note: If the slope does not
approximate -57 ±3
mV/decade, recalibration
may be necessary. If the
slope is still incorrect after
recalibration, replace the
ammonia module.
21
NITROGEN, AMMONIA IN WATER, continued
Stabilizing...
21. Add the contents of 22. The display will 23. Remove the
one Ammonia Ionic show Stabilizing... until electrode from the
Strength Adjustor the reading is stable. sample. Rinse it with
Powder Pillow to the Record or store the deionized water and blot
sample. measurement value. it dry. Place it in the
storage solution for at
least one minute.
Repeat steps 19-23 for
each sample.
22
NITROGEN, AMMONIA IN WATER, continued
Accuracy Check
To verify measurement accuracy, run a standard addition spike on
the sample. This provides more complete information about
possible inaccuracies, because the measurement includes any
interferences that may be present in the actual sample.
23
NITROGEN, AMMONIA IN WATER, continued
Measured
Measured Standard
Sample Standard
Sample Concentration
Concentration Volume (mL)
Volume (mL) (mg/L)
(mg/L)
0.1–0.6 25 100 0.1
0.6–1.0 25 100 0.2
1.0–1.5 25 100 0.3
1.5–3.0 25 100 0.5
3–6 25 100 1.0
6–10 25 100 2.0
10–15 25 100 3.0
15–25 25 1000 0.5
25–35 25 1000 0.7
35–50 25 1000 1.0
50–100 25 1000 2.0
Where:
Xs = measured value for spiked sample in mg/L
Xu = measured value for unspiked sample adjusted for dilution by the
spike, in mg/L
K = known value of the spike in the sample in mg/L
Calculations:
i X ×V u
1. X u = -----------------
-
Vu + V
Where:
Xi = measured value of unspiked sample in mg/L
Vu = volume of separate unspiked portion in mL
V = volume of spike in mL
24
NITROGEN, AMMONIA IN WATER, continued
C×V
2. K = ----------------
-
Vu + V
Where:
C = concentration of standard used in spike in mg/L
V = volume of spike in mL
Vu = volume of separate portion before spike in mL
Method Performance
Precision
In a single lab using a standard solution of 5.0 mg/L and two
electrodes with a single sension2 meter, a single operator
obtained a standard deviation of 0.095 mg/L ammonia nitrogen.
Each electrode was exposed to seven test solutions and the meter
default stability criteria was 0.5 mV/min. Between test solutions,
electrodes were rinsed and placed in the storage solution until a
reading of <0.1 mg/L was obtained.
Interferences
Low molecular weight, volatile amines cause a
positive interference.
25
NITROGEN, AMMONIA IN WATER, continued
REQUIRED REAGENTS
Quantity Required
Description Per Test Unit Cat. No.
Ammonia Electrode Filling Solution, 60 mL ......................3 mL........... 5/pkg.......... 50287-05
Ammonia Electrode Storage Solution .................................1 mL........ 500 mL .......... 25412-49
Ammonia Ionic Strength Adjustor Powder Pillows................ 1 .......... 100/pkg .......... 44471-69
Ammonia (Nitrogen) Standard Solutions:
10 mg/L NH3-N...............................................................25 mL....... 500 mL .............. 153-49
100 mg/L NH3-N.............................................................25 mL....... 500 mL .......... 24065-49
Water, deionized.................................................................100 mL..............4 L .............. 272-56
REQUIRED APPARATUS
Ammonia Electrode, BNC...................................................... 1 .................each .......... 50250-00
Beaker, 50 mL, polypropylene................................................ 1 .................each ............ 1080-41
Bottle, wash ............................................................................ 1 ........... 500 mL .............. 620-11
sension™2 Portable pH/ISE Meter........................................ 1 .................each .......... 51725-00
sension™4 Laboratory pH/ISE Meter ...................................1 .................each .......... 51775-00
Stir Bar, 22.2 x 4.76 mm......................................................... 1 .................each .......... 45315-00
Select one based on available voltage:
Stirrer, electromagnetic, 115 V, with stand and stir bar.......... 1 .................each .......... 45300-01
Stirrer, electromagnetic, 230 V, with stand and stir bar.......... 1 .................each .......... 45300-02
OPTIONAL REAGENTS
Sulfuric Acid, concentrated ...................................................................500 mL.............. 979-49
OPTIONAL APPARATUS
Air Gap Assembly ......................................................................................each .......... 50253-00
Ammonia Electrode Membrane Modules................................................ 3/pkg.......... 50251-00
Cylinder, graduated, glass......................................................................100 mL.............. 508-42
Electrode Washer ........................................................................................each .......... 27047-00
26
Method 10008
AMMONIA DISTILLATION
PROCEDURE
27
AMMONIA DISTILLATION, continued
5. Turn the heater on 6. Turn the heater off 7. Collect 250 mL of 8. Using the pH
and set the control to 10. when done steaming. sample in a 250-mL electrode, neutralize the
Heat the flask and steam graduated mixing sample in the cylinder to
out the apparatus until cylinder. Add 0.5 mL of pH 7 with 1 N sodium
the distillate shows no dechlorinating agent for hydroxide or 1 N
traces of ammonia every mg/L Cl2, if sulfuric acid.
(check with ammonia necessary. Stopper and
Note: Periodically stopper
ISE to ensure ammonia invert to mix. and invert the cylinder
is gone). several times to ensure
Note: For proof of
accuracy, use a 1.0 mg/L complete mixing.
ammonia nitrogen
standard (see Optional
Reagents) in place of the
sample.
9. Measure 12.5 mL of 10. Using the pH 11. Transfer sample to 12. Measure 25 mL of
borate buffer in a electrode, adjust the pH distillation flask using 0.04 N sulfuric acid into
graduated cylinder. Add of the solution to 9.5 by a funnel. Place a stir bar a graduated cylinder.
the measured buffer to slowly adding 6 N in the flask. Pour this into a 250-mL
the solution in the sodium hydroxide. Erlenmeyer receiving
cylinder. flask (comes with the
Note: Periodically stopper
and invert the cylinder distillation apparatus).
several times to ensure
complete mixing.
28
AMMONIA DISTILLATION, continued
13. Turn the water to 14. Distill until at least 15. Lower the 16. Place a 100-mL
the condenser on. Place 100 mL of distillate is receiving flask away beaker under the
the receiving flask so the collected or until there is from the delivery tube. delivery tube. Continue
tip of the delivery tube is at least 125 mL total Immediately cover the to distill for 1–2 minutes
below the surface of the volume in the receiving flask with parafilm or a to clean the dispenser
sulfuric acid. flask. stopper to prevent and delivery tube. Then
Note: For the Hach
ammonia contamination. turn the heater off.
Distillation Apparatus, set
the heat control to 10 and
the stir control to 5.
Ready
for
analysis
29
AMMONIA DISTILLATION, continued
REQUIRED REAGENTS
Quantity Required
Description Per Test Unit Cat. No.
Borate Buffer Solution................................................22.5 mL...........1000 mL .......... 14709-53
Sodium Hydroxide, 6 N ................................................varies .......................1L .......... 23324-53
Sodium Hydroxide, 1 N ................................................varies . .100 mL MDB* ............ 1045-32
Sodium Thiosulfate, pentahydrate, ACS......................0.5 mL ................. 454 g .............. 460-01
Sulfuric Acid, 1 N.........................................................varies .............1000 mL ............ 1270-53
Sulfuric Acid, 0.04 N................................................... 50 mL ..............500 mL.......... 23393-49
REQUIRED APPARATUS
Beaker, 400 mL................................................................ 1........................each .............. 500-48
Cylinder, graduated, mixing, 250 mL .............................. 1........................each .......... 20886-46
Distillation Apparatus, General Purpose ......................... 1........................each .......... 22653-00
Flask, volumetric, 250-mL, Nalge ................................... 1........................each .......... 14060-46
Funnel, Nalge................................................................... 1........................each ............ 1083-67
OPTIONAL REAGENTS
Nitrogen, Ammonia Standard Solution, 1.0 mg/L as N.........................500 mL............ 1891-49
OPTIONAL APPARATUS
Platinum Series pH Electrode, for use with sension™ meters, 5-pin ........each .......... 51910-00
sension™2 Portable pH Meter...................................................................each .......... 51725-00
sension™3 Laboratory pH Meter ..............................................................each .......... 51750-00
sension™4 Laboratory pH/ISE Meter .......................................................each .......... 51775-00
30
Method 10001
NITROGEN, AMMONIA IN WASTEWATER (0.1–10 mg/L NH3-N)
USPEA accepted for reporting*
Calibration
Prepare ammonia standard working solutions of 10.0, 1.0 and
0.1 mg/L ammonia nitrogen from a 100-mg/L stock solution.
