E60VGA Electrocardiograph
User`s Manual
CONTENTS
1. INTRODUCTION .............................................................................................................................................3
1.1 HOW TO USE THE USER’S MANUAL? ..........................................................................................................4
1.3 INFORMATION AND INSTRUCTIONS FOR SAFE USE.......................................................................................5
2. STRUCTURE OF THE ELECTROCARDIOGRAPH ..............................................................................7
Figure 1. Structure of the electrocardiograph............................................................................................... 7
Figure 2. External sockets view. .................................................................................................................. 8
3. DESCRIPTION OF SYMBOLS AND FUNCTIONS OF THE ELECTROCARDIOGRAPH ..............9
Figure 3. Foil keyboard................................................................................................................................ 9
Figure 4. Foil with drawing of electrodes connection.................................................................................. 9
4. TURNING THE ELECTROCARDIOGRAPH ON AND PREPARATION FOR EXAMINATION..12
4.1 INSTALLATION .........................................................................................................................................12
4.2 CONNECTING THE PATENT CABLE ............................................................................................................13
4.3 PREPARING FOR EXAMINATION ................................................................................................................13
Figure 5. Standard location of the electrodes ............................................................................................. 14
4.4 CONNECTING ECG TO THE MONITOR .......................................................................................................14
Figure 6. Connecting ECG to the monitor ................................................................................................. 15
4.4.1. Main view, electrode test mode ...........................................................................................................15
4.5 SETTING THE SPEED, SENSITIVITY AND FILTERS .......................................................................................16
4.5. SETTING UP THE REGISTRATION PARAMETERS .............................................................................................17
4.5.1. ECG Configuration Icons. .............................................................................................................17
4.5.3. Printout Format Selection. ...................................................................................................................18
4.6 OPERATING THE RECORDER, CHANGING THE ROLL OF PAPER ...................................................................19
Figure 7. Mounting the roll of paper onto the handle................................................................................. 19
5. RECORDING THE ECG TESTING.........................................................................................................20
5.1 RECORDING IN THE MAN (MANUAL) MODE ............................................................................................20
5.2 RECORDING IN THE AUTO (AUTOMATIC) MODE......................................................................................20
5.3 ANALYSIS ................................................................................................................................................21
5.4 MEASUREMENTS AND ANALYSIS OF THE HART BEAT RHYTHM .................................................................22
6. TRANSMISSION OF DATA INTO THE COMPUTER .........................................................................23
Fig. 8. PC - ECG Wiring Diagram............................................................................................................. 23
7. AUTOTEST .................................................................................................................................................24
8. OPERATING, STORAGE AND PERIODICAL MAINTENANCE ......................................................25
8.1 CHECKING, MAINTENANCE AND SAFETY TESTS ........................................................................................26
9. RECYCLING AND DISPOSAL .............................................................................................................27
10. ELECTROMAGNETIC COMPATIBILITY .....................................................................................28
11. TECHNICAL SPECIFICATIONS ......................................................................................................30
STANDARD ACCESSORIES: ..................................................................................................................................32
E60VGA Electrocardiograph
User`s Manual
1. Introduction
The Electrocardiograph E60VGA is intended for heart muscle testing in hospitals, clinics,
emergency service and private doctor’s rooms.
The Electrocardiograph E60VGA with external and internal, battery operated power
source are to be used in routine diagnostics of the heart muscle and for the 3, 6, and 12-
channel data recording of the action variables according to the widely used output
systems: Einthoven’s, Goldberger’s, Wilson’s.
The Electrocardiograph allows for the following:
- manual or automatic recording of the ECG signal
- automatic regulation of the location of the zero line inside the recording channel,
- automatic elimination of crawling of the zero line,
- automatic choice of recording sensitivity,
- digital filtration of external disturbances,
- digital filtration of disturbances occurring on the level of the patient muscles,
- signalization about disconnected electrodes,
- signalization about lack of paper or that the recorder is not closed,
- performance of 3 min II recording in order to create a characterization of the frequency of
the heart beat rhythm,
- measurements of the R-R line,
- cooperation with a computer through the inline connector RS 232,
- storage of the last test - test archiving
- repeated printout of the performed test,
- analysis of the recorded ECG signal and printout of the measurement results,
- 2 hours of continuous work for printout in AUTO mode (without analysis), when the
apparatus is powered by an internal battery.
