Surgical Services
Endovascular
Services
Design Guide
April 2016 (rev 02/23)
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
TABLE OF CONTENTS
2.0 NARRATIVE
2.1. General 2-1
2.1.1. Clinical and Operational Summary 2-1
2.1.2. VA Trends in Surgical and Interventional Service 2-32
2.1.3. The Technological Future of Surgery 2-35
2.2. Technical Considerations 2-37
2.2.1. VA Policies/Directives/Handbooks, Codes, and Standards 2-37
2.2.2. Architectural 2-39
2.2.3. Interior Design 2-41
2.2.4. Heating, Ventilation and Air Conditioning Systems (HVAC) 2-42
2.2.5. Plumbing Systems 2-46
2.2.6. Medical Gas and Vacuum Systems 2-47
2.2.7. Lighting and Power Systems 2-49
2.2.8. Telecommunications and Special Telecommunication
2.2.9. Systems, Monitoring, and Signaling Systems 2-51
2.2.10. Fire Protection and Life Safety 2-55
TABLE OF CONTENTS i
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
5.0 APPENDIX
5.1. Site Tours 5-1
5.1.1. Elmhurst Memorial Hospital, Elmhurst, Illinois 5-2
5.1.2. Advocate Sherman Hospital, Elgin, Illinois 5-5
5.1.3. Center for Care and Discovery, Chicago, Illinois 5-11
5.1.4. Rush University Medical Center, Chicago, Illinois 5-21
5.1.5. Edward Hines, Jr. VA Hospital, Hines, Illinois 5-27
5.2. Hines Interventional Suite 5-32
Surgical and Endovascular Requirements for Operative and Interventional
Procedures 5-34
Surgical and Endovascular Procedure Room Checklist 5-34
TABLE OF CONTENTS ii
Surgical and Endovascular Services Design Guide April 2016 (rev 9/21)
1.0 GENERAL
1.1. Foreword
Disclaimer: Products and materials indicated are for illustrations, they are
not endorsement for any specific product. This document is intended to be
used as the standard for design, and to supplement current VA construction
standards and other VA criteria in planning surgical and endovascular space.
Designers shall use this Design Guide as closely as possible to its context and
intent. Use of this Design Guide does not preclude the need for a functional
and physical design parameter for each specific project. It is the responsibility
of the Project Architect and the Project Engineer to develop a complete and
accurate project design that best meets the users’ needs and applicable code
requirements.
GENERAL 1-1
Surgical and Endovascular Services Design Guide April 2016 (rev 4/20)
GENERAL 1-2
Surgical and Endovascular Services Design Guide April 2016 (rev 9/21)
1.2. Acknowledgements
GENERAL 1-3
Surgical and Endovascular Services Design Guide April 2016 (rev 9/21)
• Consultants
GENERAL 1-4
Surgical and Endovascular Services Design Guide April 2016 (rev 9/21)
1.3. Introduction
The Surgical and Endovascular Services Design Guide was developed as
a design tool to assist VA clinical, contracting, planning, construction and
engineering staff to better understand the design process, and to better enhance
their collaboration with the design team throughout the planning and design
phases of a specific project.
The A/E design team shall read, understand and be familiar with the most current
editions of the applicable VA and applicable industry’s codes and standards.
Codes and standards shown in this document shall be adhered to. VA design
and construction standards may be accessed at the Office of Construction and
Facilities Management’s Technical Information Library (TIL)
(www.cfm.va.gov/TIL).
Use of this Design Guide does not supersede the project A/E’s responsibilities
to develop a complete and accurate design that meets the users’ needs and
appropriate code requirements within the budget and schedule constraints. A/E
shall confirm and verify actual dimensions, weights, and utility requirements of
equipment with manufacturers.
GENERAL 1-5
Surgical and Endovascular Services Design Guide April 2016 (rev 9/21)
GENERAL 1-6
Surgical and Endovascular Services Design Guide April 2016 (rev 9/21)
1.5. Abbreviations
GENERAL 1-7
Surgical and Endovascular Services Design Guide April 2016 (rev 9/21)
GENERAL 1-8
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
2.0 NARRATIVE
2.1. General
NARRATIVE 2-1
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-2
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
The Surgical Suite: The perioperative journey, clinical staff, the patient and
support staff.
Components of Surgical Services Team: The surgical services team
consists of following components:
1. Sterile team: it includes:
• Surgeon
• First assistant (RN or PA)
• Surgical technologist (scrub)
2. Unsterile team: it includes:
• Anesthesia: MD/CRNA/PA
• Circulator
• Environmental Services
• Sterile Processing 1, 2
1 Operating room orientation manual [Internet]. (Cited 2014 Jan 5). Available from
https://www.utmb.edu/surgery/clerks/ormanual.htm
2
Osman C. Asepsis and aseptic practices in the operating room. [Internet]. [Last updated 2000
Jul 1; Cited 2014 Jan 6\. Available from
http://www.infectioncontroltoday.com/articles/2000/07/asepsis-and-aseptic-practices-in-the-
operating-ro.aspx
NARRATIVE 2-3
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Surgical Staff: The staff typically moves from the staff lockers/lounge
through the semi-restricted corridor to the Scrub Stations and then into the
individual Operating Rooms, with exit through the semi-restricted corridor.
Possible exceptions to this flow are the “circulators,” who retrieve supplies
and equipment from the clean core, and the supervising anesthesia staff,
who are permitted to move from Operating Room to Operating Room via
the clean core.
Patient: The Patient checks in and changes into a gown in the pre-
procedure area. They are then visited by members of the surgical team,
who will start an intravenous (IV) line, review the procedure and answer any
questions the patient may have. They are then brought into the operating
room on a gurney and transferred to the operating table. While the patient is
awake during this process, the anesthesiologist may have given the patient
a drug to help them relax. Once situated on the operating table, the
anesthesiologist will give the patient drugs through their IV and a breathing
mask to induce anesthesia. When the patient is asleep the surgical
procedure begins. Once the surgery is complete the patient is moved to the
post-anesthesia care unit (PACU) until the effects of the anesthesia wear
off. Outpatients will go to the Phase 2 recovery area, where a designated
family member can come and visit with the patient while they become more
alert and are discharged to go home. Inpatients will be transferred to a
patient ward until they are discharged.
Linens: Linens are brought into the Surgical Operating Rooms and
Endovascular Procedure Rooms by way of the semi-restricted corridor.
Soiled linen is bagged and removed from each Surgical Operating Room
and Endovascular Procedure Room via the semi-restricted corridor.
The “Sterile Field”
In general, the term “sterile field” is used to describe the sterile zone in the
Operating and or Endovascular Procedure Room is approximately five feet
around the perimeter of the surgical table, which includes the space
surrounding the site of the patient’s incision. The term “sterile” indicates that
no undesirable microorganisms are present. The hands of the scrub team
are gloved and everything that enters this field must be sterile. The
operating table, the surgical instrument table, and special equipment to be
used in the sterile field are cleaned prior to each case. Imaging equipment,
surgical microscopes, and other items that are difficult to clean are draped
in sterilized plastic to maintain asepsis. The sterile field is showered by
either a laminar flow array or a combination of laminar diffusers surrounded
and encapsulated by an air curtain.
NARRATIVE 2-4
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Note: The example shown is for a General OR, the sterile field supporting
Complex OR procedures may be larger as there may be more procedural
support equipment brought into the sterile field.
The anesthesiologist and nurse anesthetist remain outside of the sterile
field and are typically separated from the sterile field by a sterile drape.
Maintaining Asepsis (Sterility)
Regarding asepsis, there are three areas of concern (to minimize the
number of pathogens present) within the Surgical and Interventional Suite
are:
• The “sterile field” itself as described above. Only fully scrubbed staff
(known as the “scrub team”) is permitted in this area in the center of
each Operating and/or Endovascular Procedure Room.
• Within the Operating and/or Endovascular Procedure Room, both
the scrub team as well as the additional staff that are not in the
sterile field must abide by strict rules established by the Medical
Center. Semi-restricted areas, including spaces such as the pre-
operative and patient holding areas, PACU, instrument workroom,
non-sterile supply storage, staff lockers/lounges/toilets/showers,
control desk, and surgery administration offices.
Internal Operating Room Circulation
It is critical to plan an Operating Room in such a way that a high level of
sterile technique can be achieved.
The circulating nurse typically addressed as “the circulator” places the
packs on the instrument table from the side of the instrument table away
NARRATIVE 2-5
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
from the sterile field. The scrub nurse removes the sterile instruments and
places them on the surgical instrument table prior to the procedure.
Consideration should be given that with some specialty surgical
procedures, the surgical instrument table is positioned toward the foot of the
operating table, but always within the sterile field. A single instrument table
may be up to 8.0 feet (2440 mm) in length, or there could be more than one
instrument table. Once the procedure starts, the scrub nurse constantly
draws from this source to supply the surgeons during the operation. The
circulator also assists the surgeon with devices used in the sterile field such
as lasers.
No one walks between the operating room table and the OR instrument
table, except those in sterile garb who have thoroughly scrubbed. When
portable imaging equipment is used, space for these items must be
considered. Surgical microscopes or video monitors on carts and other
large pieces of equipment (including robotics) may also be needed in an
Operating Room, and space for them must be considered.
The Surgical Team in a teaching VAMC is typically comprised of the
following members:
Surgeon: The surgeon is a Medical Doctor who is board certified as a
Surgeon and acts as surgical team leader to direct and supervise all
aspects of a surgery. The responsible surgeon’s eligibility to perform a
surgical procedure is based upon that surgeon’s education, training,
experience, and demonstrated proficiency. Surgeons are expected to study
and evaluate new procedures and to become knowledgeable of and
proficient with advances that are appropriate. Technical skill alone is not
sufficient to qualify a surgeon to perform new procedures. Procedural skills
should be acquired within the context of in-depth knowledge about the
disease to be treated. Qualification of a surgeon as a specialist carries the
commitment and responsibility to conduct a surgical practice that conforms
to his/her defined specialty. The appropriate surgical specialty board
recognized by the American Board of Medical Specialties (ABMS) or the
Royal College of Physicians and Surgeons of Canada (RCPSC) determines
a surgeon’s scope of practice.
Assistant Surgeon: A Medical Doctor who acts as assistant surgeon and
may be one of the following medical professionals:
• Another surgeon
• A licensed physician assistant
• A registered nurse first assistant
• A surgical resident
• A medical student
NARRATIVE 2-6
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-7
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-8
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-9
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-10
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-11
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-12
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-13
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
delivering much higher x-ray. However, their use predicated that they be
fixed / dedicated to the procedure room as access to higher input power
was required to produce higher x-ray, including radiation protection of the
procedure room which is not required for portable low emission devices.
Dedicated C-arms initially became popular in the interventional cardiac and
interventional radiology arenas as a diagnostic tool to assess flow restricting
disease, map the electrical pathway of the heart, and to provide image
guidance for the execution of interventions. These early procedure rooms
were known as “Cath Labs” and did not possess the attributes of an
operating room (OR) environment; laminar flow, higher level sterile
practices, and the ability to support open surgery.
More recently, through the growth of minimally invasive surgery on patients
with risks requiring a surgical safety net including endograft placement, and
the inception of valve replacements, a specialized interventional / surgical
environment was established. The result was the cardiovascular hybrid
operating room.
When imaging equipment is used (fixed or mobile) in a traditional operating
room and/or hybrid operating room, staff are required to wear lead aprons
and work from behind leaded glass shields. When not in use in the
operating rooms, portable imaging equipment is usually stored in alcoves in
the semi- restricted corridor so it can be efficiently moved in and out of the
operating room. The level of use of mobile imaging equipment and other
factors will be analyzed by the physicist, who will determine the need for
fixed shielding in the room.
Sterile Supplies: Sterile supplies are typically retrieved from the clean core
by the circulator. A selection of sterile supplies may also be stored within
the Operating Room.
Supplies may be bar-coded or may utilize radio-frequency identification
(RFID) chips. Both systems are designed to monitor use of supplies and to
allow computerized reorder to maintain the appropriate inventory. In
addition, scanning the bar code (or tracking supplies with RFID chips)
provides a more complete system for charging for supplies used during a
surgical procedure. It also expedites the removal of stored items in the
Operating Room when their shelf life has expired.
Case Carts: Case carts are used to bring sterile materials and instruments
from SPS to the Operating Room. A typical case cart contains specific items
required for each specific case, including all required surgical instruments
and other supplies. It is advisable for the planner to confer with nursing staff
as to the size of their case carts as each hospital is different. Note that more
complex procedures may require several case carts. Clean case carts are
typically staged in the Clean Core outside of the Operating Room prior to a
NARRATIVE 2-14
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
procedure to optimize turnover time. Some case carts may remain in the
Operating Room during the procedure and be used as back tables. After
the surgical procedure is completed, these case carts are returned via the
semi-restricted corridor to the “soiled” side of SPS on another dedicated
soiled cart lift room or dedicated soiled elevator. In the event that SPS is not
located below the Surgical Suite, an alternative traffic pattern for the case
carts must be established that isolates clean and soiled case cart traffic.
Dedicated elevators are to be sized to accommodate a set number of case
carts in the elevator cab.
NARRATIVE 2-15
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-16
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
requirements, patient areas must provide acoustical and visual privacy at all
times.
The Phase II Recovery Patient Area is utilized for recovery of the patient
after Phase I Recovery (PACU) and also for ambulatory surgery patients
who may come directly from the surgical OR if they have been anesthetized
under mild MAC sedation* where recovery is fairly quick. The Prep Area
should be co-located with the Phase II Recovery Room to provide
maximum flexibility for the patient room assignment. Since most surgical
procedures start in the morning, the Prep Area can occupy under-utilized
Phase II Recovery Space in the morning however, the same area can be
utilized for Phase II recovery in the afternoon.
*MAC Sedation: Also known as monitored anesthesia care
or conscious sedation, MAC anesthesia is a type of sedation
where the patient remains aware of their surroundings and
stays calm. The anesthetist administers it through an IV into
the skin and muscle around the area on which surgery will
be performed.
Phase I (PACU) Recovery: During Phase I care, the focus is on the
patient’s recovery from anesthesia and the return to baseline vital signs.
Consideration is given to the procedure, anesthesia care, patient
comorbidities, and the patient’s physical status to recognize, minimize and
manage any issues or complications. Phase I includes, but is not limited to,
applying PACU scoring criteria on admission and each vital signs
assessment, managing respiratory and hemodynamic changes, monitoring
the effects of the procedure (e.g., bleeding, circulation), and providing
necessary analgesia and antiemetics. While monitoring requirements are
facility-specific, they are also based on the patient’s condition. According to
ASPAN, they recommend assessing and documenting vital signs at least
every 15 minutes during the first hour and then every 30 minutes until
discharge from Phase I PACU care. After which, the patient is then
transitioned to Phase II, the inpatient setting, or the intensive care unit (ICU)
for continued care.
Patient Induction/Preparation Room: For most procedures, anesthesia is
given in the induction room located next to the operating room. In many
cases induction may be performed in the pre-operative Phase II area.
“Induction” is the start of anesthesia and usually takes less than five (5)
minutes to perform. The patient induction room is used by the
anesthesiologist to initiate anesthesia and to observe the patient prior to the
start of a procedure. It allows for parallel processing of the patient and the
Operating Room. Typical tasks in this area include nerve blocks and the
placement of intravenous (IV) lines. From here the patient is transferred to
the Operating Room.
NARRATIVE 2-17
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Corridors:
In general, corridors utilized for transporting patients as well as surgical
support equipment, (x-ray, mobile c-arms and the like) shall be minimum 8’-
0” in clear unobstructed width. Staff “back-of-house” corridors can be
minimum 3’-6” clear unobstructed width, though 5’-0” clear is
recommended.
Safe Patient Handling in Perioperative and Procedural Settings: VHA
Directive 1611 section 4. h. (12) requires: that planning, design of phases of
new and renovation construction, which includes major, minor, NRM, and
station-level equipment projects, must incorporate appropriate and
necessary safe patient handling and mobility equipment at all facilities. In
general, perioperative and procedural settings must be outfitted with ceiling
mounted patient lifting devices capable of safely handling 1,000 lb (454 kg)
loads.
Operating Room Types:
There are several types of operating rooms depending on the type of
surgery being performed. i.e. General, Orthopedic, Neuro, Spine, Urology,
Transplant. Cardiac/Thoracic and special operating rooms such as Hybrid
OR’s.
The purpose of having different types of operating rooms is that different
procedures require different equipment and personnel. The staff must have
access and enough room to move and navigate the equipment required for
each type of surgery.
1. General Operating Room: General surgery operating room is where
surgeons focus typically on numerous abdominal contents including
esophagus, stomach, small intestine, large intestine, liver, pancreas,
gallbladder, appendix, bile ducts, and often the thyroid gland (depending
on local referral patterns). Additionally, general surgery operating rooms
deal with diseases involving the skin, breast, soft tissue, trauma,
peripheral vascular disease including hernias and perform endoscopic
procedures such as gastroscopy and colonoscopy.
2. Orthopedic Operating Room: Orthopedic surgery operating room is
where surgeons perform numerous types of orthopedic surgeries on
patients. Common surgeries may include:
• Arthroscopy - a procedure using special cameras and equipment
to visualize, diagnose, and treat problems inside a joint.
• Fusion – a “welding” process by which bones are fused together
with bone grafts and internal devices – such as metal rods – to heal
into a single solid bone.
NARRATIVE 2-18
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-19
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-20
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-21
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
3Motor evoked potential monitoring for spinal cord and brain stem surgery, by Sala F, Lanter P., Bricolo
A. Advanced Tech Stand. Neurosurg 2004, 29:133-69
NARRATIVE 2-22
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Minimum clearances of the room as well as the size of the no-fly zone at
the ceiling plane will vary by model/vendor of the imaging equipment.
VHA experience demonstrates that 28 ft. (8534 mm) is the minimum
clear acceptable dimension that should be planned in the room. The
system component room will be accessible from the semi-restricted
corridor or the control room. The imaging equipment may also have a
maximum distance requirement from the system to the control module
as well as required service clearances.
The universal aspect of the hybrid operating room utilizes either a fixed
sectional Trendelenburg radiolucent (carbon fiber) surgical table or a
fixed “slab type” that is utilized for cardiac catheterization and
electrophysiology procedure rooms. The Trendelenburg surgical table
can be positioned to accommodate the type position required for the
specific surgical procedure being performed.
The perioperative team in a Hybrid OR consists of the following team
members:
• Surgical circulating nurse
• Radiology circulating nurse
• Surgical scrub person
• Radiology scrub person
• Surgeon
• Surgical first assistant
• Anesthesia care provider(s)
• Radiology technician
• Interventional radiologist
• Interventional cardiologist
• Perfusionist
There may be additional personnel present depending on the type of
procedure.
The orientation of the operating room table to the Control Room window
may be determined by the head interventionalist, who may prefer a side
(lateral) view or a toe view of the patient. However, most hybrid
operating rooms and endovascular procedure rooms are visualized from
the control room via “toe-view” though “lateral or side view” is also
utilized. The head interventionalist usually determines which view should
be utilized though room configuration and geometry may determine the
optimum view from the control room.
Contrast injector may be accomplished by three methods: 1) ceiling
mounted; 2) mobile floor unit; or 3) an OR table mounted. The
preference of contrast injector types is up to the surgical/ interventional
staff.
NARRATIVE 2-23
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Control Room:
• The Control Room is directly accessible from the Hybrid OR as well
as from the semi-restricted corridor. The counter at the view window
is minimum 2 ft. 6 in. - 762 mm (max 3’-0” - 914 mm) depth to
facilitate control modules associated with the imaging equipment as
well as charting workstations. Minimum depth of Control Room shall
not be less than 9’-0” - 2,438 mm. Width of Control Room shall
match that of Procedure Room.
• When planning the Control Room it is strongly recommended to
avoid combining endovascular procedure room control rooms!
Control rooms should be internal to the endovascular procedure
rooms (cardiac catheterization lab, electrophysiology lab,
interventional radiology lab, vascular lab and hybrid OR’s) and
separate from the outside corridor. The best and preferred solution
is to have one control room per each endovascular procedure room.
This facilitates communication between control room staff and those
in the procedure room and is not significantly more expensive than
having one control room for two endovascular rooms. The one-to-
one ratio reduces potential miscommunication since only one
patient is being discussed by one set of personnel. At the same
time, the overhearing of patient information is also reduced. A few
well-thought-out steps can help minimize endovascular procedure
room design challenges.
• The equipment for the Hybrid OR all Endovascular Procedure
Rooms may have additional requirements for the placement of the
control modules. The Medical Planner shall take into consideration
clinical staff requirements when laying out the equipment during the
planning of the room.
• Toe view verse’s lateral view: Both are acceptable, though the toe
view is most common. Lateral views typically are the result of
geometry where implementing a toe view cannot be accomplished.
It is important to understand that the clinical team always presents
to the patient the same way regardless of the view from the control
room. Consider utilizing multiple cameras in OR’s and Endovascular
Procedure Rooms so that the Control Room clinical personnel have
a clear view of the sterile field and procedural operations to
accurately document.
NARRATIVE 2-24
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-25
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-26
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-27
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Important issues for the vascular hybrid operating room include quality of
the imaging equipment, radiation burden, ease of use of the equipment,
need for specially trained personnel, ergonomics, ability to perform both
open and percutaneous procedures, sterile environments, as well as quality
and efficiency of patient care. The most important feature of working in a
dedicated hybrid vascular suite should be the ability to attain best treatment
of vascular patients. Whether the interventional radiologist or the vascular
surgeon uses the facilities is of less importance.