Prepare the standards daily before use. Higher or lower
concentration ranges (0.05–1400 mg/L NH3-N) can be obtained
by calibrating the meter with different standard solutions.
PROCEDURE
* Manual distillation is not required if comparability data on representative samples in company files
show the distillation is not necessary. Manual distillation will be required to resolve any controversies.
31
NITROGEN, AMMONIA IN WASTEWATER, continued
5. Prepare a 0.1-mg/L 6. Connect the 7. Turn the meter on. 8. Press CAL. The
NH3-N standard by Ammonia ISE to the Press ISE/mV until the display will show CAL,
pipeting 25 mL of the pH/ISE meter. Verify display shows mg/L or ?, mg/L, and the active
1.0-mg/L NH3-N that BNC is selected in other chosen keys. The units will be
standard into a 250-mL Setup 1 of the Setup concentration units. flashing. Use the arrow
volumetric flask. Dilute menu. keys to select the desired
to the mark with units, then press ENTER.
Note: One BNC and one
ammonia -free deionized five-pin connector are on
water. the back of the meter.
Choose the BNC for the
ammonia electrode.
9. Transfer 100 mL of 10. Remove the 11. Pipet 2.0 mL of 12. The display will
the 0.1-mg/L NH3-N electrode from the Ammonia ISA Solution show Standard 1? and
standard to a 150-mL storage solution. Rinse it into the standard. the value from the
beaker. Add a stir bar to with deionized water Immediately proceed to previous calibration.
the beaker. Place the and blot dry. Place the the next step. Press ENTER to accept
beaker on a magnetic electrode into the the numerical value or
stirrer and stir at a 0.1-mg/L NH3-N use the number keys to
moderate rate. standard. change the value to
match the concentration
Note: Placing the Note: Be sure no air
electrode back into the bubbles are trapped under of the standard, then
storage solution prepares the tip of the electrode. press ENTER.
it for the next standard.
Leave it in the storage
solution for at least one
minute before proceeding
to the next standard.
32
NITROGEN, AMMONIA IN WASTEWATER, continued
Repeat steps
Stabilizing... 9–14
13. Stabilizing... will 14. Rinse the electrode 15. After the last 16. The display will
be displayed until the with deionized water. standard is measured, show Store?. Press
reading is stable. The Place it in the storage press EXIT. ENTER to store the
display will show solution for one minute. calibration or EXIT to
Standard 2 and _ _ _ _ Repeat steps 9-14 for the leave the calibration
or the value of standard 1.0- and 10-mg/L mode without storing the
2 from the previous standards. calibration values.
calibration.
17. Press REVIEW. Use 18. Remove the 19. Transfer 100 mL of 20. Remove the
the Up arrow key to electrode from the last sample to a 150-mL electrode from the
scroll to the second standard. Rinse it with beaker. Add a stir bar to storage solution. Rinse it
slope value. It should be deionized water, and the beaker. Place the with deionized water
-57 ±3 mV/decade. place it in the storage beaker on a magnetic and blot dry. Place it into
Press EXIT to return to solution. stirrer and stir at a the sample.
measurement mode. moderate rate.
Note: If the slope is not -
57 ±3 mV/decade,
recalibration may be
necessary. If the slope is
still incorrect after
recalibration, replace the
ammonia module.
33
NITROGEN, AMMONIA IN WASTEWATER, continued
Repeat steps
18–22
21. Pipet 2.0 mL of 22. Stabilizing... will Note: For the distillation
Ammonia ISA solution be displayed until the procedure in this manual,
into the sample and reading is stable. Record the capture volume after
distillation and the sample
proceed immediately to or store the
volume before distillation
the next step. measurement value. is 250 mL, so the
Note: At high pH, Repeat steps 18–22 for concentration from this
ammonia solutions lose step is the actual
other samples.
ammonia to the concentration.
atmosphere, lowering the Note: Stabilization will
take about 1 to 2 minutes Note: After every hour of
concentration. It is
for concentrations above continuous use the
important to take
1.0 mg/L NH3-N and electrode should be
measurements as soon as
longer for lower placed in the storage
possible after the solution
concentrations. Slow solution for a 10 minute
is basic.
downward drift in period for thorough
Note: For most concentration indicates reconditioning. Check with
wastewater samples, 1 mL probable loss of ammonia a 10 mg/L NH3-N standard
of 10 N NaOH (or to the atmosphere. for accuracy and
equivalent ISA) is Record the highest value recalibrate if necessary.
sufficient to increase the that is stable.
pH above 11. If in doubt,
check the pH with pH
paper (Cat. No. 385-33)
and add additional NaOH
in 0.1 mL increments until
the pH exceeds 11.
34
NITROGEN, AMMONIA IN WASTEWATER, continued
Accuracy
To verify measurement accuracy, run a standard addition spike on
the sample. This provides more complete information about
possible inaccuracies, because the measurement includes any
interferences that may be present in the actual sample. The spike
should roughly double the measured concentration without
significantly diluting the sample.
35
NITROGEN, AMMONIA IN WASTEWATER, continued
Measured
Measured Standard
Sample Standard
Sample Concentration
Concentration Volume (mL)
Volume (mL) (mg/L)
(mg/L)
0.1–0.3 100 100 0.2
0.3–0.5 100 100 0.4
0.5–0.7 100 100 0.6
0.7–0.9 100 100 0.8
0.9–1.1 100 100 1.0
1.0–3.0 100 100 2.0
3.0–6.0 100 100 4.0
Calculations:
i X ×V u
1. X u = -----------------
-
Vu + V
Where:
Xi = measured value for spiked sample in mg/L
Xu = volume of separate unspiked portion in mL
V = volume of spike in mL
36
NITROGEN, AMMONIA IN WASTEWATER, continued
C×V
2. K = ----------------
-
Vu + V
Where:
C = concentration of standard used in spike in mg/L
V = volume of spike in mL
Vu = volume of separate distillate (100 mL)
mg/L × 4 mL
2. K = 100
--------------------------------------------- = 3.85 mg/L
104 mL
Method Performance
Precision
In a single lab using a standard solution of 0.8 mg/L and two
electrodes with a single sension™2 meter, a single operator
obtained a standard deviation of 0.01 mg/L ammonia nitrogen.
Each electrode was exposed to seven test solutions and a default
stability criteria was 0.5 mV/min. Between test solutions, the
electrodes were rinsed and placed in the storage solution until a
reading of <0.1 ppm was obtained.
Interferences
Low molecular weight, volatile amines give a positive
interference. Mercury and silver interfere by complexing with
ammonia. Distillation prior to ammonia analysis removes all
inorganic interferences, which complex ammonia.
37
NITROGEN, AMMONIA IN WASTEWATER, continued
38
NITROGEN, AMMONIA IN WASTEWATER, continued
REQUIRED REAGENTS
Quantity Required
Description Per Test Unit Cat. No.