The Electrocardiograph E60VGA is constructed and produced in full in the X-ray and
Medical Equipment manufacturer FARUM S.A., in accordance with the Directive about
Medical Equipment number 93/42/EEC and the ISO 9001:2000 and ISO 13485:2003
quality markers. It fulfills all the requirements of the PN-EN 60601-1:1999, PN-EN
60601-1-1:2002, PN-EN 60601-2-25:2004, PN-EN 60601-1-2:2002, PN-EN 60601-1-
4:2006, PN-EN ISO 14971:2004 standards.
It is produced under first class safety (I class), in the safety level of CF.
The patient is separated by an electro isolating barrier, which ensures the highest level of
safety and has a special protector against defibrillation impulses.
To ensure a proper and safe operation of the electrocardiograph, as well as to be able to fully
benefit from all the functions of the electrocardiograph every potential user must carefully and
thoroughly read the instructions given in the User’s Manual.
This symbol signifies the user’s duty to carefully read instructions given in a
particular point in the User’s Manual, pertaining to the topic discussed, as well
as all the warnings about any possible dangers arising for the patient and the
personnel operating the electrocardiograph.
This symbol warns about the necessity of applying special safety precautions
and the user’s duty to carefully read instructions given in a particular point of the
User’s Manual.
This symbol signifies important cautions and instructions, for example, those
cautioning against damaging the electrocardiograph or its components.
The set up of this User’s Manual allows for familiarization oneself with the proper ways of
operation of the electrocardiograph, as well as its functions in time and on the level depending
on the reader’s abilities and amount of knowledge about operation of such instruments.
- always perform the test powering the electrocardiograph from the batteries (internal power
source); disconnect the external power source cable from the power source.
- Always use the electrocardiograph according with the instructions and precautions given in the
User’s Manual.
- The electrocardiograph is equipped with the standard set of accessories. To ensure safety and
compliance with the Directive about Medical Equipment number 93/42/EEC always use only
original additional accessories accepted and approved by the manufacturer.
- The apparatus is equipped with a thermal writing head, which allows for maximum quality and
precision of the test recording on paper. To avoid frequent and expensive repairs and renovations
always use only original paper accepted and approved by the manufacturer. The manufacturer
will not be held responsible for damage caused by the usage of inappropriate paper.
- The electrocardiograph is adapted to work on mobile trolley produced by FARUM S.A. (with
arm for patient cable, equipped with two castors with brakes, the apparatus is fastened to the
trolley with the knob) – optional equipment.
- The electrocardiograph should be carried or transported to the patient in a special bag the
production of FARUM S.A. - optional equipment.
- Do not subject the apparatus to any heavy blows or shaking.
- Never allow any liquids inside the apparatus. If such occurs you must immediately bring the
apparatus to our technical service department in order to test its functionality.
- Always check if the voltage of the power source complies with that described on the information
label located on the apparatus.
- If you are using this electrocardiograph in connection with other apparatus always check that: all
connections comply with the safety regulations; all apparatus connected falls under the same
safety regulations. Lack of compliance with the safety regulations may cause bodily harm to the
patient connected to the apparatus, as well as the person operating the electrocardiograph. If it is
difficult for you to obtain information required to analyze the risks connected with each
individual connection contact the manufacturer, or refrain from using such connections.
- If you are using other apparatus connected directly or indirectly to the patient always check all
the risks, which may be caused by the sum of all the variables transmitted through the patient.
- The electrocardiograph is protected against defibrillation in compliance with the standards and
norms PN-EN 60601-2-25:2004; to ensure proper transmission of the signal always use only the
original electrodes accepted and approved by FARUM SA.