Establishment of an endovascular operating room suite has the benefit of a
sterile environment, and the possibility of performing hybrid procedures and
conversions when necessary. Moreover, angiography immediately before
treatment gives contemporary anatomical information, and after treatment
provides quality control. Consequently, better quality and service can be
provided to the individual patient. These changes in the treatment of
vascular disease require that a new type of vascular specialist, named
‘vascular hybrid surgeon’, trained to perform both endovascular and open
surgical procedures in this highly complex patient group. 44
Interventional Radiology Procedure Room: Interventional radiology (IR)
procedure room utilizes a set of techniques that allows access to the
internal structures of the body through body orifices or very small incisions
and guidance with medical imaging. Regardless of the reason for the
intervention, the procedure will likely make use of common elements such
as a puncture needle (to pass through the skin), guidewires (to guide
through structures such as blood vessels or the biliary or urinary systems),
a sheath (which slides over the guidewire and hold the path open without
injuring it), and catheters (that allow fluids to be pushed through them). 55
Interventional radiology is typically utilized for both diagnostic and
therapeutic interventional procedures and reflects the same organizational
features as the cardiac catheterization procedure room.
Configuring Endovascular Procedure Rooms:
While endovascular procedure rooms may be adjacent, it is not advisable to
have a single control room that supports multiple procedure rooms, though
the imaging equipment component room for two procedures may be co-
located within a single room.
4 The vascular hybrid room-operating room of the future, by Hudorovic-N, Rogan SA, Lovricevic I, Zovak
M, Schmidt S., Acta Clin Croat. 2010 Sep, 49(3):289-98
5 Taslakian, Befros, Ingber, Ross, Aaltonen, Eric, Horn, Jeremy, Hickey, Ryan (2019-08- 30).
“Interventional Radiology Suite: A Primer for Trainees”. Journal of Clinical Medicine 8 (9).
doi:10l3390/jcm8091347. ISSN 2077-0383, PMC 6780384. PMID 31480308
NARRATIVE 2-28
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Example follows:
Note that this configuration also applies to multiple “Universal” Hybrid OR’s
as well as multiple endovascular procedure rooms.
Work Flow
Not every patient will follow the same workflow protocol; therefore, several
different scenarios have been diagrammed to show possible routes that a
patient may take through the surgical department. It is important to note that
the major differences stem from the level of acuity of the procedure being
performed: in the case of ambulatory surgery the patient will arrive on the
day of surgery and be discharged on that same day without interaction with
any spaces outside of the surgery department. During an in-house or same
day admit surgery, intake and discharge of the patient will occur through
other hospital departments. In complex surgical cases, the patient will likely
not spend any time in pre and post procedural areas but be transferred
directly to the surgical intensive care unit. This serves to minimize patient
transfers and ultimately reduce patient risk.
Functional Adjacencies
The Surgical Suite is situated to prevent non-related traffic through the
department. It is divided into three distinct areas, which are defined by the
physical activities occurring within the area and staged in a progressive
manner to decrease the potential of cross contamination. These areas are
the unrestricted area, the semi-restricted area and the restricted area.
The unrestricted area includes the central control point that is established to
monitor the movement of patients, staff, and equipment. The semi-restricted
NARRATIVE 2-29
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
area (“red line”) includes the peripheral support areas of the surgery suite,
such as storage areas for clean and sterile supplies, instrument processing
areas, scrub sink alcoves, and the corridors leading to the restricted area.
Access to this area is restricted and appropriate surgical attire as well as
coverings for head/ facial hair is required. The restricted area includes the
operating rooms as well as the clean core. Access to this area also requires
full surgical scrub attire, foot coverings, head/hair covering, masks, surgical
gloves. All other areas in the surgery suite are non-restricted with no
specific requirements for attire and covers.
Clean Core
The surgical suite is set up to carefully orchestrate the flow of sterile and
soiled goods. A one-way flow of soiled goods out of the operating room and
sterile goods into the operating room on a separate, dedicated route is
preferred. This is best accomplished by grouping operating rooms around
the Clean Core creating a semi-restricted corridor “race-track” configuration.
The clean core should have a minimum width of 21 ft. (2438 mm). This
width will allow for 10 ft. (3048 mm) of supply storage in the middle of the
space with a 3 ft. (914 mm) wide aisle on each side as well as a 2 ft. 6 in.
(762 mm) space for case carts and other equipment, such as refrigerators
and blanket warmers.
When the provision of substantial numbers of storage cabinets in each
Operating Room is the desire of the local staff, then the space in the clean
core reserved for the storage of sterile supplies can be reduced. Case carts
are held in a clean staging area until required at the start of the surgical
procedure. Many facilities also utilize automated inventory supply units,
which store and electronically track usage of supplies.
Example follows:
NARRATIVE 2-30
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Sterile supplies are transported to the clean core via dedicated elevators
from the clean side of SPS. An optional dedicated stairway may be included
if SPS is vertically separated from the clean core by no more than two
floors.
Depending on the available space and the type of facility different clean
core arrangements can be employed. Generally, it is recommended to keep
the quantity of operating rooms around a clean core to 8 or less in number
to avoid excessive travel distances. In renovation situations where space is
not available to utilize a Clean Core Outer Racetrack concept, one corridor
outside the Operating Room may be considered a clean space where
sterile supplies are stored. Appropriate staff and materials flow must be
followed to maintain the separation of clean and soiled traffic. However, this
arrangement should only be considered for a small complement of
Operating Rooms.
Appropriate staff and materials flow must be followed to maintain the
separation of clean and soiled traffic. However, this arrangement should
only be considered for a small complement of Operating Rooms.
Waste Management
Medical Waste: Medical waste is generated in medical exam rooms or in
patient rooms where it is bagged, collected and transported using specially
designated, closed containers to the soiled utility rooms. The waste is held
NARRATIVE 2-31
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
there until it is transported via the loading dock to the medical waste
handling facility.
General Waste: General waste is generated in all spaces and is held in
waste containers for collection either in a trash receptacle within the patient
room or bulk storage in the soiled utility closet. It is then collected by cart
and transported via the loading dock to the waste handling facility.
Recycling: Means of sorting, collecting, transporting and disposing of
recyclable material should be analyzed by locality and modified to suit local
conditions and practices.
Product types used in the building: disposable vs. recycle products should
be discussed as it is. An important design consideration in alternatives that
impacts physical space for waste disposal volumes.
Refer to the NEPA Interim Guidance for Projects at
https://www.cfm.va.gov/environmental/NEPAGuidance.pdf for more
information regarding recycling requirements.
Soiled Linen: Reusable soiled linens are generated in patient rooms,
physical therapy gyms, and sometimes the medical exam rooms. They
should be collected in carts or hampers in the soiled utility rooms and
transported to a soiled linen holding room near the loading dock for pick-up.
Medical exam rooms may opt to use disposable paper products in which
case, they would be discarded after each use.
Utensils: Any medical care washable items incorporated in the plan of care
should be transported to the soiled utility room for holding. Items will then
be transported to a sterile processing department or service for cleaning
and reprocessing.
NARRATIVE 2-32
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-33
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-34
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-35
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-36
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-37
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-38
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
2.2.2. Architectural
Corridors: Shall be designed to a minimum of 8 feet clear width to
accommodate passage of equipment or beds and two stretchers and/ or
wheelchairs within the surgical department. In non-patient areas and
outpatient clinical spaces where gurneys are not utilized, corridors may be a
minimum of 5 feet in clear width.
General: Refer to VA PG-14 Room Finishes, Door, and Hardware Schedule
for requirements for ceiling materials in the restricted and semi-restricted
areas of the surgical suite.
Ceilings:
• The finished ceiling height for Surgical Operating Rooms should be a
minimum of 10 ft. (3050 mm) above the finished floor and shall be
monolithic (gypsum board) and scrubbable. The use or consideration
for lay-in ceiling panels, boards, and/or tiles are not acceptable under
any circumstances.
• The finished ceiling height for Endovascular Procedure Rooms
(hybrid OR’s, Catheterization Labs, Interventional Radiology Labs,
Electrophysiology Labs and Vascular OR’s should be a minimum of
10 ft. (3050 mm) above the finished floor and shall be monolithic
(gypsum board) and scrubbable. The use or consideration for lay-in
ceiling panels, boards, and/or tiles are not acceptable under any
circumstances.
• The finished ceiling above scrub station alcoves shall be monolithic
(gypsum board) and scrubbable. The use or consideration for lay-in
ceiling panels, boards, and/or tiles are not acceptable under any
circumstances.
• The finished ceiling height for the clean core should be a minimum of
9 ft. (2743.2 mm) above the finished floor and shall be monolithic
(gypsum board) and scrubbable. The use or consideration for lay-in
ceiling panels, boards, and/or tiles are not acceptable under any
circumstances.
Floors: Cleansability of the flooring material is of primary importance in the
surgical suite, as most areas are at high risk for the spread of infections.
Consider the following when choosing a flooring material:
NARRATIVE 2-39
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
• Ease of maintenance
• Readily cleanable
• Impact by germicidal cleaning solutions (if any)
Flooring in operating rooms shall be resinous poured flooring with integral
base seamless and of non-porous material. Special consideration should be
given to the area directly below the surgical table, where betadine staining
is a common problem. Betadine is an antiseptic frequently used in surgical
procedures that leaves a yellowish-brown stain. While some flooring
materials will hold up better than others to staining, it is generally
recommended to provide colors that will disguise betadine build-up in this
area. Also, this area may be of a different color than the rest of the flooring
in the room to demarcate the extent of the air curtain at the ceiling above
the surgical table.
Walls: Due to the large amount of cart and stretcher traffic in the surgical
and endovascular suites, consideration shall be given to the durability of
walls. It is recommended that walls are fitted with wainscot wall protection
extending at least 4’-0” above the finished floor. If budget allowable,
consideration may be given to stainless steel wall cladding.
Door swings: All doors allowing entry into the surgical OR and/ or
endovascular procedure room, including doors from the Control Room to
Procedure Room as well as doors from the Clean Core, shall swing into the
surgical OR and/or endovascular procedure room to maintain positive
pressure.
Casework: Modular moveable casework storage systems should be
considered for flexibility including the incorporation of typical dimensions for
ease of multiple re-use applications, and ease in terminal cleaning of OR
and endovascular procedure room floors. Casework systems should be
integrated with space planning to avoid corner installations and filler panels.
Countertops for all clinical and clinical support areas should be made of
solid impervious resin material (per PG 18-14: Room Finishes, Door &
Hardware Schedule) with integral sinks, which offers long-term durability,
and resists chipping and staining from medical agents expected to be used
in clinical environments. For areas where strong chemicals are used, such
as soiled utility rooms, seamless stainless steel counters with integral
backsplash should be used. Plastic laminate veneer materials may be used
in non- clinical staff and administrative areas.
Acoustics/Noise Control: Interior acoustics that support speech
intelligibility and provide comfort can be difficult to obtain in an operating
room where non-porous materials are mandated for infection control
requirements. It is important to find ways to control reverberation and noise
build-up in these spaces. The acoustical design of patient spaces shall also
NARRATIVE 2-40
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-41
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-42
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Interior Design Conditions: Interior design conditions for each space shall
be maintained throughout the year. Interior design conditions for all spaces
shall be maintained in accordance with the current version of the VA HVAC
Design Manual.
Supply Air Requirements:
The supply air volume shall be established to meet the heating and cooling
load requirements of the occupied space. The supply volume shall,
however, be modified to meet the following:
a) minimum air change requirements if this air quantity is more than
the heating and cooling load requirements,
b) maintain proper space pressurization relative to room exhaust
requirements.
For all air systems the supply air minimum airflows shall follow the current
version of the VA HVAC Design Manual.
The current version of the VA HVAC Design Manual describes an air supply
distribution system for the general operating rooms. The Manual indicates
the minimum requirements for total air changes per hour (ACH) of supply air
to be provided at the point of delivery to the room which includes the
required minimum ACH of outside air. The supply air system provided is an
air curtain system. The system is comprised of two different types of ceiling
NARRATIVE 2-43
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
diffusers located within the operating room. The area above the patient is
provided with a laminar diffuser array sized in accordance with the current
version of ASHRAE Standard 170 and provides a slow velocity downward
wash of clean supply air over the operating table surgical zone. Combined
with this diffuser array is a four-sided linear slot diffuser system using
specially designed linear slot diffusers that surrounds the operating table
surgical zone. Per the current version of the VA HVAC Design Manual, the
linear diffusers discharge a vertical airstream inclined at an outward angle
toward the sides of the operating room walls. The purpose of the linear
diffusers is to create a high velocity barrier of clean air to keep room
contaminant particles which may be recirculating from the perimeter of the
room away from the surgical zone. The linear slot diffuser curtain should be
sized to deliver the air flow rate per foot per manufacturer
recommendations. Flow rates below this quantity may not properly isolate
the surgical air zone from contaminants, while air flows above this quantity
may increase the possibility of re-entraining particles that have settled on
the floor. The percentage of total room supply air to be provided by the slot
diffusers and the unidirectional perforated diffusers is referenced in the VA
HVAC Design Manual.
Manifold-Unidirectional Air Delivery: Current and future surgical OR’s as
well as endovascular procedure rooms are utilizing fully integrated
unidirectional air flow technology (manifold system) which affords many
benefits such as: supporting ceiling suspended equipment consisting of
surgical lights, equipment booms, monitors as well as supporting ceiling
mounted imaging equipment gantries. The manifold system provides a level
stable platform which inhibits the common issue of “drift” found in
independent “red iron” supports. Another advantage of the “manifold” is the
speed to install during construction
Unlike the general operating rooms, the hybrid operating rooms utilize a
single large array of laminar flow diffusers over and beyond the sterile field
which provides 100% of the required room supply air. The laminar flow
diffuser array is located over the operating table and surrounding area to
create a large surgical zone providing a slow-moving blanket of clean air
over the patient and surrounding area. The operating rooms are the most
positively pressurized spaces in the surgery suite.
Surgical Suite Air Handling Units: Air handling units serving the surgical
suite shall be provided with emergency power per requirements of the
current version of the VA HVAC Design Manual. Air handling units shall
have the capacity for utilizing 100% outside air where required by building
code to meet purge requirements.
NARRATIVE 2-44
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
Filtration: Filtration for the Surgical Suite HVAC systems shall be provided
in conformance with the current version of the VA HVAC Design Manual.
Operating rooms and surgery support spaces shall follow the current
version of the VA HVAC Design Manual unless noted otherwise. The room
data sheets provided in the Design Standard Templates chapter of this
document refers the user to specific room types shown in the VA HVAC
Design Manual unless noted otherwise. The room type referenced in the
HVAC Design Manual is found in a table that will provide the needed
information for room design temperatures, relative humidity range, room air
changes, and space pressure relationships.
It is highly desirable to identify the supply air zone described above by
installing an area of flooring material that is a different color from the
balance of the room. This area should be located in the Operating Rooms
by aligning it with the slot diffusers described above, and for the Hybrid
Operating Rooms and Endovascular Procedure Rooms by aligning with the
perimeter of the laminar flow diffuser array. This will visually indicate the
boundaries of the sterile field.
Surgical and Endovascular Room Return Air Requirements: During a
surgical and/or interventional procedure, all of the net floor space from the
floor surface to a distance 15 in. (380 mm) above is considered
contaminated. Therefore, all return grilles must be positioned low on the
wall above the floor. Return air for both the general operating rooms and
hybrid operating rooms is collected at low wall grilles. There shall be a
minimum of four low return air grilles located at each corner of each surgical
operating room and/ or endovascular procedure room. The height of the
bottom of the return air grilles above the floor is indicated in the VA HVAC
Design Manual, typically it is eight (8) inches (203 mm) above the finished
floor surface.
Outdoor Air Requirements: The HVAC design for surgery support spaces
shall provide each space with not less than the minimum recommended
quantity of ventilation air indicated in the current version of the VA HVAC
Design Manual.
Exhaust and Return Air Requirements: The HVAC design for spaces
other than operating rooms shall provide return air or exhaust air as
required for the spaces to control the transfer of odors and provide proper
room pressurization. At a minimum, exhaust air and pressurization should
be provided as indicated in the VA HVAC Design Manual.
Noise Criteria: The HVAC design shall provide resulting sound levels in
occupied spaces not to exceed the levels shown in the current version of
the VA HVAC Design Manual.
NARRATIVE 2-45
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-46
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
• Clean Core
• Clean Utility Rooms
Floor drains and floor sinks may be installed in the following rooms:
• Housekeeping closets (HAC)
• Soiled utility rooms
• Toilets
• Staff toilet showering rooms
Note: Absolutely no drains and/or floor sinks shall be installed within a
surgical operating room or endovascular procedure room under any
circumstances.
NARRATIVE 2-47
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-48
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-49
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-50
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-51
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-52
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-53
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-54
Surgical and Endovascular Services Design Guide April 2016 (rev 02/23)
NARRATIVE 2-55
Surgical and Endovascular Services Design Guide April 2016 (rev 9/21)
3.1. General:
Not every patient will follow the same workflow protocol; therefore, several
different scenarios have been diagrammed in Figures 3.1.1, 3.1.2, 3.1.3, 3.1.4,
and 3.1.5 to indicate possible routes that a patient may take through the surgical
department.
It is important to note that the major differences stem from the level of acuity
of the procedure being performed: in the case of ambulatory surgery, whether
within the VAMC hospital or the Community Based Outpatient Clinic (CBOC)
the patient will arrive on the day of surgery and be discharged on that same day
without interacting with any other spaces outside of the surgery department.
During a hospital-based surgery, intake and discharge of the patient will occur
through other hospital departments. Though dependent on the complexity of
the surgery, the patient will have been admitted to the hospital pre-assessed
then transported to pre-operative holding, sedated and then to surgery. In
emergent surgical cases, the patient will likely not spend any time in pre- and
post-procedural areas but be transferred directly to the critical care unit. This
serves to minimize patient transfers and ultimately reduce patient risk.
Figure 3.1.1 Overall perioperative layout where the entire patient journey is
mapped from pre-operative phase then to the intra-operative phase to post-
operative phase then to patient discharge. Whether the patient’s interventional
procedure occurs as an inpatient, elective out-patient, or emergent trauma,
either within the hospital, in a stand-alone ambulatory outpatient surgical facility,
or as a component of an outpatient community based clinic, the patient’s journey
is relatively the same.
Figures 3.1.2, 3.1.3, 3.1.4, and 3.1.5 are more specific to how the patient
presents for a surgical or endovascular procedure.
• Unrestricted Area
• Semi-Restricted Area
• Restricted Area
The unrestricted area includes the central control point established to monitor
the entrance of patients, personnel, and materials. Street clothes are permitted
in this area and traffic is not limited. The front desk, patient hold, and supply are
considered semi-restricted support areas. Street clothes may be worn in these
areas. Public access to the area may be limited based on the facility’s policy
and procedures.
The semi-restricted area (“red line”) includes the peripheral support areas
of the surgical suite. This area includes storage areas for clean and sterile
supplies, work areas for processing instruments by SPS personnel, scrub sink
areas and corridors leading to the restricted areas of the surgical suite, and
the entrances to locker rooms, the preoperative admission area, the PACU,
and sterile processing. This area is entered directly from the unrestricted area
past a nurses’ station or from other areas. Traffic in semi-restricted areas is
limited to authorized personnel and patients. Note that personnel are required
to wear surgical attire and cover all head and facial hair. The care of the patient
during surgery requires movement of patients, personnel, and material within
the surgical suite. Planning and controlling these movements assist in the
containment of contamination. Only authorized personnel are allowed in the
restricted areas. “Authorized” personnel are those assigned to Surgical Services
to include:
the semi-restricted corridor that surrounds all of the surgical and interventional
procedure rooms. Noise and the number of personnel allowed to enter into the
operating room and/or endovascular procedure room is kept to a minimum,
especially with a surgical procedure in progress.
Persons entering the semi-restricted or restricted areas of the surgical suite for
a brief time for a specific purpose, such as parents or biomedical engineers,
must don proper surgical scrub attire and may don either freshly laundered
surgical attire or a single-use coverall suit designed to totally cover outside
apparel. Patients are donned with clean gowns, covered with clean linens and
hair coverings. Surgical supplies prepared for surgical procedures outside the
surgical suite (e.g. in Sterile Processing Department, SPD) are transported to
the surgical suite by means of a clean elevator from SPD to the clean core to
maintain cleanliness and sterility and to prevent physical damage.
Soiled supplies, instruments, and equipment shall not enter the clean core area.
Contaminated items shall be in closed, covered carts or containers for transport
to the decontamination area via a “soiled” elevator which terminates in the soiled
side of SPD.
The following Figures 3.2.1, 3.2.2, and 3.2.3 illustrate typical workflows in
surgical department of different complexity levels.
Notes:
General surgery as well as minimally invasive surgery (MIS) are performed in
the same size room. It should be noted that non-robotic MIS is also known as
endoscopic surgery which includes: laparoscopic surgery and thoracoscopic
surgery, also known as “keyhole” surgery. These are minimally invasive
procedures that utilize an endoscope to reach internal organs through very
small incisions. MIS surgery offers patients several benefits such as smaller
incisions, faster recovery times, and reduced pain and scarring. in many
cases, MIS surgery also offers a higher accuracy rate compared to traditional
open general surgery.
* Clean and Soiled elevators are not necessary in facilities with central sterile
processing, distribution and waste management located on the same floor with
that of perioperative services.
The surgical suite is set up to carefully orchestrate the flow of patients, clinical
and support staff, sterile and soiled goods through on-stage vs. off-stage
circulation organization. the workflow should allow for one-way flow of soiled
goods out of the operating room and sterile goods into the operating room on a
separate, dedicated routes.
This is best accomplished by grouping operating rooms around the Clean Core,
as illustrated in figures 3.2.4 through 3.2.8. Each Clean Core cluster shall be
limited to 6 to 8 operating rooms in total in order to avoid excessive patient
transportation and travel distances in the semi-restricted corridor.
Figure 3.2.7 Clean Core, 1 Pod along Racetrack Single Loaded Corridor
Figure 3.2.8 Clean Core, 2 Pods along Racetrack Single Loaded Corridor - Alternate
A double-loaded Clean Core (OR’s on each side of the clean core) shall have
a min. width of 21’-0” (6401 mm). This width will allow for supply shelving in
the middle, aisles on each side as well as case cart storage along the wall.
Sterile supplies are brought to from SPS each day via a clean elevator or
transported via a clean route to the clean core and/or sterile storage room in
the case of double loaded corridor suite design.