Ammonia Electrode Filling Solution, 60 mL...................... drops ........... 5/pkg ..........50287-05
Ammonia Electrode Storage Solution................................ 20 mL ...... 500 mL ..........25412-49
Ammonia Nitrogen Standard Sol’n, 100 mg/L NH3-N .... 100 mL ..... 500 mL ..........24065-49
Water, deionized ................................................................ 100 mL ............ 4 L ..............272-56
REQUIRED APPARATUS
Ammonia Electrode, combination, BNC ............................... 1................. each ..........50250-00
Beaker, 150 mL, polypropylene ............................................. 1................. each ............1080-44
Bottle, wash............................................................................ 1........... 500 mL ..............620-11
Flask, volumetric, Class A, 250 mL....................................... 1................. each ..........14574-46
Pipet, volumetric, Class A, 25.00 mL .................................... 2................. each ..........14515-40
Pipet, TenSette®, 1.0-10.0 mL ............................................... 1................. each ..........19700-10
sension™4 Laboratory pH/ISE Meter................................... 1................. each ..........51775-00
Stir Bar, 22.2 x 4.76 mm (7/16 x 3/16 in.) ................................ 1................. each ..........45315-00
Select one based on available voltage:
Stirrer, electromagnetic, 115 V, with stand and stir bar ......... 1................. each ..........45300-01
Stirrer, electromagnetic, 230 V, with stand and stir bar ......... 1................. each ..........45300-02
OPTIONAL REAGENTS
Ammonia Nitrogen Standard Solution, 1000 mg/L NH3-N.......................... 1L ..........23541-53
pH Paper, 9.0–12.0 pH units ............................................................ 5 rolls/pkg ..............385-33
Sulfuric Acid, concentrated................................................................... 500 mL ..............979-49
OPTIONAL APPARATUS
Air Gap Assembly...................................................................................... each ..........50253-00
Ammonia Electrode Membrane Modules ................................................ 4/pkg ..........44475-00
Cylinder, graduated, glass ..................................................................... 100 mL ..............508-42
Electrode Washer........................................................................................ each ..........27047-00
Pipet tips for 19700-01 TenSette Pipet....................................................50/pkg ..........21856-96
Pipet tips for 19700-10 TenSette Pipet....................................................50/pkg ..........21997-96
Pipet, Volumetric, Class A, 1.00 mL.......................................................... each ..........14515-35
sension™2 Portable pH/ISE Meter ........................................................... each ..........51725-00
TenSette® Pipet, 0.1-1.0 mL ...................................................................... each ..........19700-01
39
40
Method 10002
NITROGEN, AMMONIA
KNOWN ADDITION METHOD (≥0.8 mg/L NH3-N) For Wastewater
Standard Method 4500-NH3G
USEPA Accepted for reporting wastewater analyses (with distillation)*
Because an accurate measurement requires that the concentration
approximately double as a result of the addition, the approximate
sample concentration must be known within a factor of three.
Expected Sample
mL of 1000-mg/L Standard
Concentration
NH3-N Concentration
(mg/L)
0.8–4.0 2 20 mg/L
2.5–7.5 5 50 mg/L
5–15 10 100 mg/L
12–50 25 250 mg/L
25–75 50 500 mg/L
50–150 100 1000 mg/L
Slope Determination
Determine the slope before performing standard additions of the
sample. Use the 100 mg/L and 1000-mg/L NH3-N stock solutions
to determine the slope. Check the electrode occasionally to
determine if it is functioning properly and to determine its exact
slope value. The frequency of this operation depends on the
harshness of your sample.
* Manual distillation is not required if comparability data on representative samples in company files
show the distillation is not necessary. Manual distillation will be required to resolve any controversies.
41
NITROGEN, AMMONIA KNOWN ADDITION METHOD, continued
1. Prepare the electrode 2. Rinse the electrode 3. Connect the 4. Accurately transfer
as described in Section and place it in the Ammonia ISE to the 100 mL of NH3-N free
1.2 on page 10. Ammonia ISE Storage pH/ISE meter. Verify deionized water to a
Solution to condition for that BNC is selected in 150-mL beaker using a
Note: Only the Hach
sension 4 can perform at least 15 minutes.
™ Setup 1 of the Setup volumetric pipet or
Standard Additions. menu. graduated cylinder. Add
Note: Before removing the
electrode from the storage a stir bar to the beaker.
Note: Two BNC and two
solution, the reading on five-pin connectors are on Place the beaker on a
the meter should be below the back of the meter. magnetic stirrer and stir
0.1 mg/L. This insures the Choose a BNC for the at a moderate rate.
electrode is properly ammonia electrode.
conditioned. After every Note: Stirring at a
hour of continuous use, constant, moderately fast
place the electrode in the rate speeds response and
storage solution for 10 improves accuracy.
minutes to thoroughly
recondition. Check with a
10 mg/L NH3-N standard
for accuracy and
recalibrate if necessary.
42
NITROGEN, AMMONIA KNOWN ADDITION METHOD, continued
5. Pipet 1.0 mL of 6. Remove the 7. Turn the meter on. 8. The display will
10 N NaOH solution electrode from the Press STD ADDN. Press show Slope and the
into the sample. storage solution. Rinse it the arrow keys until the slope value for the last
with deionized water display shows the calibration (default is
Note: The 10-mL Hach
TenSette Pipet works and blot dry. Place the desired units. –59.2 mV). Press
conveniently for this electrode in the sample. Press ENTER. ENTER to accept the
addition, as well as the numerical value or use
Note: Be sure no air
remainder of the additions the number keys to
bubbles are trapped under
in this procedure. change the slope value.
the tip of the electrode.
Remove bubbles by lightly Press ENTER to accept
tapping the electrode or by the new value.
tilting the electrode to 20°.
9. The meter will 10. The display will 11. Use the number 12. The display will
prompt for the sample show Stabilizing... until keys to enter the volume show Standard, ?, and
volume (in mL). The the baseline reading is of standard to be used mg/L (or chosen units).
display will show mL, stable. The meter will (for example, 1.0 mL).
Sample, ?, and _ _ _ _. then prompt for the Press ENTER.
Use the number keys to standard volume. The Note: Obtain the standard
enter the sample volume, display will show mL, concentration and volume
then press ENTER. Standard, ?, and from Table 4 on page 45
_ _ _ _. after estimating the
sample concentration.
Note: Place the electrode
in the storage solution for
at least one minute to
prepare it for the next
standard.
43
NITROGEN, AMMONIA KNOWN ADDITION METHOD, continued
Stabilizing... 20 mg/L
13. Add the volume of 14. Use the number 15. Sample+Standard 16. The meter will
a known standard (listed keys to enter the and Stabilizing... will be calculate and display the
in Table 4) to the beaker concentration of the displayed until the adjusted value for the
and proceed as quickly standard used (for reading is stable. original sample. Record
as possible through the example, 1000 mg/L). or store this value as
rest of the procedure. Press ENTER. usual.
Note: When recalling data
for standard additions,
STANDARD ADDITONS
will appear in the display
with the data.
44
NITROGEN, AMMONIA KNOWN ADDITION METHOD, continued
45
NITROGEN, AMMONIA KNOWN ADDITION METHOD, continued
Calculations
i X ×V u
1. X u = -----------------
-
Vu + V
Where:
Xi = measured value of unspiked sample in mg/L
Vu = volume of separate unspiked portion in mL
V = volume of spike in mL
C × V-
2. K = ----------------
Vu + V
Where:
C = concentration of standard used in spike in mg/L
V = volume of spike in mL
Vu = volume of separate portion before spike in mL
mg/L × 4 mL
2. K = 100
--------------------------------------------- = 3.85 mg/L
100 mL + 4 mL
46
NITROGEN, AMMONIA KNOWN ADDITION METHOD, continued
Method Performance
Precision
In a single lab using a standard solution of 5.0 mg/L and two
electrodes with a single sension4 meter, a single operator
obtained a standard deviation of 0.08 mg/L ammonia nitrogen.
Each electrode was exposed to seven test solutions and a default
stability of 0.5 mV/min. Between test solutions, the electrodes
were rinsed and placed in storage solution until a reading of
0.1 mV was obtained.