- Do not allow any contact of the apparatus with extreme temperatures, excessive dirt or dust, or
very salty and humid climates; always behave in compliance with the instructions about the
environment conditions given.
- If the system of the apparatus “crashes” or “freezes” (the apparatus does not respond to the
commands given from the keyboard) press the on/off button or ,if this does not help, press the
button number 9 (reset button) shows in figure 1, which is located on the right side of the
apparatus cover in the cover hole.(using for example the plug of patient cable)
- Before starting the examinations check visually the technical state of the apparatus. Contact with
FARUM SA or with Authorized Service of the producer in all cases of finding of damage or
improper functioning of the apparatus. Contact the Authorized Technical Service Department in
every case of improper operation of the electrocardiograph.
E60VGA Electrocardiograph
User`s Manual
- To ensure a long usage of the apparatus it is important to periodically have it checked by the
Authorized Technical Service Department of the producer according to point 8 of this User’s
Manual.
- Do not use this electrocardiograph for monitoring the heart beat during
any operations.
- FARUM SA is responsible for the safe and proper operation of the electrocardiograph only
when:
o all modifications and repairs were performed by the FARUM S.A. or and Authorized
Service Department of the manufacturer;
o voltage in the building complies with that required;
o the apparatus is exploited in compliance with the instructions and precautions given in the
User’s Manual;
o all basic and additional accessories used with the apparatus have been accepted and
approved by the FARUM S.A.
E60VGA Electrocardiograph
User`s Manual
9 2 11
RESET
5
®
4
S.A. E60VGA
1
5 V1 V4
I aVR V5
10 12,5 3 II aVL V2
50 Hz Man V3 V6
20 25 III aVF
6
Auto 50 12 Auto
1 2 3
mm/mV 25 Hz Rytm
mm/s 35 Hz
RS-232 VGA
10 8 6 7
VGA RS-232
VGA interface –
RS232 interface –
monitor signal
communication with the PC
output
5 V1 V4
I aVR V5
10 12,5 aVL V2
3 II V6
50 Hz Man V3
20 25 III aVF
6
Auto 50 12 Auto
1 2 3
mm/mV 25 Hz Rytm
mm/s 35 Hz
Diode signalling
disconnected electrode
E 60
Diode signalling
connection of the
apparatus to power
supply
Diode signalling
discharged battery
E60VGA Electrocardiograph
User`s Manual
While working without a power supply, the flashing diode signals that the battery is
discharged and it is necessity to connect it to the external power source and perform a
complete battery charge, as it is described in point 4.1.
- STOP button. After pressing the button STOP the apparatus always return to
the same settings as before the test began.
I aVR V1 V4
II aVL V2 V5
III aVF V3 V6
- switches for the choice of the leads for the MAN
recording
After choosing the desired group of the leads, the apparatus will be switched over into manual
working mode MAN.
1 2 3
The electrocardiograph should be connected to the outlet of the external power source of
90÷240V 50 Hz, with the grounding bolt using the cable equipped with the apparatus.
Power supply
ON/OFF button
indicator
Battery charding
indicator
If the yellow LED diode lights up, it signalizes that the internal power source (battery) is
uncharged.
In case the level of energy in the accumulator is not sufficient and unables the proper
functioning of the apparatus a sound signal is going to be generated five times and afterwards
the apparatus will turn off. Further usage of the apparatus is going to be possible only after
plugging it into the alternating current mains.
If the battery discharges to a point when a proper ECG recording cannot be performed the
apparatus will not turn on if it’s not connected to an external power source. You must then
connect it to an external power supply and perform a complete, 16-hour battery charge
(during that time the apparatus should be turned off).
The average lifespan of the battery is approximately 500 full discharge / charge cycles.
An apparatus, which is on and left without activate the registration turn off automatically after
5 minutes
Do not remove any used batteries on your own! If the batteries must be
changed immediately contact the authorized Technical Service of FARUM
S.A.