Staging of case carts in the clean core starts each evening before the first
of the day surgical procedure, the carts are staged against the clean core
wall either singularily or in pairs contingent on the type and complexity of
the procedure to be performed. The case carts for the following procedure
are typically staged in an area proximal to the clean elevator (when SPS is
beneath surgery). The “core” technician responsibilities include maintaining
and facilitating the function of the clean core, that includes case cart
preparation with the finalization in the core for scheduled/add-on/emergent
cases for current and next day schedule, maintaining, troubleshooting, and
assuming responsibility to ensure that instrumentation and supplies are
available as needed for cases.
The following Figure 3.3.1.2 represents the same clean core configuration
with equipment location components identified. It should be noted that the
equipment shown can be oriented differently as to meet local preferences.
4.1. General
Introduction:
The Room Templates are intended as general representations of typical space, furniture
and equipment layout, as well as functional and utility supporting needs. The Room
Templates were developed as a design tool to assist the Project Team in understanding
the choices to be made during design, and to assist designers in understanding VA’s
space and functional requirements for Surgical Services. The Room Templates are
not intended to be project specific and are not meant to limit design opportunities.
However, the indicated net square feet (NSF) is the minimum acceptable square feet
(the +/- rule does not apply).
While this information is provided for a majority of spaces required, it is not possible
to foresee all possible variations or future requirements. The project-specific space
program shall be used as the basis for individual project design.
The Room Templates must be reviewed against project criteria and any special
requirements. Users shall follow other VA criteria and standards as required by VA.
Equipment manufacturers shall be consulted for the most current equipment information
such as actual dimensions, weights and utility requirements.
Legend of Symbols:
GYPSUM BOARD
3
9’-0”
HEIGHT OF CEILING FINISH ABOVE FINISH FLOOR
JSN
JSN AND EQUIPMENT NAME
EQUIPMENT NAME
WHEELCHAIR CLEARANCE
CENTERLINE
OCCUPANCY SENSOR DT
LIGHTING
2’X4’ FLUORESCENT LIGHT FIXTURE
DEVICES /
FIXTURES
2’X4’ DIRECT/INDIRECT LIGHT FIXTURE
D DDD 3 333
RECEPTACLES GROUND FAULT INTERRUPTER RECEPTACLE
DD 3 3
RECEPTACLE, CLOCK HANGER C
CB
D 3D 3
NURSE CALL DEVICE (CODE BLUE)
HVAC SUPPLY
HVAC RETURN
*LINEAR DIFFUSER
EXHAUST REGISTER
HOOD EXHAUST
PLUMBING SPRINKLER
Axonometric
140 NSF / 13,1 NSM
CT020
COUNTER, SOLID
A5082 SURFACE
DISPENSER, PAPER TOWEL, SENSOR, HANDS FREE CS140
A5180
A5077 SINK, SS,GENERAL,
TRACK, CUBICLE, SINGLE X, 10X14X ID
SURFACE MOUNTED, DISPENSER, HAND SANITIZER, HANDS FREE
A5075
WITH CURTAIN
DISPENSER, SOAP
F2010
M4665 BASKET, WASTEPAPER,
STRETCHER, 1 GFI STEP-ON
RECOVERY, SURGICAL A1012
T
a b
F0205 A5220
SIDE CHAIR WITH ARMS A5212
F3200 BRACKET, TV,
3759mm
WALL-MOUNTED,
12' - 4"
""
SCALE: 1/4” = 1’-0”
4' 8' 16' Floor Plan
140 NSF / 13,1 NSM
9' - 0"
3759mm
12' - 4"
ab
BED
11' - 4"
3454mm
Elevations 1 & 2
A5082 E0948
DISPENSER, PAPER TOWEL, SENSOR, HANDS FREE CART, GENERAL STORAGE, MOBILE
11' - 4"
3454mm M0765
REGULATOR, VACUUM
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
M0755 M0750
4-8
April 2016 (rev 2/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
ARCHITECTURAL POWER
Ceiling Type: Acoustical Ceiling Tile (SP) Normal Power: To be connected to selected
Ceiling Height: 9’-0” (2700mm) receptacles and equipment.
Ceiling Finish: Emergency Power: Critical branch of the EES
to be connected to selected
Wall Finish: Gypsum Wallboard (SC)
receptacles and equipment.
Wainscot:
Base: WSF Integral Base (min. 6”/
152 mm)
Floor Finish: Welded Seam Sheet
Flooring
Slab Depression: None Special
Sound Protection: None
Doors: None
LIGHTING COMMUNICATIONS
Refer to the VA Lighting Design Manual section Data: Yes
4.2.15 - Pre-Operative and Post- Anesthetic Care Telephone: Yes
(PACU) - for lighting design consideration.
Cable Television: No
Duress Alarm: No
Electronic Access and Door No
Control:
Intercom: No
Motion Intrusion Detection No
(MID):
Nurse Call: Yes
Code Blue: Yes
Public Address: No
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: Yes
Video Teleconferencing No
(VTEL):
Axonometric
140 NSF / 13,1 NSM
CT020
COUNTER, SOLID
SURFACE
CS140
M4665 SINK, SS,GENERAL,
STRETCHER, SINGLE X, 10X14X ID
RECOVERY, SURGICAL F2010
M7040 BASKET, WASTEPAPER,
TABLE, OVERBED STEP-ON
A5075
A5180
DISPENSER, SOAP
TRACK, CUBICLE,
A5077
SURFACE MOUNTED,
WITH CURTAIN DISPENSER, HAND
SANITIZER, HANDS FREE
a b
A5145 A5082
HOOK, GARMENT, DISPENSER, PAPER
DOUBLE TOWEL, SENSOR, HANDS
1 FREE
F0205 6" AFC
SIDE CHAIR WITH ARMS
GFI A1012
2 TELEPHONE, WALL
F3200 MOUNTED, 1 LINE
CLOCK, BATTERY, 12IN A5220
A5212
3759mm
12' - 4"
BRACKET, TV,
TV
WALL-MOUNTED,
TILT/ANGLE
M3070
M0506
HAMPER, LINEN
T TELEVISION, FLAT
SCREEN
M3072 E0948
FRAME, INFECTIOUS CART, GENERAL
WASTE BAG W/LID STORAGE, MOBILE
A1107 A5107
RAIL SYSTEM, UTILITY, DISPENSER, GLOVE,
NC
9' - 0"
3759mm
12' - 4"
ab
BED
11' - 4"
3454mm
Elevations 1 & 2
DISPENSER, SOAP A5107
CS140 DISPENSER, GLOVE, WALL-MTD
F2010
BASKET, WASTEPAPER,
STEP-ON
ELEVATION 1
11' - 4"
3454mm M0765
REGULATOR, VACUUM
4.3. PATIENT ROOM, PRE-OPERATIVE HOLDING /
M0755
M0750
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
FLOWMETER, OXYGEN,
LOW FLOW FLOWMETER, AIR
4-19
April 2016 (rev 2/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
ARCHITECTURAL POWER
Ceiling Type: Acoustical Ceiling Tile (SP) Normal Power: To be connected to selected
Ceiling Height: 9’-0” (2700mm) receptacles and equipment.
Ceiling Finish: Emergency Power: Critical branch of the EES
to be connected to selected
Wall Finish: Gypsum Wallboard (SC)
receptacles and equipment.
Wainscot:
Base: WSF Integral Base (min. 6”/
152 mm)
Floor Finish: Welded Seam Sheet
Flooring
Slab Depression: None Special
Sound Protection: 35 STC
Doors: Sliding Glass Door, Size 8’-
0” x 7’-0” (2438 mm x 2133
mm)
LIGHTING COMMUNICATIONS
Refer to the VA Lighting Design Manual section Data: Yes
4.2.15 - Pre-Operative and Post - Anesthetic Telephone: Yes
Care (PACU) - for lighting design consideration.
Cable Television: No
Duress Alarm: No
Electronic Access and Door No
Control:
Intercom: No
Motion Intrusion Detection No
(MID):
Nurse Call: Yes
Code Blue: Yes
Public Address: No
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: Yes
Video Teleconferencing No
(VTEL):
Axonometric
240 NSF / 22,3 NSM
F3200
CLOCK, BATTERY, 12IN
E0220
WORKSURFACE, W
OVERHD CAB AND
DRWRS, WALL MTD 3a
T
M1825
PRINTER, COMPUTER
M1801
COMPUTER,
MICROPROCESSING, W/
FLAT PANEL MONITOR
F0230
CHAIR, DRAFTING,
ROTARY
A1015 M20
M2055
2055
55
5334mm
17' - 6"
TELEPHONE, DESK, SHELVING,
SHELV
SHELVIING, STORAGE,
STORAGE,
MULTIPLE LINE 12x12 WIRE, CRS,
WIRE, CRS, W
F2000 ADJ
ADJUSTABLE
ADJU
USTAB
STABLELE SHELVES
SHELVES
BASKET, WASTEPAPER,
ROUND, METAL
F3050
M2055
WHITE BOARD, DRY
ERASE SHELVING, STORAGE,
WIRE, CRS, W
ADJUSTABLE SHELVES
3a
13' - 8"
4166mm
a a
9' - 0"
5334mm
17' - 6"
a S DT a
a a
13' - 8"
4166mm
ARCHITECTURAL POWER
Ceiling Type: Gypsum Wallboard (SC) Normal Power: To be connected to selected
Ceiling Height: 9’-0” (2700mm) receptacles and equipment.
Ceiling Finish: Emergency Power: Critical branch of the EES
to be connected to selected
Wall Finish: Gypsum Wallboard (SC)
receptacles and equipment.
Wainscot:
Base: Resilient Base
Floor Finish: Rubber Flooring, Welded
Seam Sheet Flooring
Slab Depression: None Special
Sound Protection: None COMMUNICATIONS
Doors: Single Door, Size 3’-8” x Data: Yes
7’-0” (1117 mm x 2133 mm)
Telephone: Yes
Wood
Cable Television: No
LIGHTING Duress Alarm: No
Anesthesia Work Room has similar lighting Electronic Access and Door
design consideration as Exam & Treatment Control:
Room. Refer to the VA Lighting Design Manual Intercom: No
section 4.2.1 - Examination and Treatment Room
- for lighting design consideration. Motion Intrusion Detection No
(MID):
Nurse Call: No
Code Blue: No
Public Address: Yes
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: No
Video Teleconferencing No
(VTEL):
Special Requirement:
Notes:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-34
4.5. ORGS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORGS1) OPERATING ROOM, GENERAL
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-35
4.5. ORGS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORGS1) OPERATING ROOM, GENERAL
FLOOR PLAN
(8.74m)
28' - 8"
M8825 A1014
(5.49m)
18' - 0"
M3070 M3150
M8970
GM
M3070
M8551
M8830 M4255
M4266
A1200 M4255
LIFT AND M4266
EXTENDER RAIL
E0954 M0630 M3080
M7650 M1801
M7845
M8800 S9755
A5108 M8920
MANIFOLD M8940
UNI-DIRECTIONAL
AIR DELIVERY IPS GM D D
ARRAY ABOVE
STERILE FIELD A5104 A5107
M4645
CLEAN CORE
OPERATING ROOM,
GENERAL
(ORGS1)
650 NSF
60.79 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-36
4.5. ORGS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORGS1) OPERATING ROOM, GENERAL
REFLECTED CEILING PLAN
LED
LED
A
LED
LED
LED
LED
(8.74m)
28' - 8"
A
TABLE
A
M7490
LED
LED
LED
LED
A
M7490
M7801
A1200
LED LED LED LED LED LED
LIFT AND
EXTENDER RAIL
GYPSUM BOARD
SOFFIT.
A A A A COORDINATE
HEIGHT WITH
EQUIPMENT.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-37
4.5. ORGS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORGS1) OPERATING ROOM, GENERAL
REFLECTED CEILING PLAN (ALTERNATIVE OPTION)
10' - 0"
M1801
A1122
6' - 2" 5' - 10" (DOCUMEN
AIR 1877mm 1781mm TATION)
CURTAIN
LAMINAR AIR
ab ab ab
4927mm
FLOW
16' - 2"
DIFFUSER
2134mm
ab ab
7' - 0"
8738mm
28' - 8"
M7801
ab ab TABLE
M7490
M7475
2134mm
7' - 0"
ab ab
3811mm
12' - 6"
M7801
U0100
ab ab ab
NOTE: OPTIONAL REFLECTED CEILING PLAN IS AN ALTERNATIVE OPTION TO THE MANIFOLD UNI-DIRECTIONAL AIR DELIVERY SYSTEM.
EQUIPMENT TAGS & REFERENCE NOTES MAY NOT BE FULLY COORDINATED WITH THE EQUIPMENT LIST.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-38
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
M7802
Monitor, Medical Grade
55-65"
A5212
LOW WALL RETURN
Bracket, Television, T H
AIR REGISTER
Wall-Mounted, Tilt/Angle
(TYPICAL)
A5108
Waste Disposal Unit, Sharps
M8910
Cart, Surgical Case
A5107
Dispenser, Glove, ELEVATION 1
Surgical/Examination, Wall
Mntd
A1014
Telephone, Wall
Mounted, 1 Line, With
A4015 Speaker
Clock, Elapsed Time, Electric M3150
F3200 Distribution System,
Clock, Battery, 12" Diameter Medication, Automatic
F3050
Whiteboard, Dry Erase
ISOLATED POWER
SYSTEM
M3080
Cabinet, Instrument,
CRS, 2 Glass Door, 6
Shelf
LOW WALL RETURN
AIR REGISTER
(TYPICAL)
F0355 ELEVATION 2
Footstool, Straight
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-40
4.5. ORGS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORGS1) OPERATING ROOM, GENERAL
ELEVATIONS
A5107
Dispenser, Glove,
Surgical/Examination, Wall
Mntd
M3080
Cabinet, Instrument, CRS, 2
Glass Door, 6 Shelf ISOLATED POWER
SYSTEM
LOW WALL RETURN
SD SD
AIR REGISTER
(TYPICAL)
M8920 A5104
Stand, Basin, CRS, Mobile, Cart, Medical Waste
Double Disposal, Mobile w/Foot
S9755
ELEVATION 3 Pedal
Suction System, Surgical, M4645
Mobile Rover Unit Patient Transfer Device
SURGICAL LIGHTING
CONTROL PANEL
M7650
Defibrillator/Monitor, Acute M8910
Care Cart, Surgical Case
E0954 LOW WALL RETURN
Cart, Emergency, Mobile, AIR REGISTER
66"H x 32"W x 22"D (TYPICAL)
M8830 M8905
Table, Instrument/Dressing, ELEVATION 4 Pail, Utility, CRS, With
Mobile Carriage
M3070 M8900
Hamper, Linen, Mobile, w/Lid Carriage, Pail, CRS,
Without Pail
M3072
Frame, Infectious Waste
Bag w/Lid
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-41
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.5 ORGS1
4.5 OPERATING ROOM, GENERAL (ORGS1)
Room Data Sheet
Notes:
1) Refer to the latest version of the VA HVAC
Design Manual for quantity and location of low
air return grilles and ceiling diffusers.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-50
4.6. OROS1
SURGICAL AND ENDOVASCULAR SERVICES
(OROS1) OPERATING ROOM, ORTHOPEDIC
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-51
4.6. OROS1
SURGICAL AND ENDOVASCULAR SERVICES
(OROS1) OPERATING ROOM, ORTHOPEDIC
FLOOR PLAN
26' - 10"
(8.18m)
8' - 0" CLEAR
SEMI-RESTRICTED
CORRIDOR
DOORS TO BE ON
AUTOMATIC DOOR
STRETCHER ALCOVE OPENERS
A5212 SCRUB ALCOVE
M7802 RETURN AIR
DUCT (TYPICAL)
LOW WALL
GM H T A4015 RETURN AIR
TV TV
F3200 REGISTER
A5107 (TYPICAL)
M8910 1
4 2 M4815
IPS
3 M8830
TABLE M8551
16' - 0" M8840
(4.88m) M8970
GM
GM
M8606 M4255 F3050
M8925 M4287 M8940
X4890 M8810 F0355
A1014
M8900 M5030
M8905 M8920
(4.27m)
(8.74m)
14' - 0"
28' - 8"
M3072 M3175
M8950 M3150
M4816
M8825
M5030
TABLE
M3070 M8840
M4255
M4266 M4255
M4266
A1200
LIFT AND M0630
EXTENDER RAIL M1801
M7845
E0954
M7650
M9110 M3080
M8800
A5104 A5108
M8940
MANIFOLD
UNI-DIRECTIONAL
AIR DELIVERY IPS GM D D
ARRAY ABOVE
STERILE FIELD M4645
M8920
S9755 A5107
CLEAN CORE
OPERATING ROOM,
ORTHOPEDIC
(OROS1)
750 NSF
69.62 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-52
4.6. OROS1
SURGICAL AND ENDOVASCULAR SERVICES
(OROS1) OPERATING ROOM, ORTHOPEDIC
REFLECTED CEILING PLAN
26' - 10"
(8.18m)
8' - 0" CLEAR FIN. CLG. 10' - 0" A.F.F.
SEMI-RESTRICTED
RETURN AIR
CORRIDOR
DUCT EACH
CORNER DIMMABLE, LED STRIP
(TYPICAL) LIGHT TYPICAL
A1122
A1130 A1120
M0765 M0750
M4280 M0755
(EQUIPMENT) M0765
4V, 1 NITROGEN, A A A A (ANESTHESIA)
1 CO2, 2O, 3V, 1A, WAGD,
10 EMERG DUPLEX, NITROUS OXIDE,
4 DATA FACE PLATES 6 EMERG DUPLEX,
4 DATA FACE PLATES,
OUTLINE OF
PROCEDURE
TABLE
A
A
LED LED LED LED
SEE LEGEND:
MANIFOLD A1122
A
A
LED
LED
LED
UNI-DIRECTIONAL M1801
LED
LED
LED
A
LED LED LED LED
(8.74m)
28' - 8"
STERILE FIELD
LED
LED
LED
LED
LED
M7490
LED
LED
LED
A
A
M7801 TABLE
LED LED LED LED
A1200
LIFT AND
LED
LED
LED
LED
LED
EXTENDER RAIL
LED
LED
LED
LED
LED
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-53
4.6. OROS1
SURGICAL AND ENDOVASCULAR SERVICES
(OROS1) OPERATING ROOM, ORTHOPEDIC
REFLECTED CEILING PLAN (ALTERNATE OPTION)
M4280
2O, 3V, 1A, WAGD, 1 M0765 M0765
NITROUS OXIDE, M8606
M0755 4V, 1 NITROGEN, 1 CO2, 10 EMER
6 EMER DUPLEX, 4 A1130
DATA FACE PLATES M0750 DUPLEX, 4 DATA FACE PLATES
(ANESTHESIA) A1120 A1122 (EQUIPMENT)
S 10' - 0"
AIR CURTAIN
LAMINAR AIR
ab ab ab ab
4927mm
16' - 2"
FLOW DIFFUSER
(DOCUMEN
TATION)
2134mm
ab ab
7' - 0"
A1122
M1801
8738mm
28' - 8"
M7475
M7490
ab ab TABLE
M7801
2134mm
7' - 0"
ab ab M7801
3811mm U0100
12' - 6"
ab ab ab ab
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-54
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
M7802 F3200
Monitor, Medical Grade Clock, Battery, 12"
55-65" Diameter
A5212 A4015
Bracket, Television, Clock, Elapsed Time,
Wall-Mounted, Tilt/Angle Electric
H T
A5108
Waste Disposal Unit, Sharps
M8910
Cart, Surgical Case
LOW WALL RETURN
AIR REGISTER
(TYPICAL)
A5107
ELEVATION 1 Dispenser, Glove,
Surgical/Examination,
Wall Mntd
A1014
Telephone, Wall
Mounted, 1 Line, With
F3050 Speaker
Whiteboard, Dry Erase M3150
ISOLATED POWER Distribution System,
SYSTEM Medication, Automatic
M3080
ELEVATION 2 Cabinet, Instrument, CRS, 2
F0355 Glass Door, 6 Shelf
Footstool, Straight LOW WALL RETURN
AIR REGISTER
(TYPICAL)
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-56
4.6. OROS1
SURGICAL AND ENDOVASCULAR SERVICES
(OROS1) OPERATING ROOM, ORTHOPEDIC
ELEVATIONS
M3080 A5107
Cabinet, Instrument, Dispenser, Glove,
CRS, 2 Glass Door, 6 Surgical/Examination,
Shelf Wall Mntd
M4645
Patient Transfer Device
S9755
Suction System, Surgical,
Mobile Rover Unit
M8920
Stand, Basin, CRS,
Mobile, Double
ISOLATED POWER
SYSTEM
SD SD
LOW WALL RETURN
AIR REGISTER
(TYPICAL)
ELEVATION 3
M7650
Defibrillator/Monitor, Acute
Care
E0954 M3072 X4890 LOW WALL RETURN
Cart, Emergency, Mobile, Frame, Infectious Waste Bag Rad/Fluoro Unit, Digital, AIR REGISTER
66"H x 32"W x 22"D w/Lid Mobile, C-Arm (TYPICAL)
A5104 M8905
Cart, Medical Waste Disposal, Pail, Utility, CRS, With
Mobile w/Foot Pedal
ELEVATION 4 Carriage
M3070 M8900
Hamper, Linen, Mobile, w/Lid Carriage, Pail, CRS,
Without Pail
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-57
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.6 OROS1
4.6 OPERATING ROOM, ORTHOPEDIC (OROS1)
Room Data Sheet
omitted subject to approval by clinical
ARCHITECTURAL leadership.
8) Facility will select number and types of
Ceiling Type: Gypsum Wallboard (SC)
scopes and other instrumentation as
Ceiling Height: 10’-0” (3048 mm)
necessitated by the unique case load.
Ceiling Finish:
9) Endoscopy equipment can be located on a
Wall Finish: Gypsum Wallboard (SC) cart or on the equipment boom.
Wainscot: 10) NSF provided for this space is the minimal
Base: RF Integral Base (min. 6”/ acceptable NSF; contact Facilities Standards
152 mm)
Services for any deviations.