Interferences
Low molecular weight, volatile amines give a positive
interference. Mercury and silver interfere by complexing with
ammonia. Distillation prior to ammonia analysis removes all
inorganic interferences, such as metals which complex ammonia.
47
NITROGEN, AMMONIA KNOWN ADDITION METHOD, continued
REQUIRED REAGENTS
Quantity Required
Description Per Test Unit Cat. No.
Ammonia Electrode Filling Solution...................................3 mL.......... 50 mL.......... 44472-26
Ammonia Electrode Storage Solution .................................5 mL........ 500 mL .......... 25412-49
Ammonia Nitrogen Standard, 1000 mg/L NH3-N.............. varies .............. 1 L.......... 23541-53
Sodium Hydroxide Solution, 10 N .....................................10 mL........500 mL .......... 25450-49
Water, deionized.................................................................100 mL............. 4 L.............. 272-56
REQUIRED APPARATUS
Ammonia Electrode ................................................................ 1 .................each .......... 50250-00
Beaker, 50 mL, polypropylene................................................ 1 .................each ............ 1080-41
Bottle, wash .......................................................................................... 500 mL .............. 620-11
Cylinder, graduated, 100-mL .................................................. 1 .................each .............. 508-42
sension™4 Laboratory pH/ISE Meter ...................................1 .................each .......... 51775-00
Stir Bar, 22.2 x 4.76 mm......................................................... 1 .................each .......... 45315-00
TenSette® Pipet, 1.0-10.0 mL ................................................. 1 .................each .......... 19700-10
Pipet tips for 19700-10 TenSette Pipet ................................................... 50/pkg .......... 21997-96
Select one based on available voltage:
Stirrer, electromagnetic, 115 V, with stand and stir bar.......... 1 .................each .......... 45300-01
Stirrer, electromagnetic, 230 V, with stand and stir bar.......... 1 .................each .......... 45300-02
OPTIONAL REAGENTS
pH Paper, pH 9.0-12.0 ..................................................................... 5 rolls/pkg .............. 385-33
Sulfuric Acid, concentrated ...................................................................500 mL.............. 979-49
OPTIONAL APPARATUS
Ammonia Electrode Membrane Modules................................................ 4/pkg.......... 44475-00
Electrode Washer ........................................................................................each .......... 27047-00
Pipet, Volumetric, Class A ....................................................................1.00 mL .......... 14515-35
Pipet, Volumetric, Class A ..................................................................10.00 mL .......... 14515-38
Pipet, Volumetric, Class A ..................................................................100.0 mL .......... 14515-42
TenSette® Pipet, 0.1-1.0 mL .......................................................................each .......... 19700-01
Pipet tips for 19700-01 TenSette Pipet .................................................. 50/pkg .......... 21856-96
sension™2 Portable pH/ISE Meter............................................................each .......... 51725-00
48
Method 8102
DIGESTION FOR LIQUIDS For Determining Ammonia Using TKN
49
DIGESTION FOR LIQUIDS, continued
5. Heat 4 more 6. Do not proceed if 7. Boil off excess 8. Take the flask off the
minutes. Do not boil sulfuric acid is not hydrogen peroxide by heater and allow to cool
sample to dryness. If visible in the flask! Add heating for one more for 5 minutes. Remove
sulfuric acid is not 10 mL of 50% hydrogen minute after addition of the fractionating column
present in the flask after peroxide to the charred the hydrogen peroxide is from the digestion flask.
the 4 minute heating, do sample via the funnel on complete. Do not heat to Note: Use finger cots to
not proceed with step 6. the fractionating dryness. remove the digestion flask
If sample evaporates to column. and place on a cooling
dryness, discard it and Note: If the digest does
pad for several seconds.
use more sulfuric acid in Then remove the column.
not turn colorless, add
step 2 or choose a Do not add water to the
5-mL increments of
smaller sample amount flask until it has cooled.
hydrogen peroxide until
for digestion in step 1. the digest becomes clear
or the color of the digest
Note: Visually confirm the does not change upon
presence of sulfuric acid addition of hydrogen
in the flask before adding peroxide.
hydrogen peroxide.
9. When the flask is 10. Add 0.5 mL of 11. Neutralize the 12. Slightly acidify the
cool enough to handle, Thymol Blue Indicator sample by adding 10 N sample by adding 1.90 N
remove the weight and to the flask. The solution NaOH dropwise, until sulfuric acid dropwise,
add about 75 mL of should turn pink. the solution just turns until one drop turns the
ammonia-free deionized blue. This will take sample yellow. Stopper
water. Stopper and invert about 6 mL of 10 N the flask and invert it
several times to mix. NaOH. several times to make
sure the yellow color
persists.
50
DIGESTION FOR LIQUIDS, continued
REQUIRED REAGENTS
Quantity Required
Description Per Digestion Unit Cat. No.
Hydrogen Peroxide, 50% ................................................... 10 mL ....... 490 mL ..........21196-49
Potassium Hydroxide Standard Solution, 1 N, SCDB* ......varies....... ...50 mL ..........23144-26
Potassium Hydroxide Standard Solution, 8 N.....................varies........ 500 mL ..............282-49
Sodium Hydroxide, 10 N .................................................... 6 mL ........ 500 mL ..........25450-49
Sulfuric Acid, ACS (conc., specific gravity 1.84).............. >3 mL ........... 2.5 L ..............979-09
Sulfuric Acid, 1.90 N, MDB ...............................................varies..... ...100 mL ..........21830-32
Water, deionized ..................................................................varies..........4 liters ..............272-56
REQUIRED APPARATUS
Boiling Chips, silicon carbide .............................................varies............500 g ..........20557-34
Dispenser, pour-out, 10 mL.................................................... 1................. each ..........22200-38
Flask, flat-bottom, volumetric, w stopper, 100 mL ................ 1................. each ..........23125-42
Pipet, serological, 10 mL........................................................ 1................. each ..............532-38
Pipet Filler, safety bulb........................................................... 1................. each ..........14651-00
Safety Glasses ........................................................................ 1................. each ..........18421-00
Safety Shield, for Digesdahl................................................... 1................. each ..........50040-00
51
DIGESTION FOR LIQUIDS, continued
1. Prepare the 2. Rinse the electrode 3. Connect the 4. Press ISE/MV until
electrode as described in and place it in the Ammonia ISE to the mg/L (or the user
Section 1.2 on page 10 Ammonia ISE storage meter. Verify that BNC selected concentration
Note: The module must
solution with the module is selected in Setup 1 of unit) appears.
be tightly screwed on. on to condition for at the Setup menu.
least 15 minutes. Note: The sension™4
Note: Use the Digestion
For Liquids procedure Note: Use Hach Ammonia meter has two BNC and
preceding this to digest Nitrogen Standards. They two five-pin connectors.
contain thymol blue to Choose one of the BNC
the sample before
indicate when the pH is connectors for the
analyzing for TKN.
high enough to begin ammonia electrode.
measurement.
53
TOTAL KJELDAHL NITROGEN, continued
Standard 1 ?
5. Press CAL. The 6. The display will 7. Fill a 50-mL beaker 8. Remove the
display will show CAL, show Standard 1 ? and to the 20-mL mark with electrode from the
?, mg/L, and the active the value of standard 1 a 1.0-mg/L NH3-N storage solution, rinse it
keys. The units will be from the previous standard. Add a stir bar with deionized water
flashing. Use the arrow calibration. If necessary, to the beaker. Place the and blot dry. Place the
keys to select the desired use the number keys to beaker on a magnetic electrode into the
units, then press ENTER. change the value to stirrer and stir at a 1.0-mg/L standard.
match the concentration moderate rate. Note: Before removing the
of the standard Note: Stirring at a electrode from the storage
(1.0 mg/L). Do not press constant rate speeds solution, the reading on
ENTER yet. response and improves the meter should be below
accuracy. For very 0.1 mg/L. This insures the
accurate work, use the electrode is properly
Hach Electromagnetic conditioned. After every
Stirrer to minimize heat hour of continuous use,
transfer to the sample. place the electrode in the
storage solution for 10
minutes to thoroughly
recondition. Check with a
10 mg/L NH3-N standard
for accuracy and
recalibrate if necessary.