If there is no more paper in the printer, or if the printer is “open,” the functional keyboard is
12,5
25
50
mm/s
blocked and diodes which are placed in the field of speed choosing signal those state
by flashing.
Pressing each button of functional keyboard and finishing the RHYTHM (3 minutes)
registration are signalling by sound signal.
The standard set includes 4 limb electrodes and 6 suction electrodes for the heart area.
The limb electrodes are placed on the signal receiving spots on the arms and legs of the
patient.
In case there is no contact or the electrodes have not been connected, electrode symbols will
flash .
Additionally lack of the contact or disconnected electrodes, may be signalized by flashing
electrode symbol on the monitor display, as follows:
R - red electrode – red plug, C1 electrode - red and white plug,
L - yellow electrode - yellow plug, C2 electrode - yellow and white plug,
F - green electrode – green plug, C3 electrode - green and white plug,
N - black electrode – black plug, C4 electrode - brown and white plug,
C5 electrode - black and white plug,
C6 electrode - purple and white plug,
Attention: If during the examination Electrode N is open, the system will print final
report only; this will help saving the paper.
It is very important to notice the following when performing the ECG recording:
- that the patient and transmitting electrodes are not touching any other persons
or transmitting objects, even if such are grounded;
- that the heart electrodes are not touching each other;
- that the patient cable is placed far from the power source cable and that they
are not intercrossing;
because such can cause severe disturbances or interrupt the recording.
Connecting monitor enables getting additional function of the apparatus like graphical
preview of disconnected electrodes, ECG signals view etc.
ECG will be functioning properly including all the basic functions also in the case, when the
monitor is disconnected.
It is recommended to connect monitor to the apparatus before turning both devices on..
VGA RS-232
Plug connecting
Monitor cable
Transformer
separating
It is recommended to use 14” or 15” LCD monitor. Using the old CRT monitors is also
possible, but because of preset frequency (50Hz) the view may flash.
Attention: If the monitor has not been plugged, the lack of connection of the
electrodes of the patient cable have not been connected or disconnection of the
electrode is present, electrode symbols will flash in the aparatus
E60VGA Electrocardiograph
User`s Manual
20
mm/mV
12,5
mm/s
50 Hz
35 Hz
3x4 TEST Directly after turning the apparatus on (producer’s
logo view) apparatus gets to the main view and
works in the electrodes test mode enabling simple
R
L identification of improper contact of the
F
N electrodes.
C1
C2
C3
C4
C5
C6
20 12,5 50 Hz
3x4 TEST
mm/mV mm/s 35 Hz
R
L
F Electrodes contact pointers. Lack of
N
C1
the contact or improper contact of the
C2 electrode with the patent is signalized
C3
C4 by the corresponding electrode
C5
C6
pointer flashing and the audible
signal. Disconnection of the patient
cable is signalized by the N electrode
flashing only.
Settings of the sensitivity, speed and filters can be changed by pressing the following buttons:
Sensitivity - mm/mV,
Speed - mm/s,
Filter 50 Hz - ON / OFF,
Filter 35 Hz or 25 Hz - ON / OFF
Turning on of the settings is signalized by LED diodes which are placed in the field of chosen
settings.
E60VGA Electrocardiograph
User`s Manual
In the top part of the display actual apparatus settings are displayed.
20 12,5 50 Hz
3x4 Manual
mm/mV mm/s 35 Hz
Cecha: 1 mV
I
II
Chart field
Recording modes;
ECG amplitude - Immediate real-time printout (MAN)
signal [mV]: - Automatic (AUTO)
5; 10; 20; - Rhythm examination (RHYTHM)
AUTO
Paper feed rate, ECG Filters: Printout format
signal scale [mm/s]: 25, 35, 50 Hz MAN: 3 channels
12.5 ; 25 ; 50 AUTO: 4x3 ; 2x6 ; 1x12
RHYTHM: 6 or 12 lines (IInd channel)
Format: 4x3
I aVR V1 V4
II aVL V2 V5
III aVF V3 V6
Format: 6 lines
0’00 ... RHYTHM: Rhythm examination.