Floor Finish: Resinous Flooring
Slab Depression: None
Sound Protection: 50 STC (to other room), 35 LIGHTING
STC (to corridor)
Doors: Double, Size 6’-0” x 7’-0” Refer to the VA Lighting Design Manual section
(1829 mm x 2133 mm) 4.2.13 - Surgery/Operating Room - for lighting
Wood w/ Narrow View design consideration.
Window; Single, Size 4’-0” x
7’-0” (1219 mm x 2133 mm)
Wood w/ Small View
Window POWER
Special Requirement:
Normal Power: Connect selected
receptacles and equipment
Notes: to Normal power IPS.
1) Shielding is to be provided as it is Emergency Power: Connect selected
determined by the Physicist on a per project receptacles and equipment
basis. to Critical Branch
emergency IPS.
2) Locate access panels as required to allow
for the maintenance of surgical booms and lights
in facilities without interstitial space. Min. size to Notes:
be 24” x 24”.
1) Provide IPS power & ground modules – 3
3) Cabinetry can be built in or free-standing. duplex receptacles & 3 ground jacks
4) Nominal wall thickness is shown at 8” (203 2) IPS Power & ground modules mounted at
mm) to account for a variety of wall-mounted +24” AFF
panels, such as isolation power unit panels, that
3) Provide Laser Receptacle Module. Module
require a thicker partition.
shall be connected to Special Equipment IPS
5) Include wall extensions at both sides of the located outside the Surgery Room.
scrub sink to protect the scrub sinks from cart
and stretcher traffic in the semi-restricted 4) Provide power connections for articulating
corridor. utility columns.
6) Coordinate structural supports, utility
connections and other requirements for surgical
lighting pendants with manufacturer.
7) Equipment and Anesthesia booms are
duplicated to provide maximum flexibility. If
duplicate booms are not desired, they can be
Notes:
1) Refer to the HVAC Design Manual for number
and location of low air return grilles and ceiling
diffusers.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-66
4.7. ORCS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCS1) OPERATING ROOM, UROLOGY / CYSTOSCOPY
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-67
4.7. ORCS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCS1) OPERATING ROOM, UROLOGY / CYSTOSCOPY
FLOOR PLAN
23' - 5"
(7.14m)
8' - 0" CLEAR
SEMI-RESTRICTED
CORRIDOR
DOORS TO BE ON
STRETCHER ALCOVE
AUTOMATIC DOOR OPENERS
A5212 SCRUB ALCOVE RETURN AIR
M7802 DUCT (TYPICAL)
LOW WALL
H T
RETURN AIR
GM TV TV
REGISTER
(TYPICAL)
A5107
1 M8910 A5108
MANIFOLD
4 2 14' - 0" UNI-DIRECTIONAL
M8830 IPS AIR DELIVERY
3 TABLE
M8900 A5108 EQ EQ ARRAY ABOVE
M8905 M8800 STERILE FIELD
M4266
GM
M3072
M4255
M8940
M4287 M8606
A1200 M4255
F3050
M5030
M3070 F3200
EQ
GM
A1014
(8.74m)
M4816
28' - 8"
F0355
C
M8940
TABLE
14' - 0"
M8825
M8925
A5104 M4255 A4015
X4890 M4266 M3150
M9110
M4815
M8810 M8840
M8950
EQ
M5512
M8810 M3175
DASHED LINE
REPRESENTS THE M8551
SURGICAL STERILE M8920
FIELD
M3080
M8825 M5030
M0630 M8920 S9755
M7845 M8830
M8970
GM
IPS D D
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-68
4.7. ORCS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCS1) OPERATING ROOM, UROLOGY / CYSTOSCOPY
REFLECTED CEILING PLAN
23' - 5"
(7.14m)
8' - 0" CLEAR
SEMI-RESTRICTED
CORRIDOR
(4.93m)
16' - 2"
TYPICAL A A
A1122
M7490 M1801
M7801 (DOCUMENTATION)
4V, 1 NITROGEN,
1 CO2,10 EMERG
DUPLEX, 4
DATA FACE
OUTLINE OF PLATES
PROCEDURE (EQUIPMENT)
EQ
(ANESTHESIA) A1122
(8.74m)
LED
LED
LED
LED
A1130
28' - 8"
2O, 3V, 1A, WAGD, A1120
LED
LED
LED
LED
M0750 M0765
LED
A
A
1 NITROUS OXIDE, M4280
M0755 LED LED LED LED
LED
LED
LED
LED
TABLE
LED
LED
LED
LED
LED
PLATES
LED LED LED LED
A1120
M0750
(3.81m)
12' - 6"
(EQUIPMENT) M0755
LED
LED
LED
LED
LED
A
LED
LED
LED
EQ
A1130
1 CO2,10 LED LED LED LED
(ANESTHESIA)
M0765
EMERG DUPLEX, 2O, 3V, 1A, WAGD,
M4280
4 DATA FACE M8606 1 NITROUS OXIDE,
PLATES 6 EMERG DUPLEX,
4 DATA FACE
PLATES
GYPSUM BOARD
A A SOFFIT.
COORDINATE
HEIGHT WITH
M7490 EQUIPMENT.
M7801
EQ EQ SEE LEGEND:
A1200 MANIFOLD
UNI-DIRECTIONAL
LIFT AND
TABLE AIR DELIVERY
EXTENDER RAIL
ARRAY ABOVE
GENERAL NOTES: OPERATING ROOM, STERILE FIELD
1.) PROVIDE EMERGENCY LIGHTING AS REQUIRED BY CODE(S) UROLOGY /
2.) REFER TO THE LEGEND FOR ADDITIONAL INFORMATION FOR THE
CYSTOSCOPY
MANIFOLD AIR DELIVERY ARRAY (ORCS1)
3.) ROOM IS TO BE PROTECTED WITH QUICK RESPONSE (QR)
650 NSF
FIRE SPRINKLERS 60.39 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-69
4.7. ORCS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCS1) OPERATING ROOM, UROLOGY / CYSTOSCOPY
REFLECTED CEILING PLAN (ALTERNATE OPTION)
M4280
NOTE: REFER TO ROOM DATA SHEET FOR ACCESS PANELS
M0765
M7490 M8606 4V, 1 NITROGEN, 1 CO2, 10 EMER
A1130 DUPLEX, 4 DATA FACE PLATES
M7801 A1122 (EQUIPMENT)
U0100
10' - 0"
AIR
CURTAIN
7' - 6 1/2" 7' - 5 1/2" M1801
2296mm 2272mm A1122
LAMINAR (DOCUMEN
AIR FLOW TATION)
DIFFUSER ab ab ab
4927mm
16' - 2"
ab ab
2285mm
7' - 6"
8738mm
28' - 8"
ab ab
TABLE
ab ab
2289mm
7' - 6"
3811mm
12' - 6"
ab ab
ab ab ab
(ANESTHESIA) A1120
2O, 3V, 1A, M1801
WAGD, M0765
1 NITROUS M0755
OXIDE, 6 EMER
M0750
DUPLEX, 4 DATA
FACE PLATES A1122 A1120 (ANESTHESIA)
(EQUIPMENT) M7475
A1130 M0765 2O, 3V, 1A, WAGD,
4V, 1 NITROGEN, 1 CO2, 10 TABLE 1 NITROUS OXIDE,
M0765 M7801 M0755
EMER DUPLEX, 4 DATA 6 EMER DUPLEX,
FACE PLATES M0750
4 DATA FACE PLATES
EQ EQ
23' - 5"
7137mm
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-70
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
M7802
Monitor, Medical Grade
55-65"
A5212
Bracket, Television,
Wall-Mounted, Tilt/Angle
A5107
Dispenser, Glove,
Surgical/Examination, Wall
Mntd
A5108
Waste Disposal Unit, Sharps LOW WALL RETURN
M8910 AIR REGISTER
Cart, Surgical Case
H T S
LOW WALL RETURN
AIR REGISTER
ELEVATION 1
F3200
Clock, Battery, 12"
Diameter
A4015
Clock, Elapsed Time,
Electric
A1014
Telephone, Wall
Mounted, 1 Line, With
Speaker
F3050
Whiteboard, Dry Erase
ISOLATION M3080
POWER Cabinet, Instrument,
SYSTEM CRS, 2 Glass Door, 6
Shelf
LOW WALL RETURN
AIR REGISTER
M3150
ELEVATION 2 Distribution System,
F0355 Medication, Automatic
Footstool, Straight
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-72
4.7. ORCS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCS1) OPERATING ROOM, UROLOGY / CYSTOSCOPY
ELEVATIONS
M4645
Patient Transfer Device
S9755
Suction System, Surgical,
Mobile Rover Unit
A5107
Dispenser, Glove, M7650
Surgical/Examination, Wall Defibrillator/Monitor,
Mntd Acute Care
SD SD
A5104
Cart, Medical Waste
Disposal, Mobile w/Foot
Pedal
M3070
Hamper, Linen, Mobile,
w/Lid
M3072
Frame, Infectious Waste
Bag w/Lid
X4890 M8830
Rad/Fluoro Unit, Digital, Table,
Mobile, C-Arm Instrument/Dressing,
Mobile
LOW WALL RETURN
AIR REGISTER LOW WALL RETURN
AIR REGISTER
M8900
ELEVATION 4 Carriage, Pail, CRS,
Without Pail
M8905
Pail, Utility, CRS, With
Carriage
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-73
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.7 ORCS1
4.7 OPERATING ROOM, UROLOGY/CYSTOSCOPY (ORCS1)
Room Data Sheet
Notes:
1) Refer to the latest version of the VA HVAC
Design Manual for quantity and location of low
air return grilles and ceiling diffusers.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-82
4.8. ORCT1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCT1) OPERATING ROOM, CARDIOTHORACIC
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-83
4.8. ORCT1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCT1) OPERATING ROOM, CARDIOTHORACIC
FLOOR PLAN
26' - 10"
(8.18m)
8' - 0" CLEAR
SEMI-RESTRICTED
CORRIDOR DOORS TO BE ON
AUTOMATIC DOOR
STRETCHER ALCOVE SCRUB ALCOVE OPENERS
RETURN AIR
DUCT (TYPICAL)
(4.88m)
16' - 0"
M8825 L0221 F0355
EQ
M4816
M4280 M8810 F3050
(8.74m)
M3175
28' - 8"
M8810 M5030 M8940
M3070 M8825
M8950 M4811 M3150
M3072 M9080
TABLE
M5030
M8940
E0954 U0105
L1095
M4812 M8551
M7650 M8800
EQ
M8940
A1200
LIFT AND
M4250 M4810
EXTENDER RAIL
M4255
M4266 M8970
M0630 M3080
X2105 M7845
A5108
E0948
M7818
E0963 TO
CLEAN
IPS a b Dc
CORE
M8830 M4645 S9755
CLEAN CORE
A5107
OPERATING ROOM,
CARDIOTHORACIC
(ORCT1)
750 NSF
69.62 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-84
4.8. ORCT1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCT1) OPERATING ROOM, CARDIOTHORACIC
REFLECTED CEILING PLAN
16' - 0"
(4.88m)
SEE LEGEND:
MANIFOLD A A A A
UNI-DIRECTIONAL EQ EQ
AIR DELIVERY
TABLE
ARRAY ABOVE
STERILE FIELD A1122
M1801
(DOCUMENTATION)
A
A
M7490
A
A
LED
(8.74m)
M7801
28' - 8"
LED
TABLE
A
A
16' - 0"
LED
LED
LED
LED
LED
LED
A
A
M7801 OUTLINE OF
M7490 LED LED LED LED LED LED LED LED
PROCEDURE
TABLE
GYPSUM BOARD
SOFFIT.
A1200
COORDINATE
LIFT AND
HEIGHT WITH
EXTENDER RAIL
A A A A EQUIPMENT.
DASHED LINE
REPRESENTS
THE SURGICAL
STERILE FIELD FIN. CLG. 10' - 0" A.F.F.
(ANESTHESIA) A1122 CLEAN CORE
2O, 3V, 1A, WAGD, A1120 (EQUIPMENT) M0765
1 NITROUS OXIDE, M0765 4V, 1 NITROGEN, 1 C02, A1130
M0755 GENERAL NOTES:
6 EMERG DUPLEX, 10 EMERG DUPLEX,
M0750 4 DATA FACE PLATES OPERATING ROOM, 1.) PROVIDE EMERGENCY LIGHTING AS REQUIRED BY CODE(S)
4 DATA FACE PLATES 2.) REFER TO THE LEGEND FOR ADDITIONAL INFORMATION FOR THE
M1801 CARDIOTHORACIC
(ORCT1) MANIFOLD AIR DELIVERY ARRAY
750 NSF
3.) ROOM IS TO BE PROTECTED WITH QUICK RESPONSE (QR)
69.62 NSM FIRE SPRINKLERS
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-85
4.8. ORCT1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCT1) OPERATING ROOM, CARDIOTHORACIC
REFLECTED CEILING PLAN (ALTERNATE OPTION)
10' - 0"
4927mm
16' - 2"
M7801
2134mm
ab ab M7801
7' - 0"
U0100
8738mm
28' - 8"
ab ab TABLE
2134mm
7' - 0"
M7490
ab ab 3811mm
12' - 6"
M7490
ab ab ab ab
(ANESTHESIA) M1801
A1130 (EQUIPMENT) NOTE: REFER
2O, 3V, 1A, WAGD, 1 NITROUS A1120
OXIDE, 6 EMER DUPLEX, 4 DATA TABLE A1122 4V, 1 NITROGEN, 1 CO2, 10 TO ROOM DATA
M0765 EMER DUPLEX, 4 DATA
FACE PLATES M0765 SHEET FOR
M0755 FACE PLATES ACCESS PANELS
EQ M0750 EQ
26' - 10"
8179mm
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-86
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
JSN Legend
M4645 PATIENT TRANSFER DEVICE
JSN DESCRIPTION
M4810 HEART/ LUNG MACHINE, BYPASS,
MODULAR
A1014 TELEPHONE, WALL MOUNTED, 1 LINE,
M4811 PUMP, INTRA-AORTIC, BALLOON
WITH SPEAKER
M4812 PACEMAKER, SINGLE CHAMBER,
A1120 COLUMN, SERVICE, PREFAB,
EXTERNAL, TEMPORARY
SURGICAL, CEILING MOUNTED
M4815 HYPO/HYPERTHERMIA UNIT,
A1122 COLUMN, EQUIPMENT ARM, CEILING
AUTOMATIC/MANUAL, MOBILE
MOUNTED, SURGERY
M4816 WARMING UNIT, PATIENT, AUTOMATIC/
A1130 CABINET, CONTROL, NITROGEN
MANUAL, AIR
A1200 LIFT SYSTEM, OVERHEAD, PATIENT
M5030 STOOL, SURGEON, REVOLVING
ROOM
M5512 LASER, SMOKE EVACUATOR
A4015 CLOCK, ELAPSED TIME, ELECTRIC
M7490 LIGHT, SURG, CEILING MTD, DUAL,
A5104 CART, MEDICAL WASTE DISPOSAL,
UNEQUAL DIA HEADS
MOBILE W/FOOT PEDAL
M7650 DEFIBRILLATOR/ MONITOR, ACUTE
A5107 DISPENSER, GLOVE, SURGICAL/
CARE
EXAMINATION, WALL-MTD
M7801 MONITOR, MEDICAL GRADE, 26”-42”
A5108 WASTE DISPOSAL UNIT, SHARPS
M7802 MONITOR, MEDICAL GRADE, 55”-65”
A5212 BRACKET, TELEVISION, WALL-MTD,
M7818 MONITOR, TRANSPORT
TILT/ANGLE
M7845 MONITOR, PHYSIOLOGICAL, BEDSIDE,
E0954 CART, EMERGENCY, MOBILE, 66”H x
4 CHANNEL
32”W x 22”D
M8551 LIGHT SOURCE, FIBEROPTIC
E0963 CART, GEN. STORAGE, MOBILE
HEADLAMP
F0355 FOOTSTOOL, STRAIGHT
M8606 ENDOSCOPY CART, FIBEROPTIC,
F3050 WHITEBOARD, DRY ERASE
W/VIDEO ACCESSORIES
F3200 CLOCK, BATTERY, 12” DIAMETER
M8800 CART, ANESTHESIA
L0221 ANALYZER, BLOOD, PORTABLE, HAND
M8810 STAND, MAYO
HELD
M8825 TABLE, INSTRUMENT/DRESSING, CRS,
L1095 CELL SAVER,
APPROX 36”x20”x34”
M0630 ANESTHESIA APPARATUS, 3 GAS
M8830 TABLE, INSTRUMENT/DRESSING,
M0750 FLOWMETER, AIR, CONNECT W/50 PSI
MOBILE
SUPPLY
M8840 TABLE, BACK, INSTRUMENT/DRESSING
M0755 FLOWMETER, OXYGEN, LOW FLOW
M8900 CARRIAGE, PAIL, CRS, WITHOUT PAIL
M0765 REGULATOR, VACUUM
M8905 PAIL, UTILITY, CRS WITH CARRIAGE
M1801 COMPUTER, MICROPROCESSING,
M8910 CART, SURGICAL CASE
W/FLAT PANEL MONITOR
M8920 STAND, BASIN, CRS, MOBILE, DOUBLE
M3070 HAMPER, LINEN, MOBILE, W/LID
M8925 STAND, BASIN, CRS, MOBILE, SINGLE
M3072 FRAME, INFECTIOUS WASTE BAG W/LID
M8940 STOOL, ANESTHESIA, WITH BACK
M3080 CABINET, INSTRUMENT, CRS, 2 GLASS
M8950 WARMER, BLOOD
DOOR, 6 SHELF
M8970 WARMER, BLOOD, HIGH VOLUME
M3109 ELECTROSURGICAL/COAGULATOR,
M9080 TABLE, OPERATING, PEDESTAL,
ARGON PLASMA
5 SECTION
M3150 DISTRIBUTION SYSTEM, MEDICATION,
R4785 ICE MAKER, SURGICAL SLUSH
AUTOMATIC
S9755 SUCTION SYSTEM, SURGICAL, MOBILE
M3175 ELECTROSURGICAL UNIT, DUAL
ROVER UNIT
OUTPUT
U0105 EXTRACORPOREAL SUPPORT SYSTEM
M4250 PUMP, SYRINGE, INFUSION
X2105 SCANNER, ULTRASOUND, CARDIAC
M4255 STAND, IV, ADJUSTABLE
(ECHO)
M4266 PUMP, VOLUMETRIC, INFUSION,
MULTIPLE LINE
M4280 PUMP, PNEUMATIC STOCKING/CUFF
M4287 IRRIGATION SYSTEM, SURGICAL
F3200
Clock, Battery, 12"
A5107 Diameter
Dispenser, Glove, A4015
Surgical/Examination, Wall Clock, Elapsed Time,
A5212 Mntd Electric
Bracket, Television,
Wall-Mounted, Tilt/Angle
M7802
Monitor, Medical Grade
55-65"
A5108
Waste Disposal Unit, Sharps
M8910
H T
Cart, Surgical Case
A1014 F3050
Telephone, Wall Mounted, 1 Whiteboard, Dry Erase
Line, With Speaker M3150
ISOLATED POWER SYSTEM Distribution System, M3080
Medication, Automatic Cabinet, Instrument,
CRS, 2 Glass Door, 6
Shelf
F0355
Footstool, Straight LOW WALL RETURN
M8940
ELEVATION 2 AIR REGISTER (TYPICAL)
Stool, Anesthesia, With Back
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-88
4.8. ORCT1
SURGICAL AND ENDOVASCULAR SERVICES
(ORCT1) OPERATING ROOM, CARDIOTHORACIC
ELEVATIONS
M7818
Monitor, Transport
E0948
Cart, General Storage,
Mobile, 42"H x 32"W x
22"D
M4645 E0963
Patient Transfer Device Cart, General Storage,
Mobile
Dc b a
A5107
Dispenser, Glove,
Surgical/Examination, Wall
Mntd
S9755 LOW WALL RETURN
Suction System, Surgical, ELEVATION 3 AIR REGISTER (TYPICAL)
Mobile Rover Unit
M8830 ISOLATED POWER SYSTEM
Table, Instrument/Dressing,
Mobile
M4812
M7650 Pacemaker, Single
Defibrillator/Monitor, Acute Chamber, External,
Care Temporary
X2105 E0954
Scanner, Ultrasound, Cardiac Cart, Emergency, Mobile,
(Echo) 66"H x 32"W x 22"D
M3070
Hamper, Linen, Mobile,
w/Lid
A5104
Cart, Medical Waste
Disposal, Mobile w/Foot
Pedal
LOW WALL RETURN
AIR REGISTER (TYPICAL)
M3072 M8900
Frame, Infectious Waste Bag Carriage, Pail, CRS,
w/Lid
ELEVATION 4 Without Pail
M8905
Pail, Utility, CRS, With
Carriage
M3070
Hamper, Linen, Mobile,
w/Lid
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-89
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.8 ORCT1
4.8 OPERATING ROOM, CARDIOTHORACIC (ORCT1)
Room Data Sheet
Notes:
1) Refer to the latest version of the VA HVAC
Design Manual for quantity and location of low
air return grilles and ceiling diffusers.
M8925 Stand, Basin, CRS, Mobile, 2 V/V Mobile single basin stand with shelf. The stand shall be
Single constructed of tubular stainless steel and mounted on 2" swivel
casters. Shall include a shelf and an 8 quart stainless steel
basin. Intended for use in ORs and treatment areas.
M8940 Stool, Anesthesia, With Back 3 V/V Anesthesia stool with back. All stainless steel with well-curved
back panel and wide conductive seat. Designed for the
anesthesiologist during surgical procedures.
M8950 Warmer, Blood 1 V/V Unit consists of a temperature regulated water bath, circulating
fluid, or dry heat with controls and an audible high temperature
alarm. The warmer provides a stable environment for the
controlled warming of blood or other fluids prior to being
transfused to a patient.
M8970 Warmer, Blood, High Volume 1 V/V Unit contains a proportional controller to regulate temperature in
the heat exchanger and an audible high temperature alarm.
Designed to provide a stable temperature for the controlled
warming of blood or other fluids prior to being transfused to a
patient. Unit may be a cuff type or circulating water heat
exchanger.