54
TOTAL KJELDAHL NITROGEN, continued
Stabilizing... ----
9. Add one 1-g scoop 10. Add the contents 11. The display will 12. The display will
of sodium sulfate to the of one Ammonia ISA show Stabilizing... until show_ _ _ _ or the value
standard. Stir until all powder pillow. the reading is stable. of standard 2 from the
the solid is dissolved. Immediately press previous calibration.
ENTER. The meter will
Note: At a high pH,
ammonia solutions lose begin measuring the
ammonia to the standard. Proceed
atmosphere, lowering the immediately to the
concentration. It is next step.
important to take
measurements as soon as
possible after the solution
is basic (turns blue).
Repeat steps
7-13
13. Place the electrode 14. Repeat steps 7-13 15. After the last 16. The display will
into the storage solution for the 10 and 100-mg/L standard is measured, show Store?. Press
for at least one minute. standards. press EXIT. ENTER to store the
Note: After the seventh
calibration or EXIT to
ISE standard, the meter leave the calibration
automatically displays the mode without storing the
STORE prompt. calibration values.
Note: Pressing REVIEW
will put the meter in
calibration review mode.
Be sure the slope is
–57 ±3 mV.
55
TOTAL KJELDAHL NITROGEN, continued
Stabilizing...
17. Fill a 50-mL 18. Remove the 19. Add the contents of 20. Stabilizing... will
beaker to the 20-mL electrode from the one powder pillow of be displayed until the
mark with digested storage solution, rinse it Ammonia ISA. reading is stable. Record
sample. Add a stir bar to with deionized water, Note: Do not add sodium
or store the
the beaker. Place the and blot dry. Place the sulfate to the digested measurement value.
beaker on a magnetic electrode into the sample. Repeat steps 17-20 for
stirrer and stir at a digested sample.
Note: At high pH, other samples.
moderate rate. Note: Before removing the ammonia solutions lose Note: It may take up to 10
electrode from the storage ammonia to the minutes for the probe to
solution, the reading on atmosphere, lowering the stabilize.
the meter should be below concentration. It is
0.1 mg/L. This insures the important to take
electrode is properly measurements as soon as
conditioned. After every possible after the solution
hour of continuous use, is basic (turns blue).
place the electrode in the
storage solution for 10
minutes to thoroughly
recondition. Check with a
10 mg/L NH3-N standard
for accuracy and
recalibrate if necessary.
56
TOTAL KJELDAHL NITROGEN, continued
57
TOTAL KJELDAHL NITROGEN, continued
Digest
Measured Digest Spiked Sample
Measured Sample Standard
Sample Volume Standard Digest Volume
Concentration (mg/L) Concentration
(mL) Volume (mL) (mL)
(mg/L)
1–3 100 1000 0.2 40
3–5 100 1000 0.4 40
5–20 100 1000 1.0 10
20-100 100 1000 5.0 10
100–250 100 1000 10 10
250–425 100 1000 15 10
425–1000 100 10000 5 1
1000–3000 100 10000 10 1
3000–5000 100 10000 25 1
58
TOTAL KJELDAHL NITROGEN, continued
Calculations:
Xi × Vu
1. X u = -----------------
-
Vu + V
Where:
Xi = measured value of unspiked sample in mL
Vu = volume of separate unspiked portion in mL
V = volume of spike in mL
C×V
2. K = -----------------
Vu + V
Where:
C = concentration of standard used in the spike, in mg/L
Vu = volume of separate portion before spike, in mL
V = volume of spike, in mL
59
TOTAL KJELDAHL NITROGEN, continued
REQUIRED REAGENTS
Quantity Required
Description Per Test Unit Cat. No.
Ammonia Electrode Filling Solution.................................. drops.......... 50 mL.......... 44472-26
Ammonia Ionic Strength Adjustor (ISA)............................ varies ...... 100/pkg .......... 44471-69
Ammonia Electrode Storage Solution ................................25 mL........500 mL.......... 25412-49
Ammonia (Nitrogen) Standard Solutions:
1 mg/L NH3-N..............................................................20 mL....... 500 mL ...........1891-49
10 mg/L NH3-N............................................................20 mL....... 500 mL .............153-49
100 mg/L NH3-N..........................................................20 mL....... 500 mL ........ 24065-49
Hydrogen Peroxide Solution, 50% .....................................10 mL........500 mL.......... 21196-49
Indicator, Thymol Blue ...................................................... 3 drops ...... 100 mL .............. 257-32
Sodium Sulfate, ACS ............................................................ 3 g............. 454 g ............ 7099-01
Sulfuric Acid, 1.900 N........................................................ varies ....... 100 mL .......... 21830-32
Water, deionized.................................................................. varies .............. 4 L.............. 272-56
REQUIRED APPARATUS
Ammonia Electrode, combination, BNC................................ 1 .................each .......... 50250-00
Beaker, 50 mL, polypropylene................................................ 1 .................each ............ 1080-41
Boiling Chips, silicone carbide............................................. 2-4 ............. 500g .......... 20557-34
Bottle, wash, 500 mL ..............................................................1 .................each .............. 620-11
Cylinder, graduated, 100 mL .................................................. 1 .................each .............. 508-42
Dropper, glass, 0.5-1.0 mL .....................................................1 ............... 5/pkg .......... 14197-05
sension™4 Laboratory pH/ISE Meter ...................................1 .................each .......... 51775-00
Pipet, volumetric, Class A, 1.00 mL....................................... 1 .................each .......... 14515-35
Pipet, volumetric, Class A, 10.00 mL..................................... 1 .................each .......... 14515-38
Scoop, 1 g ............................................................................... 1 .................each .............. 510-00
Stir Bar, 22.2 x 4.76 cm .......................................................... 1 .................each .......... 45315-00
60
TOTAL KJELDAHL NITROGEN, continued
OPTIONAL REAGENTS
Description Unit Cat. No.
Primary Standards for Kjeldahl Nitrogen.............................................. set of 3 ..........22778-00
61
62
Method 10006
AMMONIA IN SOLUTIONS CONTAINING FATS,
FATTY ACIDS, OILS, AND SURFACTANTS
Principle of Operation
The Ammonia Electrode uses a fine, porous membrane to
separate the sample solution from a thin layer of electrolyte that
coats the pH sensing bulb inside the electrode. Ammonia gas can
diffuse through, but other ions may not. When the Air Gap
Accessory is used, substances that can foul the membrane are
separated from the membrane by a small air space. Ammonia gas
diffuses through the air space and then through the membrane.
The air space has slightly different diffusion characteristics than
the membrane, which causes an increased response time.
63
AMMONIA IN SOLUTIONS CONTAINING FATS, FATTY ACIDS,
OILS, AND SURFACTANTS, continued
64
AMMONIA IN SOLUTIONS CONTAINING FATS, FATTY ACIDS,
OILS, AND SURFACTANTS, continued
Calibration
1. Connect the 2. Turn the meter on. 3. Press SETUP. Press 4. Press ENTER and
Ammonia ISE to the Press ISE/mV until the the Up arrow until the use the numerical keys
sension™4 pH/ISE display shows mg/L (or display shows to edit the display until it
meter. Verify that BNC user selected Stabilizing... and a mV shows 0.1 mV/min.
is selected in Setup 1 of concentration units). value. Press ENTER, then press
the Setup menu. EXIT.
Note: Only the sension4
meter is recommended for
this method.