0’30 ...
1’00 ... Within 3 minutes the system will record the signal from
II channel II and save it to the built-in memory. This type of
1’30 ...
2’00 ... investigation ensures determination of statistical
2’30 ... distribution of HR (R-R distance) and printout of a
histogram.
Format: 12 lines
0’00 ...
0’15 ...
0’30 ...
0’45 ...
1’00 ...
II 1’15 ...
1’30 ...
1’45 ...
2’00 ...
2’15 ...
2’30 ...
2’45 ...
E60VGA Electrocardiograph
User`s Manual
Before beginning any recording the roll of paper should be placed into the recorder and the
recorder should be loaded in the way that the printed side of the paper is touching the thermal
head with its scale. After closing the recorder table and choosing the lead it is possible to
begin the recording.
After all the paper has been used the recorder will stop and the diodes which are placed in the
chosen speed field are signalling this state by flashing.
In order to change the roll of paper push the lever of the recorder table latch. This will cause
the table to open up and will allow access to the recorder where the empty roll of paper is
located.
Table of the recorder is not fix immovable to the apparatus – it is necessary to treat
with care so as not to cause any damage, for example by dropping.
Place the new roll of paper into the recorder, pullout the end of the paper from the recorder in
the record, placing it symmetrically to the bosses on the external surface of the recorder table,
and, through pressing the table, close the record.
After changing the roll of paper the apparatus is again ready to operate.
Remember that the high quality of the print out can only be ensured if you
are using original recording paper which was bought in FARUM S.A.
Figure 4 shows how to mount the roll of paper onto the handle.
After turning on the apparatus and choosing the desired output group from the foil keyboard
(figure 3) the diode MAN light, which is placed in field of working modes.
After pressing the button the recording and printout will begin. The printout
in 3 channels will include:
- coding of the output and the calibration impulse,
- recording of the chosen output group along with the printout consisting of the following
information: MAN, measuring sensitivity, speed of paper intake, description of active
filters 35 Hz (25 Hz) and 50 Hz.
Changing the output group, sensitivity settings and the speed, as well as activation
and deactivation of filters while the recording is in progress will cause a printout of
a new information sheet.
Choosing the AUTO mode is possible by switching over the button of choose of registration
modes (rewinding) and the AUTO diode will be lighted.
Man
In similar way it is choosing the recording format (3, 6, 12 channels), the output group of
speed settings, sensitivity settings and filters settings.
After pressing the button (START) the speed, sensitivity and filters buttons will be
blocked and AUTO diode will start flashing what symbolize the registration test.
After 5 seconds of collecting data of the ECG signal a printout in the following format will
begin:
• 3 channel recording – 4 groups each 3 leads
• 6 channel recording – 2 groups each 6 leads
• 12 channel recording – 1 group each 12 leads
After the outputs are printed a report with the following information is printed:
- HR, R-R measurements results,
- Medicines used: dotting area with the place for writing of used medicines
and the others.
E60VGA Electrocardiograph
User`s Manual
The Apparatus automatically remember the last test. It enables to make the later copy of the
test and the following button can be pressed on the apparatus:
- STOP - terminates the test
- START – repeated recording
1 2 3
and it is possible to create multiple printouts copy of the last test also after switching on and
after the next switching off the apparatus.
CAUTION: In order to save paper if the N electrodes will not be connected when
performing a recording the report of „Electrode disconnected” will be
printed.