M9080 Table, Operating, Pedestal, 5 1 V/V Pedestal type major operating table with 5 sections. Table is
Section mounted on a solid base with casters and locks. Table top
surface is fabricated from radio-translucent, conductive panels
and the larger table sections are equipped with radiographic
cassette tunnels. Table includes Electro-hydraulic controls, side
rail locking system, kidney elevator, grounding receptacle and
dual arm support section. Designed for use in the operating room
in a variety of surgical procedures.
R4785 Ice Maker, Surgical Slush 1 V/V An automated surgical slush machine designed to produce a
velvet soft slush to limit the likelihood of damaging tissue due to
large or sharp ice particles during surgical procedures. Unit
operates with a temperature range of 30 degrees F to 32
degrees F. The unit is also designed to operate quietly to not
disrupt the OR environment. Casters allow for easy movement.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-100
4.9. ORHL1
SURGICAL AND ENDOVASCULAR SERVICES
(ORHL1) PUMP ROOM, CARDIOTHORACIC / HYBRID OR
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-101
4.9. ORHL1
SURGICAL AND ENDOVASCULAR SERVICES
(ORHL1) PUMP ROOM, CARDIOTHORACIC / HYBRID OR
FLOOR PLAN
9' - 0"
F3200 (2.74m)
Clock, Battery, 12" Diameter
A1014 SEMI-RESTRICTED
Telephone, Wall Mounted, 1
CORRIDOR
Line, With Speaker
E0703
Table, Process, Adj Height, 5
Drawer, 48"W x 24"D +54"
AFF T 4a
M1801 EPO
Computer, Microprocessing,
w/Flat Panel Monitor
D3295
Chair, Rotary, Laboratory,
Dental
M4810 U0105
Heart/Lung Machine, Bypass, ExtraCorporeal Support
Modular System
M0750
Flowmeter, Air, Connect w/50
PSI Supply
OPERATING
M0755
SUITE
Flowmeter, Oxygen, Low
Flow
4a
M0765
(8.84m)
Regulator, Vacuum
29' - 0"
M4810
Heart/Lung Machine, Bypass,
Modular 1
R4650
Ice Maker, Flaked, With
Dispenser E0948
Cart, General Storage,
F2000 Mobile, 42"H x 32"W x
Basket, Wastepaper, Fire 22"D
Resistant E0963
A5080 Cart, General Storage,
Dispenser, Paper Towel, SS, Mobile
Surface Mounted
CS230
Sink, SS, Double
Compartment, 10x14x16 ID
GFI
+6"AFC
A5075
Dispenser, Soap, Disposable
CT050
Countertop, Stainless Steel
4a
R7250
Refrigerator/Freezer, 20
Cubic Feet
PUMP ROOM,
CARDIOTHORACIC /
HYBRID OR
(ORHL1)
260 NSF
24.25 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-102
4.9. ORHL1
SURGICAL AND ENDOVASCULAR SERVICES
(ORHL1) PUMP ROOM, CARDIOTHORACIC / HYBRID OR
REFLECTED CEILING PLAN
9' - 0"
(2.74m)
SEMI-RESTRICTED
CORRIDOR
(8.84m)
29' - 0"
A
DIRECT/INDIRECT
FIXTURE
A
PUMP ROOM,
CARDIOTHORACIC /
HYBRID OR
(ORHL1)
260 NSF
24.25 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-103
4.9. ORHL1
SURGICAL AND ENDOVASCULAR SERVICES
(ORHL1) PUMP ROOM, CARDIOTHORACIC / HYBRID OR
ELEVATIONS
M0765
R7250 Regulator, Vacuum
Refrigerator/Freezer, 20
A5075 M0755
Cubic Feet Flowmeter, Oxygen, Low
Dispenser, Soap, F3200
CD030 Disposable Flow Clock, Battery, 12" Diameter
Cabinet, W/H, 2 Shelf, 2 A5080 M0750 A1014
DO, Sloping Top, Dispenser, Paper Towel, Flowmeter, Air, Connect Telephone, Wall Mounted, 1
38x30x13 SS, Surface Mounted w/50 PSI Supply Line, With Speaker
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-104
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
4.9 ORHL1
4.9 PUMP ROOM, CARDIOTHORACIC/HYBRID OR (ORHL1)
Room Data Sheet
ARCHITECTURAL COMMUNICATIONS
Ceiling Type: Gypsum Wallboard (SC) Data: Yes
Ceiling Height: 9’-0” (2700 mm) Telephone: Yes
Ceiling Finish: Cable Television: No
Wall Finish: Gypsum Wallboard (SC) Duress Alarm: No
Wainscot: Electronic Access and Door Control: No
Base: RF Integral Base (min. 6”/ Intercom: No
152 mm) Motion Intrusion Detection (MID): No
Floor Finish: Resinous Flooring Nurse Call: No
Slab Depression: None Code Blue: No
Sound Protection: None Public Address: Yes
Doors: Single, Size 4’-0” x 7’-0” Security Surveillance Television (SSTV): No
(1219 mm x 2133 mm) VA Satellite TV: No
Wood
Video Teleconferencing (VTEL): No
Special Requirement:
Special Requirement: No
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-109
4.10. ORNS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORNS1) OPERATING ROOM, NEUROSURGICAL
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-110
4.10. ORNS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORNS1) OPERATING ROOM, NEUROSURGICAL
FLOOR PLAN
26' - 10"
(8.18m)
SEMI-RESTRICTED
CORRIDOR DOORS TO BE ON
STRETCHER ALCOVE AUTOMATIC DOOR
A5212 SCRUB ALCOVE OPENERS
M7802 RETURN AIR
DUCT (TYPICAL)
LOW WALL
H T RETURN AIR
TV TV
A4015 REGISTER
F3200 (TYPICAL)
1
A5107
A5108
4 2
M8910
IPS
M8830 3 M8840
16' - 0" M3109
M8900
M8905 (4.88m)
M8606 M4255
M8920 M8805 M4815 M4287
M8810 F3050
M3072 M8830
(4.27m)
F0355
14' - 0"
M3070 M8970
M8940
L0221
M8920 M8825
(8.74m)
28' - 8"
MANIFOLD M8925 M5512
M8551 A1014
UNI-DIRECTIONAL M8825
M8495 M5030 M3150
AIR DELIVERY
ARRAY ABOVE M4816
STERILE FIELD M9110 TABLE
M5030
M3080
A5104
M8535 M0630
A1200 M1801
LIFT AND M7845
M8800
EXTENDER RAIL A5108
M4255
M4250
M7650
M4266
E0954
M4816 M8940
IPS a b Dc
OPERATING ROOM,
NEUROSURGICAL
(ORNS1)
750 NSF
69.65 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-111
4.10. ORNS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORNS1) OPERATING ROOM, NEUROSURGICAL
REFLECTED CEILING PLAN
SEMI-RESTRICTED
CORRIDOR
RETURN AIR
DUCT EACH DIMMABLE, LED STRIP
CORNER LIGHT TYPICAL
(TYPICAL)
A1130
M0765 A1120
M4280 M0750
A1122 M0755
(EQUIPMENT) M0765
A A A A
4V, 1 NITROGEN, 1CO2, (ANESTHESIA)
10 EMERG DUPLEX, 2O, 3V, 1A, WAGD, 1
4 DATA FACE PLATES NITROUS OXIDE,
6 EMERG DUPLEX,
4 DATA FACE
PLATES,
OUTLINE OF
PROCEDURE
TABLE
A
A
SEE LEGEND:
LED LED LED LED
MANIFOLD A1122
A
A
LED
LED
LED
LED
LED
UNI-DIRECTIONAL M1801
LED
LED
LED
LED
LED
(8.74m)
ARRAY ABOVE
28' - 8"
A
A
LED LED LED LED
STERILE FIELD
LED
LED
LED
LED
LED
A
A
TABLE
LED
M7490
LED
LED
LED
LED
LED
M7801
LED
LED
LED
LED
LED
M7490
A1200 LED LED LED LED
M7801
LIFT AND
EXTENDER RAIL
A1120
M0750
M0755 GYPSUM BOARD
M0765 SOFFIT.
M1801 A A A A COORDINATE
(ANESTHESIA) HEIGHT WITH
2O, 3V, 1A, WAGD, EQUIPMENT.
1 NITROUS OXIDE,
6 EMERG DUPLEX,
4 DATA FACE A1122
PLATES TABLE A1130
M0765
CLEAN CORE FIN. CLG. 10' - 0" A.F.F. (EQUIPMENT)
GENERAL NOTES: 4V, 1 NITROGEN,
OPERATING ROOM,
1.) PROVIDE EMERGENCY LIGHTING AS REQUIRED BY CODE(S) 1 CO2, 10 EMER
2.) REFER TO THE LEGEND FOR ADDITIONAL INFORMATION FOR THE NEUROSURGICAL DUPLEX, 4 DATA
MANIFOLD AIR DELIVERY ARRAY
(ORNS1)
FACE PLATES
3.) ROOM IS TO BE PROTECTED WITH QUICK RESPONSE (QR) 750 NSF
69.65 NSM
FIRE SPRINKLERS
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-112
4.10. ORNS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORNS1) OPERATING ROOM, NEUROSURGICAL
REFLECTED CEILING PLAN (ALTERNATE OPTION)
M4280
M0765 M0765
2O, 3V, 1A, WAGD, 1 NITROUS 4V, 1 NITROGEN, 1 CO2, 10
OXIDE, 6 EMER DUPLEX, 4 DATA M0755 M8606 EMER DUPLEX, 4 DATA
FACE PLATES M0750 A1130 FACE PLATES
(ANESTHESIA) A1120 A1122 (EQUIPMENT)
10' - 0"
AIR CURTAIN
8' - 0" 8' - 0"
LAMINAR AIR 2436mm 2441mm
FLOW DIFFUSER
4927mm
16' - 2"
ab ab ab ab
M7490
2134mm
ab ab M1801
7' - 0"
A1122
(DOCUMEN
TATION)
8738mm
28' - 8"
ab ab TABLE
M7475
2134mm
7' - 0"
ab ab
M7801
U0100
3811mm
12' - 6"
M7801
ab ab ab ab
M0765
2O, 3V, 1A, WAGD, 1 NITROUS 4V, 1 NITROGEN, 1 CO2,
M0755 M0765
OXIDE, 6 EMER DUPLEX, 4 DATA 10 EMER DUPLEX, 4 DATA
FACE PLATES M0750 TABLE A1130 FACE PLATES
(ANESTHESIA) M1801 A1122 (EQUIPMENT)
EQ A1120 EQ
26' - 10"
8179mm
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-113
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
F3200
Clock, Battery, 12"
Diameter
A4015
Clock, Elapsed Time,
Electric
M7802
Monitor, Medical Grade
55-65"
A5212
Bracket, Television,
Wall-Mounted, Tilt/Angle
A5108
Waste Disposal Unit, Sharps
A5107
Dispenser, Glove,
Surgical/Examination, Wall
H T
Mntd
LOW WALL RETURN
AIR REGISTER
(TYPICAL)
M8910 ELEVATION 1
Cart, Surgical Case
A1014
Telephone, Wall
Mounted, 1 Line, With
F3050 Speaker
Whiteboard, Dry Erase M3150
ISOLATED POWER Distribution System,
SYSTEM Medication, Automatic
F0355
Footstool, Straight ELEVATION 2
M3080 LOW WALL RETURN
Cabinet, Instrument, CRS, 2 AIR REGISTER
Glass Door, 6 Shelf (TYPICAL)
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-115
4.10. ORNS1
SURGICAL AND ENDOVASCULAR SERVICES
(ORNS1) OPERATING ROOM, NEUROSURGICAL
ELEVATIONS
M3080 A5107
Cabinet, Instrument, Dispenser, Glove,
CRS, 2 Glass Door, 6 Surgical/Examination,
Shelf Wall Mntd
M4645
Patient Transfer Device
S9755
Suction System, Surgical,
Mobile Rover Unit
X4200
Stereotactic Surgical
System
ISOLATED POWER
SYSTEM
SD S S
LOW WALL RETURN
AIR REGISTER
(TYPICAL)
M8535 M3072
Microscope, Operating,
ELEVATION 3 Frame, Infectious Waste
Portable A5104 Bag w/Lid
E0954 Cart, Medical Waste M8920
Cart, Emergency, Mobile, Disposal, Mobile w/Foot Stand, Basin, CRS,
66"H x 32"W x 22"D Pedal Mobile, Double
M7650 M3070 M8905
Defibrillator/Monitor, Acute Hamper, Linen, Mobile, Pail, Utility, CRS, With
Care w/Lid Carriage
M4816 M8900
Warming Unit, Patient, Carriage, Pail, CRS, Without
Automatic/Manual, Air
ELEVATION 4 Pail
LOW WALL RETURN M8830
AIR REGISTER Table, Instrument/Dressing,
(TYPICAL) Mobile
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-116
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.10 ORNS1
4.10 OPERATING ROOM, NEUROSURGICAL (ORNS1)
Room Data Sheet
Notes:
1) Refer to the latest version of the VA HVAC
Design Manual for quantity and location of low
air return grilles and ceiling diffusers.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-126
4.11. ORRB1
SURGICAL AND ENDOVASCULAR SERVICES
(ORRB1) OPERATING ROOM, ROBOTICS
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-127
4.11. ORRB1
SURGICAL AND ENDOVASCULAR SERVICES
(ORRB1) OPERATING ROOM, ROBOTICS
FLOOR PLAN
26' - 10"
(8.18m)
8' - 0" CLEAR
SEMI-RESTRICTED
CORRIDOR
DOORS TO BE ON
STRETCHER ALCOVE AUTOMATIC DOOR
OPENERS
A5212 A5108 SCRUB ALCOVE
M7802 RETURN AIR
DUCT (TYPICAL)
LOW WALL
GM TV H T A4015
TV RETURN AIR
F3200 REGISTER
1
A5107 (TYPICAL)
4 2
3 IPS
6' - 0" 14' - 0"
(1.83m) (4.27m)
M8560 M8606
M8910 M8920
F0355
M4815 M8840
M4255
M8830 M4287
M3175 M8810 M8940
GM
M5512 M8950 M8551
GM
(8.74m)
M8900
28' - 8"
M8925 M8970 M8560
M8905
M8560 A1014
M4816 M3150
M8830
M3072 M8825 TABLE
M5030
A1200
M4255
LIFT AND
M4266
(4.27m)
EXTENDER RAIL
14' - 0"
M4255
M4266 M0630
M1801
M9110 M7845
M3070 M3080
M8800
(1.83m)
6' - 0"
A5108 M8940
E0954
M7650
MANIFOLD
UNI-DIRECTIONAL M8920
AIR DELIVERY
ARRAY ABOVE
IPS GM Da Db
STERILE FIELD
A5104 M4645 S9755
TABLE A5107
CLEAN CORE
SURGICAL AND
ENDOVASCULAR
SERVICES
(ORRB1)
750 NSF
69.69 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-128
4.11. ORRB1
SURGICAL AND ENDOVASCULAR SERVICES
(ORRB1) OPERATING ROOM, ROBOTICS
REFLECTED CEILING PLAN
26' - 10"
(8.18m)
8' - 0" CLEAR
SEMI-RESTRICTED
CORRIDOR
RETURN AIR
DUCT EACH DIMMABLE, LED STRIP
CORNER
FIN. CLG. 10' - 0" A.F.F. LIGHT
(TYPICAL) TYPICAL
A1122
A1130 A1120
M0765 M0750
M4280 M0755
(EQUIPMENT) M0765
4V, 1 NITROGEN, 1 CO2, A A A A (ANESTHESIA)
10 EMERG DUPLEX, 2O, 3V, 1A, WAGD,
4 DATA FACE PLATES 1 NITROUS OXIDE,
6 EMERG DUPLEX,
4 DATA FACE PLATES,
OUTLINE OF
PROCEDURE
TABLE
LED LED LED LED
M7490
LED
LED
LED
LED
LED
M7801 A1122
A
A
LED
LED
LED
LED
LED
M1801
LED LED LED LED
(DOCUMENTATION)
(8.74m)
28' - 8"
A
A
LED
LED
LED
TABLE
LED
LED
LED
LED
LED
A
A1200 M7490
LIFT AND
LED
LED
LED
LED
A
M7801
EXTENDER RAIL
LED
LED
LED
LED
LED
10' - 0"
4927mm
16' - 2"
M7490
2134mm
ab ab
7' - 0"
M7475
8738mm
28' - 8"
ab ab TABLE
M7801
2134mm
7' - 0"
ab ab
3811mm
12' - 6"
M7801
U0100 ab ab ab ab
NOTE: OPTIONAL REFLECTED CEILING PLAN IS AN ALTERNATIVE OPTION TO THE MANIFOLD UNI-DIRECTIONAL AIR DELIVERY SYSTEM.
EQUIPMENT TAGS & REFERENCE NOTES MAY NOT BE FULLY COORDINATED WITH THE EQUIPMENT LIST.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-130
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
F3200
Clock, Battery, 12"
Diameter
A5107
A4015
Dispenser, Glove,
Clock, Elapsed Time,
Surgical/Examination,
M7802 Electric
Wall Mntd
Monitor, Medical Grade
55-65"
A5212
Bracket, Television,
Wall-Mounted, Tilt/Angle
M8560
Surgical System, Robotic
LOW WALL RETURN
AIR REGISTER
(TYPICAL)
H T
A5108
Waste Disposal Unit, Sharps ELEVATION 1 F3050
Whiteboard, Dry Erase
M8840 M8940
Table, Back, Stool, Anesthesia, With
Instrument/Dressing Back
F0355 A1014
Footstool, Straight Telephone, Wall
ISOLATED POWER Mounted, 1 Line, With
SYSTEM Speaker
M3080 A5107
Cabinet, Instrument, Dispenser, Glove,
CRS, 2 Glass Door, 6 Surgical/Examination,
Shelf Wall Mntd
S9755
Suction System, Surgical,
Mobile Rover Unit
M4645
Patient Transfer Device
ISOLATED POWER
SYSTEM
SD SD
E0954 M8905
Cart, Emergency, Mobile, ELEVATION 4 Pail, Utility, CRS, With
66"H x 32"W x 22"D Carriage
M3070 M8900
Hamper, Linen, Mobile, w/Lid Carriage, Pail, CRS,
Without Pail
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-133
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.11 ORRB1
4.11 OPERATING ROOM, ROBOTICS (ORRB1)
Room Data Sheet
Notes:
1) Refer to the latest version of the VA HVAC
Design Manual for quantity and location of low
air return grilles and ceiling diffusers.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-141
4.12. ORTR1
SURGICAL AND ENDOVASCULAR SERVICES
(ORTR1) OPERATING ROOM, TRANSPLANT
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-142
4.12. ORTR1
SURGICAL AND ENDOVASCULAR SERVICES
(ORTR1) OPERATING ROOM, TRANSPLANT
FLOOR PLAN
(8.74m)
M8825
28' - 8"
M4816 M3109
M8900
M8 900
M8905 M9110 M5030
M3070 M7845
M4820
A1200
LIFT AND
EXTENDER RAIL L1095
M3080
M8800
A5108
M8830
M8940 MANIFOLD
UNI-DIRECTIONAL
AIR DELIVERY
E0954 ARRAY ABOVE
M7650 STERILE FIELD
M7818
IPS a b Dc
OPERATING ROOM,
TRANSPLANT
(ORTR1)
750 NSF
69.67 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-143
4.12. ORTR1
SURGICAL AND ENDOVASCULAR SERVICES
(ORTR1) OPERATING ROOM, TRANSPLANT
REFLECTED CEILING PLAN
26' - 10"
(8.18m)
8' - 0" CLEAR
SEMI-RESTRICTED
CORRIDOR
LED
LED
A
LED
LED
LED
LED
LED
LED
LED
LED
A1122
A
A
LED LED LED LED LED LED
M1801
(DOCUMENTATION)
LED
LED
LED
LED
OUTLINE OF
(8.74m)
LED
LED
A
A
LED
LED
28' - 8"
PROCEDURE
TABLE LED LED LED LED LED LED
M7490 TABLE
A
A
LED
LED
LED
LED
M7801 M7490
LED
LED
LED
LED
M7801
A1200 LED LED LED LED LED LED
LED
LED
LED
UNI-DIRECTIONAL
LED
LED
LED
LED
AIR DELIVERY
LED LED LED LED LED LED
ARRAY ABOVE
A1120 STERILE FIELD
M0750
M0755 GYPSUM BOARD
M0765 SOFFIT.