5. Fill the Air Gap 6. Add a stir bar to each 7. Add the contents of 8. Holding the cup
Accessory sample cup to cup. Place the cup the one Ammonia Ionic firmly with one hand,
the line with 1.0-mg/L 1.0-mg/L standard on a Strength Adjustor carefully insert the
NH3-N standard. stirrer. Stir at a moderate powder pillow to the cup electrode holder into the
rate. When the speed is on the stirrer. Do not stir. sample cup with a slight
Fill another sample cup
properly adjusted, turn twist to set the O-ring.
to the line with 10-mg/L
the motor off. Bubbling at the air vent
NH3-N standard.
may occur.
Note: Stir so there is
slight vortexing of the
sample in the cup.
65
AMMONIA IN SOLUTIONS CONTAINING FATS, FATTY ACIDS,
OILS, AND SURFACTANTS, continued
Stabilizing...
9. Turn the stirrer on 10. After the two 11. The display will 12. The display will
and stir for 2 minutes. minute period, press show Standard 1, ?, and show Stabilizing... until
CAL. The display will the value of standard 1 the reading is stable. The
show CAL, ?, mg/L, from the previous display will then show
and the active keys. The calibration. Press _ _ _ _ or the value of
units will be flashing. ENTER to accept the standard 2 from the
Use the arrow keys to numerical value or use previous calibration.
select the desired units, the number keys to
Note: It may take up to
then press ENTER. change the value to five minutes for the
match the 1.0-mg/L electrode to stabilize.
concentration of the
standard, then press
ENTER.
13. Remove the 14. Rinse the electrode 15. Place the 10-mg/L 16. Holding the cup
electrode holder holder with deionized standard on the stirrer. firmly with one hand,
assembly by holding the water, including the Add the contents of one carefully insert the
sample with one hand on inside sample reservoir Ammonia Ionic Strength electrode holder into the
a horizontal surface and via the reservoir vent. Adjustor powder pillow sample cup with a slight
pulling out with a slight Gently blot the entire to the cup on the stirrer. twist to set the O-ring.
twist. Shake any excess holder. Turn the stirrer Do not stir. Bubbling at the air vent
sample off the holder off. may occur.
while keeping the holder
in a vertical position.
66
AMMONIA IN SOLUTIONS CONTAINING FATS, FATTY ACIDS,
OILS, AND SURFACTANTS, continued
Repeat steps
Stabilizing... 13-19
17. Turn the stirrer on 18. Press ENTER to 19. The display will 20. Repeat steps 13-19
and stir for 2 minutes. accept the displayed show Stabilizing... until for each additional
value for standard 2 or the reading is stable. The standard.
use the numerical keys display will then show
to change the value to _ _ _ _ or the value of
10.0 mg/L, then press standard 3 from the
ENTER. previous calibration.
21. After the last 22. The display will 23. Press REVIEW. Use 24. Fill a sample cup to
standard is measured, show Store?. Press the Up arrow key to the line with sample.
press EXIT. ENTER to store the scroll to the last slope Add a stir bar.
calibration or EXIT to value. It should be
Note: After the seventh
standard, the meter leave the calibration -57 ±3 mV/decade.
automatically displays the mode without storing the Press EXIT to return to
STORE prompt. calibration values. measurement mode.
Note: If the slope is not
approximate -57 ±3
mV/decade, recalibration
may be necessary. If the
slope is still incorrect after
recalibration, replace the
ammonia module.
67
AMMONIA IN SOLUTIONS CONTAINING FATS, FATTY ACIDS,
OILS, AND SURFACTANTS, continued
Sample Measurement
Stabilizing...
25. Place the cup on 26. Holding the cup 27. Turn the stirrer on 28. The display will
the stirrer and stir at the firmly with one hand, and stir for 2 minutes. show Stabilizing... until
same speed as the carefully insert the the reading is stable.
standards. When the electrode holder into the Record or store the
speed is adjusted sample cup and twist value.
correctly, turn the slightly to set the O-ring.
stirrer off.
Repeat steps
24-31.
29. Remove the 30. Rinse the electrode 31. Repeat steps 24-31 32. After the last
electrode holder holder with deionized for each sample. sample, gently wash and
assembly by holding the water, including the blot the entire electrode.
sample in one hand on a inside sample reservoir Turn the stirrer off.
horizontal surface and via the reservoir vent. Be
pulling out with a slight careful not to rinse oils,
twist. Shake any excess sludge or surfactants
sample off the holder back into the air gap
while keeping the holder where the membrane is
in a vertical position. (it helps to retract the
electrode from the
electrode holder before
rinsing).
68
AMMONIA IN SOLUTIONS CONTAINING FATS, FATTY ACIDS,
OILS, AND SURFACTANTS, continued
Accuracy Check
Checking Electrode Response
Note: Hach standard To verify electrode response, measure the electrode potential (in
concentrations are given mV) of two Ammonia nitrogen standard solutions, one decade
as elemental nitrogen (N), apart in concentration, that bracket the expected sample range.
not as NH3 or NH4+.
For example, use 10 and 100 mg/L ammonia nitrogen standards
to bracket an expected sample concentration of 30 mg/L. The two
standards should have potentials –57 ±3 mV apart at 25 °C.
69
AMMONIA IN SOLUTIONS CONTAINING FATS, FATTY ACIDS,
OILS, AND SURFACTANTS, continued
REQUIRED REAGENTS
Quantity Required
Description Per Test Unit Cat. No.
Ammonia Electrode Filling Solution................................. 2 drops ........ 59 mL.......... 44472-26
Ammonia Electrode Storage Solution ................................20 mL....... 500 mL .......... 25412-49
Ammonia Ionic Strength Adjustor Pillows......................... varies ...... 100/pkg .......... 44471-69
Ammonia (Nitrogen) Standard Solutions:
1 mg/L NH3-N..............................................................20 mL....... 500 mL ............ 1891-49
10 mg/L NH3-N............................................................20 mL....... 500 mL .............. 153-49
100 mg/L NH3-N..........................................................20 mL....... 500 mL ......... 24065-49
Water, deionized.................................................................100 mL............. 4 L.............. 272-56
REQUIRED APPARATUS
Air Gap Accessory Kit............................................................ 1 .................each .......... 50253-00
Ammonia Electrode, BNC...................................................... 1 .................each .......... 50250-00
Bottle, wash, 500 mL ..............................................................1 .................each .............. 620-11
sension™4 Laboratory pH/ISE Meter .................................... 1 .................each .......... 51775-00
Select one based on available voltage:
Stirrer, electromagnetic, 115 V, with stand and stir bar.......... 1 .................each .......... 45300-01
Stirrer, electromagnetic, 230 V, with stand and stir bar.......... 1 .................each .......... 45300-02
70
SECTION 3 ANALYTICAL METHODS (mV meters)
71
72
DIRECT MEASUREMENT METHOD (with Graph Paper)
For meters that measure pH and millivolts, this method creates a
calibration curve by measuring the electrode potential of the
sample, then comparing this potential with potentials of
standards. The electrode potentials vs. concentrations are plotted
on a graph to make a calibration curve (see Figure 3). A best-fit
line is drawn through the points on the graph. Some ion or ISE
meters provide direct reading of the ammonia concentration. This
eliminates the need to plot a calibration curve unless the user
wants to verify the calibration curve.
Note: Most direct reading meters can only display results in the linear
part of the calibration curve. The advanced sension™2 and 4
pH/ISE meters can perform nonlinear calibrations, accurately
tracking electrode potentials in the nonlinear response region. This
eliminates manual plotting of calibration curves.
73
DIRECT MEASUREMENT METHOD (with Graph Paper), continued
To prepare standards:
74
DIRECT MEASUREMENT METHOD (with Graph Paper), continued
2. Fill to the mark with ammonia-free deionized water. Mix
well. This is a ten-fold dilution of the original standard
(e.g., 10 mg/L).
75
DIRECT MEASUREMENT METHOD (with Graph Paper), continued
Procedure for Analysis Using a Manually Plotted Calibration Curve
1. Place the assembled 2. Select standards that 3. Pour the least 4. Add a clean
electrode in Ammonia cover the concentration concentrated standard magnetic stir bar. Place
Electrode Storage range of the samples. into a 50-mL plastic the beaker on a magnetic
Solution. Measure 25 mL of a beaker. stirrer and stir at a
standard solution in a moderate rate.