5.3 Analysis
After the printout is finished in the AUTO (automatic) mode the following button will be
pressed:
- STOP - terminates the test
- START – repeated recording
1 2 3
After pressing button and after 5 seconds the following will be performed:
- a printout of standardized outputs with the vertical markers showing the beginning and the
end of each break will be created
- a printout of calculated time parameters will be created
- a printout of calculated amplitude will be created
- stop the print out and the following button will be pressed:
In case of detection by the apparatus the impulses of pacemaker (or appearing a big periodic
disturbances, the analysis will not be done. However, on screen of the apparatus appear the
statement:
Analysis failed
Pacemaker is detected
CAUTION! ECG research, in which can appear the pacemaker impulses, it has to done when
the filters 25, 35, 50 Hz are turn on. Turning on in apparatus the filters 25, 35, 50
Hz, it cause deformity and filtration the impulses of the pacemaker, what the
reason it what they can not be seen in printout.
Choosing the RHYTHM mode is possible by switching over the button of choose of
registration modes (rewinding) till the RHYTHM diode will be lighted.
Man
Diode which Lever to change
signalling the Auto
the registration
Rytm
RHYTHM choose choose mode
working mode
After inputting the patient identification data (not mandatory) and pressing the (start)
button the speed, sensitivity and filters buttons will become blocked and
RHYTHM diode will start flicker, what signals the data acquisition.
After the 3-minute recording of the II output a printout of the measured contractions in the 6-
channel format will be created and the following button will be pressed:
:
- STOP – terminates the test
- START – repeated recording
1 2 3
1 2 3
After pressing the (analysis) button the following calculations will be printed out:
R-R interval parameters and a histogram of time trends of the recorded test will be created.
E60VGA Electrocardiograph
User`s Manual
The Apparatus has a socket used for PC connecting by RS 232 interface (figure. 1 point 6).
The functions built into the apparatus enable transmission of the ECG signal data
to a personal computer in compliance with the RS232 standard. This connection
can be operated only by qualified personnel whose training and experience
assures safe and proper use of the apparatus.
The PC must be loaded with the E600Win program delivered by the producer of the apparatus
to enable such transmission. The program enables archiving of data after the recording in
AUTO (automatic) or RHYTHM modes is finished.
Required PC parameters:
Minimal hardware requirements:
- The processor: compatible with Pentium 400 MHz,
- Memory: 256 MB RAM,
- Resolution: 256 colours, 800x600 dpi,
- Windows 95
During the examination the apparatus must not be connected to the PC.
PC
PC
RS 232
7. Autotest
The “Autotest” function allows the user to turn on the automatic testing and checking of the
apparatus functions operation.
Testing of the apparatus is activated upon pressing and holding the measurement button for 3
seconds after turning on the apparatus.
The triangular line on the width of the paper will be printed.
The test of the apparatus will be made and in case of proper working the symbol will be
printed.
In case of mistakes and damages will be appeared the symbol will be printed.
It means that the apparatus should be given to the service.
Avoid storing and operating the apparatus in the close proximity of X-ray apparatus,
diathermy apparatus, large transformers or engines.
Before cleaning the apparatus always remember to disconnect from the power source.
Never clean the apparatus using a cleaning detergent or sterilize it using boiling / hot
water, steam or hot air.
The apparatus may be externally cleaned using a solution of water and a mild
detergent with a soft cloth.
After each use of the apparatus the electrodes should be cleaned using the solution of
water and soap. You must check if the sucker of the suction electrode is completely
dry. Sterilization can also be performed using the ethylene oxide or, for example,
Alhydex, Vygon.
The patient cable should not be subjected to any heavy pressure. While disconnecting
the outputs you must always hold the plugs, not the cable. The cables should be placed
in a way that they would not cause any persons to trip over them or any carts wheeling
them over.
If sterilization is required it should be performed using only gas, never steam. In order
to disinfect the cable should be wiped with a soft cloth dampened with one of the
following detergents: Inicidin GG, Amocid, Lysoformin, Alhydex.
The longevity of the apparatus is seven (7) years on condition that maintenance and annual
technical safety tests of the apparatus will be done
Caution: If the apparatus was transported in temperature lower than 0°C, after getting out it
from packing, it is needed to warm it to the working temperature (10 °C ÷ 40°C)
before connecting the apparatus to the mains or switching on it using starting
button. That time depends on freezing grade of the apparatus and it could be
approximately from one to few hours. After that time it is needed to full recharge of
the battery according to point 4.1 of the User`s Manual.