M1801 COORDINATE
A A A A
2O, 3V, 1A, WAGD, HEIGHT WITH
1 NITROUS OXIDE, EQUIPMENT.
6 EMERG DUPLEX,
4 DATA FACE PLATES
(ANESTHESIA) RETURN AIR
DUCT EACH CORNER
TABLE (TYPICAL)
CLEAN CORE FIN. CLG. 10' - 0" A.F.F. A1122
GENERAL NOTES: A1130
1.) PROVIDE EMERGENCY LIGHTING AS REQUIRED BY CODE(S) OPERATING ROOM, M0765
2.) REFER TO THE LEGEND FOR ADDITIONAL INFORMATION FOR THE 4V, 1 NITROGEN,
TRANSPLANT
MANIFOLD AIR DELIVERY ARRAY (ORTR1) 1 CO2
3.) ROOM IS TO BE PROTECTED WITH QUICK RESPONSE (QR) 10 EMER DUPLEX,
750 NSF
FIRE SPRINKLERS 69.67 NSM
4 DATA FACE PLATES
(EQUIPMENT)
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-144
4.12. ORTR1
SURGICAL AND ENDOVASCULAR SERVICES
(ORTR1) OPERATING ROOM, TRANSPLANT
REFLECTED CEILING PLAN (ALTERNATE OPTION)
M4280
M0765 M0765
2O, 3V, 1A, WAGD, 1 NITROUS OXIDE, 6 M0755 M8606
4V, 1 NITROGEN, 1 CO2, 10 EMER
EMER DUPLEX, 4 DATA FACE PLATES M0750 A1130 DUPLEX, 4 DATA FACE PLATES
(ANESTHESIA) A1120 A1122 (EQUIPMENT)
10' - 0"
M7475
8' - 0" 8' - 0"
AIR CURTAIN
2436mm 2440mm
LAMINAR AIR
FLOW DIFFUSER
ab ab ab ab
4927mm
16' - 2"
M7490
M1801
A1122
2134mm
ab ab (DOCUMEN
7' - 0"
TATION)
8738mm
28' - 8"
M7475
ab ab TABLE
M7801
2134mm
7' - 0"
ab ab
3811mm
12' - 6"
M7801
U0100 ab ab ab ab
(ANESTHESIA) A1120
A1122 (EQUIPMENT)
GASES: 2O, 3V, 1A, WAGD, M1801
1 NITROUS OXIDE A1130 4V, 1 NITROGEN, 1 CO2, 10 EMER
M0765 TABLE
M0765 DUPLEX, 4 DATA FACE PLATES
M0755
12' - 4 3/4" M0750 12' - 5 1/4"
3776mm 3793mm
26' - 10"
8179mm
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-145
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
F3200
Clock, Battery, 12"
Diameter
A5107
Dispenser, Glove, A4015
A5212 Surgical/Examination, Wall Clock, Elapsed Time,
Bracket, Television, Mntd Electric
Wall-Mounted, Tilt/Angle
M7802
Monitor, Medical Grade
55-65"
A5108
Waste Disposal Unit, Sharps
M8910
Cart, Surgical Case
LOW WALL RETURN
AIR REGISTER EPO
T H S
(TYP)
ELEVATION 1
ISOLATED POWER F3050
SYSTEM Whiteboard, Dry Erase
A1014 A5107
Telephone, Wall Dispenser, Glove,
Mounted, 1 Line, With Surgical/Examination,
Speaker Wall Mntd
F0355
Footstool, Straight ELEVATION 2 M3080 LOW WALL RETURN
M8940
Cabinet, Instrument, CRS, 2 AIR REGISTER
Stool, Anesthesia, With Back
Glass Door, 6 Shelf (TYP)
M3150
Distribution System,
Medication, Automatic
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-147
4.12. ORTR1
SURGICAL AND ENDOVASCULAR SERVICES
(ORTR1) OPERATING ROOM, TRANSPLANT
ELEVATIONS
M4645
Patient Transfer Device
S9755
M3080 A5107 Suction System, Surgical,
Cabinet, Instrument, Dispenser, Glove, Mobile Rover Unit
CRS, 2 Glass Door, 6 Surgical/Examination, E0963
Shelf Wall Mntd Cart, General Storage,
Mobile
M8551
Light Source, Fiberoptic
Headlamp
ISOLATED POWER
SYSTEM
U0105
ExtraCorporeal Support
System
A5104
Cart, Medical Waste
SD SD SD
Disposal, Mobile w/Foot
Pedal
M8830 M8905
Table, Instrument/Dressing, ELEVATION 4 Pail, Utility, CRS, With
Mobile Carriage
M3070 M8905
Hamper, Linen, Mobile, w/Lid Pail, Utility, CRS, With
M3072 Carriage
Frame, Infectious Waste Bag
w/Lid
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-148
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.12 ORTR1
4.12 OPERATING ROOM, TRANSPLANT (ORTR1)
Room Data Sheet
Notes:
1) Refer to the latest version of the VA HVAC
Design Manual for quantity and location of low
air return grilles and ceiling diffusers.
CONTROL
ROOM
SYSTEMS
COMPONENTS
ROOM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-158
4.13. ORHY1
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY1) OPERATING ROOM, MONOPLANE HYBRID
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-159
4.13. ORHY1
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY1) OPERATING ROOM, MONOPLANE HYBRID
FLOOR PLAN
SEMI-RESTRICTED CORRIDOR
32' - 2"
DOORS TO BE ON
(9.8m) AUTOMATIC
A5107 DOOR OPENERS
STRETCHER ALCOVE M8940 A5108 SCRUB ALCOVE RETURN AIR
DUCT
(TYPICAL)
GM IPS A4015 LOW WALL
H T
EPO F3200 RETURN AIR
1
REGISTER
A1200 (TYPICAL)
4 2
LIFT AND
M8910
MANIFOLD 20' - 0" 3 EXTENDER RAIL
UNI-DIRECTIONAL TO CONTROL
(8.74m)
(6.1m)
28' - 8"
AIR DELIVERY ROOM
M8810 M8825
ARRAY ABOVE M3175
F3050 M4255 M8800 M8920 R4785
STERILE FIELD A5108
M4250 M8825
M4266
M8920
E0948
M8900
M8905
M8606
GM
EQ
M8925 BORROWED
M5030 LIGHT
M3070
(3.66m)
A5104
12' - 0"
TABLE
F0355
M3070 M4255
L1095 M8830
M4810 M4287
M4645
EQ
GM D D
CLEAN CORE
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-160
4.13. ORHY1
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY1) OPERATING ROOM, MONOPLANE HYBRID
REFLECTED CEILING PLAN
32' - 2"
(9.8m)
(EQUIPMENT)
M7490
4V, 1 NITROGEN, 1 CO2, M7801
10 EMERG DUPLEX, 4 DATA FIN. CLG. 10' - 2" A.F.F.
FACE PLATES
A1122 A1200
A1130 LIFT AND
M0765 EXTENDER RAIL
(DOCUMEN-
TATION)
(EQUIPMENT)
A1122 4V,
M1801 A A A A
1 NITROGEN,
SEE LEGEND: 1 CO2,
MANIFOLD 10 EMERG
UNI- DUPLEX,
DIRECTIONAL 4 DATA
AIR DELIVERY FACE PLATES
EQ
(ANESTHESIA)
X6190 2O, 3V, 1A,
LED
LED
LED
LED
LED
LED
LED
LED
LED
M7860
WAGD,
A
LED
LED
LED
LED
LED
LED
LED
LED
LED
LED
A
A5212 1 NITROUS
(8.74m)
28' - 8"
M7802
LED LED LED LED LED LED
OXIDE,
ATV
A
6 EMERG
LED
LED
LED
LED
LED
LED
LED
LED
DUPLEX,
TABLE
LED
LED
LED
LED
LED
LED
LED
LED
4 DATA FACE
A TV
A
(EQUIPMENT) LED LED LED LED LED LED PLATES
4V, A1120
LED
1 NITROGEN, M0750
LED
LED
LED
LED
LED
LED
LED
LED
LED
A
1 CO2, M0755
LED
LED
LED
LED
LED
LED
LED
LED
LED
A
10 EMERG M0765
DUPLEX,
LED LED LED LED LED LED
(PERFUSION)
4 DATA FACE
2O, 3V, 2A,
PLATES
1 NITROUS
EQ
A1122
OXIDE,
A1130
M0765 6 EMERG
DUPLEX,
(ANESTHESIA) 4 DATA FACE
2O, 3V, 1A, A A A A PLATES
WAGD, A1122
1 NITROUS A1130
OXIDE, M0750
6 EMERG M0755
DUPLEX, M0765
4 DATA FACE
PLATES TABLE
GYPSUM BOARD SOFFIT. M7490
A1120
M0750 COORDINATE HEIGHT WITH M7801
M0755 EQUIPMENT.
M0765 18' - 0"
GENERAL NOTES:
(5.49m) 1.) PROVIDE EMERGENCY LIGHTING AS REQUIRED BY CODE(S)
OPERATING ROOM,
MONOPLANE HYBRID 2.) REFER TO THE LEGEND FOR ADDITIONAL INFORMATION FOR THE
(ORHY1) MANIFOLD AIR DELIVERY ARRAY
3.) ROOM IS TO BE PROTECTED WITH QUICK RESPONSE (QR)
900 NSF
83.65 NSM FIRE SPRINKLERS
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-161
4.13. ORHY1
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY1) OPERATING ROOM, MONOPLANE HYBRID
REFLECTED CEILING PLAN (ALTERNATE OPTION)
4V, 1
NITROGEN,
1 CO2,
10 EMER
DUPLEX,
4 DATA FACE
PLATES
M0765
A1130
10' - 2" A1122
(EQUIPMENT)
U0100
M7801
2O, 3V, 2A,
1 NITROGEN,
4343mm
14' - 3"
6 EMER DUPLEX,
M1801 ab ab ab ab ab ab 4 DATA FACE
A1122 PLATES
(DOCUMEN M0765
TATION) M0755
M7490 ab ab M0750
A1130
A1122
(PERFUSION)
M7860 ab TABLE
M4280
M0765
M8606
A1130
ab ab A1122
(EQUIPMENT)
2O, 3V, 1A, WAGD, 4V, 1 NITROGEN,
1 NITROUS 1 CO2,
4394mm
14' - 5"
OXIDE, 6 EMER ab ab ab ab ab ab 10 EMER DUPLEX,
DUPLEX, 4 DATA 4 DATA FACE
FACE PLATES PLATES
(ANESTHESIA)
A1120
A1120 M0765
M1801 M0755
M0765 M0750
M0755 (ANESTHESIA)
M0750 2O, 3V, 1A, WAGD,
1 NITROUS
OXIDE, 6 EMER
15' - 8 3/4" 16' - 5 1/4" DUPLEX, 4 DATA
FACE PLATES
4797mm 5007mm
M7801
TABLE M7475
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-162
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
A5107
Dispenser, Glove, F3200
Surgical/Examination, Wall Clock, Battery, 12"
Mntd Diameter
M8940 A5108 A4015
Stool, Anesthesia, With Back Waste Disposal Unit, Sharps Clock, Elapsed Time,
Electric
EPO
H T S
ELEVATION 1
M8910
Cart, Surgical Case ISOLATED POWER
SYSTEM
LOW WALL RETURN
AIR REGISTER (TYPICAL)
A1014
Telephone, Wall
Mounted, 1 Line, With
F0355 Speaker
Footstool, Straight A5107
E0948 Dispenser, Glove,
Cart, General Storage, Surgical/Examination,
Mobile, 42"H x 32"W x 22"D Wall Mntd
BORROWED
LIGHT
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-164
4.13. ORHY1
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY1) OPERATING ROOM, MONOPLANE HYBRID
ELEVATIONS
A5107 M3165
Dispenser, Glove, Cabinet, Catheter Storage
Surgical/Examination, M3150
Wall Mntd Distribution System,
Medication, Automatic
SD SD
M3150 F3050
Whiteboard, Dry Erase
Distribution System, ELEVATION 4
Medication, Automatic M8905
E0954 Pail, Utility, CRS, With
Cart, Emergency, Mobile, Carriage
66"H x 32"W x 22"D M8900
M3072 Carriage, Pail, CRS,
Frame, Infectious Waste Bag Without Pail
w/Lid
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-165
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.13 ORHY1
4.13 OPERATING ROOM, MONOPLANE HYBRID (ORHY1)
Room Data Sheet
Notes:
1) Refer to the latest version of the VA HVAC
Design Manual for quantity and location of low
air return grilles and ceiling diffusers.
2) Room does not contain multiple slot
diffusers and uses laminar flow perforated face
outlets only for supply air
3) Suggested minimum laminar flow array over
the entire sterile field area as imaging equipment
gantry creates excessive turbulence. Mechanical
design engineer shall be responsible to design
the array in such manner as to minimize
turbulence and to maintain the sterile aseptic
field
CONTROL
ROOM
SYSTEMS
COMPONENTS
ROOM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-175
4.14. ORHY2
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY2) OPERATING ROOM, BIPLANE HYBRID
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-176
4.14. ORHY2
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY2) OPERATING ROOM, BIPLANE HYBRID
FLOOR PLAN
SEMI-RESTRICTED CORRIDOR
32' - 2"
MANIFOLD (9.8m) DOORS TO BE ON
UNI-DIRECTIONAL AUTOMATIC DOOR
AIR DELIVERY OPENERS
ARRAY ABOVE STRETCHER ALCOVE A5107 RETURN AIR
SCRUB ALCOVE
STERILE FIELD M8940 A5108 DUCT
(TYPICAL)
GM IPS A4015 LOW WALL
H T
EPO F3200 RETURN AIR
1 REGISTER
(TYPICAL)
4 2
M8910
20' - 0" 3 TO CONTROL
(8.74m)
28' - 8"
(6.1m) M8920 ROOM
F3050 R4785 A1200
M4255 LIFT AND
M4250 M5030 M8825 E0948
M8900
M8 900 EXTENDER RAIL M3175
M4266 M4280
M8905 M8551
M8825
M8810
GM M8800 M8606
GM
(3.66m)
M3070
12' - 0"
A5108 M8925
BORROWED
TABLE L1095
LIGHT
A5104 F0355
X6175
M3072 M9080 M4255 M8830
M4645 M4287
M4810 M4815 X2106
M8920 M8940 M4816
M5030
E0954
M4812 X3145
M7818
U0105
M7650
M0630 M4811
M1801 M8551 M8810 M8810 M8840 S9755
M8970 M8950 M3109 M5512
M7845
COMPONENT
EQUIPMENT
M8940 L0221
IPS
18' - 0"
A1014
IMAGE
ROOM
(5.49m)
A5107
GM D D
CLEAN CORE
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-177
4.14. ORHY2
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY2) OPERATING ROOM, BIPLANE HYBRID
REFLECTED CEILING PLAN
(EQUIPMENT) (EQUIPMENT)
4V, 32' - 2" 4V,
1 NITROGEN, (9.8m) 1 NITROGEN,
1 CO2, 1 CO2,
10 EMERG A1200 10 EMERG
DUPLEX, M7490 LIFT AND DUPLEX,
4 DATA M7801 EXTENDER RAIL 4 DATA FACE
FACE PLATES PLATES
A1122 A1122
A1130 A1130
M0765 M0765
A A A A
M4280
(DOCUMEN-
TATION)
A1122
M1801
(ANESTHESIA)
2O, 3V, 1A,
WAGD,
1 NITROUS
OXIDE,
X6175 LED LED LED LED LED LED
6 EMERG
DUPLEX,
LED
LED
LED
LED
LED
LED
LED
LED
LED
4 DATA FACE
A
A
LED
LED
LED
LED
LED
LED
LED
LED
LED
LED
PLATES
A5212
LED LED LED LED LED LED
A1120
M7802 M0750
(8.74m)
A
28' - 8"
A TV TV A
M0755
LED
LED
LED
LED
LED
LED
LED
TABLE M0765
LED
LED
LED
LED
LED
LED
LED
LED
A
M7802
LED LED LED LED LED LED (PERFUSION)
A5212 2O, 3V, 2A,
1 NITROUS
LED
LED
LED
LED
LED
LED
LED
LED
LED
LED
A
A
LED
LED
LED
LED
LED
LED
LED
LED
MANIFOLD 6 EMERG
UNI- LED LED LED LED LED LED
DUPLEX,
DIRECTIONAL 4 DATA FACE
AIR DELIVERY PLATES
ARRAY ABOVE
A1122
STERILE FIELD A1130
M0750
(ANESTHESIA) M0755
2O, 3V, 1A, M0765
A A A A
WAGD,
1 NITROUS
OXIDE, M7490
6 EMERG M7801
DUPLEX,
4 DATA FACE
PLATES GYPSUM BOARD TABLE
FIN. CLG. 10' - 2" A.F.F.
A1120 SOFFIT. COORDINATE
M0750 HEIGHT WITH EQUIPMENT.
M0755 18' - 0"
GENERAL NOTES:
M0765 (5.49m) OPERATING ROOM, 1.) PROVIDE EMERGENCY LIGHTING AS REQUIRED BY CODE(S)
M1801
BIPLANE HYBRID 2.) REFER TO THE LEGEND FOR ADDITIONAL INFORMATION FOR THE
(ORHY2) MANIFOLD AIR DELIVERY ARRAY
900 NSF 3.) ROOM IS TO BE PROTECTED WITH QUICK RESPONSE (QR)
83.67 NSM FIRE SPRINKLERS
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-178
4.14. ORHY2
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY2) OPERATING ROOM, BIPLANE HYBRID
REFLECTED CEILING PLAN (ALTERNATE OPTION)
4V, 1 NITROGEN,
1 CO2, 10 EMER
DUPLEX, 4 DATA
FACE PLATES
M0765
10' - 2"
A1130
A1122
(EQUIPMENT)
U0100
M7801
M7490
4369mm
2O, 3V, 2A, 1
14' - 4"
ab ab ab ab ab NITROGEN, 6 EMER
DUPLEX, 4 DATA
FACE PLATES
M1801
M0765
A1122
ab M0755
(DOCUMEN ab
M0750
TATION)
A1130
A1122
(PERFUSION)
M7860 ab TABLE
M4280
M0765
M8606
2O, 3V, 1A, WAGD, A1130
1 NITROUS ab
ab A1122
OXIDE, 6 EMER
DUPLEX, 4 DATA (EQUIPMENT)
FACE PLATES 4V, 1 NITROGEN,
1 CO2, 10 EMER
4369mm
(ANESTHESIA)
14' - 4"
ab ab ab ab ab DUPLEX, 4 DATA
A1120 FACE PLATES
M1801
M0765 A1120
M0755 M0765
M0750 M0755
M0750
(ANESTHESIA)
2O, 3V, 1A, WAGD,
1 NITROUS
OXIDE, 6 EMER
DUPLEX, 4 DATA
13' - 8" 18' - 6" FACE PLATES
4165mm 5639mm
M7475
TABLE M7801
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-179
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
A4015
Clock, Elapsed Time, Electric
F3200
Clock, Battery, 12" Diameter
ISOLATED POWER
SYSTEM
EPO
H T S
BORROWED
LIGHT
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-181
4.14. ORHY2
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY2) OPERATING ROOM, BIPLANE HYBRID
ELEVATIONS
M3165
A5107 Cabinet, Catheter Storage
Dispenser, Glove, M3150
Surgical/Examination, Distribution System,
Wall Mntd Medication, Automatic
SD SD
M3080
Cabinet, Instrument, CRS, 2
Glass Door, 6 Shelf
M8920
M7650 Stand, Basin, CRS,
Defibrillator/Monitor, Acute Mobile, Double
Care M4645
M4812 M7802 Patient Transfer Device
Pacemaker, Single Chamber, Monitor, Medical Grade M3070
External, Temporary 55-65" Hamper, Linen, Mobile,
M7818 A5212 w/Lid
Monitor, Transport Bracket, Television,
ISOLATED POWER Wall-Mounted, Tilt/Angle
SYSTEM
F3050
M3150 Whiteboard, Dry Erase
Distribution System, ELEVATION 4 M8905
Medication, Automatic Pail, Utility, CRS, With
E0954 Carriage
Cart, Emergency, Mobile,
M8900
66"H x 32"W x 22"D Carriage, Pail, CRS,
M3072 Without Pail
Frame, Infectious Waste Bag
w/Lid
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-182
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.14 ORHY2
4.14 OPERATING ROOM, BIPLANE HYBRID (ORHY2)
Room Data Sheet
Notes:
1) Refer to the latest version of the VA HVAC
Design Manual for quantity and location of low
air return grilles and ceiling diffusers.
2) Room does not contain multiple slot
diffusers and uses laminar flow perforated face
outlets only for supply air
3) Suggested minimum laminar flow array over
the entire sterile field area as imaging equipment
gantry creates excessive turbulence. Mechanical
design engineer shall be responsible to design
the array in such manner as to minimize
turbulence and to maintain the sterile aseptic
field
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-193
SURGICAL AND ENDOVASCULAR SERVICES 4.15 ORHC1
(ORHC1) CONTROL ROOM, HYBRID OR, CATHETERIZATION LAB
ELECTROPHYSIOLOGY LAB, INTERVENTIONAL RAD LAB, VASCULAR LAB
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-194
SURGICAL AND ENDOVASCULAR SERVICES 4.15 ORHC1
(ORHC1) CONTROL ROOM, HYBRID OR, CATHETERIZATION LAB
ELECTROPHYSIOLOGY LAB, INTERVENTIONAL RAD LAB, VASCULAR LAB
FLOOR PLAN
9' - 0"
(2.73m)
F3200 SEMI-RESTRICTED
Clock, Battery, 12" Diameter CORRIDOR
A5145
D
Hook, Garment, Double, SS,
Surface Mounted
TYP.
T
TO PROCEDURE
ROOM
X3150
Rack, Apron/Gloves, Wall
Mounted
A1012
(8.84m)
29' - 0"
M1801
Computer, Microprocessing,
1
w/Flat Panel Monitor
M7860 2
Monitoring System, Cardiac
Catheterization Lab
F3050
M1825
Whiteboard, Dry Erase
Printer, Computer
M1801
Computer, Microprocessing,
w/Flat Panel Monitor X4112
CT030 Console, PACS, Remote
Countertop, High Pressure View, w/Two 2MP
Laminate Monitors
X1425
Imager, Laser (1024 X
1024) (Din/PACS)
CONTROL ROOM,
HYBRID OR, CATH LAB,
EP LAB, IR LAB AND
VASCULAR LAB
(ORHC1)
260 NSF
CONSULT MANUFACTURE WALL REINFORCEMENT SPECIFICATIONS 24.16 NSM
FOR OBJECT: X3150 - RACK, APRON/GLOVES, WALL MOUNTED.
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-195
SURGICAL AND ENDOVASCULAR SERVICES 4.15 ORHC1
(ORHC1) CONTROL ROOM, HYBRID OR, CATHETERIZATION LAB
ELECTROPHYSIOLOGY LAB, INTERVENTIONAL RAD LAB, VASCULAR LAB
REFLECTED CEILING PLAN
9' - 0"
(2.73m)
(8.84m)
29' - 0"
CONTROL ROOM,
HYBRID OR, CATH LAB,
EP LAB, IR LAB AND
VASCULAR LAB
(ORHC1)
260 NSF
24.16 NSM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-196
SURGICAL AND ENDOVASCULAR SERVICES 4.15 ORHC1
(ORHC1) CONTROL ROOM, HYBRID OR, CATHETERIZATION LAB
ELECTROPHYSIOLOGY LAB, INTERVENTIONAL RAD LAB, VASCULAR LAB
ELEVATIONS
F3200
Clock, Battery, 12" Diameter
X3150
Rack, Apron/Gloves, Wall
Mounted
S S SD
ELEVATION 1
X3150
Rack, Apron/Gloves, Wall
Mounted
A5145
Hook, Garment, Double, SS,
A6110 A1012 Surface Mounted
Counter, Console, Telephone, Wall Mounted, 1 TYP.
Communications Line
BORROWED
BORROWED LIGHT
LIGHT
X4112 M1825
Console, PACS, Remote Printer, Computer F0275
View, w/Two 2MP Chair, Swivel, High Back
CT030
Monitors Countertop, High M1801
Pressure Laminate Computer,
ELEVATION 2 M1801
Microprocessing, w/Flat
Panel Monitor
Computer, M7860 F2000
Microprocessing, w/Flat Monitoring System, Basket, Wastepaper, Fire
Panel Monitor Cardiac Catheterization Resistant
Lab
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-197
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
4.15 ORHC1
4.15 CONTROL ROOM, HYBRID OR (ORHC1)
Room Data Sheet
ARCHITECTURAL COMMUNICATIONS
Ceiling Type: Gypsum Wallboard (SC) Data: Yes
Ceiling Height: 9’-0” (2700 mm) Telephone: Yes
Ceiling Finish: Cable Television: No
Wall Finish: Gypsum Wallboard (SC) Duress Alarm: No
Wainscot: Electronic Access and Door Control: No
Base: RF Integral Base (min. 6”/ Intercom: No
152 mm) Motion Intrusion Detection (MID): No
Floor Finish: Resinous Flooring Nurse Call: No
Slab Depression: None Code Blue:
Sound Protection: None Public Address: Yes
Doors: Single, Size 3’-0” x 7’-0” Security Surveillance Television (SSTV): No
(914 mm x 2133 mm) Wood VA Satellite TV: No
Special Requirement: Video Teleconferencing (VTEL): No
Notes:
1) Shielding is to be provided in the wall to the
HEATING, VENTILATING AND AIR
hybrid operating room.
CONDITIONING
2) Consult with equipment manufacturer for
countertop requirements. General Requirement: Refer to Controls &
Communication Center Room data sheet in the
current version of the VA HVAC Design Manual
for room temperatures, humidity range, room air
LIGHTING change requirements, and pressurization
Refer to the VA Lighting Design Manual section
4.2.13 – Surgery/Operating Room – Design PLUMBING AND MEDICAL GASES
Parameter – Control Room - for lighting design Cold Water: No
consideration.
Hot Water: No
Waste: No
Reagent Grade Water: No
POWER Medical Air No
Normal Power: To be connected to selected Medical Vacuum No
receptacles and equipment. Oxygen No
Emergency Power: Critical branch of the EES to
be connected to selected
receptacles and equipment. FIRE PROTECTION AND LIFE SAFETY
Notes: Fire Alarm: Yes
1) Provide a duplex receptacle per computer Sprinkler: Yes
and location connected to Emergency. Hazard Type: Light Hazard
2) Provide convenience duplex receptacle on
normal power.
Axonometric
850 NSF / 79,0 NSM
SEMI-RESTRICTED
CORRIDOR
M7802 SCRUB ALCOVE
F3200 A1014 A5212 RETURN AIR
DUCT (TYPICAL)
b GM TV
T H
A5077 C LOW WALL
RETURN AIR
A5107 L1085 REGISTER
X-RAY IN USE M3150 E0703 (TYPICAL)
IPS
SIGNAGE
TO R6200 M3155
L0220
CONTROL M4092
GLUCOMETER
ROOM
M8810 E0954
M4812 M8940 M8940 M7650
M3175 M7845
M8825 A1014
M8800
F0355 A1130
BORROWED
LIGHT 2V, 2A, 2O,
M4255 1 NITROGEN
GM
M4266
8738mm
28' - 8"
M3072 M4250
M3070 M4645
A5104
GM
F2020 X6190
X2125 M4645
LAMINAR FLOW (PART OF
M8826 SYSTEM)
STERILE FIELD M7818
PERIMETER AIR
1 M4255
CURTAIN ABOVE
M8940 M4266
4 2
X2106
F3050 3 X6195
A5107 M3155
A5077
IPS
Da Db Dc GM
30' - 4"
9245mm
CLEAN CORE
(DOCUMENTATION)
U0100
10' - 2"
M7801
M1801 c c c c
A1122
2A, 2V,
2O,1NITROGEN
4572mm
15' - 0"
10 EMER DUPLEX,
4 DATA FACE PLATES, ab ab ab ab ab ab
LASER OUTLET
M7801
M0765
M0755 ab ab
1826mm
M0750
6' - 0"
M7860
A1122
8738mm
28' - 8"
A1130
(EQUIPMENT) ab
TABLE
LAMINAR
AIR FLOW
1829mm
6' - 0"
DIFFUSER
ab ab
4166mm
13' - 8"
ab ab ab ab ab ab
14' - 0"
4267mm
c c c c
30' - 4"
9245mm
JSN Legend
JSN DESCRIPTION M7802 MONITOR, HD, LCD, FP, MEDICAL
GRADE, 55 INCH
A1014 TELEPHONE, WALL MOUNTED, 1 LINE, M7818 MONITOR, TRANSPORT
WITH SPEAKER M7860 MONOTORING SYSTEM, CARDIAC
A1122 COLUMN, EQUIPMENT ARM, CEILING CATHETERIZATION LAB
MOUNTED, SURGERY M8800 CART, ANESTHESIA
A1130 CABINET, CONTROL, NITROGEN M8810 STAND, MAYO
A5077 DISPENSER, HAND SANITIZER, HANDS M8825 TABLE, INSTRUMENT/DRESSING
FREE M8826 TABLE, INSTRUMENT/DRESSING CRS,
A5104 CART, WASTE DISPOSAL, MOBILE W/ 60X24X34
FOOT PEDAL M8940 STOOL, ANESTHESIA, WITH BACK
A5107 DISPENSER, GLOVE, WALL-MTD R6200 REFRIGERATOR, U/C OR F/S, 5 CU FT
A5212 BRACKET, TELEVISION, WALL-MTD, U0100 OR INTEGRATION
TILT/ANGLE X2106 SCANNER, ULTRASOUND, INTRA-
E0703 TABLE, PROCESS, 5 DRAWERS CARDIAC ECHO
E0954 CART, EMERGENCY, MOBILE X2125 SCANNER, ULTRASOUND, PORTABLE
F0355 FOOTSTOOL, STRAIGHT X6190 RADIOGRAPHIC/FLOURO UNIT,
F2020 CAN, TRASH, 44 GALLON CARDIAC, 100kW, DIGITAL
F3050 WHITE BOARD, DRY ERASE X6195 INJECTOR, ANGIOGRAPHIC
F3200 CLOCK, BATTERY, 12IN
L0220 ANALYZER, HEMOGLOBIN, PORTABLE,
HAND HELD
L1085 ANALYZER, COAGULATION, AUTO
M0750 FLOWMETER, AIR”
M0755 FLOWMETER, OXYGEN, LOW FLOW
M0765 REGULATOR, VACUUM
M1801 COMPUTER, MICROPROCESSING
W/ FLAT PANEL MONITOR
M3070 HAMPER, LINEN
M3072 FRAME, INFECTIOUS WASTE BAG W/LID
M3150 DISTRIBUTION SYSTEM, MEDICATION,
AUTOMATIC
M3155 CABINET, INVENTORY MANAGEMENT
SYSTEM
M3175 ELECTROSURGICAL UNIT, DUAL OUPUT
M4250 PUMP SYRINGE, INFUSION
M4255 STAND IV
M4266 PUMP, VOLUMETRIC, INFUSION,
MULTIPLE LINES
M4645 PATIENT TRANSFER DEVICE
M4812 PACEMAKER, SINGLE CHAMBER,
ETERNAL, TEMPORARY
M7480 LIGHT, SURGICAL, CEILING, SINGLE,
MEDIUM
M7650 DEFIBRILLATOR/ MONITOR/ RECORDER
AUTO
M7801 MONITOR, HD, LCD, FP, MEDICAL
GRADE, 26 INCH
Elevations 1 & 2
DISPENSER, GLOVE, ANALYZER,
WALL-MTD COAGULATION, AUTO
LOW WALL RETURN
ROOM TEMPLATES
Surgical and Endovascular Services Design Guide
Endovascular Services
A1130 DEVICE
CABINET, CONTROL, M7818
NITROGEN MONITOR, TRANSPORT
ISOLATION POWER X2106
SYSTEM
4.16. CARDIAC CATHETERIZATION LABORATORY (XCCE1)
SCANNER, ULTRASOUND,
M7650 INTRA-CARDIAC, ECHO
DEFIBRILLATOR/ M0750
MONITOR/ RECORDER
INTERVENTIONAL RADIOLOGY (IR) LABORATORY (XACR1)
FLOWMETER, AIR
AUTO
M0755
E0954
FLOWMETER, OXYGEN,
CART, EMERGENCY, LOW FLOW
MOBILE
M0750
A1014
FLOWMETER, AIR
TELEPHONE, WALL
MOUNTED, 1 LINE, WITH
SPEAKER
4-205
April 2016 (rev 7/17)
Elevations 3 & 4
SANITIZER, HANDS FREE
A5107
ELEVATION 3
F2020
CAN, TRASH, 44 GALLON 28' - 8"
F3050 8738mm
WHITE BOARD, DRY A5107
ERASE DISPENSER, GLOVE,
ROOM TEMPLATES
WALL-MTD
Surgical and Endovascular Services Design Guide
A5107
Endovascular Services
DISPENSER, GLOVE, A5077
WALL-MTD DISPENSER, HAND
A5077 SANITIZER, HANDS FREE
DISPENSER, HAND BORROWED LIGHT
4.16. CARDIAC CATHETERIZATION LABORATORY (XCCE1)
SANITIZER, HANDS
FREE LOW WALL RETURN
LOW WALL RETURN AIR REGISTER
AIR REGISTER
F0355
INTERVENTIONAL RADIOLOGY (IR) LABORATORY (XACR1)
4-206
April 2016 (rev 7/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
4.16. CARDIAC CATHETERIZATION LABORATORY (XCCE1)
INTERVENTIONAL RADIOLOGY (IR) LABORATORY (XACR1)
Equipment List
Axonometric
900 NSF / 83,7 NSM
SEMI-RESTRICTED
CORRIDOR
A1014
M7802
F3200 A5212 E0703 SCRUB ALCOVE
RETURN AIR
DUCT (TYPICAL)
GM TV
T H
A5107 LOW WALL
RETURN AIR
C
A5077 REGISTER
(TYPICAL)
L1085 M3155
M3150 M7845
IPS
L0220 M0630
TO R6200 X2106 GLUCOMETER
CONTROL M3150 M8940
ROOM M4812 E0954
M8810 M7650
M8940
M8825 A1014
F0355 M3175
M8800 X3145
BORROWED M8940
GM
A5108
LIGHT
M4255
M3072
M4266
A1130
8738mm
M4250
28' - 8"
M3070 2V, 2A, 2O,
1 NITROGEN
A5104
GM
F2020
X6175
X2125
M8826
LAMINAR FLOW
STERILE FIELD M4645
PERIMETER AIR M4255
CURTAIN ABOVE M4266 M7818
F3050 1 M4086
M4645
4 2
3
X6195 M3155
A5107
A5077
Dc IPS GM
32' - 0"
9754mm
CLEAN CORE
4572mm
M1801
15' - 0"
(ANESTHESIA)
ab ab ab ab ab ab
ab ab
1829mm
(PART OF
6' - 0"
SYSTEM)
X6175
8738mm
28' - 8"
LAMINAR AIR ab TABLE
FLOW DIFFUSER
1829mm
6' - 0"
ab ab
4166mm
13' - 8"
ab ab ab ab ab ab
c c c c
10' - 2"
A1122
A1130
M0765
(EQUIPMENT - DEVICES) M0755 MONITOR BANK (PART OF X6175)
2O, 2V, 2A, 1 NITROGEN, M0750 M7475
6 EMER DUPLEX, 4 M3190
DATA FACE PLATES M7801 32' - 0"
9754mm
Elevations 1 & 2
FREE MANAGEMENT SYSTEM
LOW WALL RETURN LOW WALL RETURN
28' - 8"
A1014 8738mm
TELEPHONE, WALL
MOUNTED, 1 LINE, WITH
M0755
SPEAKER
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
FLOWMETER, OXYGEN,
M7650 LOW FLOW
DEFIBRILLATOR/ M0765
MONITOR/ RECORDER
REGULATOR, VACUUM
AUTO
M4645
4-222
April 2016 (rev 2/17)
ISOLATION
X6195 POWER SYSTEM
INJECTOR, A5077
ANGIOGRAPHIC
DISPENSER, HAND
M3155 SANITIZER, HANDS FREE
Elevations 3 & 4
CABINET, INVENTORY A5107
MANAGEMENT SYSTEM DISPENSER, GLOVE,
ELEVATION 3
28' - 8"
8738mm
F3050
M3072
WHITE BOARD, DRY
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
ERASE FRAME, INFECTIOUS
WASTE BAG W/LID
A5107
A5077
DISPENSER, GLOVE,
WALL-MTD DISPENSER, HAND
A5077 SANITIZER, HANDS FREE
4-223
April 2016 (rev 2/17)
LIGHTING Notes:
Refer to the VA Lighting Design Manual section 1) Provide IPS power & ground modules – 3
4.2.13 - Surgery/Operating Room - for lighting duplex receptacles & 3 ground jacks
design consideration. 2) IPS Power & ground modules mounted at
+24” AFF
3) Provide Laser Receptacle Module. Module
shall be connected to Special Equipment IPS
located outside the Surgery Room.
4) Provide power connections for articulating
utility columns.
COMMUNICATIONS
Data: Yes
Telephone: Yes
Cable Television: No
Duress Alarm: No
Electronic Access and Door Yes
Control:
Intercom: Yes (Phone)
Motion Intrusion Detection No
(MID):
Nurse Call: Yes
Code Blue: Yes
Public Address: No
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: No
Video Teleconferencing No
POWER (VTEL):
Normal Power: Connect selected recep- Special Requirement:
tacles and equipment to
Notes:
Normal power IPS.
1) Provide connections for articulating utility
Emergency Power: Connect selected recep-
columns.
tacles and equipment to
Critical Branch emergency 2) Provide connections for video monitor
IPS. pendants. Video monitor pendants will be part of
the video integration system. The extent of the
system is to be selected on a project basis.
Axonometric
350 NSF / 32,6 NSM
17' - 3"
5258mm
A5082 F3200 SEMI-RESTRICTED
CORRIDOR
F2010 P3100
Da
M3072 A5145
GFI
TO SCOPE
M3070 A5075 STORAGE
ROOM
F2017 A5107
A5077
A5180
X2105
M8940
M7401
1
A1132
2A,2V,2O 4 2
6096mm
20' - 0"
M4255 NC 3 F0355
M4266 M9006
M4116
F0206 TO SOILED ROOM OR
A5108 SCOPE PROCESSING
M3150 U4093 ROOM
E0954 R6200
M7650 M3150 F0280
M1801
E0948 A1015
E0210
A5180
3048mm
10' - 0"
9' - 0"
6096mm
20' - 0"
TABLE
3048mm
10' - 0"
DIMMABLE LIGHTING
17' - 3"
5258mm
Elevations 1 & 2
M3072 A5107
FRAME, INFECTIOUS DISPENSER, GLOVE,
ELEVATION 1
20' - 0"
6096mm
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
M7401
4.18. TRANSESOPHAGEAL ECHOCARDIOGRAPH (TEE)
E0210
LIGHT, EXAM, MOBILE
WORKSURFACE,
ELEVATION 2
4-240
April 2016 (rev 2/17)
M3155 17' - 3"
5258mm M3150
CABINET, INVENTORY
MANAGEMENT SYSTEM DISTRIBUTION SYSTEM, MEDICATION, AUTOMATIC
E0210 R6200
WORKSURFACE, REFRIGERATOR, U-C OR
W/OVERHEAD CAB, F-S, 5 CU FT
WALL MTD, 48" W
M1801 M3150
Elevations 3 & 4
COMPUTER, DISTRIBUTION SYSTEM,
MICROPROCESSING, W/ MEDICATION, AUTOMATIC
M4266
PUMP, VOLUMETRIC,
INFUSION, MULTIPLE
M0755 LINES
FLOWMETER, OXYGEN, M0755
LOW FLOW
20' - 0"
6096mm FLOWMETER, OXYGEN,
M0750
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
LOW FLOW
FLOWMETER, AIR M0750
M0765
FLOWMETER, AIR
4.18. TRANSESOPHAGEAL ECHOCARDIOGRAPH (TEE)
REGULATOR, VACUUM
M0765
M4116
4-241
April 2016 (rev 2/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
ARCHITECTURAL LIGHTING
Ceiling Type: Acoustical Ceiling Tile (SP) Refer to the VA Lighting Design Manual section
Ceiling Height: 9’-0” (2700mm) 4.2.5 - Procedure Room - for lighting design
consideration.
Ceiling Finish:
Wall Finish: Gypsum Wallboard (SC)
Wainscot:
Base: WSF Integral Base (min. 6”/
152 mm)
Floor Finish: Welded Seam Sheet
Flooring
Slab Depression: None
Sound Protection: 50 STC (to other room), 35
STC (to corridor)
Doors: Double, Size 6’-0” x 7’-0”
(1829 mm x 2133 mm) Wood
w/ Narrow View Window;
Single, Size 4’-0” x 7’-0”
(1219 mm x 2133 mm) Wood
w/ Small View Window
Special Requirement:
Notes:
1) Facility will select number and types of
scopes and other instrumentation as necessi-
tated by the unique case load.
2) Scope Cleaning can occur in SPD or in the
Equipment Cleaning Room. Facility to select
adjacency of this room to either the Equipment
Cleaning Room or a Soiled Hold Room, where
scopes can be stored until pick-up and process-
ing by SPD. POWER
Normal Power: To be connected to selected
receptacles and equipment.
Emergency Power: Critical branch of the
EES to be connected to
selected receptacles and
equipment.
COMMUNICATIONS
Data: Yes
Telephone: Yes
Cable Television: No
Duress Alarm: No
Electronic Access and Door Yes
Control:
Intercom: Yes (Phone)
Motion Intrusion Detection No
(MID):
Nurse Call: Yes
Code Blue: Yes
Public Address: No
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: No
Video Teleconferencing No
(VTEL):
Special Requirement:
Notes:
Axonometric
165 NSF / 15,5 NSM
A5082
9' - 6"
DISPENSER, PAPER
A5107 2896mm TOWEL, SENSOR, HANDS
DISPENSER, GLOVE, FREE
WALL-MTD P3100
LAVATORY, VITREOUS
A5077 CHINA, SLAB TYPE
DISPENSER, HAND A5075
SANITIZER, HANDS
6" AFC GFI DISPENSER, SOAP
FREE
2286mm
F2010
7' - 6"
CT050 CLEAN
STORAGE BASKET, WASTEPAPER,
COUNTER, STAINLESS 3 STEP-ON
STEEL
OSa
S1905
M3161
SLIDING SERVICE
CABINET, STORAGE, WINDOW
ULTRASOUND, TEE PROBE S2610
DISINFECTOR, AUTOMATED,
M2100 6" AFC GFI PROBE, TEE
1 A5082
CART, SHELVING,
STORAGE, MOBILE, SS DISPENSER, PAPER
2 TOWEL, SENSOR, HANDS
A5107 FREE
3048mm
10' - 0"
A5077 A1195
6" AFC GFI
DISPENSER, HAND COUNTER, CLEAN-UP,
SANITIZER, HANDS WITH 2 OR 3 SINKS
FREE F2020
CAN, TRASH, 44 GALLON
9' - 0"
CLEAN STORAGE
2286mm
7' - 6"
a
5486mm
9' - 0" 18' - 0"
DECONTAMINATION
3048mm
10' - 0"
S a
9' - 6"
2896mm
Elevations 1 & 2
165 NSF / 15,5 NSM
M2100
CART, SHELVING, STORAGE, MOBILE, SS
A1195
COUNTER, CLEAN-UP, WITH 2 OR 3 SINKS
ELEVATION 1
STORAGE, CLEAN TEE PROBE (TRTE3)
10' - 0"
3048mm
CD030
CABINET, W-H, 2 SHELF,
4.19. DECONTAMINATION, TEE PROBE (TRTE2)
2 DO
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
A5082
DISPENSER, PAPER
TOWEL, SENSOR, A5075
HANDS FREE DISPENSER, SOAP
ELEVATION 2
SCALE: 3/16” = 1’-0”
4-252
April 2016 (rev 2/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 2/17)
TEE PROBE
M3161
S1905
ELEVATION 3
2896mm
9' - 6"
16'
COUNTER, STAINLESS
AUTOMATED, PROBE,
STEEL
DISINFECTOR,
TEE
CT050
S2610
8'
SCALE: 3/16” = 1’-0”
4'
Elevation 3
165 NSF / 15,5 NSM
ARCHITECTURAL POWER
Ceiling Type: Gypsum Wallboard (SC) Normal Power: To be connected to selected
Ceiling Height: 9’-0” (2700mm) receptacles and equipment.
Ceiling Finish: Emergency Power: Critical branch of the EES
to be connected to selected
Wall Finish: Gypsum Wallboard (SC)
receptacles and equipment.
Wainscot:
Base: Resilient Base
Floor Finish: Rubber Flooring, Welded
Seam Sheet Flooring
Slab Depression: None COMMUNICATIONS
Data: No
Sound Protection: None
Telephone: No
Doors: Single, Size 3’-0” x 7’-0” (914
mm x 2133 mm) Wood Cable Television: No
Duress Alarm: No
LIGHTING Electronic Access and Door No
Refer to the VA Lighting Design Manual section Control:
4.2.14 - Surgical Corridor and Scrub Area - for Intercom: No
lighting design consideration. Motion Intrusion Detection No
(MID):
Nurse Call: No
Code Blue: No
Public Address: Yes
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: No
Video Teleconferencing No
(VTEL):
Special Requirement: No
Axonometric
185 NSF / 17,2 NSM
CLEAN CORE
P3100
12' - 0"
3658mm LAVATORY, VITREOUS
CHINA, SLAB TYPE
A5075
DISPENSER, SOAP
A1012
TELEPHONE, WALL A5082
MOUNTED, 1 LINE
DISPENSER, PAPER
CT050 TOWEL, SENSOR, HANDS
COUNTER, STAINLESS FREE
3a
STEEL
S0125
STERILIZER, STM, VAC,
1DO, CAB
54" AFF
4724mm
15' - 6"
1
A5107
CT050 6" AFC
GFI DISPENSER, GLOVE,
COUNTER, STAINLESS WALL-MTD
STEEL
M3070
CS150
HAMPER, LINEN
SINK, SS, SINGLE
COMPARTMENT
F2010
BASKET, WASTEPAPER, T 3
STEP-ON
A5077
DISPENSER, HAND
SANITIZER, HANDS SEMI-RESTRICTED
FREE CORRIDOR
A5107
DISPENSER, GLOVE,
WALL-MTD
9' - 0"
2438mm
8' - 0"
4724mm
DT S 15' - 6"
a
2134mm
7' - 0"
12' - 0"
3658mm
S0125
CA020 STERILIZER, STM, VAC, 1DO, CAB
CABINET, OPEN, W-H, 2 SHELF, 38X24X13
CS150
SINK, SS, SINGLE COMPARTMENT
CT050
C0040
COUNTER, STAINLESS STEEL
RAIL, APRON, 4x54x1
C05F0
CT050
ROOM TEMPLATES
C03F0
4.20. IMMEDIATE USE STERILIZATION ROOM (ORSR1)
ELEVATION 1
4-261
April 2016 (rev 2/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
ARCHITECTURAL POWER
Ceiling Type: Gypsum Wallboard (SC) Normal Power: To be connected to selected
Ceiling Height: 9’-0” (2700mm) receptacles and equipment.
Ceiling Finish: Emergency Power: Critical branch of the EES
to be connected to selected
Wall Finish: Gypsum Wallboard (SC)
receptacles and equipment.
Wainscot:
Notes:
Base: RF Integral Base (min. 6”/
1) Connect sterilizer to emergency power -
152 mm)
Critical Branch of the EES.
Floor Finish: Resinous Flooring
Slab Depression: None Special
Sound Protection: None
Doors: Single Door, Size 3’-8” x COMMUNICATIONS
7’-0” (1117 mm x 2133 mm) Data: Yes
Wood w/ Half View Window
Telephone: Yes
LIGHTING Cable Television: No
Refer to the VA Lighting Design Manual section Duress Alarm: No
7.6 - Sterile Processing and Distribution Clean Electronic Access and Door No
Side - for lighting design consideration. Control:
Intercom: No
Motion Intrusion Detection No
(MID):
Nurse Call: No
Code Blue: No
Public Address: Yes
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: No
Video Teleconferencing No
(VTEL):
Special Requirement: No
Notes:
Special Requirement:
Notes:
1) Canopy hood exhaust provided at sterilizer.
Axonometric
110 NSF / 10,3 NSM
A5180 M4665
TRACK, CUBICLE, STRETCHER, RECOVERY,
SURFACE MOUNTED, SURGICAL
WITH CURTAIN E0948
F2010 CART, GENERAL
BASKET, WASTEPAPER, 1 STORAGE, MOBILE
STEP-ON M7040
M3072 TABLE, OVERBED
3353mm
11' - 0"
FRAME, INFECTIOUS M4266
WASTE BAG W/LID PUMP, VOLUMETRIC,
A5108 INFUSION, MULTIPLE
LINES
WASTE DISPOSAL UNIT,
SHARPS M4255
M7845 STAND IV
MONITOR, M1801
PHYSIOLOGICAL, COMPUTER,
BEDSIDE MICROPROCESSING, W/
NC
9' - 0"
3353mm
11' - 0"
ab
10' - 0"
3048mm
ROOM TEMPLATES
STAND IV
ELEVATION 1
4.21. PATIENT BAY, PACU / PHASE I RECOVERY (RRBP1)
4-269
April 2016 (rev 2/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
ARCHITECTURAL
Ceiling Type: Acoustical Ceiling Tile (SP)
Ceiling Height: 9’-0” (2700mm)
Ceiling Finish:
Wall Finish: Gypsum Wallboard (SC)
Wainscot:
Base: WSF Integral Base (min. 6”/
152 mm)
Floor Finish: Welded Seam Sheet
Flooring
Slab Depression: None
Sound Protection: 35 STC (Room)/ None (Bay) POWER
Doors: Sliding Glass Door, Size 8’- Normal Power: To be connected to selected
0” x 7’-0” (2438 mm x 2133 receptacles and equipment.
mm) ; Single, Size 3’-0” x Emergency Power: Critical branch of the EES
7’-0” (914 mm x 2133 mm) to be connected to selected
Wood w/ Half View Window receptacles and equipment.
Special Requirement:
Notes:
1) Verify with life safety requirements that sliding
glass doors can break out of the room
2) Sliding door and footwall can be omitted to
create an open patient cubicle if desired by the
facility
LIGHTING
COMMUNICATIONS
Refer to the VA Lighting Design Manual section Data: Yes
4.2.15 - Pre-Operative and Post-Anesthetic Care Telephone: Yes
(PACU) - for lighting design consideration.
Cable Television: No
Duress Alarm: No
Electronic Access and Door No
Control:
Intercom: No
Motion Intrusion Detection No
(MID):
Nurse Call: Yes
Code Blue: Yes
Public Address: No
Axonometric
150 NSF / 14,0 NSM
M4665
A5180
a b T
A5145
P3100
A1012
1524mm
5' - 0"
F0205
GFI
1
F2010
F3200
2 A5075
A5145 A5080
A1012 c
M7040
A5077
A5077
BORROWED
LIGHT E0948
3
CS140
2134mm
7' - 0"
A5108
CT020
A5107
A5107
LOW WALL RETURN
M3070 6" AFC AIR REGISTER
GFI
A5080
NC
M1801
A5075 M4255
M3072 M4266
A1107 M7845
6' - 6" 12' - 6"
1981mm 3810mm
A5180
TRACK, CUBICLE,
SURFACE MOUNTED,
WITH CURTAIN
9' - 0"
3658mm
12' - 0"
9' - 0"
2134mm
ab
7' - 0"
JSN DESCRIPTION
A1012 TELEPHONE, WALL MOUNTED, 1 LINE
A1107 RAIL SYSTEM, UTILITY, GAS
AND ELECTRIC
A5075 DISPENSER, SOAP
A5077 DISPENSER, HAND SANITIZER,
HANDS FREE
A5080 DISPENSER, PAPER TOWEL
A5107 DISPENSER, GLOVE, WALL-MTD
A5108 WASTE DISPOSAL UNIT, SHARPS
A5145 HOOK, GARMENT, DOUBLE
A5180 TRACK, CUBICLE, SURFACE
MOUNTED, WITH CURTAIN
CS140 SINK, SS,GENERAL, SINGLE X,
10X14X ID
E0948 CART, GENERAL STORAGE, MOBILE
F0205 SIDE CHAIR WITH ARMS
F2010 BASKET, WASTEPAPER, STEP-ON
F3200 CLOCK, BATTERY, 12IN
M0750 FLOWMETER, AIR
M0755 FLOWMETER, OXYGEN, LOW FLOW
M0765 REGULATOR, VACUUM
M1801 COMPUTER, MICROPROCESSING,
W/ FLAT PANEL MONITOR
M3070 HAMPER, LINEN
M3072 FRAME, INFECTIOUS WASTE BAG
W/LID
M4255 STAND IV, ADJUSTABLE
M4266 PUMP, VOLUMETRIC, INFUSION,
MULTIPLE LINES
M4665 STRETCHER, RECOVERY, SURGICAL
M7040 TABLE, OVERBED
M7845 MONITOR, PHYSIOLOGICAL, BEDSIDE,
4 CHANNEL
P3100 LAVATORY, VITREOUS CHINA,
SLAB TYPE
Elevations 1 & 2
DISPENSER, SOAP
E0948
CART, GENERAL STORAGE, MOBILE
12' - 6"
M0755 3810mm M0755
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
FLOWMETER, OXYGEN, LOW FLOW FLOWMETER, OXYGEN, LOW FLOW
M0765 M0765
REGULATOR, VACUUM REGULATOR, VACUUM
4-279
April 2016 (rev 2/17)
Elevation 3
150 NSF / 14,0 NSM
6' - 6"
1981mm
PACU / PHASE I RECOVERY (RRIR1)
CB020
CABINET, W-H, 2 SHELF, 1 DO, 38x24x13 LH A5075
A5080 DISPENSER, SOAP
DISPENSER, PAPER TOWEL CT020
CS140 COUNTER, SOLID SURFACE
SINK, SS,GENERAL, SINGLE X, 10X14X ID M3070
C0036 HAMPER, LINEN
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
RAIL, APRON, 4x30x1 C01D0
F2010 CABINET, U-C-B, 4 DR, 36x18x22
BASKET, WASTEPAPER, STEP-ON
4-280
April 2016 (rev 2/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
ARCHITECTURAL
Ceiling Type: Acoustical Ceiling Tile (SP)
Ceiling Height: 9’-0” (2700mm)
Ceiling Finish:
Wall Finish: Gypsum Wallboard (SC)
Wainscot:
POWER
Base: WSF Integral Base (min. 6”/
Normal Power: To be connected to selected
152 mm)
receptacles and equipment.
Floor Finish: Welded Seam Sheet
Emergency Power: Critical branch of the EES
Flooring
to be connected to selected
Slab Depression: None receptacles and equipment.
Sound Protection: 35 STC (Room)/ None (Bay)
Doors: Sliding Glass Door, Size 8’-
0” x 7’-0” (2438 mm x 2133
mm) ; Single, Size 3’-0” x
7’-0” (914 mm x 2133 mm)
Wood w/ Half View Window
Special Requirement:
Notes:
1) Verify with life safety requirements that sliding COMMUNICATIONS
glass doors can break out of the room Data: Yes
LIGHTING Telephone: Yes
Refer to the VA Lighting Design Manual section Cable Television: No
4.2.15 - Pre-Operative and Post-Anesthetic Care Duress Alarm: No
(PACU) - for lighting design consideration. Electronic Access and Door No
Control:
Intercom: No
Motion Intrusion Detection No
(MID):
Nurse Call: Yes
Code Blue: Yes
Public Address: No
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: No
Video Teleconferencing No
(VTEL):
Axonometric
120 NSF / 11,2 NSM
R6111 A5108
FREEZER, CRYOGENIC, WASTE DISPOSAL UNIT,
CHEST SHARPS
E0703 M3072
TABLE, PROCESS, 5 FRAME, INFECTIOUS
DRAWERS WASTE BAG W/LID
A1015
A5077
TELEPHONE, DESK,
MULTIPLE LINE 12x12 DISPENSER, HAND
SANITIZER, HANDS FREE
D3295
CHAIR, ROTARY,
LABORATORY, DENTAL A5107
F3200 DISPENSER, GLOVE,
CLOCK, BATTERY, 12IN WALL-MTD
OSa
L0100 A5082
3048mm
10' - 0"
MICROSCOPE, DISPENSER, PAPER
BINOCULAR TOWEL, SENSOR, HANDS
1
L8590 FREE
STAINER, SLIDE, CS140
AUTOMATIC, 6" AFC
SINK, SS,GENERAL,
HEMATOLOGY
SINGLE X, 10X14X ID
GFI
L4200
BATH, WATER,
SEROLOGY, ELECTRIC P1965
L3400 EYEWASH, EYE-FACE,
MICROTOME, ROTARY, 12' - 0" SINK MOUNTED,
TILT, BENCH MOUNTED 3658mm HANDS-FREE
L9025 A5075
FREEZE-DRYER, TISSUE DISPENSER, SOAP
CT060
E0721
COUNTER, MODIFIED
TABLE, PROCESS, 5 DRAWERS
EPOXY RESIN
3048mm
10' - 0"
a a
9' - 0"
12' - 0"
3658mm
SEROLOGY, ELECTRIC
BASKET, WASTEPAPER,
Surgical and Endovascular Services Design Guide
ROOM TEMPLATES
ROUND, METAL E0721
C0036 TABLE, PROCESS, 5
RAIL, APRON, 4x30x1 DRAWERS
CT060 C01D0
4-292
April 2016 (rev 2/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
ARCHITECTURAL POWER
Ceiling Type: Gypsum Wallboard (SC) Normal Power: To be connected to selected
Ceiling Height: 9’-0” (2700mm) receptacles and equipment.
Ceiling Finish: Emergency Power: Critical branch of the EES
to be connected to selected
Wall Finish: Gypsum Wallboard (SC)
receptacles and equipment.
Wainscot:
Notes:
Base: Resilient Base
1) Connect lab equipment to Critical Branch of
Floor Finish: Resinous Flooring, Welded the EES.
Seam Sheet Flooring
Slab Depression: None Special
Sound Protection: None
Doors: Single Door, Size 3’-0” x COMMUNICATIONS
7’-0” (914 mm x 2133 mm) Data: Yes
Wood
Telephone: Yes
LIGHTING Cable Television: No
Refer to the VA Lighting Design Manual Duress Alarm: No
section 4.2.4 - Laboratory - for lighting design
Electronic Access and Door Yes
consideration.
Control:
Intercom: Yes (Phone)
Motion Intrusion Detection No
(MID):
Nurse Call: No
Code Blue: No
Public Address: No
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: No
Video Teleconferencing No
(VTEL):
Special Requirement: No
Axonometric
120 NSF / 11,2 NSM
F2010 A1012
BASKET, WASTEPAPER, TELEPHONE, WALL
STEP-ON MOUNTED, 1 LINE
A5082
A5077
DISPENSER, PAPER 54" AFF DISPENSER, HAND
TOWEL, SENSOR,
SANITIZER, HANDS FREE
HANDS FREE
A5075 S9755
DISPENSER, SOAP 2 1 SUCTION SYSTEM, SURGICAL
MOBILE ROVER UNIT
P3100
3658mm
OSa
S9756
12' - 0"
LAVATORY, VITREOUS
SUCTION SYSTEM, SURGICAL
CHINA, SLAB TYPE GFI 6" AFC ROVER DOCKING STATION
P2451
M3072
VALVE, MIXING,
THERMOSTATIC, EYEWASH FRAME, INFECTIOUS
WASTE BAG W/LID
P2000
P6350
EYEWASH, WALL
SINK, FLUSHING RIM, CHINA
NC
MOUNTED,
HANDS-FREE
A5107
DISPENSER, GLOVE, 10' - 0"
WALL-MTD 3048mm
M3070
HAMPER, LINEN
9' - 0"
3658mm
12' - 0"
10' - 0"
3048mm
Elevations 1 & 2
120 NSF / 11,2 NSM
S9755 42" SIDE WALL
SUCTION SYSTEM, SURGICAL
MOBILE ROVER UNIT
U0126 P6350
SUCTION SYSTEM, SURGICAL SINK, FLUSHING RIM, CHINA
ROVER DOCKING STATION
ELEVATION 1
12' - 0"
3658mm
CD030
CABINET, W-H, 2 SHELF, 2 DO
P2451
4.24. UTILITY ROOM, SURGICAL SOILED (USCL7)
ROOM TEMPLATES
THERMOSTATIC, EYEWASH DISPENSER, SOAP
P2000 A5082
EYEWASH, WALL MOUNTED, HANDS-FREE DISPENSER, PAPER
ELEVATION 2
SCALE: 3/16” = 1’-0”
4-302
April 2016 (rev 2/17)
Surgical and Endovascular Services Design Guide April 2016 (rev 11/18)
ARCHITECTURAL POWER
Ceiling Type: Gypsum Wallboard (SC) Normal Power: To be connected to selected
Ceiling Height: 9’-0” (2700mm) receptacles and equipment.
Ceiling Finish: Emergency Power: Not required.
Wall Finish: Gypsum Wallboard (Epoxy)
Wainscot:
Base: Resilient Base
COMMUNICATIONS
Floor Finish: Resinous Flooring, Welded
Seam Sheet Flooring Data: Yes
Slab Depression: None Special Telephone: Yes
Sound Protection: None Cable Television: No
Duress Alarm: No
Doors: Single Door, Size 3’-0” x
7’-0” (914 mm x 2133 mm) Electronic Access and Door Yes
Wood Control:
Intercom: Yes (Phone)
LIGHTING
Motion Intrusion Detection No
Refer to the VA Lighting Design Manual section (MID):
7.3 - Soiled Utility Room - for lighting design
consideration. Nurse Call: Yes
Code Blue: No
Public Address: No
Security Surveillance Televi- No
sion (SSTV):
VA Satellite TV: No
Video Teleconferencing No
(VTEL):
CONTROL
ROOM
IMAGING EQUIPMENT
COMPONENT ROOM
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-307
4.25 ORHY3
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY3) OPERATING ROOM, HYBRID OR ROBOTIC
INTERACTIVE 3D PDF
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-308
4.25 ORHY3
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY3) OPERATING ROOM, HYBRID OR ROBOTIC
FLOOR PLAN
SEMI-RESTRICTED CORRIDOR
32' - 2"
DOORS TO BE ON
(9.8m)
AUTOMATIC DOOR
A5107 OPENERS
STRETCHER ALCOVE A5108 SCRUB ALCOVE RETURN AIR
M8940 DUCT
(TYPICAL)
GM IPS
TV T H A4015 LOW WALL
EPO
F3200 RETURN AIR
1 REGISTER
M8900
M8 900 (TYPICAL)
M8905 M8910 4 2
25' - 6"
MANIFOLD (7.77m)
(8.74m)
X6190 3 A1200 TO CONTROL
28' - 8"
UNI- DIRECTIONAL M4255 LIFT AND
M8920 M8810 ROOM
AIR DELIVERY
(4.62m)
R4785
ARRAY ABOVE M4266
STERILE FIELD M8825
E0948
M8606
GM
(3.66m)
12' - 0"
M8925
M8825
M3175 BORROWED
M5030
M3070 LIGHT
A5104 F0355
TABLE M4645 M8800 M4255 M8830
A5108 M4287
M3072 M4810
U0105 M4815 M8830
E0954
M4812 M8940 M4816 X2106
M7818 M9080
GM M7650 X3145
M8940 L0221
IPS
18' - 0"
A1014
IMAGE
ROOM
(5.49m)
A5107
GM D D
TABLE
M3150 M3165 M3080
OPERATING ROOM,
HYBRID OR ROBOTIC
(ORHY3)
900 NSF
83.67 NSM
CLEAN CORE
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-309
4.25 ORHY3
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY3) OPERATING ROOM, HYBRID OR ROBOTIC
REFLECTED CEILING PLAN
1 NITROUS
(8.74m)
28' - 8"
A
LED
LED
LED
LED
LED
LED
LED
LED
LED
OXIDE,
A
LED
LED
LED
LED
LED
LED
LED
LED
LED
6 EMERG
LED
DUPLEX,
A
LED LED LED LED LED LED
4 DATA FACE
A
TABLE PLATES
LED
LED
LED
LED
LED
LED
LED
LED
LED
(ANESTHESIA)
A
LED
LED
LED
LED
LED
LED
LED
LED
LED
A1120
A
M0750
LED LED LED LED LED LED
A
M0755
LED
LED
LED
LED
LED
LED
LED
LED
LED
LED
(EQUIPMENT) M0765
A
LED
LED
LED
LED
LED
LED
LED
LED
LED
4V,
1 NITROGEN, (PERFUSION)
LED LED LED LED LED LED
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-310
4.25 ORHY3
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY3) OPERATING ROOM, HYBRID OR ROBOTIC
JSN LEGEND
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-311
4.25 ORHY3
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY3) OPERATING ROOM, HYBRID OR ROBOTIC
ELEVATIONS
EPO
T H S
ELEVATION 1 A5107
Dispenser, Glove,
M8910
Surgical/Examination,
Cart, Surgical Case
Wall Mntd
LOW WALL RETURN
ISOLATED POWER
AIR REGISTER (TYPICAL)
SYSTEM
F0355 A1014
Footstool, Straight Telephone, Wall
E0948 Mounted, 1 Line, With
Cart, General Storage, Speaker
Mobile, 42"H x 32"W x 22"D
A5107
Dispenser, Glove,
Surgical/Examination,
Wall Mntd
BORROWED
LIGHT
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-312
4.25 ORHY3
SURGICAL AND ENDOVASCULAR SERVICES
(ORHY3) OPERATING ROOM, HYBRID OR ROBOTIC
ELEVATIONS
M3165
A5107 Cabinet, Catheter Storage
Dispenser, Glove, M3150
Surgical/Examination, Distribution System,
Wall Mntd Medication, Automatic
SD SD
M7650
Defibrillator/Monitor, Acute
Care M4645
M4812 Patient Transfer Device
Pacemaker, Single Chamber, M3070
External, Temporary Hamper, Linen, Mobile,
M7818 w/Lid
Monitor, Transport X6190
ISOLATED POWER F3050 Radiographic/Fluoro Unit,
SYSTEM Whiteboard, Dry Erase Cardiac, 100 kW, Digital
M3150 M8905
Distribution System, ELEVATION 4 Pail, Utility, CRS, With
Medication, Automatic Carriage
E0954 M8900
Cart, Emergency, Mobile, Carriage, Pail, CRS,
66"H x 32"W x 22"D Without Pail
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS. 4-313
Surgical and Endovascular Services Design Guide April 2016 (rev 11/22)
4.25 ORHY3
4.25 OPERATING ROOM, HYBRID OR ROBOTIC (ORHY3)
Room Data Sheet
Notes:
1) Refer to the latest version of the VA HVAC
Design Manual for quantity and location of low
air return grilles and ceiling diffusers.
2) Room does not contain multiple slot
diffusers and uses laminar flow perforated face
outlets only for supply air
3) Suggested minimum laminar flow array over
the entire sterile field area as imaging equipment
gantry creates excessive turbulence. Mechanical
design engineer shall be responsible to design
the array in such manner as to minimize
turbulence and to maintain the sterile aseptic
field.