Note: Before removing the
electrode from the storage graduated cylinder.
solution, the reading on
the meter should be below
0.1 mg/L. This insures the
electrode is properly
conditioned. After every
hour of continuous use,
place the electrode in the
storage solution for 10
minutes to thoroughly
recondition. Check with a
10 mg/L NH3-N standard
for accuracy and
recalibrate if necessary.
5. Place the electrode 6. Add the contents of 7. Prepare the next 8. Remove the
in the standard. one Ammonia ISA standard as in steps electrode from the
Powder Pillow, or add 2 and 3. previous standard. Place
0.10 g of ISA powder to it in Ammonia Electrode
the beaker. Record the Storage Solution for one
potential when it is minute.
stable.
76
DIRECT MEASUREMENT METHOD (with Graph Paper), continued
9. Rinse with distilled 10. Add the same 11. Repeat the above 12. On a sheet of
water and blot dry with a amount of ISA as was steps for all standards semilogarithmic graph
paper towel. Immerse added in step 6. Record used in making the paper (2-7 cycles), plot
the electrode in the new the potential when the calibration curve. the ammonia
standard. reading is stable. concentration values vs.
the potentials. Plot
concentrations on the
logarithmic (vertical)
axis. Each cycle is equal
to a 10-fold change in
concentration. Plot the
mV potential on the
linear (horizontal) axis.
See Figure 5 on page 80.
Note: Concentration units
may be ppm, mg/L,
moles/L or other
concentration units.
77
DIRECT MEASUREMENT METHOD (with Graph Paper), continued
Measuring Samples
1. Place the electrode 2. Add a magnetic stir 3. Remove the 4. Add the contents of
into Ammonia Storage bar to the beaker. Place electrode from the one Ammonia ISA
Solution. Measure the beaker on a magnetic Ammonia Storage Powder Pillow or 0.10 g
25 mL of sample with a stirrer and stir at a Solution. Rinse with of ISA powder to the
graduated cylinder. Pour moderate rate. deionized water or a beaker with the sample.
it into a 50-mL plastic portion of the next
beaker. sample. Blot dry with a
paper towel (do not rub).
78
DIRECT MEASUREMENT METHOD (with Graph Paper), continued
79
DIRECT MEASUREMENT METHOD (with Graph Paper), continued
Example
A water sample, measured using the above procedure, had a
potential of -108 mV. What is the ammonia concentration of
the sample?
80
DIRECT MEASUREMENT METHOD (with Graph Paper), continued
REQUIRED REAGENTS AND APPARATUS
Quantity Required
Description Per Test Unit Cat. No.
Ammonia Electrode, BNC ..................................................... 1................. each ..........50250-00
Ammonia Electrode Filling Solution, 60 mL......................varies............5/pkg ..........50287-05
Ammonia Electrode Storage Solution.................................varies....... 500 mL ..........25412-49
Ammonia Ionic Strength Adjustor Powder Pillows ............... 3............100/pkg ..........44471-69
Ammonia (Nitrogen) Standard Solutions:
10 mg/L NH3-N ......................................................... 20 mL ....... 500 mL ..............153-49
100 mg/L NH3-N ....................................................... 20 mL ....... 500 mL ..........24065-49
Beaker, 50 mL, polypropylene ............................................... 1................. each ............1080-41
Bottle, wash, 500 mL ............................................................. 1................. each ..............620-11
Graph Paper, semilogarithmic ................................................ 1............100/pkg ..........23062-01
sension™1 Portable pH Meter.............................................. 1................. each ..........51700-00
sension™3 Laboratory pH Meter ......................................... 1................. each ..........51750-00
Stir Bar, 22.2 x 4.76 mm ........................................................ 1................. each ..........45315-00
Water, deionized .................................................................varies............... 4 L ..............272-56
Select one based on available voltage:
Stirrer, electromagnetic, 115 V, with stand and stir bar ......... 1................. each ..........45300-01
Stirrer, electromagnetic, 230 V, with stand and stir bar ......... 1................. each ..........45300-02
81
82
ANALYSIS WITH a pH/mV METER AND CALCULATOR
Also known as a linear calibration using two standards that
bracket the sample, this method is often used to calibrate for
direct measurements. The two-point calibration can be repeated
quickly and easily during each series of measurements. Figure 5
on page 80 is an example of a linear calibration curve.
Note: Use the two-point The two standards chosen should be close to the expected
calibration only in the linear concentration range of the samples. The standards often differ by
region. See Measuring only one decade.
Hints.
83
ANALYSIS WITH a pH/mV METER AND CALCULATOR, continued
Repeat steps
1–4
E1 and E2 = potentials of
standards 1 and 2.
9. Add 0.5 mL of 10. Remove electrode 11. Place the electrode 12. Calculate the
Ammonia ISA Solution from Storage Solution. into the sample (stirring sample ammonia
or 0.5 mL of Ammonia Rinse and blot the increases accuracy and concentration using the
ISA Solution. Stir well. electrode. response time). Record equations following
the mV value when the this step.
display is stable.
84
ANALYSIS WITH a pH/mV METER AND CALCULATOR, continued
Calculating the Ammonia in Samples
∆E = Ex – E 1
Where:
Ex = sample potential
Where:
Cx = sample concentration
Example:
Standard 1 (C1) = 10.0 mg/L
Accuracy Check
To verify measurement accuracy, see Accuracy Check on page 23
in the Nitrogen-Ammonia in Water procedure (Method 10000).
85
ANALYSIS WITH a pH/mV METER AND CALCULATOR, continued
REQUIRED REAGENTS AND APPARATUS
Quantity Required
Description Per Test Unit Cat. No.
Ammonia Electrode, BNC...................................................... 1 .................each .......... 50250-00
Ammonia Electrode Filling Solution, 60 mL ..................... drops............ 5/pkg .......... 50287-05
Ammonia Electrode Storage Solution ................................20 mL....... 500 mL .......... 25412-49
Ammonia Ionic Strength Adjustor Powder Pillows................ 3 .......... .100/pkg .......... 44471-69
Ammonia (Nitrogen) Standard Solutions:
10 mg/L NH3-N ..........................................................20 mL........500 mL.............. 153-49
100 mg/L NH3-N ........................................................20 mL........500 mL.......... 24065-49
Beaker, 50 mL, polypropylene................................................ 1 .................each ............ 1080-41
Bottle, wash, 500 mL ..............................................................1 .................each .............. 620-11
sension™1 Portable pH Meter .............................................. 1 .................each .......... 51700-00
sension™3 Laboratory pH Meter.......................................... 1 .................each .......... 51750-00
Stir Bar, 22.2 x 4.76 mm......................................................... 1 .................each .......... 45315-00
Water, deionized ................................................................. varies ...............4 L .............. 272-56
Select one based on available voltage:
Stirrer, electromagnetic, 115 V, with stand and stir bar.......... 1 .................each .......... 45300-01
Stirrer, electromagnetic, 230 V, with stand and stir bar.......... 1 .................each .......... 45300-02
86
Method 10007
MEASURING LOW LEVELS OF AMMONIA
For samples below 0.5 mg/L NH3-N
The electrode response is nonlinear below 0.5 mg/L NH3-N
(3.6 X 10-5 M). Low ammonia concentrations in the nonlinear
range can be determined by plotting a calibration curve, or by
using the sension™2 or sension™4 pH/ISE Meter. Use several
standards per decade to increase accuracy. Repeat or check the
calibration more often in this less stable, less reproducible region.
Use one of the calibration standards as a frequent check on the
stability of the calibration.
The method below measures all the standards in one beaker, with
spikes of ammonia standard added to incrementally raise the
ammonia level. Response time will be slightly longer in the
low-level range. If using a Hach sension4 meter, set the stability
criteria to 0.1 mV/min. Stirring at a constant rate is important and
using a chart recorder helps visualize the change in the potential.
87
MEASURING LOW LEVELS OF AMMONIA, continued
PROCEDURE
88
MEASURING LOW LEVELS OF AMMONIA, continued
89
MEASURING LOW LEVELS OF AMMONIA, continued
Sample Analysis
1. To determine the 2. Add at least 50 mL 3. Place stir bar in the 4. Place the electrode
sample concentration, of sample to a 100-mL sample beaker and stir at in the sample. Record
rinse the electrode and beaker. Add the a moderate rate on an the potential when it is
blot it dry. appropriate amount of electromagnetic stirrer. stable. Read the
ISA for the sample concentration from the
volume (one ISA ISE meter or the
Powder Pillow for every calibration curve.
50 mL or the amount
used in step 2 to adjust
for the sample pH).
90
KNOWN ADDITION METHOD WITH A pH/mV METER
This method may be used when only a few samples need analysis.
It does not require a calibration curve and is a quick way to obtain
results on a few samples. This method produces the best results
when the ammonia level is doubled by the standard addition. This
requires knowing the approximate concentration of ammonia in
the sample, either from prior knowledge with the sample type or
estimating it from the potential using direct measurement (a
calibration curve is necessary).
1. Place the electrode 2. Using a volumetric 3. Add the contents of 4. Place the electrode
in Ammonia Electrode pipet, measure 50 mL of two Ammonia ISA into the sample.
Storage Solution. sample into a beaker. Powder Pillows (or 0.20
Note: Before removing the
g of Ammonia ISA) to
electrode from the storage the beaker. Add a stir bar
solution, the reading on and place the beaker on
the meter should be below a magnetic stirrer. Stir
0.1 mg/L. This insures the moderately.
electrode is properly
conditioned. After every
hour of continuous use,
place the electrode in the
storage solution for 10
minutes to thoroughly
recondition. Check with a
10 mg/L NH3-N standard
for accuracy and
recalibrate if necessary.
91
KNOWN ADDITION METHOD, continued
5. Record the potential 6. Estimate the sample 7. Calculate the volume 8. Add the desired
when it is stable. ammonia concentration of ammonia standard to volume of standard into
using a calibration curve be added to roughly the sample.
or from previous double the ammonia
experience with the concentration:
same sample type. Vstd = Cs x Vs ÷ Χstd
Where:
Vs = sample volume
Cs = sample
concentration
92
KNOWN ADDITON METHOD, continued
∆E = E 2 – E 1
Where:
E1 = the sample potential
E2 = the potential after the spike
V std × C std
Sample ammonia = ------------------------------------------------------------------------------------------
∆E ⁄ s
( V sample + V std ) × 10 – V sample
Example
The sample potential is -53.9 mV. After adding 0.20 mL of a
1000-mg/L standard, the potential measures -72.2 mV.
E1 = -53.9 mV
E2 = -72.2 mV
∆E = -72.2 -(-53.9) = -18.3 mV
Vstd = 0.20 mL
Cstd = 100 mL
Vsample = 50 mL
s = -58.6 mV/decade (experimentally determined)
0.20 × 1000
C NH3-N = ------------------------------------------------------------------------------ = 3.8 mg/L
– 18.3 ⁄ – 58.6
( 50 + 0.2 ) × 10 – 50
93
KNOWN ADDITION METHOD, continued
REQUIRED REAGENTS
Quantity Required
Description Per Test Units Cat. No.
Ammonia Electrode Filling Solution, 5/pkg....................... varies ......... 59 mL.......... 50287-05
Ammonia Electrode Storage Solution ................................20 mL....... 500 mL .......... 25412-49
Ammonia Ionic Strength Adjustor Powder Pillows................ 2 .......... 100/pkg .......... 44471-69
Ammonia Standard Solution, 100 mg/L NH3-N ................ varies ........500 mL.......... 24065-49
Water, deionized.................................................................. varies .............. 4 L.............. 272-56
REQUIRED APPARATUS
Ammonia Electrode, BNC...................................................... 1 .................each .......... 50250-00
Beaker, 50 mL, polypropylene................................................ 1 .................each ............ 1080-41
Bottle, wash, 500 mL ..............................................................1 .................each .............. 620-11
Pipet, volumetric, 50 mL, Class A.......................................... 1 .................each .......... 14515-41
Pipet, TenSette®, 0.1 to 1.0 mL .............................................. 1 .................each .......... 19700-01
Pipet Tips, for 19700 TenSette® Pipet ............................... varies ......... 50 pkg .......... 21856-96
sension™2 Portable pH/ISE Meter........................................ 1 .................each .......... 51725-00
sension™4 Laboratory pH/ISE Meter ...................................1 .................each .......... 51775-00
Stir Bar, 22.2 x 4.76 mm......................................................... 1 .................each .......... 45315-00
Select one based on available voltage:
Stirrer, electromagnetic, 115 V, with stand and stir bar.......... 1 .................each .......... 45300-01
Stirrer, electromagnetic, 230 V, with stand and stir bar.......... 1 .................each .......... 45300-02
OPTIONAL APPARATUS
Electrode Washer ........................................................................................each .......... 27047-00
94
SECTION 4 ELECTRODE MAINTENANCE
If the electrode is accidently left in the air and erratic results are
observed, the space between the inside of the membrane and the
sensing element may be dry. If this occurs, withdraw the glass
electrode from the membrane by gently pulling back on the
cables. New filling solution will flow into the space.
95
96
SECTION 5 ELECTRODE CHARACTERISTICS
Where:
E = measured electrode potential
Eo = reference potential
OH- = hydroxide concentration in solution
S = electrode slope (–59.2 mV/decade)
98
SECTION 5, continued
5.3 Reproducibility
Reproducibility is limited by factors such as temperature
fluctuations, drift, and noise. Within the operating range of the
electrode, reproducibility is independent of concentration. With
calibration every hour, measurements to ±2% can be obtained.
0 -54.20
5 -55.20
10 -56.18
15 -57.17
20 -58.16
25 -59.16
30 -60.15
35 -61.14
40 -62.13
99
SECTION 5, continued
5.6 Interferences
Volatile amines interfere with electrode measurement. Most gases
do not interfere as they are converted to ionic form in basic
solutions. Ionic species cannot cross the gas-permeable
membrane and are not direct electrode interferences. However,
the level of ions in solution can change the solubility of ammonia.
Standards and samples should have about the same level of ions
and dissolved species.
The Henry's Law constant, Kh, varies both with temperature and
the level of dissolved species. For example, the constant is about
20% lower in 1 M NaCl than in distilled water.
101
102
SECTION 6 TROUBLESHOOTING
This section contains a series of decision trees based on the
symptoms that result from a problem or a combination of
problems. The symptoms are listed below along with the page
number of the decision tree that will help solve the problem(s)
that may cause the symptom.
Before using a decision tree, check the pH bulb per Section 6.1
Symptom Page
Inaccurate readings 104
Imprecise readings 104
Low slope or no slope 105
Drift in the meter reading 107
Slow response time 107
Error messages or no key response See meter manual
103
SECTION 6, continued
104
SECTION 6, continued
105
SECTION 6, continued
106
SECTION 6, continued
107
SECTION 6, continued
108
ELECTRODE SERVICE REQUEST QUESTIONNAIRE
If the electrode still malfunctions after following the
troubleshooting guide, our service representatives want to help
solve any problems as quickly as possible. To help them with this
process, please have the following information when calling:
109
110
GLOSSARY
Acid —A chemical compound that contributes hydrogen ions
(H+) to an aqueous solution or causes the pH to be less than 7.0.
111
GLOSSARY, continued
112
GENERAL INFORMATION
113
114
APPENDIX CONVERSION TABLES
115
116
HOW TO ORDER
117
REPAIR SERVICE
118
WARRANTY
119