E60VGA Electrocardiograph
User`s Manual
If the storage time of the apparatus is longer than one month, it is needed to
recharge the battery according to point 4.1 of the User`s Manual after each month
of storage.
Thermo print head SHALL BE cleaned with switched off apparatus. Correct frequency of
cleaning ensures long time of printing and its correctness.
Proceed as follows:
remove the shield of the paper container
clean the heat points of the thermoprint head with the small piece of cotton-wool
saturated with spirit, regarding not to touch the head with hand or with other
thing!
Every 12 months the apparatus should be given a safety test performed by the FARUM
S.A. Technical Service Department or Authorized Service of the producer.
The test should include the following:
- thorough overview of the apparatus,
- thorough overview of the components,
- test of the safety grounding, test of the isolation,
- measurement (analysis) of the current flow,
- checking of all the functions of the apparatus.
The results of the tests should be attached to the documentation pertaining to the apparatus.
All repairs can be performed only by FARUM S.A. Service or by Authorized Service of the
producer.
E60VGA Electrocardiograph
User`s Manual
For extra-European markets comply with the laws in force considering that the product
contains:
- Battery
- Lead
- Electronic units
- Ferroalloys, aluminium and copper
- ABS
E60VGA Electrocardiograph
User`s Manual
A class
Voltage fluctuations and PN – EN 61000-3- Positive
flicker emission 3:2002+A2
2:2002
PN-EN 61000-4-
5:1998+A1
Conducted RF PN-EN 60601-1- Positive The best quality of
immunity 2:2002 registered test could be
PN-EN 61000-4- reached by disconnection,
6:1999+A1 during registration, power
supply and using internal
accumulator.
Magnetic field PN-EN 60601-1- Positive It is not allowed to use
immunity 2:2002 devices which generate
PN-EN 61000-4- magnetic field during test
8:1998+A1 in view of possibility of
increase interference level
visible in printout.
Voltage dips, PN-EN 60601-1- Positive Apparatus is equipped
interruptions 2:2002 with internal accumulator
immunity tests PN-EN 61000-4- which ensue continuity of
11:2005 activity.
ECG amplifier:
Input impedance >100 MΩ
CMRR >100dB
Registration band 0,05 ÷ 150Hz
A/C converter 12 bits
Frequency sampling 800 Hz/channel
Maximal level of constant ±300 mV
voltage
Time constant ≥ 3,2 s
Measurement sensitivity 5; 10; 20 mm/mV or AUTO
Paper intake speed 12,5; 25; 50 mm/s
Working modes Manual (3 leads choice)
Automatic:
Basic test and RHYTHM
Signal memory 5 s in AUTO mode and 3 minutes in RHYTHM mode
Pacemaker recognition AUTO and RHYTHM modes – automatic recognition
In MAN mode the impulses are visible in EKG record
after turning on the filters 25/35/50 Hz
Filfil demands of IEC 60601-2-51:2003
Signalisation of disconnected
Yes
electrodes
Co-operation with PC Use of one COM port
Available of archivization and A4 print out.
Work with the monitor Standard VGA socket,
Resolution: 640 x 480; Refresh frequency: 50Hz.
Observation of ECG signal on
Yes
PC monitor
Programme of ECG parameters
Yes
measurement (Analysis)
Histogram for RHYTHM test Yes
Digital muscle filter 35 Hz or 25Hz
Digital mains filter 50 Hz
Thermal head 104 mm, 8 points./mm
Thermoactive paper Width 110÷112 mm on roll of ∅ 54 mm
Serial interface RS-232 port
Keyboard functional
Quantity of channels 3 / 6 / 12
Screen: Graphical colour display, 640 x 480, 6.25 inch,
presenting 3, 6, 12 channels
Wright of the apparatus ~ 2 kg
Standard accessories:
Additional accessories: